Waters High Strength Silica Columns User Manual

[ Care and Use ManUal ]
HIGH stRenGtH sILICa HPLC CoLUMns
Contents
I. IntRoDUCtIon
II. ConeCtInG tHe CoLUMn to tHe HPLC sYsteM
III. WateRs sMaLL PaRtICLe sIze (3.5 �M) CoLUMns – Fast CHRoMatoGRaPHY
V. CoLUMn InstaLLatIon PRoCeDURe
VI. CoLUMn PeRFoRManCe VaLIDatIon
VII. InItIaL CoLUMn eFICIenCY DeteRMInatIon
VIII. CoLUMn UsaGe
IX. CoLUMn CLeanInG, ReGeneRatInG anD stoRaGe
X. tRoUbLesHootInG
I. IntRoDUCtIon
Thank you for choosing a High Strength Silica (HSS) HPLC column.
The manufacture of HSS HPLC columns begins with ultrapure
reagents to control the chemical composition and purity of the
final product. HSS HPLC columns are manufactured in a cGMP,
ISO 9001:2000 certified plant with each step being conducted
within narrow tolerances. Every column is individually tested and
Certificates of Batch Analysis and a Performance Chromatogram are
provided with each column.
HSS HPLC columns are based on the same particle technology and
chemistries as ACQUITY UPLC HSS columns, thus enabling seamless
transferability between HPLC and UPLC separations.
Waters recommends the use of Sentry
column lifetime and protect the column from contaminants.
guard columns to maximize
HSS HPLC C olumns
1
[ Care and Use ManUal ]
II. CONNECTING THE COLUMN TO THE HPLC SYSTEM
a. Column Connection
Handle the column with care. Do not drop or hit the column on a hard
surface as this may disturb the bed and affect its performance.
1. Correct connection of 1/16 inch outer diameter stainless steel
tubing leading to and from the column is essential for high-
quality chromatographic results.
2. An arrow on the column identification label indicates correct
direction of solvent flow.
3. When using standard stainless steel compression screw fittings,
it is important to ensure proper fit of the 1/16 inch outer
diameter stainless steel tubing. When tightening or loosening the
compression screw, place a 5/16 inch wrench on the compression
screw and a 3/8 inch wrench on the hex head of the column
endfitting.
Caution: If one of the wrenches is placed on the column flat
during this process, the endfitting will be loosened and leak.
Under-tightening compression screws or using worn ferrules can
lead to solvent leaking. Care should be taken to check all column
connections for leaks to avoid exposure to solvents and the
hazards associated with such exposure including risks to health
and electrical
connections.
4. If a leak occurs between the stainless steel compression screw fit-
ting and the column endfitting, a new compression screw fitting,
tubing and ferrule must be assembled.
ferrule. If a non-Waters style column is presently being used, it is
critical that ferrule depth be reset for optimal performance prior to
installing a HSS HPLC column. In a proper tubing/column connection
(Figure 2), the tubing touches the bottom of the column endfitting,
with no void between them.
Attention: A void will occur if tubing with a Parker ferrule is con-
nected to a Waters style endfitting (Figure 3). This will dramatically
reduce the efficiency of the column and cause peak shape distortion.
Figure 1: Waters and Parker Ferrule Types
Waters Ferrule Setting Parker Ferrule Setting
Figure 2: Proper Tubing/Column Connection
Tubing touches the bottom of the column endfitting, with no void
between them.
Attention: A void will occur if tubing with a Parker ferrule is con-
nected to a Waters style endfitting (Figure 3). This will dramatically
reduce the efficiency of the column and cause peak shape distortion.
Figure 3: Parker Ferrule in a Waters Style Endfitting
It is important to realize that extra column peak broadening can
destroy a successful separation. The choice of appropriate column con-
nectors and system tubing is discussed in detail below.
b. Column Connectors and System Tubing Considerations
Due to the absence of an industry standard, various column man
facturers have employed different types of chromatographic column
connectors. The chromatographic performance of the separation can
be negatively affected if the style of the column endfittings does not
match the existing tubing ferrule setting. This section explains the
differences between Waters style and Parker style ferrules and endfit-
tings (Figure 1). Each endfitting style varies in the required length
of the tubing protruding from the ferrule. The HSS HPLC column is
equipped with Waters style endfittings, which require a 0.130 inch
HSS HPLC C olumns
To fix this problem: Cut the end of the tubing with the ferrule, place a
new ferrule on the tubing and make a new connection. Before tighten-
ing the screw, make sure that the tubing bottoms out in the endfitting
of the column.
Conversely, if tubing with a Waters ferrule is connected to a column
with Parker style endfitting, the end of the tubing will bottom out
before the ferrule reaches its proper sealing position. This will leave a
gap and create a leak (Figure 4).
Caution: The connection will leak if a Waters ferrule is connected to a
column with a Parker style endfitting.
2
[ Care and Use ManUal ]
Diluted/Distorted Sample Band
0.005 inches
0.020 inches
0.040 inches
Figure 4: Waters Ferrule in a Parker Style Endfitting
There are two ways to fix the problem:
1. Tighten the screw a bit more. The ferrule moves forward, and
reaches the sealing surface. Do not overtighten since this may
break the screw.
2. Cut the tubing, replace the ferrule and make a new connection.
Alternatively, replace the conventional compression screw fitting with
an all-in-one PEEK
fitting (Waters Part Number PSL613315) that
allows resetting of the ferrule depth. Another approach is to use a
Thermo Hypersil
Keystone, Inc. SLIPFREE® connector to ensure the
correct fit. The fingertight SLIPFREE connectors automatically adjust to fit
all compression screw type fittings without the use of tools (Figure 5).
Figure 5: Single and Double SLIPFREE Connectors
Figure 6: Effect of Connecting Tubing on System
d. Measuring System Bandspread Volume
This test should be performed on an HPLC system with a single wave-
length UV detector (not a Photodiode Array (PDA)).
1. Disconnect column from system and replace with a zero dead
volume union.
2. Set flow rate to 1 mL/min.
3. Dilute a test mix in mobile phase to give a detector sensitivity
0.5-1.0 AUFS (system start up test mix can be used which
contains uracil, ethyl and propyl parabens; Waters Part Number
WAT034544).
Table 1. Waters Part Numbers for SLIPFREE Connectors
SLIPFREE Type and Tubing Length
0.005” 0.010” 0.020”
Single 6 cm PSL 618000 PSL 618006 PSL 618012
Single 10 cm PSL 618002 PSL 618008 PSL 618014
Single 20 cm PSL 618004 P SL 618010 PSL 618016
Double 6 cm PSL 618001 PSL 618007 PSL 618013
Double 10 cm P SL 618003 PSL 618009 PSL 618015
Double 20 cm P SL 618005 PSL 618001 PSL 618017
c. Band Spreading Minimization
Internal tubing diameter influences system band spreading and peak
shape. Larger tubing diameters cause excessive peak broadening and
lower sensitivity (Figure 6).
Tubing Internal Diameter
4. Inject 2 to 5 μL of this solution.
5. Using 5 sigma method measure the peak width at 4.4% of peak
height:
Band Spreading (μL) = Peak Width (min) x Flow Rate (μL/min)
= 0.1 min x 1000 μL/min
= 100 μL
Figure 7: Determination of System Bandspread Volume Using 5-Sigma Method
In a typical HPLC system, the Bandspread Volume should be 100 μL ± 30 μL .
HSS HPLC C olumns
3
Loading...
+ 5 hidden pages