Waters Acquity UPLC User Manual

[ CARE AND USE MANUAL ]
ACQUITY UPLC Peptide CSH C18, 130 Å, 1.7 µm and XP 2.5 µm Columns and ACQUITY PREMIER Peptide CSH C18, 130 Å, 1.7 µm Columns
CONTENTS
I. INTRODUCTION
II. GETTING STARTED
a. Column Connectors b. Column Installation c. Column Equilibration d. Procedure for Using New, Out-of-Box Columns e. eCord Installation f. Column QR Code g. Initial Column Efficiency Determination h. VanGuard Pre-Columns i. Installation Instructions
III. COLUMN USE
a. Sample Preparation b. pH Range c. Solvents d. Pressure e. Temperature
IV. COLUMN CLEANING, REGENERATION, AND STORAGE
a. Cleaning and Regeneration b. Storage
V. eCORD INTELLIGENT CHIP TECHNOLOGY
VI. REPRESENTATIVE TEST CHROMATOGRAPH AND CONDITIONS FOR SEPARATION OF PROTEIN DIGEST
I. INTRODUCTION
Thank you for choosing a Waters™ ACQUITY™ UPLC™ and/ or ACQUITY PREMIER Peptide CSH™ Column. Both feature Waters Charged Surface Hybrid (CSH) Technology which provides excellent peak shape, high efficiency and loading capacity for basic compounds when using acidic, low ionic strength mobile phases. This same particle technology is used in the XSelect™ Peptide CSH HPLC columns, thus enabling seamless transferability between HPLC and UPLC system platforms. The ACQUITY UPLC and ACQUITY PREMIER Peptide CSH C18, 130 Å packing materials are manufactured in a cGMP, ISO 9001 certified manufacturing facility using ultra-pure reagents. Each batch of Peptide CSH C18, 130 Å material is tested chromatographically with acidic, basic, and neutral analytes as part of qualification for use in peptide mapping. The results are held to narrow specification ranges to assure excellent, reproducible performance. ACQUITY UPLC and ACQUITY PREMIER Peptide CSH C18, 130 Å batches are also QC tested with a gradient separation of a tryptic digest of cytochrome c using
0.1% formic acid containing eluents. Finally, every shipped column is individually tested for packed bed efficiency and a Performance Chromatogram and Certificate of Batch Analysis are provided on the eCord™ Intelligent Chip. ACQUITY UPLC and ACQUITY PREMIER Peptide CSH C18, 130 Å Columns will exhibit maximum chromatographic performance and benefits ONLY when used on holistically-designed ACQUITY UPLC Systems since these systems and columns were created and designed to operate together.
VII. ADDITIONAL INFORMATION
VIII. CAUTIONARY NOTE
1
[ CARE AND USE MANUAL ]
II. GETTING STARTED
Each ACQUITY UPLC and ACQUITY PREMIER Peptide CSH C18, 130 Å Column comes with a Certificate of Analysis and a Performance Test Chromatogram embedded within the eCord Intelligent Chip. The Certificate of Analysis is specific to each batch of packing material contained in the ACQUITY UPLC and ACQUITY PREMIER Peptide CSH C18, 130 Å Column and includes the gel batch number, analysis of unbonded particles, analysis of bonded particles, and chromatographic results and conditions. The Performance Test Chromatogram is specific to each individual column and contains such information as: gel batch number, column serial number, USP plate count, USP tailing factor, capacity factor, and chromatographic conditions. These data should be stored for future reference.
a. Column Connectors
The ACQUITY UPLC System utilizes tubing and gold plated compression screws which have been designed to meet stringent tolerance levels and to minimize extra column volumes.
Note: Waters recommends using finger-tight fittings when connecting Waters Peptide CSH C18, 130 Å, XP 2.5 µm column to an HPLC. To ensure a proper connection to your XP 2.5 µm column, Waters recommends p/n: 700003139 to connect to the inlet and p/n: 700004841 to connect to the outlet of your column. This will ensure correct seating of the HPLC system’s connection tubing ferrules to your XP 2.5 µm column and minimize undesired band broadening.
Optimized column inlet tubing (p/n: 430001084) is supplied with the ACQUITY UPLC System. The inject valve end of the tubing is clearly marked with a blue shrink tube marker. Insert the opposite end of the tubing into the ACQUITY UPLC and AQCUITY PREMIER Column and tighten the compression fitting using two 5/16-inch wrenches.
3. When the mobile phase is flowing freely from the column outlet, stop the flow and attach the column outlet to the detector. This prevents entry of air into the detection system and gives more rapid baseline equilibration.
4. Gradually increase the flow rate as described in Step 2.
5. Once a steady backpressure and baseline have been achieved, proceed to the next section.
Note: If mobile-phase additives are present in low concentrations (e.g., ion-pairing reagents), 100 to 200 column volumes may be required for complete equilibration. In addition, mobile phases that contain formate (e.g., ammonium formate, formic acid, etc.) may also require longer initial column equilibration times.
c. Column Equilibration
ACQUITY UPLC and ACQUITY PREMIER Peptide CSH C18, 130 Å Columns are shipped in 100% acetonitrile. It is important to ensure mobile-phase compatibility before changing to a different mobile-phase system. Equilibrate the column with a minimum of 10-column volumes of the mobile phase to be used (refer to Table 1 for a list of column volumes).
Table 1. Empty column volumes in mL (Multiply by 10 for flush solvent volumes)
Column length (mm) Internal diameter 2.1 mm
50 0.2 mL 100 0.4 mL 150 0.5 mL
To avoid precipitating mobile-phase buffers on your column or in your system, flush the column with five-column volumes of a water/organic solvent mixture, using the same or lower solvent content as in the desired buffered mobile phase. (For example, flush the column and system with 60% methanol in water prior to introducing 60% methanol/40% buffer mobile phase.)
For information on the correct column outlet tubing, please refer to the relevant detector section in the ACQUITY UPLC System Operator’s Guide (p/n: 71500082502).
b. Column Installation
1. Purge the pumping system, with HPLC-grade water, of any buffer-containing mobile phases and connect the inlet end of the column to the injector outlet.
2. Flush column with 100% organic mobile phase (methanol or acetonitrile) by setting the pump flow rate to 0.1 mL/min and increase the flow rate to 0.5 mL/min over five minutes.
d. Procedure for Using New, Out-of-Box Columns
Prior to using a new column, it is important to confirm that it produces reproducible chromatography and the desired level of chromatographic resolution. To this end, it is useful to benchmark column performance with a sample that is representative of the intended application. The number of injections necessary to achieve reproducible performance may be dependent on sample characteristics and system type. Method variables like pH, mass load, ionic strength, and ion pairing, could also have impact. The ACQUITY PREMIER Columns have MaxPeak™ High Performance Surfaces that reduce the number of injections necessary to achieve desired performance due to the improved hardware inertness.
2ACQUITY UPLC and ACQUITY PREMIER Peptide CSH Columns
[ CARE AND USE MANUAL ]
e. eCord Installation
The eCord button should be attached to the side of the column heater module. The eCord button is magnetized and does not require specific orientation.
f. Column QR Code
The quick reference (QR) code that is located on the column label provides column-specific information (i.e., the part and serial numbers that are unique identifiers for the column), and its encoding follows a widely adopted industry-standard.
ACQUITY UPLC Column VanGuard Pre-Column
Place wrench here
Ferrule
Collet
Place wrench here
Flow
1. Scan QR code using any device that is capable of
scanning QR codes (i.e., for smart phones and tablets, use the built-in camera app).
2. Be directed to the column’s information hub
on waters.com.
3. Access technical and scientific information for the column
(i.e., certificate of analysis, application notes).
g. Initial Column Efficiency Determination
1. Perform an efficiency test on the column before using it. Waters recommends using a suitable solute mixture, as found in the “Performance Test Chromatogram”, to analyze the column upon receipt.
2. Determine the number of theoretical plates (N) and use this value for periodic comparisons.
3. Repeat the test at predetermined intervals to track column performance over time. Slight variations may be obtained on two different UPLC systems due to the quality of the connections, operating environment, system electronics, reagent quality, column condition, and operator technique.
Figure 1. VanGuard Pre- Column installation diagram.
i. Installation Instructions
1. Remove VanGuard Pre-column from box and shipping tube and remove plastic plug.
2. Orient pre-column so that male end is facing up and carefully remove rubber O-ring that holds collet and ferrule in place during shipping (collet and ferrule are not yet permanently attached).
3. Orient ACQUITY UPLC and ACQUITY PREMIER Peptide CSH C18, 130 Å Column perpendicular to work surface so that column inlet is on the bottom (column outlet on top).
4. From below, insert VanGuard Pre-column into ACQUITY UPLC and ACQUITY PREMIER Peptide CSH C18, 130 Å Column inlet and hand-tighten (collet and ferrule are not yet permanently attached).
5. While pushing the VanGuard Pre-column into the column inlet, turn assembled column and pre-column 180° so that pre-column is now on top.
h. VanGuard Pre-Columns
VanGuard™ Pre-columns are 2.1 mm I.D. x 5 mm length guard column devices designed specifically for use in the ACQUITY UPLC Systems. VanGuard Pre-columns are packed with the same chemistries and frits as our 2.1 mm I.D. ACQUITY UPLC Peptide CSH C18, 130 Å Columns. VanGuard Pre-columns are designed to be attached directly to the inlet side of an ACQUITY UPLC Peptide CSH C18, 130 Å Column.
Note: In order to ensure void-free and leak-free connections, the VanGuard Pre-column is shipped with the collet and ferrule NOT permanently attached. Care must be taken when removing the O-ring that holds these two pieces on the pre-column tubing.
3ACQUITY UPLC and ACQUITY PREMIER Peptide CSH Columns
[ CARE AND USE MANUAL ]
III. COLUMN USE
To ensure the continued high performance of ACQUITY UPLC and ACQUITY PREMIER Peptide CSH C18, 130 Å Columns, follow these guidelines:
a. Sample Preparation
1. Sample must be dissolved in a diluent compatiable with initial strength of mobile phase.
2. Sample must be completely in solution and free of particulates.
3. To remove particulates the sample may be filtered with a
0.2 µm membrane. If the sample is dissolved in a solvent that contained an organic modifier (e.g., acetronitrile, methanol, etc.) ensure that the membrane material does not dissolve in the solvent. Contact the membrane manufacturer with solvent compatibility questions. Alternatively, centrifugation for 20 minutes at 8000 rpm, followed by the transfer of the supernatant liquid to an appropriate vial, could be considered.
b. pH Range
The recommended operating pH range for ACQUITY UPLC and ACQUITY PREMIER Peptide CSH C18, 130 Å Columns is 1 to 11. A listing of commonly used buffers and additives is given in Table 2. Additionally, the column lifetime will vary depending upon the operating temperature, the type and concentration of buffer used. For example, the use of phosphate buffer at pH 8 or above in combination with elevated temperatures will lead to shorter column lifetimes.
c. Solvents
To maintain maximum column performance, use high quality chromat-ography grade solvents. Filter all aqueous buffers prior to use. Pall Gelman Laboratory Acrodisc® filters are recommended. Solvents containing suspended particulate materials will generally clog the outside surface of the inlet distribution frit of the column. This will result in higher operating pressure and poorer performance.
d. Pressure
ACQUITY UPLC and ACQUITY PREMIER Peptide CSH C18, 130 Å Columns can tolerate operating pressures up to 18,000 psi (1241 bar or 124 MPa).
Note: Working at the extremes of pressure, pH and/or temperature will result in shorter column lifetimes.
e. Temperature
Temperatures up to 80 °C are recommended for operating ACQUITY UPLC and ACQUITY PREMIER Peptide CSH C18, 130 Å Columns in order to enhance selectivity, lower solvent viscosity, and increase mass transfer rates. When operating at high pH, lower operating temperatures are recommended for longer column lifetime. Working at high temperatures (e.g, >70 °C) may also result in shorter column lifetimes.
Note: Working at the extremes of temperature, pressure and/or pH will result in shorter column lifetimes.
Important Note:
Waters ACQUITY UPLC Peptide CSH C18, 130 Å material is produced from our ACQUITY UPLC Peptide CSH C18, 130 Å particles that undergo a surface modification by the addition of a low concentration of weakly basic ionizable silanes, followed by C18 bonding and end capping. The optimal surface concentration of the ionizable silane groups is more than an order of magnitude lower than that of the primary C18 bonded phase. The weakly basic silane groups are protonated at a low pH and are neutral at pH greater than 7. Consequently, and while the ACQUITY UPLC Peptide CSH C18, 130 Å particles can tolerate a pH range from 1 to 11, it will not provide frequently desired positive surface charge benefits when used at a pH greater than six.
4ACQUITY UPLC and ACQUITY PREMIER Peptide CSH Columns
[ CARE AND USE MANUAL ]
Table 2. Buffer recommendations for using ACQUITY UPLC and ACQUITY PREMIER Peptide CSH C18, 130 Å Columns from pH 1 to 11
Used for
Mass Spec
Comments
Ion pair additive, can suppress MS signal, used in the 0.02–0.1% range.
Maximum buffering obtained when used with ammonium acetate salt. Used in 0.1–1.0% range.
Maximum buffering obtained when used with ammonium formate salt. Used in 0.1–1.0% range.
Used in the 1–10 mM range.
Note that sodium or potassium salts are not volatile. Used in the 1–10 mM range.
Note that sodium or potassium salts are not volatile.
Above pH 7, reduce temperature/concentration and use a guard column to maximize lifetime.
Keep concentration below 10 mM and temperatures below 30 °C.
Used in the 5–10 mM range (for MS work keep source >150 °C ). Adjust pH with ammonium hydroxide or acetic acid. Good buffering capacity at pH 10.
Note: Use ammonium bicarbonate (NH4HCO3), not ammonium carbonate ((NH4)2CO3).
Zwitterionic buffer, compatible with acetonitrile, used in the 1–10 mM range. Low odor.
Zwitterionic buffer, can give longer lifetimes than borate buffer.
Zwitterionic buffer, compatible with acetonitrile, used in the 1–10 mM range. Low odor.
Additive/Buffer pK
a
range
Buffer
Volatility
(±1 pH unit)
TFA 0.3 Volatile Yes
Acetic acid 4.76 Volatile Yes
Formic acid 3.75 Volatile Yes
Acetate (NH4CH2COOH)
4.76 3.76–5.76 Volatile Yes
Formate (NH4COOH) 3.75 2.75–4.75 Volatile Yes
Phosphate 1 2.15 1.15–3.15 Non-volatile No Traditional low pH buffer, good UV transparency.
Phosphate 2 7.2 6.20–8.20 Non-volatile No
4–Methylmorpholine ~8.4 7.4–9.4 Volatile Yes Generally used at 10 mM or less.
Ammonia (NH4OH) 9.2 8.2–10.2 Volatile Yes
)
3
+
)
6.8–11.3 Volatile Yes
4
Ammonium Bicarbonate
10.3 (HCO
9.2 (NH
6.3 (H2CO3)
Ammonium (Acetate) 9.2 8.2–10.2 Volatile Yes Used in the 1–10 mM range. Ammonium (Formate) 9.2 8.2–10.2 Volatile Yes Used in the 1–10 mM range.
CAPSO 9.7 8.7–10.7 Non-volatile No
Glycine 2.4, 9.8 8.8–10.8 Non-volatile No
1–Methylpiperidine
10.2 9.3–11.3 Volatile Yes Used in the 1–10 mM range.
CAPS 10.4 9.5–11.5 Non-volatile No
Note: Working at the extremes of pH, temperature, and/or pressure will result in shorter column lifetimes.
5ACQUITY UPLC and ACQUITY PREMIER Peptide CSH Columns
[ CARE AND USE MANUAL ]
IV. COLUMN CLEANING, REGENERATION, AND STORAGE
a. Cleaning and Regeneration
Changes in peak shape, peak splitting, shoulders on the peak, shifts in retention, change in resolution or increasing backpressure may indicate contamination of the column. Flushing with a neat organic solvent, taking care not to precipitate buffers, is usually sufficient to remove the contaminant. If the flushing procedure does not solve the problem, purge the column using the following cleaning and regeneration procedures.
Use the cleaning routine that matches the properties of the samples and/or what you believe is contaminating the column (see Table 3). Flush columns with 20-column volumes of HPLC-grade solvents. Increasing mobile-phase temperature to 35–55 °C increases cleaning efficiency. If the column performance is poor after regenerating and cleaning, call your local Waters office for additional support.
Table 3. Column Cleaning Sequence
Polar Samples Proteinaceous Samples
Option 1: Inject repeated 100 µL aliquots
Water
Methanol
Isopropanol
Note: To avoid potentially damaging precipitation within your column (e.g., if your separation eluent contains phosphate buffer), be certain to flush column with 5 to 10 column volumes of water BEFORE using suggested organic eluent column wash procedures.
of dimethylsulfoxide (DMSO) using a reduced flow rate delivering 50% Eluent A and 50% Eluent B
Option 2: Gradient of 10% to 90% B where: A = 0.1% trifluoroacetic acid (TFA) in water, B = 0.1% trifluoroacetic acid (TFA) in acetonitrile (CH3CN)
Option 3: Flush column with 7 M guanidine hydrochloride, or 7 M urea
step could result in precipitation of the buffer salt in the column when 100% acetonitrile is introduced. Completely seal column to avoid evaporation and drying out of the bed.
Note: If a column has been run with a mobile phase that contains formate (e.g., ammonium formate, formic acid, etc.) and is then flushed with 100% acetonitrile, slightly longer equilibration times may be necessary when the column is re-installed and run again with a formate-containing mobile phase.
V. eCORD
a. Introduction
The eCord Intelligent Chip provides the history of a column’s performance throughout its lifetime. The eCord is permanently attached to the column to assure that the column’s performance history is maintained in the event that the column is moved from one instrument to another.
eCord Intelligent Chip
Figure 1. eCord Intelligent Chip.
At the time of manufacture, tracking, and quality control information will be downloaded to the eCord. Storing this information on the chip will eliminate the need for a paper Certificate of Analysis. Once the user installs the column, the software will automatically download key parameters into a column history file stored on the chip. The eCord provides a solution to easily track the history of column usage.
b. Storage
For periods longer than four days at room temperature, store the column in 100% acetonitrile. For elevated temperature applications, store immediately after use in 100% acetonitrile for the best column lifetime. Do not store columns in buffered eluents. If the mobile phase contained a buffer salt, flush the column with 10-column volumes of HPLC-grade water (see Table 1 for common column volumes) and replace with 100% acetonitrile for storage. Failure to perform this intermediate
eCord inserted into side of column heater
Figure 2. eCord inserted into side of column heater.
6ACQUITY UPLC and ACQUITY PREMIER Peptide CSH Columns
[ CARE AND USE MANUAL ]
b. Installation
Install the column into the column heater. Plug the eCord into the side of the column heater. Once the eCord is inserted into the column heater, the identification and overall column usage information will be available in Empower™ and MassLynx™ Software allowing the user to access column information on their desktop.
c. Manufacturing Information
Figure 3. The eCord chip provides the user with an overview of the bulk material QC test results.
Figure 4. The eCord chip provides the user with QC test conditions and results on the column run by the manufacturer. The information includes mobile phases, running conditions, and analy tes used to test the columns. In addition, the QC results and acceptance is placed onto the column.
d. Customer Use Information
Figure 5. An example of column use information provided by the eCord chip.
The eCord will automatically capture column use data. The top of the screen identifies the column including chemistry type, column dimensions, and serial number. The overall column usage information includes the total number of samples, total number of injections, total sample sets, date of first injection, date of last injection, maximum pressure, and temperature. The information also details the column history by sample set including date started, sample set name, user name, system name, number of injections in the sample set, number of samples in the sample set, maximum pressure, and temperature in the sample set and if the column met basic system suitability requirements.
7ACQUITY UPLC and ACQUITY PREMIER Peptide CSH Columns
[ CARE AND USE MANUAL ]
1.20
1.10
1.00
0.90
0.80
0.70
0.60
0.50
0.40
0.30
0.20
0.10
0.00
2345678910 11 12
min.
VI. REPRESENTATIVE TEST CHROMATOGRAPH AND CONDITIONS FOR SEPARATION OF A CYTOCHROME c TRYPTIC DIGEST
Chromatographic Conditions:
Column: ACQUITY UPLC Peptide CSH C18,
130 Å, 1.7 µm, 2.1 x 50 mm
(p/n: 186006936) Mobile phase A: 0.1% formic acid in 100% water Mobile phase B: 0.085% formic acid in 75% acetonitrile
reconstituted w/ 200 µL of MPA,
5.0 µL injection
Sample: Tryptic digest of bovine cytochrome c
(p/n: 186006371) Flow rate: 0.2 mL/min Gradient: 1–21% B in 5 min; 21–31.4% B in 2.5 min;
31.4–95% B in 2.5 min; 95% B for 1.5 min;
1% B for 2.5 min Temp.: 40 °C UV detection: 214 nm
Peak identification
T1 N-AcGDVEK T13-T14 KYIPGTK T14 YIPGTK T4 IFVQ K T9-T10 KTGQAPGFSYTDANK T10 TGQAPGFSYTDANK T8 TGPNLHGLFGR T15 MIFAGIK T19C EDLIAY T19 EDLIAYLK T12-T13 GITWGEETLMEYLENPKK T12 GITWGEETLMEYLENPK T5 CAQCHTVEK (heme attached)
Retention time (min) Result
T19C 6 .71
Retention time difference (min) Result
T4 and T14 0.30 T10 and T9-T10 0.77 T19 and T19C 0.47 T12 and T12-T13 0.28
VII. ADDITIONAL INFORMATION
Tips for maximizing ACQUITY UPLC and ACQUITY PREMIER Peptide CSH C18, 130 Å Column lifetimes.
1. To maximize ACQUITY UPLC and ACQUITY PREMIER Peptide CSH C18, 130Å Column lifetime, pay close attention to:
■
Water quality (including water purification system)
■
Solvent quality
■
Mobile-phase preparation, storage, and age
■
Sample, buffer, and mobile-phase solubilities
■
Sample quality and preparation
2. When problems arise, often only one improper practice must be changed.
3. Always remember to:
■
Use in-line filter unit or, preferably, a VanGuard Pre-column.
■
Discourage bacterial growth by minimizing the use of 100% aqueous mobile phases where possible.
■
Change aqueous mobile phase every 24–48 hours (if 100% aqueous mobile phase use is required).
■
Discard old 100% aqueous mobile phases every 24–48 hours to discourage bacterial growth.
8ACQUITY UPLC and ACQUITY PREMIER Peptide CSH Columns
[ CARE AND USE MANUAL ]
[ CARE AND USE MANUAL ]
■
Add 5%–10% organic modifier to mobile phase A and adjust gradient profile.
■
Filter aqueous portions of mobile phase through
0.2 µm filter.
■
Maintain your water purification system so that it is in good working order.
■
Only use ultra pure water (18 megohm-cm) water and highest quality solvents possible. HPLC grade water is not UPLC-grade water.
■
Consider sample preparation (e.g., solid-phase extraction, filtration, etc).
4. Avoid (where possible):
■
100% aqueous mobile phases (if possible).
■
HPLC-grade bottled water.
■
“Topping off” or adding “new” mobile phase to “old” mobile phase.
■
Old aqueous mobile phases. Remember to rinse bottles thoroughly and prepare fresh every 24–48 hours.
■
Using phosphate salt buffer in combination with high ACN concentrations (e.g., >70%) due to precipitation.
5. Don’t: assume a “bad” column is the culprit when high back pressure or split peaks are observed. Investigate cause of column failure:
■
Backpressure
■
Mobile phase(s), bacteria, precipitation, and/or samples
■
Peak splitting
■
Sample quality
■
Injection solvent strength
6. Remember: UPLC flow rates are often much lower and, therefore, mobile phases last much longer (only prepare what you need or store excess refrigerated).
7. Mobile-phase related questions to ask:
■
Am I using 100% aqueous mobile phases? Am I able to add a small amount of organic modifier to my mobile phase A?
■
Do I filter my aqueous mobile phases through
0.2 µm filters?
■
How old is my mobile phase? Do I label the bottle with preparation date?
■
Do I “top off” or do I prepare fresh mobile phases every 24–48 hours?
■
What is the quality of my water? Has the quality recently changed? How is my water purification system working? When was it last serviced?
■
Am I working with a pH 7 phosphate buffer (which is VERY susceptible to bacterial growth)?
8. Sample-related questions to ask:
■
If I inject neat standards prepared in mobile phase, do I observe these problems?
■
If I prepare my standards in water and prepare them like samples (e.g., SPE, filtration, etc.), do I still observe these problems?
■
Has the quality of my samples changed over time?
VIII. CAUTIONARY NOTE
Depending on user’s application, these products may be classified as hazardous following their use and as such are intended to be used by professional laboratory personnel trained in the competent handling of such materials. Responsibility for the safe use and disposal of products rest entirely with the purchaser and user. The Safety Data Sheet (SDS) for this product is available at www.waters.com/sds.
Waters, The Science of What’s Possible, ACQUITY, UPLC, XSelect, Empower, MassLynx, CSH, eCord, and VanGuard are trademarks of Waters Corporation. Acrodisc is a trademark of Pall Life Sciences. All other trademarks are the proper ty of their respective owners.
©2020 Waters Corporation October 2020 720004544EN Rev D IH-PDF
Waters Corporation
34 Maple Street Milford, MA 01757 U.S.A. T: 1 508 478 2000 F: 1 508 872 1990 www.waters.com
Loading...