hydroxide, and organic amines (such as diethylamine and
triethylamine) and ammoniated methanol. Typical concentrations
of 20 mM or 0.2% can be used safely. Consider the volatility,
solubility and detector compatibility when choosing an
appropriate additive. Additives tend to improve peak shape and
control the retention characteristics of analytes, but can also
impart different selectivities. It is recommended to flush and
remove additives and salts from the column before placing the
column into storage.
d. Pressure
ACQUITY UPC2 Column hardware and packing materials are
packed and tested to pressures of up to 6,000 psi (414 bar or
41 Mpa).
3. If the sample is not dissolved in the mobile-phase modifier,
ensure that the sample, solvent and mobile phases are miscible
in order to avoid sample precipitation. Filter sample with
0.2 μm membranes to remove particulates. If the sample is
ACQUITY UPC2 Torus Columns Care & Use
4
[ CARE AND USE MANUAL ]
Diluted/Distorted Sample Band
Dimension
100% CO
Predicted Max Flow Rate
(mL/min)
2
Predicted Max Pressure
(psi)
Predicted Max Flow Rate
(mL/min)
60/40 CO2/MeOH
Predicted Max Pressure
2.1 x 50 mm3.4555812.006 011
2.1 x 75 mm2.6559701.355938
2.1 x 100 mm2.0559691.056060
2.1 x 150 mm1.4 059510.706002
3.0 x 50 mm4.0041793.605946
3.0 x 75 mm3.7548972.605959
3.0 x 100 mm3.4554992.005919
3.0 x 150 mm2.7059771.406023
Table 2. Useable Flow Rate Ranges and Expected ACQUITY UPC2 System Pressures.
e. Useable Flow-Rate Ranges and Expected ACQUITY UPC2
System Pressures
Table 2 estimates maximum flow rate and system pressure which
can be achieved without over-pressuring the ACQUITY UPC
System. The ACQUITY UPC2 System has a maximum flow rate of
4 mL/minute and maximum pressure of 6000 psi (~400 Bar).
2
IV. TROUBLESHOOTING
1. One of the most common problems with regards to columns
is incorrect or insufficient priming of the co-solvent/modifier
pump. If no peaks are observed after an injection or unusually
long retention times are observed, check the priming of the
co-solvent/modifier pump first.
(psi)
The predicted flow rates and expected backpressure information
is based on the ACQUITY UPC2 System BPR (Backpressure
Regulator) set at 1800 psi (~120 Bar).
f. Band Spreading Minimization
The ACQUITY UPC2 System has been designed to minimize band
spreading. Deviation from Waters specified tubing could result
in deterioration of chromatographic performance due to band
spreading induced by inappropriate tubing i.d. Figure 2 shows the
influence of tubing internal diameter on system band spreading
and peak shape. As can be seen, the larger tubing diameter causes
excessive peak broadening and lower sensitivity.
0.005 inches
0.020 inches
0.040 inches
2. If you see frosting on the column at the inlet or outlet, tighten
the compression screw on that side. If tightening doesn’t work,
depressurize the system and the column, and then replace the
fitting that is not sealing correctly.
3. If you continue to see frosting on the column, turn off the CO2
and vent the system. Allow the column to depressurize fully
before disconnecting the inlet or outlet. Please contact your
Waters representative for additional support.
V. COLUMN CLEANING, EFFECTS OF ADDITIVES,
AGING, AND STORAGE
Changes in peak shape, peak splitting, shoulders on the
peak, shifts in retention, change in resolution or increasing
backpressure may indicate contamination of the column.
Flushing with high concentrations of organic solvent, taking care
not to precipitate buffers, is usually sufficient to remove the
contaminant. If the flushing procedure does not solve the
problem, purge the column using the following cleaning and
regeneration procedures.
Figure 2: Effect of connecting tubing on system.
ACQUITY UPC2 Torus Columns Care & Use
5
[ CARE AND USE MANUAL ]
a. Cleaning
Use the cleaning routine that matches the properties of the
samples and/or what you believe is contaminating the column
(see Table 4). Placing the column on an ACQUITY UPLC® System
(or HPLC system) will enable the use of a wide range of solvents
and water to improve cleaning. Flush columns with 20 column
volumes of solvent. Increasing column temperature increases
cleaning efficiency. If the column performance is poor after
cleaning and regenerating, call your local Waters office for
additional support.
Note: Using an ACQUITY UPC2 System is not recommended for
the cleaning and regeneration procedures because recommended
solvents and high concentrations of water are immiscible with
supercritical CO2. Rather, the use of an ACQUITY UPLC System
is recommended for cleaning and regeneration.
Polar SamplesNon-Polar Samples
1. Water (could also be
a mixture of acetonitrile
and water)
2. Methanol2. Tetrahydrofuran
3. Tetrahydrofuran3. Dichloromethane
4. Methanol4. Hexane
5. Water
6. Mobile Phase6. Mobile Phase
Table 3: Column Cleaning Sequence.
1. Isopropanol (or an
appropriate water/
Isopropanol mixture)
5. Isopropanol (followed
by an appropriate water/
isopropanol mixture)
b. Effects of Additives
Use of co-solvent additives, such as ammonia or organic
amines, can be used successfully with all Torus columns.
Consider the volatility and detector compatibility when choosing
an appropriate additive. Additives tend to improve peak shape
and control the retention characteristics of analytes, but can
also impart different selectivities. An ammonia additive works
particularly well on Torus DIOL and Torus 1-AA to improve peak
shapes and reduce retention variability. Ion pairing additives
should be used with caution on Torus phases, particularly
those with exposed amines (e.g. Torus 2-PIC and Torus DEA).
Counter ions, such as trifluoroacetate, can bind very strongly or
irreversibly with the stationary phase. It is recommended that
dedicated columns be used with particular ion pairing additives.
c. Aging
Research at the Waters Corporation confirms that there may be
changes in retention with SFC columns due to packing material
surface changes. This is commonly referred to by SFC industry
experts as “Aging or retention drift.” Mobile-phase additives
and methanol can cause changes to the silica or hybrid surface
resulting in retention shifts. When working with a new column,
the column can be flushed with CO2/methanol to eliminate
or reduce further shifts in retention. Storage in 100% CO2 is
recommended to ensure the column retention and selectivity
will be the same from the end of one test to the beginning of
another. The 100% CO2 will halt the aging process and results
in no additional retention shifts.
Retest the column after using the cleaning procedure to
determine if the specific problem has been fixed. If so, continue
using the column, avoiding samples and solvents that may clog
the column inlet.
Torus columns were designed and developed to not show
the characteristic retention and selectivity changes related
to aging. When exposed to CO2/Methanol mobile phases,
these columns show little or no shifts in retention over extended
periods of time.
d. Storage
Purging out any additives with a CO2/Methanol mobile phase
is recommended and storing columns in pure CO2.
ACQUITY UPC2 Torus Columns Care & Use
6
[ CARE AND USE MANUAL ]
VI. INT RODUCING ECORD INTELLIGENT
CHIP TECHNOLOGY
a. Introduction
The eCord Intelligent Chip will provide the history of a column’s
performance throughout its lifetime. The eCord is permanently
attached to the column to assure that the column’s performance
history is maintained in the event that the column is moved from
one instrument to another.
Figure 3. eCord Intelligent Chip.
At the time of manufacture, tracking and quality control
information will be downloaded to the eCord. Storing this
information on the chip will eliminate the need for a paper
Certificate of Analysis.
c. Manufacturing Information
Once the user installs the column, the software will automatically
download key parameters into a column history file stored on
the chip. In this manual, we explain how the eCord will provide
a solution for easily tracking the history of the columns, reduce
the frustration of paperwork trails, and give customers the
reassurance that a well-performing column is installed onto
their instruments.
b. Installation
Install the column into the ACQUIT Y® Column Manager. Plug
the eCord into the side of the column heater. Once the eCord is
inserted into the column heater the identification and overall
column usage information will be available allowing the user to
access column information on their desktop.
eCord Fob
Figure 5: Manufacturing results stored on an eCord.
Figure 4. Installing the eCord Intelligent Chip.
ACQUITY UPC2 Torus Columns Care & Use
7
[ CARE AND USE MANUAL ]
d. Column Use Information
The eCord Chip provides the customer with column use data,
column dimensions and serial number. The overall column
usage information includes the total number of samples, total
number of injections, total sample sets, date of first injection,
date of last injection, maximum pressure and temperature.
The information also details the column history by sample set
including date started, sample set name, user name, system
name, number of injections in the sample set, number of
samples in the sample set, maximum pressure and temperature
in the sample set and if the column met basic system suitability
requirements. Up to 50 sample sets can be stored on the eCord
Chip. In addition, the eCord provides two-way communications
between the eCord Chip and Empower® Software.
VII. ACQUITY UPC2 COLUMN FAMILY
The ACQUITY UPC2 Torus Columns Care & Use Manual is one of
three manuals for the ACQUIT Y UPC2 Column family. The other
two Care & Use Manual are available at www.waters.com/chemcu
ACQUITY UPC2 Trefoil™ Columns Care & Use Manual
- 720004828EN
ACQUITY UPC2 BEH, CSH, and HSS Columns Care & Use Manual
- 720004349EN
Figure 6. Column use information.
Waters, The Science of W hat’s Possible, ACQUITY U PC,2 UPC,2 Sep-Pak, Oasis, Waters Quality Parts, ACQUITY, and Empower
are registered trademarks of Waters Corporation. Trefoil, eCord, UltraPerformance Convergence Chromatograp hy, VanGuard,
and Torus are trademarks of Waters Corporation. All other trademarks are the property of their respective owners.