The information in this document is subject to change without notice and should not be
construed as a commitment by Waters Corporation. Waters Corporation assumes no
responsibility for any errors that may appear in this document. This manual is believed to
be complete and accurate at the time of publication. In no event shall Waters Corporation
be liable for incidental or consequential damages in connection with, or arising from, the
use of this manual.
Millennium, PowerStation, and Waters are registered trademarks and busLAC/E is a
trademark of Waters Corporation.
All other trademarks are the sole property of their respective owners.
The quality management system of Waters’ chromatography applications software design
and manufacturing facility, Milford, Massachusetts, complies with the International
Standard ISO 9001 Quality Managemen t and Quality Assurance Stand ards. Waters’
quality management system is periodically audited by the registering body to ensure
compliance.
STOP
Attention:
instrument.
When you use the instrument, follow generally accepted procedures for quality control and
methods development.
If you observe a change in the retention of a particular compound, in the resolution
between two compounds, or in peak shape, immediately take steps to determine the
reason for the changes. Until you determine the cause of a change, do not rely upon the
results of the separations.
This is a highly sensitive instrument. Read this user's manual before using the
STOP
STOP
Caution:
same type and rating.
Attention:
responsible for compliance could void the user’s authority to operate the equipment.
Attention:
A digital device, pursuant to Part 15 of the FCC rules. These limits are designed to provide
reasonable protection against harmful interference when the equipment is operated in a
commercial environment. This equipment generates, uses, and can radiate radio
frequency energy, and, if not installed and used in accordance with the instruction manual,
may cause harmful interference to radio communications. Operation of this equipment in a
residential area is likely to cause harmful interference, in which case you must correct the
interference at your own expense.
Shielded cables must be used with this unit to ensure compliance with Class A FCC limits.
Note:
Level II category pertains to equipment that receives its electrical power from a local level,
such as an electrical wall outlet.
For continued protection against fire hazard, replace fuses with those of the
Changes or modifications to this unit not expressly approved by the party
This equipment has been tested and found to comply with the limits for a Class
The Installation Category (Overvoltage Category) for t his instrument is Level II. The
Canadian Emissions Notice
This digital apparatus does not exceed the Class A limits for radio noise emissions from
digital apparatus set forth in the Radio Interference Regulations of the Canadian
Departme nt of Communications.
Le présent appareil numérique n’émet pas de bruits radioélectriques dépassant les limites
applicables aux appareils numériques de la classe A prescrites dans les règlements sur le
brouillage radioélectrique édictés par le Ministère des Communications du Canada.
Symbols Used on the Waters 996 Photodiode Array Detector
Direct current
Alternating current
Protective conductor
terminal
Frame or chassis terminal
Caution, risk of electric
shock (high voltage)
Caution or refer to
manual
Caution, hot surface
UV
Ultraviolet light
Table of Contents
How to Use This Guide..................................................................... 10
D-1Mobile Phase Absorbance Measured Against Air or Water...... 70
Table of Contents9
How to Use This Guide
Purpose of This Guide
The
Waters 996 PDA Detector Operator’s Guide
maintaining, and troubleshooting the Waters
optics and the principles of Spectral Contrast used in the Millennium
analyzing the data from the PDA detector. Also included is information on vector analysis,
mobile phase absorbance, specifications, and the warranty.
Audience
This guide is intended for individuals who need to install, operate, maintain, and
troubleshoot the Waters 996 PDA Detector. It is also intended for users who need to
understand the Spectral Contrast principles underlying the processi ng of PDA detector
data by Millennium
32
soft wa r e .
Structure of This Guide
The
Waters 996 PDA Detector Operator’s Guide
Each chapter and appendix page is marked with a tab and a footer to help you quickly
access information.
The following table describes the mater ial covered in each chapter and appendix of this
guide.
describes the procedures for installing,
®
996 PDA Detector . It also describes detector
is divided into chapters and appendixes.
32
software for
Chapter/AppendixDescription
Chapter 1, InstallationDe scr ibes how to install and set up the 996 detector.
Chapter 2, Diagnostics
and Calibration
Chapter 3,
Maintenance
Chapter 4, Princi pl e s of
the 996 PDA Detector
Optics
Chapter 5, Spectral
Contrast Theory
10How to Use This Guide
Describes how to troubleshoot the 996 detector.
Describes how to replace the flow cell, the lamp, and the fuse.
Explains the principles involved in resolving spectral data,
measuring light at the photodiode, verifying wavelengths, and
computing absorbance data.
Describes the calculations used for Spectral Contrast.
Chapter/AppendixDescription
Appendix A, Detector
Provides the specifications of the Waters 996 PDA detector.
Specifications
Appendix B, Spare
Provides a list of recommended and optional spare par ts.
Parts
Appendix C, Warranty
Includes warranty and service information.
Information
Appendix D, Mobile
Phase Absorbance
Provides a table of absorbances at several wavelengths for
common mobile phases.
Related Documentation
The table below lists other guides related to the
.
Guide
TitleDescription
Millennium32 Online
Help
Describes all Millennium32 windows, menus, menu selections,
and dialog boxes. Also includes reference information and
procedures for performing all tasks required to use the
Millenn iu m
32
software. Included as part of the Millennium32
software.
Waters 996 PDA Detector Operator’s
Millennium32 Software
Getting Started Guide
Millennium32 PDA
Software Getting
Started Guide
Provides an introduction to the Millennium32 System.
Describes the basics of how to use Millennium
32
software to
acquire data, develop a processing method, review results,
and print a report. Also covers basics for managing projects
and configuring systems.
Describes the basics of how to use Millennium32 PDA soft ware
to develop a PDA processing method and to review PDA
results
How To Use This Guide11
TitleDescription
Millennium32 System
Insta l latio n /
Configuration Guide
Describes Millen niu m32 software inst a llat io n, includin g the
stand-alone workstation, PowerStation
system, and
™
client/server system. Discusses how to configure the
computer and chromatographic instruments as part of the
Millenn iu m
32
system. Also covers the installa ti o n ,
configuration, and use of the busLAC/E card and LAC/E
Acquisition Server.
Wa te r s 9 9 6 PDA
Describes qualification procedures for the PDA detector.
Detector Qualification
workbook
Related Adobe™ Acrobat Reader Documentation
For detailed information about using the Adobe Acrobat Reader, refer to the
Acrobat Reader Online Guide
. This Online Guide covers procedures such as viewing,
navigating and printing electronic documentation from Adobe Acrobat Reader.
Adobe
Printing From This Electronic Document
Adobe Acrobat Reader lets you easily print pages, pages ranges, or the entire electronic
document by selecting
recommends that you specify a Postscript printer driver for your printer. Ideally, use a
printer that supports 600 dpi print resolution.
Print
from the File menu. For optimum print quantity, Waters
32
Conventions Used in This Guide
This guide uses the following conventions to make text easier to understand.
Purple Text
•
Press 0, then press
•
Italic
text denotes new or important words, and is also used for emphasis. For
example:
An
12How to Use This Guide
indicates user action. For example:
instrument method
Enter
for the remaining fields.
tells the software how to acquire data.
• Underlined, Blue Color text indicates hypertext cross-references to a specific
chapter, section, subsection, or sidehead. Clicking this topic using the hand symbol
automatically brings you to this topic within the electronic document. Right-clicking
and selecting
originating topic. For example:
Before you replace the detector lamp, check the lamp usage indicator located to
the front of the lamp retainer, as described in Section 3.2.1,
Usage.
Notes, Attentions, and Cautions
• Notes call out information that is important to the operator. For example:
Record your results before you proceed to the next step.
Note:
• Attentions provide information about preventing possible damage to the system or
equipment. For example:
Go Back
from the popup context menu brings you back to the
Checking Lamp
STOP
Attention:
To avoid damaging the detector flow cell, do not touch the flow cell
window.
• Cautions provide information essential to the safety of the operator. For example:
Caution:
To avoid chemical or electrical hazards, always observe safe laboratory
practices when operating the system.
How To Use This Guide13
1
1
Installation
The Waters® 996 Photodiode Array (PDA) Detector operates in any standard laboratory
environment. The detector requires electrical power, sample and waste fluidic lines, and
the Millennium®
communication with chart recorders, data integrators, and other instruments that are not
compatible with Millennium software control.
1.1 Instal lation Site Requirements
Install the Waters 996 PDA Detector (Figure 1-1) at a site that meets the specifications
listed in Table 1-1
Table 1-1 Site Requirements
Ambient temperature4 to 40° C (39 to 104° F)
Relativ e hum idi ty20 to 80 percent,
32
Workstation. Optional connections on the detector rear panel allow
.
FactorSpecification
noncondensing
Bench spaceWidth: 11.5 in. (29 cm)
Bench supportCapable of supporting
ClearanceAt least 4 in. (10 cm) on
14Installation
Depth: 24 in. (61 cm)
Height: 8.125 in. (22
cm)
31.5 pounds (14.3 kg)
the back and left sides
for ventilation
11.5 in. (29 c m )
WATERS 996
Photodiode Array
8.5 in.
(22 cm)
24 in.
(61 cm)
Figure 1-1 Waters 996 PDA Detector Dimensions
1.2 Power Connections
Ensure that power connections for the 996 PDA Detector are made according to the
procedures that follow.
1
LAMP
Sample Inlet
Sam ple Out let
Drain Line
Operating Voltage
The 996 PDA Detector has a universal input power supply that requires no voltage
adjustment. The electrical power requirements for the Waters 996 PDA Detector are:
Voltage range:
•
Frequency range:
•
95 to 240 Vac (±10%)
50 to 60 Hz (±3 Hz)
Fuses
The Waters 996 PDA Detector is shipped with fuses rated for North American operation. If
you operate the Waters 996 PDA Detector in another location, install the IEC-rated fuses
(supplied in the Waters 996 Detector Startup Kit) in the fuse holder in the rear of the
detector (refer to Section 3.3,
Replacing the Fuses).
Power Connections15
1
Connecting the Power Cord
Connect one end of the 996 detector power cord to the rear panel power receptacle
(Figure 1-2
) and the other end to a power outlet.
Power Cord
Receptacle
TP01452
Figure 1-2 Detector Rear Panel
1.3 Millennium32 Workstation Connections
The 996 detector requires signal connections to the Millennium32 Workstation over the
IEEE-488 bus. All detector control and data acquisition communications take place over
the IEEE-488 bus.
If an inject start signal is not available over the IEEE-488 bus, you must provide a
Note:
signal at the Event In terminals on the 996 detector rear panel (see Section 1.4.2,
Connecting Event Cables).
1.3.1 Connecting the IEEE-488 Cable
If the 996 detector is to be rack-mounted or stacked on other instruments, use the
Note:
right-angle adaptor included in the Star t up Kit when you make the IEEE-488 connection.
16Installation
To connect the 996 detector to a Millennium32 Workstation:
1. Con nect one end of the IEEE-488 cable to the IEEE-488 receptacle on the rear
panel of the 996 detector. Connect the other end of the cable (stackable
connector for daisy-chaining additional instruments) to the IEEE-488 connector
on any of the other instruments in your chromatographic system (Figure 1-3
Waters
IEEE-488
Cable
).
1
2690
32
IEEE-488 Cabl e
Wate r s 996
Detector
TP01544
IEEE-488
Connector
Millennium
busLAC/E Card
(on Workstati on)
Wate rs
Separations Module
Figure 1-3 Example of IEEE-488 Cable Connections
The order in which you connect IEEE-488 devices to the busLAC/E card on
Note:
the workstation is not important. For example, you can connect the 2690
separations module before or after the 996 detector.
2. Use another IEEE-488 cable to connect to the stackable connector on the first
instrument and the IEEE-488 connector on another instrument.
3. Repeat step 2 for each IEEE-488 i nst rument in your chromatographic system, up
to a maximum of 14 IEEE-488 instruments.
Keep in mind cable-length limitations when you set up your system. For a list
Note:
of IEEE-488 interface guidelines, refer to the Millennium
4. Ensure that all IEEE-488 cable screws are fastened finger-tight.
Millennium32 Workstation Connections17
1.3.2 Setting the IEEE-488 Address
To set the IEEE-488 address for the 996 detector:
1
1. Use a small screwdriver (or similar device) to set the DIP switches on the detector
rear panel (Figure 1-4
) to the IEEE-488 address of the 996 detector. The address
must be a number from 2 to 29 and must be unique within your network.
Refer to the
Millennium32 System Installation/Configuration Guide,
Section 2.3.1,
Connecting IEEE-488 Devices, for the correct IEEE-488 DIP switch settings.
IEEE-488 Cable
Connection
12345
IEEE-488
Address
Switches
– OPEN –
IEE 488 ADDRESS
2. To instruct the 996 detector to accept the new IEEE-488 address, power off, then
power on the detector (see Section 1.6,
18Installation
TP01457
Figure 1-4 Locating the IEEE-488 Address Switches
Startup/S hutdown).
1.4 Non-IEEE-488 Communication Connections
Non-IEEE-488 communication connections on the 996 detector include:
Analog Ou t pu ts
STOP
•
absorbance-unit analog output channels to integrators, chart recorders, or other
components.
Event Inputs and Outputs
•
signals to and from other instruments.
You make all non-IEEE-488 communication cable connections to the 996 detector at the
rear panel (see Figure 1-2
Attention:
disturbances that may affect the performance of this instrument, do not use cables longer
than 9.8 feet (3 meters) when you make connections to the screw-type barrier terminal
strips. In addition, ensure that you always connect the shield of each cable to chassis
ground at one instrument only.
To me et the regulator y requirem ents of immunity from external electrical
– The 996 detector provides two, unattenuated, 1 volt-per
– The 996 detector sends and receives contact closure
).
1.4.1 Connecting Analog Output Cables
The values of the analog output signals generated by the 996 detector are specified
by parameter valu es set from the Millen niu m
“Waters 996 Detector Properties” topic in the
Required Materials
•One small, flat-blade screwdriver
32
Workstation. For details, refer to the
Millennium32 Online Help
Find tab.
1
•One electrical insulation stripping tool
•Analog signal cables (from Startup Kit)
Procedure
To connect the Waters 996 detector to a device that receives analog output signals from
the 996 detector
1. Pull off the Analog output term inal str ip from the 996 detector rear panel
(Figure 1-5
). This step simplifies the following steps.
Non-IEEE-488 Communication Connections19
1
Removable An a l og O utput
Term inal Strip
+
Analog Ou t 1
–
IEE 488 ADDRESS
123456
+
Analog Out 2
–
Figure 1-5 Analog Output Terminals
2. Insert the bare wires at one end of an analog signal cable into the positive (+)
and negative (–) terminals of Analog Out 1 (see Fi gure 1-5
screws to secure the + and – wires.
3. Connect the other end of the analog signal cable to the appropriate analog input
terminal on the external device, being sure to maintain negative-to-negative and
positive-to-positive continuity.
4. Reinstall the Analog Output strip.
1.4.2 Connecting Event Cables
The 996 detector has four terminal strip connections for contact closure signals:
TP01456
). Tighten the two
• Two input (inject start) signal terminals
• Two output (programmable event table) signal terminals
If an inject start signal is not avail a ble over the IEEE-488 bus, you must provide a signal at
an Event In terminal on the 996 detector rear panel. Manual injectors such as the
Rheodyne 7725i provide a cable that connects the injector to an Event In terminal on the
996 detector rear panel.
20Installation
The values of the event output signals generated by the 996 detector are specified by
parameter values set from the Millennium
996 Detector Properties” topic in the
32
Workstation. For details, refer to the “Waters
Millenn iu m32 Online Help
Find tab.
Electrical Specifications
Before you connect an external device to an event input or output terminal, refer to the
electrical specifications in Table 1-2
Table 1-2 Event In (Inject Start) Ter m in al Specifications on TTL or Switch Closure
ParameterSpecification
Low trigger<1.8 V
High trigger>3.0 V
Protected to±30 V
Minimum pulse width30 msec
Maximum current5 mA
Table 1-3 Event Out Ter minal Spec ifications on Contact Closure
ParameterSpecification
and Table 1-3.
1
Maximum power10 W
Maximum current0.5 A at 20 V
Maximum voltage24 V RMS
Attention:
STOP
electrical connections as outlined in this section.
Required Materials
•Small flat-blade screwdriver
•One electrical insulation stripping tool
•Event signal cables (in Startup Kit)
To avoid damage to the 996 detector electronics, be sure you make the proper
Non-IEEE-488 Communication Connections21
Making Event Input/Output Connections
To connect the 996 detector to an external event input or output device:
1
1. Pull off the Event Input/Output terminal strip from the rear panel (Figure 1-6
simplifies the following steps.
Removable Event
Input/Output
Terminal Strip
Event
Out 2
Event
Out 1
Event
–
In 2
+
1234567890
–
Event
+
In 1
IEE 488 ADDRESS
TP01455
Figure 1-6 Event Input/Output Terminal Strip
). This
2. Insert the bare wires at one end of the event signal cable into the positive (+)
and negative (–) slots of the appropriate event input or output terminal (see
Figure 1-6
). Tighten the two screws to secure the +and – wires.
3. Connect the other end of the event signal cable to the appropriate event input or
event output terminal on the external device.
4. Reinstall the Event Input/Output strip.
1.5 Fluidic Connections
Caution:
handing solvents. Refer to the Material Safety Data Sheets for the solvents in use.
22Installation
To avoid chemical hazards, always observe safe laboratory practices when
Required Materials
•5/16-inch open-end wrench
•0.009-inch (0.23 mm) I.D. stainless steel tubing (in Startup Kit)
•Stainless steel tubing cutter or scribing file
•Pliers, plastic-covered, or with cloth
•Compression screw assembles, three
Procedure
To make fluidic connections to the 996 detector:
1. Measure the lengths of tubing needed to connect:
• The column outlet to the 996 detector inlet.
Be sure that you keep the length of this tubing as short as possible to
Note:
prevent band broadening.
• The 996 detector outlet to a waste collection bottle.
Ensure the length of this tubing is at least 1 to 2 feet (30 to 60 cm) to
Note:
prevent air bubbles from forming in the flow cell.
2. Cut the two lengths of tubing as follows:
• Use a Waters 1/16-inch stainless steel tubing cutter or a file with a cutting edge
to scribe the circumference of the tubing at the desired break point.
1
• Grasp the tubing on both sides of the scribed mark with cloth- or
plastic-covered pliers (to prevent marring the surface), then gently work the
tubing back and forth until it separates.
• File the tubing ends smooth and straight to minimize dead volume and band
broadening.
3. Assemble a compression fitting (as shown in Figure 1-7
column outlet line and at one end of the detector outlet line.
) at both ends of the
Fluidic Connec tions23
Loading...
+ 53 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.