Waters Atlantis T3, DC18 and HILIC Silica Columns User Manual

[ Care and Use ManUal ]
AtLAntIs t3, dc18 And HILIc sILIcA coLuMns
I. IntroductIon
Thank you for choosing an Atlantis® Column. The manufacture of Atlantis
columns begins with ultrapure reagents to control the chemical composition
ISO 9001:2000 certified plant with each step being conducted within nar-
row tolerances. Every column is individually tested and Certificates of Batch
Analysis and a Performance Chromatogram are provided with each column.
Waters recommends the use of Sentry
lifetime and protect the column from contaminants.
guard columns to maximize column
contents
I. IntroductIon
II. connectInG tHe coLuMn or cArtrIdGe to tHe
III. systeM ModIfIcAtIon recoMMendAtIons
HPLc systeM
IV. WAters sMALL PArtIcLe sIze (3 µM) coLuMns
– fAst cHroMAtoGrAPHy
V. coLuMn equILIbrAtIon
VI. coLuMn InstALLAtIon Procedure
VII. coLuMn PerforMAnce VALIdAtIon
VIII. InItIAL coLuMn effIcIency deterMInAtIon
IX. coLuMn usAGe
XII. HILIc GettInG stArted
Atlantis Columns
XIII. coLuMn cLeAnInG, reGenerAtInG And storAGe
XIV. troubLesHootInG
1
[ Care and Use ManUal ]
II. CONNECTING THE COLUMN OR CARTRIDGE TO THE HPLC SYSTEM
a. Column Connection
Handle the column with care. Do not drop or hit the column on a hard
surface as this may disturb the bed and affect its performance.
1. Correct connection of 1/16 inch outer diameter stainless steel
tubing leading to and from the column is essential for high-qual-
ity chromatographic results.
2. An arrow on the column identification label indicates correct
direction of solvent flow.
3. When using standard stainless steel compression screw fittings,
it is important to ensure proper fit of the 1/16 inch outer diameter
stai nless s teel t ubing. W hen t igh teni ng or l oose ning the c omp res -
sion screw, place a 5/16 inch wrench on the compression screw
and a 3/8 inch wrench on the hex head of the column endfitting.
Caution: If one of the wrenches is placed on the column flat during this
process, the endfitting will be loosened and leak. Under-tightening
compression screws or using worn ferrules can lead to solvent
leaking. Care should be taken to check all column connections
for leaks to avoid exposure to solvents and the hazards associ-
ated with such exposure including risks to health and electrical
connections.
4. If a leak occurs between the stainless steel compression screw fit-
ting and the column endfitting, a new compression screw fitting,
tubing and ferrule must be assembled.
b. Cartridge Connection
Handle the cartridge with care. Do not drop or hit the cartridge on a
hard surface as it may disturb the bed and affect its performance. Refer
to Figure 1 for an exploded view of an Atlantis cartridge column with
a Sentry guard column.
1. Unscrew end connectors from the old cartridge. Leave them con-
nected to the inlet and outlet lines of the instrument.
2. Attach new cartridge column between connectors so that the
direction of the flow arrow on the label is pointing in the direc-
tion of mobile phase flow (toward detector).
3. Fingertighten all fittings.
Caution: Under-tightening the connectors can lead to solvent leak-
ing. Care should be taken to check all column connections for leaks
to avoid exposure to solvents and the hazards associated with such
exposure including risks to health and electrical connections.
4. Check for leaks once flow has been initiated. If a leak occurs
between the connector and the column endfitting, the column
may be misaligned in the connector or the Kalrez O-ring must be
replaced in the connector.
It is important to realize that extra column peak broadening can destroy
a successful separation. T he choice of appropriate column connectors
and system tubing is discussed in detail below.
c. Column Connectors and System Tubing Considerations
Due to the absence of an industry standard, various column man
facturers have employed different types of chromatographic column
connectors. The chromatographic performance of the separation can
be negatively affected if the style of the column endfittings does
not match the existing tubing ferrule setting. This section explains
the differences between Waters style and Parker style ferrules and
endfittings (Figure 2). Each endfitting style varies in the required
length of the tubing protruding from the ferrule. The Atlantis column
is equipped with Waters style endfittings, which require a 0.130 inch
ferrule. If a non-Waters style column is presently being used, it is
critical that ferrule depth be reset for optimal performance prior to
installing an Atlantis column. In a proper tubing/column connection
(Figure 3), the tubing touches the bottom of the column endfitting,
with no void between them.
Figure 1: Installation of Atlantis Cartridge Column with Sentry Guard Column
Spacer
Sentry™ Guard
0.25 mm (0.009 inch) Tubing Atlantis® Cartridge Column
Atlantis Columns
Column
C-Clip
2
End Connector
[ Care and Use ManUal ]
Diluted/Distorted Sample Band
0.005 inches
0.020 inches
0.040 inches
Figure 2: Waters and Parker Ferrule Types
Waters Ferrule Setting Parker Ferrule Setting
Figure 3: Proper Tubing/Column Connection
Tubing touches the bottom of the column endfitting, with no void between them.
Attention: A void will occur if tubing with a Parker ferrule is connected
to a Waters style endfitting (Figure 4). This will dramatically reduce the
efficiency of the column and cause peak shape distortion.
Figure 4: Parker Ferrule in a Waters Style Endfitting
There are two ways to fix the problem:
1. Tighten the screw a bit more. The ferrule moves forward, and
reaches the sealing surface. Do not overtighten since this may
break the screw.
2. Cut the tubing, replace the ferrule and make a new connection.
Alternatively, replace the conventional compression screw fitting with
an all-in-on e PE EK
fitting (Waters Part Number PSL613315) that allows
resetting of the ferrule depth. Another approach is to use a Thermo
Hypersil
Keystone, Inc. SLIPFREE® connector to ensure the correct
fit. The fingertight SLIPFREE connectors automatically adjust to fit all
compression screw type fittings without the use of tools (Figure 6).
Figure 6: Single and Double SLIPFREE Connectors
To fix this problem: Cut the end of the tubing with the ferrule, place a
new ferrule on the tubing and make a new connection. Before tighten-
ing the screw, make sure that the tubing bottoms out in the endfitting
of the column.
Conversely, if tubing with a Waters ferrule is connected to a column
with Parker style endfitting, the end of the tubing will bottom out
before the ferrule reaches its proper sealing position. This will leave a
gap and create a leak (Figure 5).
Caution: The connection will leak if a Waters ferrule is connected to a
column with a Parker style endfitting.
Figure 5: Waters Ferrule in a Parker Style Endfitting
Table 1. Waters Part Numbers for SLIPFREE Connectors
SLIPFREE Type and Tubing Internal Diameter Tubing Length 0.005” 0.010” 0.020”
Single 6 cm PSL 618000 PSL 618006 PSL 618012 Single 10 cm PSL 618002 PSL 618008 PSL 618014 Single 20 cm PSL 618004 PSL 618010 PSL 618016 Double 6 cm PSL 618001 PSL 618007 PSL 618013 Double 10 cm PSL 618003 PSL 618009 PSL 618015 Double 20 cm PSL 618005 PSL 618001 PSL 618017
d. Band Spreading Minimization
Internal tubing diameter influences system band spreading and peak
shape. Larger tubing diameters cause excessive peak broadening and
lower sensitivity (Figure 7).
Figure 7: Effect of Connecting Tubing on System
Atlantis Columns
3
Loading...
+ 7 hidden pages