[ Care and Use ManUal ]
ACCQ•Fluor reagent kIt
I. IntroduCtIon
The Waters AccQ•Fluor™ Reagent Kit contains five sets of the deriva-
tizing reagents used to perform the AccQ•Tag™ Amino Acid Analysis
Method. A set consists of one vial each of:
– AccQ•Fluor Borate Buffer - Vial 1. You add the buffer to the sam-
ples to generate the proper pH environment for the derivatizatlon.
– AccQ•Fluor Reagent Powder - Vial 2A. The reagent powder Is the
6-aminoquinolyl-N-hydroxysuccinimidyI carbamate (AQC) derivatiz-
ing reagent. It Is shipped dry for maximum stability.
– AccQ•Fluor Reagent Diluent - Vial 2B. Use the diluent(acetonitrile)
to reconstitute the reagent for derivatization.
The Waters AccQ•Fluor Reagent Kit (Figure 1) contains sufficient
reagent for up to 250 analyses.
Contents
I. IntroduCtIon
II. Waters accQ•tag Method overvIeW
III. reagent PurIty
Iv. storage
v. reconstItutIng accQ•Fluor reagent
vI. usIng Waters accQ•Fluor reagent
vII. derIvatIzIng the calIbratIon standard
vIII. reconstItutIng the hydrolyzed saMPle
IX. derIvatIzIng the saMPle
X. storIng derIvatIves
Figure 1: AccQ•Fluor Reagent Kit
AccQ•Fluor Reagent Kit 1
XI. sPecIFIcatIons
XII. technIcal lIterature
XIII. orderIng InForMatIon
[ Care and Use ManUal ]
II. Waters accQ•tag Method overvIeW
The AccQ•Tag Method is a precolumn derivatization and analysis
technique for peptide and protein hydrolysate amino acids. To
perform the AccQ-Tag Method:
1. Use Waters AccQ•Fluor Reagent (US Patent 5,296,599
and European Patent EP 0 533 200 B1) to derivatize the
amino acids.
2. Separate the derivatives by reversed-phase HPLC.
3. Quantitate the derivatives by fluorescence detection.
III. reagent PurIty
The Waters AccQ•Fluor Reagent Kit is manufactured in Waters’- ISO
9001-2006 certified facilities under strict guidelines to ensure the
lowest possible amino acid background. To retain this level of purity,
handle all reagents carefully.
6. Heat Vial 2A on top of heating block, vortex occasionally, until
the powder dissolves. Do not heat longer than 10 minutes.
Reconstituted AccQ•Fluor Reagent is approximately 10 mM
AQC in acetontrile.
Note: You can store the reconstituted AccQ•Tag reagent in a desiccator
at room temperature for up to one week. The AccQ•Tag reagent reacts
with atmospheric moisture so seal the container tightly when not in
use and DO NOT refrigerate.
Reconstituted AccQ•Fluor Reagent is approximately 10 mM AQC in
acetontrile.
IV. storage
If unopened, the AccQ•Fluor Reagent Kit may be stored at room
temperature for up to one year.
v. reconstItutIng accQ•Fluor reagent
To reconstitute the AccQ•Fluor Reagent (Figure 2):
1. Preheat a heating block to 55 °C.
2. Tap Vial 2A lightly before opening to ensure all AccQ•Fluor
Reagent Powder is at the bottom of the vial.
3. Rinse a clean micropipette by drawing and discarding 1 ml of
AccQ•Fluor Reagent Diluent from Vial 2B.
WARNING: Acetonitrile is flammable and toxic. Refer to the Material
Safety Data Sheet.
4. Transfer 1.0 ml of AccQ•Fluor Reagent Diluent From Vial 2B to
the AccQ•Fluor Reagent Powder In Vial 2A.
Figure 2: Reconstituting AccQ•fluor Reagent
storage
Store reconstituted AccQ•Fluor Reagent at room temperature in a
desiccator for up to one week.
Note: AccQ•Fluor reagent reacts with atmospheric moisture. Seal the
container tightly when not in use.
5. Cap Vial 2A lightly and vortex for 10 seconds.
AccQ•Fluor Reagent Kit 2