1. Remove the clogged emitter and the column from the union on the
universal sprayer
2. Remove the PEEK nuts on either end
3. Loosen the screw on the sliding portion of the sprayer and remove
the union
5. Even if the union does not appear to be clogged it should be cleaned to
remove any small particulates
Cleaning the Union (Method #1)
As the union is 100% stainless steel, 20% nitric acid in water can be
used to clean it. This aggressive procedure helps ensure that the union
becomes free of even the strongest contaminations. In addition, it
provides excellent electric contact for electrospray ionization.
Wash the outside of the outlet capillary and fitting with water/acetron-
itrile (50/50) before reconnecting to union.
Note: Do not mix nitric acid with organic acids, as they will react and
produce high volumes of CO2, which could cause the explosions
of gas tight bottles.
Note: Again for demonstration purposes only, these photos show
fittings and tubing being handled with bare fingers, which is known
to introduce contamination in the system. It is therefore critical that
customers ALWAYS use powder-free, clean room and compatible
gloves when changing tubing or handling fittings.
4. Hold the union up to a light (above). Light should shine through an
unclogged union. If the union is completely clogged the obstruction
must be removed.
To order the Union: Waters PN 700002843
To order Peek Fitting: Waters PN 700002842.
Cleaning the Union (Method #2)
The union can be cleaned by sonication in a solution of IPA/water for at
least 15 minutes
After sonication, remove any residual solvent by flushing the union
with air or nitrogen
After cleaning, examine the union with a microscope and ensure that
any particulate matter has been removed. While looking at the union
be sure the internal threads and opening have not been scored or dam-
aged in any way
If the obstruction within the union cannot be removed or it appears
damaged please replace the union.
Note: In many cases it is best to simply REPLACE the union whenever
a new emitter is used.
nanoACQUITY UPLC Column Care and Use6
Page 7
[ Care and Use ManUal ]
Replace the Emitter
Place the pin plug at the back side of the union. T his will mark the
center of the union (right)
Thread the replacement emitter (back end first) through the front of the
sprayer and into the union
Tighten the PEEK nut around the union carefully (do not over-tighten)
Remove the pin plug and replace it with the transfer line from the
nanoACQUITY column (again, do not over-tighten the fitting).
About poorly cut fittings
DO NOT cut or modify the emitter or column tubing in any way.
To reduce the occurrence of clogging
Make sure the union between the column and emitter is clean
Do not over tighten the fittings on the union (finger tight only)
Make sure the UPLC solvents are clean
Keep a flow on the tip even when the instrument is not in use (as low as
200 nL/min)
Try using a TaperTip instead of a Pulled Tip Emitter if clogging persists
TaperTip Emitter
A TaperTip (P/N 186003932) is offered as an alternative to a
PicoTip and is the default emitter in the universal sprayer kit. (July,
2007). The TaperTip has an ID of 20 µm but unlike the PicoTip the ID
remains unchanged thought out the emitter. This allows the emitter
to be implemented for higher flow rates. Furthermore, the TaperTip
may experience a lower frequency of particle clogging.
All tubing and columns supplied by Waters is pre-cut and polished.
Improperly cut tubing can lead to "shedding" and introduction of
fused silica particles downstream of nanoACQUITY 75, 100, or
150um I.D. column. This consequently can cause undesired system
backpressure increases.
Note: Compared to the PicoTip Emitter, Waters TaperTip Emitter does
not produce as stable a spray at 250-400 nL/min, which is the flow
rate range most customers use with 75 µm or 100 µm ID columns.
As indicated above, use of the TaperTip Emitter is the suggested
when a more robust device is preferred.
Column IDOptimal Flow Rate (nL/min)Emitter Type
75 µm300PicoTip
100 µm400-500PicoTip / TaperTip
150 µm1000TaperTip
300 µm4000ESI Source required
nanoACQUITY UPLC Column Care and Use7
Page 8
[ Care and Use ManUal ]
What flow rate can I used with each type of emitter?
The flow rate of an LC method should be determined by the optimal
flow rate for the particular column. The optimal flow rate is directly
related to the column ID
A PicoTip is ideally suited for nanospray applications while the TaperTip
is suited for microspray applications (>1 µl/min)
Waters currently does not condone the use of a nanospray source for
the 300 µm column format.
VIII. LONG TERM, NANOACQUITY COLUMN STORAGE
When not used, it is recommended to maintain flow through the
nanoACQUITY column using 100% eluent B to maximize bed stability
and minimize potential bed drying
If prolonged column storage required or if the column needs to be
removed from system, it is recommended that the nanoACQUITY
column is flushed with 100% organic solvent containing no TFA
or FA (e.g. simply store in 100% acetonitrile).
XI. ORDERING INFORMATION
Description
nanoACQUITY UPLC Columns (10,000 psi)
Symmetry C
Symmetry C
Symmetry C
Symmetry C
Symmetry C
Symmetry C
Symmetry C
Symmetry C
Atlantis dC
Atlantis dC
Atlantis dC
Atlantis dC
Atlantis dC
Atlantis dC
Atlantis dC
Atlantis dC
nanoACQUITY UP LC BEH130 C18 75 µm100 mm 1.7 µm186003542
nanoACQUITY UP LC BEH130 C18 75 µm150 mm 1.7 µm186003543
nanoACQUITY UP LC BEH130 C18 75 µm200 mm 1.7 µm186003544
nanoACQUITY UP LC BEH130 C18 75 µm250 mm 1.7 µm186003545
nanoACQUITY UP LC BEH130 C18 100 µm 100 mm 1.7 µm186003546
nanoACQUITY UP LC BEH130 C18 150 µm 100 mm 1.7 µm186003550
nanoACQUITY UP LC BEH300 C
nanoACQUITY UP LC BEH300 C18 100 µm 100 mm 1.7 µm186003811
nanoACQUITY UP LC BEH300 C18 150 µm 100 mm 1.7 µm186003812
nanoACQUITY UP LC BEH300 C18 75 µm150 mm 1.7 µm186003813
nanoACQUITY UP LC BEH300 C18 75 µm200 mm 1.7 µm186003814
nanoACQUITY UP LC BEH300 C18 75 µm250 mm 1.7 µm186003815
Inner
Diameter Length
Particle
Size
75 µm100 mm 3.5 µm186003491
75 µm150 mm 3.5 µm186003492
100 µm100 mm 3.5 µm186003493
100 µm150 mm 3.5 µm186003494
150 µm100 mm 3.5 µm186003495
150 µm150 mm 3.5 µm186003496
300 µm100 mm 3.5 µm186003497
300 µm150 mm 3.5 µm186003498
75 µm100 mm 3 µm186003499
75 µm150 mm 3 µm186003500
100 µm100 mm 3 µm186003501
100 µm150 mm 3 µm186003502
150 µm100 mm 3 µm186003503
150 µm150 mm 3 µm186003504
300 µm100 mm 3 µm186003505
300 µm150 mm 3 µm186003506
75 µm100 mm 1.7 µm186003810
18
Part No.
nanoACQUITY UPLC Column Care and Use8
Protein Separation Technology nanoACQUITY UPLC
Columns (10,000 psi)
nanoACQUITY UP LC BEH300 C4 75 µm100 mm 1.7 µm186004639
nanoACQUITY UP LC BEH300 C4100 µm100 mm 1.7 µm186004640
nanoACQUITY UP LC BEH300 C4150 µm100 mm 1.7 µm186004641
For use with nanoACQUITY UPLC systems rated to 10,000 psi only.Not for use with nanoACQUIT Y UPLC systems rated to 5,000 psi.