Veris 196 User Manual

196
Digital Multimeter
Instruction Manual
TABLE OF CONTENTS
page
A. INTRODUCTION
1. Congratulations··············· 4
2. Product Description ············ 4
3. EC Declaration of Conformity ······ 5
B. SAFETY CONSIDERATIONS ········· 6
1. Features and Benefits ··········· 8
2. Product Applications ··········· 9
3. Specifications ················ 10
D. MEASUREMENT TECHNIQUES
1. Controls and Functions ········· 15
2. Power on Options ············· 16
3. Step by Step Procedures: a) Measuring DC Volts ········· 18 b) Measuring AC Volts ········· 20 c) Measuring DC Amps ········· 21 d) Measuring AC Amps ········· 23 e) Measuring Resistance ········ 24 f) Measuring Diodes··········· 25 g) Continuity Buzzer ··········· 26 h) Measuring Frequency ········ 27 i) Record Mode·············· 28 j) Compare Mode ············ 29 k) Relative %Mode ············ 30 l) Two-Hold System ··········· 31 m) Using the Current Output
Functions ················ 32
1) Source Mode ············ 32
2) Simulate Mode··········· 32
3) Producing a Steady mA Output 33
2
A. INTRODUCTION
INTRODUCTION
1. Congratulations!!
Thank you for purchasing a TPI Digital Multimeter. The Triple Display196 is an innovative new concept in DMM design. The unique display enables you to view more than one event at a time. This eliminates the hassle of switching back and forth to review minimum, maximum or preset comparative values. The meter is also easy to use and built to last.
2. Production Description
The DMM 196 is a hand-held autoranging DMM. The backlit LCD can display three readings at one time. In addition to basic function of AC/DC V, AC/DC A, Ohm, Diode test, continuity, Capacitance, Frequency, there is the adaptor function. The DMM 196 also has RS232 output and software for recording information into a PC.
The DMM 196 also features:
REC Records Min/Max and Average readings during
specified measurement intervals.
COMP Compare actual reading to preset HI and LOW
value for Pass/Fail testing of component.
HOLD Tow hold system automatically holds the previous
stable reading when a new one is obtained.
The DMM 196 comes complete with the following accessories:
Battery Rubber Boot Test Lead Set Instruction Manual
4
4) Manually Stepping the mA
Output ················ 34 n) Loop Power Supply Mode ····· 36 o) RS-232C interface ·········· 37
E. ACCESSORIES ················· 40 F. MAINTENANCE ················· 41
G. TROUBLE SHOOTING GUIDE ········ 42
3
B. SAFETY CONSIDERATIONS
WARNING!: Please follow manufacturers test
procedures whenever possible. Do not attempt to measure unknown voltages or components until a complete understanding of the circuit is obtained.
GENERAL GUIDELINES
ALWAYS
• Test the DMM 196 before using to make sure it is operating property.
• Inspect the test leads before using to make sure there are no breaks or shorts.
• Double-check all connections before testing.
• Have someone check on you periodically if working alone.
• Have a complete understanding of the circuit under test.
• Disconnect power to circuit, and then connect test leads to the DMM 196, and then to circuit being measured.
NEVER
• Attempt to measure unknown high voltage.
• Attempt to measure current with the meter in parallel to the circuit.
• Connect to the test leads to a live circuit before setting up the instrument.
• Touch any exposed metal part of the test lead assembly.
6
3. EC Declaration of Conformity
This is to certify that model DMM 196 conforms to the protection requirement of the council directive 89/336/EEC, in the approximation of laws of the member states relating to Electromagnetic compatibility and 73/23/EEC, the low voltage Directive by application of the following standards:
EN 50081-1 1992 Emissions standard EN 50082-1 1992 Immunity standard EN 61010-1 1993 Safety standard EN 61010-2-031 1995 Safety standard
To ensure conformity with these standards, this instruction must be operated in accordance with the instruction and specifications given in this manual.
CAUTION:
Even though this instrument complies with the immunity standards, the accuracy can be affected by strong radio emissions not covered in the above standards. Sources such as hand held radio transceivers, radio and TV transmitters, vehicle radios and cellular phones generate electromagnetic radiation that could be induced into the test leads of the instrument. Care should be taken to avoid such situations or alternatively, check to make sure that the instrument is not being influenced by these emissions.
5
C. TECHNICAL DATA
1. Features and benefits
Safety Meets CE and IEC 1010 requirements. UL
Listed to U.S. and Canadian Safety Standards.
True RMS Needed to accurately measure non-
sinusoidal AC voltage and current waveforms found on many controls and circuits.
Triple Display Shows more than one reading
simultaneously.
Two-Hold Holds two readings on the display at System thesame time.
Auto Power Off
Automatically powers instrument down after 15 minutes of inactivity, yet will continue acquiring data in its various modes.
Record Records Min/Max and Average values
Compare Compares stored value with measured
value for matching components.
Relative Displays measured value as a % of stored Percentage value for checking component tolerances.
RS232 Output Transfers data directly to a PC while
measuring.
Back Light Allows viewing in any light condition.
Auto range Automatically selects the best range for
the measurement.
8
INTERNATION SYMBOLS
CAUTION: RISK OF ELECTRIC SHOCK
AC (ALTERNATING CURRENT)
DC (DIRECT CURRENT)
REFER TO INSTRUCTION MANUAL
GROUND
FUSE
DOUBLE INSULATION
7
3. Specifications
IEC 1010 Over Voltage:
CAT II-1000VDC, 750VAC CATIII-600V Pollution Degree 2
INSTALLATION I·II·III INSTALLATION CATEGORY(OVERVOLTAGE CATEGORY) I :
Signal level, special equipment or parts of equipment, telecommunication, electronic etc., with smaller transient overvoltages than INSTALLATION CATEGORY II.
INSTALLATION CATEGORY(OVERVOLTAGE CATEGORY) II :
Local level, appliances, PORTABLE EQUIPMENT etc., with smaller transient overvoltages than INSTALLATION CATEGORY III.
INSTALLATION CATEGORY(OVERVOLTAGE CATEGORY) III :
Distribution level, fixed installation, with smaller transient overvoltages than INSTALLTION CATEGORY IV.
10
2. Product Applications
Perform the following tests and/or measurements with the DMM196 and the appropriate function:
HVAC/R DC mV • Gas application. ACA • Heat anticipator current in thermostats. ACV • Line voltage. ACV or DCV • Control circuit voltage. DcuA • Flame safeguard control current. OHMS • Heating element resistance (continuity). OHMS • Compressor winding resistance. OHMS • Contactor and relay coil resistance. OHMS • Continuity of wiring. Hz/FREQ • Frequency of line and control voltage. CAP • Motor start and run capacitance. REC • Record min/max voltage of controls and line
voltages.
ELECTRICAL ACV • Measure line voltage. OHMS • Continuity of circuit breakers. Hz/FREQ • Frequency of line and control voltage. DCV • Voltage of direct drive DC motors.
ELECTRONIC ACV • Measure power supply voltage. OHMS • Continuity of circuit breakers. REL% • Match components COMP • Compare readings in circuit or components Hz/FREQ • Frequency of line and control voltage.
9
12
3. Specifications (cont.)
General Specifications
11
3. Specifications (cont.)
f. OHM(Resistance)
Range Resolution Accuracy Overload
protection
500 0.01 +/-0.05% of reading, 600V DC or AC 5K 0.0001K+/- 5digits peak 50K 0.001K 500K 0.01K 5M 0.0001MΩ +/-1.0% of reading, 50M 0.001MΩ +/- 10digits
g. Diode test
Range Resolution Accuracy Overload protection
3V Approx. 1mA 600V DC or AC peak
d. DCA
Range Resolution Accuracy
Overload protection
5mA 0.0001mA +/-0.5% of reading, Fuse(fast blow) 50mA 0.001mA +/- 5digits F600V, 0.5A 31CM 500mA 0.0001A +/-0.75% of reading, Fuse(fast blow) 1A 0.01mA +/- 5digits F600V, 1A 31CM
e. ACA
Range Resolution Accuracy
Overload protection
5mA 0.0001mA +/-0.75% of reading, Fuse(fast blow) 50mA 0.001mA +/- 5digits F600V, 0.5A 31CM 500mA 0.0001A +/-1.5% of reading, Fuse(fast blow) 1A 0.01mA +/- 5digits F600V, 1A 31CM
b. DCmV
Range Resolution Accuracy Impedance
50mV 0.001mV +/-0.1% of reading, +/- 10digits 500mV 0.01mV
c. ACV(45Hz to 450Hz)
Range Resolution Accuracy Impedance
5V 0.0001V +/-0.4% of reading, +/- 40digits 10M 50V 0.001V 500V 0.01V 1000V 0.1V
a. DCV
Range Resolution Accuracy Impedance
5V 0.0001V +/-0.05% of reading, 10M 50V 0.001V +/- 4digits 500V 0.01V 1000V 0.1V +/-0.1% of reading, +/- 10digits