Waters CORTECS 1.6 um User Manual

CORTECS Columns
CONTENTS
I. GETTING STARTED
a. Column Connectors
b. Column Installation
c. Column Equilibration
d. eCord Installation
e. Initial Column Efficiency Determination
f. VanGuard Pre-Columns
a. Sample Preparation
b. pH Range
c. Solvents
d. Pressure
e. Temperature
III. COLUMN CLEANING, REGENERATING
AND STORAGE
a. Cleaning and Regeneration
b. Storage
Thank you for choosing a Waters CORTECS™ Column. The CORTECS packing materials were designed specifically for use with the Waters ACQUITY UPLC® System and are manufactured in a cGMP, ISO 9001 certified plant using ultra pure reagents. Each batch of CORTECS material is tested chromatographically with acidic, basic and neutral analytes and the results are held to narrow specification ranges to assure excellent, reproducible performance. Every column is individually tested and a Performance Chromatogram and Certificate of Batch Analysis are provided on
the eCord
CORTECS Columns were designed and tested specifically for use on the ACQUITY UPLC System. CORTECS Columns will exhibit maximum chromatographic performance and benefits when used on the holistically-designed ACQUITY UPLC System since they were created and designed to operate with it.
Intelligent Chip.
IV. INTRODUCING eCORD INTELLIGENT
CHIP TECHNOLOGY
a. Introduction
b. Installation
c. Manufacturing Information
d. Column Use Information
V. ADDITIONAL INFORMATION
a. Tips for Maximizing CORTECS Column Lifetimes
b. Getting Started with CORTECS HILIC Columns
I. GETTING STARTED
Each CORTECS Column comes with a Certificate of Analysis and a Performance Test Chromatogram embedded within the eCord intelligent chip. The Certificate of Analysis is specific to each batch of packing material contained in the CORTECS Column and includes the gel batch number, analysis of unbonded particles, analysis of bonded particles, and chromatographic results and conditions. T he Performance Test Chromatogram is specific to each individual column and contains such information as: gel batch number, column serial number, USP plate count, USP tailing factor, capacity factor, and chromatographic conditions. These data should be stored for future reference.
Note: If mobile-phase additives are present in low concentrations (e.g., ion-pairing reagents), 100 to 200 column volumes may be required for complete equilibration. In addition, mobile phases that contain formate (e.g., ammonium formate, formic acid, etc.) may also require longer initial column equilibration times.
c. Column Equilibration
CORTECS Columns are shipped in 100% acetonitrile. It is important to ensure mobile-phase compatibility before changing to a different mobile-phase system. Equilibrate the column with a minimum of 10 column volumes of the mobile phase to be used (refer to Table 1 for a list of column volumes). The column may be considered thermally equilibrated once a constant backpressure is achieved.
a. Column Connectors
The ACQUITY UPLC System utilizes tubing and gold plated compression screws which have been designed to meet stringent tolerance levels and to minimize extra column volumes.
Optimized column inlet tubing (Part Number 430001084) is supplied with the ACQUITY UPLC System. The inject valve end of the tubing is clearly marked with a blue shrink tube marker. Insert the opposite end of the tubing into the CORTECS Column and tighten the compression fitting using two 5/16-inch wrenches.
For information on the correct column outlet tubing, please refer to the relevant detector section in the ACQUITY UPLC System Operator’s Guide (Part Number 71500082502).
b. Column Installation
Note: The flow rates given in the procedure below are for a typical
1.6 µm, 2.1 mm ID by 50 mm length column. Scale the flow rate up or down accordingly based upon the column pressure and system limits.
1. Purge the pumping system of any buffer-containing mobile phases and connect the inlet end of the column to the injector outlet.
2. Flush column with 100% organic mobile phase (methanol or acetonitrile) by setting the pump flow rate to 0.1 mL/min and increase the flow rate to 0.5 mL/min over 5 minutes.
3. When the mobile phase is flowing freely from the column outlet, stop the flow and attach the column outlet to the detector. This prevents entry of air into the detection system and gives more rapid baseline equilibration.
Table 1. Column Volumes (mL)
Column Length (mm)
30 0.1 0.2
50 0.2 0.4
75 0.3 0.5
100 0.4 0.8
150 0.5 1.0
Internal Diameter
2.1 mm 3.0 mm
To avoid precipitating mobile-phase buffers on your column or in your system, flush the column with five column volumes of a water/organic solvent mixture, using the same or lower solvent content as in the desired buffered mobile phase. (For example, flush the column and system with 60% methanol in water prior to introducing 60% methanol/40% buffer mobile phase).
For CORTECS HILIC Columns, flush with 50 column volumes of 50:50 acetonitrile/water with 10 mM final buffer concentration. Prior to the first injection, equilibrate with 20 column volumes of initial mobile-phase conditions (refer to Table 1 for a list of column volumes). See “Getting Started with CORTECS HILIC Columns” for additional information.
d. eCord Installation
The eCord button should be attached to the side of the column heater module. The eCord button is magnetized and does not require specific orientation.
4. Gradually increase the flow rate as described in step 2.
5. Once a steady backpressure and baseline have been achieved, proceed to the next section.
2
CORT ECS Columns
e. Initial Column Efficiency Determination
1. Perform an efficiency test on the column before using it. This test may consist of: a. An analyte test mixture that is commonly used in your laboratory, and/or b. An analyte mixture as found on the “Performance Test Chromatogram” which accompanied your column.
Note: If b. is performed, the isocratic efficiencies measured in your laboratory may be less than those given on the Waters “Performance Test Chromatogram”. This is normal. CORTECS Columns are QC tested on ACQUITY UPLC I-Class Systems which have extremely low system volumes. This presents a more challenging test of how well the column was packed and guarantees the highest quality packed column.
2. Determine the number of theoretical plates (N) and use this value for periodic comparisons.
3. Repeat the test at predetermined intervals to track column performance over time. Slight variations may be obtained on two different UPLC® Systems due to the quality of the connections, operating environment, system electronics, reagent quality, column condition and operator technique.
f. VanGuard Pre-Columns
VanGuard™ Pre-columns are 2.1 mm ID x 5 mm length guard column devices designed specifically for use in the ACQUITY UPLC System. VanGuard Pre-columns are packed with the same CORTECS chemistries and frits as our 2.1 mm ID CORTECS Columns. VanGuard Pre-columns are designed to be attached directly to the inlet side of a CORTECS Column.
Note: In order to ensure void-free and leak-free connections, the VanGuard Pre-column is shipped with the collet and ferrule NOT permanently attached. Care must be taken when removing the O-ring that holds these two pieces on the pre-column tubing.
CORT ECS Column
Place wrench here
Ferrule
Flow
VanGuard Pre-Column
Collet
Place wrench here
Installation Instructions
1. Remove the VanGuard Pre-column from its box and shipping tube and remove plastic plug.
2. Orient the pre-column so that male end is facing up and carefully remove rubber O-ring that holds collet and ferrule in place during shipping (collet and ferrule are not yet permanently attached).
3. Orient the CORTECS Column perpendicular to the work surface so that column inlet is on the bottom (column outlet on top).
4. From below, insert the VanGuard Pre-column into the CORTECS Column inlet and hand-tighten (collet and ferrule are not yet permanently attached).
5. While pushing the VanGuard Pre-Column into the column inlet, turn assembled column and pre-column 180˚ so that pre-column is now on top.
6. Tighten with two 5/16” wrenches placed onto the CORTECS Column flats and the VanGuard Pre-column hex nut (male end) as shown above.
7. Tighten 1/4 turn to set collet and ferrule.
8. Check that the ferrule is set by loosening the connection and inspecting the ferrule depth. A properly set ferrule depth will resemble other connections in the ACQUITY UPLC System.
3
9. Reattach pre-column, apply mobile-phase flow and inspect for leaks.
CORT ECS Columns
Table 2: Buffer Recommendations for Using CORTECS Columns
Additive/Buffer pKa
TFA 0.30 Volatile Yes
Acetic Acid 4.76 Volatile Yes
Formic Acid 3.75 Volatile Yes
Acetate (NH
COOH)
4CH2
Formate (NH
COOH)
4
Phosphate 1 2.15 1.15 – 3.15 Non-volatile No Traditional low pH buffer, good UV transparency.
Phosphate 2 7.20 6.20– 8.20 Non-volatile No
4.76 3.76 – 5.76 Volatile Yes
3.75 2.75 – 4.75 Volatile Yes
Buffer Range
Volatility (±1 pH unit)
Used for Mass Spec
Comments
Ion pair additive, can suppress MS signal, used in the
0.02-0.10% range. Maximum buffering obtained when used with ammonium acetate
salt. Used in 0.1-1.0% range. Maximum buffering obtained when used with ammonium formate
salt. Used in 0.1-1.0% range. Used in the 1-10 mM range. Note that sodium or potassium salts
are not volatile. Used in the 1-10 mM range. Note that sodium or potassium salts
are not volatile.
Above pH 7, reduce temperature/concentration and use a guard column to maximize lifetime.
II. COLUMN USE
To ensure the continued high performance of CORTECS Columns, follow these guidelines:
a. Sample Preparation
1. Sample impurities often contribute to column contamination. One option to avoid this is to use Waters Oasis® Solid-Phase Extraction Cartridges/Columns or Sep-Pak® Cartridges of the appropriate chemistry to cleanup the sample before analysis. For more information, visit www.waters.com/sampleprep
2. It is preferable to prepare the sample in the operating mobile phase or a mobile phase that is weaker than the mobile phase for the best peak shape and sensitivity.
3. If the sample is not dissolved in the mobile phase, ensure that the sample, solvent and mobile phases are miscible in order to avoid sample and/or buffer precipitation.
4. Filter sample with 0.2 µm membranes to remove particulates. If the sample is dissolved in a solvent that contains an organic modifier (e.g., acetonitrile, methanol, etc.) ensure that the membrane material does not dissolve in the solvent. Contact the membrane manufacturer with solvent compatibility questions. Alternatively, centrifugation for 20 minutes at 8000 rpm, followed by the transfer of the supernatant liquid to an appropriate vial, could be considered.
5. For Hydrophilic Interaction Chromatography (HILIC) separations, the samples must be prepared in a high percentage of organic solvents (e.g., 95% acetonitrile). See “Getting Started with CORTECS HILIC Columns”.
b. pH Range
The recommended operating pH range for CORTECS Columns is 2 to 8 for the C See Table 2 for more characteristic details of CORTECS chemistries.
+
and C18 chemistries, and 1 to 5 for the HILIC chemistry.
18
4
CORT ECS Columns
A listing of commonly used buffers and additives is given in Table 2. Additionally, the column lifetime will vary
OSi
O
O
OSi
O
O
depending upon the operating temperature, the type and concentration of buffer used.
Note: Working at the extremes of pH, temperature and/or pressure will result in shorter column lifetimes.
Table 3: CORTECS Chemistry Characteristics
C
18
C18+ HILIC
Chemistry
General purpose, high-efficiency,
Intended Use
reversed-phase column. Balanced retention of acids, bases and neutrals at low and mid-range pH.
Ligand Type Trifunctional C
General purpose, high-efficiency, reversed-phase column. A positively charged surface delivers excellent peak shape for basic compounds at low pH.
18
Trifunctional C
18
High-efficiency column designed for retention of extremely polar analytes. Offers ort hogonal selectivity vs. C
None
Surface Charge Modification None + None
Endcap Style Proprietary Proprietary None
Carbon Load 6.6% 5.7% Unbonded
Ligand Density 2.7 µmol/m
2
2.4 µmol/m
2
N/A
pH Range 2-8 2-8 1-5
Temperature Limits
1
Low pH = 45 °C High pH = 45 °C
Low pH = 45 °C High pH = 45 °C
Low pH = 45 °C High pH = 45 °C
columns.
18
c. Solvents
To maintain maximum column performance, use high quality chromatography grade solvents. Filter all aqueous buffers prior to use through a 0.2 µm filter. Solvents containing suspended particulate materials will generally clog the outside surface of the inlet distribution frit of the column. This will result in higher operating pressure and poorer performance. See Section V for more information.
d. Pressure
CORTECS Columns can tolerate operating pressures up to 18000 psi (1241 bar or 124 MPa).
Note: Working at the extremes of pressure, pH and/or temperature will result in shorter column lifetimes.
e. Temperature
Temperatures between 20–45 °C are recommended for operating CORTECS Columns in order to enhance selectivity, lower solvent viscosity and increase mass transfer rates. When operating at high pH, lower operating temperatures are recommended for longer column lifetime. Working at high temperatures (e.g. > 40 ˚C) may also result in shorter column lifetimes.
Note: Working at the extremes of temperature, pressure and/or pH will result in shorter column lifetimes.
5
CORT ECS Columns
III. COLUMN CLEANING, REGENERATING AND STORAGE
a. Cleaning and Regeneration
Changes in peak shape, peak splitting, shoulders on the peak, shifts in retention, change in resolution or increasing backpressure may indicate contamination of the column. Flushing with a neat organic solvent, taking care not to precipitate buffers, is usually sufficient to remove the contaminant. If the flushing procedure does not solve the problem, purge the column using the following cleaning and regeneration procedures.
Use the cleaning routine that matches the properties of the samples and/or what you believe is contaminating the column (see Table 4). Flush columns with 20 column volumes of solvent. Increasing column temperature increases cleaning efficiency. If the column performance is poor after regenerating and cleaning, call your local Waters office for additional support.
Flush CORTECS HILIC Columns with 50:50 acetonitrile/water to remove polar contaminants. If this flushing procedure does not solve the problem, purge the column with 5:95 acetonitrile/water.
Table 4. Reversed-Phase Column Cleaning Sequence
Polar Samples
1. water
2. methanol
3. tetrahydrofuran (THF)
4. methanol 4. hexane
5. water
6. mobile phase 6. mobile phase
Non-polar Samples**
1. isoproanol (or an appropriate isopropanol/water mixture*)
2. tetrahydrofuran (THF)
3. dichloromethane
5. isopropanol (fol­lowed by an appro­priate isopropanol/ water mixture*)
Proteinaceous Samples
Option 1: Inject repeated aliquots of dimethylsulfoxide (DMSO)
Option 2: gradient of 10% to 90% B where: A = 0.1% trifluoroacetic acid (T FA) in water B = 0.1% trifluoroacetic acid (T FA) in acetonitrile
CN)
(CH
3
Option 3: Flush column with 7M guanidine hydrochloride, or 7M urea
b. Storage
For periods longer than four days at room temperature, store reversed-phase and CORTECS HILIC Columns in 100% acetonitrile. For elevated temperature applications, store immediately after use in 100% acetonitrile for the best column lifetime. Do not store columns in buffered eluents. If the mobile phase contained a buffer salt, flush the column with 10 column volumes of HPLC grade water (see Table 1 for common column volumes) and replace with 100% acetonitrile for storage. Failure to perform this intermediate step could result in precipitation of the buffer salt in the column when 100% acetonitrile is introduced. Completely seal column to avoid evaporation and drying out of the bed.
For periods longer than four days, store CORTECS HILIC Columns in 95:5 acetonitrile/water. Do not store in buffered solvent. If the mobile phase contained a buffered salt, flush the column with 10 column volumes of 95:5 acetonitrile/water (see Table 1 for common column volumes).
Note: If a column has been run with a mobile phase that contains formate (e.g., ammonium formate, formic acid, etc.) and is then flushed with 100% acetonitrile, slightly longer equilibration times may be necessary when the column is re-installed and run again with a formate-containing mobile phase.
* Use low organic solvent content to avoid precipitating buffers.
** Unless a Hexane Tetrahydrofuran Compatibility Kit (P/N 205000464) has been installed, running solvents such as THF or hexane should only be considered when the column cannot be cleaning by running neat, reversed-phase organic solvents such as acetonitrile. Reduce flow rate, lower operating temperatures and limit system exposure to THF and/or hexane.
6
CORT ECS Columns
IV. INTRODUCING eCORD INTELLIGENT CHIP TECHNOLOGY
a. Introduction
The eCord Intelligent Chip provides the history of a column’s performance throughout its lifetime. The eCord will be permanently attached to the
Figure 1. Waters eCord Intelligent Chip
column to assure that the column’s performance history is maintained in the event that the column is moved from one instrument to another.
At the time of manufacture, tracking and quality control information will be downloaded to the eCord. Storing this information on the chip will eliminate the need for a paper Certificate of Analysis. Once the user installs the column, the software will automatically download key parameters into a column history file stored on the chip. In this manual, we explain how the eCord will provide a solution for easily tracking the
eCord inserted into side of column heater
history of the columns, reduce the frustration of paperwork trails, and give customers the reassurance that a well-performing column is installed onto
Figure 2. eCord Inserted into Side of Column Heater
their instruments.
b. Installation
Install the column into the column heater. Plug the eCord into the side of the column heater. Once the eCord is inserted into the column heater the identification and overall column usage information will be available allowing the user to access column information on their desktop.
Waters eCord - intelligent chip
c. Manufacturing Information
The eCord chip provides the user with an overview of the bulk material QC test results.
d. Column Use Information
The eCord chip provides the customer with column use data. The top of the screen identifies the column including chemistry type, column dimensions and serial number. The overall column usage information includes the total number of samples, total number of injections, total sample sets, date of first injection, date of last injection, maximum pressure and temperature. The information also details the column history by sample set including date started, sample set name, user name, system name, number of injections in the sample set, number of samples in the sample set, maximum pressure and temperature in the sample set and the column.
The eCord chip provides QC test conditions and results on the column run by the manufacturer. The information includes mobile phases, running conditions and analytes used to test the columns. In addition the QC results and acceptance is placed onto the column.
7
CORT ECS Columns
V. ADDITIONAL INFORMATION
a. Tips for Maximizing CORTECS Column Lifetimes
1. To maximize CORTECS Column lifetime, pay close attention to:
 Water quality (including water purification system)  Solvent quality  Mobile-phase preparation, storage and age  Sample, buffer and mobile-phase solubilities  Sample quality and preparation.
2. W hen problems arise, often only one improper practice must
be changed.
3. Always remember to:
Use in-line filter unit or, preferably, a VanGuard Pre-column. Discourage bacterial growth by minimizing the use of 100% aqueous mobile phases where possible. Change aqueous mobile phase every 24 – 48 hours (if 100% aqueous mobile phase use is required). Discard old 100% aqueous mobile phases every 24-48 hours to discourage bacterial growth. Add 5% -10% organic modifier to mobile phase A and adjust gradient profile. Filter aqueous portions of mobile phase through
0.2 µm filter. Maintain your water purification system so that it is in good working order. Only use ultra pure water (18 megohm-cm) water and highest quality solvents possible. HPLC grade water is not UPLC grade water. Consider sample preparation (e.g., solid-phase extraction, filtration, etc).
4. Avoid (where possible):
100% aqueous mobile phases (if possible) HPLC-grade bottled water “Topping off” your mobile phases Old aqueous mobile phases. Remember to rinse bottles
thoroughly and prepare fresh every 24 to 48 hrs Using phosphate salt buffer in combination with high ACN concentrations (e.g., > 70%) due to precipitation.
5. Don’t: assume a “bad” column is the culprit when high
backpressure or split peaks are observed: Investigate cause of column failure
• Backpressure
 Mobile phase(s), bacteria, precipitation and/or samples  Peak splitting  Sample quality  Injection solvent strength.
8
CORT ECS Columns
6. Remember: the diameter of CORTECS Column (2.1 and 3.0 mm ID) are often lower than that of a conventional HPLC column and therefore, mobile phases last much longer. To reduce the chances of mobile-phase contamination or degradation, only prepare what you need for analysis or store excess bulk quantities in a refrigerated environment.
7. Mobile-phase related questions to ask: Am I using 100% aqueous mobile phases? Am I able to add a small amount of organic modifier to my mobile phase A? Do I filter my aqueous mobile phases through 0.2 µm filters? How old is my mobile phase? Do I label the bottle with preparation date? Do I “top off” or do I prepare fresh mobile phases every 24 – 48 hrs? What is the quality of my water? Has the quality recently? changed? How is my water purification system working? When was it last serviced? Am I working with pH 7 phosphate buffer (which is VERY susceptible to bacterial growth)?
8. Sample-related questions to ask: If I inject neat standards prepared in mobile phase do I observe these problems? If I prepare my standards in water and prepare them like samples (e.g. SPE, filtration, etc.) do I still observe these problems? Has the quality of my samples changed over time?
b. Getting Started with CORTECS HILIC Columns
1. Due the fact that CORTECS HILIC Columns do not possess a bonded phase, the pH operating range is 1 to 5, and they can be operated at temperatures up to 45 °C.
2. As with any LC column, operating at the extremes of pH, pressures and temperatures will result in decreased column lifetime.
Column Equilibration
1. When the column is first received, flush in 50% acetonitrile/ 50% water with 10 mM final buffer concentration for 50 column volumes.
2. Equilibrate with 20 column volumes of initial mobile-phase conditions before making first injection.
3. If gradient conditions are used, equilibrate with 8-10 column volumes between injections.
4. Failure to appropriately equilibrate the column could result in drifting retention times.
Mobile-Phase Considerations
1. Always maintain at least 3% polar solvent in the mobile phase or gradient (e.g., 3% aqueous/3% methanol or 2% aqueous/1% methanol, etc.). This ensures that the CORTECS particle is always hydrated.
2. Maintain at least 40% organic solvent (e.g., acetonitrile) in your mobile phase or gradient.
Miscellaneous Tips
1. CORTECS HILIC Columns are designed to retain very polar bases. Acidic, neutral and/or non-polar compounds will have limited retention.
2. Optimal flow rates for small (<200 daltons) very polar bases are in the 0.4 to 0.8 mL/min range with the CORTECS HILIC Columns.
3. Avoid phosphate salt buffers to avoid precipitation in HILIC mobile phases. Phosphoric acid is okay.
4. Buffers such as ammonium formate or ammonium acetate will produce more reproducible results than additives such as formic acid or acetic acid. If an additive (e.g., formic acid, etc.) must be used instead of a buffer, use 0.2% (v:v) instead of 0.1%.
5. For best peak shape, maintain a buffer concentration of 10 mM in your mobile phase/gradient at all times.
Injection Solvents
1. If possible, injection solvents should be 95% acetonitrile. The polar solvent (i.e., water, methanol, isopropanol) should be minimized to 25% of the total volume.
2. A generic injection solvent is 75:25 acetonitrile/methanol. This is a good compromise between analyte solubility and peak shape.
3. Avoid water and dimethylsulfoxide (DMSO) as injection sol­vents. These solvents will produce very poor peak shapes.
4. Exchange water or DMSO with acetonitrile by using reversed-phase solid-phase extraction (SPE). If this is not possible, dilute the water or DMSO with organic solvent.
3. In HILIC, it is important to remember that water is the strongest solvent. Therefore, it must be eliminated or minimized in the injection solvent.
4. For initial scouting conditions, run a gradient from 95% acetonitrile to 50% acetonitrile. If no retention occurs, run isocratically with 95:3:2 acetonitrile/methanol/aqueous buffer.
5. Alternate polar solvents such as methanol, acetone or isopropanol can also be used in place of water to increase retention.
6. If using an ACQUITY UPLC System, the weak needle wash should closely match the % organic present in the initial mobile-phase conditions, otherwise, analyte peak shape or retention may suffer.
©2013 Waters Corporation. Waters, ACQUITY UPLC, UPLC, Oasis and SepPak are registered trademarks of Waters Corporation. CORT ECS, VanGuard, eCord and The Science of What’s Possible are trademarks of Waters Corporation. All ot her trademarks are the property of their respective owners.
June 2013 720004750EN Rev. A VW-PDF
Waters Corporation 34 Maple Street Milford, MA 01757 U.S.A. T: 1 508 478 2000 F: 1 508 872 1990 www.waters.com
Loading...