Vishay DXP10, DXP15 Operator's Manual

WEIGH SYSTEM TECHNOLOGY
BLH
DXP10/15 Weigh Transmitters
Operator’s Manual
NOTICE
BLH makes no representation or warranties of any kind whatsoever with respect to the contents hereof and specifically disclaims any implied warranties or merchantability or fitness for any particular purpose. BLH shall not be held liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this publication or its contents.
BLH reserves the right to revise this manual at any time and to make changes in the contents hereof without obligation to notify any person of such revision or changes.
Call (781) 298-2216 for BLH Field Service
Table of Contents
SECTION 1. General Information ...................................................................................................................... 1-1
1.1 INTRODUCTION ...................................................................................................................................... 1-1
1.1.1 General Description. ............................................................................................................................. 1-1
1.1.2 Standard Instrument. ............................................................................................................................ 1-1
1.2 OPTIONS ................................................................................................................................................. 1-1
1.2.1 Mounting Options. ................................................................................................................................ 1-1
1.2.2 Optional Terminal Computer Interface. ................................................................................................ 1-1
1.2.3 Optional Modbus RTU Interface. .......................................................................................................... 1-2
1.2.4
Optional Fisher ProVox Protocol .......................................................................................................... 1-2
1.2.5 Optional Allen-Bradley Remote I/O ...................................................................................................... 1-2
1.2.6 Analog Option. ...................................................................................................................................... 1-2
1.2.7 Software Filter Option. .......................................................................................................................... 1-2
1.3 DXp-10/15 SPECIFICATIONS ................................................................................................................. 1-2
1.4 ORDERING SPECIFICATIONS ............................................................................................................... 1-4
1.5 WARRANTY POLICY ............................................................................................................................... 1-4
1.6 FIELD ENGINEERING ............................................................................................................................. 1-4
SECTION 2. Installation..................................................................................................................................... 2-1
2.1 INTRODUCTION ...................................................................................................................................... 2-1
2.1.1 General. ................................................................................................................................................ 2-1
2.2 MOUNTING .............................................................................................................................................. 2-1
2.2.1 Standard Units. ..................................................................................................................................... 2-1
2.2.2 Explosion-Proof and Division 2 Options. .............................................................................................. 2-1
2.3 ELECTRICAL ........................................................................................................................................... 2-2
2.3.1 Transducer Inputs. ................................................................................................................................ 2-2
2.3.2 Serial Communication. ......................................................................................................................... 2-2
2.3.3 Analog Output. ...................................................................................................................................... 2-3
2.3.4 Mains (AC) Power (Figure 2-4)............................................................................................................. 2-3
2.3.5 Auxiliary I/O Port. .................................................................................................................................. 2-3
SECTION 3. Configuration ................................................................................................................................ 3-1
3.1 GENERAL ................................................................................................................................................ 3-1
3.2 SET-UP .................................................................................................................................................... 3-1
3.2.1 Power Selection .................................................................................................................................... 3-1
3.2.2 Excitation Voltage. ................................................................................................................................ 3-1
3.2.3 Serial Format, Address, and Baud Rate. .............................................................................................. 3-1
3.2.4 Input Range Selection. ......................................................................................................................... 3-1
3.3 CALIBRATION .......................................................................................................................................... 3-1
iii
3.3.1 Digital Calibration. ................................................................................................................................ 3-1
3.3.2 Analog Calibration. ............................................................................................................................... 3-2
3.3.3 Monitor Mode Calibration. .................................................................................................................... 3-2
3.3.4 LCp-40 Calibration. ............................................................................................................................... 3-2
3.4 CHECK CAL ............................................................................................................................................. 3-5
3.5 FILTER SELECTION ................................................................................................................................ 3-6
3.5.1 Standard Filter. ..................................................................................................................................... 3-6
3.5.2 Optional Digital Filtering. ...................................................................................................................... 3-6
3.6 Factory Default Calibration ....................................................................................................................... 3-7
SECTION 4. Serial Communication .................................................................................................................. 4-1
4.1 GENERAL ................................................................................................................................................ 4-1
4.1.1 LCp-40 Digi System Network. .............................................................................................................. 4-1
4.1.2 Standard Simplex Output (Continuous Output). ................................................................................... 4-1
4.1.3 Computer/Terminal Interface (Optional). .............................................................................................. 4-2
4.1.4 MODBUS BTU Protocol (Optional). ..................................................................................................... 4-4
4.1.5 Fisher ProVox Protocol (Optional) ........................................................................................................ 4-7
4.1.6 Allen-Bradley Remote I/O (Optional). ................................................................................................... 4-8
SECTION 5. Operation ...................................................................................................................................... 5-1
5.1 GENERAL ................................................................................................................................................ 5-1
5.2 GROSS WEIGHT WEIGHING ................................................................................................................. 5-1
5.3 ZERO OPERATION ................................................................................................................................. 5-1
5.4 NET WEIGHT WEIGHING ....................................................................................................................... 5-1
5.5 TARE OPERATION .................................................................................................................................. 5-1
5.6 ERROR DETECTION ............................................................................................................................... 5-1
5.7 CHECK CAL OPERATION ....................................................................................................................... 5-1
iv
SECTION 1. General Information
1.1 INTRODUCTION
1.1.1 General Description
The DXp-10 and DXp-15 (Figure 1-1) are microprocessor based weight transmitters. DXp­10/15 transmitters sum and convert the millivolt signal from up to four strain gage type load cells into a digital and/or analog voltage or current signal (up to eight cells can be summed using an external 308A summing unit). Bridge excitation voltage is factory selectable for 10 or 15 volts. Units are available in NEMA 4, NEMA 4X or explosion proof enclosures for field mounting. The unit operates from either 110 or 220 VAC at 50/60 Hz, and is equipped with screw terminal connections for power input, and serial/analog outputs. Set-up and calibration is per-formed using three internal switches (see SECTION III). Communication baud rate and instrument address values are selected via a bank of internal DIP switches (see SECTION III).
Class I, II, Division 2, Group A-G hazardous location requirements and are optionally available with FM and CSA approvals and certifications.
1.2 OPTIONS
1.2.1 Mounting Options
For corrosive, hose down, or sanitary environments, a NEMA 4X stainless steel enclosure is available. An explosion proof enclosure is available for Class I, II, Division 1, Group B-G locations. Note: BLH 404 or 405 Intrinsic Safety Barriers must be specified for weigh systems located in a Division I area.
The DXp-10 Transmitter performance specifications are designed for inventory and other static weighing applications where moderate resolution and speed is required. The DXp-15 Process Transmitter is equipped with high performance circuitry that provides the greater resolution and speed for dynamic process applications. Externally and in regard to set-up and operation, each model is identical
Using RS 485 serial communication protocol, up to sixteen DXp transmitters can be networked to an LCp-40, 41, or 42 Network Controller. The serial format can also be configured to provide point to point or network communication to a customer supplied computer.
1.1.2 Standard Instrument
The standard instrument includes an RS 485 serial port with BLH Digi-System network or a simplex output protocols, a four cell summing circuit, 10 or 15 volt excitation, averaging filter, and a NEMA 4 mild steel, painted enclosure. Standard instruments are designed to meet
Figure 1-1. DXp-10 Weight Transmitter
1.2.2 Optional Terminal Computer Interface.
The terminal/computer interface option provides a simple mnemonic half-duplex ASCII communications protocol via a built-in macro language consisting of 1 to 3 character command strings (reference Table 4-3).
This powerful feature allows direct keyboard control (using easily remembered commands) of
1-1
DXp-10/15 calibration, and programming or recall of weight variables (gross, net, tare, zero, etc.) An on-line help function is available by transmitting an ASCII Ir.
Easily learned macro language syntax greatly simplifies the writing of a host computer communication interface (customer supplied).
1.2.3 Optional Modbus RTU Interface.
The Modbus interface option provides a simple seamless communication link to any PC, PLC or DCS with a mod-bus RTU Driver Option. Use of this industry standard protocol (see paragraph
4.1.4) eliminates the requirement for custom
software driver development.
1.2.4
DXp-15 transmitters may be ordered with the Fisher ProVox protocol. ProVox protocol allows the DXp-15 to interface directly with a Fisher ProVox distributed control system (DCS).
Optional Fisher ProVox Protocol
1.2.5 Optional Allen-Bradley Remote I/O
Optional Allen-Bradley Remote I10 (RIO) allows 13Xp¬15 transmitters direct access to a PLC master processor. In essence, DXp-15 transmitters become direct coprocessors with the PLC. DXp-15 data is read directly into the
main ladder logic program without using BCD or serial interface cards. A simple three wire RIO network connection ties all units together, even if multiple DXp15s are used. Within the RIO network link, each DXp-15 is addressed as 1/4 'logical rack'. Since a PLC logical rack consists of 128 input and 128 output bits (or points), DXp units communicate 32 input and 32 output bits; 1/4 logical rack format. For a full description of logical rack addressing and data communications formatting, reference the Allen Bradley Remote I/O technical manual, BLH part number TM010.
1.2.6 Analog Option.
An optional analog output provides both a 4-20 mA and a 0-10 VDC output representing the gross weight value. The circuit uses a 12 bit D-A converter providing up to 4096 counts of resolution. The current output drives up to a 1000 ohm load and the voltage output will operate with a 25,000 ohm minimum load resistance.
1.2.7 Software Filter Option.
The digital filtering option offers eight software filtering selections. The filter algorithms dampen noise by averaging successive A-D conversions. Digital filtering allows East response to true weight changes by setting limits on averaging based upon the magnitude of signal change. Filter selections are presented in SECTION III. Note: BLH tech note TD-071 provides a full discussion of filter operation.
1.3 DXp-10/15 SPECIFICATIONS
PERFORMANCE
Resolution: DXp-10 20,000 counts DXp-15 50,000 counts Sensitivity DXp-10 1.0 μV/count
DXp-15 0.5 μV/count Full Scale Range 25 of 35 mV (selectable) Dead Load Range 100% Input Impedance 10 M-ohms, max. Load Cell Excitation 10 V for up to eight 350 ohm load cells (250 mA) (Factory Selectable) 15 V for up to six 350 ohm load cells (250 mA) Linearity ± 0.01% of full scale
1-2
Temperature Coefficient
Span 2ppm/°C Zero ± 2ppm/°C Common Mode Rel. 100 db or better at or below 35Hz Normal Mode Rej. 100 db or better at or below 35Hz Conversion Speed DXp-10 400 msec
DXp-15 50 msec
Environment
Operating Temperature -10 to 55°C (12 to 131°F) Storage Temperature -20 to 85°C (-4 to 185°F) Humidity 5 to 90% rh, non-condensing Voltage 117/230 1: 15% 50/60 Hz Power 10 watts max Parameter Storage EEPROM EMVRFI Shielded from typical industrial interference
ENCLOSURE
Dimensions (NEMA 414X) 11.5 x 8.0 x 4.3 HWD Explosion Proof 12.875 x 10.875 x 8.188 HWD
OPTIONS
Isolated Analog Output
Type 12 bit Digital to Analog Conversion Voltage 0 to 10 volt (25K ohm min load) Current 4 to 20 mA (1000 ohm max load)
SERIAL COMMUNICATION
LCp-40 Network (Standard)
Type RS 485 Half Duplex (Multi-Drop) Baud 56.7k
Simplex Data Output (Standard)
Interface Type RS 485 (Simplex) Data Format Simplex ASCII Data
7 Data Bit Even Parity 1 Stop Bit
Terminal/Computer Interface (Optional)
Interface Type RS 485 Half Duplex (Standard) Baud 1200 or 9600 Protocol Duplex Command/Response Format
MODBUS RTU Protocol (Optional) Fisher ProVox Protocol (Optional) Allen-Bradley Remote I/O (Optional) Consult factory for details
1-3
1.4 ORDERING SPECIFICATIONS
DXp-10 or DXp-15 [M]-[C]-[P]-[S] Includes: RS 485 Serial Output
[M] Mounting
(1) NEMA 4 Painted standard (2) NEMA 4X Stainless Steel (3) NEMA 7 & 9 Explosion-Proof Class I, H, Div. 1,2 Grp. BG (8) #1 & FM/CSA Approved [Class l, Div 2, Group ABCD FG] (9) #2 & FM/CSA Approved [Class l, Div 2, Group ABCD FG]
[C] Communication
(1) RS 485 LCp-40 Network standard (2) RS 485 LCp-40 Network and Terminal/Computer Interface (4) Allen Bradley Remote I/O (DXp-15 only) (5) MODBUS RTU Protocol (D4-15 only)
[P]Process Output
(1) No Process Output standard (2) 0-10 V & 4-20 mA analog
[S]Software
(1) Standard (2) Dynamic Digital Filtering (DXp-15 only)
apply beyond their normal span of life to any
1.5 WARRANTY POLICY
BLH warrants the products covered hereby to be free from defects in material and workmanship. BLH's liability under this guarantee shall be limited to repairing or furnishing parts to replace, f.o.b. point of manufacture, any parts which, within three (3) years from date of shipment of said product(s) from BLH's plant, fail because of defective workmanship or material performed or furnished by BLH. As a condition hereof, such defects must be brought to BLH's attention for verification when first discovered, and the material or parts alleged to be defective shall be returned to BLH if requested. BLH shall not be liable for transportation or installation charges, for expenses of Buyer for repairs or replacements or for any damages from delay or loss of use for other indirect or consequential damages of any kind. BLH may use improved designs of the parts to be replaced. This guarantee shall not apply to any material which shall have been repaired or altered outside of BLWs plant in any way so as, in BLH's judgment, to affect its strength, performance or reliability, or to any defect due in any part to misuse, negligence, accident or any cause other than normal and reasonable use, nor shall it
materials whose normal span of life is shorter than the applicable period stated herein. In consideration of the forgoing guarantees, all implied warranties are waived by the Buyer, BLH does not guarantee quality of material or parts specified or furnished by Buyer, or by other parties designated by buyer, if not manufactured by BLH. If any modifications or repairs are made to this equipment without prior factory approval, the above warranty can become null and void.
1.6 FIELD ENGINEERING
The field service department at BLH is the most important tool to assure the best performance from your application. The expertise and understanding of BLH's Field Engineers can solve even the most perplexing installation problem. Precise calibration and start-up procedures, performed by a qualified, experienced field engineer, assure not only the reliability of BLH components, but the integrity of the entire weigh system.
Call (Factory Number) (781) 298-2200
Ask for Field Service
Canada (416) 251-2554
1-4
SECTION 2. Installation
2.1 INTRODUCTION
2.1.1 General.
The DXp-10/15 is designed to 'installed within the length of the load cell(s) cable which is normally 35 ft or less. The standard NEMA 4 or optional NEMA 4X enclosures are suitable for an outdoor or wash down type environment. Both enclosures are provided with pre-punched holes for installation of conduit or cable fittings and holes for mounting to a bracket or wall.
2.2 MOUNTING
2.2.1 Standard Units.
The NEMA 4 and NEMA 4X enclosures are equipped with four pre-punched holes for mounting to a wall or bracket. A U-bolt can be used for mounting to a pipe support. The
instrument should be installed in a vibration-free location within the normal length of the load cell cables. If conduit is used, drains should be provided to reduce the possibility of condensate entering the enclosure. Outline dimensions for the standard DXp¬10/15 transmitter are presented in Figure 2-1.
2.2.2 Explosion-Proof and Division 2 Options.
DXp units may be ordered with an optional explosion-proof enclosure for use in Division I hazardous locations. Dimensions for the optional explosion-proof enclosure are provided in Appendix C.
For Division 2 applications, FM approved units are available as non-incendive devices in NEMA 4/4X enclosures with dimensions identical to those presented in Figure 2-1.
Figure 2-1. DXp-10 Outline Dimensions
2-1
Figure 2-2. Load Cell Connections.
2.3 ELECTRICAL
2.3.1 Transducer Inputs.
Up to four load cells can be connected to the summing circuit within the DXp. Connect individual load cells directly to the circuit board connectors as shown in Figure 2-2. Excitation and signal connection locations are dearly marked according to function and standard color code.
If more than four cells are required, an external 308A summing junction box must be used. Make all load cell connections in the 3084 unit, not the DXp-10/15. Connect the output leads of the 3084 summing box to the load cell #1 terminal blocks in the DXp. Sense leads from the 3084 must be connected to the sense terminal blocks in the DXp-10/15 (Figure 2-2). Jumpers JF1 and JP2 must be removed.
2.3.2 Serial Communication.
If a deadweight or substitution method of calibration is being used, the load cal cable can be shortened as required. The leads should be re­tinned before the final connection is made.
NOTE: If tension or universal load cells are used, red (-signal) and white (+ signal) leads may need to be reversed.
A terminal connector is provided for RS 485 wire connections (Figure 2-3). Multiple DXp transmitters, networked together, are wired in a parallel configuration with a termination jumper installed on the last instrument. A pair of twisted wires (14-20 gauge) is all that is required for interconnection. Communication lines should not be run near ac voltage power lines.
Figure 2-4. Ac Power Connections and Fuse
2-2
2.3.3 Analog Output.
When the analog option is installed (shown, Figure 2-3), a three position terminal connector is provided for 4-20 mA, 0-10 V, and common connections. As with serial communication, the wiring should be routed away from ac power lines and other sources of EMI. The current output is essentially immune to noise and can be transmitted long distances. The voltage output is susceptible to EMI/RFI and should be used only for short distances.
2.3.4 Mains (AC) Power (Figure 2-4).
A screw terminal is provided for permanent transmitter power connection. DXp transmitters
can be switch selected to operate at 115 or 220 VAC (see SW1, Figure 2-4). Before connecting power to the unit, verify that the proper power selection has been made. The two position terminal block is equipped with a clear plastic cover to prevent operator injury. Cable can be either solid or stranded 12 or 14 gage with a ground conductor.
The transmitter is protected with a 114 amp slow blow fuse, located adjacent to the mains terminal block. If the fuse opens, replace it with the same type and current rating.
2.3.5 Auxiliary I/O Port.
The auxiliary I/0 port connection is a factory test port and is not useful to an operator.
2-3
SECTION 3. Configuration
3.1 GENERAL
Set-up and calibration of the DXp-10/15 is accomplished by an operator without programming using the DIP switches, push­buttons, and toggle switches within the unit. Units connected to an LCp-40 or computer ter­minal can be set-up and calibrated remotely via the serial port (see Appendix E).
3.2 SET-UP
3.2.2 Excitation Voltage.
All units are shipped from the factory set for 10 volt excitation.
3.2.3 Serial Format, Address, and Baud Rate.
DIP switch selections for transmitter address, baud rate and serial interface format are presented and defined in SECTION IV, Serial Communication.
3.2.1 Power Selection
All units are shipped from the factory configured for 115 VAC operation. To change the voltage selection to 230 VAC, change SW1 (see Figure 2-4) to the 230V setting. The unit will operate within specification at 50 or 60 Hz.
3.2.4 Input Range Selection.
Units are shipped configured for a 25 mV input range. To increase this range to 35 my, remove the jumper shunt from JP2 on the A-D converter board as shown in Figure 3-1 (see Appendix A photo for A-D board location).
Figure 3-1. 35 mV Input Conversion Instructions
3.3 CALIBRATION
Note: All switches mentioned in paragraphs
3.3.1 – 3.3.4 are depicted in Figure 3-1.
3.3.1 Digital Calibration.
The transmit only serial output can be digitally calibrated using the DIGITAL selector, 1NC/DEC
and ENTER switches mounted on the DXp­10/L5 control panel. For those systems where applying a full capacity dead weight or input signal is not practical, the DXp-10/15 will automatically calculate a linear full span calibration based on a single span point. See Table 3-1 for instructions.
3-1
3.3.2 Analog Calibration.
The analog output is calibrated independently of the digital calibration and can be set using span points anywhere between zero and full scale capacity. The 0-10 V and 4-20 mA outputs CAN NOT be calibrated independently. See Table 3-2 for complete analog calibration instructions.
3.3.3 Monitor Mode Calibration.
The terminal/computer interface option enables the DXp-101L5 to be calibrated remotely from a host computer or terminal. The monitor mode functionally accesses the software routines used to provide remote access via the keypad of an LCp-40 Network Controller. In response, the DXp-10/15 transmits two lines of information similar to the two line display on an LCp-40. In this mode the capacity, graduations, decimal
point, 5 point linearization and other values can be established from the host device. Help messages also can be accessed at any time to aid in the set-up process. The procedure recorded in Table 3-3 shows the command sequence required to perform remote calibration. Consult SECTION IV for serial communication details.
3.3.4 LCp-40 Calibration.
When networked to an LCp-40, 41, or 42 controller, the DXp-10/15 can be remotely calibrated using the controller display and keypad. In this configuration, up to five linearized span points can be entered. Appendix E presents step by step flow diagrams for remote LCp-40 calibration and parameter set­up.
Figure 3-2. Set-up and Calibration Switch Locations (Shade)
Table 3-1. DXp-10 Digital Calibration
Instructions
[1]. Connect Load Cells and Remote Terminal.
Install the load cells and wire them to the DXp-10/15. Connect the RS 485 serial output to a compatible terminal/readout device. Select a compatible serial output format and establish that the communication link is operating (paragraph
3.2.3).
[2]. Establish Zero.
Make sure that all dead weight (vessel, mixer, pipes, etc) is applied as it will be during normal operation. Move the five position slide switch (Figure 3-2) to the digital zero position. Press the enter button to acquire zero and wait approximately 5 seconds for the zero reference value to be stored. The red OPTION STATUS LED will resume flashing when the storage procedure is complete.
3-2
Loading...
+ 29 hidden pages