Waters BioSuite Columns User Manual

[ Care and Use ManUal ]
BIosuIte columns
I. IntroductIon
Thank you for choosing a Waters BioSuite™ column. Every column is indi­vidually tested to assure maximum product quality. Waters BioSuite high performance chemistries are dedicated to the isolation, analysis, and charac­terization of proteins and peptides. For the latest list on ion-exchange, size exclusion, hydrophobic interaction and reversed-phase go to www.waters. com/lifesciences
contents
I. IntroductIon
a. Mobile Phase
b. Flow Direction
II. connectIng your column to the hPlc system
III. Waters BIosuIte Pc
and PPhenyl rPc columns
18
IV. Waters BIosuIte Phenyl hIc columns
V. Waters BIosuIte ultra hIgh resolutIon (uhr), hIgh
resolutIon (hr), and sIze exclusIon (sec) columns
VI. Waters BIosuIte Q-PeeK and sP-PeeK columns
VII. Waters BIosuIte deae and sP nP columns
VIII. Waters BIosuIte Q, deae, sP, and cm columns
BioSuite Columns 1
[ Care and Use ManUal ]
0.175 inches
Void
a. Mobile Phase
Prevent air bubbles from entering the column during its installation, use,
and storage since this may cause degradation of column performance
through the formation of channels in the packed bed. Mobile phases must be
thoroughly degassed before use. This can be accomplished by vacuum filtra-
tion, helium sparging or by in-line vacuum degassing. In addition to degas-
sing the solvent, vacuum filtration of the solvent will also prevent small
particles from plugging the column frit. You can use 0.20 or 0.45 µm mem-
brane to filter aqueous and aqueous/organic mobile phases. (Note: Consult
with filtration membrane manufacturer for details on solvent compatibility.)
Note: Use high quality reagents, HPLC grade water, and HPLC grade solvents
for preparing buffers to maximize column efficiency especially when using
Waters BioSuite Non-Porous (NP) columns.
The useful column lifetime is a function of numerous factors including: the
cleanliness and composition of the mobile phase and the sample; the flow
rate and pressure used; and the temperature. Refer to the section below
about “Cleaning” for information on extending column life.
Note: Cleaning, however, is not effective when the column is damaged by
irreversible sample adsorption, channeling, or exposure of the packing mate-
rial to excessive heat or shock.
tubing and the column that will result in undesired peak broadening (Figure
B). To correct this problem, Cut the tubing, place a new ferrule on it and re-
make the connection (i.e., 0.175” depth) to the Waters BioSuite column.
Figure A
Figure B
In a proper tubing/column connection, the tubing touches the bottom of the
column end fitting with no void between them. (Figure C)
b. Flow Direction
The recommended flow direction through the column is indicated by the
arrow on the tag. Operating the column with the flow in the reverse direc-
tion is only recommended as part of a cleaning procedure when removing
particulates from a clogged frit.
II. connectIng your column to the hPlc system
Due to the absence of an industry standard, various column manufacturers
have employed different styles of chromatographic column connectors. The
chromatographic performance of your separation can be negatively affected
if the style of column end fitting does not match the existing instrumenta-
tion tubing ferrule setting. Waters BioSuite columns require end fittings
that have a 0.175” depth between the ferrule and the end of the extending
stainless steel tubing (Figure A) that must perfectly seat to the end of the
BioSuite column. Connecting a non-BioSuite style connector to a Waters
BioSuite column will leave a gap between the end of the stainless steel
BioSuite Columns 2
Figure C
After manufacturing and quality control, the column has been flushed with
storage solvent and closed with caps to prevent solvent evaporation. When
installing the column to the Liquid Chromatograph, it is important to prevent
air from entering the column. The following steps can be used to minimize
this potential problem:
• Remove the cap from the column’s inlet side. Solvent should be visible at
the inlet fitting (if not, see below).
• StartsolventflowthroughtheliquidchromatographBEFOREconnecting
the Waters BioSuite column to the system. Turn off the pump after a
steady flow of solvent from the system is observed. You can now connect
the waters BioSuite column to the liquid chromatograph and slowly begin
delivering solvent to the column (Note: If the Waters BioSuite column
inlet fitting appears dry, we recommend that you first disconnect the
bottom cap and hook up the column exit to the injector. Then slowly start
[ Care and Use ManUal ]
the flow in this reversed flow direction until a few drops of mobile phase
exit from the column. Turn off the flow, let the pressure go to zero, and
disconnect the column from the system. Turn the column around and hook
it up so that the flow is now in the direction of the arrow. Start the flow
at a low setting and stop it as soon as the mobile phase exits from the
bottom fitting. Now you can hook up the column to the detector inlet and
increase the flow to the desired setting.)
• Besuretosettheflowwithintherecommendedrangeshowninthe
following tables.
Sample
If possible, always dissolve your sample in mobile phase or the starting
mobile phase when operating under gradient conditions. Alternatively, try
to match the pH, salt concentration and organic solvent of the sample with
those of the mobile phase and run a test to ensure that no precipitate, sus-
pension or flocculate is formed. Finally, before making an injection, filter the
sample through a 0.20-0.45 µm porosity membrane.
III. Waters BIosuIte Pc18 and PPhenyl rPc columns
This section contains recommended operating conditions and specifications
for the Waters polymer-based, C
and phenyl reversed-phase columns.
18
Operating Conditions
1. Shipping Solvent: 70% Methanol - 30% Water
Note: When using a high viscosity buffer and ambient temperature, the maxi­mum flow rate may have to be reduced to avoid exceeding the maximum recommended operating pressure for the column. When changing solvents, use 50% of the maximum flow rate (20% for the 21.5 mm i.d. preparative
column).
3. Recommended pH Range: 2 - 12 (Avoid extended operation at pH >
10 to maximize column life.)
4. Recommended Organic Concentration Range: 0 - 100% (Note:
Reduced organic concentrations may be required if buffers are present
in the eluent in order to avoid salt precipitation.)
5. Recommended Temperature Range: 5 - 50 °C (Note: Reduce flow rate
when operating at low temperatures (e.g.. 5 C) to avoid excessive
column pressure.)
6. Recommended Cleaning Solvents/Procedures:
A) Acetonitrile or Methanol, or
B) Aqueous buffer in organic solvent, or
C) 0.1 - 0.2M NaOH, or
D) 20 - 40% acetic acid in HPLC grade water.
Clean the column regularly with any of the listed solvents. Cleaning
solvents (C) and (D) can be injected onto the column in 250 µl incre-
ments for a total of one column volume. Use proportionally larger
volumes for the 21.5 mm i.d. prep column. Following use of cleaning
solvent, rinse the column with 3 to 5 column volumes of HPLC grade
water before storage.
2. Recommended Operating Flow Rates/Max Pressure:
Description Matrix i.d.
(mm)L (mm)
BioSuite pC NP RPC
BioSuite pC 7 µm RPC
BioSuite pC 7 µm RPC
BioSuite pC 13 µm RPC
BioSuite pPhenyl, 1000, 10 µm RPC
BioSuite pPhenyl, 1000, 10 µm RPC
BioSuite pPhenyl, 1000, 13 µm RPC
BioSuite Columns 3
, 2.5 µm
18
, 500,
18
, 500,
18
, 500,
18
polymer 4.6 35 1.0 - 1.6 mL/min /
polymer 4.6 150 0.08 - 0.22 mL/min
polymer 4.6 150 0.5 - 1.2 mL/min /
polymer 21.5 150 3.0 - 8.0 mL/min /
polymer 2.0 75 0.05 – 0.12 mL/min
polymer 4.6 75 0.5 - 1.2 mL/min /
polymer 21.5 150 6.0 - 8.0 mL/min /
Flow Rates / Max
Pressure
3000 PSI
/ 1500 PSI
1750 PSI
500 PSI
/ 150 PSI
450 PSI
450 PSI
7. Recommended Storage: For overnight storage, continuously flush the
column with the mobile phase at 10 - 20% of the maximum recom-
mended flow rate. Store the column in the shipping solvent when the
column will not be used within 48 hours.
[ Care and Use ManUal ]
IV. Waters BIosuIte Phenyl hIc columns
This section contains recommended operating conditions and specifica-
tions for the Waters polymer-based, Phenyl Hydrophobic-interaction
Chromatography columns.
Operating Conditions
1. Shipping Solvent: HPLC grade Water
2. Recommended Operating Flow Rates/Max Pressure:
Description Matrix i.d.
(mm)L (mm)
BioSuite Phenyl, 1000, 10 µm HIC
BioSuite Phenyl, 1000, 13 µm HIC
polymer 7.5 75 0.5 - 1.2 mL/min /
polymer 21.5 150 4.0 - 8.0 mL/min /
Note: When using a high viscosity buffer and ambient temperature, the maxi-
mum flow rate may have to be reduced to avoid exceeding the maximum
recommended operating pressure for the column. When changing solvents,
use 50% of the maximum flow rate (20% for the 21.5 mm i.d. preparative
column).
3. Recommended pH Range: 2 - 12 (Avoid extended operation at pH > 10
to maximize column life.)
4. Recommended Organic Conc.: less than 50% (Note: Reduced organic
concentrations may be required if buffers are present in the eluent in
order to avoid salt precipitation.)
5. Recommended Salt Conc.: Less than or equal to 4.0 Molar
Flow Rates /
Max Pressure
300 PSI
300 PSI
mended flow rate. Store the column in the HPLC grade water
when it will be used within 24 hrs or in either 20% ethanol or metha-
nol for long term storage. (Caution: Rinse column in HPLC
grade water before switching to 20% organic solvent to avoid salt
precipitation in column.)
V. Waters BIosuIte ultra-hIgh resolutIon (uhr), hIgh resolutIon (hr), and sIze-exclusIon (sec) columns
This section contains recommended operating conditions and specifications
for the Waters silica-based, ultra-high Resolution (UHR), high-resolution (HR),
and standard size-exclusion (SEC) columns.
Operating Conditions
1. Shipping Solvent: 0.05% Sodium Azide or 0.1M Sodium Sulphate in
0.1M phosphate buffer, pH 6.7
2. Recommended Operating Flow Rate/Max Pressure:
Description Matrix i.d.
(mm)L (mm)
BioSuite 125, 4 µm UHR SEC
BioSuite 250, 4 µm UHR SEC
BioSuite UHR Guard SEC silica 4.6 35 0.1 - 0.4 mL/min /
silica 4.6 300 0.1 - 0.4 mL/min /
silica 4.6 300 0.1 - 0.4 mL/min /
Flow Rates /
Max Pressure
1800 PSI
1800 PSI
1800 PSI
6. Recommended Temperature: 10-50 °C. Reduce flow rate when
operating at low temperatures (e.g. 10 °C) to avoid excessive column
pressure.
7. Recommended Cleaning Solvents/Procedures:
A) 0.1 - 0.2M NaOH, and, if not successful,
B) 20 - 40% acetic acid in HPLC grade water.
Clean the column regularly by injecting up to one column volume 0.1
- 0.2M NaOH in 250 µl increments. Use proportionally larger volumes
for the 21.5 mm i.d. preparative column. Following use of cleaning
solvent, rinse the column with 3 to 5 column volumes of HPLC grade
water before storage.
8. Recommended Storage: For overnight storage, continuously flush the
column with the mobile phase at 10 - 20% of the maximum recom-
BioSuite Columns 4
BioSuite 125, 5 µm HR SEC
BioSuite 250, 5 µm HR SEC
BioSuite 450, 8 µm HR SEC
BioSuite HR Guard SEC silica 6 40 0.5 - 1.2 mL/min /
BioSuite 125, 10 µm SEC
BioSuite 125, 13 µm SEC
BioSuite 250, 10 µm SEC
silica 7.8 300 0.5 - 1.2 mL/min /
1050 PSI
silica 7.8 300 0.5 - 1.2 mL/min /
1050 PSI
silica 7.8 300 0.5 - 1.2 mL/min /
525 PSI
1050 PSI
silica 7.5 300 0.5 - 1.2 mL/min /
300 PSI
silica 21.5 300 3.0 - 8.0 mL/min /
150 PSI
silica 7.5 300 0.5 - 1.2 mL/min /
375 PSI
[ Care and Use ManUal ]
Description Matrix i.d.
(mm)L (mm)
BioSuite 250, 13 µm SEC
BioSuite 450, 13 µm SEC
BioSuite 450., 17 µm SEC
BioSuite Guard SEC silica 7.5 75 0.5 - 1.2 mL/min /
BioSuite Guard SEC silica 21.5 75 3.0 - 8.0 mL/min /
silica 21.5 300 3.0 - 8.0 mL/min /
silica 7.5 300 0.5 - 1.2 mL/min /
silica 21.5 300 3.0 - 8.0 mL/min /
Flow Rates /
Max Pressure
225 PSI
225 PSI
300 PSI
600 PSI
450 PSI
Note: When using a high viscosity buffer and ambient temperature, the maxi-
mum flow rate may have to be reduced to avoid exceeding the maximum
recommended operating pressure for the column. When changing solvents,
use 50% of the maximum flow rate.
3. Recommended pH Range: 2.5 - 7.5
4. Recommended Salt Conc.: Less than or equal to 0.5 Molar
5. Recommended Organic Conc.: 0 - 100% (Caution: Many proteins and
peptides are insoluble at elevated organic concentrations. Prior to
chromatography, test to ensure the sample does not precipitate at the
organic concentration to be used for the chromatography.)
before switching to 20% organic solvent to avoid salt precipitation in
column.)
9. Recommended Column Protection: Use of guard columns is recom-
mended to prolong the life of the analytical column life. Guard column
life depends greatly on sample cleanliness. As a general rule, guard
columns should be replaced after every 30-40 sample injections,
when the peaks become excessively wide, or when the peaks show
splitting.
VI. Waters BIosuIte Q-PeeK and sP-PeeK columns
This section contains recommended operating conditions and specifications
for the Waters polymer-based, anion and cation-exchange materials con-
tained in PEEK
columns.
Note: Compared to use of stainless-steel columns, connect the PEEK column
to the HPLC or Liquid Chromatographic System with a 10-32 polymer nut
and ferrule.
Operating Conditions:
1. Shipping Solvent: 20% Ethanol containing 20 mM Tris-HCl buffer
(pH 8.0)
6. Recommended Temperature: 10-30 °C. Reduce flow rate when
operating at low temperatures (e.g., 10 °C) to avoid excessive column
pressure.
7. Recommended Cleaning Solvents/Procedures:
A) A concentrated salt solution at low pH, e.g. 0.5M Na
2SO4
,
pH 2.7
B) A low concentration of methanol or acetonitrile (e.g., 20%) in
HPLC grade water
C) A buffered solution of SDS, urea, or guanidine
Note: Choose a cleaning solvent based on sample properties, e.g. use (A) to
remove basic proteins, and (B) to remove hydrophobic proteins. Chaotrophic
agents can solvate strongly adsorbed proteins via hydrogen bond disruption.
8. Recommended Storage: For overnight storage, continuously flush the
column with the mobile phase at 10 - 20% of the maximum recom-
mended flow rate. Store the column in the HPLC grade water when
it will be used within 24 hrs or in either 20% ethanol or methanol
for long term storage. (Caution: Rinse column in HPLC grade water
2. Recommended Operating Flow Rates/Max. Pressure:
Description Matrix i.d.
(mm)L (mm)
BioSuite Q-PEEK 10 µm AXC
BioSuite SP-PEEK 7 µm CXC
polymer 4.6 50 0.3 - 1.2 mL/min /
polymer 4.6 50 0.3 - 1.0 mL/min /
Flow Rates /
Max Pressure
375 PSI
375 PSI
Note: When using a high viscosity buffer and ambient temperature, the maxi-
mum flow rate may have to be reduced to avoid exceeding the maximum
recommended operating pressure for the column. When changing solvents,
use 50% of the maximum flow rate.
3. Recommended pH Range: 2 - 12 (Avoid extended operation at pH
values less than 3 or greater than 10 to maximize column life.)
4. Recommended Salt Conc.: Less than or equal to 2.5 Molar
5. Recommended Organic Conc.: 0 - 30%. (Caution: Some large
molecular weight proteins may be insoluble at elevated organic
BioSuite Columns 5
[ Care and Use ManUal ]
concentrations. Prior to chromatography, test to ensure the sample
does not precipitate at the organic concentration to be used for the
chromatography.)
6. Recommended Temperature: 4 - 60 °C. Reduce flow rate when
operating at low temperatures (e.g., 4 °C) to avoid excessive column
pressure.
7. Recommended Cleaning Solvents/Procedures:
A) 0.1 - 0.5M NaOH in HPLC grade water and if not successful,
B) 20 - 40% acetic acid in HPLC grade water or,
C) 30% acetonitrile or 30% methanol in HPLC grade water or,
D) 30% ethanol containing 0.5M NaOH or,
E) 8M Urea, 6M Guanidine or other nonionic surfactant HPLC
grade water
Clean the column regularly by injecting up to one column volume of
solution (A) (i.e., 0.1 - 0.5M NaOH in HPLC grade water) in 100 to
250 microliter increments. To maximize cleaning effect, reverse the
direction of eluent flow through the column and reduce the flow rate by
25%.
If desired, the HPLC column can also be cleaned by delivering, in
reverse eluent flow direction and at a 25 % standard flow rate, one of
the alternative cleaning solvents (i.e., 20 - 40% acetic acid in HPLC
grade water, or 30% acetonitrile or methanol in HPLC grade water, or
30% ethanol containing 0.5M NaOH, or either 8M Urea, 6M Guanidine
or other nonionic surfactant in HPLC grade water).
8. Recommended Storage: For overnight storage, continuously flush the
column with the mobile phase at 10 - 20% of the maximum recom-
mended flow rate. Store the column in the HPLC grade water when
it will be used within 24 hrs or in 20% ethanol containing 20mM
Tris-HCl buffer, pH 8 for long term storage.
VII. Waters BIosuIte deae and sP nP columns
This section contains recommended operating conditions and specifications
for the Waters non-porous, polymer-based, anion-and cation-exchange col-
umns.
Operating Conditions:
1. Shipping Solvent: The columns are shipped in HPLC grade water. Upon
receiving, flush the column with 30% acetonitrile/70% HPLC grade
water for 15 minutes at 1 mL/min. This is followed by equilibrating the
columns with the starting buffer.
Use high quality reagents, HPLC grade water, and HPLC grade solvents
for preparing buffers to maximize Waters BioSuite non-porous (NP)
column efficiency and life. Filter all eluents through a 0.20 or 0.45
µm membrane before use (Consult the with filtration membrane
manufacturer for details on solvent compatibility.) Fouling of the
non-porous resin, leading to a loss in retention and/or efficiency,
occurs faster compared to use of porous packing materials due to the
small surface area of non-porous resin particles. The useful column
lifetime is a function of numerous factors including the cleanliness and
composition of the mobile phase and the sample, the flow
rate and operating temperature, and the operating pressure. While
column cleaning is possible (see below), cleaning is not effective when
the column is damaged by irreversible sample
adsorption, channeling, or exposure of the packing material to exces-
sive heat of shock.
2. Recommended Operating Flow Rates/Max Pressure:
Description Matrix i.d.
(mm)L (mm)
BioSuite DEAE 2.5 µm NP AXC
BioSuite SP 2.5 µm NP CXC
Note: When using a high viscosity buffer and ambient temperature, the maxi-
mum flow rate may have to be reduced to avoid exceeding the maximum
recommended operating pressure for the column. When changing solvents,
use 50% of the maximum flow rate.
polymer 4.6 35 1.0 - 1.6 mL/min /
polymer 4.6 35 1.0 - 1.6 mL/min /
Flow Rates /
Max Pressure
3000 PSI
3000 PSI
BioSuite Columns 6
3. Recommended pH Range: 2 - 12 (Avoid extended operation at pH
values less than 3 or greater than 10 to maximize column life.)
4. Recommended Salt Conc.: Less than or equal to 1.0 Molar
5. Recommended Organic Conc.: 0 - 20%.
[ Care and Use ManUal ]
6. Recommended Temperature: 4 - 60 °C. Reduce flow rate when
operating at low temperatures (e.g., 4 °C) to avoid excessive column
pressure.
7. Recommended Cleaning Solvents/Procedures:
A) 0.1 - 0.2M NaOH in HPLC grade water and if not successful,
B) 20 - 40% acetic acid in HPLC grade water or,
C) 30% acetonitrile, 30% methanol, or 30% ethanol in HPLC
grade water or,
D) 8M Urea, 6M Guanidine or other nonionic surfactant HPLC
grade water
Clean the column regularly by injecting up to one column volume of
solution (A) (i.e., 0.1 - 0.2M NaOH in HPLC grade water) in 100 to
250 microliter increments. To maximize cleaning effect, reverse the
direction of eluent flow through the column and reduce the flow rate by
25%.
If desired, the HPLC column can also be cleaned by delivering, in
reverse eluent flow direction and at a 25% standard flow rate, one of
the alternative cleaning solvents (i.e., 20 - 40% acetic acid in HPLC
grade water, or 30% acetonitrile, 30% methanol, or 30% ethanol in
HPLC grade water, or either 8M Urea, 6M Guanidine or other nonionic
surfactant in HPLC grade water).
8. Recommended Storage: For overnight storage, continuously flush the
column with the mobilephase at 10 - 20% of the maximum recom-
mended flow rate. Store the column in the HPLC grade water when it
will be used within 24 hrs or in 20% acetonitrile in HPLC grade water
for long term storage.
VIII. Waters BIosuIte Q, deae, sP, and cm columns
This section contains recommended operating conditions and specifications
for the Waters polymer-based, anion- and cation-exchange columns.
Operating Conditions:
1. Shipping Solvent: HPLC grade Water
2. Recommended Operating Flow Rate/Max Pressure:
Description Matrix i.d.
(mm)L (mm)
BioSuite Q 10 µm AXC polymer
BioSuite Q 13 µm AXC polymer 21.5 150 4.0 - 8.0 mL/min /
BioSuite DEAE 10 µm AXC
BioSuite DEAE 13 µm AXC
BioSuite SP 10 µm CXC polymer 7.5 75 0.5 - 1.2 mL/min /
BioSuite SP 13 µm CXC polymer 21.5 150 40 - 8.0 mL/min /
BioSuite CM 10 µm CXC polymer 7.5 75 0.5 - 1.2 mL/min /
BioSuite CM 13 µm CXC polymer 21.5 150 4.0 - 8.0 mL/min /
polymer 7.5 75 0.5 - 1.2 mL/min /
polymer 21.5 150 4.0 - 8.0 mL/min /
7.5
75 0.5 - 1.2 mL/min /
Note: When using a high viscosity buffer and ambient temperature, the maxi-
mum flow rate may have to be reduced to avoid exceeding the maximum
recommended operating pressure for the column. When changing solvents,
use 50% of the maximum flow rate (20% for the 21.5 mm i.d. preparative
column).
Flow Rates /
Max Pressure
300 PSI
300 PSI
225 PSI
375 PSI
225 PSI
375 PSI
225 PSI
375 PSI
BioSuite Columns 7
3. Recommended pH Range: 2 - 12 (Avoid extended operation at pH
values less than 3 or greater than 10 to maximize column life.)
4. Recommended Salt Conc.: Less than or equal to 1.0 Molar
5. Recommended Organic Conc.: 0 - 20%.
6. Recommended Temperature: 4 - 40 °C. Reduce flow rate when
operating at low temperatures (e.g., 4 °C) to avoid excessive column
pressure.
7. Recommended Cleaning Solvents/Procedures:
A) 0.1 - 0.2M NaOH in HPLC grade water and if not successful,
B) 20 - 40% acetic acid in HPLC grade water, or
[ Care and Use ManUal ]
C) 30% acetonitrile, 30% methanol, or 30% ethanol in HPLC
grade water, or
D) 8M Urea, 6M Guanidine or other nonionic surfactant HPLC
grade water
Clean the column regularly by injecting up to one column volume
of solution (A) (i.e., 0.1 - 0.2M NaOH in HPLC grade water) in 250
microliter to 2 mL increments. To maximize cleaning effect, reverse the
direction of eluent flow through the column and reduce the flow rate
by 25%.
If desired, the HPLC column can also be cleaned by delivering, in
reverse eluent flow direction and at a 25% standard flow rate, one of
the alternative cleaning solvents (i.e., 20 - 40% acetic acid in HPLC
grade water, or 30% acetonitrile, 30% methanol, or 30% ethanol in
HPLC grade water, or either 8M Urea, 6M Guanidine or other nonionic
surfactant in HPLC grade water).
8. Recommended Storage: For overnight storage, continuously flush the
column with the mobile phase at 10 - 20% of the maximum recom-
mended flow rate. Store the column in the HPLC grade water when it
will be used within 24 hrs or in 20% acetonitrile in HPLC grade water
for long term storage.
Austria and European Export (Central South Eastern Europe, CIS and Middle East) 43 1 877 18 07
Australia 61 2 9933 1777 Belgium 32 2 726 1000 Brazil 55 11 4134 3788 Canada 1 800 252 4752 x2205 China 86 21 6879 5888 CIS/Russia +497 727 4490/290 9737 Czech Republic 420 2 617 1 1384 Denmark 45 46 59 8080 Finland 09 5659 6288 France 33 1 30 48 72 00 Germany 49 6196 400600 Hong Kong 852 29 64 1800 Hungary 36 1 350 5086 India and India Subcontinent 91 80 2837 1900 Ireland 353 1 448 1500 Italy 39 02 265 0983 Japan 81 3 3471 7191 Korea 82 2 6300 4800 Mexico 52 55 5524 7636 The Netherlands 31 76 508 7200 Norway 47 6 384 60 50 Poland 48 22 833 4400 Puerto Rico 1 787 747 8445 Singapore 65 6273 7997 Spain 34 93 600 9300 Sweden 46 8 555 11 500 Switzerland 41 56 676 70 00 Taiwan 886 2 2543 1898 United Kingdom 44 208 238 6100 All other countries:
Waters Corporation U.S.A.
1 508 478 2000 1 800 252 4752
www.waters.com
©2009 Waters Corporation, Waters, The Science of W hat’s Possible, and BioSuite are trademarks of Waters Corporation. PEEK is a trademark of Victrex plc.
March 2009 715000652 Rev C K K-PDF
BioSuite Columns 8
Waters Corporation
34 Maple Street Milford, MA 01757 U.S.A. T: 1 508 478 2000 F: 1 508 872 1990 www.waters.com
Loading...