The information in this document is subject to change without notice and should not be
construed as a commitment by Waters Corporation. Waters Corporation assumes no
responsibility for any errors that may appear in this document. This manual is believed to
be complete and accurate at the time of publication. In no event shall Waters Corporation
be liable for incidental or consequential damages in connection with, or arising from, the
use of this manual.
Millennium, PowerStation, and Waters are registered trademarks and busLAC/E is a
trademark of Waters Corporation.
All other trademarks are the sole property of their respective owners.
The quality management system of Waters’ chromatography applications software design
and manufacturing facility, Milford, Massachusetts, complies with the International
Standard ISO 9001 Quality Managemen t and Quality Assurance Stand ards. Waters’
quality management system is periodically audited by the registering body to ensure
compliance.
STOP
Attention:
instrument.
When you use the instrument, follow generally accepted procedures for quality control and
methods development.
If you observe a change in the retention of a particular compound, in the resolution
between two compounds, or in peak shape, immediately take steps to determine the
reason for the changes. Until you determine the cause of a change, do not rely upon the
results of the separations.
This is a highly sensitive instrument. Read this user's manual before using the
STOP
STOP
Caution:
same type and rating.
Attention:
responsible for compliance could void the user’s authority to operate the equipment.
Attention:
A digital device, pursuant to Part 15 of the FCC rules. These limits are designed to provide
reasonable protection against harmful interference when the equipment is operated in a
commercial environment. This equipment generates, uses, and can radiate radio
frequency energy, and, if not installed and used in accordance with the instruction manual,
may cause harmful interference to radio communications. Operation of this equipment in a
residential area is likely to cause harmful interference, in which case you must correct the
interference at your own expense.
Shielded cables must be used with this unit to ensure compliance with Class A FCC limits.
Note:
Level II category pertains to equipment that receives its electrical power from a local level,
such as an electrical wall outlet.
For continued protection against fire hazard, replace fuses with those of the
Changes or modifications to this unit not expressly approved by the party
This equipment has been tested and found to comply with the limits for a Class
The Installation Category (Overvoltage Category) for t his instrument is Level II. The
Canadian Emissions Notice
This digital apparatus does not exceed the Class A limits for radio noise emissions from
digital apparatus set forth in the Radio Interference Regulations of the Canadian
Departme nt of Communications.
Le présent appareil numérique n’émet pas de bruits radioélectriques dépassant les limites
applicables aux appareils numériques de la classe A prescrites dans les règlements sur le
brouillage radioélectrique édictés par le Ministère des Communications du Canada.
Symbols Used on the Waters 996 Photodiode Array Detector
Direct current
Alternating current
Protective conductor
terminal
Frame or chassis terminal
Caution, risk of electric
shock (high voltage)
Caution or refer to
manual
Caution, hot surface
UV
Ultraviolet light
Table of Contents
How to Use This Guide..................................................................... 10
D-1Mobile Phase Absorbance Measured Against Air or Water...... 70
Table of Contents9
How to Use This Guide
Purpose of This Guide
The
Waters 996 PDA Detector Operator’s Guide
maintaining, and troubleshooting the Waters
optics and the principles of Spectral Contrast used in the Millennium
analyzing the data from the PDA detector. Also included is information on vector analysis,
mobile phase absorbance, specifications, and the warranty.
Audience
This guide is intended for individuals who need to install, operate, maintain, and
troubleshoot the Waters 996 PDA Detector. It is also intended for users who need to
understand the Spectral Contrast principles underlying the processi ng of PDA detector
data by Millennium
32
soft wa r e .
Structure of This Guide
The
Waters 996 PDA Detector Operator’s Guide
Each chapter and appendix page is marked with a tab and a footer to help you quickly
access information.
The following table describes the mater ial covered in each chapter and appendix of this
guide.
describes the procedures for installing,
®
996 PDA Detector . It also describes detector
is divided into chapters and appendixes.
32
software for
Chapter/AppendixDescription
Chapter 1, InstallationDe scr ibes how to install and set up the 996 detector.
Chapter 2, Diagnostics
and Calibration
Chapter 3,
Maintenance
Chapter 4, Princi pl e s of
the 996 PDA Detector
Optics
Chapter 5, Spectral
Contrast Theory
10How to Use This Guide
Describes how to troubleshoot the 996 detector.
Describes how to replace the flow cell, the lamp, and the fuse.
Explains the principles involved in resolving spectral data,
measuring light at the photodiode, verifying wavelengths, and
computing absorbance data.
Describes the calculations used for Spectral Contrast.
Chapter/AppendixDescription
Appendix A, Detector
Provides the specifications of the Waters 996 PDA detector.
Specifications
Appendix B, Spare
Provides a list of recommended and optional spare par ts.
Parts
Appendix C, Warranty
Includes warranty and service information.
Information
Appendix D, Mobile
Phase Absorbance
Provides a table of absorbances at several wavelengths for
common mobile phases.
Related Documentation
The table below lists other guides related to the
.
Guide
TitleDescription
Millennium32 Online
Help
Describes all Millennium32 windows, menus, menu selections,
and dialog boxes. Also includes reference information and
procedures for performing all tasks required to use the
Millenn iu m
32
software. Included as part of the Millennium32
software.
Waters 996 PDA Detector Operator’s
Millennium32 Software
Getting Started Guide
Millennium32 PDA
Software Getting
Started Guide
Provides an introduction to the Millennium32 System.
Describes the basics of how to use Millennium
32
software to
acquire data, develop a processing method, review results,
and print a report. Also covers basics for managing projects
and configuring systems.
Describes the basics of how to use Millennium32 PDA soft ware
to develop a PDA processing method and to review PDA
results
How To Use This Guide11
TitleDescription
Millennium32 System
Insta l latio n /
Configuration Guide
Describes Millen niu m32 software inst a llat io n, includin g the
stand-alone workstation, PowerStation
system, and
™
client/server system. Discusses how to configure the
computer and chromatographic instruments as part of the
Millenn iu m
32
system. Also covers the installa ti o n ,
configuration, and use of the busLAC/E card and LAC/E
Acquisition Server.
Wa te r s 9 9 6 PDA
Describes qualification procedures for the PDA detector.
Detector Qualification
workbook
Related Adobe™ Acrobat Reader Documentation
For detailed information about using the Adobe Acrobat Reader, refer to the
Acrobat Reader Online Guide
. This Online Guide covers procedures such as viewing,
navigating and printing electronic documentation from Adobe Acrobat Reader.
Adobe
Printing From This Electronic Document
Adobe Acrobat Reader lets you easily print pages, pages ranges, or the entire electronic
document by selecting
recommends that you specify a Postscript printer driver for your printer. Ideally, use a
printer that supports 600 dpi print resolution.
Print
from the File menu. For optimum print quantity, Waters
32
Conventions Used in This Guide
This guide uses the following conventions to make text easier to understand.
Purple Text
•
Press 0, then press
•
Italic
text denotes new or important words, and is also used for emphasis. For
example:
An
12How to Use This Guide
indicates user action. For example:
instrument method
Enter
for the remaining fields.
tells the software how to acquire data.
• Underlined, Blue Color text indicates hypertext cross-references to a specific
chapter, section, subsection, or sidehead. Clicking this topic using the hand symbol
automatically brings you to this topic within the electronic document. Right-clicking
and selecting
originating topic. For example:
Before you replace the detector lamp, check the lamp usage indicator located to
the front of the lamp retainer, as described in Section 3.2.1,
Usage.
Notes, Attentions, and Cautions
• Notes call out information that is important to the operator. For example:
Record your results before you proceed to the next step.
Note:
• Attentions provide information about preventing possible damage to the system or
equipment. For example:
Go Back
from the popup context menu brings you back to the
Checking Lamp
STOP
Attention:
To avoid damaging the detector flow cell, do not touch the flow cell
window.
• Cautions provide information essential to the safety of the operator. For example:
Caution:
To avoid chemical or electrical hazards, always observe safe laboratory
practices when operating the system.
How To Use This Guide13
1
1
Installation
The Waters® 996 Photodiode Array (PDA) Detector operates in any standard laboratory
environment. The detector requires electrical power, sample and waste fluidic lines, and
the Millennium®
communication with chart recorders, data integrators, and other instruments that are not
compatible with Millennium software control.
1.1 Instal lation Site Requirements
Install the Waters 996 PDA Detector (Figure 1-1) at a site that meets the specifications
listed in Table 1-1
Table 1-1 Site Requirements
Ambient temperature4 to 40° C (39 to 104° F)
Relativ e hum idi ty20 to 80 percent,
32
Workstation. Optional connections on the detector rear panel allow
.
FactorSpecification
noncondensing
Bench spaceWidth: 11.5 in. (29 cm)
Bench supportCapable of supporting
ClearanceAt least 4 in. (10 cm) on
14Installation
Depth: 24 in. (61 cm)
Height: 8.125 in. (22
cm)
31.5 pounds (14.3 kg)
the back and left sides
for ventilation
11.5 in. (29 c m )
WATERS 996
Photodiode Array
8.5 in.
(22 cm)
24 in.
(61 cm)
Figure 1-1 Waters 996 PDA Detector Dimensions
1.2 Power Connections
Ensure that power connections for the 996 PDA Detector are made according to the
procedures that follow.
1
LAMP
Sample Inlet
Sam ple Out let
Drain Line
Operating Voltage
The 996 PDA Detector has a universal input power supply that requires no voltage
adjustment. The electrical power requirements for the Waters 996 PDA Detector are:
Voltage range:
•
Frequency range:
•
95 to 240 Vac (±10%)
50 to 60 Hz (±3 Hz)
Fuses
The Waters 996 PDA Detector is shipped with fuses rated for North American operation. If
you operate the Waters 996 PDA Detector in another location, install the IEC-rated fuses
(supplied in the Waters 996 Detector Startup Kit) in the fuse holder in the rear of the
detector (refer to Section 3.3,
Replacing the Fuses).
Power Connections15
1
Connecting the Power Cord
Connect one end of the 996 detector power cord to the rear panel power receptacle
(Figure 1-2
) and the other end to a power outlet.
Power Cord
Receptacle
TP01452
Figure 1-2 Detector Rear Panel
1.3 Millennium32 Workstation Connections
The 996 detector requires signal connections to the Millennium32 Workstation over the
IEEE-488 bus. All detector control and data acquisition communications take place over
the IEEE-488 bus.
If an inject start signal is not available over the IEEE-488 bus, you must provide a
Note:
signal at the Event In terminals on the 996 detector rear panel (see Section 1.4.2,
Connecting Event Cables).
1.3.1 Connecting the IEEE-488 Cable
If the 996 detector is to be rack-mounted or stacked on other instruments, use the
Note:
right-angle adaptor included in the Star t up Kit when you make the IEEE-488 connection.
16Installation
To connect the 996 detector to a Millennium32 Workstation:
1. Con nect one end of the IEEE-488 cable to the IEEE-488 receptacle on the rear
panel of the 996 detector. Connect the other end of the cable (stackable
connector for daisy-chaining additional instruments) to the IEEE-488 connector
on any of the other instruments in your chromatographic system (Figure 1-3
Waters
IEEE-488
Cable
).
1
2690
32
IEEE-488 Cabl e
Wate r s 996
Detector
TP01544
IEEE-488
Connector
Millennium
busLAC/E Card
(on Workstati on)
Wate rs
Separations Module
Figure 1-3 Example of IEEE-488 Cable Connections
The order in which you connect IEEE-488 devices to the busLAC/E card on
Note:
the workstation is not important. For example, you can connect the 2690
separations module before or after the 996 detector.
2. Use another IEEE-488 cable to connect to the stackable connector on the first
instrument and the IEEE-488 connector on another instrument.
3. Repeat step 2 for each IEEE-488 i nst rument in your chromatographic system, up
to a maximum of 14 IEEE-488 instruments.
Keep in mind cable-length limitations when you set up your system. For a list
Note:
of IEEE-488 interface guidelines, refer to the Millennium
4. Ensure that all IEEE-488 cable screws are fastened finger-tight.
Millennium32 Workstation Connections17
1.3.2 Setting the IEEE-488 Address
To set the IEEE-488 address for the 996 detector:
1
1. Use a small screwdriver (or similar device) to set the DIP switches on the detector
rear panel (Figure 1-4
) to the IEEE-488 address of the 996 detector. The address
must be a number from 2 to 29 and must be unique within your network.
Refer to the
Millennium32 System Installation/Configuration Guide,
Section 2.3.1,
Connecting IEEE-488 Devices, for the correct IEEE-488 DIP switch settings.
IEEE-488 Cable
Connection
12345
IEEE-488
Address
Switches
– OPEN –
IEE 488 ADDRESS
2. To instruct the 996 detector to accept the new IEEE-488 address, power off, then
power on the detector (see Section 1.6,
18Installation
TP01457
Figure 1-4 Locating the IEEE-488 Address Switches
Startup/S hutdown).
1.4 Non-IEEE-488 Communication Connections
Non-IEEE-488 communication connections on the 996 detector include:
Analog Ou t pu ts
STOP
•
absorbance-unit analog output channels to integrators, chart recorders, or other
components.
Event Inputs and Outputs
•
signals to and from other instruments.
You make all non-IEEE-488 communication cable connections to the 996 detector at the
rear panel (see Figure 1-2
Attention:
disturbances that may affect the performance of this instrument, do not use cables longer
than 9.8 feet (3 meters) when you make connections to the screw-type barrier terminal
strips. In addition, ensure that you always connect the shield of each cable to chassis
ground at one instrument only.
To me et the regulator y requirem ents of immunity from external electrical
– The 996 detector provides two, unattenuated, 1 volt-per
– The 996 detector sends and receives contact closure
).
1.4.1 Connecting Analog Output Cables
The values of the analog output signals generated by the 996 detector are specified
by parameter valu es set from the Millen niu m
“Waters 996 Detector Properties” topic in the
Required Materials
•One small, flat-blade screwdriver
32
Workstation. For details, refer to the
Millennium32 Online Help
Find tab.
1
•One electrical insulation stripping tool
•Analog signal cables (from Startup Kit)
Procedure
To connect the Waters 996 detector to a device that receives analog output signals from
the 996 detector
1. Pull off the Analog output term inal str ip from the 996 detector rear panel
(Figure 1-5
). This step simplifies the following steps.
Non-IEEE-488 Communication Connections19
1
Removable An a l og O utput
Term inal Strip
+
Analog Ou t 1
–
IEE 488 ADDRESS
123456
+
Analog Out 2
–
Figure 1-5 Analog Output Terminals
2. Insert the bare wires at one end of an analog signal cable into the positive (+)
and negative (–) terminals of Analog Out 1 (see Fi gure 1-5
screws to secure the + and – wires.
3. Connect the other end of the analog signal cable to the appropriate analog input
terminal on the external device, being sure to maintain negative-to-negative and
positive-to-positive continuity.
4. Reinstall the Analog Output strip.
1.4.2 Connecting Event Cables
The 996 detector has four terminal strip connections for contact closure signals:
TP01456
). Tighten the two
• Two input (inject start) signal terminals
• Two output (programmable event table) signal terminals
If an inject start signal is not avail a ble over the IEEE-488 bus, you must provide a signal at
an Event In terminal on the 996 detector rear panel. Manual injectors such as the
Rheodyne 7725i provide a cable that connects the injector to an Event In terminal on the
996 detector rear panel.
20Installation
The values of the event output signals generated by the 996 detector are specified by
parameter values set from the Millennium
996 Detector Properties” topic in the
32
Workstation. For details, refer to the “Waters
Millenn iu m32 Online Help
Find tab.
Electrical Specifications
Before you connect an external device to an event input or output terminal, refer to the
electrical specifications in Table 1-2
Table 1-2 Event In (Inject Start) Ter m in al Specifications on TTL or Switch Closure
ParameterSpecification
Low trigger<1.8 V
High trigger>3.0 V
Protected to±30 V
Minimum pulse width30 msec
Maximum current5 mA
Table 1-3 Event Out Ter minal Spec ifications on Contact Closure
ParameterSpecification
and Table 1-3.
1
Maximum power10 W
Maximum current0.5 A at 20 V
Maximum voltage24 V RMS
Attention:
STOP
electrical connections as outlined in this section.
Required Materials
•Small flat-blade screwdriver
•One electrical insulation stripping tool
•Event signal cables (in Startup Kit)
To avoid damage to the 996 detector electronics, be sure you make the proper
Non-IEEE-488 Communication Connections21
Making Event Input/Output Connections
To connect the 996 detector to an external event input or output device:
1
1. Pull off the Event Input/Output terminal strip from the rear panel (Figure 1-6
simplifies the following steps.
Removable Event
Input/Output
Terminal Strip
Event
Out 2
Event
Out 1
Event
–
In 2
+
1234567890
–
Event
+
In 1
IEE 488 ADDRESS
TP01455
Figure 1-6 Event Input/Output Terminal Strip
). This
2. Insert the bare wires at one end of the event signal cable into the positive (+)
and negative (–) slots of the appropriate event input or output terminal (see
Figure 1-6
). Tighten the two screws to secure the +and – wires.
3. Connect the other end of the event signal cable to the appropriate event input or
event output terminal on the external device.
4. Reinstall the Event Input/Output strip.
1.5 Fluidic Connections
Caution:
handing solvents. Refer to the Material Safety Data Sheets for the solvents in use.
22Installation
To avoid chemical hazards, always observe safe laboratory practices when
Required Materials
•5/16-inch open-end wrench
•0.009-inch (0.23 mm) I.D. stainless steel tubing (in Startup Kit)
•Stainless steel tubing cutter or scribing file
•Pliers, plastic-covered, or with cloth
•Compression screw assembles, three
Procedure
To make fluidic connections to the 996 detector:
1. Measure the lengths of tubing needed to connect:
• The column outlet to the 996 detector inlet.
Be sure that you keep the length of this tubing as short as possible to
Note:
prevent band broadening.
• The 996 detector outlet to a waste collection bottle.
Ensure the length of this tubing is at least 1 to 2 feet (30 to 60 cm) to
Note:
prevent air bubbles from forming in the flow cell.
2. Cut the two lengths of tubing as follows:
• Use a Waters 1/16-inch stainless steel tubing cutter or a file with a cutting edge
to scribe the circumference of the tubing at the desired break point.
1
• Grasp the tubing on both sides of the scribed mark with cloth- or
plastic-covered pliers (to prevent marring the surface), then gently work the
tubing back and forth until it separates.
• File the tubing ends smooth and straight to minimize dead volume and band
broadening.
3. Assemble a compression fitting (as shown in Figure 1-7
column outlet line and at one end of the detector outlet line.
) at both ends of the
Fluidic Connec tions23
1
Tubing
Compression
Screw
Ferrule
End Must Be Straight
and Smoot h to Prevent
Dead Volume
TP01139
Figure 1-7 Compression Screw Assembly
4. Bottom one end of the column outlet tubing in the fitting seat of the column
outlet, then tighten the compress ion screw about 3/4-turn past finger-tight ( using
the 5/16-inch open-end wrench).
5. Bottom the other end of the tubing in the fitting seat of the detector inlet, then
tighten the compression screw as in step 4.
6. Bottom one end of the detector outlet tubing with the compression fitting in the
fitting seat of the detector outlet, then tighten the compression screw about
3/4-turn past finger-tight. Insert the other end of the tubing in t he waste
container.
To avoid damage to the flow cell, avoid pressures approaching the maximum
STOP
Attention:
pressure for the analytical flow cell, 1000 psi (70 kg/cm
1.6 Startup/Shutdown
Follow the procedures in this section to ensure reliable detector performance.
Distan ce D etermi n e d by
the Union or Column Fitting
2
).
Startup
To star t up the 996 detecto r :
1. In your instrument method, set the solvent delivery system or pump to deliver 1
mL/min of degassed mobile phase. For details, refer to the “Waters 2690
Separations Module Proper ties” or the “Waters 600 Properties” topic in the
Millennium32 Online Help
24Installation
Find tab.
Use only thoroughly degassed HPLC-grade solvents. Gas in the mobile
Note:
phase may form bubbl es in the fl ow cell and cause the detector to fail the Ref erence
Energy diagnostic.
2. Flush the detector for 10 minutes or until no bubbles appear in the outlet line.
3. Press the 0/1 (Off/On) switch on the front panel of the detector (Figure 1-8
1
(On) position.
) to the
4. Observe the Lamp and Status indicator lights on the front panel of the detector
(Figure 1-8
).
• If both lights remain illu mi nated , the detector pas sed the internal diagnostic s.
• If either indicator light blinks or is off, refer to the troubleshooting tables in
Chapter 2,
Diagnostics and Calibration.
WATERS 996
Photodiode Array
Detector
On/Off
Switch
LAMP
1
Status
Indicator
Lamp
Indicator
TP01460
Figure 1-8 996 Detector Indicator Lights
5. Wait 1 hour for the deuterium lamp to stabilize before you attempt to acquire
data at low absorbances.
Startup/Shutdown25
Shutdown
To shut down the 996 detector:
1
1. If the mobile phase contains buffers, set the solvent delivery system or pump to
deliver 1 mL/min of HPLC-grade water for 10 minutes. Otherwise, set the solvent
delivery system or pump to deliver 1 mL/min of degassed methanol for 10 minutes.
2. Press the 0/1 (Off/On) Switch on the front panel of the detector to the 0 (Off)
position.
26Installation
2
Diagnostics and Calibration
The Waters 996 Photodiode Array Detector automatically runs a series of internal
diagnostics upon start up. The indicator lights on the front of the detector and messages at
the Millennium
(Figure 2-1
32
Workstation show the results of the start up internal diagnosti cs
).
If you need to determine the cause of a problem during operation of the detector, you can
run the same internal diagnostics from the Millennium
information about the performance of the detector is also available through the PDA
Calibration window, accessed from QuickSet in the Millennium
If you encounter a problem that you cannot troubleshoot (see Section 2.1,
Diagnostics), contact Waters Technical Service at (800) 252-4752,
customers only
Technical Service Representative, or call Wat ers c orporate headquarters for assistance at
1-508-478-2000 (U.S.).
. Other customers, call your local Waters subsidiary or your local Waters
2.1 Startup Diagnostics
Refer to Table 2-1 to troubleshoot problems encountered during startup diagnostics and
during detector operation.
32
Workstation. Additional
32
software .
Startup
U.S. and Canadian
2
Startup Diagnostics27
2
WATERS 996
Photodiode Array
Detector
Figure 2-1 996 PDA Detector Indicator Lights
Table 2-1 996 Detector Troubleshooting
SymptomPossible CauseCorrective Action
On/Off
Switch
LAMP
Status
Indicator
Lamp
Indicator
TP01460
Status light off No power.1. Check line cord connections.
Status light blinks and
lamp light off
28Diagnostics and Calibration
2. Check outlet for power.
Blown fuse.Replace fuse (see Section 3.3,
Replacing the Fuses).
Detector is st ill
Wait for diagnostics to end.
performing
diagnostics
Fail ed startup
diagnostics.
1. Check that lamp door is secure.
2. Replace lamp. If replacing the lamp
fails to correct the problem, contact
Waters Technical Service.
Insufficient energy
reaching photodiode
array because of air
bubble or dirty flow
cell can cause shutter
diagnostic to fail.
Weak lamp.Replace lamp (see Section 3.2,
Shutter failure.Run the Shutter diagnostic. For
Detector not
connected to
busLAC/E or to
32
LAC/E
server in the
Millennium
Workstation
acquisition
32
Flush the flow cell (see Section 3.1.1,
Flushing the Flow Cell).
Flush the flow cell (see S ec tion 3.1.1,
Flushing the Flow Cell).
To prevent air bubbles from forming,
check that there is a 1- to 2-f oot (30- to
60-cm) length of 0.009-inch (0.23-mm)
I.D. tubing connected to the detector
waste outlet.
Replacing the Lamp).
details, refer to the “PDA Diagnostics
Window” topic in the
Flush the flow cell, or apply slight
backpressure on the detector waste
outlet.
To prevent air bubbles, check that
there is a 1- to 2-foot (30- to 60-cm)
length of 0.009-inch (0.23-mm) I.D.
tubing connected to the detector
waste outlet.
Rebuild flow cell with a new gasket
(see Section 3.1.3,
Cleaning the Flow Cell).
Check fittings for overtightening or
undertightening , and replace fittings if
necessary.
Disassembling and
There are two types of user-initiated PDA diagnostic tests:
Internal Tests
•
source of a malfunction. These tests do not require connections to external devices.
Interactive Tests
•
communications between the detector and connected external devices. These tests
require connections to pump flow and/or test equipment.
You cannot perform diagnostics on a 996 detector while it is acquiring data.
Note:
The system administrator can restrict access to the 996 detector diagnostics by
Note:
disabling user access to Quick Set. For details, refer to the “User Type Properties Dialog
Box” topic in the Millennium
30Diagnostics and Calibration
– Tests run by the instrument firm ware that help you determine the
– Tests that check analog output and event input/output signal
32
Online Help Find tab.
You can run all user-initiated dia gnos tics from Qui ckSet in the Millen nium32 software. For
more information on QuickSet and PDA diagnostics, refer to the “PDA Diagnostics
Window” topic in the
Millennium32 Online Help
Find tab.
If you encounter a problem that you cannot troubleshoot (see Section 2.1,
Diagnostics), contact Waters Technical Service at (800) 252-4752,
customers only
Technical Service Representative, or call Wat ers c orporate headquarters for assistance at
1-508-478-2000 (U.S.).
. Other customers, call your local Waters subsidiary or your local Waters
2.3 PDA Calibration
You can adjust, or calibrate, the 996 detector to ensure that wavelength readings are
accurate. Recalibrate the 996 detector
Internal Diagnostics tests) fails.
You calibrate the 996 detector using the PDA Calibration window, which you access from
QuickSet and which allows you to:
• View the effects of exposure time on photodiode saturation for a given wavelength
range.
• Verify the wavelength location of the deuterium spectrum Balmer lines (486.0 nm
and
656.1 nm).
• Recalibrate to set the 486-nm peak at the proper wavelength.
• Ensure precise data for library matching.
Startup
U.S. and Canadian
only
if the Wavelength Accuracy diagnostic (in the
2
The system administrator can restrict access to the PDA Calibration window.
Note:
Recalibrating the wavelength requires that spectral libraries be reentered.
Required Materials
• HPLC-grade methanol
• HPLC-grade water
Preparing for Calibration
Ensure that the flow cell is clean before you check calibration. (See Section 3.1.1,
Note:
Flushing th e F low Cell .)
PDA Calibration31
2
To prepare for calibration:
1. Set the pump to deliver 1 mL/min of degassed methanol for 10 minutes. If methanol
is not miscible with the previous solvent, flush with a miscible sovent before
switching to methanol.
2. If you have been using buffers, flush with HPLC-quality water at 1 mL/min for 10
minutes, then switch to methanol for 10 minutes.
Ensure that the solvent is miscible with the previous mobile phase.
Note:
For information on performing calibration, refer to the “PDA Calibration Window” topic in
Millennium32 Online Help
the
Find tab.
32Diagnostics and Calibration
3
Maintenance
This chapter covers maintenance of the Waters 996 Photodiode Array Detector flow cell,
lamp, and fuse.
Caution:
supply covers. The power supply does not contain user-serviceable component s.
To avoid the po ssibility of elec tric shock, do no t rem ove the 996 de tec t or power
3.1 Fl ow Cell Maintenance
The flow cell requires maintenance when:
• The reference spectrum changes.
• The Lamp diagnostic (in the Millennium
lamp status light i s on (see Table 2-1
• The 996 detector causes high backpressure.
Conditions other than a dirty flow cell may cause decreased lamp intensity . F or
Note:
more information, refer to Chapter 2,
Flow cell maintenance consists of:
• Flushing the flow cell
• Removing the flow cell
• Cleaning the flow cell
• Installing the flow cell a ssembly
Diagnostics and Calibration.
32
PDA Diagnostics window) fails, and the
).
3
3.1.1 Flushing the Flow Cell
Required Materials
•HPLC grade water
•HPLC grade methanol
If the flow cell requires cleaning, the first cleaning method to try is flushing the flow cell
with solvent.
Flow Cell Maintenance33
To flush the flow cell:
1. Select a solvent compatible with the samples and mobile phases that you have been
using. If you have been using buffers, flush with HPLC-grade water f or 10 minutes at
1 mL/min, then switch to a low-surface-tension solvent such as methanol.
3
STOP
Attention:
2. Set p ump flow to 1 mL/min, then run the pump 10 minutes.
3. Test the lamp energy by performing the Lamp diagnostic tes t. For details, refer to
the “PDA Diagnostics Window” topic in the
If the lamp diagnostic fails and the lamp has not been used more than 1000
hours, disassemble the flow cell and clean the flow cell components using the
procedure described in Section 3.1.2,
Ensure that the solvent is miscible with the previous mobile phase.
3.1.2 Removing the Flow Cell
You do not need to shut down the 996 detector to remove and replace the flow cell.
Note:
Required Materials
•5/16-inch open-end wrench
•Phillips screwdriver
Procedure
To remove the 996 detector flow cell:
1. Set the flow to
2. Power off the solvent delivery system or pump.
0.0
mL/min.
Millennium
Removing the Flow Cell.
32
Online Help
Find tab.
Caution:
inlet or outlet fluidic lines while there is pressure in the chromatographic system.
Always vent your system before disconnecting fluidic lines.
3. Use the 5/16-inch wrench to disconnect the fluidic lines at the front of the
detector.
4. Lift up the 996 detector front cover and pull the front cover from the detector
chassis.
5. Open the flow cell access door by pulling the black thumbtab, then pull the door
gently toward you (Figure 3-1
34Maintenance
To avoid the possibility of leaking mobile phase, do not disconnect the
).
Flow C ell Access
Door
Thumbtabs
LAMP
STATUS
TP01461
Figure 3-1 Flow Cell Access Door
6. Use th e P h illips s crewd river to lo osen the two thu m b scr ews tha t ho ld t h e f low
cell assembly to the optics bench and the thumbscrew that secures the bracket
holding the fluidic connections, then detach the bracket (Figure 3-2
).
3
Thumbs c rews
Holding Flow
Cell Assembly
and Bracket
Fluidic Connec ti ons
(Inside Bracket)
TP01462
Figure 3-2 Removing the Flow Cell Assembly
Flow Cell Maintenance35
7. Pull the flow cell assembly and fluidic connection bracket gently toward you to
remove it from the detector (Figure 3-3
).
Flow Cell Body
Lens Holder
Assembly
3
Fluidic
Connections
Figure 3-3 Flow Cell and Fluidic Connections Assemblies
3.1.3 Disassembling and Cleaning the Flow Cell
The lens surface finish and the alignment of the lenses are critical to the performance of
the 996 detector. Be careful not to touch or damage the lenses and the lens holders.
STOP
STOP
Attention:
inspecting, cleaning, or replacing parts within the flow cell or when removing or replacing
the flow cell within its assembly.
Attention:
inspecting, cleaning, or replacing parts within the flow cell or when removing or replacing
parts within the flow cell or when removing or replacing the flow cell within its
subassembly.
To prevent contamination, use powder-free gloves when disassembling,
To prevent contamination, use powder-free gloves when disassembling,
Bracket
TP01463
36Maintenance
Required Materials
•TORX T10 screwdriver
•Small, flat-blade screwdriver
•Lens tissue or nonparticulating swab
•HPLC-grade methanol
•Belleville spring washer
•Flow cell gas ket
Procedure
To disassemble and clean the flow cell (and lenses):
1. Use the TORX T10 screwdriver to remove the three screws that secure one of the
lens holder assemblies (Figure 3-4
).
Screws
Flow Cell Disk
Lens Assembl y
Flow Cell
Body
Spring
Washer
Gasket
Slot for
Removing
Lens
Assembly
Lens Assembly
Flow Cell Disk
Belleville
Spring
Washer
TP01464
Figure 3-4 Disassembling the Flow Cell
2. Use the small, flat-blade screwdriver to gently pry the lens assembly from the
flow cell body at the slots.
3
Flow Cell Maintenance37
3
Attention:
STOP
norma l use, the gasket protects the lens holder from solvents.
3. Use a lens tissue or a nonpar ticulating swab to wipe the lens with methanol.
4. Remove and discard the gasket.
5. Repeat steps 1 through 4 to remove, disassemble, and clean the other lens
holder assembly.
6. Use methanol and a nonparticulating swab to clean the flow cell body.
Solvents other than methanol may damage a disassembled flow cell. In
Reassembling the Flow Cell
To reassemble the flow cell (see Figure 3-4):
1. Insert a replacement gasket into one side of the flow cell body.
2. Align the screw holes of the lens assembly with the holes in the flow cell body.
3. Place the new Belleville spring washers (wit h t h e c o ncave side facing out) onto
the lens assembly.
4. Place the flow cell disk over the lens assembly.
5. Insert the three screws using the TORX T10 driver to gradually tighten each
screw, alternating between the screws in a clockwise pattern. Tighten until the
screws meet the flow cell disk, then tighten each screw 1/4-turn. If a torque
screwdriver is available, tighten the screws to 16 in-oz (0.113 N-m).
STOP
Attention:
6. Repeat steps 1 through 5 to reassemble the other side of the flow cell.
Be careful not to overtighten the screws.
3.1.4 Installing the Flow Cell Assembly
Attention:
STOP
38Maintenance
operation. Be careful not to damage the flow cell body.
To install the flow cell assembly:
1. While you hold the flow cell assembly in a vertical orientation (see Figure 3-3
2. Gently push the front of the assembly until it seats on the front alignment pins.
3. Hand-tighten the thumbscrews.
4. Secure the fluidic connection bracket.
5. Reconnect the fluidic lines.
The alignment of the flow cell in the optics bench is critical to detector
the assembly into the optics bench. Note that the flow cell is self-aligning and uses
the guide pins on the optics bench.
), insert
6. Replace the front cover.
7. Flush the flow cell (refer to Section 3.1.1,
3.2 Replacing the Lamp
Replace the lamp in the 996 detector when either of the following conditions exists:
• Intensity is low enough that sensitivity is not sufficient for your method.
• The sampling rate requires an exposure time shorter than the minimum exposure
time you can set with the current lamp.
Before you replace the detector lamp, check the lamp usage indicator located to the front
of the lamp retainer, as described in Section 3.2.1,
3.2.1 Checking Lamp Usage
The Waters 996 detector lamp is designed to provide adequate energy for more than 1000
hours of operation. You can monitor lamp usage by checking the lamp usage indicator, a
mercury column with a scale of 0 to 10, where 10 represents 1000 hours. As the lamp
ages, the bubble in the mercury column moves toward the 10.
Caution:
disconnect the power cord before you begin this procedure. Note that the lamp and
housing are extremely hot. To avoid the possibility of contacting hot surfaces, allow the
lamp to cool for 15 minutes before you handle the lamp assembly or surfaces close to the
lamp.
To avoid electrical hazards and exposure to UV light, turn off the power and
Flushing the Flow Cell).
Checking Lamp Usage.
3
Procedure
To inspect the lamp usage indicator:
1. Power off the 996 detector, remove the pow er cord, and allow the lamp to cool for at
least 15 minutes.
2. Lift up the front panel cover and pull it away from the chassis.
3. Open the lamp access door by pulling t he t humbtab, then pulling the door toward
you (Figure 3-5
).
Replacing the Lamp39
Thumbtabs
LAMP
STATUS
Figure 3-5 Lamp Access Door
Lamp
Access Door
TP01461
3
4. Examine the lamp usage indicator (Figure 3-6
Figure 3-6 Lamp Usage Indicator
).
Lamp Us age
Indicator
TP01466
40Maintenance
If lamp intensity is low , but the lamp has not been used for 1000 hours, you
S
Note:
may be able to increase lamp intensity by cleaning the flow cell (see Section 3.1,
Flow Cell Maintenance).
Absorbance by the mobile phase also affects the apparent lamp intensity. For
example, acetonitrile is more transparent than methanol at wavelengths under 220
nm.
3.2.2 Removing the Lamp
Attention:
STOP
lamp glass damages the lamp and reduces life expectancy.
Attention:
TOP
the lamp.
Required Materials
TORX T20 screwdriver
Caution:
the 99 6 d etect or and disc onne ct the power cord.
Procedure
To replace the lamp in the 996 detector:
1. Power off the 996 detector, remove the power cord, and allow the lamp to cool for
at least 15 minutes.
Caution:
15 minutes after powering off the detector before you handle the lamp.
2. Lift up the front panel cover and pull it away from the chassis.
3. Open the lamp access door by pulling the thumbtab, then pull the door toward
you (see Figure 3-5
Do not touch the lamp glass while unpacking or inserting the lamp. Touching
To prevent contamination, use powder-free gloves when removing or replacing
To avoid electrical hazards when you perform the following procedure, power off
3
To avoid the possibility of contacting hot surfaces, wait at least
).
4. Disconnect the lamp power connector (Figure 3-7
).
Replacing the Lamp41
Lamp Alignment Notch
3
Lamp Mounting
Screws
Figure 3-7 Lamp Power Cord and Mounting Screws
5. Use the TORX T20 screwdriver to unscrew the two T20 lamp mounting screws.
6. Grip the metal base of the lamp, pull the lamp out, and set it aside.
7. Carefully unpack the replacement lamp.
8. While wearing powder-free gloves and holding the lamp by its base, orient the
lamp so that the notch in the base aligns with the positioning pin in the optics
bench.
Lamp Power
Connector
TP01467
9. Insert the lamp and secure it with the two T20 screws. Make sure that the lamp
base is flush against the lamp housing.
10. Reconnect the lamp power connector (see Figure 3-7
11. Close the lamp door and secure it with the thumbtab.
12. Install the front panel cover.
13. Reconnect the power cord and power on the 996 detector.
42Maintenance
).
3.3 Replacing the Fuses
Replace the fuses under the conditions indicated in the troubleshooting table (see Section
2.1, Start up Diagnostics). The 996 detector requires two 4 A, 250 V fuses (5 mm ×20
mm).
Caution:
cord before you perform the following procedure.
To avoid electrical hazards, power off the 996 detector and disconnect the power
Procedure
To replace the two fuses in the 996 detector:
1. Power off the 996 detector and remove the power cord.
2. Locate the fuse block above the power cord plug (Figure 3-8
Squeeze Side C li p s
to Access Fuses
Figure 3-8 Fuse Block
3. Squeeze the two side clips on the fuse block while you pull out the block.
4. Remove the fuses from the block, then install the new fuses.
) on the rear panel.
3
5. Orient the fuse block with the small tab pointing down, then push in the block
until the side clips engage.
6. Connect the power cord, then power on the 996 detector.
Replacing the Fuses43
4
Pr incipl es of the 996 PDA
Detector Optics
To use the Millennium32 PDA software effectively, you must be familiar with the principles
of operation of the optics and electronics of the Waters 996 PDA Detector.
4.1 996 Detector Optics
The 996 detector is an ultraviolet/visible light (UV/Vis) spectrophotometer with:
• 512 diodes
• Optical resolution of 1.2 nm per diode
• Operating wavelength range from 190 nm to 800 nm
4
The light path through the optics assembly of the 996 detector is shown in Figure 4-1
.
44Principles of the 996 PDA Detector Optics
Photodiode
Array
50-mm
Aperture
Assembly
Spectrographic
Mirror and Mask
Shutter
Flow Cell
Assembly
Lamp and
Lamp Optics
Figure 4-1 Optics Assembly Light Path
Grating
Beamsplitter
Assembly
Reference
Diode
Table 4-1
describes the optics assembly components in the 996 detector.
Table 4-1 Optics Assembly Components
ComponentFunction
Lamp and lamp
optics
Beamsplitter and
reference diode
Focuses light from the deuterium source lamp through a
beamsplitter to the flow cell.
Reflects par t of the light back to a reference diode, which
measures the intensity of the light emitted by the lamp. The
detector uses this measurement to keep the lamp output constant.
4
996 Detector Optics45
Table 4-1 Optics Assembly Components (Continued)
ComponentFunction
Flow cell
assembly
Spectrograph
mirror and mask
ApertureControls wavelength resolution and intensity of light striking the
Shutter assemblyPrevents light from reaching the photodiode array except during
GratingDisperses the light into bands of wavelengths and focuses those
Second-order
filter
Houses the segment of the flow path (containing eluent and
sample) through which the polychromatic light beam passes. This
arrangement of optical components, with the flow cell positioned
between the lamp and the grating, is commonly called reversed
optics.
The mirror focuses light transmitted through the flow cell onto the
aperture at the entrance to the spectrographic portion of the optics.
The mirror mask defines the beam of light focused on the
spectrograph mirror.
photodiodes. The width of the aperture is 50 µm.
sampling and calibration. For details on the dark current, see
Section 4.4.1,
wavelength bands onto the plane of the photodiode array.
Reduces the contribution of second-order reflection of UV light
(less than
350 nm) to the light intensity observed at visible wavelengths
(greater than 350 nm).
Calculating Absorbance.
4
Photodiode arrayAn array of 512 diodes arranged linearly. The diode width and
spacing provide a single wavelength resolution of 1.2 nm.
4.2 Resolving Spectral Data
The ability to distinguish similar spectra depends on photodiode spacing and the
bandwidth of the light striking the photodiode. The bandwidth of the light striking the
photodiodes depends on the aperture width.
The aperture width determines:
• Attainable wavelength bandwidth at the photodiode array
• Intensity of the light reaching the photodiode array (optical throughput)
46Principles of the 996 PDA Detector Optics
The aperture creates a narrow beam that reflects from the grating to the photodiode array.
The wavelength that strikes a particular diode depends on the angle of reflection from the
grating.
Figure 4-2
using the standard 50-µm aper ture. In this spectrum, the wavelength resolution is
sufficient to resolve five principal benzene absorption peaks.
shows an absorbance spectrum of benzene obtained from the 996 detector
Absorbance
nm
Figure 4-2 Benzene Spectrum at 1.2 nm Resolution
4.3 Measuring Light at the Photodiode
The Waters 996 Photodiode Array Detector measures the amount of light striking the
photodiode array to determine the absorbance of the sample in the flow cell.
The array consists of 512 photodiodes arranged in a row. Each photodi ode acts as a
capacitor by holding a fixed amount of charge.
Light striking a photodiode discharges the diode (Figure 4-3
discharge depends on the amount of light striking the photodiode.
Measuring Light at the Photodiode47
). The magnitude of the
4
Light from gra ti ng
dispersed in to 1.2 nm
wav elength beams
continuously
discharges diodes.
Grating
Sample in flow cell
absorbs at spe cific
wav elengths
100%
Charge
0 %
Deuterium Lamp
Flow
Cell
Mirror
4
Figure 4-3 Photodiodes Discharged by Light
The 996 detector measures the amount of current required to recharge each photodiode.
The current is proportional to the amount of light transmitted through the flow cell over the
interval specified by the diode exposure time.
Exposure Time
The 996 detector recharges each diode and reads the recharging current one diode at a
time. The interval between two readings of an individual diode is the exposure time. The
996 detector requires 11 msec to sequentially read all of the diodes in the array. The
minimum exposure time is 11 msec. You can set exposure time from 11 to 500 msec.
For example, if an exposure time is set to 50 milliseconds, the Waters 996 detector:
1. Recharges diode 1 and reads the current required to recharge diode 1.
2. Recharges diode 2 and reads the current required to recharge diode 2.
3. Sequentially recharges and reads the current required to recharge all the
remaining 510 photodiodes.
48Principles of the 996 PDA Detector Optics
4. After all of the diodes have been recharged and read (11 msec), the detector
waits 39 msec before beginning the recharge-and-reading sequence with
diode 1.
You set the exposure time parameter in the General tab of the 996 PDA Instrument
Method Editor. You can specify either Auto Exposure or Exposure Time. For details, refer
to the “Waters 996 PDA Detector Properties” topic in the
tab.
Millennium32 Online Help
Using the Auto Exposure Parameter
The Auto Exposure time parameter allows the 996 detector optics to calculate the
optimum exposure time needed to recharge the diodes based on lamp energy, lamp
spectrum, mobile phase absorbance, and the chosen wavelength range. To minimize
detector noise, Auto Exposure adjusts the exposure time to 80 to 90 percent of full scale.
The Auto Exposure time setting ensures that the photodiodes are:
• Not saturating due to overexposure
• Operating above the range of normal, dark current discharge
With auto exposure enabled, the 996 detector:
• Calculates exposure time at the start of a run based on maximum light intensity
within the wavelength range
• Limits the exposure so that no diode within the given wavelength range is
discharged more than 80%
Find
• Provides proper settings for signal-to-noise and dynamic range for each run
The Auto Exposure time setting may not support certain sampling rates or wavelength
ranges required for your analysis. If this is the case, you can set the exposure time
manually to adjust the exposure time from experiment to experiment.
Using the Exposure Time Parameter
The Exposure Time parameter enables you to manually set the length of time the
photodiodes are exposed to light before they are read. The supported range is 11 to
500 msec.
Be aware that increasing the Exposure Time parameter has the potential to saturate the
photodiodes. A longer exposure time may cause the 996 detector to lose the signal at
certain wavelengths because of diode saturation. When specifying the Exposure Time,
select a value that provides settings for an optimum signal-to-noise ratio over the
wavelength range of your analysis (see “
Optimizing the Signal-to-Noise Ratio” below).
Measuring Light at the Photodiode49
4
Optimizing the Signal-to-Noise Ratio
To optimize signal-to-noise ratios, choose an acquisition wavelength range that includes
only the wavelengths of interest and over which the mobile phase absorbs minimally (see
Appendix D,
Mobile Phase Absorbance).
4.4 Computing Absorbance Data Points
4
The 996 detector calculates absorbance values before transmitting the data to the
Millenn iu m
32
database. To calculate absorbance, the 996 detector:
• Computes the absorbance at each diode using the dark current and reference
spectrum (see Section 4.4.1,
• Aver ages the absorbances at a particular wavelength as specified in the spectra per
second sample rate and repor ts the average as a single data point (see Section
4.4.2, Resolution).
• Can apply a filter that acts like an analog filter (see Section 4.4.3,
4.4.1 Calculating Absorbance
The 996 detector computes absorbance by subtracting the dark current and reference
spectrum from the acquired spectrum . Absorbanc e is based on the pr inciple s of Beer's
Law.
Beer’s Law
The relationship between the quantity of light of a particular wavelength arriving at the
photodiode and the concentration of the sample passing through the flow cell is described
by the Beer-Lambert Law (commonly called Beer’s Law). Beer’ s Law i s ex pressed as
ε
A
=
lc
Where:
Calculating Absorbance).
Filtering Data).
A
= absorbance
ε
= molar absorptivity
l
= path length (1.0 cm in the 996 detector normal flow cell)
c
= molar concentration
50Principles of the 996 PDA Detector Optics
Beer’s Law applies only to well-equilibrated dilute solutions. It assumes that the refractive
index of the sample remains constant, that the light is monochromatic, and that no stray
light reaches the detector element. As concentration increases, the chemical and
instrumental requirements of Beer's law may be violated, resulting in a deviation from
(absorbance versus concentration) linearity (Figure 4-4
can reduce the linear range by the amounts shown in Appendix D,
Absorbance.
Absorbance
). The absorbance of mobile phase
Mobile Phase
Ideal
Actual
Working Ran ge
AbsorbanceBackground
Concentrati on
Figure 4-4 Absorbance as a Function of Concentration
Dark Current
Photodiodes lose charge over time even when they are not exposed to light. The amount
of charge lost is called
At the start of a chromatographic run, the 996 detecto r closes the shutter to take a dark
current reading for each diode. The shutter closes after the exposure time is calculated
and stays closed for the same interval as the exposure time.
The detector subtracts the dark current values from the current values recorded during
absorbance measurements for both the sample and the reference spectra.
dark current
.
Computing Absorbance Data Points51
4
Refer ence Spectrum
Immediately after the dark current measurement and before any components are eluted,
the 996 detector records a reference spectrum. The reference spectrum is a measure of
lamp intensity and mobile phase absorbance over the interval specified in the exposure
time taken with the shutter open.
For best results, the reference spectrum should be representative of the initial
Note:
mobile phase.
For extremely long exposure times, the dark current and reference spectrum
Note:
readings may take several minutes to finish.
Absorbance
The 996 detector calculates the absorbance for each diode at the end of each exposure
time using the following equation:
Sn Dn
–()
Absorbance
where:
S
= Signal obtained during sample analysis
D
= Signal obtained during the dark test
R
= Signal obtained from the reference spectrum
n
= Diode number
log=
-------------------------
Rn Dn
–()
4
4.4.2 Resolution
The data reported by the 996 detector to the Millennium32 database can be the average of
a number of data points. After calculating absorbance, the detector averages absorbance
values based on:
• Spectral resolution
• Sample rate
Averaging Spectral Data Based on Resolution
Spectral resolution (or bandwidth) is the wavelength interval (in nanometers) between
data points in an acquired spectrum. The minimal resolution of the 996 detector is 1.2 nm.
Spectral resolution with the 996 detector is always a multiple of 1.2 nm. For example, the
996 detector averages three diodes for each reported wavelength when the spectral
resolution is set in the Millennium
52Principles of the 996 PDA Detector Optics
32
software to 3.6 nm.
Averaging Chromatographic Data Based On Sample Rate
Sample rate is the number of data points per second reported to the Millennium32
database. The number of times the photodiodes are read during the sample rate interval is
dependent on the exposure time.
For ex ample, if e xposure time is 25 msec, and sample rate is 1 second, then readings per
data point are
1000 msec
-------------------------40=
25 msec
The readings are averaged and reported as a single data point.
Combining Spectral Resolution and Sample Rate
Spectral resolution and sample rate have opposite effects on noise and spectral detail.
Increasing the value of the spectral resolution parameter and decreasing the number of
spectra per second decrease the size of the data file.
The data storage rate is based on wavelength range, spectral resolution, and
Note:
sample rate, which are set in the General tab of the 996 PDA Instrument Method Editor.
For details, ref er to the “Waters 996 PDA Detector Properties” topic in the
Millennium
32
Online Help Find tab.
4.4.3 Filtering Data
Use the Channel 1 tab of the 996 PDA Instrument Method Editor (for details, refer to the
“Waters 996 Detector Properties” topic in the
an optional noise filter (the Filter Response parameter) to the dat a sent to the Mi llennium
software database. A noise filter of 1 second is the default value, which provides a good
signal-to-noise ratio for most chromatographic separations.
Note the following with regard to filtering data:
• The noise filter is similar in function to an analog RC filter.
• The filter calculates a data point that is a modified rolling average for a wavelength
over a number of readings.
• The filter values are comparable to the effects of a 1-, 2-, or 3-second RC filter.
Millennium32 Online Help
Computing Absorbance Data Points53
Find tab) to apply
32
4
5
Spectral Contrast Theory
This chapter explains the theory behind the Spectral Contrast technique, which is used to
compare UV/Vis absorbance spectra collected by the 996 detector. Spectral Contrast
makes use of the fact that different compounds have differently shaped absorbance
spectra. This chapter describes how Spectral Contrast represents absorbance spectra as
vectors. When applied to the UV/Vis absorbance data collected by the 996 detector, the
Spectral Contrast technique determines whether differences between spectra are due to
the presence of multiple compounds in the same peaks (coelution) or due to nonideal
conditions such as noise, photometric error, or solvent effects.
5.1 Comparing Absorbance Spec tra
The shape of an absorbance spectrum is determined by the relative absorbance at
different wavelengths. The shape of a compound’s absorbance spectrum is a
characteristic of that compound at the solvent and pH conditions under which the
absorbance spectrum is measured.
5
Figure 5-1
absorbance at 245 nm to the absorbance at 257 nm is approximately 2.2 for compound A
and 0.7 for compound B.
The absorbance ratios of two wavele ngth pairs is a limited spectral comparison. For more
information, you need to compare the absorbance ratios of multiple wavelength pairs.
54Spectral Contrast Theory
shows the absorbance spectra for the two compounds, A and B. The ratio of the
Compound A:
245 nm
257 nm
Compound A
Compound B:
Compound B
Figure 5-1 Comparing Spectra of Two Compounds
5.2 Representing Spectra as Vectors
Ab
245
--------------2.2=
Ab
257
Ab
245
--------------0.7=
Ab
257
The Spectral Contrast technique uses vectors to quantify differences in the shapes of
spectra. Spectral Contrast converts baseline-corrected spectra to vectors and then
compares the vectors. Spectral vectors have two properties:
Length
•
Direction
•
– Proportiona l to analyte concentration.
– Determined by the relative absorbance of the analyte at all wavelengths
(its absorbance spectrum). Direction is independent of concentration.
Vector direction contributes to the identification of a compound, since the direction is a
function of the absorbance spectrum of the compound. The ability of spectral vectors to
differentiate compounds depends on the resolution of spectral features. As both
wavelength range and spectral resolution increase, the precision of a spectral vector for
the resultant spectrum increases. A vector derived from the Waters 996 PDA Detector can
include absorbances in any range between 190 nm and 800 nm with a spectral resolution
of 1.2 nm.
Representing Spectra as Vectors55
5
5.2.1 Vectors Derived from Two Wavelengths
The Spectral Contrast algorithm uses vectors to characterize spectra (Figure 5-2). To
understand the vector principle, consider two vectors (Figure 5-2
depicted in Figure 5-1
AU at 257 nm
.
AU at 245 nm
Figure 5-2 Plotting Vectors for Two Spectra
) based on the spectra
5
The axes in Figure 5-2
the absorbance ratio shown in Figure 5-1
intersection of the absorbance values (for Compound A) at the two wavelengths
represented by each axis. The other vector is similarly derived for the spectrum of
Compound B.
The vector for Compound B points in a direction different from that of the vector for
Compound A. The difference in direction, which reflects the difference in the absorbance
ratios of the two compounds at wavelengths 245 nm and 257 nm, is called the Spectral
Contrast Angle. A Spectral Contrast angle (e.g., θ in Figure 5-2
indicates a shape difference between spectra (see Section 5.3,
The length of the vector is proportional to the concentration.
are in absorbance units at the two wavelengths used to calculate
. The head of the vector for Compound A is at the
5.2.2 Vectors Derived from Multiple W av elengths
When absorbance ratios are limited to two wavelengths, the chance that two different
spectra will have the same absorbance ratio is much greater than if comparison is made
using absorbance ratios at many wavelengths. Therefore, the Spectral Contrast technique
uses absorbances from multiple wavelengths to form a vector in an
n
space, where
56Spectral Contrast Theory
is the number of wavelengths from the spectrum.
)
greater than zero
Spectral Contrast Angles).
n
-dimensional vector
To compare two spectra, the Spectral Contrast technique forms a vector for each spectrum
n
in an
compute the angle between the two vectors.
Just as in the 2-wavelength comparison, a Spectral Contrast angle of zero in
n
match. Conversely, if any comparison of ratios does not match, then the corresponding
vectors point in different directions.
-dimensional space. The two spectral vectors are compared mathematically to
-dimensional space means that all ratios of absorbances at corresponding wavelengths
5.3 Spectral Contrast Angles
Spectra that have the same shape have vectors that point in the same direction. Spectra
that have different shapes have vectors that point in different directions. The angle
between the two vectors of any two spectra, the Spectral Contrast angle, quantifies the
magnitude of the shape difference between the spectra. The
the difference in direction between the spectral vectors of two spectra.
A Spectral Contrast angle can vary from 0 degrees to 90 degrees. A Spectral Contrast
angle near 0 degrees indicates little shape difference between the compared spectra.
Matching a spectrum to itself produces a Spectral Contrast angle of exactly 0 degrees.
The maximum Spectral Contrast angle, 90 degrees, indicates that the two spectra do not
overlap at any wavelength.
Spectral Co ntrast an gle
is
To illustrate the relationship between Spectral Contrast angle and spectral shape
differences, consider the pairs of spectra shown in Figure 5-3
, Figure 5-4, and Figure 5-5.
Spectra with Different Shapes
In Figure 5-3, the absorbance spectra of two compounds, A and B, are distinctly different,
and therefore, have a large Spectral Contrast angle (62.3°).
Spectral Contrast Angles57
5
Compound B
Normalized Absorbance
Figure 5-3 Spectra with a Large Spectral Contrast Angle
Spectra with Similar Shapes
Spectral Contrast Angle: 62.3°
Compound A
Wavelength (nm)
5
In Figure 5-4, the absorbance spec tra of two compounds, A and B, are similar, and
therefore, have a small Spectral Contrast angle (3.0°).
58Spectral Contrast Theory
Spectral Cont rast Angle: 3.0
Compound A
Compound B
Normalized Absorba nce
Wavelength (nm)
Figure 5-4 Spectra with a Small Spectral Contrast Angle
°
Differences Between Spectra of the Same Compound
Small but significant differences between absorbance spectra can occur because of
factors other than those due to the absorbance properties of different compounds. For
same
example, multiple spectra of the
of detector noise, photometric error, high sample concentration, or variations in solvent
conditions. The spectra in Figure 5-5
the shape of an absorbance spectrum of one compound. This effect is most likely to occur
at low concentrations where the signal-to-noise ratio is lo w . Note that the Spectral Contrast
angle between these absorbance spectra of the same compound is 3.4°.
compound may exhibit slight differences because
, for example, show how instrument noise can affect
Spectral Contrast Angles59
5
Normalized Spectra of a Compound at Different
Concentratio ns
Nor m alized Ab so r b ance
Figure 5-5 Absorbance Spectra of a Compound at Two Concentrations
5.4 Nonidealities
Shape differences between absorbance spectra can be caused by one or more of the
following nonideal phenomena:
• Detector noise
• Photometric error caused by high sample concentration
Spectral Contrast Angle: 3.4
Wavelength (nm)
°
5
• V ariation in solvent composition
These sources of spectral variation can cause chemically pure, baseline-resolved peaks
to exhibit a small level of spectral inhomogeneity. You c an assess the significance of
spectral inhomogeneity by comparing a Spectral Contrast angle to a
Section 5.4.4).
5.4.1 Detector Noise
Statistical and thermal variations add electrical noise to the absorbance measurem ent s
made by the 996 detector. The noise manifests itself as fluctuations in the baseline, known
baseline noise
as
thermal variations can be predicted from the instrument noise in the baseline region of a
chromatogram.
60Spectral Contrast Theory
. The magnitude of any absorbance differences caused by statistical and
Threshold angle
(see
5.4.2 Photometric Error
At high absorbances (generally greater than 1 AU), a combination of effects can produce
slight departures (about 1%) from Beer’s Law due to photometric error. Although
photometric errors at this level may have a negligible effect on quantitation, they can be a
significant source of spectral inhomogeneity . To minimize the effects of photometric error
for all Spectral Contrast operations, the maximum spectral absorbance of a compound
should be less than 1 AU. Keep in mind that the absorbance of the mobile phase reduces
the working linear dynamic range by the amount of mobile phase absorbance at each
wavel ength. For e xamples of mobile phase absorbance, see Appendix D,
Absorbance.
For more information about the effects of the photometric error curve, refer to
Note:
Principles of Instrumental Analysis, third edition, by Douglas A. Skoog, Saunders College
Publishing,1985, pp 168–172.
5.4.3 Solvent Changes
As long as solvent concentration and composition do not change (isocratic operation), the
background absorbance, if any, by the solvent remains constant. A change, however, in
solvent pH or composition, such as occurs in gradient operation, can affect the intrinsic
spectral shape of a compound, as shown in Figure 5-6
5.4.4 Threshold Angle
Mobile Phase
.
In addition to computing Spectral Contrast angles, the Spectral Contrast technique also
computes a Threshold angle. The Threshold angle is the maximum Spectral Contrast
angle between spectra that can be attributed to nonideal phenomena.
Comparison of a Spectral Contrast angle to its Threshold angle can assist in determining if
the shape difference between spectra is genuine, that is, generated by mixtures that are
dissimilar. In general, a Spectral Contrast angle less than its Threshold angle indicates
that shape differences can be attributed to nonideal phenomena alone, and that there is
no evidence for genuine differences between the spectra. A Spectral Contrast angle
greater than its Threshold angle indicates that the shape differences are due to genuine
differences between the spectra.
Nonidealities61
5
Effect of pH
Effect of pH
p
-Aminobenzoic Acid
p - Aminobenzoic Acid
pH 3.1pH 5.1p H 6.9
+
NH
3
COOH
NH
2
COOH
pH 6.9
NH
COO–
COOH
2
Absorbance
pH 3.1
200220240260280300320340
nM
Effect of Acetonitrile Concentration
Effect of Acetonitrile Concentrationp - Aminobenzoic Acid
p
-Aminobenzoic Acid
2–10% Acetonitrile
Absorbance
Wa vel ength
pH 5.1
NH
2
COOH
5
200220240260280300320340
Figure 5-6 Effects of pH and Solvent Concentration
on the Absorbance Spectrum of
62Spectral Contrast Theory
Wavelength (nm)
p
-Aminobenzoic Acid
Append ix A
Detector Specifications
Table A-1 lists the Waters 996 PDA Detector specifications.
Table A-1 996 Detector Specifications
ItemSpecification
Dimensions11.5 × 22 inches (29 × 56 cm)
Weight31 lbs (14.3 kg)
Wavelength range190 to 800 nm
Wavelength accuracy±1 nm
Linearity range
Spectral resolution1.2 nm
Baseline noise±1.5 × 10
Drift1 × 10
1
5% at 2.0 AU, propylparaben,
at 258 nm
–5
AU peak-to-peak,
dry,
at 254 nm
–3
AU/hour
at 254 nm (after warmup)
A
Flow cells
Standard
Semi-preparative
Microbore
Inert
1
Per ASTM 685-79
Pathlength (mm): Tubing (I.D . )
10 0.009 in
3 0.040 in
30.005 in
100.010 in
Appendix A63
Append ix B
Spare Parts
The s par e parts lis ted in Table B-1 are those parts recommended for customer installation.
Damage incurred by performing unauthorized work on your 996 detector may invalidate
certain warranties.
Table B-1 Spare Parts
Item Part Number
Flow cell, standardWAT057919
Flow cell, semi-preparativeWAT057463
Flow cell, microboreWAT057462
Flow cell, inertWAT057461
Gasket, flow cell ( 2 )WAT057924
Belleville wash er (2)WAT05 7925
Lens mount and lens (2)WAT057923
Semi-prep lens kitWAT057968
Deuterium lampWAT057760
Fuse, 4 A (5 × 20 mm)WAT057337
Waters Erbium Perchlorat e Wavelength
Accuracy Solution
WAT042885
B
Waters Absorbance Detector Linearity
Solution
Waters 996 PDA Detector Qualification
Workbooks
WAT042881
WA T509-01
Appendix B64
Append ix C
Warranty Information
This appendix includes information on:
• Limited express warranty
• Shipments, damages, claims, and returns
C.1 Limited Express Warranty
Waters® Corporation provides this limited express warranty (the Warranty) to protect
customers from nonconformity in the product workmanship and materials. The Warranty
covers all new products manufactured by Waters.
Waters warrants that all products that it sells are of good quality and workmanship. The
products are fit for their intended purpose(s) when used strictly in accordance with Waters’
instructions for use during the applicable warranty period.
Limited Warranty
C
Waters Corporation warrants that the Waters 996 PDA Detector is a Class I medical
device under 21 CFR 862.2260, as now in effect, and is for general purpose use and is not
for use in clinical diagnostic procedures, and that during the Warranty period, the
performance of all components of the Waters 996 PDA Detector [other than Third-Party
Components (non-Waters named)], will not deviate materially from the Specifications for
such detectors. Warranties, if any, that may be applicable to Third-Party Components shall
be provided by the respective manufacturers or suppliers of such Third-Party
Components, and Waters Corporation shall use reasonable efforts to assist Customer in
securing the benefits of any such warranties.
Exclusions
The foregoing warranty does not apply to any material deviation from the Specifications by
any component of the Waters 996 PDA Detector that results from (a) use of the Waters
996 PDA Detector for any purpose other than general purpose use and sp ecifically
excluding use of the Waters 996 PDA Detector in clinical diagnostic procedures, or use of
the Waters 996 PDA Detector for investigational use with or without confirmation of
diagnosis by another, medically established diagnostic product or procedure, (b) errors or
defects in any
Appendix C65
Third-Party Component, (c) modification of the Waters 996 PDA Detector by anyone other
than Waters Corporation, (d) failure by Customer to install any Standard Enhancement in
accordance with an update procedure, release of firmware or any operating system
release, (e) any willful or negligent action or omission of Customer, (f) any misuse or
incorrect use of the Waters 996 PDA Detector, (g) any malfunction of any information
system or instrument with which the Waters 996 PDA Detector may be connected, or (h)
failure to establish or maintain the operating environment for the W aters 996 PDA Detector
in accordance with the operator’s manual.
Exclusive Remedy
In the event of any failure of the Waters 996 PDA Detector to perform, in any material
respect, in accordance with the warranty set forth herein, the only liability of Waters
Corporation to Customer, and Customer’s sole and exclusive remedy, shall be the use, by
Waters Corporation, of commercially reasonable efforts to correct for such deviations, in
Waters Corporation’s sole discretion, replacement of the purchased Waters 996 PDA
Detector, or refund of all amounts theretofore paid by Customer to Waters Corporation for
the Waters 996 PDA Detector.
Disclaimers
C
THE LIMITED W ARRANTY SET FO RTH HEREIN IS EXCLUSIVE AND IN LIEU OF, AND
CUST OMER HER EBY W AI VES , ALL O THER R EPRES ENTATIONS, W ARRANT IES AN D
GU AR ANTEES, EXPRESS OR IMPLIED, IN C LUDING WITHOUT LIMITATION ANY
IMPLIED WARRANTIES OF MERCHANTABILITY, FITNESS OF THE WATERS 996 PDA
DETECTOR FOR A PARTICULAR PURPOSE, IN CL U D ING FITNESS FOR USE IN
CLINICAL DIAGNOSTIC PROCEDURES OR FOR INVESTIGATIONAL USE WITH OR
WITHOUT CONFIRMATION OF DIAGNOSIS BY ANOTHER MEDICALLY ESTABLISHED
DIAGNOSTIC PRODUCT OR PROCEDURE, OR NONINFRINGEMENT, AND ANY
WARRANTIES ARISING OU T OF CO UR S E OF DEAL IN G OR COU RSE OF
PERFORMANCE. CUSTOMER EXPRESSLY ACKNOWLEDGES THAT BECAUSE OF
THE COMPLEX NATURE OF THE WATERS 996 PDA DETECTOR AND ITS
MANUFACTURE, WATERS CORPORATION CANNOT AND DOES NOT W ARRANT
THAT THE OPERATION OF THE WATERS 996 PDA DETE CTOR WILL BE WITH OUT
DEFECT. CUSTOMER EXPRESSLY ACKNO WLEDGES THAT CUSTOMER IS SOLELY
RESPONSIBLE FOR USE OF THE WATERS 996 PDA DETECTOR IN CLINCIAL
DIAGNOSTIC PROCEDURES OR FOR INVESTIGATIONAL USE WITH OR WITHOUT
CONFIRMATION OF DIAG N O SIS BY AN OTHER MEDICALLY ESTABLISHED
DIAGNOSTIC PRODUCT OR PROCEDURE.
66 Appendix C
Wa rr anty Service
Warranty service is performed at no charge and at Waters’ option in one of three ways:
• A service representative is dispatched to the customer facility.
• The product is repaired at a Waters repair facility.
• Replacement part s with appropriate installation instructions are sent to the
customer.
Nonconforming products or parts are repaired, replaced with new or like-new parts, or
refunded in the amount of the purchase price, when the product is returned. Warranty
service is performed only if the customer notifies Waters during the applicable warranty
period.
Unless otherwise agreed at the time of sale, warranty service is not provided by
dispatching a service represent ative when the equipment has been removed from the
initial installation location to a new location outside the home country of the selling
company.
AM
Warranty service is provided during business hours (8
Friday). Service is not available when Waters offices are closed in observance of legal
holidays.
to 5PM, EST, Monday through
Wa rr anty Service Exceptions
Warranty service is not performed on:
C
• Any product or part that has been repaired by others, improperly installed, altered,
or damaged in any way.
• Products or parts identified prior to sale as not manufactured by Waters. In such
cases, the warranty of the original manufacturer applies.
• Products that malfunction because the customer has failed to perform maintenance,
calibration checks, or observe good operating procedures.
• Products that malfunction due to the use of unapproved parts and operating
supplies.
Repair or replacement is not made:
• For expendable items such as gaskets, windows, lenses, and fuses, if such items
were operable at the time of initial use.
• Because of decomposition due to chemical action.
• For used equipment.
• Because of poor facilities, operating conditions, or utilities.
Appendix C67
Wa rr anty Period
The warranty period begins when the product is installed or, in the case of a customer
installation, 15 days after shipment from Waters.
In no case does the warranty period extend beyond 15 months from date of shipment. If
an item is replaced during its warranty period, the replacement par t is warranted for the
balance of the original warranty period. Table C-1
applicable components.
Table C-1 Warranty Periods
ItemWarranty
Waters 996 PDA Detector1 year
Deuterium lamp1000 hours
summarizes the warranty periods for
C
Gaskets
Windows
Lenses
Fuses
Plunger seals
Tubing and fittings
Not
warranted
items
Lamp Replacement Warranty
The Waters 996 PDA Detector deuterium lamp is warranted to light and pass powerup
verification tests for 1000 hours.
68 Appendix C
C.2 Shipments, Damages, Claims, and Returns
Shipments
As all shipments are made Free On Board (FOB) shipping point, we suggest insurance be
authorized on all shipments. Instruments and major component s are packed and shipped
via surface, unless otherwise required. Supplies and/or replacement parts are packed and
shipped via United Parcel Service (UPS), UPS Blue, air parcel post, or parcel post unless
otherwise requested.
Damages
The Interstate Commerce Commission has held that carriers are as responsible for
concealed damage as for visible damage in transit. Unpack shipment promptly after
receipt as there may be concealed damage even though no evidence of it is apparent.
When concealed damage is discovered, cease further unpacking of the unit involved and
request immediate inspection by local agent or carrier and secure written report of his
findings to support claim. This request must be made within 15 days of receipt. Otherwise,
the claim will not be honored by the carrier. Do not return damaged goods to the factory
without first securing an inspection report and contacting Waters for a return merchandise
authorization number (RMA).
Claims
C
After a damage inspection report is secured, Waters cooperates fully in supplying
replacements and handling of a claim which may be initiated by either party.
Returns
No returns may be made without prior notification and authorization. If for any reason it is
necessary to return material to Waters, please contact Waters Customer Service or your
nearest Waters subsidiary or representative for a return merchandise authorization (RMA)
number and forwarding address.
Appendix C69
Append ix D
Mobile Phase Absorban ce
This appendix provides a list of the absorbances at several wavelengths for commonly
used mobile phases. Choose your mobile phase carefully to reduce baseline noise.
The best mobile phase for your application is one that is transparent at the chosen
detection wavelengths. Such a mobile phase ensures that any absorbance is due only to
the sample. Absorbance by the mobile phase also reduces the linear dynamic range of the
detector by the amount of absorbance that is autozeroed out. Wavelength, pH and
concentration of the mobile phase will affect its absorbance. Examples of several mobile
phases are provided below.
Table D-1 Mobile Phase Absorbance Measured Against Air or Water
claim s
damages
disclaimers
exclusions
information
lamp replacement
limited
period
returns
ser vi c e
service exceptions
Waters 996
absorbance calculations
aperture width
dark current
detector optics, overview
hardware theory
photodiode array overview
reference spectrum
spare parts
specifications
spectral resolution
Waters liability
Waters Technical Service, contacting
Wa velength
accur acy
derived vectors
mobile phase absorbances
69
69
66
65
65–68
68
65
68
69
67
67
50, 52
47
51
44–46
44–53
47
52
64
63
47
66
31
56, 56
70
27, 31
V
Vectors
derived from multiple wavelengths
derived from two wavelengths
spectra, representing
spectral contrast
76Index
55
55
56
56
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.