Walchem WCT400 Series, WCT410 Series Instruction Manual

Page 1
W
W A L C H E M
IWAKI America Inc.
CT400/410 Controllers
WCT400/410 Series
Cooling Tower Conductivity Controller
Instruction Manual
Five Boynton Road Hopping Brook Park Holliston, MA 01746 USA
TEL: 508-429-1110 FAX: 508-429-7433 WEB: www.walchem.com
Page 2
Notice
© 2014 WALCHEM, Iwaki America Inc. (hereinafter “Walchem”) 5 Boynton Road, Holliston, MA 01746 USA (508) 429-1110 All Rights Reserved Printed in USA
Proprietary Material
The information and descriptions contained herein are the property of WALCHEM. Such information and descriptions may not be copied or reproduced by any means, or disseminated or distributed without the express prior written permission of WALCHEM, 5 Boynton Road, Holliston, MA 01746.
This document is for information purposes only and is subject to change without notice.
Statement of Limited Warranty
WALCHEM warrants equipment of its manufacture, and bearing its identification to be free from defects in workmanship and material for a period of 24 months for electronics and 12 months for mechanical parts and electrodes from date of delivery from the factory or authorized distributor under normal use and service and otherwise when such equipment is used in accordance with instructions furnished by WALCHEM and for the purposes disclosed in writing at the time of purchase, if any. WALCHEM's liability under this warranty shall be limited to replacement or repair, F.O.B. Holliston, MA U.S.A. of any defective equipment or part which, having been returned to WALCHEM, transportation charges prepaid, has been inspected and determined by WALCHEM to be defective. Replaceable elastomeric parts and glass components are expendable and are not covered by any warranty.
THIS WARRANTY IS IN LIEU OF ANY OTHER WARRANTY, EITHER EXPRESS OR IMPLIED, AS TO DESCRIPTION, QUALITY, MERCHANTABILITY, FITNESS FOR ANY PARTICULAR PURPOSE OR USE, OR ANY OTHER MATTER.
180324. K April 2014
Page 3
TABLE OF CONTENTS
1.0 INTRODUCTION ........................................................................................................................ 1
2.0 SPECIFICATIONS ..................................................................................................................... 2
2.1 Measurement Performance .................................................................................................... 2
2.2 Electrical: Input/Output ........................................................................................................... 2
2.3 Mechanical ............................................................................................................................. 2
2.4 WCT Variables and their Limits .............................................................................................. 3
3.0 UNPACKING & INSTALLATION ................................................................................................ 4
3.1 Unpacking the unit ................................................................................................................. 4
3.2 Mounting the electronic enclosure ......................................................................................... 4
3.3 Installation .............................................................................................................................. 4
3.4 Icon Definitions ....................................................................................................................... 4
3.5 Electrical installation ............................................................................................................... 8
4.0 FUNCTION OVERVIEW .......................................................................................................... 12
4.1 Front Panel ........................................................................................................................... 12
4.2 Display ................................................................................................................................. 12
4.3 Keypad ................................................................................................................................. 13
4.4 Access Code ........................................................................................................................ 13
4.5 Startup .................................................................................................................................. 13
4.6 Shut Down ............................................................................................................................ 13
5.0 OPERATION ............................................................................................................................ 14
5.1 Main Menu ........................................................................................................................... 14
5.2 Conductivity Menu ................................................................................................................ 16
5.3 Temperature Menu ............................................................................................................... 18
5.4 Bleed Menu .......................................................................................................................... 19
5.5 Feed Menu ........................................................................................................................... 21
5.6 WM1 and WM2 Menus ......................................................................................................... 24
5.7 Bio1 and Bio2 Menus ........................................................................................................... 25
5.8 Time Menu ........................................................................................................................... 29
5.9 Conductivity Alarm Menu ..................................................................................................... 30
5.10 4-20mA Menu ....................................................................................................................... 31
5.11 Access Code Menu .............................................................................................................. 32
5.12 Datalog Menu ....................................................................................................................... 33
5.13 Config Menu ......................................................................................................................... 34
5.14 Upgrade Menu ..................................................................................................................... 35
6.0 MAINTENANCE ....................................................................................................................... 36
6.1 Electrode Cleaning ............................................................................................................... 36
6.2 Replacing the Fuses ............................................................................................................ 37
7.0 TROUBLESHOOTING ............................................................................................................. 37
7.1 Error Messages .................................................................................................................... 37
7.2 Conductivity Readout Does Not Change ............................................................................. 38
7.3 Procedure for Evaluation of Conductivity Electrode ............................................................. 39
8.0 SERVICE POLICY ................................................................................................................... 39
Page 4
1.0 INTRODUCTION
The Walchem WCT400 Series controllers offer conductivity and watermeter-based control of cooling tower water, control of corrosion/scale inhibitor feed, and with the WCT410 models, control of two biocide pumps. The inhibitor pump may be selected to operate in one of the following modes: Feed and Bleed Feed and Bleed with Lockout Feed as a percent of Bleed Feed as a percent of Time Feed based on a Water Contactor input Feed based on a Paddlewheel water meter input
The WCT series cooling tower controllers are supplied with a temperature compensated carbon electrode with a cell constant of 1.0. The controllers are microprocessor driven industrial type with on/off control outputs. A timed sample mode may be selected, and on small towers can reduce installation costs by eliminating the need for a sampling bypass line. An optional isolated 4-20 mA output that is proportional to the conductivity reading is available for all models.
Any set point may be viewed without interrupting control. Each set point change will take effect as soon as it is entered. An access code is available to protect set point parameters, while still allowing settings to be viewed.
The model WCT410 biocide outputs are scheduled on a user selectable 1, 2, or 4 week cycle. Each biocide output is independent and may be programmed for one add per day. (in 1, 2 or 4 week cycle) or up to 10 times per day (in daily cycle). It is possible to add both chemicals in the same day, however, as a safeguard they will not occur at the same time. Biocide outputs may use independent pre-bleeds and lockout times. All outputs are interlocked with a flow switch input.
An alarm relay is provided with WCT410 models. It is triggered by: Low Conductivity reading High Conductivity reading No Flow situation Bleed Timeout Sensor Error Temp Error
Our unique USB feature provides the ability to upgrade the software in the controller to the latest version.
An advanced USB capability option is available. The Config file feature allows you to save all the set points from a controller onto a USB flash disk, and then import them into another controller, making the programming of multiple controllers fast and easy. The data logging feature allows you to save the last 2 month’s readings and events to a USB flash disk.
1
Page 5
2.0 SPECIFICATIONS
2.1 Measurement Performance
Conductivity Range 0 - 10,000 µS/cm (microSiemens/centimeter)
Conductivity Resolution 1 µS/cm
Conductivity Accuracy 10 - 10,000 µS/cm ±1% of reading
Temperature Range 32 - 158°F (0 - 70°C)
Temperature Resolution 0.1°C
Temperature Accuracy ± 1% of reading
2.2 Electrical: Input/Output
Input Power
Input Signals
Conductivity electrode 1.0 cell factor, 10K thermistor
Flow Meter (optional) Isolated, dry contact closure required (i.e. relay, ,reed switch)
Flow Switch (optional) Isolated, dry contact closure required (i.e. reed switch)
Outputs
Mechanical Relays Pre-powered on circuit board switching line voltage
4 - 20 mA (optional) Internally powered
Agency Approvals
Safety UL 61010-1:2012 3rd Ed.
CSA C22.2 No. 61010-1:2012 3rd Ed. IEC 61010-1:2010 3rd Ed. EN 61010-1:2010 3rd Ed.
EMC IEC 61326-1:2005 EN 61326-1:2006 Note: For EN61000-4-6, EN61000-4-3 the controller met performance criteria B. *Class A equipment: Equipment suitable for use in establishments other than domestic, and those directly connected to a low voltage (100-240 VAC) power supply network which supplies buildings used for domestic purposes.
0 - 10 µS/cm ±20% of reading
100-240 VAC, 50/60 Hz, 8A Fuse: 1.0 ampere, 5 x 20 mm
6 A (resistive), 1/8 HP All relays are fused together as one group, total current for this group must not exceed 6A
Fully isolated 600 Ohm max resistive load Resolution .001% of span Accuracy ± 1% of reading
2.3 Mechanical
Enclosure Material Polycarbonate
Operating Ambient Temp 32 – 122°F (0 – 50°C)
Storage Temperature -20 – 180°F (-29 – 80°C)
Graphite electrode pressure rating 150 psi
Stainless steel electrode pressure rating 150 psi
High pressure electrode & flow switch
manifold pressure rating
Flow switch manifold pressure rating 150 psi
Flow switch manifold connections ¾" NPTF
NEMA Rating NEMA 4X
Dimensions 8.5" x 6.5" x 5.5"
Display 2 x 16 character backlit liquid crystal
300 psi
2
Page 6
2.4 WCT Variables and their Limits
Low Limit High Limit
Conductivity menu
Temperature Menu
Bleed Menu
Set Point 0 µS/cm 10,000 µS/cm Dead Band 5 µS/cm 500 µS/cm Bleed Limit Time (set in hours/minutes) 1 minute 8 hrs: 20 min (enabled)
Bleed Volume 1 (gallons or liters) 9,999 (gallons or liters)
Feed Menu
Feed Lockout Timer (Mode A) 1 second 99 min: 59 sec Percent of Bleed (Mode B) 5 % 99 % Feed Time Limit (Mode B) 1 minute 99 min: 59 sec Percent of Time (Mode C) 0.1 % 99 % Feed Cycle Time (Mode C) 10 minutes 59 min: 59 sec Time per Contact (Mode D) 1 second 59 min: 59 sec ÷ Contacts by (Mode D) 1 contact 100 contacts Time Limit (Mode D & E) 1 minute 99 min: 59 sec Time/Vol (Mode E) 1 second 59 min: 59 sec Vol to Initiate Feed (Mode E) 1 9,999
WM1 and WM2 Menus
Gallons per Contact 1 gal/contact 500 gal/contact Liters per Contact 1 L/contact 500 L/contact K Factor (Paddlewheel) 0.01 pulse/vol 9,999.99 pulses/vol
Biocide Menus (WCT410 only)
Pre-Bleed (0 disables pre-bleed) 1µS/cm 9.999 µS/cm Lockout 0 minutes 9 hrs:59 min Add Time Daily Adds 0 minutes 144 minutes All other modes 0 minutes 1440 minutes
4-20 mA Menu 4 & 20 mA Settings 0 µS/cm 10,000 µS/cm
Access Code Menu New Value 0 9999
Alarm Menu* High & Low (set to zero to disable) 1% 50%
Datalog Menu (Optional)
Config Menu (Optional)
Upgrade Menu
*Note: The Alarm relay is non-programmable. Refer to the Main Menu diagram on page 16 for the list of error conditions that trigger the alarm relay.
PPM Conversion Factor 0.200 ppm/µS/cm 1.000 ppm/µS/cm
Interval Time (sampling) 5 minutes 24:00 hours
Duration Time (sampling) 1 minute 59 min: 59 sec
% Calibration Range -50 +50
Makeup Volume 1 (gallons or liters) 9,999 (gallons or liters)
No variables
0=unlimited (disabled)
No variables
No variables
No variables
3
Page 7
3.0 UNPACKING & INSTALLATION
3.1 Unpacking the unit
Inspect the contents of the carton. Please notify the carrier immediately if there are any signs of damage to the controller or its parts. Contact your distributor if any of the parts are missing. The carton should contain: a WCT400 series controller and instruction manual. Any options or accessories will be incorporated as ordered.
3.2 Mounting the electronic enclosure
The WCT series controller is supplied with mounting holes on the enclosure. It should be wall mounted with the display at eye level, on a vibration-free surface, utilizing all four mounting holes for maximum stability. Use M6 (1/4" diameter) fasteners that are appropriate for the substrate material of the wall. The enclosure is NEMA 4X rated. The maximum operating ambient temperature is 122°F (50°C); this should be considered if installation is in a high temperature location. The enclosure requires the following clearances: Top: 2" (50 mm) Left: 8" (203 mm) Right: 4" (102 mm) Bottom: 7" (178 mm)
3.3 Installation
Once the WCT series controller is mounted, the metering pumps may be located at any distance from the controller. The conductivity electrode should be placed as close to the controller as possible, to a maximum distance of 250 ft. Less than 25 ft is recommended. The cable must be shielded from background electrical noise. Always route low voltage (sensor) signals with at least a 6” separation from AC voltage wiring.
Locate the electrode tee where an active sample of cooling tower water is available and where the electrode can easily be removed for cleaning. It must be situated so that the tee is always full and the electrode is never subjected to a drop in water level resulting in dryness. Refer to Figure 1 for typical installation.
IMPORTANT: To avoid cracking the female pipe threads on the supplied plumbing parts, use no more than 3 wraps of Teflon tape and thread in the pipe FINGER tight plus 1/2 turn! Do not use pipe dope to
seal the threads of the flow switch because the clear plastic will crack!
3.4 Icon Definitions
Symbol Publication Description
IEC 417, No.5019 Protective Conductor Terminal
IEC 417, No. 5007 On (Supply) IEC 417, No. 5008 Off (Supply)
ISO 3864, No. B.3.6 Caution, risk of electric shock
ISO 3864, No. B.3.1 Caution
4
Page 8
SOLENOID VALVE
COOLING TOWER
WATER METER
HEAT EXCHANGER
METERING PUMPS
FLOW SWITCH
10.0" (254 mm)
Cooling Tower Contr oller
Cooling Tower Controller
PREV NEXT
EXIT
ENTER
www.walchem.com
CIRCULATING PUMP
FLOW
FLOW
PREV NEXT
EXIT
ENTER
www.walchem.com
ELECTRODE
SAMPLE VALVE
1/4” POLYPRO PANEL
Figure 1 Typical Installation
5
17" (432 mm)
Page 9
(F)
Cooling Tower Controller
TYPICAL INSTALLATION
INTERMITTENT SAMPLING
TO
CONSTANT
110 VOLTS
PREV NEXT
ENTER
www.walchem.com
PE DISCHARGE
CHEMICAL
PUMP
EXIT
TUBING
CHEMICAL
TANK
TO ELECTRODE TO SOLENOID
(A)
TO
SANITARY
SERVER
(A)
(D)
(B)
(F)
(C)
(E)
SPRA Y T OWE R
(A) Unions; installed for replacement of solenoid, strainer or probe
(B) ½" solenoid (C) ½" Y-strainer
SHUTOFF VALVE
CITY WATER
MAKEUP
SPRAY PUMP
(D) ¾" PVC threaded tee probe (supplied with controller)
(E) ½" PVC bulkhead fitting
winter drain
Figure 1a Typical Installation
Intermittent Sampling
6
Page 10
Cooling Tower Controller
PREV NEXT
ENTER
TYPICAL INSTALLATION
SUBMERSION ELECTRODE
EXIT
TO
110 VOLT AC
OUTLET
CHEMICAL
PUMP
CHEMICAL
TANK
www.walchem.com
XYZX
VALVE
TO
SANITARY
SERVER
(OPTION "A")
CIRCULATION
PUMP
(X) Unions; installed for replacement of solenoid or strainer (Y) ½" Y-strainer
(Z) ½"solenoid
PROBE
COOLING
TOWER
(OPTION "B")
OPEN SUMP
XYZX
VALVE
TO
SANITARY
SERVER
NOTE: LOCA TION OF SOLENOID VALVE USING OPTION "A" OR OPTION "B" DEPENDS ON SYSTEM CONFIGURATION.
DETAIL OF
PROBE ASSEMBL Y
CLIP PROBE ASSEMBLY TO SUMP WALL FOR EASY REMOVAL FOR CLEANING
¾" PROBE IN PVC TEE
TO
CONTROLLER
1" PVC PIPE
1" NPTF PVC COUPLING
INSTALL APPROXIMA TELY HALF-WAY IN SUMP VO LU ME
AVOID INSTALLATION NEAR PUMP SUCTION INLET
Figure 1b Typical Installation
Submersion Electrode
7
Page 11
UG
O
3.5 Electrical installation
The various standard wiring options are shown in figure 2. Your WCT series controller will arrive from the factory prewired or ready for hardwiring. Depending on your configuration of controller options, you may be required to hardwire some or all of the input/output devices. Refer to figures 3 and 4 for circuit board layout and wiring.
Note: when wiring the optional flow meter contactor input, the 4-20 mA output or a remote flow switch, it is advisable to use stranded, twisted, shield pair wire between 22-26 AWG. Shield should be terminated at the controller ground stud (see figures 3 and 4).
1. There are live circuits inside the controller even when the power switch on the front panel is in the OFF position! The front panel must never be opened before power to the controller is REMOVED! If your controller is prewired, it is supplied with a 8 foot, 18 AWG power cord with USA style plug. A tool (#1 Phillips driver) is required to open the front panel.
2. When mounting the controller, make sure there is clear access to the disconnecting device!
3. The electrical installation of the controller must be done by trained personnel only and conform to all applicable National, State and Local codes!
CAUTION
4. Proper grounding of this product is required. Any attempt to bypass the grounding will compromise the safety of persons and property.
5. Operating this product in a manner not specified by Walchem may impair the protection provided by the equipment.
CONDUCTIVITY ELECTRODE
FLOW SWITCH (OPTIONAL)
Figure 2a WCT400 Conduit Wiring
POWER
BLEED
FEED
PL
SPARES
4-20mA (OPTIONAL)
FLOW METER (OPTIONAL)
CONDUCTIVITY ELECTRODE
FLOW SWITCH (OPTIONAL)
Figure 2b WCT410 Conduit Wiring
POWER
BLEED
FEED
BIO 1
PLUG
BI
ALARM
2
SPARES
4-20mA (OPTIONAL)
8
FLOW METER (OPTIONAL)
Page 12
F1
FEED
L2L2
BLEED
N.C.
N.O. N.C. N.O.
FLOW SW 1
IN-IN+
Contact Closure:
Polarity not critical
FLOW MTR 1
IN-
IN+
FLOW MTR 2
IN+ IN-
Hall Effect
Flow Meter
L1 L2/N
GROUND STUD
G
R
G
N
R
N
1
/
2
Y
0
E
V
L
2
4
0
V
COND
T-T+
W
G
R
H
E
I
T
E
E
N
BLKRED
+5V
SHIELD
B
B
L
R
N
2
4
0
V
F2
L1 L2/N
K
1
2
0
V
WHT 120V BLU 240V
Reed Switch
Flow Meter
Polarity not Critical
Conductivity
Electrode
Figure 3 Inputs
9
Power Supply
(115 VAC or 230 VAC)
Page 13
Chart Recorder
GROUND
STUD
V 0 2
1
N
R
G
R
G
FEED
L2L2
L2/N
L1
V
0
4
2
L
E
Y
/
N
GRN 120V GRN/YEL 240V
BLEED
N.C.
N.O. N.C. N.O.
BLEED
N.C.
N.O. N.C. N.O.
FEED
WHT 120V BLU 240V
NOTE: When connecting a motorized ball valve, the pre-wired pigtail must be removed and the valve requires two wires, one to N.O. to open the valve and one to N.C. to close the valve.
Bleed Solenoid/
Motorized Ball Valve
Figure 4a WCT400 Outputs
W
H
B
T
L
U
2
4
BLK 120V BRN 240V
IF MOTORIZED
BALL VALVE
1
2
0
V
0
V
BLK 120 BRN 240V
PUMP
10
Page 14
)
A
Chart Recorder
GROUND
STUD
L2 L2 L2 L2 L2 L2
L1 L2/N
BLEED
FEED
N.C.
N.O. N.C.
BOI 1
N.O. N.C. N.O.
N.O.N. C.
N.C.
BIO 2
ALARM
N.C.
N.O.
N.O.
GRN 120V GRN/YEL 240V
L2/N L2/N L2/N L2/N L2/N
WHT 120V
WHT 120V
WHT 120V
BLU 240V
WHT 120V
BLU 240V
BLU 240V
NOTE: When connecting a motorized ball valve, the pre-wired pigtail must be removed and the
Bleed Solenoid/
Motorized Ball Valve
valve requires two wires, one to N.O. to open the valve and one to N.C. to close the valve.
L2/N
BLU 240V
IF MOTORIZED
BALL VALVE
BLEED
N.C.
N.O. N.C.
FEED
N.O.N.C.
BLK 120V BRN 240V
GRN 120V GRN/YEL 240V TO GROUND STUD
BIO 1
N.O. N.C. N.O.
N.C.
BLK 120V BRN 240V
GRN 120V GRN/YEL 240V TO GROUND STUD
BIO 2
BLK 120V BRN 240V
GRN 120V GRN/YEL 240V TO GROUND STUD
GRN 120V GRN/YEL 240V
N.O.
larm
N.C.
ALARM
N.O.
Pump(s
Figure 4b WCT410 Outputs
11
Page 15
4.0 FUNCTION OVERVIEW
4.1 Front Panel
Setting Adjustment Keys
On/Off Power Switch
Cooling Tower Controller
Figure 5 Front Panel
Backlit LCD Display
Output LEDs (WCT400 only has Bleed and Feed LEDs)
Menu/Function Keys
USB LED
USB Connector
4.2 Display
A summary screen is displayed while the WCT controller is on. This display shows a bar graph of the conductivity relative to the set point, the actual conductivity value, and current operating conditions. The center of the bar graph is at the (S), which represents the set point. For each 1% rise above the set point a vertical bar appears to the right of the (S). For each 1% drop below the set point a vertical bar appears to the left of the (S). There are small breaks in the bars at each 5%. The bar graph is limited to displaying 20% above or below the set point. When these limits are reached an (L) or (H) may appear at the end of the bar graph to indicate a low or high alarm. The operating conditions that are displayed on the bottom line of this display are Bleed, Chem Feed, Cond Hi/Lo Alarm, Sampling, Waiting, No Flow, Biocide Pre Bleed, Bleed Timeout, Biocide Add, Biocide Lockout, Sensor Error, Temperature Error, and Normal. Normal just means there is nothing unusual to report.
L
S
Normal
1546
S
1728
Normal
Figure 6 Summary Screen
S
Normal
2010
12
Page 16
4.3 Keypad
The keypad consists of 4 directional arrow keys and 4 function keys. The arrows are used to move the adjustment cursor and change settings, while the function keys are used to enter values, and navigate the various menu screens. The function keys are ENTER, EXIT, NEXT, and PREV (previous). NEXT and PREV scroll through the various menu choices. ENTER is used to enter a submenu and to enter a value. EXIT is used to back up one menu level. If you are at the main menu level, EXIT will return you to the Summary Display.
To change a value in a submenu, the left/right arrow keys move the cursor left and right to each digit or option that can be changed. The up/down arrows will change numeric values up or down, or scroll through option choices. Press ENTER only when you have finished making all of the changes for that menu screen.
4.4 Access Code
The WCT series controller is shipped with the access code disabled. If you wish to enable it, see Section 5.11 for operation. With the access code enabled, any user can view parameter settings, but not change them. Note that this provides protection only against casual tampering. Use a lock on the cover latch if you need more protection.
4.5 Startup
Initial Startup
After having mounted the enclosure and wired the unit, the controller is ready to be started.
Plug in the controller and turn on the power switch to supply power to the unit. The display will briefly show the WCT model number and then revert to the normal summary display. Scroll through the menus and calibrate the conductivity reading, temperature, and set the control parameters detailed in Section 5, Operation.
To return to the summary display, press the EXIT key until you return to this screen. The controller will automatically return to this screen after 10 minutes.
Normal Startup
Startup is a simple process once your set points are in memory. Simply check your supply of chemicals, turn on the controller, calibrate it if necessary and it will start controlling.
4.6 Shut Down
To shut the WCT controller down, simply turn off the power. Programming remains in memory.
13
Page 17
5.0 OPERATION
These units control continuously while power is applied. Programming is accomplished via the local keypad and display.
To view the top level menu, press any key. The menu structure is grouped by inputs and outputs. Each input has its own menu for calibration and unit selection as needed. Each output has its own setup menu including set points, timer values and operating modes as needed. After ten minutes of inactivity in the menu, the display will return to the summary display. Keep in mind that even while browsing through menus, the unit is still controlling.
5.1 Main Menu
The exact configuration of your WCT controller determines which menus are available as you scroll through the settings. Certain menus are only available on the WCT410 controller, and others will change if you select certain options. All settings are grouped under the following main menu items.
Conductivity Temperature Bleed Feed WM1 WM2 Bio 1 WCT410 Only Bio 2 WCT410 Only Time Alarm 4-20mA Only if 4-20mA option installed Access Code Datalog Only if advanced USB feature is in model code Config Only if advanced USB feature is in model code Upgrade
The NEXT key travels forward through this list while the PREV key travels backwards through the list. Pressing ENTER will Enter the lower level menu that is currently displayed.
14
Page 18
Main Menu
2000 S 67°F
WM2 0 GAL
2000 S 67°F
Access Code DIS
2000 S 67°F
WM1 0 GAL
2000 S 67°F
Trips diagnostic alarm relay
in WCT410 model
* Stops conductivity control
Temp Error
Sensor Error*
No Flow
Bleed Timeout*
Biocide Add*
Bleed
Waiting
Sampling
Biocide Lockout*
Biocide Pre Bleed
Chem Feed
Cond Hi/Lo Alarm
Normal
Feed OFF
2000 S 67°F
Bleed OFF
Possible status screens
2
2000 S 67°F
4-20mA 4.0 mA
2000 S 67°F
Cnd Alrm Lo Alrm
2000 S 67°F
Time: Sun 3:00
2000 S 67°F
Upgrade
2000 S 67°F
Config
Press Enter key to enter menu.
Press Exit key to exit menu.
Blinking fields may be edited with the adjust arrows.
Press Enter when modification is complete to return
to Main Menu Level.
Operation
2000
S
Normal
L
2000 S 67°F
Temperature
ENTEREXIT
2000 S 67°FConductivity
2000 S 67°F
Bio 2 OFF
1
2000 S 67°FBio 1 OFF
1
2000 S 67°FDatalog
Legend
Menu only appears in WCT410 model.
1
Menu only appears when 4-20 mA hardware is installed.
2
Figure 7 Main Menu
15
Page 19
5.2 Conductivity Menu
The conductivity menu provides the following settings: Calibration, Self Test, Unit selection, and sampling mode setup. Additional settings are also discussed below. Refer to figure 8, Conductivity Menu Chart.
Calibrate
Self Test
Units
ppm C.F.
Sample Mode C / I
To Calibrate the conductivity, use either a hand held meter, or a buffer solution, and adjust the WCT controller to match. Once Calibrate is entered, the unit continuously displays conductivity readings. Press an arrow key to change the value displayed to match the hand held meter or the buffer solution. You must press ENTER to activate the new calibration. You must press the EXIT key to exit calibration. The Bleed output is unaffected until the calibration menu is exited, so if it was ON when you entered calibration it will stay on until you exit.
Press ENTER to begin self test. Press any key to stop. Self Test internally simulates a conductivity sensor and should always give the reading 1000 µS/cm ± 100 µS. If it does not, there is a problem with the electronics and unit should be serviced. If the self-test is in the expected range, and there is a problem calibrating, then the sensor or its wiring is at fault.
You may choose to display conductivity in µS/cm or in ppm. Press ENTER and then use the Up and Down arrows to change the units. If you change the units, you will be warned to check your settings. This is important. Set points are not automatically translated from µS/cm to ppm. If you change the units you will need to change your Bleed settings.
This is the ppm Conversion Factor (or multiplier). This is typically 0.666 but can be changed to accommodate various requirements.
Press enter to choose Continuous sampling or Intermittent sampling. A 'C' at the end of the display means that sampling is continuous, while an 'I' indicates intermittent sampling. Use Continuous sampling with a traditional bypass line installation of the conductivity sensor.
Choose Intermittent sampling to use the bleed solenoid valve for timed sampling of the conductivity. Intermittent sampling installations read the conductivity at set intervals for a given sample duration. If the
conductivity is above the set point, the valve that controls the sampling will stay open until the conductivity falls below the set point. If the time the valve stays open goes beyond the sample duration, the controller will display Xtnd on the top status line, as well as the amount of time extended. A limit on this amount of time may be imposed; see Figure 10 Bleed Menu.
Continuous
Interval Duration RawCond
Intermittent
If Intermittent sampling is chosen, the Flow Switch input will be ignored, and the following two settings will become available:
This sets the amount of time between samples. This is set in Hours:Minutes.
This is the length of each sample. This is set in Minutes:Seconds.
This menu is for information only and displays the temperature-compensated conductivity reading before the user’s calibration has been applied.
16
Page 20
P
ibl
y
2000 S 68° F Conductivity
ENTEREXIT
Cond 1000µS Calibrate
ENTER
EXIT
oss
Cond Wait 23:50 Cond Samp 10:00
Cond Xtnd 10:00
Cond 1000µS Self T est
ENTER
EXIT
e status screens
Cond 1000µS Units S/cm
EXIT
ENTER
Conductivity Menu
NEXTPREV.
1
Cond 1000µS ppm C.F. 0.666
Cond 1000µS Sample Mode C
EXIT
ENTER
Calibrate
S 2000
Calibrate ppm 1333
1
Self Test Cond 1000 S
Self T est Cond 666 ppm
1
Units S/cm
S
Units ppm ppm
1
. . . . WARNING . . . .
Adjust Set Point
2 2
Cond 1000µS Interval (H) 3:59
Legend
Menu choices that appear when ppm units are
1
selected.
Operation
Press Enter key to enter menu. Press Exit key to exit menu. Blinking fields may be edited with the adjust arrows.
Menu choices that appear when intermittent
2
sampling is selected.
Press Enter when modification is complete to return to Conductivit
Figure 8 Conductivity Menu
Cond 1000µS Duration 2:00
Menu Level.
Sample Mode C Continuous
Sample Mode I Intermittent
2
Cond 1000µS RawCond 2000µS
17
Page 21
5.3 Temperature Menu
The Temperature menu contains the following settings: Calibration and Units (if the Temp element is detected when the unit is powered on) or Manual Temp and Units (if no Temp sensor is detected at power up). Refer to the Temperature Menu chart, figure 9.
If a "Temp Error" message appears or if the "Man Temp" menu appears after the unit is powered on, it indicates the temp element is not working properly. Refer to the Troubleshooting Section.
Calibrate
Man Temp Units
Mode
2000 s 68° F Temperature
ENTEREXIT
Temp 68°F Calibrate
To Calibrate the Temperature, use a thermometer to measure the fluid temperature and adjust the WCT controller to match. Once Calibrate is entered, the unit continuously displays temperature readings. Press the Up or Down arrow key to change the value displayed to match the thermometer. You must press ENTER to activate the new calibration. You must press the EXIT key to exit calibration.
This menu appears only if no temperature element is connected at power-up. Use the arrow keys to adjust the temperature displayed to match that of the water.
You may choose to display temperature in °C or °F. Press ENTER and the Up or Down Arrow keys to change the temperature units for display.
Press ENTER and use the UP or DOWN arrow keys to select between Manual Temperature Compensation (used with sensors that do not have a temperature measuring element in them) or Automatic Temperature Compensation. If Automatic is selected, and the temperature element is not detected, the controller will display a Temp Error and revert to Manual Temp Comp until the signal is restored.
Possible Status Screen
Temperature Err
NEXTPREV.
1
Temp 68°F Man Temp 68
2
Temp 68°F Units °F
Temperature Menu
Temp 68°F Mode Manual
ENTER
EXIT
Calibrate Temp °F 68
Calibrate Temp °C 20.1
Legend
1
Menu wording that appears when Automatic Temperature Compensation is selected.
2
Menu wording that appears when Manual Temperature Compensation is selected.
ENTER
EXIT
Units °F °F
Units °C °C
Figure 9 Temperature Menu
ENTER
EXIT
Mode Man Manual Temp Comp
Mode Man Auto Temp Comp
18
Page 22
5.4 Bleed Menu
The Bleed Menu provides the following settings: Set Point, Dead Band, Bleed Mode, Bleed Vol, Makeup Vol, HOA. The Bleed menu will be indicated on the display by one of the following: (The 'A' indicates that the output is being controlled automatically.) Hint: For easiest programming, program Bleed Mode first, then step through the rest of the bleed parameters. The modes are defined as follows:
H High Set Point
L Low Set Point
B Bleed Volume on Makeup Volume
Bleed A
High set point is the default setting, where the bleed relay activates when the conductivity rises above the set point value.
Low set point, or reverse set point, is where the bleed relay activates when the conductivity drops below the set point value.
In bleed volume on makeup volume, a flow of a programmable volume of makeup water triggers the relay to activate, and it stays activated until the bleed water meter detects the programmed volume of water.
Indicates that the bleed output is currently OFF.
OFF
Bleed A Bleed A
Bleed A Bleed A
Set Point
Dead Band
Time Limit
Reset Timer Bleed Vol
Makeup Vol Bleed Mode
H / L / B H O A
Indicates the length of time that the Bleed output has been ON.
10:00
NO FLOW
LOCKOUT
TIMEOUT
This is the conductivity value at which the bleed solenoid valve is turned ON. The factory default setting for the WCT controller is for the Bleed output to turn on when the conductivity is higher than the set point. This may be changed at the Control Direction screen.
This is the conductivity value that when combined with the set point determines when the bleed output turns OFF. Assuming that the control direction is set for normal operation (High Set Point) the bleed output will turn off when the conductivity drops below the set point minus the Dead Band. For example: The set point is 1500 µS/cm and the Dead Band is 200 µS/cm. The bleed output turns ON when the conductivity reading is greater than 1500 but does not turn OFF until the conductivity drops below 1300.
This menu allows you to set a maximum amount of time for the bleed. The limit time is program med in hours and minutes and can be set between 1 minute and 8 hrs: 20 minutes. If the time limit is set to zero, then the valve may be open indefinitely. If the maximum time is exceeded, the bleed valve will close and will not re-open until the "Reset Timer" menu is reset by an operator.
Only appears if the time limit above has been exceeded. Use the up or down arrow to change "N" to "Y", then press ENTER.
This menu is only displayed if the Bleed Mode is set to BD Vol on MU Vol. Use the arrow keys to select the volume of water to bleed once the relay is triggered.
This menu is only displayed if the Bleed Mode is set to BD Vol on MU Vol. Use the arrow keys to select the volume of makeup water required to trigger the bleed relay.
This menu allows you to set the bleed mode as described above. Press ENTER and then use the Up or Down arrow keys to select between High Set Point, Low Set Point or BD Vol on MU Vol.
The "Hand Off Auto" screen allows you to select the operating mode of the bleed output. In Hand (manual) mode, the output is turned on immediately for a maximum of 10 minutes. If you walk away the output will return to Auto mode at the end of that time. In Off mode the output will stay Off indefinitely. In Auto mode the bleed output will respond to changes in conductivity based on the set point. The HOA mode of the bleed output is indicated on the bleed status lines.
Indicates that bleed control has been suspended because there is presently no flow past the flow switch.
Indicates that the output is currently locked out due to a biocide add or biocide lockout. Indicates that the bleed output has been on longer than the Time Limit.
19
Page 23
Bleed Menu
p
A
AAA
A
NEXTPREV.
Bleed 1:01:15
Reset Timer N
2
Bleed 1:01:15
Time Limit 1:15
Press Enter key to enter menu.
Press Exit key to exit menu.
Blinking fields may be edited with the adjust arrows.
Press Enter when modification is complete to return
Operation
A
HAND OFF > AUTO
Bleed 1:01:15
MU VOL 1000g
A
Bleed 1:01:15
Bleed VOL 1000g
3 3
Bleed 1:01:15
ENTER
A A
EXIT
Bleed Mode B
Bleed 1:01:15
Bleed Mode H
Bleed Mode H
Bleed Mode L
High Set Point
BD Vol on MU Vol
Low Set Point
to Bleed Menu Level.
Bleed 1:01:15
Bleed No Flow
Bleed Lockout
Bleed OFF
Bleed Timeout
Possible status screens
A A A
Bleed 1:01:15
Dead Band 200
1
Volume Mode.
ENTEREXIT
A
2000 S 68° F
Bleed 1:01:15
A
Bleed 1:01:15
Appears only in Continuous
Sample Mode.
Appears only if Limit Ti mer
has expired.
Appears only in Bleed V olume
Set Point 2000
1
Legend
2
or Makeu
3
Figure 10 Bleed Menu
20
Page 24
5.5 Feed Menu
The Feed Menu adapts to the selected Feed output mode. Hint: For easiest programming, program Feed Mode first, then step through the rest of the feed parameters. The modes are defined as follows:
A Bleed and Feed with Optional
B Feed % of Bleed
C Feed % of Time
D Feed based on Water Contactor
E Feed based on Paddlewheel Input
The Feed menu will be indicated on the display by one of the following: (The 'A' indicates that the feed is being controlled automatically.)
Feed A Feed A
Feed A
Feed A Feed A
Bleed and Feed Mode
Bleed and Feed Mode turns the Feed output On and Off at the same time as the Bleed output.
Lockout
NO FLOW
TIMEOUT
LOCKOUT
The lockout setting determines the maximum allowable time for the Feed output. If this time is exceeded the Feed output is turned off and Locked out until the Bleed output turns off.
Feed % of Bleed Mode tracks the length of time that the Bleed output is on. When the bleed turns off the feed output is energized for a user defined proportion of the bleed time.
Feed % of Time Mode turns on the Feed output for a user definable % of a timed cycle. The time cycle length is adjustable from 10 to 60 minutes.
Feed based on Water Contactor Input Mode turns on the Feed output for a user definable time each time a water contactor pulse is detected. This contactor input can be divided to
Input
accommodate a large variety of water meters. Contacts will accumulate feed time so that all contacts are accounted for.
Turns on the Feed output for a user definable time each time a defined volume of flow is detected. This paddlewheel input has a programmable K Factor to work with a large variety of
Mode
Hall Effect (square wave, not sine wave) flow meters.
Indicates that the Feed output is currently OFF.
OFF
Indicates the length of time that the Feed output has been ON or the length of time that the
10:00
Feed output will be ON. Indicates that Feed control has been suspended because there is presently no flow past the flow
switch. Indicates that the feed lockout timer in the Bleed and Feed mode has expired. Indicates that the output is currently locked out due to a biocide add or biocide lockout.
Lockout
Set this for the Feed Lockout Time. The lockout time is the maximum length of time that the feed output can be on. If the lockout time is set to 0:00, the lockout timer is no longer used and the feed output will be on for as long as the bleed is on.
Feed % of Bleed Mode
% of Bleed
This is the % value that is multiplied times the accumulated bleed time to determine how long the feed will be. For example, if the bleed was on for 10 minutes and this setting was 50%, the feed output would be on for 5 minutes.
Max Time
This is similar to the lockout time above in that the feed output will not exceed this maximum length.
Feed % of Time Mode
% of Time
This is the % value that is multiplied times the cycle length to determine the length of time that the feed output is ON. If the cycle length were 10 minutes and this setting was 40%, the feed output would be on for 4 minutes, then off for 6 minutes and then repeat the cycle.
Cycle Time
This determines the length of the cycle to be used.
21
Page 25
Feed Based on Water Contactor Mode
Time/Cont.
(Time per contact.) This determines the length of time that the feed pump should be on for each contact that is received.
÷ Contacts By
This setting allows a divider to be entered. The divider will count actual contacts from the meter until the setting is reached before a contact is considered to be received. For example, if the divider is set to 10 and the Time/Cont is set to 5:00, then the feed output would turn on for 5:00 minutes after 10 contacts were received.
Assign Meter
Press ENTER and then use the Up or Down arrow keys to assign the water meter to be used for controlling this relay. Select between Watermeter 1, Watermeter 2, and Both.
Time Limit
This setting puts a limit on the amount of time that can be accumulated by the water meter input. Once this setting has been reached, all contacts will be ignored until the accumulated feed time expires. By setting Time Limit = Time/Cont., the accumulation of contacts can be disabled.
Feed Based on Paddlewheel Mode
Time/Vol
This setting determines the pump on-time once a given volume of water has passed through the paddlewheel sensor. The volume required to initiate feed is set below.
Vol to Init.
This setting determines the volume of makeup water that will initiate chemical feed.
Assign Meter
Press ENTER and then use the Up or Down arrow keys to assign the water meter to be used for controlling this relay. Select between Watermeter 1, Watermeter 2, and Both.
Time Limit
This setting puts a limit on the amount of time that can be accumulated by the water meter input. Once this setting has been reached, all contacts will be ignored until the accumulated feed time expires. By setting Time Limit = Time/Vol., the accumulation of contacts can be disabled. This is set in minutes and seconds.
The following settings are for all feed modes.
Chem Feed Mode
This allows the user to select the chemical feed mode as described above.
A / B / C / D / E H O A
This sets the Hand Off Auto for the feed output. This was explained in the Bleed Menu section and functions similarly. In Off position, the output will not turn ON regardless of the feed mode selected.
22
Page 26
P
oss
ibl
A
A
p
A
A
A
2000 S/cm 68° F
A
Feed 21:05
ENTEREXIT
Menu choices depend on feed mode
Feed Lockout 99:00
1
Feed % of Bleed 10
22
Feed % of Time 10
3
Feed Time/cont. 1:00
4
Feed
Time/Vol 5:00
TIMEOUT
A
A
TIMEOUT
A
TIMEOUT
A A
TIMEOUT
A
TIMEOUT
1
Feed Max Time 99:00
Feed Cycle Time 10:00
3
Feed
Contacts by 1
4
Feed
Vol to Init 5:00
A
A
A
TIMEOUT
TIMEOUT
TIMEOUT
TIMEOUT
e status screens
A
Feed Off Feed 21:05
A A
Feed Lockout
A
Feed No Flow
A
Feed Timeout
3
A
Feed
TIMEOUT
ssign Meter 1
ENTER
EXIT
Assign Meter 1 Wateremeter1
Assign Meter 2 Wateremeter2
Assign Meter B Both
4
A
Feed
TIMEOUT
Feed
ENTER
EXIT
Time Limit 99:00
Assign Meter
3
Feed Time Limit 99:00
4
Feed Timeout
TIMEOUT
Time Limit 99:00
NEXTPREV.
TIMEOUT
Feed Menu
(Hint: program Mode first)
Feed Chem Feed Mode
Chem Feed Mode Bleed & Feed
1
Chem Feed Mode Feed % of Bleed
2
Chem Feed Mode Feed % of Time
3
Chem Feed Mode Water Contactor
4
Chem Feed Mode Paddlewheel
EXIT
TIMEOUT
ENTER
Feed
HAND OFF >AUTO
TIMEOUT
Assign Meter 1 Wateremeter1
Assign Meter 2 Wateremeter2
Assign Meter B Both
Legend
1
Menu choices that appear when Feed as % of Bleed mode is selected.
Operation
Press Enter key to enter menu. Press Exit key to exit menu. Blinking fields may be edited with the adjust arrows.
2
Menu choices that appear when Feed as % of Time mode is selected.
3
Menu choices that appear when Feed based on Water contactor input is selected.
4
Menu choices that appear when Feed based on Paddlewheel in
ut is selected.
Press Enter when modification is complete to return to Feed Menu Level. Press Enter or Adjust arrow to turn on/off output at Hand menu.
Figure 11 Feed Menu
23
Page 27
5.6 WM1 and WM2 Menus
The Water Meter 1 and 2 menus are used to set up the type of meter and how to convert the water meter signal into volume. The Water Meter menus will be indicated on the display as follows:
WM1 100 Gal or Liters
Reset Total
Vol/cont
Only appears if the WM Type mode is Water contactor.
K Factor
Only appears if the WM Type mode is Paddlewheel
Tot Units WM Type
This allows you to restart the totalizer display. Press the Up or Down arrow key to change the N to Y and press ENTER to reset the totalizer to 0 gal. The totalizer will count up to a maximum of 99,999,999. After that it will reset itself to zero.
This allows you to set the volume of makeup per contact from the water meter.
Enter the number of pulses per unit volume that the paddlewheel sensor sends out. This value is usually printed on the sensor’s flow cell or in its instructions.
This is used to set the units of measure for the totalizer. Press ENTER, then use the up and down arrow keys to toggle between "Gallons" and "Liters." Press ENTER when the desired choice is displayed.
Press ENTER and then use the Up or Down arrow keys to select between Water Contactor, Paddlewheel, or Not Used.
1000 S 68° F
WM1 1000Gal
ENTEREXIT
WM1 1000Gal Reset Total N
Legend
1
Menu choice appears if WM Type is Water Contactor
2
Menu choice appears if WM Type is Paddlewheel
3
This is the only menu shown if WM Type is Not Used.
NEXTPREV.
1
WM1 1000Gal Gal/Cont. 10
or
WM1 1000Gal Lit/Cont. 10
Figure 12 WM1 and WM2 Menus
2
WM1 1000Gal KFactor 999.9
WM1 and WM2 Menus
3
WM1 1000Gal Tot Units Gal
ENTEREXIT
Tot Units Gal Gallons
Tot Units Gal Liters
WM1 1000Gal WM Type P
WM Type P Paddlewheel
WM Type C Water Contactor
WM Type N Not Used
ENTEREXIT
24
Page 28
5.7 Bio1 and Bio2 Menus
These menus are only available on the WCT410 controller. The Bio 1 and Bio 2 menus are separate from each other but operate in exactly the same way. Each menu provides the following independent settings: Prog Bio Adds, Bio Pre-Bleed, Bio Lockout Time, Set Bio Mode, HOA. Bio menus may display the following screens:
Bio 1 A Bio 1 A
Bio 1 A Bio 1 A
Bio 1 A
Note: When programming the unit for the first time, it is recommended that the Biocide Add Mode be chosen first, then step through the rest of the Biocide Add menu indicated below and in figures 13 and
14. This will make programming simple and logical.
A number of built-in interlocks are part of the biocide feed program. When one biocide relay activates, the other biocide will be locked out until the first biocide finishes its control cycle. Similarly, the bleed relay is locked out once the biocide cycle begins (except for the pre-bleed portion of the cycle). Bear this in mind, especially if the biocide feed time is very long or very frequent...very little time is left in the day to control the conductivity.
The inhibitor feed relay interaction is more complex.
For Bleed and Feed mode For Feed as % Bleed
For Feed as % Time
For Water Contactor or Paddlewheel mode
OFF
PENDING
PRE BLD
4:50
NO FLOW
Indicates that the Bio 1 output is off. Indicates that Bio 1 is ready to begin a biocide cycle but is unable to begin due to NO FLOW,
or Bio 2 already ON. Indicates that Bio 1 is in the Pre-Bleed portion of its cycle. Indicates that Bio 1 is adding and has 4 minutes and 50 seconds remaining to add, or that there
are 4 hours and 50 minutes of Bio 1 Lockout remaining. If the Bio 1 LED is ON, the timer is for the Add. If the LED is OFF, the timer is for remaining lockout time.
Indicates that there is no biocide activity and none is pending and that there is no flow past the flow switch.
Since the bleed relay is locked out, the feed relay will be locked out as well. If the bleed is adjacent to the start of the biocide cycle, or is during the pre-bleed part of the
cycle, the feed time does not occur. The feed relay is locked out during the biocide add, the feed events are not stored in
memory, but the feed cycle time is reset once the biocide add cycle is over, so that as soon as the biocide cycle is complete, one feed cycle will occur.
The water flow is stored in memory, up to 256 contacts, and the correct feed time is activated once the biocide cycle has completed.
25
Page 29
Prog Bio 1 Add
Pre Bleed
Lockout
Bio 1 Add Mode 1
Press ENTER here to see a list of all of the biocide adds presently scheduled. The first screen displayed is for Today's Biocide add. Use the NEXT key to see Tomorrow's add or the PREV key to see Yesterday's add. Use the arrow keys to edit the starting time or length of the add. The screen should look similar to the one below: Prog Bio 1 Add 2Mon 10:00 45
The bottom line indicates that on the second Monday in a multi-week cycle, Bio 1 will begin its Biocide Addition cycle at 10 AM and will add for 45 minutes. If the biocide output does not come on at exactly 10 AM, there may be NO FLOW, or there may be a conflict with Biocide output 2. If either of these conditions is true, the output will be delayed until the flow has been restored, or the Biocide 2 lockout time is completed. The times are set in a 24 hour format, so 1 PM is entered as 13:00. The WDT supports daily cycles, 1 week cycles, 2 week cycles and 4 week cycles. It is not necessary for both biocide outputs to be on the same cycle.
Press the EXIT key to exit the Prog Bio 1 menu. There are separate Bio 1 and Bio 2 Program menus and separate program memories, allowing addition of each biocide independently.
This setting provides a Pre Biocide Add Bleed Conductivity Set Point. Pre Bleed is the first step in a biocide addition. The operating conductivity must be lower than this setting before the biocide output will turn ON. If the conductivity is higher than this set point, the bleed output is turned on to lower the conductivity. This setting is independent of the bleed set point in the Bleed Menu. There are separate Bio 1 and Bio 2 Pre Bleed settings. To disable the Pre Bleed, set this value higher than your operating set point or set the value to zero.
This setting determines how long bleed and additional biocide adds will be locked out after completing a biocide add. This value is set in hours and minutes. There are separate Bio 1 and Bio 2 settings. This can be set to 0 if it is not desired.
The last digit indicates that Bio 1adds are scheduled on a 1 week repeating cycle. To change this, press the ENTER key and then use the Up or Down arrow keys to select the appropriate choice. Biocide cycles may be daily (up to ten times per day) or 1, 2, or 4 weeks long. This is provided to save you some programming effort. If you add the same thing every week, use the 1 week cycle. If you wish to add Biocide 1 only once every two or four weeks use the 2 or 4 week cycle. If you wish to add Biocide 1 more than once a day, choose the daily cycle.
The Bio 1cycle length is not affected by the Bio 2 cycle length. In weekly cycle modes, the WDT can schedule one add per day for each Biocide output for up to 4 weeks. If a daily cycle is chosen, that biocide can be added up to ten times per day, every day.
1 Week 2 Week 4 Week Daily
H O A
This is the Hand Off Auto selector screen for the Biocide output. There are separate settings for Bio 1 and Bio 2. In the Off position Biocide adds can be missed. See the description in the Bleed menu for more details about HOA settings.
26
Page 30
P
oss
ibl
y
e status screens
Bio 1 10:00
2000 S 68° F
Bio 1
PRE BLD
ENTEREXIT
Bio 1 2:00
Bio 1 OFF Bio 1 Bio 1
1
Bio 1 Prog Bio 1 Adds
Prog Bio 1 Adds 1 Sun 12:00 99
1
Prog Bio 1 Adds 1 Sun 12:00 10
2
Prog Bio 1 Adds 1 Sun 12:00 0
3
Prog Bio 1 Adds Daily A 12:00 99
EXIT
PRE BLD
ENTER
Bio 1 Pre Bleed (µS) 1800
NEXTPREV.
7 days, total
14 days, total
28 days, total
up to ten times/day
PRE BLD
Prog Bio 1 Adds 1 Sat 12:00 0
Prog Bio 1 Adds 2 Sat 12:00 0
Prog Bio 1 Adds 4 Sat 12:00 0
Prog Bio 1 Adds Daily J 12:00 30
Legend
Menu choices that appear when 1 Week Biocide
1
Cycle mode is selected.
ADD
LCK
NO FLOW
PENDING
Bio 1 Lockout (H) 2:00
PRE BLD
1
2
3
Biocide 1 Menu
NEXTPREV.
(Hint: program Mode first)
Bio 1 Bio 1 Add Mode
Bio 1 Add Mode
1
1 Week Cycle
Bio 1 Add Mode
2
2 Week Cycle Bio 1 Add Mode
3
4 Week Cycle Bio 1 Add Mode
Daily Adds
EXIT
PRE BLD
ENTER
Bio 1
HAND OFF >AUTO
PRE BLD
Operation
Press Enter key to enter menu. Press Exit key to exit menu. Blinking fields may be edited with the adjust arrows. Press Enter when modification is complete. Use Next and Prev keys to change days in Prog.
Menu choices that appear when 2 Week Biocide
2
Cycle mode is selected. Menu choices that appear when 4 Week Biocide
3
cle mode is selected.
C
Figure 13 Biocide 1 Menu
27
Page 31
Possible status screens
y
Bio 2 10:00
2000 S 68° F
Bio 2
PRE BLD
ENTEREXIT
Bio 2 2:00
Bio 2 OFF Bio 2 Bio 2
1
Bio 2 Prog Bio 2 Adds
EXIT
Prog Bio 2 Adds 1 Sun 12:00 99
1
Prog Bio 2 Adds 2 Sun 12:00 10
2
Prog Bio 2 Adds 4 Sun 12:00 0
3
Prog Bio 2 Adds Daily A 12:00 99
PRE BLD
ENTER
Bio 2 Pre Bleed (µS) 1800
7 days, total
14 days, total
28 days, total
up to ten times/day
PRE BLD
NEXTPREV.
Prog Bio 2 Adds 1 Sat 12:00 0
Prog Bio 2 Adds 2 Sat 12:00 0
Prog Bio 2 Adds 4 Sat 12:00 0
Prog Bio 2 Adds Daily J 12:00 30
Legend
Menu choices that appear when 1 Week Biocide
1
Cycle mode is selected.
ADD
LCK
NO FLOW PENDING
Bio 2 Lockout (H) 2:00
PRE BLD
1
2
3
Biocide 2 Menu
NEXTPREV.
(Hint: program Mode first)
Bio 2 Bio 2 Add Mode
Bio 2 Add Mode
1
1 Week Cycle Bio 2 Add Mode
2
2 Week Cycle
Bio 2 Add Mode 4 Week Cycle
Bio 2 Add Mode Daily Adds
EXIT
PRE BLD
ENTER
Bio 2
HAND OFF >AUTO
PRE BLD
Operation
Press Enter key to enter menu. Press Exit key to exit menu. Blinking fields may be edited with the adjust arrows. Press Enter when modification is complete. Use Next and Prev keys to change days in Prog.
2
Menu choices that appear when 2 Week Biocide Cycle mode is selected.
Menu choices that appear when 4 Week Biocide
3
C
cle mode is selected.
Figure 14 Biocide 2 Menu
28
Page 32
5.8 Time Menu
This menu has only one choice, to set the Time used for Biocide adds (WCT410) or Datalogs. This menu will appear as follows:
Time: 2Mon 10:20
Set Time
2000 S 68° F
Time Sat 12:15
ENTER
EXIT
Press ENTER to set the Time. Use the arrow keys to adjust the day and time and then press ENTER to store or EXIT to discard. If either biocide is set for a 2 or 4 week cycle, you can change the current week in the cycle by changing the week number. If the longest biocide cycle is 2 weeks, the 3Sun through 4Sat choices are not shown.
Time Menu
Time Menu is only present if Biocide Option is installed.
Time Sat 12:15 Set Clock
ENTER
EXIT
Set Clock Set Jan/ 3/98
ENTEREXIT
Set Clock Set Thu 12:15
Figure 15 Time Menu
29
Page 33
5.9 Conductivity Alarm Menu
A
A
This menu is available for both WCT400 and WCT410 series controllers, however, only the WCT410 has a relay output for the alarm. On both units the settings here determine when the (L) and (H) indicators appear on the summary screen. The Alarm menu screen will appear as follows:
Cnd Alrm OFF Cnd Alrm LOW ALRM Cnd Alrm HI ALRM
Set % Low
Set % High
2000 S 68° F Cnd Alrm
Cnd Alrm
larm % Low 20
This is the % below the bleed set point that the LOW ALARM will activate. If the conductivity set point is 1000 and the % Low setting is 20 then the Low alarm will activate at 800. The alarm can be disabled by setting it to zero.
This is the % above the bleed set point that the HIGH ALARM will activate. If the conductivity set point is 1000 and the % High setting is 20 then the High alarm will activate at 1200. The alarm can be disabled by setting it to zero.
HI ALRM
ENTER EXIT
HI ALRM
Figure 16 Conductivity Alarm Menu
Possible status screens
Cnd Alrm OFF Cnd Alrm Cnd Alrm HI
NEXTPREV.
Cnd Alrm
larm % High 20
HI ALRM
Cnd Alrm Menu
LO ALRM
ALRM
30
Page 34
5.10 4-20mA Menu
This menu is only available if the 4-20mA output is installed in the controller. The 4-20mA output is available for either the WCT400 or WCT410 series controllers. This menu provides for scaling and calibrating the output. The 4-20mA menu screen appears as follows:
4-20mA 9.20mA
This indicates that the current output of the 4-20mA card is 9.20 mA.
Set 4mA Pt Set 20mA Pt Calibrate
2000 S 68° F
4-20mA 12.4mA
This conductivity setting will correspond to a 4 mA output from the controller.
This conductivity setting will correspond to a 20mA output from the controller.
This will provide fixed 4mA and fixed 20mA outputs to allow you to calibrate connected equipment.
4-20mA Menu
ENTEREXIT
NEXTPREV.
4-20mA 12.4mA Set 4mA Pt 0
4-20mA 12.4mA Set 20mA Pt 0
4-20mA 12.4mA Calibrate 4-20mA
ENTER
EXIT
Calibrate 4-20mA Fixed 4mA Output
ENTER
EXIT
Calibrate 4-20mA Fixed 20mA Output
Figure 17 4-20 mA Menu
4-20mA menu is only present if 4-20mA hardware is installed.
31
Page 35
5.11 Access Code Menu
This menu determines whether the access code feature of the controller is enabled or disabled and allows you to customize the access code to your own value. The access code controls whether or not you are allowed to change the parameters in the controller. With the access code disabled, any user may change any parameter. With the access code enabled, any user can view any parameter, but cannot change them. Once an attempt is made to change a parameter, the display will prompt the user to enter the access code. If the correct access code is entered, the parameters can be changed. If the wrong access code is entered the parameters cannot be changed. Once the access code has been correctly entered, it will remain valid until there is a period of 10 minutes without a key being pressed. The access code menu will appear as shown below:
The access code menu will appear as shown below:
Access Code Access Code Access Code
REQ
Enable N / Y New Value
Indicates that the access code is disabled. No access code is required to change any setting.
DIS
Indicates that the access code is required to alter settings. Indicates that the access code is required and has been entered correctly.
OK
Press the Up or Down arrow key to change the N to Y and press ENTER to enable the access code feature. If the access code is enabled you must first enter the access code to disable it.
Press ENTER to display the current access code value and use the arrow keys to change it to any value between 0 and 9999. If the access code has been enabled, you will be prompted to enter the current access code before being allowed to change it. You must remember the access code if you enable it.
The Factory default Access code is 1995. If you change the access code and can't remember it follow this procedure:
1. Turn off power to the controller.
2. Wait 10 seconds.
3. Press and Hold the UP and DOWN arrow keys while turning on the power.
4. Read the access code on the display.
5. Release the keys, and the access code will disappear.
Figure 18 Access Code Menu
32
Page 36
5.12 Datalog Menu
This menu is available on both the WCT400 and WCT410, if the data logging option has been purchased. This is indicated in the model code by the letter U at the end of the model code. This menu allows you to save data from the controller to a USB flash drive.
The controller has four logs, the Current Datalog, the Backup Datalog, the Event Log, and the Reset Log. All files are in a CSV format that may be opened in a spreadsheet such as Microsoft Excel.
Current Datalog
Backup Datalog
Event Log
Reset Log
Current or Backup Datalog
Copy Event Log
Contains the following data taken at 10 minute intervals: Conductivity Temperature Water Meter 1 Total
Water Meter 2 Total When the current datalog is downloaded to a USB stick, it is erased and a new log file is started. If the current datalog is not downloaded before it reaches its maximum size (at least 60 days of
data) the oldest data is overwritten by the newest data. Contains the same data as the current log but it is never erased. When the backup log reaches its
maximum size (at least 60 days of data), the oldest data is overwritten by the newest data. Contains columns for each relay and flow switch input, as well as the date and time. Each time
any of these change state, the date and time is updated and it will show a 1 if the relay is on and 0 if it is off, and a 1 if the flow switch indicates no flow, 0 if there is flow. Tens of thousands of events will be recorded before the oldest data is overwritten by the newest, the number varying with the controller’s configuration.
Consists of time stamps of when power was lost, when it was returned, and the cause of the reset.
Place a USB flash drive with at least 10 MB capacity into the USB port on the front panel of the controller. Press the Enter key to download the file from the controller to the disk. The file name for the Current Datalog will be Datalog<serial number><date><time>.csv using the date and time it was downloaded. The file name for the Backup Datalog will be Datalog<serial number><date><time> .csv using the date and time it was created.
The controller will display the progress of the file download process. If the file was successfully copied to the USB disk the controller will display Transfer Success.
Place a USB flash drive with at least 10 MB capacity into the USB port on the front panel of the controller. Press the Enter key to download the file from the controller to the stick. The file name will be Eventlog<serial number><date><time>.csv.
The controller will display the progress of the file download process. If the file was successfully copied to the USB disk the controller will display Transfer Success, otherwise Transfer Fail 1.
Copy Reset Log
2000 µS 67°F Datalog
ENTER EXIT
Datalog Current Datalog
Possible Status Screens
Transfer Success Transfer Fail 1
ENTER EXIT ENTER EXIT ENTER EXIT
Datalog Transfer Success
Transfer Success Transfer Fail 1
Place a USB flash drive with at least 10 MB capacity into the USB port on the front panel of the controller. Press the Enter key to download the file from the controller to the stick. The file name will be Resetlog<serial number><date><time>.csv.
The controller will display the progress of the file download process. If the file was successfully copied to the USB disk the controller will display Transfer Success.
Transfer Success Transfer Fail 1
Datalog Menu
Figure 19 Datalog Menu
Next
Datalog Backup DataLog
Prev
ENTER EXIT
Datalog Transfer Success
Next
Datalog Copy Event Log
Prev
Datalog Transfer Success
33
Next
Prev
Datalog Copy Reset Log
Datalog Transfer Success
Page 37
5.13 Config Menu
This menu is available on both the WCT400 and WCT410 models. It allows you to export a file that contains all of the set points in the controller to a USB flash disk drive, and then later import the set points into another controller.
Export Config
Import Config
Place a USB flash drive with at least 10 MB capacity into the USB port on the front panel of the controller. Press the Enter key to export the configuration file from the controller to the stick. The file name will be UCF.ini. If you are exporting files with different set points you may rename the file to something that describes it, as long as it has an ini extension.
The controller will display the progress of the file download process. If the file was successfully exported to the USB disk the controller will display Transfer Success, otherwise Transfer Fail 1.
Transfer Success Transfer Fail 1
Place a USB flash drive that contains only one configuration file stored on the root directory of the stick into the USB port on the front panel of the controller. Press the Enter key to import the configuration file from the stick to the controller. The file name must have an ini extension in its name.
The controller will display the progress of the file import process. If the file was successfully imported from the USB disk the controller will display one of the messages below:
Import Failure
Import Success: Any key to reboot
File Open Failed
File Read Failed Invalid CFG File Invalid Model Wrong SW Version
Corrupt CFG File Wrong file Size
Indicates that there were problems connecting to or accessing the USB stick.
The configuration file import succeeded and will be ready for use after reboot.
A config file could not be found on the USB stick or the USB stick file system could not be accessed.
The config file is too short (incomplete) or empty.
The imported file is not a valid config file.
The imported config file is not for this controller model.
The version of the imported config file is not compatible with this controller software version.
The imported config file is corrupt. (The checksum failed.)
The size of the imported config file is wrong.
Figure 20 Config Menu
34
Page 38
5.14 Upgrade Menu
This menu is available in both WCT400 and WCT410 models. It is used to upgrade the software to a newer version. If a new version of the software is available, an upgrade file will be posted on our web site. Save this file to a USB flash disk drive. It needs to be the only executable (.exe file extension) file stored on the root directory of the stick. Press the Enter key to import the software upgrade file from the stick to the controller.
The controller will display the progress of the file import process. If the file was successfully imported from the USB disk the controller will display Transfer Success. The controller will automatically reboot and come up with the new software installed.
Upgrade
To check that it was successful, turn off power to the controller, then press the Enter key while turning power on. The controller will show the software version, which should match the name of the upgrade file that you used.
The controller will display the progress of the file import process. If the file was successfully imported from the USB disk the controller will display Transfer Success. The controller will automatically reboot and come up with the new software installed.
If the software upgrade fails, you will see one of the following messages:
UpgradFileInvald
No Upgrade File
CorrptUpgradFile Flash Failure
The file found on the USB stick is for the wrong product, or is corrupt. Try getting the correct upgrade file and make sure it’s the only upgrade file on the stick.
There is no upgrade file stored on the stick, or the file is named incorrectly.
Try getting a new copy of the file.
The flash memory on the processor board has a problem. Repair or replace the front panel assembly.
Figure 21 Upgrade Menu
35
Page 39
6.0 MAINTENANCE
The WCT controller itself requires very little maintenance. Wipe with a damp cloth. Do not spray down the controller unless the enclosure door is closed and latched.
6.1 Electrode Cleaning
NOTE: The controller must be recalibrated after cleaning the electrode.
Frequency
The electrode should be cleaned periodically. The frequency required will vary by installation. In a new installation, it is recommended that the electrode be cleaned after two weeks of service. To determine how often the electrode must be cleaned, follow the procedure below.
1. Read and record the conductivity.
2. Remove, clean and replace the conductivity electrode.
3. Read conductivity and compare with the reading in step 1 above.
If the variance in readings is greater than 5%, increase the frequency of electrode cleaning. If there is less than 5% change in the reading, the electrode was not dirty and can be cleaned less often.
Cleaning Procedure
The electrode can normally be cleaned using a cloth or paper towel and a mild detergent. If coated with scale, clean with a dilute (5%) hydrochloric acid solution. Occasionally a electrode may become coated with various substances that require a more vigorous cleaning procedure. Usually the coating will be visible, but not always. To clean a coated electrode, use fine grit abrasive, such as emery paper. Lay the paper on a flat surface and move the electrode in a back and forth motion. The electrode should be cleaned parallel to the carbon electrodes, not perpendicular.
CLEAN IN
THIS DIRECTION
Figure 22 Cleaning the Electrode
36
Page 40
6.2 Replacing the Fuses
CAUTION: Disconnect power to the controller before opening front panel!
Locate the fuses on the circuit board at the back of the controller enclosure. (See figure 3.) Gently remove the old fuse from its retaining clip and discard. Press the new fuse into the clip, secure the front panel of the controller and return power to the unit.
Warning: Use of non-approved fuses can affect product safety approvals. Fuse ratings depend on controller power rating. Specifications are shown below. To insure product safety certifications are maintained, it is recommended that a Walchem fuse be used.
F1 Fuse Walchem P/N F2 Fuse Walchem P/N 5 x 20 mm, 1.0A, 250V 103163 5 x 20 mm, 6A, 250V 102834
7.0 TROUBLESHOOTING
CAUTION: Disconnect power to the controller before opening front panel!
Troubleshooting and repair of a malfunctioning controller should only be attempted by qualified personnel using caution to ensure safety and limit unnecessary further damage. Contact the factory.
7.1 Error Messages
TEMP ERROR
This error condition will stop both conductivity and pH control. It indicates that the temperature signal from the conductivity electrode is no longer valid. This prevent controlling based upon a false pH or conductivity reading.
Possible Cause Correction Action
Green or white electrode wire disconnected. Reconnect. Faulty electrode. Replace electrode. Revert to manual temperature compensation by
SENSOR ERROR
This error condition will stop conductivity control. It indicates that the conductivity signal from the electrode is no longer valid. This prevents controlling based upon a false conductivity reading.
Possible Cause Correction Action
Back or red electrode wire shorted. Disconnect short. Faulty electrode. Replace electrode. Faulty controller. Verify via failed self test.
HIGH ALARM
The summary screen will display an H at the right end of the bar graph if the conductivity rises above the high conductivity alarm set point. If your unit is wired for alarm output, the alarm relay will trip. The controller will continue to check the conductivity, and the bleed and/or feed outputs will be allowed to be activated.
Possible Cause Correction Action
Dirty electrode Clean electrode (see Sect. 6.1) Faulty solenoid valve Repair or replace solenoid valve. Faulty electrode Evaluate (see Sect. 7.3). Check Temp display. Improper wiring of valve or controller Correct wiring. See Section 3.4. Conductivity rose over alarm limit while biocide lockout occurred. Allow normal bleed to occur. Clogged Y-strainer in bleed line Clean Y-strainer. Faulty bleed relay Replace relay. (Consult factory.)
cycling power off and on.
37
Page 41
BLEED TIMEOUT
This error condition will stop conductivity control. It is caused by the bleed output being activated for longer than the programmed Bleed Time Limit.
Possible Cause Corrective Action
Programmed value too low for normal conditions. Increase Bleed Time Limit. Bleed flow rate too low.
Bleed valve not opening. Check for faulty bleed valve.
Check for clogged strainer. Check for insufficient pressure differential.
Check bleed valve wiring. Check controller relay.
LOW ALARM
The summary screen will display an L at the left end of the bar graph and the alarm relay will trip. The controller will continue to check the conductivity and feed inhibitor as programmed.
Possible Cause Corrective Action
Sensor disconnected Reconnect. Check cable for continuity. Sensor dry Check tee for obstruction. Verify flow. Change location of electrode. Pre-bleed set too low Check pre-bleed setting compared to % low. Solenoid valve stuck open Repair or replace solenoid valve. (Consult your distributor). Faulty electrode Evaluate (see Section 7.3). Replace if necessary. Improper wiring of electrode Correct wiring. See Section 3.4. Faulty bleed relay Replace relay. (Consult factory.)
NO FLOW
This error message will appear if the flow/no flow switch closure is opened. The error message will be displayed and the alarm relay will trip. The conductivity is updated but all outputs are deactivated.
Possible Cause Corrective Action
No flow Check recirculating pump, valves, etc. Faulty flow switch/cable Check for open circuit. Disconnect flow switch and short the input
Insufficient pressure drop through sample stream Check piping.
with a piece of wire. If "no flow" message disappears, then switch or cable may be faulty.
CAL FAIL
The controller allows a maximum of ±50% correction to the conductivity reading. This is a cumulative correction factor which means that if at one time the electrode is calibrated by 30%, the next time it is calibrated it will only accept an adjustment of +20% before reporting a Cal Fail.
The most likely problem is that the electrode needs to be cleaned. Refer to section 6.1.
TimeErr:SetTime!
This error message is triggered by corruption of the clock chip. Biocide additions will be locked out until the clock time is reset.
Possible Cause Corrective Action
Power spikes, high frequency noise If this occurs frequently, install a power line filter.
7.2 Conductivity Readout Does Not Change
If the readout is stuck at or near zero:
Possible Cause Corrective Action
Dry electrode Check for flow through system. Electrode is disconnected. Check wiring to electrode. Go to self-test menu, as described in
If the readout is stuck at another number:
Possible Cause Corrective Action
Dirty or faulty electrode Evaluate electrode (section 7.3). Stagnant sample Check system for proper flow.
section 5.2. If readout changes to 900-1100, the problem is with electrode or connections. See section 7.3 If still at zero, problem is with controller. Consult the factory.
38
Page 42
7.3 Procedure for Evaluation of Conductivity Electrode
Can be used for troubleshooting low conductivity, high conductivity, conductivity stuck at 0, and/or conductivity stuck at a number other than 0.
Try cleaning the electrode first (refer to Sect. 6.1).
To find out if the electrode or the controller is faulty, step through the Self-Test menu, as described in section 5.2. The display should read 1000 ±100 µS/cm. This indicates that the controller is OK and the problem is in the electrode or its connections. If it does not read 1000 ±100 µS/cm, return the control module for repair.
To check the electrode, check the electrode connections to the terminal strip (refer to Figure 3). Make sure that the correct colors go to the correct terminals, and that the connections are tight. Restore power and see if the conductivity is back to normal. If not, replace the electrode.
8.0 SERVICE POLICY
The WCT series Cooling Tower Conductivity Controller has a 2-year warranty on electronic components and a 1-year warranty on mechanical parts (keypad, terminal strip and relays).
We stock circuit boards for immediate exchange after we have isolated the cause of the problem.
Factory authorized repairs that are received by next-day-air will be returned within 24 hours. Normal priority for returns is two weeks.
Out of warranty repairs or circuit board exchanges are done on a flat fee basis after the warranty is expired.
39
Page 43
Model Code
WCT400 WCT410
VOLTAGE
1 = 120 VAC, prewired 5 =
100-240 VAC, conduit
OUTPUT
N = No data output 4 = Single 4-20 mA output 2 = Dual 4-20 mA output (WDT)
SENSOR
N = No electrode
WCT400/410
1 = PP/Graphite electrode & tee, 20ft. (6.1m) cable (for in-line or submersion mounting) 2 = PP/Graphite electrode & flow switch manifold mounted on PP panel, 5 ft. (1.5m) cable 4 = High pressure electrode (up to 300 psi), 20 ft. (6.1m) cable 5 = High pressure electrode & flow switch manifold on PP panel, 5 ft. (1.5m) cable 6= PP/SS electrode & tee, 20 ft. (6.1m) cable 7= PP/SS electrode & flow switch manifold on PP panel, 5 ft. (1.5m) cable
USB FEATURES
U = Integrated datalogging, event/reset logging, and configuration file import/export
FIVE BOYNTON ROAD HOPPING BROOK PARK HOLLISTON, MA 01746 USA
Voltage Output Sensor U (USB) -
TEL: 508-429-1110 FAX: 508-429-7433 Web: www.walchem.com
40
Loading...