VEGA TOR622 User Manual

Operating Instructions
VEGATOR 620, 621, 622
12 43
5
Level and Pressure
max.
0
5
min.
010
VEGATOR 622
56 78
9 1011121314
!
out
in
Contents
Safety information ........................................................................ 2
Note Ex area ................................................................................ 2
1 Product description
1.1 Function and configuration .................................................. 3
1.2 Approvals ............................................................................. 3
1.3 Types and versions ............................................................. 4
1.4 Technical data and dimensions .......................................... 5
2 Mounting and installation instructions ................................ 9
3 Electrical connection
3.1 Connection instructions ..................................................... 10
3.2 Connection instructions for Ex approved
applications ........................................................................ 10
3.3 Connection diagrams ........................................................ 11
4Setup
4.1 Indicating and adjustment elements ................................ 13
4.2 Setup sequence ................................................................. 14
4.3 Switching point adjustment ............................................... 15
Contents
5 Diagnostics
5.1 Maintenance ....................................................................... 17
5.2 Repair .................................................................................. 17
5.3 Failure rectification ............................................................. 18

Safety information

Please read this manual carefully, and also take note of country-specific installation standards (e.g. the VDE regulations in Germany) as well as all prevailing safety regulations and acci­dent prevention rules. For safety and warranty reasons, any internal work on the instruments, apart from that in­volved in normal installation and electrical con­nection, must be carried out only by qualified VEGA personnel.
2 VEGATOR 620 … 622

Note Ex area

Please note the attached safety instructions containing important information on installation and operation in Ex areas. These safety instructions are part of the operating instructions and come with the Ex approved instruments.
Product description

1 Product description

1.1 Function and configuration

VEGATOR 620 … 622 signal conditioning instruments are module units with plug-in socket, suitable for carrier rail mounting (DIN 46 277). In conjunction with sensors they are used for level detection and switching com­mand output, which is triggered by relay and transistor switches.
Typical applications are double point con­trols, e.g. pump controls (on/off) and monitor­ing functions such as overfill or dry run protection.
The instruments are equipped with adjust­able integration time. The integrated fault monitoring (not with VEGATOR 620) detects short circuit or interruption of the measuring circuit. In case of failure, a failure LED lights up. The level relay and the transistor output deenergise.
To construct a measuring system you need a VEGATOR … level switch and a sensor with analogue measured data transmission, i.e.
- capacitive meas. probes,
- hydrostatic pressure transmitters or
- process pressure transmitters.

1.2 Approvals

If measuring systems are installed according to the following approvals, the appropriate official documents have to be noted and the regulations observed. The documents are supplied with the respective measuring sys­tem.
WHG approval
VEGATOR 621 and 622 signal conditioning instruments with capacitive meas. probes or pressure transmitters as part of an overfill protection according to WHG.
General type approval
no. Z.-65.13-105
Ex approval
When using measuring systems in Ex and StEx areas, the instruments must be suitable and approved for these applications and explosion zones. The suitability is checked by approval authorities and certified in ap­proval documents. VEGATOR 621 Ex and 622 Ex are provided with Ex approval ac­cording to CENELEC (PTB no. Ex-96.D.2068).
12 43
max.
0
min.
010
on
VEGATOR 622
56 78
91011121314
5
10
5
!
p
p
CE conformity
VEGATOR 620, 621 and 622 signal condi­tioning instruments see „1.4 Technical data“.
VEGATOR …
VEGATOR 620 … 622 3
Product description
off
on
off
on

1.3 Types and versions

VEGATOR 620
Single point level switch (fixed switching hysteresis) with selectable mode A/B (A = overfill protection, B = dry run protection) and adjustable integration time Inputs:
- active 1 capacitive probe, 1 hydrostatic pressure transmitter or 1 process pressure transmitter
- passive for connection to an active circuit
Output: 1 relay (spdt) Application: level detection, overfill protection, dry run
protection
When using two VEGATOR 620 connected to one sensor, it is possible to provide a dual single point control. It is possible to connect up to 10 VEGATOR 620 to one sensor, see „3.3 Connection diagrams“.
Level
Time t
VEGATOR 620 single point level switch
Level
on 1
1
2
off 1
on 2
off 2
off 1
on 2
Time t
Double single point control with two VEGATOR 620
VEGATOR 621
Level
Single point level switch (fixed switching hysteresis) with fault monitoring, selectable mode A/B (A = overfill pro­tection, B = dry run protection) and adjustable integration time Input: 1 capacitive probe or
1 hydrostatic pressure transmitter or
Outputs: 1 relay (spdt)
1 process pressure transmitter
1 transistor
Approval: [EEx ia] IIC,
as part of an overfill protection acc. to WHG
Ex area
Non-Ex area
VEGATOR 621 single point level switch
Time t
Application: level detection, overfill protection, dry run
protection
4 VEGATOR 620 … 622
Product description
VEGATOR 622
Double point level switch with fault monitoring, adjustable switching hysteresis, se­lectable mode A/B (A = overfill protection, B = dry run pro­tection) and with adjustable integration time Input: 1 capacitive electrode
1 hydrostatic pressure transmitter 1 process pressure transmitter
Outputs: 1 relay (spdt)
1 transistor
Approvals: [EEx ia] IIC,
as part of an overfill protection acc. to WHG
Application: Min./max. control, overfill protection, dry run
protection

1.4 Technical data and dimensions

General data
Power supply
Operating voltage 20 … 72 V DC
20 … 250 V AC, 50/60 Hz (sine) in case of emergency power supply with curve form extremely deviating from sine: max.
125 V AC (rectangle) Power consumption max. 3 W (3 … 18 VA) Fuse
- supply range T 1 A, 250 V
Level
Max.
Min.
Ex area Non-Ex area
VEGATOR 622 double point level switch
off
on
off
Time t
Measurement data input, general
Number 1 current input Input type active two-wire input analogue Range 4 … 20 mA Sensor capacitive probes
hydrostatic pressure transmitters
process pressure transmitters Sensor supply voltage
- VEGATOR 620 24 V DC
- VEGATOR 621, 622 15 … 18 V DC Switching threshold 4 … 20 mA adjustable Min. hysteresis
- VEGATOR 620, 621 80 µA
- VEGATOR 622 80 … 16000 µA Current limitation at 24 mA, permanently shortcircuit proof Temperature error 0.05 %/10 K of range Connection cable 2-wire (standard cable) Resistance per wire
- VEGATOR 620 max. 250 Ohm
- VEGATOR 621, 622 max. 35 Ohm
VEGATOR 620 … 622 5
Product description
Additional measurement data input on VEGATOR 620
Number 1 additional current input Input type passive two-wire input analogue Range 4 … 20 mA Inner resistance 22 Ohm Application switching in series with active inputs of e.g.
VEGATOR 620, VEGAMET 601 Hysteresis 80 µA Temperature error 0.05 %/10 K of range
Relay output
Number 1 output Contact 1 spdt
(12 - 13 open, 12 - 14 close) Contact material AgCdO and Au plated Switching voltage min. 10 mV DC
max. 250 V AC, 250 V DC Switching current min. 10 µA DC
max. 3 A AC, 1 A DC Breaking capacity max. 750 VA, 54 W
Transistor output (not on VEGATOR 620)
Number 1 output
(synchronously switching with the relay outputs) Switching voltage U Switching current I
CE
C
max. 36 V DC
max. 60 mA DC (short circuit proof, transistor
limits the current in case of short circuit to a
stand-by current of approx. 5 mA) Voltage loss U Blocking current I
CE
O
- 1.5 V at IB = 60 mA
< 10 µA
Electrical connection
Screw terminal max. 1.5 mm
2
Electrical protective measures
Protection
- instrument IP 30
- socket IP 20 Protection class II Overvoltage category II
Mechanical data
Mounting carrier rail mounting acc. to DIN 46 277, p. 3 Dimensions W = 36 mm, H = 118.5 mm, D = 134 mm Weight approx. 170 g
6 VEGATOR 620 … 622
Product description
Ambient conditions
Permissible ambient temperature -20°C … +60°C
VEGATOR 620: With an operating voltage of 60 V DC … 72 V DC the permissible ambient temperature declines linearly from 60°C to
40°C. Storage and transport temperature -40°C … +70°C Option specially protected (tropicalised) electronics
Indicating elements
LED in front plate green on: operating voltage on
yellow: switching point control
red: fault signal
Functions
Mode (selectable) overfill protection (A)
dry run protection (B) Integration time (adjustable) 0.2 … 20 sec.
Adjustment elements
Front plate one or two potentiometers with scale 0 … 10
for switching point adjustment On top cover, to the side DIL for adjustment of:
- mode A/B
(A = overfill protection, B = dry run protection)
- integration time
- switch on/off delay
Electrical separation measures
Reliable separation acc. to VDE 0106, part 1; between power supply, measurement data input,
level relay and transistor output
- reference voltage 250 V
- insulation resistance 2.3 kV
- test voltage 2.3 kV
CE conformity
VEGATOR 620 … 622 signal conditioning instruments meet the protective regulations of EMC (89/336/EWG) and NSR (73/23/EWG). The conformity has been judged acc. to the following standards: EMC Emission EN 50 081 - 1: 1992
Susceptibility EN 50 082 - 2: 1995
NSR EN 61 010 - 1: 1993
VEGATOR 620 … 622 7
Product description
Ex technical data
Measurement data input (intrinsically safe signal circuit)
Classification [EEx ia] IIC Max. values
- voltage UO - 20 V
- current IO - 125 mA
- wattage PO - 624 mW Characteristics linear Effective internal inductance L Effective internal capacitance C
i
i
Max. permissible external inductance LO0.5 mH 1 mH 1.5 mH 2 mH
Max. permissible external capacitance CO97 nF 78 nF 68 nF 200 nF
The intrinsically safe circuits are reliably separated from the non intrinsically safe circuits up to a peak value of 375 V. In case of failure, the max. voltage on the non intrinsically safe circuits must not exceed 250 V.
Carrier rail mounting 35 x 7.5 or 35 x 15 acc. to EN 50 022
Transparent cover
negligible negligible
[EEx ia] IIC [EEx ib] IIC
12 34
118,5
on
54,5
134
56 78
9 1011121314
36
8 VEGATOR 620 … 622
12 4
56 78
9 1011121314
3
A o B o C o 1 o 2 o 3 o
o o o
7 o 8 o 9 o
o o o
N L1
VEGA
0...10V
Mounting and installation instructions

2 Mounting and installation instructions

Mounting
Each series 600 signal conditioning instru­ment consists of a terminal socket for carrier rail mounting DIN 46 277 and a module unit.
The supply voltage can be connected to the terminals 9 and 10 of the terminal socket.
If there are adjoining series 600 instruments, it is possible to continue the connection L1 and N via the supplied jumpers.
Note!
The jumpers must never be used on single instruments or at the respective end of a row of instruments, as there is the danger of coming into contact with the supply voltage, or of causing a short circuit.
The signal conditioning instrument must al­ways be mounted outside hazardous areas or special Ex protective measures must be taken.
Transparent cover
To protect the instrument against unauthor­ised or inadvertent use, the front plate of VEGATOR can be provided after setup with a lockable transparent cover. To remove transparent cover, see drawing below.
Coding
To avoid inadvertent swapping of the differ­ent signal conditioning instruments, the termi­nal socket is provided with pins and the signal conditioning instrument with appropri­ate gaps (mechanical coding). An instrument coding with coded pins in different positions ensures that the various signal conditioning instruments cannot get mixed up.
Ex instruments
An Ex coding with fixed coded pin ensures that non Ex and Ex instru­ments cannot get mixed up.
The coding is part of the explosion protec­tion: On VEGATOR… Ex the supplied coded pins (instrument coded pin and Ex coded pin) must be inserted by the user according to the table below.
Instrument Ex
coding coding
VEGATOR 620 2 ––
VEGATOR 621 2 ––
VEGATOR 622 2 ––
VEGATOR 621 Ex 2 A
VEGATOR 622 Ex 2 A
Separating wall (must be always plugged on Ex instruments)
Ex coding
VEGATOR 620 … 622 9
Instrument coding VEGATOR 620 … 622
Plug

3 Electrical connection

Electrical connection

3.1 Connection instructions

The following connection plans are valid for standard as well as for Ex versions of the signal conditioning instruments. Please ob­serve the following instructions:
- the relay contacts are shown in their con­dition in the absence of current
- if strong electromagnetic interferences are expected, we recommend the use of shielded cable for the signal cables
- the shielding must be grounded only on one sensor or signal conditioning instru­ment side
- if there is the danger of overvoltage, the use of VEGA overvoltage arresters is recommended
- the connection should be made according to country-specific installation standards (e.g. in Germany acc. to the VDE regulations).

3.2 Connection instructions for Ex approved applications

For the following applications, certified instru­ments are required:
- in hazardous areas (if necessary, observe
the special national regulations)
- as part of an overfill protection system
acc. to WHG
- in the shipbuilding industry
- for pressurized vessels.
For these applications, the respective official documents (test reports, test certificates, general type approval and conformity certifi­cates) as well as the applicable mounting and operating instructions should be ob­served. The official documents are supplied with the respective instrument.
In Ex applications, the voltage supply of the sensor must be delivered only through an intrinsically safe circuit. This can be done by means of:
- VEGATOR series 600 signal conditioning
instrument in Ex version
- not certified VEGATOR series 600 signal
conditioning instrument with VEGA safety barrier type 145
Please also note the official documents of these additional instruments.
10 VEGATOR 620 … 622
Electrical connection

3.3 Connection diagrams

Sensor
Current input passive
Sensor Sensor
-+
+
12 4
VEGATOR 620
910111213
+
Ð NL1
L1
-
3
+
12
-
VEGATOR 621
-
+
56
14
910
+
11
12 13
14
Ð N
Transistor outputs in conjunction with a PLC
I
C
+
U
CE
-
U
-
B
E 1.0
+
-
+
12
VEGATOR 622
-
+
56
11
910
+
Ð
L1 N
Transistor output
Relay output
12
14
13
Supply voltage
I
C
+
U
CE
E 1.1
+
PLC
VEGATOR 621 VEGATOR 622
-
VEGATOR 620 … 622 11
Extension example for VEGATOR 620
Instrument 1 = VEGATOR 620 … 622 switched to active mode Instrument 2 through (max.) 10: VEGATOR 620 switched to passive mode
Sensor
Electrical connection
(active)
-
+
-
+
12
VEGATOR 620
-
+
43
-
+
12
VEGATOR 620
-
+
43
(passive) (passive)
-
+
12 43
VEGATOR 620 … 622
Instrument 1 Instrument 2 ..................... through ....... (max.) instrumen t 10
As the above extension example shows, it is possible to control several signal conditioning instruments with one sensor in order to detect different levels.
This wiring layout can be extended to include 10 instruments (loop current).
12 VEGATOR 620 … 622
Setup
;
;
;
;
;
;
;

4 Setup

4.1 Indicating and adjustment elements

12 43
5
1
010
2
3
on
VEGATOR 620
56 78
9 1011121314
View, laterally on top
;;;;;
;;;;;
;;;;;
;;;;;
;;;;;
;;;;;
;;;;;
4
5
1
12 43
5
min.
010
4
5
12 43
max.
min.
8
2
3
6 6 6
7
VEGATOR 621
56 78
9 1011121314
11
12
13
!
on
1 Potentiometer for switching point adjustment 2 Status indication of the outputs 3 LED, supply voltage 4 Separating wall 5 Terminals, inputs 6 Terminals, transistor output 7 Terminals, supply voltage
and relay output 8 LED, fault signal 9 Potentiometer for max. switching point 10 Potentiometer for min. switching point
11 Bar code, serial number 12 DIL switch block 13 Screws 14 Ventilation opening 15 Transparent cover
2
3
7
on
VEGATOR 622
56 78
91011121314
14
5
10
0
5
010
!
4
5
9
10
8
7
VEGATOR 620 … 622 13
15
Potentiometer
The potentiometers (1, 9 and 10), allow continuous adjustment of the switching points.
Signal lamps
Green (3)
- supply voltage available
- ready for operation Yellow (2)
- status indication of the relay and transistor outputs LED lights (relay energised, transistor conductive), LED extinguished (relay deenergised, transistor blocked)
Red (8)
- failure indication
DIL switch block (12)
When VEGATOR is installed, the terminal socket covers the DIL switch block. Adjust­ments are only possible when the instrument is dismounted.
VEGATOR 620 VEGATOR 621, 622
12
t sec
6
off
2 ze za
B
A
t sec
12 6
off
2 ze za A
B
The individual switches are assigned as follows:
- switch A/B: mode switch A - max. detection (overfill protection) B - min. detection (dry run protection)
- switch ze and za ze - switch on delay za - switch off delay
- switch 2, 6, 12 With these you can adjust the delay time. The time intervals of the activated time switches sum up. When the on delay/off delay switches are switched on together, the set time is valid for both delay modes (see the following table).
Setup
Switch on/off Switch delay za ze 2 6 12
0,2 s off off off off off
0,3 s on on off off off
2 s on on on off off
6 s on on off on off
8 s ononononoff
12 s on on off of on
14 s on on on off on
18 s on on of f on on
20 s ononononon

4.2 Setup sequence

The following list explains briefly the essential setup steps
- mount the terminal socket
- wire the terminal socket according to your control requirements
- cover the input terminals with the separat­ing wall (4)
- set the requested mode (A/B) on the DIL switch block (12)
- deactivate the on delay/off delay modes (switches ze, za, 2, 6, 12 to position „off“) on the DIL switch block (12)
- mount the instrument module onto the ter­minal socket
- switch on the power supply, the green LED (3) lights
- adjust the switching points acc. to „4.3 Switching point adjustment“
- if an integration time (on delay/ off delay) should be necessary, the instrument must be removed again to set the time on the DIL switch block (12).
14 VEGATOR 620 … 622
Setup

4.3 Switching point adjustment

Adjustment example 1, mode A (overfill protection)
- VEGATOR 620, 621
- single point control
- sensors: vertically mounted measuring probes, hydrostatic pressure transmitters or process pressure transmitters
Capacitive meas. probes
Hydrostatic pressure transmitters
Level/Pressure
Switching point
Mode A
Mode B
Procedure
• Set the potentiometer (1) to 10.
• Fill the vessel up to the requested level or pressure.
• Turn the potentiometer (1) very slowly anticlockwise until the status indication (2) changes, i.e. the status indication extin­guishes.
• The signal conditioning instrument is now ready for operation. Please close with the transparent cover (15).
Process pressure transmitters
Rel.
t
12 13 14
12 13 14
Transistor
conductive
blocked
conductive
blocked
Adjustment example 2, mode B (dry run protection)
- VEGATOR 620, 621
- single point control
- sensors: horizontally mounted capacitive meas. probes or capacitive meas. probes for adhesive products
Capacitive meas. probes
Procedure
• Set the potentiometer (1) to 0.
• Fill the vessel up to 100 mm below the electrode.
100 mm
• Turn the potentiometer (1) very slowly clockwise until the status indication (2) changes, i.e. lights in mode A, extinguishes in mode B. Note the dial position of the potentiometers (1).
• Continue filling the vessel until the measur­ing probe is completely covered. With measuring probes for adhesive products, the sensitive tip must be covered com­pletely (standard 100 mm above the elec­trode). The status indication (2) changes due to the additional filling.
100 mm
VEGATOR 620 … 622 15
Setup
• Turn the potentiometer (1) very slowly clockwise until the status indication (2) changes, i.e. lights in mode A, extinguishes in mode B. Note the position of the potenti­ometer.
• Determine the average value and adjust it on the potentiometer (1).
• The signal conditioning instrument is now ready for operation. Please close with the transparent cover (15).
The relay and transistor output as well as the status indication behave in mode A and B as previously shown in „Adjustment example 1“, see top left.
Adjustment example 3
- two VEGATOR 620
- as dual single point control (see „3.3 Connection plans, extension example“)
- vertically mounted capacitive measuring probes
Level 2 2nd VEGATOR…
Level 1 1st VEGATOR…
Procedure
• Set the potentiometers (1) of the two signal conditioning instruments to 10.
• Fill the vessel up to level 1.
• Turn the potentiometer (1) of the first VE­GATOR … very slowly anticlockwise until the status indication (2) changes, i.e. lights in mode B, extinguishes in mode A.
• Now fill the vessel up to level 2.
• Now turn the potentiometer (1) of the sec­ond VEGATOR … very slowly anticlockwise until the status indication (2) changes, i.e. lights in mode B, extinguishes in mode A.
• The two signal conditioning instruments are now ready for operation. Please close each with the transparent cover (15).
Adjustment example 4
- VEGATOR 622
- double point control
- sensors: vertically mounted capacitive meas. probes or hydrostatic pressure transmitters
Max. level
Min. level
Level/Pressure
Switching points
2nd VEGATOR Mode A
Mode B
1st VEGATOR
Mode A
Mode B
Level 2
Level 1
t
Rel.
12 13 14
12 13 14
12 13 14
12 13 14
Transistor
conductive
blocked
conductive
blocked
conductive
blocked
conductive
blocked
Level/Pressure
Switching points
Mode A
Mode B
Max. level
Min. level
Rel.
t
12 13 14
12 13 14
Transistor
conductive
blocked
conductive
blocked
16 VEGATOR 620 … 622
Setup / Diagnostics
Procedure
• Set both potentiometers (9 and 10) to 0.
• Fill the vessel to min. level.
• Turn the min. potentiometer (10) very slowly clockwise until the status indication (2) changes, i.e. lights in mode A, extin­guishes in mode B.
• Now turn the potentiometer (9) to 10.
• Continue filling the vessel to max. level, the status indication (2) remains unchanged.
• Turn the max. potentiometer (9) very slowly anticlockwise until the status indication (2) changes, i.e. extinguishes in mode A, lights in mode B.
• The signal conditioning instrument is now ready for operation. Please close with the transparent cover (15).
Note:
VEGATOR 622 can also be used for single point control. The max. potentiometer must be set to 0, the min. potentiometer then acts as single point adjustment.

5 Diagnostics

5.1 Maintenance

The instrument is maintenance free.

5.2 Repair

Repairs require accessing internal parts of the instrument to correct an instrument de­fect. For safety and warranty reasons, repair must only be carried out by qualified VEGA personnel.
In case of a defect, please return respective instrument with a short description of the flaw to our repair department.
Failures are short-term malfunctions caused by wrong adjustment or defects in the sensor or in the connection cables.
Failures, their possible causes and their rectification are listed under „5.3 Failure recti­fication“.
VEGATOR 620 … 622 17
Diagnostics

5.3 Failure rectification

Failu re Rectification / Measure
Instrument is dead/ Green operating lamp extinguishes Check the voltage supply and the mains connections as described
VEGATOR 620 Check the sensor inputs for the following failures: functions incorrectly - short-circuit on the input or not at all - sensor not connected correctly or polarity-reversed
Red failure LED of VEGATOR 621 Check if the sensor is connected correctly. or 622 lights - failures in the sensor, effecting a current change to below 2 mA or
under „Connection instructions“. If the instrument still does not function, please call our service department.
- sensor line interrupted
- supply voltage too low
above 23 mA, generate a fault signal on VEGATOR.
15 … 18 V (24 V with VEGATOR 620)
V
VEGATOR
12 43
Measure the current in the connection cable to the sensor.
4 … 20 mA
+
-
mA
For Ex systems, make sure that the Ex protection is not degraded by the measuring instruments.
a. Current value < 2 mA
- Check the supply voltage on the connection cable to the sensor. The voltage should be at least 17 V. Should a value below 17 V be measured, it indicates a defect in the signal conditioning instrument. In this case, return the instrument for repair to VEGA.
- If at least 17 V are available, separate the signal conditioning instru­ment from the connection cable and connect to the sensor input of the signal conditioning instrument a resistor of 2.2 kOhm. If the failure lamp continues to light, the signal conditioning instrument is defective. In such a case, return the instrument for repair to VEGA.
- Should the failure lamp extinguish, reconnect the signal conditioning instrument. Separate the sensor from the connection cable and con­nect a resistor of 2.2 kOhm in its place.
- Should the failure lamp continue to light, there is probably an inter­ruption of the connection line. Check the connection cable to the sensor.
- If the failure lamp extinguishes, the sensor is defective. Check the connected sensor.
18 VEGATOR 620 … 622
Diagnostics
Failu re Rectification / Measure
VEGATOR 620 functions incorrectly or not at all
Red failure LED of VEGATOR 621 or 622 lights b.Current value > 22 mA
- Check all connections and the connection cable to the sensor.
- Should the red failure lamp continue to light, separate the sensor from the connection cable and connect a resistor of 2.2 kOhm in its place. If the failure lamp extinguishes, the sensor is defective. Check the connected sensor.
- Should the failure lamp continue to light, reconnect the sensor. Sepa­rate the signal conditioning instrument from the connection cable and connect a resistor of 2.2 kOhm to its sensor input.
- Should the failure lamp continue to light, the signal conditioning instru­ment is defective. In such a case, return the instrument for repair to VEGA.
- If the failure lamp extinguishes, there is probably a short circuit in the connection line. Check the connection cable to the sensor.
Note:
A flawlessly operating VEGATOR… has the following switching point values: Potentiometer position Sensor current
0 approx. 4 mA 5 approx. 12.5 mA
10 approx. 21 mA
VEGATOR 620 … 622 19
VEGA Grieshaber KG Am Hohenstein 113 77761 Schiltach/Germany Phone +49 (0) 7836 50-0 Fax +49 (0) 7836 50-201 E-Mail info@de.vega.com
www.vega.com
ISO 9001
The statements on types, application, use and operating con­ditions of the sensors and processing systems correspond to the latest information at the time of printing.
Technical data subject to alterations
2.19 950 / July 2001
Loading...