Waters MassPREP Micro Desalting Column User Manual

[ CARE AND USE MANUAL ]
MassPREP Micro Desalting Column
CONTENTS
I. INTRODUCTION
II. CONNEC TING THE MASSPREP MICRO DESALTING COLUMN WITH T HE CHOSEN LC SYSTEM
III. LC/MS SYSTEM CONFIGURATION TO MINIMIZE MS SOURCE CONTAMINATION
IV. USING MASSP REP MICRO DESALTING COLUMN TO MINIMIZE PROTEIN CARRYOVER BETWEEN INJECTIONS
V. GUIDELINES FOR USE
VI. RECOMMENDED STORAGE
I. INTRODUCTION
Mass spectrometry is a powerful tool for intact protein analysis. However, the quality of analyses can be compromised by salts in the sample that interfere with ionization and/or form adducts with the investigated proteins. It is, therefore, important to remove or significantly minimize the introduction of extraneous salt ions into the mass spectrometer. The use of on-line, desalting devices can efficiently perform this function with minimal sample handling.
Waters MassPREP™ Micro Desalting Columns (2.1 x 5 mm, P/N
186004032) contain a polymer-based, reversed-phase packing material (20 µm particle size, 1000 Å pore size) that can be used to effectively desalt protein samples prior to mass spectrometry. The phenyl phase material contained in these on-line devices successfully retains proteins allowing the salts to be washed to waste prior to elution of the desalted protein into the mass spectrometer. With an optimized LC/MS method, cycle times as low as 4 minutes are achievable. These on-line devices, when used with an appropriate LC system injector and injector wash solution, can be used for the sequential analyses of different samples.
II. CONNECTING THE MASSP REP MICRO DESALTING COLUMN WITH T HE CHOSEN LC SYSTEM
The MassPREP Micro Desalting Column can be used with an appropriately performing LC system to perform the desalting functions so long as proper fittings/tubings are used to make connections. However, optimized performances are achieved only when the cartridge is connected to a Waters ACQUITY or nanoACQUITY UPLC system. For the use of the cartridge with ACQUITY or nanoACQUITY, pre-assembled tubing kits (Intact Mass Tubing Kit, p/n 205000548) are provided to connect the column to the UPLC systems and mass spectrometer.
There are two configurations to connect the cartridge into the LC fluidic path. The inlet of the cartridge can be directly screwed onto the Valco® UPLC injection valve with no tubing connection.
[ CARE AND USE MANUAL ]
This configuration would leave the cartridge operating at room temperature. The other configuration is to screw the cartridge into one end of the included union connector. The connection of proper tubing to the other end of the union connector would allow the cartridge to be placed into a column heater to control the desalting temperature.
III. LC/MS SYST EM CONFIGURATION TO MINIMIZE MS SOURCE CONTAMINATION
Non-volatile salts (e.g., NaCl) can suppress protein ionization and foul the ionization source of the mass spectrometer. It is, therefore, important to avoid the introduction of these compounds into the mass spectrometer. Figure 1 details an LC/MS configuration that can effectively be used to load sample onto the column and elute salt to waste prior to valve switching and protein elution into the mass spectrometer.
Position 1: Sample Loading and Salt Elution to Waste
MassPREP
Gradient UPLC/HPLC System
Micro Desalting Column
Mass Spectrometer
Position 2: Sample Elution to Mass Spectrometer
MassPREP
Gradient UPLC/HPLC System
Figure 1. LC/MS Valve Switching Configuration for Use of MassPREP Micro Desalting.
Micro Desalting Column
Mass Spectrometer
Waste
Waste
IV. USING MASSPREP MICRO DESALTING COLUMN TO MINIMIZE PROTEIN CARRYOVER BETWEEN INJECTIONS
LC injector design, composition of injector wash solution, and number of injector wash cycles between injections are some of the factors that can effectively minimize protein carryover between sample analyses. Waters MassPREP Micro Desalting Column, when used with an appropriately performing UPLC/HPLC system and injector wash protocol, can be used for the sequential analyses of different samples due to the removal of protein carryover from prior injections.
V. GUIDELINES FOR USE
1. Initial Column Conditioning. Prior to use, the MassPREP Micro Desalting Column reversed-phase packing material must be “wet” with organic solvent prior to equilibration with 95% Eluent A / 5% Eluent B (e.g., Eluent A = 0.1% formic acid in MS grade H20 and Eluent B = 100% Acetonitrile). This can be accomplished by running a “blank injection” (i.e., Eluent A injection followed by running the separation gradient) or by washing the column with a minimum of five column volumes (~20 µL) of 50% Eluent B followed by a ten column volume equilibration at 95% Eluent A / 5% Eluent B prior to sample injection.
2. The maximum recommended operating pressure for MassPREP On-Line Desalting Cartridge, when configured WITHOUT pre-column tubing, is approximately 150 psi. Use of pressure generating tubing immediately before the MassPREP On-Line Desalting Cartridge is required to generate sufficient UPLC System back pressure for proper functioning. Thus, the total reported UPLC System backpressure generated when BOTH the pre-column tubing and MassPREP On-Line Desalting Cartridge are configured for an application is approximately 5700 psi at a flow of 0.5 mL/min with a 5% ACN/0.1% FA eluent.
3. Results obtained with Waters MassPREP Micro Desalting Columns can vary depending upon the physical characteristics of the analyzed proteins as well as on the performance characteristics of the HPLC (or UPLC/MS System), injector, and injector wash protocol used.
MassPREP Micro Desalting Column
2
Loading...
+ 1 hidden pages