Supplementary documents appropriate to the ordered version come
with the delivery. You can find them listed in chapter "Product
description".
Instructions manuals for accessories and replacement parts
Tip:
To ensure reliable setup and operation of your VEGAVIB 63, we offer
accessories and replacement parts. The corresponding documentations are:
l 31086 - External housing - VEGAVIB
l 30172 - Electronics module VEGAVIB series 60
l 34296 - Protective cover
l 29753 - Lock fitting for VEGAVIB 63 unpressurized
l 29754 - Lock fitting for VEGAVIB 63 16 bar
Editing status: 2012-04-12
29280-EN-120418
VEGAVIB 63 • - two-wire3
1 About this document
1 About this document
1.1 Function
This operating instructions manual provides all the information you
need for mounting, connection and setup as well as important
instructions for maintenance and fault rectification. Please read this
information before putting t he instrument into operation and keep this
manual accessible in the immediate vicinity of the device.
1.2 Target group
This operating instructions manual is directed to trained qualified
personnel. The contents of this manual should be made available to
these personnel and put into practice by them.
1.3 Symbolism used
Information, tip, note
This symbol indicates helpful additional information.
Caution: If this warning is ignored, faults or malfunctions can
result.
Warning: If this warning is ignored, injury to persons and/or serious
damage to the instrument can result.
Danger: If this warning is ignored, serious injury to persons and/or
destruction of the instrument can result.
Ex applications
This symbol indicates special instructions for Ex applications.
lList
The dot set in front indicates a list with no implied sequence.
àAction
This arrow indicates a single action.
1Sequence
Numbers set in front indicate successive steps in a procedure.
4VEGAVIB 63 • - two-wire
29280-EN-120418
2 For your safety
2 For your safety
2.1 Authorised personnel
All operations described in this operating instructions manual must be
carried out only by trained specialist personnel authorised by the plant
operator.
During work on and with the device the required personal protective
equipment must always be worn.
2.2 Appropriate use
The VEGAVIB 63 is a sensor for level detection.
You can find detailed information on the application range in chapter
"Product description".
Operational reliability is ensured only if the instrument is properly used
according to the specifications in the operating instructions manual as
well as possible supplementary instructions.
For safety and warranty reasons, any invasive work on the device
beyond that described in the operating instructions manual may be
carried out only by personnel authorised by the manufacturer. Arbitrary
conversions or modifications are explicitly forbidden.
2.3 Warning about misuse
Inappropriate or incorrect use of the instrument can give rise to
application-specific hazards, e.g. vessel overfill or damage to system
components through incorrect mounting or adjustment.
2.4 General safety instructions
This is a high-tech instrument requiring the strict observance of
standard regulations and guidelines. The user must take note of the
safety instructions in this operating instructions manual, the countryspecific installation standards as well as all prevailing safety
regulations and accident prevention rules.
The instrument must only be operated in a technically flawless and
reliable condition. The operator is responsible for trouble-free
operation of the instrument.
During the entire duration of use, the user is obliged to determine the
compliance of the necessary occupational safety measures with the
current valid rules and regulations and also take note of new
regulations.
29280-EN-120418
VEGAVIB 63 • - two-wire5
2 For your safety
2.5 Safety label on the instrument
The safety approval markings and safety tips on the device must be
observed.
2.6 CE conformity
This device fulfills the legal requirements of the applicable EC
guidelines. By attaching the CE mark, VEGA provides a confirmation
of successful testing. You can find the CE conformity declaration in the
download area of "
www.vega.com".
2.7 SIL conformity
VEGAVIB 63 meets the requirements of functional safety according to
IEC 61508. Further information is available in the Safety Manual
"VEGAVIB series 60".
2.8 Safety instructions for Ex areas
Please note the Ex-specific safety information for installation and
operation in Ex areas. These safety instructions are part of the
operating instructions manual and come with the Ex-approved
instruments.
2.9 Environmental instructions
Protection of the environment is one of our most important duties. That
is why we have introduced an environment management system with
the goal of continuously improving company environmental protection.
The environment management system is certified according to DIN
EN ISO 14001.
Please help us fulfil this obligation by observing the environmental
instructions in this manual:
l Chapter "Packaging, transport and storage"
l Chapter "Disposal"
- Supplementary instructions manual "Plug connector for level
sensors" (optional)
- Ex-specific "Safety instructions" (with Ex versions)
- if necessary, further certificates
The VEGAVIB 63 consists of the components:
l Housing cover
l Housing with electronics
l Process fitting with vibrating rod
Fig. 1: VEGAVIB 63 - with plastic housing
1Housing cover
2Housing with electronics
3Process fitting
Type label
The type label contains the most important data for identification and
use of the instrument:
l Article number
l Serial number
l Technical data
l Article numbers, documentation
l SIL identification (with SIL rating ex works)
29280-EN-120418
VEGAVIB 63 • - two-wire7
3 Product description
Application area
With the serial number, you can access the delivery data of the
instrument via
search". In addition to the type label outside, you can also find the
serial number on the inside of the instrument.
www.vega.com, "VEGA Tools" and "serial number
3.2 Principle of operation
VEGAVIB 63 is a point level sensor with vibrating rod for level
detection.
It is designed for industrial use in all areas of process technology and
is preferably used for bulk solids.
Typical applications are overfill and dry run protection. Thanks to its
simple and robust measuring system, VEGAVIB 63 is virtually
unaffected by the chemical and physical properties of the bulk solid.
It also works when subjected to strong external vibrations or changing
products.
Solid detection in water
If VEGAVIB 63 was ordered for solid detection in water, the vibrating
rod is calibrated to the density of water. If covered by water (density:
1 g/cm³/0.036 lbs/in) VEGAVIB 63 signals "uncovered". Only if the
vibrating element is also covered with solids (e.g. sand, sludge, gravel
etc.) will the sensor signal "covered".
Fault monitoring
The electronics module of VEGAVIB 63 monitors continuously the
following criteria:
l Correct vibrating frequency
l Line break to the piezo drive
If one of these faults is detected, the electronics signals this via a
defined current to the signal conditioning instrument. The connection
cable to the vibrating element is also monitored.
Functional principle
Voltage supply
8VEGAVIB 63 • - two-wire
The vibrating rod is piezoelectrically energised and vibrates at its
mechanical resonance frequency of approx. 360 Hz. When the
vibrating rod is submerged in the product, the vibration amplitude
changes. This change is detected by the integrated electronics module
and converted into a switching command.
Depending on your requirements, VEGAVIB 63 with two-wire
electronics can be connected to different signal conditioning instruments. Compatible signal conditioning instruments are listed in chapter
"Technical data".
The data for power supply are specified in chapter "Technical data".
29280-EN-120418
3 Product description
3.3 Operation
With the factory setting, products with a density of > 0.05 g/cm³
(0.002 lbs/in³) can be measured. It is possible to adapt the instrumentfor products with lower density > 0.02 g/cm³ (0.0007 lbs/in³).
On the electronics module you will find the following indicating and
adjustment elements:
l Signal lamp for indication of the switching condition (green/red)
l Potentiometer for adaptation to the product density
l Mode changeover for selection of the output current
3.4 Storage and transport
Packaging
Transport
Transport inspection
Storage
Storage and transport
temperature
Your instrument was protected by packaging during transport. Its
capacity to handle normal loads during transport is assured by a test
according to DIN EN 24180.
The packaging of standard instruments consists of environmentfriendly, recyclable cardboard. In addition, the sensor is provided with
a protective cover of cardboard. For special versions PE foam or PE
foil is also used. Dispose of the packaging material via specialised
recycling companies.
Transport must be carried out under consideration of the notes on the
transport packaging. Nonobservance of these instructions can cause
damage to the device.
The delivery must be checked for completeness and possible transit
damage immediately at receipt. Ascertained transit damage or
concealed defects must be appropriately dealt with.
Up to the time of installation, the packages must be left closed and
stored according to the orientation and storage markings on the
outside.
Unless otherwise indicated, the packages must be stored only under
the following conditions:
l Not in the open
l Dry and dust free
l Not exposed to corrosive media
l Protected against solar radiation
l Avoiding mechanical shock and vibration
l Storage and transport temperature see chapter "Supplement -
Technical data - Ambient conditions"
l Relative humidity 20 … 85 %
29280-EN-120418
VEGAVIB 63 • - two-wire9
4 Mounting
4 Mounting
4.1 General instructions
Suitability for the process conditions
Switching point
Moisture
Make sure that all parts of the instrument exposed to the process, in
particular the sensor element, process seal and process fitting, are
suitable for the existing process conditions. These include above all
the process pressure, process temperature as well as the chemical
properties of the medium.
You can find the specifications in chapter "Technical data" and on the
type label.
In general, VEGAVIB 63 can be installed in any position. The
instrument only has to be mounted in such a way that the vibrating
element is at the height of the desired switching point.
Use the recommended cables (see chapter "Connecting to power
supply") and tighten the cable gland.
You can give your instrument additional protection against moisture
penetration by leading the connection cable downward in front of the
cable entry. Rain and condensation water can thus drain off. This
applies mainly to outdoor mounting as well as installation in areas
where high humidity is expected (e.g. through cleaning processes) or
on cooled or heated vessels.
Fig. 2: Measures against moisture penetration
Transport
Pressure/Vacuum
10VEGAVIB 63 • - two-wire
Do not hold VEGAVIB 63 on the vibrating element. Especially with
flange and tube versions, the sensor can be damaged by the weight of
the instrument.
Remove the protective cover just before mounting.
The process fitting must be sealed if there is gauge or low pressure in
the vessel. Before use, check if the seal material is resistant against
the measured product and the process temperature.
29280-EN-120418
4 Mounting
The max. permissible pressure is specified in chapter "Technical data"
or on the type label of the sensor.
Handling
Agitators and fluidiza-
tion
Inflowing medium
The vibrating level switch is a measuring instrument and must be
treated accordingly. Bending the vibrating element will destroy the
instrument.
Warning:
The housing must not be used to screw the instrument in! Applying
tightening force can damage internal parts of the housing.
Use the hexagon above the thread for scr ewing in.
4.2 Instructions for installation
Due to the effects of agitators, equipment vibration or similar, the level
switch can be subjected to strong lateral forces. For this reason, do not
use an overly long extension tube for VEGAVIB 63, but check if you
can mount a short level switch on the side of the vessel in horizontal
position.
Extreme vibration caused by the process or the equipment, e.g.
agitators or turbulence in the vessel, e.g. by fluidisation, can cause the
extension tube of VEGAVIB 63 to vibrate in resonance. This leads to
increased stress on the upper weld joint. Should a longer tube version
be necessary, you can provide a suitable support directly above the
vibrating element to secure the extension tube.
This measure applies mainly to applications in Ex areas. Make sure that
the tube is not subject to bending stress due to this measure.
If VEGAVIB 63 is mounted in the filling st ream, unwanted false
measurement signals can be generated. For this reason, mount
VEGAVIB 63 at a position in the vessel where no disturbances, e.g.
from filling openings, agitators, etc., can occur.
This applies particularly to instrument types with long extension tube.
29280-EN-120418
VEGAVIB 63 • - two-wire11
4 Mounting
Fig. 3: Inflowing medium
Lock fitting
Socket
Material cone
VEGAVIB 63 can be mounted with a lock fitting for height adjustment.
Take note of the pressure information of the lock fitting.
The vibrating element should protrude into the vessel to avoid buildup.
For that reason, avoid using mounting bosses for flanges and screwedfittings. This applies particularly to use with adhesive products.
In silos for bulk solids, material cones can form and change the
switching point. Please keep this in mind when installing the sensor in
the vessel. We recommend selecting an installation location where the
vibrating rod detects an average value of the material cone.
The vibrating rod must be mounted in a way that takes the
arrangement of the filling and emptying apertures into account.
To compensate measurement errors caused by the material cone in
cylindrical vessels, the sensor must be mounted at a distance of d/6
from the vessel wall.
29280-EN-120418
12VEGAVIB 63 • - two-wire
dd
d
6
d
6
Fig. 4: Filling and emptying centred
d
d
6
1
2
3
4 Mounting
Fig. 5: Filling in the centre, emptying laterally
1VEGAVIB 63
2Discharge opening
3Filling opening
29280-EN-120418
VEGAVIB 63 • - two-wire13
> 125 mm
(> 5")
> 120 mm
(> 4.7")
4 Mounting
Baffle protection against
falling rocks
In applications such as grit chambers or settling basins for coarse
sediments, the vibrating element must be protected against damage
with a suitable baffle.
This baffle must be manufactured by you.
Fig. 6: Baffle for protection against mechanical damage
14VEGAVIB 63 • - two-wire
29280-EN-120418
5 Connecting to power supply
5 Connecting to power supply
5.1 Preparing the connection
Note safety instructions
Take note of the
safety instructions for Ex
applications
Voltage supply
Connection cable
Connection cab-
le for Ex applications
Always keep in mind the following safety instructions:
l Connect only in the complete absence of line voltage
In hazardous areas you must take note of the respective regulations,
conformity and type approval certificates of the sensors and power
supply units.
Connect the operating voltage according to the following diagrams.
Take note of the general installation regulations. As a rule, connect
VEGAVIB 63 to vessel ground (PA), or in case of plastic vessels, to the
next ground potential. On the side of the instrument housing there is a
ground terminal between the cable entries. This connection serves to
drain off electrostatic charges. In Ex applications, the installation
regulations for hazardous areas must be given priority.
The data for power supply are specified in chapter "Technical data".
The instrument is connected with standard two-wire cable without
screen. If electromagnetic interference is expected which is above the
test values of EN 61326 for industrial areas, screened cable should be
used.
Use cable with round cross-section. A cable outer diameter of 5 … 9 mm
(0.2 … 0.35 in) ensures the seal effect of the cable gland. If you are
using cable with a different diameter or cross-section, exchange the
seal or use a suitable cable gland.
In hazardous areas, only use approved cable connections for VEGAVIB
63.
Take note of the corresponding installation regulations for Ex
applications.
Cover all housing openings conforming to standard according to
EN 60079-1.
5.2 Connection procedure
With Ex instruments, the housing cover may only be opened if there is
no explosive atmosphere present.
Proceed as follows:
1 Unscrew the housing cover
2 Loosen compression nut of the cable entry
3 Remove approx. 10 cm (4 in) of the cable mantle, strip approx.
1 cm (0.4 in) of insulation from the ends of the individual wires
4 Insert the cable into the sensor through the cable entry
29280-EN-120418
VEGAVIB 63 • - two-wire15
5 Connecting to power supply
5 Lift the opening levers of the terminals with a screwdriver (see
following illustration)
Fig. 7: Connection steps 5 and 6
6 Insert the wire ends into the open terminals according to the wiring
plan
7 Press down the opening levers of the terminals, you will hear the
terminal spring closing
8 Check the hold of the wires in the terminals by lightly pulling on
them
9 Tighten the compression nut of the cable entry. The seal ring must
completely encircle the cable
10 If necessary, carry out a fresh adjustment
11 Screw the housing cover back on
The electrical connection is finished.
5.3 Wiring plan, single chamber housing
The following illustrations apply to the non-Ex as well as to the EEx-d
version.
16VEGAVIB 63 • - two-wire
29280-EN-120418
1
4
44
23
1
Housing overview
5 Connecting to power supply
Fig. 8: Material versions, single chamber housing
1Plastic (not with EEx d)
2Aluminium
3Stainless steel (not with EEx d)
4Filter element for pressure compensation or blind stopper with version IP 66/
IP 68, 1 bar (not with EEx d)
Wiring plan
For connection to a VEGATOR (Ex) signal conditioning instrument.
The sensor is powered by the connected VEGATOR signalconditioning instrument. Further information is available in chapter
"Technical data", "Ex-technical data" are available in the supplied
"Safety information manual".
The wiring example is applicable for all suitable signal conditioning
instruments.
If the mode switch of VEGAVIB 63 is correctly set to "max.", the control
lamp on VEGAVIB 63 lights.
l red - with submerged vibrating element
l green - with uncovered vibrating element
Take note of the operating instructions manual of the signal
conditioning instrument. Suitable signal conditioning instruments are
listed in chapter "Technical data".
Fig. 9: Wiring plan
1Voltage supply
29280-EN-120418
VEGAVIB 63 • - two-wire17
1
2
5 Connecting to power supply
Wire assignment connection cable
5.4 Wiring plan - version IP 66/IP 68, 1 bar
Fig. 10: Wire assignment connection cable
1brown (+) and blue (-) to power supply or to the processing system
2Shielding
18VEGAVIB 63 • - two-wire
29280-EN-120418
6 Set up
3
4
1
2
5
6.1 General information
The figures in brackets refer to the following illustrations.
6 Set up
Function/Configuration
Electronics and connec-
tion compartment
On the electronics module you will find the following indicating and
adjustment elements:
l Potentiometer for adaptation to the product density (1)
l DIL switch for mode adjustment - min./max. (2)
l Signal lamp (5)
Note:
As a rule, always set the mode before starting to set up VEGAVIB 63 .
If used on a VEGATOR signal conditioning instrument, always set themode switch (2) on VEGAVIB 63 to max. mode.
The mode is selected on the signal conditioning instrument with the
mode switch.
The switching output will change if you set one of the two mode
switches afterwards. This could possibly trigger other connected
instruments or devices.
6.2 Adjustment elements
Fig. 11: Electronics and connection compartment - two-wire output
With the potentiometer you can adapt the switching point to the solid. It
is already preset and must only be modified in special cases.
As a default setting, the potentiometer of VEGAVIB 63 is set to the
complete right position (> 0.3 g/cm³/0.011 lbs/in³). In very light solids
you have to turn the potentiometer to the complete left position
(0.02 … 0.1 g/cm³ or 0.0007 … 0.0036 lbs/in³). By doing this,
VEGAVIB 63 will be more sensitive and light solids can be detected
more reliably.
For instruments detecting solids in water, these settings are not
applicable. The switching point adaptation is preset and must not be
changed.
With the mode adjustment (min./max.) you can determine the output
current.
When using a VEGATOR signal conditioning instrument, always set
the mode switch to max. mode.
In this case you select the requested mode according to the "Function
chart" (max. - max. detection or overfill protection, min. - min.detection
or dry run detection) on the VEGATOR signal conditioning instrument.
When used on a control system, the following values apply:
l Mode min.
- Vibrating element uncovered - 16 mA ±1 mA
- Vibrating element covered - 8 mA ±1 mA
l Mode max.
- Vibrating element uncovered - 8 mA ±1 mA
- Vibrating element covered - 16 mA ±1 mA
Control lamp for indication of the switching status
l green = 8 mA
l red = 16 mA
l red (flashing) = Failure (< 2.3 mA)
6.3 Function chart
Level switch VEGAVIB 63
The following chart provides an overview of the switching conditions
depending on the adjusted mode and level.
20VEGAVIB 63 • - two-wire
29280-EN-120418
6 Set up
Mode on the
sensor
Max.
Overflow protec-
tion
Max.
Overflow protec-
tion
Min.
Dry run protection
Min.
Dry run protection
Fault message
(min./max. mode)
VEGAVIB 63 level switch with signal conditioning instrument
LevelSignal current -
Sensor
8 mA
approx. 16 mA
approx. 8 mA
approx. 16 mA
any< 2.3 mA
Signal lamp sensor
Green
Red
Green
Red
flashes red
1)
The following chart provides an overview of the switching conditions
depending on the adjusted mode of the signal conditioning instrument
and the level.
Note:
Keep in mind that the mode switch of VEGAVIB 63 must be always set
to "max.".
Mode on the
signal conditioning instrument
Mode A
Overflow pro-
tection
1)
You will find suitable signal conditioning instruments in chapter "Technical
data".
LevelSignal cur-
rent - Sensor
approx. 8 mA
Signal lamp sensor
Green
Signal lamp signal conditioning instrument
29280-EN-120418
VEGAVIB 63 • - two-wire21
6 Set up
General information
Mode on the
signal conditioning instrument
Mode A
Overflow pro-
tection
Mode B
Dry run pro-
tection
Mode B
Dry run pro-
tection
Fault message
(mode A/B)
LevelSignal cur-
rent - Sensor
approx. 16 mA
approx. 16 mA
approx. 8 mA
any< 2.3 mA
Signal lamp sensor
Red
Red
Green
flashes red
Signal lamp signal conditioning instrument
6.4 Recurring test and function test (SIL)
The VEGAVIB 63 is qualified for use in measuring chains of level SIL2
according to IEC 61508 (redundant, level SIL3).
SIL
The measuring system can be used for level detection of bulk solids
and meets the special requirements of safety technology.
This is possible up to SIL2 in a single channel architecture (1oo1D),
and up to SIL3 in a multiple channel, redundant architecture.
The following instrument combinations meet the requirements
according to SIL:
VEGAVIB 63 Ex with
l Oscillator VB60Z
l Signal conditioning instrument VEGATOR 636Ex or SPLC (safety-
oriented PLC)
Implementation - Function test
There are the following possibilities to carry out the recurring function
test:
1 Filling of the vessel up to the switching point
2 Dismounting of the sensor and immersion in the the original
product
22VEGAVIB 63 • - two-wire
29280-EN-120418
6 Set up
3 Short interruption of the supply line to the sensor
4 Pushing the test key on the signal conditioning instrument
1 Filling the vessel up to the switching point
If this does not cause any problems, you can fill the vessel up to the
switching point and monitor the correct sensor reaction.
2 Dismounting of the sensor and immersion in the original
product
You can dismount the sensor for test purposes and check its proper
functioning by immersing it in the original product.
3 Short interruption of the supply line to the sensor
The recurring function test according to IEC 61508 can be carried out
through a short interruption (> 2 seconds) of the supply line to the
sensor. This starts a test sequence.
The correctness of the subsequent switching conditions on the
indications of the SPLC must be monitored. The sensor must neither
be dismounted nor triggered by filling the vessel.
You can carry out the function test with the outputted current values
also directly via a safety PLC or a process control system.
4 Pushing the test key on the signal conditioning instrument
A test key is lowered in the front plate of the signal conditioning
instrument. Push the test key for > 2 seconds with a suitable object.
Hence a test is started. Hence the correctness of the subsequent
switching conditions must be monitored via the two LEDs on the signal
conditioning instrument as well as the connected facilities. The sensor
must neither be dismounted, nor controlled by filling the vessel.
Test without filling or
dismounting the sensor
(3, 4)
This test is valid if you cannot change the vessel filling or cannot
dismount the sensor.
The recurring function test according to IEC 61508 can be carried out
by pushing the test key on a respective signal conditioning instrument
or briefly (> 2 seconds) interrupting the su pply line to the sensor.
The correctness of the subsequent switching conditions must be
monitored via the two LEDs on the signal conditioning instrument as
well as the connected devices. The sensor must neither be
dismounted nor triggered by filling the vessel.
This applies for VEGAVIB 63 with two-wire electronics module VB60Z.
You can carry out the function test with the outputted current valuesalso directly via a safety PLC or a process control system.
A function test can be carried out with measurement setups in
conjunction with the two-wire electronics module VB60Z EX.
29280-EN-120418
VEGAVIB 63 • - two-wire23
16
I/mA
8
2
t/s
1
2
3
0,61,51,5
6 Set up
If you are using a signal conditioning instrument of type VEGATOR for
this purpose, you can carry out the test with the integrated test key.
The test key is recessed in the front plate of the signal conditioning
insturment. Push the test key for > 2 seconds with a suitable object
(screwdriver, pen, etc.).
When the VEGAVIB 63 is connected to a processing system or an
SPLC, you have to interrupt the connection cable to the sensor for > 2seconds. The switching delay must be set to 0.5 s.
After releasing the test key or interrupting the connection cable to the
sensor, the complete measuring system can be checked on correct
function. The following operating conditions are simulated during the
test:
l Fault message
l Empty signal
l Full signal
Fig. 35: Flow chart of the function test
1Full signal
2Empty signal
3Fault message
Check if all three switching conditions occur in the correct sequence
and the stated time period. If this is not the case, there is a fault in the
measuring system (see also the operating instructions manual of the
signal conditioning instrument). Keep in mind that connected instruments are activated during the function test. By doing this, you can
check the correct function of the measuring system.
Note:
Test procedure
Keep in mind that the starting time t
the time up to the first switching (e.g. VEGATOR 636: +1 s)
After releasing the button or after a brief line break.
of the voltage supply can extend
A
The speci fied times apply with a tolerance of ±20 %.
24VEGAVIB 63 • - two-wire
29280-EN-120418
6 Set up
1. Fault sig-
nal
approx. 0.6 s
2)
+ t
A
2. Empty
signal
(approx.
1.5 s)
3. Full signal
(approx.
1.5 s)
4. Return to
current operating condition
Sensor
current Sensor
< 2 mAcurrentless
approx.
8 mA
approx.
16 mA
Level relay
A - overfill
protection
energized
currentlessenergizedenergized
Note:
When used in measuring chains according to IEC 61508, mode B is
not permitted.
Signal
lamp A -Overfill
protection
Level relay
B - dry run
protection
currentlesscurrentless
currentlessenergized
Signal
lamp B -Dry run
protection
Fail safe
relay
Control
lamp
Test assessment (SPLC)
Test passed
l Interference signal (< 3.6 mA) ≥ 400 ms
l Uncovered (approx. 8 mA) ≥ 1 s
l Covered (approx. 16 mA) ≥ 1 s
Test not passed
l Interference signal (< 3.6 mA) < 400 ms / ≥ 750 ms
l Uncovered (approx. 8 mA) < 1 s / ≥ 2 s
l Covered (approx. 16 mA) < 1 s / ≥ 2 s
2)
Starting time of the voltage supply
29280-EN-120418
VEGAVIB 63 • - two-wire25
7 Maintenance and fault rectification
7 Maintenance and fault rectification
7.1 Maintenance
If the instrument is used properly, no special maintenance is required
in normal operation.
7.2 Remove interferences
Reaction when malfunc-
tions occur
Failure reasons
The operator of the system is responsible for taking suitable measures
to rectify faults.
VEGAVIB 63 offers maximum reliability. Nevertheless, faults can occur
during operation. These may be caused by the following, e.g.:
l Sensor
l Process
l Voltage supply
l Signal processing
Fault rectification
The first measure to be taken is to check the output signal. In many
cases, the causes can be determined this way and the faults rectified.
24 hour service hotline
Should these measures not be successful, please call in urgent cases
the VEGA service hotline under the phone no. +49 1805 858550.
The hotline is available to you 7 days a week round-the-clock. Since
we offer this service world-wide, the support is only available in the
English language. The service is free of charge, only the standard
telephone costs will be charged.
Checking the switching
signal
ErrorCauseRectification
VEGAVIB 63 signals
"covered" without being
submerged (overfill protection)
VEGAVIB 63 signals
"uncovered" when being
submerged (dry run protection)
Operating voltage too low Check operating voltage
Electronics defectivePress the mode switch. If the instrument then changes
Unfavourable installation
location
Buildup on the vibrating
element
the mode, the vibrating element may be covered with
buildup or mechanically damaged. Should the switching
function in the correct mode still be faulty, return the
instrument for repair.
Press the mode switch. If the instrument then does not
change the mode, the electronics module may be
defective. Exchange the electronics module.
Mount the instrument at a location in the vessel where
no dead zones or mounds can form.
Check the vibrating element and the sensor if there is
buildup and remove it.
29280-EN-120418
26VEGAVIB 63 • - two-wire
ErrorCauseRectification
2
1
Wrong mode selectedSet the mode switch on VEGAVIB 63 to "max". Set the
Signal lamp flashes redError on the vibrating ele-
ment
Interference on the electronics module
instrument defectiveExchange the instrument or send it in for repair
correct mode on the signal conditioning instrument (A:
overfill protection; B: dry run protection).
Check if the vibrating element is damage or extremely
corroded.
Exchanging the electronics module
7 Maintenance and fault rectification
Reaction after fault rectification
Depending on the reason for the fault and the measures taken, the
steps described in chapter "Set up" may have to be carried out again.
7.3 Exchanging the electronics module
In general, all electronics modules of series VB60 can be inter-
changed. If you want to use an electronics module with a different
signal output, you can download the corresponding operating
instructions manual from our homepage under Downloads.
With EEx d instruments, the housing cover may only be opened if there
is no explosive atmosphere present.
Proceed as follows:
1 Switch off power supply
2 Unscrew the housing cover
3 Lift the opening levers of the terminals with a screwdriver
4 Pull the connection cables out of the terminals
5 Loosen the two screws with a screw driver (Torx size T10 or slot 4)
Fig. 46: Loosening the holding screws
1Electronics module
2Screws (2 pcs.)
29280-EN-120418
VEGAVIB 63 • - two-wire27
7 Maintenance and fault rectification
6 Pull out the old electronics module
7 Compare the new electronics module with the old one. The type
label of the electronics module must correspond to that of the old
electronics module. This applies particularly to instruments used in
hazardous areas.
8 Compare the settings of the two electronics modules. Set the
adjustment elements of the new electronics module to the same
setting of the old one.
Information:
Make sure that the housing is not rotated during the electronics
exchange. Otherwise the plug may be in a different position later.
9 Insert the electronics module carefully. Make sure that the plug is
in the correct position.
10 Screw in and tighten the two holding screws with a screwdriver
(Torx size T10 or Phillips 4)
11 Insert the wire ends into the open terminals according to the wiring
plan
12 Press down the opening levers of the terminals, you will hear the
terminal spring closing
13 Check the hold of the wires in the terminals by lightly pulling on
them
14 Check cable gland on tightness. The seal ring must completely
encircle the cable.
15 Screw the housing cover back on
The electronics exchange is now finished.
7.4 Instrument repair
If a repair is necessary, please proceed as follows:
You can download a return form (23 KB) from our homepage at
vega.com under: "Downloads - Forms and certificates - Repair form".
By doing this you help us carry out the repair quickly and without
having to call back for needed information.
l Print and fill out one form per instrument
l Clean the instrument and pack it damage-proof
l Attach the completed form and, if need be, also a safety data
sheet outside on the packaging
l Please ask the agency serving you for the address of your return
shipment. You can find the respective contact data on our website
www.vega.com under: "Company - VEGA worldwide"
28VEGAVIB 63 • - two-wire
www.
29280-EN-120418
8 Dismounting
8 Dismounting
8.1 Dismounting steps
Warning:
Before dismounting, be aware of dangerous process conditions such
as e.g. pressure in the vessel, high temperatures, corrosive or toxic
products etc.
Take note of chapters "Mounting" and "Connecting to power supply"
and carry out the listed steps in reverse order.
With Ex instruments, the housing cover may only be opened if there is
no explosive atmosphere present.
8.2 Disposal
The instrument consists of materials which can be recycled by
specialised recycling companies. We use recyclable materials and
have designed the electronics to be easily separable.
WEEE directive 2002/96/EG
This instrument is not subject to the WEEE directive 2002/96/EG and
the respective national laws. Pass the instrument directly on to a
specialised recycling company and do not use the municipal collecting
points. These may be used only for privately used products according
to the WEEE directive.
Correct disposal avoids negative effects on humans and the environment and ensures recycling of useful raw materials.
Materials: see chapter "Technical data"
If you have no way to dispose of the old instrument properly, please
Fig. 51: VEGAVIB 63, threaded version G1 A (DIN ISO 228/1)
LSensor length, see chapter "Technical data"
29280-EN-120418
VEGAVIB 63 • - two-wire35
22mm
(
55
/
64
")
33mm
(1
19
/
64
")
L
125mm (4
59
/
64
")
ø16mm (5/8")
G1½A
ø 29mm (1 9/64")
9 Supplement
Fig. 52: VEGAVIB 63, threaded version G1½ A (DIN ISO 228/1)
36VEGAVIB 63 • - two-wire
LSensor length, see chapter "Technical data"
29280-EN-120418
ø 34 mm
(1 11/32")
178 mm (7
1
/
64
")
9 Supplement
Fig. 53: Temperature adapter
29280-EN-120418
VEGAVIB 63 • - two-wire37
9 Supplement
9.3 Industrial property rights
VEGA product lines are global protected by industrial property rights.
Further information see http://www.vega.com.
Only in U.S.A.: Further information see patent label at the sensor
housing.
VEGA Produktfamilien sind weltweit geschützt durch gewerbliche
Schutzrechte.
Nähere Informationen unter http://www.vega.com.
Les lignes de produits VEGA sont globalement protégées par des
droits de propriété intellectuelle.
Pour plus d'informations, on pourra se référer au site http://www.vega.
com.
VEGA lineas de productos están protegidas por los derechos en el
campo de la propiedad industrial.
Para mayor información revise la pagina web http://www.vega.com.
Линии продукции фирмы ВЕГА защищаются по всему миру
правами на интеллектуальную собственность.
Дальнейшую информацию смотрите на сайте http://www.vega.com.