This document provides information regarding the general care and
use of the Waters GlycoWorks™ Single Use Sample Preparation Kit
for release of N-glycans. T his protocol can be used as a guideline and
is validated using IgG but can also be used and optimized for other
N-linked glycoproteins. This sample preparation offers an inclusive
protocol that allows the user to source many of the critical components
needed to perform the preparation from one vendor.
The Single Use Catridge Pack contains 20 1cc catridges, good for
10 analyses.
For comparison results using this protocol versus the 96 well plate
and for recovery information with the control standard
1,2
.
General Guideline of Sample Preparation from Glycoprotein to
enrich FLR Labeled Glycans Using Reductive Amination Reaction.
Glycoproteins
Denaturation and Deglycosylation
1
2
3
4
If using MALDI please refer to Product Solution, Literature Code 720004660EN
RapiGest™ (if necessary)
DTT Reduction (if necessary),
IAM Alkylation (if necessary)
Deglycosylation by PNGase F
Extraction of Released Glycans
GlycoWorks HILIC 1 cc Cartridge
FLR Labeling Reaction of Glycans
Excess Labeling Reagent Removal
GlycoWorks HILIC 1 cc Cartridge
FLR Labeled Glycans
MALDI MS
LC-FLR
LC-FLR-MS
The guide provides the following:
■
An instruction template for new users that highlights the use of a
control standard (included in the kit) for first time validation as
well as subsequent troubleshooting
■
Protocol for LC/FLR and LC/FLR/MS
■
Tips and tricks throughout each section to help aid in the success
■
An online training video is also available
GlycoWorks High-throughput Sample Preparation Kit
3
Figure 1: Sample Preparation from Glycoprotein to Labeled Glycans
1
[ CARE AND USE MANUAL ]
Table 1: Kit Components
Kit ComponentPurpose
GlycoWorks
HILIC 1 cc
Cartridge
GlycoWorks
Glycoprotein
Control
Standard
RapiGest SFAn enzyme friendly detergent used to denature
Manifold
Waste Tray
GlycoWorks
Reagent Kit
Extract and clean up of free or FLR labeled glycans
■
Capture of glycans
■
Removing contaminants like salt and
detergents from hydrophilic analytes, such
as carbohydrates, prior to MS
■
Removing excess labeling reagent
■
The 10 mg sorbent was test to have maximum
binding capacity for 400 µg glycans
Unlabeled IgG (100 µg/lyophilized) to
use as a control for process validation and
troubleshooting
the glycoproteins prior to deglycosylation
reaction. For stability purposes, there is enough
RapiGest included for initial use of the kit with
5 samples, more may need to be ordered for
processing all 10 samples.
For collection of non-releasing glycans steps.
This is disposable but can be cleaned and used
sever al times befo re acquiring a new one as long
as it is rinsed out.
Contains ready to use reagents for both
denaturation, reduction, and alkylation as well as
raw materials for FLR tagging.
II. STORAGE AND DISPOSAL OF CARTRIDGES
Cartridges stored in their original sealed pouch remain stable for
long periods. For long term storage of unused cartridges, keep
cartridges sealed and in a desiccator.
Note: Dispose of used cartridges safely in accordance with applicable
government or local regulations
III. ABBREVIATIONS
DMSO DimethylsulfoxideIAMIodoacetamide
DTTDithiothreitol2AB2-aminobenzamide
IV. USING THE GLYCOWORKS SINGLE USE KIT
Below, are general guidelines for utilizing the GlycoWorks Single
Use Kit for complete extraction, purification, and concentration of
N-linked glycans validated using IgG. Also included is a check list of
accessory materials that will be needed to complete the protocol from
start to finish as well as tips and tricks throughout the instructions.
Table 2: Checklist of Additional Materials Needed Before You Begin
Material Description
Ammonium
bicarbonate
Ammonium acetate
Formic acid (to make
1% [v/v] formic acid,
prepared in 50%
acetonitrile)
15,000 units
(500,000 units/mL)
PNGase F*
Acetonitrile
25 mM ammonium
bicarbonate in water,
pH 8 -9
100 mM ammonium
acetate in 5% [v/v]
acetonitrile, pH 7
85% [v/v] acetonitrileSee AboveYes No
2-AB**Sigma P/N A89804Yes No
SPE vacuum pump
Centrifugal vacuum
evaporator
Heat blockN/AYes No
Fisherbrand Premium
1.5 mL MCT graduated,
RNase/DNase free
* PNGase F comparison table: https://www.neb.com/tools-and-resources/
Protect the final preparation of labeling reagent from light.
■
The sodium cyanoborohydride once it is solubilized, should be
used within an hour.
■
Left over reagent which contains sodium cyanoborohydride has
to be disposed separately in a clearly marked bottle due to its
toxicity (see MSDS).
■
To ensure high yield of labeled glycans, it may be necessary,
for some samples, to increase the concentrations of the
reducing and labeling reagents.
[ CARE AND USE MANUAL ]
Step 4. Removal of Excess Label
1) Condition a NEW well of the GlycoWorks cartridge by adding
1,000 μL of Milli-Q water to each cartridge and aspirate using the
vacuum manifold.
2) Add 1,000 μL of 85% acetonitrile to each well and aspirate using
the vacuum manifold or positive pressure manifold.
3) Add 200 μL of acetonitrile to the 20 μL of 2AB labeled glycan
sample and load into the cartridge using low vacuum setting
(1-2 in Hg).
4) Wash each well with 400 μL of 85% acetonitrile. Remove
filtrate to waste.
5) Replace the waste tray with sterile 1.5 mL Eppendorf tubes.
6) Elute glycans with 100 μL of 100 mM ammonium acetate in 5%
acetonitrile.
Note: The concentration of ammonium acetate and acetonitrile in the elu-
tion buffer may require optimization for some glycan samples.
7) Repeat elution two more times.
8) Place 96-well collection plate in a centrifugal vacuum evaporator
and dry the glycans to a concentrate of ≤ 2 µL in volume.
Note: If a rotor is not available for plates, transfer each eluent to a sterile
1.5 (or 0.5) mL Eppendorf and place in the vacuum evaporator. This may
contribute to a small sample loss due to extra transfer.
9) Glycans can be stored in Milli-Q water at -20 °C until required.
V. PREPARING LABELED GLYCANS FOR HILIC-FLR WITH
AN ACQUITY UPLC BEH GLYCAN COLUMN
The following steps are suggested for preparing the labeled glycans for
HILIC-FLR with an ACQUITY UPLC BEH Glycan column when using this
kit with the Waters Total Glycan Application Solution.
1. Reconstitute the labeled glycans in 100 µL of Milli-Q water by
gently aspirating, avoiding vortexing and sonication.
2. Add 150 µL of ACN to the reconstituted glycans immediately
prior to and in preparation for analysis by HILIC-FLR.
3. Inject 4 µL of the labeled glycan containing 60% acetonitrile
solution onto the column.
Note: Preparations of non-IgG glycoproteins or glycoprotein samples with
concentrations other than 1 mg/mL may need to be reconstituted with
different volumes to achieve desired results. Similarly, more concentrated
samples can be obtained by reconstituting with up to 10 times lower volumes.
The optimum injection volume for this application is ≤4 µL.
VI. REFERENCES
1. Lauber MA, Koza S, Fountain K J. Optimization of HILIC SP E for the Quantitative and Robust Recovery of N-Linked Glycans. Waters Application Note
720004710EN. 2013 June.
2. Lauber MA, Koza S, Fountain KJ.Single-Use and High-Throughput HILIC SPE
Device Formats and an IgG Control Standard for Facilitating N-Glycan Analyses Waters Technology Brief 720004711EN. 2013 June.
3. Training Video: http://www.waters.com/waters/en_US/Glycan-Standards/nav.
htm?cid=134640534
Step 4 Tips and Tricks
■
Extended periods of time between incubation and analysis
may result in desialylation of labeled glycans and consequently
The Glycan Performance Test Standard is a 2-AB labeled
human IgG-like standard that is QC verified to contain the
components needed to benchmark and evaluate ACQUITY
UPLC BEH Glycan, 1.7 µm Columns.
Dextran C alibration Ladder
The 2-AB labeled, Dextran Calibration Ladder is used
to calibrate the HILIC column from retention time to GU
values. This calibration ladder provides good peak shape
and reliable identification from 2 to 30 Glucose Units.
GlycoWorks HILIC 1cc Flangeless Catridges Pack
This pack of 20 catridges, good for 10 analyses, are
offered for those customers who prefer to use the single
use devices with the positive pressure manifold (PPM)
176 00 3119
186007080
186001860
186007033
186007034
176003090
186002780
186001860
186007033
186007034
600001282
186006349
186006841
186007239
Glycan Analysis System Standards and Mobile Phases
Description UsageVolumePart No.
Alliance with Fluorescence
Qualification Standards Test Kit
Kit contains seven 10 mL vials of varying concentrations of
anthracene in 80:20 acetonitrile/water:
Fluorescence Detector
Performance P
Standard Solution
0.10 mg/L anthracene in
70:30 acetonitrile/water
Ammonium Formate Concentrate
5000 mM Ammonium Formate in a
Water w/ 3.8% Formic acid matrix.
This allows the end user to dilute the
concentrate 10 mL to 1L before using to
reach a 50 mM concentration.