matographic results and conditions. T hese data should be stored for
future reference.
), and chromatographic results and conditions.
18
VIII. InItIAL coLuMn effIcIency deterMInAtIon
1. Perform an efficiency test on the column before using it. Waters rec-
ommends using a suitable solute mixture, as found in the “Perfor-
mance Test Chromatogram”, to verify the performance of the column
upon receipt.
2. Determine the number of theoretical plates (N) and use for peri-
odic comparison.
3. Repeat the test periodically to track column performance over
time. Slight variations may be obtained on two different HPLC
systems due to the quality of the connections, operating environ-
ment, system electronics, reagent quality, column condition and
operator technique.
4. When the mobile phase is changed, gradually increase the flow
rate of the new mobile phase from 0.0 mL/min to 1.0 mL/min in
0.1 mL/min increments.
5. Once a steady backpressure and baseline have been achieved, the
column is ready to be used (or equilibrated).
Note: If mobile phase additives are present in low concentrations
(e.g., ion-pairing reagents), 100 to 200 column volumes may be
required for complete equilibration. In addition, mobile phases
that contain formate (e.g., ammonium formate, formic acid, etc.)
may also require slightly longer initial column equilibration times.
Please see additional equilibration information for Atlantis HILIC
Silica columns in “HILIC Getting Started.”
VII. coLuMn PerforMAnce VALIdAtIon
Each Atlantis column comes with a Certificate of Batch Analysis and
a Performance Test Chromatogram. The Certificate of Analysis is spe-
cific to each batch of packing material and includes the batch number,
analysis of unbonded particles, analysis of bonded particles (Atlantis
Note: If 1) is performed, the isocratic efficiencies measured in your
laboratory may be less than those given on the Waters “Perfor-
mance Test Chromatogram.” This is normal. The Waters isocratic
column testing systems have been modified in order to achieve
extremely low system volumes. This presents a more challeng-
ing test of how well the column was packed. This guarantees the
highest quality packed column. These special testing systems have
been modified to such an extent that they are not commercially
viable and have limited method flexibility other than isocratic
column testing.
IX. coLuMn usAGe
Caution: Accumulation of particulates from solvents, samples, or pump
seals may cause the column backpressure to increase over time. This
may lead to a system shutdown or leaking of column connections.
Accumulation of contaminants from “dirty” samples at the column
inlet may lead to a loss of resolution or ion suppression in a mass
spectrometer, resulting in erroneous results.
Atlantis Columns
6
Page 7
[ Care and Use ManUal ]
To ensure the continued high performance of Atlantis columns and
cartridges, follow these guidelines:
a. Guard Columns
Use a Waters Sentry guard cartridge of matching i.d., chemistry and
particle size between the injector and main column. For best results, the
guard column should be replaced prior to the observation of a substantial
loss in resolution or increase in system backpressure. It is important to use
a high-performance matching guard column to protect the main column
while not compromising or changing analytical resolution.
b. Sample Preparation
1. Sample impurities often contribute to column contamination. Use
®
Waters Oasis
of the appropriate chemistry to cleanup the sample before analysis.
2. For reversed-phase separations (Atlantis T3 and dC
sample in mobile phase or a solvent that is weaker (less organic
modifier) than the mobile phase. For Hydrophilic Interaction
Chromatography (HILIC) separations (Atlantis
samples must be prepared in 100% organic solvents (e.g., aceto-
nitrile). See “HILIC Getting Started” for additional information.
3. If the sample is not dissolved in the mobile phase, ensure that the
sample and diluent are miscible in the mobile phase(s) in order to
avoid sample and/or diluent precipitation.
4. Filter sample through a 0.2 µm membrane to remove particulates.
If the sample is dissolved in a solvent that contains an organic
modifier (e.g., acetonitrile, methanol, etc.) ensure that the
membrane material does not dissolve in the solvent. Contact the
membrane manufacturer with solvent compatibility questions.
c. Recommended pH Range
Atlantis HILIC Silica: 1-5 Atlantis T3: 2-8 Atlantis dC
Column lifetime will vary depending upon the temperature, type and
concentration of buffer used. A listing of recommended and non-rec-
ommended buffers is given in Table 3. Please use this as a guideline
when developing methods.
or Sep-Pak® solid-phase extraction cartridges/columns
) prepare the
18
HILIC Silica), the
: 3-7
18:
Attention: Operating at the upper or lower end of the pH range in com-
bination with elevated temperatures will lead to shorter column lifetime
and/or may result in the column generating high backpressure.
Table 3: Buffer recommendations for using Atlantis columns
from pH 1 to 7
Additive or Buffer range Used for
Buffer pKa (±1 pH unit) Volatility Mass Spec? Comments
TFA 0.3 Volatile Yes Ion pair additive, can suppress MS signal.
Used in the 0.01-0.1% range.
Formic Acid 3.75 Volatile Yes Maximum buffering obtained when used with
Ammonium Formate salt. Used in 0.1-1.0% range.
Acetic Acid 4.76 Volatile Yes Maximum buffering obtained when used with
Ammonium Acetate salt. Used in 0.1-1.0% range.
Formate 3.75 2.75 – 4.75 Volatile Yes Used in the 1-10mM range. Note: sodium or
COOH) potassium salts are not volatile.
(NH
4
Acetate 4.76 3.76 – 5.76 Volatile Yes Used in the 1-10mM range. Note: sodium or
COOH) potassium salts are not volatile.
(NH
4CH2
Phosphate 1* 2.15 1.15 – 3.15 Non-volatile No Traditional low pH buffer, good UV transparency
Phosphate 2* 7.2 6.20 – 8.20 Non-volatile No Much shorter colum lifetimes will be realized
using phosphate at pH 7
* Phosphate salt buffers are not recommended for HILIC (phosphoric acid is OK) due to phosphate buffer
salt insolubility at high acetonitrile concentrations.
d. Solvents
To maintain maximum column performance, use high quality chrom-
atography grade solvents. Filter all aqueous buffers prior to use. The
addition of at least 5% organic to neutral pH buffers is recommended
®
to prevent bacterial growth. Pall Corporation Acrodisc
will generally clog the outside surface of the inlet frit of the column.
This will result in higher operating pressure and poorer performance.
Degas all solvents thoroughly before use to prevent bubble formation
in the pump and detector. The use of an on-line degassing unit is also
recommended. This is especially important when running low pressure
gradients since bubble formation can occur as a result of aqueous and
organic solvent mixing during the gradient.
e. Pressure
Atlantis columns can tolerate pressures of up to 6,000 psi (400 bar
or 40 Mpa) although pressures greater than 4,000 - 5,000 psi should
be avoided in order to maximize column and system lifetimes, and the
risk of system shutdowns and leaking.
Atlantis Columns
7
Page 8
[ Care and Use ManUal ]
f. Temperature
Temperatures between 20 ˚C - 45 ˚C are recommended for operating
Atlantis columns in order to enhance selectivity, lower solvent viscosity
and increase mass transfer rates. However, any temperature rise above
ambient will have a negative effect on lifetime which will vary depend-
ing on the pH and buffer conditions used. The combination of operating
at elevated temperatures and at pH extremes should be avoided.
g. Scaling Up/Down Isocratic Methods
The following formulas will allow scale up or scale down, while main-
taining the same linear velocity (retention time), and provide new
sample loading values:
If column i.d. and length are altered: F2 = F1(r2/r1)
or: Load2 = Load1(r2/r1)2(L2/L1)
or: Inj vol
Where: r = Radius of the column, in mm
F = Flow rate, in mL/min
L = Length of column, in mm
1 = Original, or reference column
2 = New column
2
= Inj vol2 (r2/r1)2 (L2/L1)
1
XII. HILIc GettInG stArted
a. Equilibration of Atlantis HILIC Silica Columns
1. Upon receipt, equilibrate in 50% acetonitrile/50% aqueous buffer
(10 mM final buffer concentration) for 50 column volumes.
2. Prior to first injection, equilibrate with 20 column volumes of
initial mobile phase conditions.
3. When running gradients, equilibrate with 10 column volumes
between injections.
Failure to appropriately equilibrate the column could result in drifting
retention times.
2. Maintain at least 40% organic solvent (e.g., acetonitrile) in your
mobile phase or gradient.
3. Avoid phosphate salt buffers to avoid precipitation in HILIC mobile
phases (phosphoric acid is OK).
4. Buffers such as ammonium formate or ammonium acetate, will
produce more reproducible results than additives such as formic
acid or acetic acid. If an additive (e.g., formic acid) must be used
instead of a buffer, use 0.2% (v:v) instead of 0.1%.
5. For best peak shape, maintain a buffer concentration of 10 mM in
your mobile phase/gradient at all times.
c. Injection Solvents for HILIC
1. If possible, injection solvents should be 100% organic solvent.
Water must be eliminated or minimized. Choose weak HILIC
solvents such as acetonitrile, isopropanol, methanol, etc.
2. A generic injection solvent is 75:25 acetonitrile methanol. This is
a good compromise between analyte solubility and peak shape.
3. Avoid water and dimethylsulfoxide (DMSO) as injection solvents.
These solvents will produce very poor peak shapes.
4. Exchange water or DMSO with acetonitrile by using reversed-
phase solid-phase extraction. If this is not possible, dilute the
water or DMSO with organic solvent.
d. Additional HILIC Recommendations
1. For initial scouting conditions, run a gradient from 95% acetoni-
trile to 50% acetonitrile. If no retention occurs, run isocratically
with 95:3:2 acetonitrile:methanol:aqueous buffer.
2. Alternate polar solvents such as methanol, acetone or isopro-
panol can also be used in place of water in the mobile phase to
increase retention.
b. HILIC Mobile Phase Considerations
1. Always maintain at least 5% polar solvent in the mobile phase
or gradient (e.g., 5% water, 5% methanol or 3% methanol/2%
aqueous buffer, etc.). T his ensures that the Atlantis HILIC Silica
particle is always hydrated.
Atlantis Columns
3. Be sure that your needle wash solvent/purge solvent contains the
same high organic solvent concentration as your mobile phase,
else peak shapes will suffer.
8
Page 9
[ Care and Use ManUal ]
XIII. coLuMn cLeAnInG, reGenerAtInG And storAGe
a. Cleaning and Regeneration
A sudden increase in pressure or shift in retention or resolution may
indicate contamination of the column.
Atlantis T3 and dC
the non-polar contaminant(s). If this flushing procedure does not solve
the problem, purge the column with a sequence of progressively more
non-polar solvents. For example, switch from water to tetrahydrofuran
to methylene chloride. Return to the standard mobile phase conditions
by reversing the sequence.
Atlantis HILIC Silica – Flush with 50:50 acetonitrile:water to remove
the polar contaminant(s). If this flushing procedure does not solve the
problem, purge the column with 5:95 acetonitrile:water.
Guard columns require replacement at regular intervals as determined
by sample contamination. When system backpressure increases above
a set pressure limit, it is usually an indication that the guard column
should be replaced. A sudden appearance of split peaks is also indica-
tive of a need to replace the guard column.
– Flush with a neat organic solvent to remove
18
b. Storage
Atlantis T3 and dC18 – For period s longer than four days, store the c olumn
in 100% acetonitrile. Do not store columns in buffered eluents. If the
mobile phase contained a buffer salt, flush the column with 10 column
volumes of HPLC grade water (see Table 2 for common column volumes)
and replace with 100% acetonitrile for storage. Failure to perform this
intermediate step could result in precipitation of the buffer salt in the
column when 100% acetonitrile is introduced.
Atlantis HILIC Silic a – For periods longer than four days, store the column
in 95:5 acetonitrile:water. Do not store in buffered eluents. If the mobile
phase contained a buffer salt, flush the column with 10 column volumes of
95:5 acetonitrile:water (refer to Table 2 for a listing of standard column
volumes) prior to storage.
Completely seal column to avoid evaporation and drying out of the bed.
Note: If a column has been run with a formate-containing mobile phase
(e.g., ammonium formate, formic acid, etc.) and is flushed to remove the
buffer, slightly longer equilibration times may be required after the column
is re-installed and run again with a formate-containing mobile phase.
XIV. Troubleshooting
Changes in retention time, resolution, or backpressure are often due to
column contamination (refer to “Column Cleaning, Regenerating and
Storage”). Information on column troubleshooting problems may be
found in HPLC Columns Theory, Technology and Practice, U.D. Neue,
(Wiley-VCH, 1997) or the Waters HPLC Troubleshooting Guide (Litera-