Waters IC-Pak Column and Guard User Manual

[ Care and Use ManUal ]
IC-pak Column and guard Column
I. IntroduCtIon
Waters IC-Pak™ anion (A) and cation (C) columns are designed for use
in Ion Chromatography (IC) applications to separate ions.
a. Anion Columns
The stainless steel anion columns separate and quantify many
anions, such as F-, CI-, NO
-
, Br-, NO
2
ppb levels. Table 1 lists the characteristics of IC-Pak anion columns.
Table 1: Anion Column Characteristics
Anion Column Dimensions Particle Size Capacity
IC-Pak A 4.6 x 50 mm 10 µm 30 ± 3 µeq/mL
IC-Pak A HC
(High Capacity)
IC-Pak A HR
(High Resolution)
4.6 x 150 mm 10 µm 30 ± 3 µeq/mL
4.6 x 75 mm 6 µm 30 ± 3 µeq/mL
The packing material is a polymethacrylate resin with a quaternary
ammonium functional group. The columns are shipped in 1.3 mM
gluconic acid/1.3 mM boric acid (pH 8.5).
-
, HPO
3
2-
, and SO
4
2-
anions at
4
Contents
I. IntroduCtIon
II. InstallatIon
III. eluent and sample guIdelInes
IV. operatIon
V. Care and maIntenanCe
VII. Warranty/serVICe
b. Cation Columns
The non-metallic cation columns separate and quantify Li+, Na+,
+
NH
, K+, Mg2+, Ca2+, Sr2+ and Ba2+ cations at ppb levels. T he Waters
4
cation guard column, placed in the solvent stream before the
sample injector, is used only to aid monovalent cation detection by
removing polyvalent cations from the solvent stream.
Table 2 lists the characteristics of IC-Pak cation and cation guard
columns.
IC-Pak Column and Guard Column 1
[ Care and Use ManUal ]
COMPRESSION SCREW OR NUT
T
UBE
FERRULE
END MUST BE STRAIGHT AND SMOOTH TO ACHIEVE MAXIMUM COLUMN EFFICIENCY
CRITICAL DISTANCE TO BE DETERMINED BY THE UNION OR COLUMN FITTING, ETC.
Table 2: Cation Column Characteristics
Cation Column Dimensions Particle Size Capacity
IC-Pak C 4.6 x 50 mm 10 µm 12.0 ± 0.2 µeg/mL
Guard 4.6 x 50 mm n/a 2.0 ± 0.2 meg/mL
The packing material is a styrene divinylbenzene resin with a
sulfonic acid functional group. The columns are shipped in 2 mM
nitric acid.
II. InstallatIon
Before attaching the column in the flow path:
1. Directly connect the HPLC injector to the detector by replacing the old
column with a zero-dead-volume union.
2. Flush the lines free of microparticulates and previous solvents. Flush
the injector loop if applicable.
3. Remove the union.
4. Install the column.
To prepare a new tubing/ferrule connection:
1. Scribe the circumference of the tubing at the desired break using either
a file with a cutting edge or a tube cutter.
2. Grasp the tubing on both sides of the scribe mark with cloth-covered
pliers (to prevent marring the tube surface), and gently work the tube
back and forth until it separates.
Note: Ensure that the tubing end is straight, open, and free of burrs.
3. Slide the compression fitting, followed by the ferrule (large end of the
taper first) over the tube.
Note: Properly bottom the tubing in the fitting seat. Otherwise, dead
volume could result in sample band spreading.
Figure 1: Ferrule and Compression Assembly (Stainless Steel Column)
a. Installing a Stainless Steel Column
Remove the end plugs from the column and save them for use when
the column is removed from the system and stored. The column
outlet is indicated by an arrow on the label (showing the direction
solvent should flow).
To install the column, thread the inlet and outlet fittings into the
column until finger tight, and then tighten the fittings 1/4-to-1/2
turn. Do not over-tighten. Over-tightening will damage the
connection. A properly prepared and assembled compression fitting
in good condition is all that is required.
Prepare a new tubing/ferrule connection (Figure 1) when a new
column is connected or when a damaged compression screw or worn
ferrule are removed.
Note: The tubing distance beyond the ferrule may differ for different
column types. Re-size the tubing to the correct distance by replacing
the ferrule.
IC-Pak Column and Guard Column 2
b. Installing Non-Metallic Columns
Remove the end plugs from your column and save them for use when
the column is removed from the system and stored. The column
outlet is indicated by an arrow on the label (showing the direction
solvent should flow).
To install the column, thread the inlet and outlet fittings into the
column until finger tight, and then tighten the fittings 1/4-to-1/2
turn. Use caution with non-metallic columns. Over-tightening these
fittings may cause damage which results in a leaking connection.
A properly prepared and assembled compression fitting in good
condition is all that is required.
[ Care and Use ManUal ]
COMPRESSION SCREW
.021-inch (i.d.)
TEFZEL TUBING
TEFLON® FACE
BEVEL
FERRULE
STAINLESS STEEL
FAC E
FROM SOLVENT
RESERVOIR
PUMP
Inj
OPTIONAL GUARD-PAK AND HOLDER
IC-PAK
A
COLUMN
DETECTOR
TO WASTE
HIGH SENSITIVITY
NOISE FILTER
PUMP
Inj
HIGH SENSITIVITY
NOISE FILTER
ELUENT
EDTA/ HNO
3
GUARD COLUMN Inj
IC-PAK C
430/431
PUMP
Inj
HIGH SENSITIVITY
NOISE FILTER
ELUENT
EDA/ HNO
3
Inj
IC-PAK C
430/431
To replace a fitting:
1. Use a sharp razor blade to make a long tapering cut in the tubing as
shown in Figure 2. Alternately, if tubing is not in short supply, stretch
the tubing (after warming over low heat) between two sets of pliers and
cut the tube at thinnest point. The resulting taper will slip through the
fittings more easily.
2. Pass the tubing through the compression screw and ferrule. Be sure the
stainless steel side of the ferrule is facing the screw. Pull the tubing
with pliers to ensure the bevel or taper is completely past the ferrule.
Figure 5: IC Systems for Divalent Cation Analysis
III. eluent and sample guIdelInes
a. Preparing the Eluent Requirements
3. Hold the ferrule with a second set of pliers and give the tubing a half
twist to lock the ferrule. Cut the tubing flush with the end of the ferrule.
Figure 2: Ferrule and Compression Assembly (Non-Metallic Column)
c. Configuring the Column In the System
Install stainless steel or non-metallic anion columns relative to the
other major components as shown in Figure 3.
Water-miscible organic solvents, which may be used as modifiers to
reduce hydrophobic interaction between sample and packing, must
not exceed the concentrations listed below.
IC-Pak A Columns 12% acetonitrile in water
IC-Pak C and Guard Columns 10% acetonitrile in water
Note: Organic solvent concentration should not exceed 20%. It is
possible, however, to use organic solvents up to 50% concentration
by gradual introduction via gradient method. Perform solvent
changeovers at 0.1 mL/min over 6 hours. Dedicate column to use in
the organic concentrations over 20%.
Use high performance liquid chromatography (HPLC) grade solvents
that have been filtered to remove microparticulate matter greater
than 0.45 µm in size. Filtering will ensure optimum long-term
performance of the column. Vacuum filtration, sonication, or sparging
may be used to remove dissolved gas which could affect results.
Before filtering or clarifying any solvents, flush the filters with
100 mL of the eluent. Discard the filtered eluent. Surfactants used to
treat some filters may otherwise affect analysis.
Figure 3: IC System for Anion Analysis
Install cation columns with a guard column for detecting monovalent
cations (Figure 4) or without a guard column for polyvalent cations
(Figure 5).
Figure 4: IC Systems for Monovalent Cation Analysis
IC-Pak Column and Guard Column 3
An in-line filter may be used to remove particulates in the solvent
and contaminants which the 0.45 µm filter does not remove.
Note: Do not use tetrahydrofuran (THF) or methanol with IC-Pak A columns.
Loading...
+ 7 hidden pages