Pub# OM20-U3-pH
6-21
Procedure # 12 U Series pH troubleshooting
1. Unit doesn’t cycle in automatic mode.
-Does ‘Fix Window’ show GPS, DGPS, or RTK status?
-Solution: connect GPS to system; check baud and herz rate; settings
-Is EC/OM communication light green, showing communication between V.Sense controller and
PC?
-Solution: establish communication between the PC and controller; may require reconnecting and/or
rebooting both devices
-Are EC 0-2’ readings positive?
Solution: move ahead a short distance to allow EC coulters to penetrate
-Are you more than the minimum sample distance from previous sample?
-Solution: move ahead at least the minimum sample distance (user selectable under Controller
Settings)
-Does State window have any text in it? System will not cycle again until it has completed the current
cycle
-Solution: Depending on the state indicated, you may need to manually reset pH system by turning
both controllers off and on again; may need to move sampler shoe to up position using manual
mode
-Does sampler move in manual mode?
If yes but not in automatic mode, check items above: GPS, communication, etc
2. Sampler doesn’t move in manual mode.
Solution: check power to controller and battery connections to pulling vehicle; if ok, additional
troubleshooting with voltmeter will be required.
-Does the sampler only move one direction? Remove hydraulic cover and check LED’s on lower hyd
block DIN connections. Left is Up Sol, Right is down Sol.
3. pH electrodes not cleaning properly;
-check jets manually; are they both spraying, are they aligned?
-if pump not working, prime pump as outlined in pH Operations section; clean filter
-lengthen wash cycle (see Controller setup)
4. pH readings not agreeing with each other
-manually wash electrodes for several seconds; do they agree when clean?
-are both electrodes at same depth? Has one or both moved up into holder?
-perform 4 and 7 pH buffer solution test
5. pH readings taking too long to settle
-check soil/electrode contact; is i-Scan runner penetrating deep enough? Is crop residue entering
slot measurement area?
-Check movement of pH electrode parallel linkage; does it move freely and have full travel?