Waters MassPREP Phosphopeptide Standards User Manual

[ Care and Use ManUal ]
massPreP PhosPhoPePtIde standards
I. IntroduCtIon
Phosphorylated proteins and peptides are difficult to detect and
characterize by mass spectrometry due to their low abundance and
low ionization efficiency. A set of standard samples can aid in devel-
oping methods for optimized phosphopeptide detection. As a tool
for screening sample preparation and sample enrichment methods
for phosphopeptides, Waters offers MassPREP™ Phosphopeptide
Standards containing synthetic phosphopeptides based on modified
peptides from tryptic yeast enolase digest (SwissProt P00924).
Commonly used phosphorylated samples are ß-casein digests or
standards which contain only two phosphopeptides, both modified
atserine. T he four modified enolase peptides in the mixture are either
singly phosphorylated at serine, threonine, or tyrosine or doubly
phosphorylated at serine.
II. synthetIC enolase PhosPhoPePtIdes
Peptide Sequence Isotopic mass,
[M+2H]2+
[M+H]+
Contents
I. IntroduCtIon
II. synthetIC enolase PhosPhoPePtIdes
III. reCommended usage
IV. samPle PreParatIon
V. examPle sPeCtra
VII. related aCCessorIes
T18 1P NVPLpY K 813.3912 407.1995
T19 1P HLADL pSK 863.4028 432.2053
T43 1P VNQIG pTLSES IK 1368.6776 684.8428
T43 2P VNQIG TLpSEpS IK 1448.6439 724.8259
Phosphopeptide Standards 1
[ Care and Use ManUal ]
III. reCommended usage
There are three sample kits available (#1-3 below). Each kit contains
1 nmol each of four synthetic enolase phosphopeptides - T18 1P
(phosphotyrosine), T19 1P (phosphoserine), T43 1P (phosphothreo-
nine and T43 2P, double phosphoserine).
Part Number (PN) MassPREP Product
Information
186003285 #1 - Phosphopeptide
Standard – Enolase (1 nmol each)
186003286 #2 - Enolase Digest with
Phosphopeptides Mix (1 nmol each, equimolar mix)
186003287 #3 - Phosphopeptide
Sample Kit – Enolase (contains PNs 186003285 and
186002325)
The sample kits are intended for use in optimizing phosphopeptid
detection in mass spectrometry (MS) or liquid chromatography (LC).
These standards can be used on many instrument systems, including
Waters MALDI micro MX™, Alliance, ZQ™, Q-Tof™, LCT Premier™, and
nanoACQUITY UPLC™ systems. Either UV- or MS- based detection is
possible.
Suggested Application
Test detection sensitivity and sample preparation methods with no ion suppression from non-phosphorylated peptides
Compare phosphopeptide detection before and after treatment or enrichment pro­cedures without conducting a lengthy digestion procedure
Conduct control experiments with phosphopeptides only and enolase digest only and make samples with different phosphopeptide to unmodi­fied peptide ratios to test enrichment procedures
IV. samPle PreParatIon
The following methods are provided as a general guideline. For opti-
mum performance, use of the highest purity commercially available
matrices (Waters MassPREP) and solvents (HPLC grade or better) is
recommended.
Procedure
1) Add 100 μL of water (or 0.1% TFA or 0.1% formic acid) to vial
and vortex to make a 10 pmol/μL solution.
2) Make the desired dilutions in preferred solvent.
3) Prior to LC or MS analysis, use enolase digest to adjust instru-
ment settings for optimum peak shape, resolution and sensitiv-
ity. Calibrate instrument.
4) For MALDI-MS, make a solution of 20 mg/mL 2,5-DHB matrix
in 4:1 (v/v) acetonitrile (ACN):water with 0.2 to 2% phosphoric
acid (PA).
a) For Waters MassPREP matrix containing 10 mg/vial, add
400 μL ACN and 100 μL 2% PA in water for overall
concentration of 0.4% PA.
b) Apply 1 μL (or less) of the sample (~ 1 pmol/μL) onto
a clean MALDI target. Add an equal volume of matrix solu-
tion to the sample solution on the target plate. Dry at room
temperature before submitting target for MALDI analysis.
5) For LC analysis, transfer desired amount of solution to appropri-
ate sample vial and subject to LC-UV or LC-ESI-MS analysis.
a) For separations on ~2 mm diameter columns, the recom-
mended injection volume is 15 μL or more of a solution
10 pmol/μL or more.
Phosphopeptide Standards 2
b) For separations on columns less than 200 μm in diameter,
the recommended injection volume is 1 μL of a solution
50 fmol/μL or more.
Loading...
+ 2 hidden pages