ntary documents appropriate to the ordered version come
with the delivery. You can find them listed in chapter "Product
description".
Instructions manuals for accessories and replacement parts
Tip:
re reliable setup and operation of your VEGATOR 532 Ex, we
To ensu
offer accessories and replacement parts. The associated documents
are:
l Operating instructions manual "Housing type 505"
l Operating instructions manual "Carrier BGT 596"
oning instrument3
1 About this document
1 About this document
1.1 Function
is operating instructions manual provides all the information you
Th
need for mounting, connection and setup as well as important
instructions for maintenance and fault rectification. Please read this
information before putting the instrument into operation and keep this
manual accessible in the immediate vicinity of the device.
1.2 Target group
This operating instructions manual is directed to trained qualified
personnel. The contents of this manual should be made available to
these personnel and put into practice by them.
1.3 Symbolism used
Information, tip, note
This
symbol indicates helpful additional information.
Caution: If this
result.
Warning: If this warning is ignored, injury to persons and/or serious
damage to the instrument can result.
Danger: If this warning is ignored, serious injury to persons and/or
destruction of the instrument can result.
applications
Ex
Th
symbol indicates special instructions for Ex applications.
is
lList
The dot set in front indicates a list with no implied sequence.
warning is ignored, faults or malfunctions can
àAction
This a
rrow indicates a single action.
1Sequence
Numbers set in front indicate successive steps in a procedure.
23435-EN-090130
4VEGATOR 53
2 Ex • Signal conditioning instrument
2 For your safety
2 For your safety
2.1 Auth
All operations described in this operating instructions manual must be
carried out only by trained specialist personnel authorised by the plant
operator.
During work on and with the device the required personal protective
equipment must always be worn.
orised personnel
2.2 Appropriate use
VEGATOR 532 Ex is a universal signal conditioning instrument for
connection of a level switch.
You can find detailed information on the application range in chapter
"Product description".
Operational reliability is ensured only if the instrument is properly used
according to the specifications in the operating instructions manual as
well as possible supplementary instructions.
For safety and warranty reasons, any invasive work on the device
beyond that described in the operating instructions manual may be
carried out only by personnel authorised by the manufacturer. Arbitrary
conversions or modifications are explicitly forbidden.
2.3 Warning about misuse
Inappropriate or incorrect use of the instrument can give rise to
application-specific hazards, e.g. vessel overfill or damage to system
components through incorrect mounting or adjustment.
23435-EN-090130
VEGATOR 532 Ex • Signal conditi
2.4 General safety instructions
This is a high-tech instrument requiring the strict observance of
standard regulations and guidelines. The user must take note of the
safety instructions in this operating instructions manual, the countryspecific installation standards as well as all prevailing safety
regulations and accident prevention rules.
The instrument must only be operated in a technically flawless and
reliable condition. The operator is responsible for trouble-free
operation of the instrument.
During the entire duration of use, the user is obliged to determine the
compliance of the required occupational safety measures with the
current valid rules and regulations and also take note of new
regulations.
oning instrument5
2 For your safety
2.5 CE conformi
This device fulfills the legal requirements of the applicable EC
guidelines. By attaching the CE mark, VEGA provides a confirmation
of successful testing. You can find the CE conformity declaration in the
download area of
ty
www.vega.com.
2.6 Safety instructions for Ex areas
Please note the Ex-specific safety information for installation and
operation in Ex areas. These safety instructions are part of the
operating instructions manual and come with the Ex-approved
instruments.
2.7 Environmental instructions
Protection of the environment is one of our most important duties. That
is why we have introduced an environment management system with
the goal of continuously improving company environmental protection.
The environment management system is certified according to DIN
EN ISO 14001.
Please help us fulfil this obligation by observing the environmental
instructions in this manual:
l Chapter "Packaging, transport and storage"
l Chapter "Disposal"
6VEGATOR 53
23435-EN-090130
2 Ex • Signal conditioning instrument
3 Product description
on
VEGA
!
1
2
3 Product description
ion
Scope of delivery
3.1 Configurat
The scope of delivery encompasses:
l VEGATOR 532 Ex signal conditioning instrument
l Transparent cover (lead-sealable)
l Documentation
The type label contains the most important data for identification and
use of the instrument:
l Article number
l Serial number
l Technical data
l Article numbers documentation
The serial number allows you to access the delivery data of the
instrument via
search".
www.vega.com, "VEGA T
ools" and "serial number
Application area
23435-EN-090130
VEGATOR 532 Ex • Signal conditi
3.2 Principle of operation
VEGATOR 532 Ex is a single signal conditioning instrument with two
channels for processing the signals of conductive probes.
oning instrument7
3 Product description
Functional principle
Power supply
Packaging
Transport
The VEGATOR 53
2 Ex signal conditioning instrument powers
connected instruments and simultaneously processes their measuring
signals.
When the medium reaches the switching point of the sensor, the
voltage between sensor and VEGATOR 532 Ex drops. The output
relays then switch according to the set mode.
Detailed information about the power supply can be found in chapter
"Technical data".
3.3 Operation
The mode (A/B) can be preset via the DIL sw itch on the signal
conditioning instrument.
A potentiometer for switch point adjustment is located on the front
plate of VEGATOR 532 Ex.
3.4 Packaging, transport and storage
Your instrument was protected by packaging during transport. Its
capacity to handle normal loads during transport is assured by a test
according to DIN EN 24180.
The packaging of standard instruments consists of environmentfriendly, recyclable cardboard. For special versions, PE f oam or PE foil
is also used. Dispose of the packaging material via specialised
recycling companies.
Transport must be carried out under consideration of the notes on the
transport packaging. Nonobservance of these instructions can cause
damage to the device.
Transport inspection
The delivery must be checked for completeness and possible transit
damage immediately at receipt. Ascertained transit damage or
concealed defects must be appropriately dealt with.
Storage
Up to the time of installation, the packages must be left closed and
stored according to the orientation and storage markings on the
outside.
Unless otherwise indicated, the packages must be stored only under
the following conditions:
l Not in the open
l Dry and dust free
l Not exposed to corrosive media
l Protected against solar radiation
l Avoiding mechanical shock and vibration
Storage and transport
temperature
l Storage and transport temperature see chapter "Supplement -
Technical data - Ambient conditions"
8VEGATOR 53
23435-EN-090130
2 Ex • Signal conditioning instrument
l Relative humidity 20 … 85 %
3 Product description
23435-EN-090130
VEGATOR 532 Ex • Signal conditi
oning instrument9
4 Mounting
4 Mounting
Installation location
Transparent cover
Mounting in single
housing type 505 Ex
4.1 General ins
VEGATOR 532 Ex can be optionally mounted into a single housing
type 505 Ex or with a module in a carrier BGT 596 or BGT 596 Ex.
Module
Module consisting of:
Multipoint connector DIN 41612, series F, 33-pole with coded pin andmounting material for mounting in carrier BGT 596.
Ex module
Ex-module consisting of:
Multipoint connector DIN 41612, series F, 33-pole with coded pins, Ex-
separating chamber and mounting material for mounting into carrier
BGT 596 Ex.
Single housing
Plastic housing type 505 Ex for single mounting of signal conditioning
instruments with an instrument width of 5 TE (25.4 mm/1 in).
To protect VEGATOR 532 Ex against unauthorised and inadvertent
adjustment, the front plate can be covered with a lead-sealable
transparent cover.
tructions
4.2 Mounting instructions
The socket of the single housing type 505 Ex can either be screwed
directly to the mounting plate or can be plugged onto a carrier rail
TS35 x 7.5 according to EN 50022 or TS32 according to EN 50035.
You can find further information on mounting in the operatinginstructions manual of the single housing type 505 Ex.
A VEGATOR 532 Ex in E
instrument and must not be installed in explosion-endangered areas.
Mounting in carrier rail
BGT 596 (Ex)
10VEGATOR 53
Mount the respective module (standard or Ex version) on your carrier
BGT 596 or BGT 596 Ex.
The female multipoint connector is available in the following
connection versions:
l Wire-Wrap standard connection 1 x 1 mm
l Plug connection 2,8 x 0.8 mm
l Termi-Point standard connection 1,6 x 0.8 mm
l Soldering connection
l Screw terminals 2 x 0.5 mm²
You can find further information on mounting in the operating
instructions manual of the carrier.
x version is an auxiliary, intrinsically safe
23435-EN-090130
2 Ex • Signal conditioning instrument
on
1
4 Mounting
Ex separating chamber
When you are mounting VEGATO
R 532 Ex with Ex approval in a carrier,
you have to use a VEGA Ex module.
In Ex applications, a protection of IP 20 must be maintained. Cover the
gaps or free modules from the front with appropriate blind covers.
Keep a distance of at least 2 TE (10 mm/0.4 in) from the module cards
of other manufacturers.
If you want to mount VEGATOR 532 Ex in the complete left position in
the carrier, you have to mount a blind cover with at least 4 TE (20 mm/
0.8 in) in front of the module of VEGATOR 532 Ex.
Fig. 2: Distance to the
1Blind cover
carrier side
The Ex separating chamber must be plugged in before starting with the
setup of the Ex versions of VEGATOR 532 Ex.
The Ex separating chamber is part of module 33/Ex-33.
Safe operation can be only ensured if the operating instructionsmanual and the EU type approval certificate are observed. VEGATOR532 Ex must not be opened.
You can find further information on mounting the Ex separating
chamber in the operating instructions manual of the module (module
33/Ex-33).
Instrument coding
23435-EN-090130
VEGATOR 532 Ex • Signal conditi
All signal conditioning instruments are provided with different gaps
dependent on type and version (mechanical coding).
The module is provided with coded pins that can be inserted to prevent
accidental interchanging of the various instrument types.
The coded pins are part of module 33/Ex-33.
With a VEGATOR 532 Ex in Ex version, the supplied coded pins (type
coded pins and Ex coded pin) must be inserted by the user according
to the below chart.
oning instrument11
1
bzd
ca
3
5
7
9
11
13
15
17
19
21
23
25
27
29
31
1
3
2
AB
4 Mounting
You can find furth
er information on mounting the coded pins in the
operating instructions manual of the module (module 33/Ex-33).
Fig. 3: Instrument coding - Multipoint connector VEGATOR 532 E
1Coding - Signal conditioning instrument family VEGATOR (c7)
2Ex coding with Ex version (c23)
3Coding - Signal conditioning instrument type VEGATOR 532 Ex (a21)
A Instrument coding
B Function coding
x
12VEGATOR 53
23435-EN-090130
2 Ex • Signal conditioning instrument
5 Connecting to power supply
5 Connecting to power supply
Note safety instructions
Take note of
safety instructions for Ex applications
Select power supply
Selecting connection
cable
Cable screening and
grounding
5.1 Preparin
Always keep in mind the following safety instructions:
l Connect only in the complete absence of line voltage
l If overvoltage surges are expected, overvoltage arresters should
be installed
Tip:
We recomm
VEGATOR 532 Ex) and B62-36G (sensor supply).
In hazardous areas you should take note of the appropriate
regulations, con
and power supply units.
The power supply can be 20 … 53 V AC, 50/60 Hz or 20 … 72 V DC.
Detailed information for power supply is specified in chapter
"Technical data".
Power supply of VEGATOR 532 Ex is connected with standard cable
according to the national installation standards.
The sensors are connected with standard two-wire cable without
screen. If electromagnetic interference is expected which is above the
test values of EN 61326 for industrial areas, screened cable should be
used.
From a cable length of 50 m use one cable for each signal conditioning
instrument.
If you are using a common cable, the min. and max. cables must be
screened. The screen must be connected the ground on both sides.
g the connection
end VEGA overvoltage arresters B61-300 (power supply
formity and type approval certificates of the sensors
Select connection cable for Ex
applications
23435-EN-090130
VEGATOR 532 Ex • Signal conditi
Take note of the corresponding installation regulations for Ex
Control lamps (LED) in the front plate indicate operation, switching
status and fault signal.
l Green
- Operating control lamp
- Mains voltage on, instrument is
l Red
- Failure lamp
- Fault on the sensor circuit due to sensor failure or line break
- Th
e relay deenergises in case of failure
l Yellow
- Relay
- The yellow relay control lamp reacts depending on the set
- Generally the relay control lamp shows the activated
- A dark
oning instrument19
control lamp
mode (A/B)
(energized) condition of the relay
relay control lamp means that the relay is deenergised
(transistor
blocks)
operating
1
2
3
6 Set up
Potentiometer for
switching point adjustment
DIL switch - Mode
Mode A
Two potentiometers for switching point adaptation are located on the
front plate
of the signal conditioning instrument. With this potenti-
ometer you can adapt the measuring system to the conductivity of the
medium separately for each channel.
Use a small screwdriver to carry out adjustments on the potentiometer.
One slide switch per channel is located on the circuit board of the
signal conditioning instrument. Set the requested mode before
inserting VEGATOR 532 Ex because the switch will no longer be
accessible in assembled condition.
l A - Max. detection or overflow protection
l B - Min. detection or dry run detection
Selection of the mode
You can set mode A or B by means of the selection switch.
Preferrably as overflow protection, compulsory as overfill protection.
Fig. 11: Mode A - Overfill protection
with covered max. electrode:
Means
l Relay (channel 1) denergizes, connection d10/b10 connected
through relay
l Transistor output (channel 1) blocks
l Control lamp output (channel 1) extinguishes
Means with uncovered max. electrode (level detection) or min.
electrode (two-point control):
l Relay (channel 1) energizes, connection d10/z10 connected
through relay
l Transistor output (channel 1) conducts
l Control lamp output (channel 1) lights
20VEGATOR 53
23435-EN-090130
2 Ex • Signal conditioning instrument
1
2
3
6 Set up
Mode B
Fault signal adjustment,
bridge
23435-EN-090130
VEGATOR 532 Ex • Signal conditi
Preferrably as dry run protection system.
Fig. 12: Mode B - Dry run protection
Mean
s with covered max. electrode:
l Relay (channel 1) energizes, connection d10/z10 connected
through relay
l Transistor output (channel 1) conducts
l Control lamp output (channel 1) lights
Means with uncovered max. electrode (single point control) or min.
electrode (double point control):
l Relay (channel 1) denergizes, connection d10/b10 connected
through relay
l Transistor output (channel 1) blocks
l Control lamp output (channel 1) extinguishes
To monitor the electrodes and their circuits, a resistor of 220 kΩ must
be mounted between connection 1 and connection 2 in the connection
housing of the probe, i.e. with single point control, the measuring and
earth electrodes are monitored, and with double point control, the max.
and earth electrodes are monitored.
If you want to use both channels for two separate measurements (no
two-point control), the measuring electrodes are monitored, e.g.
ground and max. electrode of channel 1 (connection d30 and d32) and
ground and min. electrode of channel 2 (connection z30 and z32).
By doing this, a failure is triggered separately for each channel. In case
of failure:
l the channel-specific control lamp fault signal lights
l the relay output of the concerned channel deenergises
l the transistor output of the concerned channel is blocked
l The functions of the inteferred channel remain
For probes with 220 kΩ resistor, a soldering bridge must be closed on
the circuit board of VEGATOR 532 Ex.
l Channel 1 - Soldering bridge BR1
oning instrument21
BR2
BR1
1
2
6 Set up
l Channel 2 - Soldering
Fig. 13: Fault message adjustment - Soldering
1Soldering bridge for fault message adjustment of channel 1 (BR1)
2Soldering bridge for fault message adjustment of channel 2 (BR2)
bridge BR2
bridges
Note:
The line
If a measuring system is used as part of an overfill prote
monitoring and the fault signal are inactive with this bridge.
ction system,
the bridge on the signal conditioning instrument must not be closed.
With Ex probes, the resistor in the probe housing is already present.
6.3 Switching point adjustment
Level detection for max.
signal
The setup refers to channel 1. The procedure is the same for channel
2.
Prerequirements
l Set the selection switch (channel 1) to mode A
l Connect the signal conditioning instrument to power supply
l Turn the potentiometer (channel 1) to position 0
Switching point max.
signal
l Fill the vessel until the max. electrode is covered approx. 1 cm by
the product
l Turn the potentiometer slowly clockwise until the yellow LED
extinguishes
The max. electrode is now adapted to the conductivity of the medium,
i.e. the relay (channel 1) deenergises at max. level and the transistor
blocks (channel 1).
Level detection for min.
signal
22VEGATOR 53
Prerequirements
l Set the selection switch to mode B
l Connect the signal conditioning instrument to power supply
l Turn the potentiometer to position 10
23435-EN-090130
2 Ex • Signal conditioning instrument
6 Set up
Switching point min.
signal
Two-point control (min./
max.), mode A (overfill
protection)
Adjust the switching
point
Two-point control (min./
max.), mode B (dry run
protection)
Adjust the switching
point
l Empty the vessel until the min. electrode
is covered approx. 1 cm
by the product
l Turn the potentiometer slowly anticlockwise until the yellow signal
lamp extinguishes
The min. electrode is now adapted to the conductivity of the medium, i.
e. the relay deenergises at min. level and the transistor blocks.
Prerequirements
l First of all, connect only ground and max. electrode to the signal
conditioning instrument (terminals d28 and d30)
l Set the selection switch to mode A
l Connect the signal conditioning instrument to power supply
l Turn the potentiometer (channel 1) to position 0
l Fill the vessel until the max. electrode is covered approx. 1 cm by
the product
l Turn the potentiometer slowly clockwise until the yellow signal
lamp extinguishes
l Connect the min. electrode to terminal d32 of the signal
conditioning instrument
The two-point control is now adapted to the conductivity of the
medium, i.e. the relay deenergises at max. level and the transistor
blocks.
This switching condition remains until the level falls below the position
of the min. electrode, i.e. the relay energizes at min. level and the
transistor conducts.
The relay deenergises again at max. level and the transistor blocks.
Prerequirements
l First of all, connect only ground and max. electrode to the signal
conditioning instrument (terminals d28 and d30)
l Set the selection switch to mode B
l Connect the signal conditioning instrument to power supply
l Turn the potentiometer to position 10
l Fill the vessel until the max. electrode is covered approx. 1 cm by
the product
l Turn the potentiometer slowly clockwise until the yellow signal
lamp extinguishes
l Connect the min. electrode to terminal d32 of the signal
conditioning instrument
The two-point control is now adapted to the conductivity of the
medium, i.e. the relay deenergises at max. level and the transistor
blocks.
This switching condition remains until the level falls below the position
of the min. electrode, i.e. the relay energizes at min. level and the
transistor conducts.
The relay deenergises again at max. level and the transistor blocks.
23435-EN-090130
VEGATOR 532 Ex • Signal conditi
oning instrument23
6 Set up
Note:
Keep in mind
moment when the switching point is reached and the moment when
the switching function is triggered, there is a damping i.e. delay of
0.5 sec. The potentionmeter must therefore be turned slowly.
In all applications in conjunction with an integrated resistor of 220 kΩ,
line monitoring of the ground and max. electrode is ensured.
for all switching point adjustments that, between the
24VEGATOR 53
23435-EN-090130
2 Ex • Signal conditioning instrument
7 Maintenance and fault rectificati
7 Maintenance and fault rectification
on
Causes of malfunction
Fault rectification
24 hour service hotline
Failure
7.1 Maintena
When used in the correct way, no special maintenance is required in
normal operation.
nce
7.2 Remove interferences
A maximum of reliability is ensured. Nevertheless, faults can occur
during operation. These may be caused by the following, e.g.:
l Measured value from sensor not correct
l Power supply
l Interference on the cables
The first measure to be taken is to check the input/output signal. In
many cases, the causes can be determined this way and the faults
rectified.
However, should these measures not be successful, call the VEGA
service hotline in urgent cases under the phone no. +49 1805 858550.
The hotline is available to you 7 days a week round-the-clock. Since
we offer this service world-wide, the support is only available in the
English language. The service is free of charge, only the standard
telephone costs will be charged.
? The red failure control lamp (LED) of the signal conditioning
instrument lights
l Sensor not connected correctly
à Check the electrical connection by means of the wiring plans
l Line break
à Check the electrical connection cables from the probe to the
signal conditioning instrument
l Resistance for line monitoring missing
à Check if a resistor of 220 kΩ is mounted in the sensor housing
between terminals 1 and 2 or if the respective soldering bridge
to the fault message adjustment on the signal conditioning
instrument is closed.
à You can find further information in chapter "Set up - Failure
adjustment, bridge".
In Ex systems, make
measuring instruments used.
sure that the Ex protection is not influenced by the
23435-EN-090130
VEGATOR 532 Ex • Signal conditi
oning instrument25
7 Maintenance and fault rectificati
on
? The signal conditioning instrument does not switch when the
respective
l Conductivity of the product too low
à Check if the conductivity value of your product is at least
probe is covered or uncovered
7.5 µS/cm
7.3 Instrument repair
If a repair is necessary, please proceed as follows:
You can download a return form (23 KB) from our Internet homepage
www.vega.com under: "Dow
form".
By doing this you help us carry out the repair quickly and without
having to call back for needed information.
l Print and fill out one form per instrument
l Clean the instrument and pack it damage-proof
l Attach the completed form and, if need be, also a safety data
sheet outside on the packaging
l Please ask the agency serving you for the address of your return
shipment. You can find the respective agency on our website
www.vega.com under: "Compa
nloads - Forms and certificates - Repair
ny - VEGA worldwide"
26VEGATOR 53
23435-EN-090130
2 Ex • Signal conditioning instrument
8 Dismounting
8 Dismounting
8.1 Dism
Take note of chapters "Mounting" and "Connecting to power supply"
and carry out the listed steps in reverse order.
ounting steps
8.2 Disposal
The instrument consists of materials which can be recycled by
specialised recycling companies. We use recyclable materials and
have designed the electronics to be easily separable.
WEEE directive 2002/96/EG
This instrument is not subject to the WEEE directive 2002/96/EG and
the respective national laws. Pass the instrument directly on to a
specialised recycling company and do not use the municipal collecting
points. These may be used only for privately used products according
to the WEEE directive.
Correct disposal avoids negative effects to persons and environment
and ensures recycling of useful raw materials.
Materials: see chapter "Technical data"
If you have no possibility to dispose of the old instrument
professionally, please contact us concerning return and disposal.
23435-EN-090130
VEGATOR 532 Ex • Signal conditi
oning instrument27
9 Supplement
9 Supplement
9.1 Technical
General data
Series19" module card, multipoint connector according to
Weight150 g (5.3 oz)
Power supply
Operating voltage20 … 53 V AC, 50/60 Hz, 20 … 72 V DC
Max. power consumption2 W, 3 VA
Fuse
- Supply areaT 1 A, 253 V
- Switching off possibilitymin. 35 A at 253 V AC or 125 V DC
Sensor input (channel)
Quantity2 x level detection or 2 x two-point control (min./
Response resistor1 … 200 kΩ adjustable
Parallel resistor for fault monitoring220 kΩ
Measuring circuitmax. 5 V eff., max. 1 mA
Permissible line capacitance1 x 100 nF or 2 x 70 nF with min./max. control
Switching hysteresis15 %
Relay outputs
Quantity2
ModeA/B switch (A - max. detection or overfill protection,
Damping500 ms
Contact1 spdt for each output
Contact materialAgNi, hard gold-plated
Turn-on voltage≥ 10 mV DC, ≤ 253 V AC/DC
Switching current≥ 10 µA DC, ≤ 3 A AC, 1 A DC
Breaking capacity≤ 750 VA, ≤ 54 W DC
data
DIN 41612, including transparent cover (lockable)
max.)
B - min. detection or dry run protection) adjustable
separately for each channel
Transistor outputs
Number, function1 output, synchronously switching with the relay
Galvanic separationFloating
28VEGATOR 53
2 Ex • Signal conditioning instrument
23435-EN-090130
9 Supplement
Maximum values
- U
- I
B
B
≤ 36 V DC
≤ 60 mA
Transistor voltage loss (UCE)approx. 1.5 V at IB60 mA
Inverse current (I0)< 10 µA
Adjustment elements
DIL switchfor preadjustment of the mode
Potentiometerfor switching point adjustment
Control lamps in the front plate
- Status indication operating voltageSignal lamp green (LED)
- Status indication fault signalSignal lamp red (LED)
- Status indication switching point controlSignal lamp yellow (LED)
Ambient conditions
Ambient temperature-20 … +60 °C (-4 … +140 °F)
Storage and transport temperature-40 … +70 °C (-40 … +158 °F)
Electromechanical data
Electrical connection
- Carrier BGT 596 Ex33-pole multipoint connector, series F (d, b, z) with
coding holes
- Housing type 505 ExScrew terminal for wire cross-section up to 1.5 mm²
(AWG 16)
Electrical protective measures
Protection
- Signal conditioning instrument - not
IP 00
mounted
- mounted into BGT 596 Ex - front side
IP 30
(completely equipped)
- mounted into BGT 596 Ex - upper and
IP 20
lower side
- mounted into BGT 596 Ex - wiring sideIP 00
- mounted into housing type 505 ExIP 30
Overvoltage categoryII
Protection classII
Approvals
1)
ATEXATEX II (1) GD [EEx ia] IIC
1)
Deviating data in Ex applications: see separate safety instructions.