particulate materials will generally clog the outside surface
of the inlet distribution frit of the column. This will result in
higher operating pressure, reduced column lifetime, and
compromised performance.
ACQUITY UPC2 Torus Columns Care & Use
4
[ CARE AND USE MANUAL ]
ACQUITY UPC2 Columns, 1.7 µm
100% CO
Dimension
2.1 x 50 mm3.4555812.006 011
2.1 x 75 mm2.6559701.355938
2.1 x 100 mm2.0559691.056060
2.1 x 150 mm1.4 059510.706002
3.0 x 50 mm4.0041793.605946
3.0 x 75 mm3.7548972.605959
3.0 x 100 mm3.4554992.005919
3.0 x 150 mm2.7059771.406023
Predicted Max Flow Rate
(mL/min)
2
Predicted Max Pressure
(psi)
Predicted Max Flow Rate
(mL/min)
60/40 CO2/MeOH
Predicted Max Pressure
ACQUITY UPC2 Columns, 3.5 µm
100% CO
Dimension
2.1 x 50 mm4.0035494.004963
2.1 x 75 mm4.0040963.755973
2.1 x 100 mm3.8545303.005979
2.1 x 150 mm3.5053862.105946
3.0 x 50 mm4.0030424.003703
3.0 x 75 mm4.0032754.004283
3.0 x 100 mm4.0035284.004906
3.0 x 150 mm4.0040633.805965
Predicted Max Flow Rate
(mL/min)
2
Predicted Max Pressure
(psi)
Predicted Max Flow Rate
(mL/min)
60/40 CO2/MeOH
Predicted Max Pressure
(psi)
(psi)
Table 3. Useable Flow Rate Ranges and Expected ACQUITY UPC2 System Pressures.
IV. TROUBLESHOOTING
1. One of the most common problems with regards to columns
is incorrect or insufficient priming of the co-solvent/modifier
pump. If no peaks are observed after an injection or unusually
long retention times are observed, check the priming of the
co-solvent/modifier pump first.
2. If you see frosting on the column at the inlet or outlet, tighten
the compression screw on that side. If tightening doesn’t work,
depressurize the system and the column, and then replace the
fitting that is not sealing correctly.
3. If you continue to see frosting on the column, turn off the CO2
and vent the system. Allow the column to depressurize fully
before disconnecting the inlet or outlet. Please contact your
Waters representative for additional support.
.
V. COLUMN CLEANING, AGING, REGENERATING,
AND STORAGE
Changes in peak shape, peak splitting, shoulders on the
peak, shifts in retention, change in resolution or increasing
backpressure may indicate contamination of the column.
Flushing with high concentrations of organic solvent, taking
care not to precipitate buffers, is usually sufficient to remove
the contaminant. If the flushing procedure does not solve the
problem, purge the column using the following cleaning and
regeneration procedures.
a. Cleaning
Use the cleaning routine that matches the properties of the
samples and/or what you believe is contaminating the column
(see Table 4). Placing the column on an ACQUITY UPLC® System
(or HPLC system) will enable the use of a wide range of solvents
and water to improve cleaning. Flush columns with 20 column
ACQUITY UPC2 Torus Columns Care & Use
5
[ CARE AND USE MANUAL ]
volumes of solvent. Increasing column temperature increases
cleaning efficiency. If the column performance is poor after
cleaning and regenerating, call your local Waters office for
additional support.
Note: Using an ACQUITY UPC2 System is not recommended for
the cleaning and regeneration procedures because recommended
solvents and high concentrations of water are immiscible with
supercritical CO2. Rather, the use of an ACQUITY UPLC System is
recommended for cleaning and regeneration.
Polar SamplesNon-Polar Samples
1. Water (could also be
a mixture of acetonitrile
and water)
2. Methanol2. Tetrahydrofuran
3. Tetrahydrofuran3. Dichloromethane
4. Methanol4. Hexane
5. Water
6. Mobile Phase6. Mobile Phase
Table 4: Column Cleaning Sequence.
1. Isopropanol (or an
appropriate water/
Isopropanol mixture)
5. Isopropanol (followed
by an appropriate water/
isopropanol mixture)
c. Regeneration
If you have noticed a significant change in retention or selectivity,
an additional regeneration process may restore your column’s
performance. There are two main sources of retention or
selectivity change for ACQUITY UPC2 Columns, assuming no
damage has been done to the column bed. Mobile-phase additives
and methanol can cause changes to the silica or hybrid surface.
The effects can be gradual or expeditious and may be permanent.
Columns showing peak shape, retention or selectivity changes
are best cleaned and regenerated on an ACQUIT Y UPLC System.
Before placing the column to be regenerated on a LC system,
it should be free of mobile-phase co-solvents and additives.
To do this, flow 100% CO2 through the column (3.0 x 100 mm) at
1 mL/min for five minutes (40 °C and 2000 psi at the ABPR).
For other column dimensions, the flow rate should be recalculated
appropriately to avoid overpressuring the system.
Based on the stationary phase to be regenerated, different rinsing
agents should be used as to not damage the packing material.
High-pH rinsing agents (pH > 8) should be avoided for cleaning
silica-based materials. Low-pH rinsing agents (pH < 1) should be
avoided for cleaning silica-based materials with ligand bondings.
Retest the column after using the cleaning procedure to
determine if the specific problem has been fixed. If so, continue
using the column, avoiding samples and solvents that may clog
the column inlet.
b. Aging
Research at the Waters Corporation confirms that there may be
changes in retention with SFC columns due to packing material
surface changes. This is commonly referred to by SFC industry
experts as “Aging.” Mobile-phase additives and methanol can
cause changes to the silica or hybrid surface resulting in retention
shifts. When working with a new column, the column can be
flushed with CO2/methanol to eliminate or reduce further shifts
in retention. Storage in 100% CO2 is recommended to ensure the
column retention and selectivity will be the same from the end of
one test to the beginning of another. The 100% CO2 will halt the
aging process and results in no additional retention shifts.
Step 1. Unbonded and bonded silica-based materials
should be rinsed with 200+ column volumes of pure water.
All hybrid-based materials (e.g., BEH and CSH) can be
rinsed with either 200+ column volumes of pure water or
30+ column volumes of 0.1% N,N-Dimethylbutylamine
(or similar tertiary amine) in water.
Step 2. All columns should be rinsed with an additional
10-column volumes of 90/10 IPA/water to remove the mobile-
phase additives.
Step 3. All columns should be rinsed with 10-column
volumes of isopropanol to remove any residual water before
returning the column to an ACQUIT Y UPC2 System.
After regeneration, please reinstall and equilibrate the ACQUIT Y
UPC2 Column on the ACQUIT Y UPC2 System. Retest the column
after using the cleaning procedure to determine if the specific
problem has been fixed.
ACQUITY UPC2 Torus Columns Care & Use
6
[ CARE AND USE MANUAL ]
d. Storage
Purging and storing a column in pure CO2 is recommended.
This will eliminate or minimize retention changes of the
column during storage. Storing the column in methanol is
not recommended.
VI. INT RODUCING ECORD INTELLIGENT
CHIP TECHNOLOGY
a. Introduction
The eCord Intelligent Chip will provide the history of a column’s
performance throughout its lifetime. The eCord is permanently
attached to the column to assure that the column’s performance
history is maintained in the event that the column is moved from
one instrument to another.
eCord Fob
Figure 4. Installing the eCord Intelligent Chip.
c. Manufacturing Information
Figure 3. eCord Intelligent Chip.
At the time of manufacture, tracking and quality control
information will be downloaded to the eCord. Storing this
information on the chip will eliminate the need for a paper
Certificate of Analysis.
Once the user installs the column, the software will automatically
download key parameters into a column history file stored on
the chip. In this manual, we explain how the eCord will provide
a solution for easily tracking the history of the columns, reduce
the frustration of paperwork trails, and give customers the
reassurance that a well-performing column is installed onto
their instruments.
b. Installation
Install the column into the ACQUIT Y® Column Manager. Plug
the eCord into the side of the column heater. Once the eCord is
inserted into the column heater the identification and overall
column usage information will be available allowing the user to
access column information on their desktop.
Figure 5: Manufacturing results stored on an eCord.
ACQUITY UPC2 Torus Columns Care & Use
7
[ CARE AND USE MANUAL ]
d. Column Use Information
The eCord Chip provides the customer with column use data,
column dimensions and serial number. The overall column
usage information includes the total number of samples, total
number of injections, total sample sets, date of first injection,
date of last injection, maximum pressure and temperature.
The information also details the column history by sample set
including date started, sample set name, user name, system
name, number of injections in the sample set, number of
samples in the sample set, maximum pressure and temperature
in the sample set and if the column met basic system suitability
requirements. Up to 50 sample sets can be stored on the eCord
Chip. In addition, the eCord provides two-way communications
between the eCord Chip and Empower® Software.
VII. ACQUITY UPC2 COLUMN FAMILY
UPC2 BEH, CSH, and HSS Columns Care & Use Manual is one of
three manuals for the ACQUIT Y UPC2 Column family. The other
two Care & Use Manual are available at www.waters.com/CU
ACQUITY UPC2 Trefoil™ Columns Care & Use Manual
- 720004828EN
ACQUITY UPC2 Torus™ Columns Care & Use Manual
- 720005203EN
Figure 6. Column use information.
Waters, The Science of W hat’s Possible, ACQUITY U PC,2 Sep-Pak, Oasis, Waters Quality Parts, ACQUITY,
ACQUITY U PLC, and Empower are registered trademarks of Waters Corporation. Trefoil, eCord, VanGuard, and
Torus are trademarks of Waters Corporation. All other trademarks are the property of their respective owners.