[ Care and Use ManUal ]
ACCQ•Fluor reagent kIt
I. IntroduCtIon
The Waters AccQ•Fluor™ Reagent Kit contains five sets of the deriva-
tizing reagents used to perform the AccQ•Tag™ Amino Acid Analysis
Method. A set consists of one vial each of:
– AccQ•Fluor Borate Buffer - Vial 1. You add the buffer to the sam-
ples to generate the proper pH environment for the derivatizatlon.
– AccQ•Fluor Reagent Powder - Vial 2A. The reagent powder Is the
6-aminoquinolyl-N-hydroxysuccinimidyI carbamate (AQC) derivatiz-
ing reagent. It Is shipped dry for maximum stability.
– AccQ•Fluor Reagent Diluent - Vial 2B. Use the diluent(acetonitrile)
to reconstitute the reagent for derivatization.
The Waters AccQ•Fluor Reagent Kit (Figure 1) contains sufficient
reagent for up to 250 analyses.
Contents
I. IntroduCtIon
II. Waters accQ•tag Method overvIeW
III. reagent PurIty
Iv. storage
v. reconstItutIng accQ•Fluor reagent
vI. usIng Waters accQ•Fluor reagent
vII. derIvatIzIng the calIbratIon standard
vIII. reconstItutIng the hydrolyzed saMPle
IX. derIvatIzIng the saMPle
X. storIng derIvatIves
Figure 1: AccQ•Fluor Reagent Kit
AccQ•Fluor Reagent Kit 1
XI. sPecIFIcatIons
XII. technIcal lIterature
XIII. orderIng InForMatIon

[ Care and Use ManUal ]
II. Waters accQ•tag Method overvIeW
The AccQ•Tag Method is a precolumn derivatization and analysis
technique for peptide and protein hydrolysate amino acids. To
perform the AccQ-Tag Method:
1. Use Waters AccQ•Fluor Reagent (US Patent 5,296,599
and European Patent EP 0 533 200 B1) to derivatize the
amino acids.
2. Separate the derivatives by reversed-phase HPLC.
3. Quantitate the derivatives by fluorescence detection.
III. reagent PurIty
The Waters AccQ•Fluor Reagent Kit is manufactured in Waters’- ISO
9001-2006 certified facilities under strict guidelines to ensure the
lowest possible amino acid background. To retain this level of purity,
handle all reagents carefully.
6. Heat Vial 2A on top of heating block, vortex occasionally, until
the powder dissolves. Do not heat longer than 10 minutes.
Reconstituted AccQ•Fluor Reagent is approximately 10 mM
AQC in acetontrile.
Note: You can store the reconstituted AccQ•Tag reagent in a desiccator
at room temperature for up to one week. The AccQ•Tag reagent reacts
with atmospheric moisture so seal the container tightly when not in
use and DO NOT refrigerate.
Reconstituted AccQ•Fluor Reagent is approximately 10 mM AQC in
acetontrile.
IV. storage
If unopened, the AccQ•Fluor Reagent Kit may be stored at room
temperature for up to one year.
v. reconstItutIng accQ•Fluor reagent
To reconstitute the AccQ•Fluor Reagent (Figure 2):
1. Preheat a heating block to 55 °C.
2. Tap Vial 2A lightly before opening to ensure all AccQ•Fluor
Reagent Powder is at the bottom of the vial.
3. Rinse a clean micropipette by drawing and discarding 1 ml of
AccQ•Fluor Reagent Diluent from Vial 2B.
WARNING: Acetonitrile is flammable and toxic. Refer to the Material
Safety Data Sheet.
4. Transfer 1.0 ml of AccQ•Fluor Reagent Diluent From Vial 2B to
the AccQ•Fluor Reagent Powder In Vial 2A.
Figure 2: Reconstituting AccQ•fluor Reagent
storage
Store reconstituted AccQ•Fluor Reagent at room temperature in a
desiccator for up to one week.
Note: AccQ•Fluor reagent reacts with atmospheric moisture. Seal the
container tightly when not in use.
5. Cap Vial 2A lightly and vortex for 10 seconds.
AccQ•Fluor Reagent Kit 2

[ Care and Use ManUal ]
vI. usIng Waters accQ•Fluor reagent
Note: The Waters AccQ•Fluor Reagent Kit is not for In vitro diagnostic
use. The Waters AccQ•Fluor Reagent Kit is for research use only.
Use reconstituted AccQ•Fluor Reagent to convert samples and
standards to highly stable fluorescent
derivatives. T his section describes the procedures for:
– Derivatizing a calibration standard
– Derivatizing a hydrolyzed sample
Note: Pyrolyze all glassware al 500 °C for at least 4 hours.
vII. derIvatIzIng the calIbratIon standard
The calibration standard contains 100 pmol/μl of each amino
acid (except cystine, at 50 pmol/μl). To derivatize the calibration
standard:
1. Preheat a heating block to 55 °C.
Derivitizing a Sample
Before derivatizing a vacuum-dried sample, you must reconstitute it.
vIII. reconstItutIng the hydrolyzed saMPle
To reconstitute the sample:
1. Prepare a 20 mM HCI solution by adding 10 μL constant-boiling
(6N) HCI to 3 ml HPLC grade water in a clean autosampler vial.
2. Add 20 μl of 20 mM constant-boiling HCI to the sample.
3. Vortex thoroughly.
IX. derIvatIzIng the saMPle
To derivatize the reconstituted sample:
I . Preheat a heating block to 55 °C.
2. Use a clean syringe to deliver 10 μl of calibration standard to
the bottom of a clean sample tube.
3. Use a micropipette to add 70 μl of AccQ•Fluor Borate Buffer
(Vial 1) to the sample tube. Vortex briefly.
4. Use a micropipette with a clean tip to add 20 μl of reconstituted
AccQ•Fluor Reagent to the sample tube. Vortex immediately for
several seconds.
5. Allow to incubate for 1 minute at room temperature.
6. Use a 9-Inch Pasteur pipette to transfer the contents of the
sample tube to the bottom of an autosampler vial limited
volume Insert (LVI). Cap with a siliconelined septum.
7. Heat the vial in a heating block for 10 minutes at 55 °C.
Note: Heating converts a minor side product of Tyr to the major
mono-derivatized compound. Conversion occurs more slowly at room
temperature, with a half life of approximately 1 hour.
A 5 μl injection of derivatized standard contains 50 pmol of each
amino acid derivative (except) cystine at 25 pmol.
2. Add 60 μl of AccQ•Fluor Borate Buffer to the reconstituted
sample in the 6 x 50 mm sample tube and vortex.
3. Add 20 μl of AccQ•Fluor reagent. Vortex immediately for
several seconds.
4. Allow to incubate for one minute at room temperature.
5. Transfer the contents of the tube to an autosampler vial low
volume insert (LVI). Cap with a silicone-lined septum.
6. Heat the vial In a heating block for 10 minutes at 55 °C.
X. storIng derIvatIves
Amino acid derivatives may be stored at room temperature for up to
one week if tightly sealed and protected from evaporation.
AccQ•Fluor Reagent Kit 3

[ Care and Use ManUal ]
XI. sPecIFIcatIons
Waters manufactures the AccQ•Fluor Reagent Kit to exacting analyti-
cal specifications to minimize amino acid background. A 10 μL injec-
tion of a property prepared derivatizatlon blank (20 μL reconstituted
AccQ•Fluor Reagent plus 80 μL. AccQ•Fluor Borate Buffer) contains
less than 2 pmol of any of the hydrolysate amino acids (less than 4
pmol Cys).
Ensure all glassware is clean. For high sensitivity analyses, pyrolyze
all sample tubes at 500 °C for at least 4 hours.
XII. technIcal lIterature
Cohen, S.A. and D.P. Michaud, (1993) “Synthesis of a Fluorescent
Derivatizing Reagent, 6-Aminoquinolyl-N- Hydroxysuccinimidyl Car-
bamate, and its Application for the Analysis of Hydrolysale Amino
Acids via High Performance Liquid Chromatography,” Anal. Biochem.
211, pp 279.287.
XIII. orderIng InForMatIon
Item Part Number
AccQ•Fluor Reagent Kit WAT052880
AccQ•Tag Chemistry Package WAT052875
AccQ•Tag Column, 3.9 x 150 mm WAT052885
AccQ•Tag Eluent A Concentrate, 950 ml WAT052890
AccQ•Tag EIuent B, 950 ml WAT052895
Amino Acid Hydrolysale Standard WA T0 8 8122
Aulosampler Vials, 4 ml, Glass, 48/pkg WAT 072710
Caps for 4 ml Aulosampler Vials, 144/pkg W AT 0 7 9 711
Septa for 4 mL Aulosampler Vials, 1440/pkg WAT073005
150 μL Low Volume Inserts, 144/pkg WAT072704
Springs for Low Volume Inserts, 144/pkg WAT072708
Sample tubes, 6 x 50 mm, 72/pkg WAT00757I
Cohen, S.A., De Antonis, K., and D.P. Michaud, (1993) “Compositional
Protein Analysis Using 6-Aminoquinolyl-N-Hydroxysuccinimidyl
Carbamate, a Novel Derivatization Reagent,” Techniques In Protein
Chemistry IV, pp 289.298, (R.H. Angeletil, Ed.) Academic Press, San
Diego, CA.
AccQ•Fluor Reagent Kit 4

[ Care and Use ManUal ]
Sales Offices:
Austria and European Export
(Central South Eastern Europe,
CIS and Middle East) 431 877 18 07
Australia 2 9933 1777
Belgium 32 2 726 1000
Brazil 55 11 5094 3788
Canada 800 252 4752
China 8621 6495 6999
CIS/Russia +7 495 3367000
Czech Republic 42 02 617 11384
Denmark 45 46 59 8080
Finland +358 9 5659 6288
France (33) 1 30 48 72 00
Germany 49 6196 400600
Japan (81) 3 3471 7191
Korea (82) 2 820 2700
Mexico 5255 5200 1860
The Netherlands +31 (0)76-50 87 200
Norway 47 63 84 60 50
Poland (48) 22 833 4400
Puerto Rico 787 747 8445
Singapore 65 6273 1221
Spain 34 93 600 93 00
Sweden 46 8 555 11500
Switzerland 41 62 889 2030
Taiwan 886 2 2543 1898
United Kingdom 44 208 238 6100
Hong Kong 852 29 64 1800
Hungary 36 1 350 5086
India and India Subcontinent
91 80 2 837 1900
Ireland 353 1 448 1500
Italy 39 02 274 211