Waters Dextran Calibration Ladder Standard User Manual

[ CARE AND USE MANUAL ]
Dextran Calibration Ladder Standard
CONTENTS
I. INTRODUCTION
II. STORAGE & STABILITY
III. RECONSTITUTION OF DEXTRAN CALIBRATION LADDER
IV. BENEFITS OF USING A CALIBRATION STANDARD
V. EXAMPLES OF USING THE DEXTRAN CALIBRATION
LADDER ON THE ACQUITY UPLC
VI. ORDERING INFORMATION
I. INTRODUCTION
Hydrophilic Interaction Liquid Chromatography (HILIC) coupled with Fluorescent Detection is widely used for fluorescent (FLR) labeled glycans. Waters ACQUIT Y® BEH Glycan column operated in HILIC mode shows drastic improvement in separation e.g., peak resolution and the ability to separate both neutral and acidic glycans, compared to the conventional HPLC methods in HILIC mode (Ref. 1). The ACQUITY BEH Glycan column enables scientists to get robust and reproducible glycan separation data with less time spent in method optimization.
In order to take full advantage of the BEH glycan column, Waters created the 2-aminobenzamide labeled Dextran Calibration Ladder Standard to assist glycan profiling: glycan profile obtained from UPLC-HILIC/FLR system can be calibrated against the 2AB-labeled Dextran ladder and assigned with glucose unit (GU) values (Ref.2). Waters 2AB-labeled Dextran Calibration Ladder is different than the existing commercial offerings. The average molecular weight of the glucose homopolymer is higher (~4,500 Dalton), therefore, the “workable” GU value range is twice as much as the other Dextran ladder standards; the observed GU goes from 2 to 30. This feature improves the large glycan retention time assignment. The purity and structural integrity of the ladder is assessed by normal phase HPLC and MS.
Dextran Calibration Ladder Standard
II. STORAGE AND STABILITY
The lyophilized powder is shipped at ambient temperatures, but it is highly recommended upon receipt of the standard to store it refrigerated (4°C) until the expiration date printed on the label. Once reconstituted in should be used immediately for best results.
III. RECONSTITUTION OF THE DEXTRAN CALIBRATION LADDER
One vial of the Dextran Calibration Standard contains 200 µg of a lyophilized, solid powder. It is contained in a Waters Max Recovery Vial for supreme ease of use so that it can be directly solubilized and put on the system for analysis and injection.
1
[ CARE AND USE MANUAL ]
EU
For reconstitution, the sample can be diluted with 100 µl of Milli-Q water plus 100 µl of Acetonitrile for a 200 µl total dilution (1µg/µl).
IV. BENEFITS OF USING A CALIBRATION STANDARD
Each individual glycan structure has a GU value which is directly related to the number of linkage of its constituent monosaccharides. The GU value can be used to predict structures because each monosaccharide in a specific linkage adds a given amount to the GU value of a given glycan.
The elution times of glycans are expressed in glucose units (GU) by reference to a dextran ladder. This ladder is used to calibrate the LC runs against day-to-day or system-to-system changes. The GU value is calculated by fitting a fifth order polynomial distribution curve to the dextran ladder, then using this curve to allocate GU values from retention times. The GU values for neutral N-glycans are very reproducible with standard deviations of <0.3 between columns, this allows direct comparison with database values collected from a range of instruments over a period of time (Ref. 3).
V. EXAMPLES OF USING THE DEXTRAN CALIBRATION LADDER
ON THE ACQUITY UPLC
IV. ORDERING INFORMATION
Description Part Number
Dextran Calibration Ladder
186006841
ACQUITY UPLC BEH Glycan 1.7µm 2.1 x 50mm 186004740
ACQUITY UPLC BEH Glycan
1.7µm 2.1 x 100mm
ACQUITY UPLC BEH Glycan
1.7µm 2.1 x 150mm
186004741
186004742
ACQUITY UPLC BEH Glycan, Vanguard 3-pack 186004739
References
1. Joomi Ahn, Jonathan Bones, Ying Qing Yu, Pauline M. Rudd, Martin Gilar. “Separation of 2-aminobenzamide labeled glycans using hydrophilic interaction chromatography columns packed with 1.7 µm sorbent”, Journal of Chromatography B, 878 (2010) 403–408.
2. Matthew P. Campbell, Louise Royle, C atherine M. Radcliffe, Raymond A. Dwek and Pauline M. Rudd. “GlycoBase and autoGU:tools for H PLC-based glycan analysis”, BIOINFORMATICS A PPLICATIONS NOTE, Vol. 24, no. 9 2008, 1214-1216.
3. Geoffrey R. Guile, Pauline M. Rudd, David R. Wing, Sally B. Prime, Raymond A. Dwek. “A Rapid High-Resolution High-Performance Liquid Chromatographic Method for Separating Glyc an Mixtures and Analyzing Oligosaccharide Profiles” Analytic al Bioc hemistry, 240 (1996) 210-226.
Below is a reference chromatogram with conditions to help provide an example of the Dextran Calibration Ladder. The elution time of glycans is expressed in glucose units (GU) by reference to a dextran ladder, refer to Figure 1. If the column size is different, the gradient/ injection volume can be scaled accordingly.
Figure 1: Example Chromatogram of the Dextran Calibration Ladder with glucose units (GU) labeled peaks scaled from 0.8 – 46.5 minutes.
System Conditions: Injection Volume: 1.5 µl Injection Mode: Partial loop needle overfill Column: ACQUIT Y UPLC BEH Glyc an, 2.1 x 150mm Eluent A: 100mM Ammonium Formate Buffer pH 4.5 Eluent B: Acetonitrile Weak Needle Wash: Acetronitrile/HPLC Grade Water, (90:10 v/v) Strong Needle Wash: Acetronitrile/H PLC Grade Water, (10:90 v/v) Seal Wash: Acetronitrile/Water (50:50 v/v) Temperature: 60°C Detection: Flourescence: λex= 330 nm, λem= 420 nm Gradient: Time Flow Rate (min) (mL/min) %A %B Curve initial 0.50 25 75 6
46.5 0.50 50 50 6 48 0.25 100 0 6 49 0.25 100 0 6 50 0.50 25 75 6 63 0.50 25 75 6
34.00
GU = Glucose Unit
32.00
30.00
28.00
26.00
24.00
22.00
20.00
GU-2
18.00
16.00
14.00
12.00
10.00
8.00
6.00
4.00
2.00
0.00
0 10 15 20
GU-5
5 25 30 35 40 45 min
GU-10
GU-15
GU-20
Dextran Calibration Ladder Standard 2
Loading...
+ 1 hidden pages