4.1 EXAMINATION AND ADJUSTMENT BEFORE TAKING MEASUREMENTS
4.2 MEASUREMENT OF VOLTAGE
4.3 TIME MEASUREMENTS
4.4 PHASE DIFFERENCE MEASUREMENTS
4.5 X – Y MODE APPLICATIONS
5 . ACCESSORIES
1
WARNING
1. SAFETY PRECAUTION
The A oscilloscope is easy to operate and highly reliable. It is an ideal instrument for research,
production, education, and development. It is a compact and portable oscilloscope with a
frequency bandwidth of 10MHz and sensitivity of 5mV/DIV-5V/DIV. Supplied as standard,
with a switchable X10: X1 probe which extends the sensitivity up to 50V/div. The horizontal
timebase is 0.1S/DIV to 0.1µS/ D IV.
The instrument is designed and tested in accordance with EN publication 61010, CAT II,
Pollution degree II and Overvoltage 600V. The instrument has been tested in accordance to the
following EC Directives (EMC) :
a. EN50082
b. EN55011
c. EN610000-3-2
d. EN61000-3-3
The instrument complies with the requirements of the European Council Directive 89/336/EEC
(EMC Directive) and 73/23/EEC (Low Voltage Directive). To ensure that the instrument is
used safely, follow all safety and operating instructions in this manual. If the instrument is not
used as described in this manual, the safety features might be impaired.
Non compliance with the warnings and/or the instructions for use may damage the
instrument and/or its components or injure the operator.
Take extreme care under the following conditions when using the instrument:
For your own safety and that of the instrument, you must follow the procedures described
in this instruction manual and especially read all the notes proceeded by the symbol
carefully.
Do not use this instrument in a location where there is explosive gas in the vicinity. The use
of this instrument in a location where there is explosive gas could result in explosion.
If there is any smoke, abnormal odor, or abnormal sound coming from this instrument,
immediately switch off the power and disconnect the power cord. Continuous using of this
instrument under these conditions could result in electrical shock or fire. After
disconnecting the power cord, contact the service offices for repair. Repair by the user is
2
dangerous and should be strictly avoided.
Sensitivity
5mV/DIV. ~5V/DIV ±3%
Trimming Ratio
2.5:1
Rise Time
35ns
Bandwidth(-3dB)
DC:0~10MHz AC:10Hz~10MHz
Input Impedance
1MΩ±3%,30 pF ±5pF
Max. Input Voltage
400V pk
Trigger Sensitivity
Int 1 div., Ext 0.3V
Ext. Trigger Input Impedance
1MΩ 30pF
Ext. Trigger Max. Input Voltage
400Vpk
Trigger Sources
Int, Line, Ext
Trigger Mode
Norm, AUTO, TV
Sweep Time
0.1S / DIV ~0.1µS / DIV ±3%
Trimming Ratio
2.5:1
Take care not to allow water to get into this instrument. The use of this instrument in a wet
state could result in electrical shock or fire. If water or other foreign matter has penetrated
this instrument, first switch the power off, remove the power cord and call for repair.
Do not place this instrument on an unstable or slanting surface . The dropping or turning
over of this instrument could result in electrical shock, injury or fire. If this instrument has
been dropped or its cover has been damaged, switch the power off, remove the power cord
and call for repair.
Do not allow any foreign matter such as metal or inflammable substance to get into the
instrument via the air holes. The penetration of any foreign matter from the ventilation
holes could result in fire, electrical shock, or power failure.
Use this instrument with the rated AC power source. Use of this instrument with a voltage
other than specified could result in electrical shock, fire or power failure. The usable
power voltage range is marked on the rear panel.
Do not remove either the cover or panel.
Take sufficient care when measuring high voltages.
Do not modify this instrument.
Avoid use of damaged cables .
2. SPECIFICATIONS
2.1 VERTICAL SYSTEM
2.2 TRIGGER SYSTEM
2.3 HORIZONTAL SYSTEM
3
Sensitivity
0.2V/DIV〜0.5V/DIV
Bandwidth(-3dB)
DC: 0~1MHz AC: 10Hz~1MHz
Waveform
Symmetric Square Wave
Range
05.V ±2%
Frequency
1kHz ±2%
Display Area
8 × 10DIV 1DIV=6mm
Accelerating Voltage
1200V
Display Color
Green
Voltage Range
110V±10% , 220V±10%
Frequency
50Hz ±2Hz
Power Consumption
25W
Weight
3kg
Dimensions (H x W x D)
190 ×130 ×270mm
Working temperature
5°C ~ 40°C
Storage Environment
-30°C ~60°C, 10~80%RH
Working Altitude
≤2000m
Pressure-proof test
1500V 1min
2.4 X-Y MODE
2.5 CALIBRATION SIGNAL
2.6 CRT
2.7 POWER SOURCE
2.8 PHYSICAL FEATURES
2.9 WORKING ENVIRONMENT
2.10 PRESSURE-PROOF TEST
4
3.1.1 Front Panel Fig 3-1
3.1.2 Rear Panel Fig 3-2
3. CONTROL AND INDICATORS
3.1 Control Panel Position
5
NO.
SWITCHES
FUNCTIONS
1
POWER SWITCH
Power on/off
2
POWER LIGHT
Lights when power on
3
INTENSITY
Controls brightness of display
4
FOCUS
After obtaining appropriate brightness with INTENSITY,
adjust FOCUS for clearest line
5
CALIBRATION
Provide symmetric square wave for 0.5V range,
adjusting the vertical & horizontal sensitivity.
6
TIME/DIV
Selects the sweep rate
7
Horizontal
POSITION
Horizontal positioning control of trace on the screen
8
HORIZONTAL
VA R
Provides continuously variable sweep rate, turn clock wise
to the end is the calibrating position
9
LEVEL
Control signal trigger to sweep at certain level
10
+/-
+: Triggering occurs when trigger signal crosses trigger
TV: used to show TV signals
11
AUTO/NORM
AUTO : a single trace shown on screen even no signal.
EXT/X: X-Y switch
12
INT/EXT/LINE
Switch to select the Trigger Source INT/EXT/LINE
13
Ext. Trig Input
Terminal
When switching [11] to EXT/X, it’s X-Y input terminal;
When switching [12] to EXT, it’s Ext. Trig. Input terminal.
14
ATT
Continuously adjusting X-Y mode sensitivity, turn
clockwise to obtain the highest sensitivity
15
VOLTS/DIV
Adjusting sensitivity of vertical system
16
Vertical POSITION
Control vertical position of trace on the screen
17
VERTICAL VAR
Continuously adjusting the sensitivity of vertical system,
turn clockwise to the end is the calibration position
18
coupling options
(AC⊥DC)
Selects input coupling options
19
INPUT
Vertical input terminal
20
POWER INPUT
AND FUSE
power input connector (refer to the rear panel for voltage)
3.2 FUNCTIONS OF CONTROL SWITCHES
frequency=1KHz. Used for adjusting 10:1 Capacitor and
TV
EXT/X: X-Y
level in a positive-going direction.
-: Triggering occurs when trigger signal crosses trigger
level in a negative-going direction
Automatically reverts to triggered sweep
operation when adequate triggered signal is
present. Needs to adjust the Level.
NORM: No trace on screen if no signal. Trace is only
generated when adequate trigger signal is present
CONNECTOR
6
110V±10% , 220 V ±10%;
CONTROL SWITCHES
POSITION
INTENSITY [3]
Center
FOCUS [4]
Center
POSITION [7] [16]
Center
VOLTS/DIV [15]
0.1V
VAR [8] [17]
Calibration position
AUTO/NORM [11]
Auto
TIME/DIV [6]
0.5mS
SLOPE +/- [10]
+
INT/EXT/LINE [12]
Int
AC⊥DC [18]
DC
3.3 OPERATING INSTRUCTIONS
3.3.1 VOLTAGE CHECKING
A oscilloscope is set for 110V±10% , 220V ±10% voltage. Before connecting the unit to the
mains supply, make sure that the correct voltage is being used. Incorrect mains voltage may
cause damage to the instrument.
3.3.2 BASIC OPERATION
(1) POSITION FOR CONTROL SWITCHES
(2) OPERATION
a. Power on [1],
b. Power lights at [2],
c. Allow a warm-up period of 5 minutes, then adjust the intensity[3],
d. Adjust focus [4] for clearest line. If unstable, adjust level [9] .
(3) Horizontal Level Adjustment :
a. Slightly turn screws [21][23] counter clockwise, but do not remove;
b. Insert a straight end screw driver to [22], and whilst observing the waveform,
turn the screw driver in order to set the waveform parallel with the horizontal
line ;
c. Tighten screws [21] [23].
(4) Connect the calibration signal [5] to Y input terminal[19] with 1:1 probe
(5) Adjust positions [7] [16] to obtain waveform like Fig 3-3.
7
Fig 3-3
3.3.3 VERTICAL SYSTEM OPERATION
(1) VOLTS/DIV switch should be turned to the correct position following the input signal
range. Adjust position [16] to show the whole waveform within the available area.
Adjust with VAR [17] if necessarily, trimming ratio is 2.5:1.
(2) Input coupling options: “DC”is used for observing a signal with a direct current
content such as logic or static signals, “DC” must be used with low frequency. “AC”
is used for observing the AC component of signals. “⊥” (ground) is used to establish
a trace at a zero volt reference.
(3) X-Y OPERATION: When [11] set at EXT/X, The oscilloscope is used for X-Y
operation, at this moment Input [19] is Y-axis with the same sensitivity, Input [13] as
X-axis, [14] can be adjusted continuously within 0.2V/DIV~0.5V/DIV.
3.3.4 TRIGGER SOURCE
In Fig 3-1, [12] provides 3 sources for selection, INT trigger, EXT trigger input from
[13], LINE input from power source.
3.3.5 HORIZONTAL SYSTEM OPERATION
(1) Sweep setup : turn the sweep switch to the correct position according to the signal
frequency and adjust POSITION [7] to show the whole waveform within the available
area. Adjust with VAR [8] if necessarily, trimming ratio is 2.5:1.
(2) There are 3 kinds of Trigger sources: [11] “AUTO”auto sweep, when a triggering
signal is applied, Level [9] will adjust to correct position. The screen showing a steady
8
free run waveform requires a frequency higher than 20Hz; [11] “NORM”while
waiting for sweep with no trace. When an input signal occurs the scope is triggered to
sweep
(3) and show the waveform . [10] “TV” used to determine TV signals.
(4) SLOPE selection: Used to select whether the trigger signal crosses trigger level in a
positive or a negative-going direction.
(5) Level setup : Use to adjust signal sweep on a level at a starting reference.
3.3.6 SIGNAL CONNECT
(1) Probe operation :
Use 10:1 to 1:1 switchable probes. When using 10:1 probe, input impedance is 10MΩ
16 pF. While if 1:1 is used for observing a signal, input impedance is 1MΩ 30 pF. At
this stage, please consider the affect of the probe in certain circuits.
(2) Probe Adjustment
Before using , 10:1 probe must be adjusted correctly, see Point 4.1.2
4 Measurement
4.1 EXAMINATION AND ADJUSTMENT BEFORE MEASUREMENT
In order to ensure accuracy and prevent errors, the following observations should be made
before taking a measurement.
4.1.1 TRACE ROTATION
The horizontal trace on the screen should be parallel with the horizontal line. Due to the
earth’s magnetic field and other factors which causes horizontal trace leaning you must
examine the following before taking measurements:
(1) Adjust the front panel control to obtain a horizontal trace on the screen .
(2) Adjust the vertical position control to ensure the horizontal line in is centered.
(3) If the horizontal line/trace is not level, then please following point 3.3.2(3) to correct
it .
4.1.2 PROBE COMPENSATION
Adjustment of the probe is required to compensate for the variations caused by difference
in oscilloscope input characteristics:
(1) Follow step 3.3.2, setup front panel controls, and obtain a sweep baseline.
(2) Set VOLTS/DIV to 10mV/DIV.
(3) Connect CH1 10:1 probe to input terminal, and connect to the “CAL” terminal.
(4) Follow Chapter 3 to operate relative controls, to obtain a waveform on the screen as
9
shown in figure 4.1
(5) Observe the waveform compensation and adjust the LF compensation controls as
shown in figure 4-2
Good Compensate Over Compensate Under Compensate
Fig 4-1
LF COMP
Fig 4-2
4.2 MEASUREMENT
4.2.1P-P Voltage Measurements
Step :
(1) Input signal to INPUT [19] terminal.
(2) Setup VOLTS/DIV and observe waveform, set waveform display on the screen within
5 divisions, and turn VAR clockwise to the calibration position.
(3) Adjust level to make waveform steady.
(4) Adjust sweep controls to show at least one cycle of the waveform is on the screen.
10
(5) Adjust vertical position to ensure the bottom of waveform lies on a horizontal axis on
Fig 4-3
A
B
the screen. Fig 4-3A.
(6) Adjust horizontal position to ensure the top of waveform lies center of vertical axis.
Fig 4-3B.
(7) Read the divisions between A-B on vertical direction.
(8) Calculate the signal Vp-p using the formula below :
Vp-p= DIV of vertical direction× Sensitivity
For example, In Fig 4-3, vertical divisions of A-B is 4.1 DIV, sensitivity of the 10:1
probe is 2V/DIV, then Vp-p=2×4.1=8.2(V)
4.2.2 DC VOLTAGE MEASUREMENT
STEP:
(1) Setup front panel connector to obtain a sweep baseline on the screen.
(2) Setup input coupling options as “⊥”.
(3) Setup POSITION, let sweep baseline to coincide with horizontal center, define it as
the zero reference level.
(4) Input signal into terminal.
(5) Set input coupling to “DC”, adjust VOLTS/DIV, so that the waveform is shown
centrally on the screen, turn VAR to the calibration position.
(6) Read the divisions between the zero reference level to the waveform from the test
input.
(7) Calculate the DC voltage:
V= divisions on vertical axis × sensitivity × direction(+/-)
Shown in Figure 4-4, zero reference level at the center, use 10:1probe, sensitivity is
2V/Div, 2 points as A & B, A is 1.5 Div. over the zero reference level, B is 3Div.
below the zero reference level. DC voltage level of the 2 points are :
VA = 1.5 x 2 x (+) = 3 V VB = 3 x 2 x (-) = -6 V
11
Fig 4-4
Fig 4-6
B
A
Zero-volt
Reference
Level
Level
A
B
4.3 TIME MEASUREMENTS
4.3.1 TIME SPACE MEASUREMENTS
This is a procedure for making time (period) measurements between two points on a
waveform:
(1) Connect the signal to be measured to the input terminal [19].
(2) Adjust level to obtain steady waveform.
(3) Turn VAR clockwise to the calibration position, and set sweep controls to obtain a
normal display of 1-2 signal cycles.
(4) Using the vertical and horizontal positions, set two points to be measured in the
waveform on the same horizontal level.
(5) Measure the distance between the two points , the measurement is calculated by the
following equation:
Distance between 2 points (DIV) × rate of sweep (t/Div)
TIME (S) = --------------------------------------------------------------------
Horizontal factor
Shown in figure 4-6, distance between A & B is 8 Div. sensitivity is 2 µS/Div, Horizontal
factor x 1, then Time measurement is 16µS
12
4.3.2 CYCLE & FREQUENCY MEASUREMENTS
Fig. 4-7
10%
90%
100%
2.4DIN
Shown in Fig 4-6, frequency measurements are made by measuring the time period of
one cycle of waveform (T), and calculating the frequency that equals the reciprocal of the time
period. For example, T=16µS, then frequency is: 1
F =1/T = ------------------ = 62.5 KHz
16 ×10-6
4.3.3 PULSE RISE TIME AND FALL TIME MEASUREMENTS
For rise time and fall time measurements, the 10% and 90% amplitude points are used as
starting and ending reference points.
(1) Apply a signal to the input jack [19].
(2) Use the VOLTS/DIV and VAR controls to adjust the waveform peak to peak height to
five divisions.
(3) Adjust vertical position so that the tops of the waveform coincide with 100% point,
while the bottoms of the waveform coincide with 0% point.
(4) Adjust Sweep switch to obtain the positive-going direction or negative-going direction
of the waveform on the screen.
(5) Use the horizontal POSITION control to adjust the 10% points to coincide with a
vertical reference line.
(6) Measure the horizontal distance in divisions between the 10% and 90% points on the
waveform (divisions).
(7) Pulse rise time and fall time measurement is calculated by the following equation: