VEGA TOR521, TOR522, TOR523, TOR527 User Manual

Operating Instructions
VEGATOR 521, 522, 523, 527
Level and Pressure
5
010
on
521
5
010
max
5
010
on
522
5
010
1
5
010
2
on
523
5
010
1
5
010
2
on
527
p
Contents
Safety information ........................................................................ 2
Note Ex area ................................................................................ 2
1 Product description
1.1 Function and configuration .................................................. 3
1.2 Types and versions ............................................................. 3
1.3 Approvals ............................................................................. 3
1.5 Dimensions ........................................................................... 6
2 Mounting
2.1 Mounting instructions ........................................................... 7
3 Electrical connection
3.1 Connection instructions ....................................................... 9
3.2 Connection schematic ......................................................... 9
4Setup
4.1 Indicating and adjustment elements ................................ 11
4.2 Principle of operation ......................................................... 13
4.3 Switching point adjustment ............................................... 14
Contents
5 Diagnostics
5.1 Maintenance ....................................................................... 21
5.2 Repair.................................................................................. 21
5.3 Failure rectification ............................................................. 22

Safety information

Please read this manual carefully, and also take note of country-specific installation standards (e.g. the VDE regulations in Germany) as well as all prevailing safety regulations and accident prevention rules.
For safety and warranty reasons, any internal work on the instruments, apart from that involved in normal installation and electrical connection, must be carried out only by qualified VEGA personnel.
2 VEGATOR

Note Ex area

Please note the approval documents (yellow binder), and especially the included safety data sheet.
Product description

1 Product description

1.1 Function and configuration

VEGATOR 521, 522, 523 and 527 level switches are signal conditioning instruments in European size (DIN 41 494) for level detection. Typical applications are overfill protection in vessels and dry run protection or double point control of pumps. Depending on the instrument type, up to 2 sensors with analogue transmission of measured data can be connected to the input. 2 relays (spdt) and 2 floating transistors are available as outputs. All instruments are equipped with a fail-safe relay and transistor. The instruments are provided with fault monitoring and adjustable integration time.
All VEGATORs are available with Ex approvals or as overfill protection systems acc. to WHG and VbF.
Function
VEGATOR signal conditioning instruments enable detection of min./max. levels. The measured data of the sensor are converted into a switching command, and connected instruments can be switched on or switched off. The integrated fault monitoring detects short-circuit or break of the measuring cable as well as selected failures from the transmitter. The failure lamp lights and the output relays and output transistors are deenergised in case of failure.

1.2 T ypes and versions

VEGATOR 521 (Ex)
Signal conditioning instrument as level switch for single point control. 1 sensor input for capacitive electrodes or pressure transmitters. 1 relay output (spdt) and 1 transistor output with fixed switching hysteresis.
VEGATOR 522 (Ex)
Signal conditioning instrument as level switch for one sensor for double point control. 1 sensor input for capacitive electrodes or pressure transmitters. 1 relay output (spdt) and 1 transistor output with adjustable switching hysteresis.
VEGATOR 523 (Ex)
Signal conditioning instrument as level switch for one sensor for dual single point control. 1 sensor input for capacitive electrodes or pressure transmitters. 2 relay outputs (spdt) and 2 transistor outputs with fixed switching hysteresis.
VEGATOR 527 (Ex)
Signal conditioning instrument as level switch for two sensors for single point control. 2 sensor inputs for capacitive electrodes or pressure transmitters. 2 relay outputs (spdt) and 2 transistor outputs with fixed switching hysteresis.
In case of mains failure, the relays deenergise.
Configuration
For configuration of a measuring system, a VEGATOR and one or two sensors with analogue measured data transmission, e.g. a capacitive electrode or a pressure transmitter are required.
VEGATOR 3

1.3 Approvals

VEGATOR is available with the following approvals:
- explosion protection: respective instrument with intrinsically safe circuit(s) [EEx ia] IIC
- overfill protection acc. to WHG applied for
For these applications, please note the rele­vant offical documents (test report, test certificates, type approvals and conformity certificates). These are supplied with the respective instrument.
Product description
1.4 Technical data
General
Version European size DIN 41 612 Dimensions W = 25.4 mm (5 TE); H = 128.4 mm; D = 162 mm Weight approx. 180 g Ambient temperature -20°C … +60°C Storage and transport temperature -40°C … +70°C
Power supply
Operating voltage 20 … 53 V AC, 50/60 Hz or 20 … 72 V DC Power consumption max. 4 VA
Electrical protective measures
Protection class II Overvoltage class II Protection
- mounted in carrier BGT 596 with module IP 00
- front side completely equipped IP 30
- upper and lower side IP 20 (only Ex carrier)
- wiring side (rear) IP 00
- single housing IP 20
Inputs
Transmission analogue Switching threshold adjustable 4 … 20 mA Hysteresis (fixed / min.) 80 µA Current limitation 24 mA (permanently short-circuit proof) Connection cable to the sensor 2-wire Sensor supply voltage approx. 15 … 18 V DC Resistance per wire max. 35 Temperature error 0.05 %/10 K (relating to range) Integration time 0.2 … 20 s
Relay output
Number of
- VEGATOR 521, 522 1 level relay
- VEGATOR 523, 527 2 level relay
Modes A/B-switch
Contact floating spdt Contact material AgNi and Au plated Turn-on voltage min. 10 mV
Switching current min. 10 µA
Breaking capacitance max. 500 VA AC, 54 W DC
4 VEGATOR
1 fail-safe relay
1 fail-safe relay
A - max. detection or overfill protection B - min. detection or dry run protection
max. 250 V AC/DC
max. 3 A AC, 1 A DC
Product description
Transistor ouput
Number of
- VEGATOR 521, 522 2 (synchronous switching with relays)
- VEGATOR 523, 527 3 (synchronous switching with relays) Galvanic separation floating Operating voltage max. 36 V DC Operating current max. 60 mA Transistor voltage loss U
- 1.5 V
CE
Approvals
VEGATOR 521 Ex, 522 Ex, 523 Ex, 527 Ex with input in classification Intrinsic safety EEx ia IIC, EEx ia IIB
EEx ib IIC, EEx ib IIB
Max. values U
= 20 V
O
IK = 126 mA P = 626 mW
Characteristics linear
EEx ia IIC EEx ia IIB EEx ib IIC EEx ib IIB
Max. permissible external inductance (mH) 0.5 1 1.5 2 2 9
Max. permissible external capacitance (nF) 97 78 68 486 200 1000
The intrinsically safe circuits are reliably (galvanically) separated from the non-intrinsically safe circuits up to a peak value of the nominal voltage of 375 V. On instruments with 2 channels (VEGATOR 527 Ex) the intrinsically safe circuits are reliably separated.
CE conformity
VEGATOR 521 Ex, 522 Ex, 523 Ex and 527 Ex signal conditioning instruments meet the protective regulations of EMC (89/336/EW) and NSR (73/23/EWG). Conformity has been judged acc. to the following standards: EMC Emission EN 50 081 - 1: 1993
Susceptibility EN 50 082 - 2: 1995
NSR EN 61 010 - 1: 1993
VEGATOR 5

1.5 Dimensions

Circuit board 100 x 160 x 1.5 European size
Multiple plug
Product description
5 TE
9
5,5
162
100
128,4
on
25,4
6 VEGATOR
Mounting

2 Mounting

2.1 Mounting instructions

VEGATOR can be mounted either in a card slot in a carrier BGT 596 or in a single housing type 505.
Card slot
Multipoint connector DIN 41 612, series F, 33­pole (d, b, z) with coded pin and mounting material for mounting in carrier BGT 596.
Card slot Ex
Multipoint connector DIN 41 612, series F, 33­pole (d, b, z) with coded pins, Ex separating chamber and mounting material for mounting in carrier BGT 596 Ex.M.
Single housing
Plastic housing type 505 for individual mounting of signal conditioning instruments with instrument width 5 TE (25.4 mm).
Mounting carrier
Mount the respective module (standard or Ex version) on your carrier BGT 596. Wire the connections of the multipoint connector according to the connection schematic on the following page.
The multipoint connector is available as follows:
- Wire-Wrap standard connection
1.0 x 1.0 mm
- Plug connection 2.8 x 0.8 mm
- Termi-Point standard connection
1.6 x 0.8 mm
- Soldering connection
- Screw terminals 2 x 0.5 mm
For further information on mounting, use the operating instructions of the carrier.
2
Mounting single housing
The socket can be either screwed directly to the mounting plate or plugged to a carrier rail (TS 35 x 7.5 acc. to EN 50 022 or TS 32 acc. to EN 50 035). Connect the terminals of the basic plate according to the connection schematic on the following page. For further information on mounting, use the operating instructions of the housing.
Coding
To prevent inadvertent swapping of the various instruments, the multiple plug of the instrument is provided with one or several coding holes.
By means of a fixed coded pin on the Ex module, it is ensured that only Ex instruments can be inserted.
To prevent inadvertent swapping of the various instruments, two additional coded pins are included with the instrument. Insert these coded pins into the predetermined holes on the multipoint connector. The coded pins must be inserted into the following positions:
Instrument Ex coding
coding
VEGATOR 521 a9 / c7 ––
VEGATOR 521 Ex a9 / c7 c23
VEGATOR 522 a11 / c7 ––
VEGATOR 522 Ex a11 / c7 c23
VEGATOR 523 a13 / c7 ––
VEGATOR 523 Ex a13 / c7 c23
VEGATOR 527 a15 / c7 ––
VEGATOR 527 Ex a15 / c7 c23
VEGATOR 7
Mounting
z b d
a c
o 1 o o 3 o o 5 o
VEGATOR 521 a9 VEGATOR 522 a11 VEGATOR 523 a13 VEGATOR 527 a15
o 7 o o 9 o o11o o13o o15o o17o o19o o21o o23o o25o o27o o29o o31o
c7
VEGATOR
c23
Ex-Codierung
Transparent cover
To protect the instrument against unauthor­ized or inadvertent use, the front plate of VEGATOR can be provided with a lockable transparent cover.
Ex version
Ex separating chamber
To ensure sufficient "air and creeping distance", an Ex separating chamber must be mounted to the connections of VEGATOR. Lead the cables through the Ex separating chamber and connect the cables. Fasten the Ex separating chamber with the lower holding screw. Note the operating instructions of the carrier BGT 596 Ex.M.
Mounting in carriers
When mounting VEGATOR with Ex approval in a carrier, a VEGA-Ex card slot must be used. Keep a distance of at least 10 mm (2 TE) to modules of other manufacturers. When mounting VEGATOR to the complete left position in the carrier, a blind cover of at least 20 mm (4 TE) must be mounted in front of the card slot of the instrument.
2
on
VEGATOR
Blind cover
Protection for Ex applications
For Ex applications, a protection of IP 20 must be maintained. Cover the gaps or free card slots with blind covers from the front.
8 VEGATOR
Electrical connection

3 Electrical connection

3.1 Connection instructions

The electrical connection should be carried out according to the following connection schematics. If only one channel is used on the VEGATOR 527 signal conditioning instruments, connect a resistor of 1 k (0.5 W) to the free connection pin of the se­cond channel. The resistor prevents the triggering of a fault signal caused by a mis­sing sensor.
For each relay output, an additional parallel­operating transistor output is available.
Reset of alarm functions
For deactivation of a signalling device (horn, diode etc.) in case of alarm (e.g. reaching the max. permissible level), an additional key can be connected to VEGATOR 521 or 522. This key can deactivate a level alarm. In case of failure (e.g. cable breakage), the alarm cannot be deactivated.
If the key is connected, the fail-safe relay has the same function as the level relay. However, it is possible to reset the fail-safe relay by pressing the key for reset of alarm functions.
If, for example, the max. level is reached and an acoustic warning system is activated, this system can be deactivated by pressing the key for reset of alarm functions. The second output (level relay) still signals the process control that the max. level has been reached.
Note
If very strong electromagnetic interferences are expected, screened cable is recommended. The screening of the cable should only be grounded on one sensor side (e.g. capacitive electrode).
The following illustrations show the currentless condition.

3.2 Connection schematic

VEGATOR 521 (Ex), 522 (Ex)
d b z
+-
Power supply
Fail safe transistor
Level transistor
+
-
Sensor 1
L1 (+) N (-)
2
6
10 12
+
16
-
18
+
20
-
22 24
28 30 32
Fail safe relay
Level relay
Key for reset of alarm functions (opener function)
+
-
VEGATOR 9
Electrical connection
VEGATOR 523 (Ex)
Power supply
Fail safe transistor
Level transistor 1
+
-
Sensor 1
L1 (+) N (-)
2
6
10 12
16 18
20 22 24
28 30 32
VEGATOR 527 (Ex)
Power supply
Fail safe transistor
Level transistor 1
L1 (+) N (-)
2
6
10 12
16 18 20 22 24
d b z
+-
+
­+
+
-
-
+
-
d b z
+-
+
­+
+
-
-
Fail safe relay
Level relay 1 Level relay 2
Level transistor 2
Fail safe relay
Level relay 1 Level relay 2
Level transistor 2
Ex version
Please note the instructions in the attached approval documents as well as the valid installation regulations for connection of Ex certified instruments.
Keep in mind that the Ex separating chamber must be mounted on the multipoint connector. Always lead the cables through the Ex separating chamber. Check also the operating instructions of the carrier BGT 596 Ex.M and the Ex instructions on page 8.
Wiring of the transistor outputs
(example VEGATOR 527 Ex)
VEGATOR 527 Ex PLC
+
I
B
U
CE
±
R
S
+
±
R
S
+
±
The resistor RS is used as short-circuit protection
±
+
0.1
+
0.2
+
0.3
Example:
UB = 24 V, UCE - 1.5 V, IB = max. 60 mA
UB – U
R
= ––––––––– = ––––––––––––– = 375
S
I
B
CE
24 V – 1.5 V
60 mA
Sensor 1
28
+
+
+
-
30
-
-
32
+
-
Sensor 2
10 VEGATOR
Setup

4 Setup

4.1 Indicating and adjustment elements

VEGATOR 521 VEGATOR 522 VEGATOR 523 VEGATOR 527
5
010
on
521
VEGATOR 527
B--
123456
zaze26
A B--
123456
zaze26
A
5
010
5
010
on
sec
0
00
12
sec sec
0
00
12
sec
5
max
min
522
010
1
5
010
2
on
523
5
010
1
5
010
2
on
527
12
3
5
4
010
5
6
on
off
on
off
7 8
9
1
5
010
2
on
527
16 13 1012
1 Connection plan 2 Lockable screw 3 Potentiometer as switching point
adjuster for max. (on VEGATOR 527
switching point adjuster for channel 1) 4 LED relay output 1 5 LED fault signal channel 1
6 Potentiometer as switching point
adjuster for min. (on VEGATOR 527
switching point adjuster for channel 2) 7 LED relay output 2 8 LED fault signal channel 2 9 LED power supply 10 Fixing screw 11 Multiple plug
111415
12 Min./Max. switch (VEGATOR 527) 13 Mode switch for variable fail-safe
adjustment (VEGATOR 522) 14 DIL switch block for channel 1 15 DIL switch block for channel 2 16 Transparent cover
VEGATOR 11
Setup
Potentiometer
The switching point of VEGATOR or the hysteresis (VEGATOR 522) can be changed via potentiometer (3 and 6). For potentiometer setting, use a small screwdriver.
Signal lamps
LEDs in the front plate indicate, operation, switching conditions and failures.
Green
- operation control lamp (9)
- mains voltage on, instrument in operation.
Red
- failure lamp (5 und 8)
- failure on the respective sensor current circuit by sensor failure or cable breakage.
- if the fail-safe relay is deenergised, the red signal lamp lights.
Yellow
- relay control lamp (4 and 7)
- the yellow control lamp reacts according to the adjusted mode (A/B).
- in general, the relay control lamp signals the activated (energised) condition of the relay.
- a dark signal lamp means that the relay is deenergised.
Min./Max. switch
(only on VEGATOR 527) The min./max. switch (12) is used for connection of both sensor inputs to one common min./max. signal.
DIL switch block
One DIL switch block (14 and 15) per output with 6 switches each is located on the circuit board of VEGATOR. The individual switches are assigned as follows: 1 A/B mode
A - max. detection (overfill protection)
B - min. detection (dry run protection) 2 Switch off delay (za) 3 Switch on delay (ze) 4 Integration time 2 s 5 Integration time 6 s 6 Integration time 12 s
You can set the mode (A - overfill protection, B - dry run protection) with switch 1. When used as part of a certified overfill protection system, switch 1 must be set to A.
With switches 2 and 3, the switch on and/or off delay can be adjusted independently. The switch on or off delay relates to the switching conditions of the relay and transistor outputs.
As example in the following illustration, mode A (max. detection or overfill protection) is selected (switch 1). The switch on delay is activated (switch 3) and the integration time is set to 8 seconds (switch 4, 5 and 6).
sec
0
00
zaze26
sec
12
1A/B mode 2 Switch off delay 3 Switch on delay 4 Integration time +2 s 5 Integration time +6 s 6 Integration time +12 s
B- -
123456
A
12 VEGATOR
Setup
With the switches 4, 5 and 6 the integration time can be adjusted respectively. The times of the activated time switches sum up, see illustration 10. When the switch on and off delay are switched on together, the set time is valid for both delay modes.
Switch
Time
0.2 sA/B00000
0.5 s A/B 0 0 0
2 s A/B 12 0 0
6 s A/B 060
8 s A/B 260
12 s A/B 0 0 12
14 s A/B 2 0 12
18 s A/B 0 6 12
20 s A/B 2 6 12
123456
adjusted delay mode.
The times are valid for the
3 to "-".
Either set switch 2 or switch
Changeover switch for variable fail­safe adjustment
(only on VEGATOR 522, VEGATOR 522 Ex)
The switch for the variable fail-safe adjustment (13) is located on the circuit board of VEGATOR 522. For detection of a measured value reduction due to failure, e.g. break of the probe or damage of the electrode, this function can be activated.
With separate sensor housing, a full signal or failure information is triggered, if the connection cable is broken.
As soon as the actual value falls below a previously adjusted measured value, a fault signal is triggered. By this means, a single point control with safety monitoring can be achieved (see "4.3 Switch point adjustment VEGATOR 522").

4.2 Principle of operation

The following illustrations show mode A (max. detection, overfill protection).
VEGATOR 521
Signal conditioning instrument as level switch for a sensor with single point control. A horn can be activated, if e.g. a certain level is exceeded. If the level falls below the set value, the horn is switched off. A typical application for this level switch is an overfill protection or dry run protection with one sensor. The point that can be adjusted has a fixed hysteresis.
off on
VEGATOR 522
Signal conditioning instrument as level switch for one sensor with double point control (adjustable hysteresis). The instrument has two potentiometers. Both points, the switch on and off point can be adjusted independently. This ensures that, e.g. a filling or emptying pump or a valve control can be achieved.
off off
on
VEGATOR 13
Setup
VEGATOR 523
Signal conditioning instrument as level switch for a sensor with dual single point control. The instrument has two potentiometers. Therefore it is possible to achieve with one sensor, e.g. an overfill protection and a dry run protection or an overfill protection with pre- or main alarm. The two points that can be adjusted have a fixed hysteresis.
off1
1
2
on1
off2on2 on2
VEGATOR 527
Signal conditioning instrument as level switch for single point control with two sensors or double point control with two sensors (see illustration 18). With VEGATOR 527, two sensors can be monitored at the same time with one signal conditioning instrument. The two points that can be adjusted have a fixed hysteresis.

4.3 Switching point adjustment

The following control modes are possible with the individual instruments:
- level detection [521, 522, 523, 527]
- double point control with one sensor and
adjustable hysteresis (min./max. control)
[522]
- single point control with one sensor and 2
level switches [523]
- two single point controls with 2 sensors
[527]
- double point control with 2 sensors [527]
Note
In general, the relay control lamp shows the activated (energised) condition of the relay. A dark control lamp means that the relay is deenergised. This means also that in case of power failure the relay always returns to deenergised condition.
VEGATOR 521 Single point control
Please note for single point control also "4.1 Indicating and adjustment elements" (the numbers in brackets refer to the illustration).
off1
1
2
12
14 VEGATOR
on1
on2 off2
Setup
Vertically installed measuring probes and pressure transmitters
Fill the vessel to the requested level.
On the DIL switch block (14) you select the mode (A - overfill protection, B - dry run protection).
Make sure that the integration time is switched off. The switches 2 - 6 on the DIL switch block (14) must be in Off position. Set the potentiometer (6) to 0.
Turn the potentiometer (6) very slowly clockwise until the relay control lamp (7) lights (mode A) or extinguishes (mode B).
If necessary, choose the following adjustments on the DIL switch block (14):
- switch off delay, switch on delay
- integration time (0.2 … 20 s)
VEGATOR is now ready for operation.
Horizontally mounted meas. probes and meas. probes for very adhesive products
On the DIL switch block (14) you select the mode (A - overfill protection, B - dry run protection). Make sure that the integration time is switched off. The switches 2 - 6 on the DIL switch block (14) must be in Off position. Set the potentiometer (6) to 0.
Fill the vessel until the measuring probe is covered completely. In case of measuring probes for very adhesive products, the sen­sitive tip (standard 100 mm) must be totally covered.
Turn the potentiometer (6) very slowly clockwise until the relay control lamp (7) lights (mode A) or extinguishes (mode B). Now note the adjusted value.
Set the potentiometer (6) to the average value of the two noted values.
If necessary, choose the following adjustments on the DIL switch block (14):
- switch off delay, switch on delay
- integration time (0.2 … 20 s)
VEGATOR is now ready for operation.
VEGATOR 522 Double point control with one sensor and adjustable hysteresis
For switch point adjustment, please also note "4.1 Indicating and adjustment elements" (the numbers in brackets refer to the illustration).
1
Lower the level at least down to 100 mm below the measuring probe.
Turn the potentiometer (6) very slowly clockwise until the relay control lamp (7)
2
lights (mode A) or extinguishes (mode B). Now note the adjusted value.
VEGATOR 15
Setup
Vertically mounted measuring probes or pressure transmitters
Fill the vessel to the required max. level.
Make sure that the integration time is switched off. The switches 2 - 6 on the DIL switch block (14) and the changeover switch (13) must be in Off position.
On the DIL switch block (14) you select the mode (A - overfill protection, B - dry run protection).
Set the min. potentiometer (6) to 0 and the max. potentiometer (3) to 10.
If you have chosen mode A, the relay control lamp (7) must light. If the control lamp does not light, turn the min. potentiometer (6) shortly to 10, to activitate the relay control lamp. Then turn the min. potentiometer again to 0.
If you have chosen mode B, the relay control lamp (7) should not light. If the control lamp lights, turn the min. potentiometer (6) shortly to 10, to deactivate the relay control lamp. Then turn the min. potentiometer again to 0.
Turn the max. potentiometer (3) very slowly anticlockwise until the relay control lamp lights or extinguishes.
VEGATOR 522 Single point control with safety monitoring
Note for the switch point adjustment "4.1 Indicating and adjustment elements" (the numbers in brackets refer to the illustration).
!
Only possible with horizontally mounted meas. probes or meas. probes for very adhesive products
Make sure that the integration time is switched off. The switches 2 - 6 on the DIL switch block (14) must be in Off position.
Lower the medium until the probe is uncovered. The level should be at least =100 mm below the probe.
Lower the medium to the required min. level.
Turn the min. potentiometer (6) very slowly clockwise until the relay control lamp lights or extinguishes.
If necessary, choose the following adjustments on the DIL switch block (14):
- switch off delay, switch on delay
- integration time (0.2 … 20 s)
VEGATOR is now ready for operation.
16 VEGATOR
Set the changeover switch for the variable fail-safe adjustment (13) to "on". Set the two potentiometers (3 and 6) to 0. The red failure lamp (8) extinguishes.
Turn the max. potentiometer (3) to 10. The red failure lamp (8) lights. Turn the max. potentiometer very slowly clockwise until the relay control lamp (7) extinguishes. Turn the max. potentiometer (3) another step anticlockwise.
Turn the min. potentiometer (6) very slowly clockwise until the relay control lamp (7) lights or extinguishes. Note this value.
Setup
Fill the vessel until the measuring probe is covered completely or is submerged into the medium by approx. 100 mm. The relay control lamp (7) changes switching condition.
Turn the min. potentiometer (6) very slowly clockwise until the relay control lamp (7) lights or extinguishes.
Note the value of the potentiometer. Set the potentiometer (6) to the average value of the two noted values.
If necessary, choose the following adjustments on the DIL switch block (14):
- switch off delay, switch on delay
- integration time (0.2 … 20 s)
VEGATOR is now ready for operation.
VEGATOR 523 Single point control with one sensor and two level switches
Note for switch point adjustment "4.1 Indicating and adjustment elements" (the numbers in brackets refer to the illustration).
For min. and max. information, max. and max. alarm or min. and min. alarm information
Make sure that the integration time is switched off. The switches 2 - 6 of the DIL switch blocks (14 and 15) must be in Off position.
Choose on the DIL switch blocks (14 and 15) the mode (A - overfill protection, B - dry run protection).
Set the two potentiometers (3 and 6) to 0. Lower the medium to the required min. level.
Turn the min. potentiometer (6) very slowly clockwise until the relay control lamp (7) lights or extinguishes.
Fill the vessel to the required max. level.
Turn the max. potentiometer (3) very slowly clockwise until the relay control lamp (4) lights or extinguishes.
If necessary, choose the following adjustments on the DIL switch blocks (14 and 15):
- switch off delay, switch on delay
- integration time (0.2 … 20 s)
VEGATOR is now ready for operation. (see
1
2
VEGATOR 17
also illustration 7 c).
Setup
VEGATOR 527 Single point control with 2 sensors
For switch point adjustment also note "4.1 Indicating and adjustment elements" (the numbers in brackets refer to the illustration).
Vertically mounted measuring probes and pressure transmitters
1
2
12
Choose on the DIL switch blocks (14 and 15) the mode (A - overfill protection, B - dry run protection). Make sure that the integration time is switched off. The switches 2 - 6 of the DIL switch blocks (14 and 15) and the min./ max. changeover switch (12) must be in Off position. Set the two potentiometers (3 and 6) to 0.
Channel 1
Fill the vessel to the required level. Turn the potentiometer (3) very slowly clockwise until the relay control lamp (4) lights (mode A) or extinguishes (mode B).
Channel 2
Carry out the adjustment for channel 1. Then repeat the adjustment with the adjustment elements for channel 2.
If necessary, choose the following adjustments on the DIL switch blocks (14 and 15):
- switch off delay, switch on delay
- integration time (0.2 … 20 s)
VEGATOR is now ready for operation.
Horizontally mounted meas. probes and meas. probes for very adhesive products
1
2
Choose on the DIL switch blocks (14 and 15) the mode (A - overfill protection, B - dry run protection). Make sure that the integration time is switched off. The switches 2 - 6 of the DIL switch blcoks (14 and 15) and the min./ max. changeover switch (12) must be on in Off position. Set the two potentiometers (3 and 6) to 0.
Channel 1
Lower the level at least 100 mm below electrode 1.
Turn the potentiometer (3) very slowly clockwise until the relay control lamp (4) lights (mode A) or extinguishes (mode B). Note the value adjusted now.
Fill the vessel until electrode 1 is covered completely. With electrodes for very adhesive products, the sensitive tip (standard 100 mm) must be covered completely.
18 VEGATOR
Setup
Turn the potentiometer (3) very slowly clockwise until the relay control lamp (4) lights (mode A) or extinguishes (mode B). Note the value adjusted now.
Set the potentiometer (3) to the average value of the two noted values.
Channel 2
Carry out the adjustment like for channel 1. Then repeat the adjustment with the adjustment elements for channel 2.
If necessary, choose the following adjustments on the DIL switch blocks (14 and 15):
- switch off delay, switch on delay
- integration time (0.2 … 20 s)
VEGATOR is now ready for operation.
VEGATOR 527 Double point control with 2 sensors
For switch point adjustment also note "4.1 Indicating and adjustment elements" (the numbers in brackets refer to the illustration).
Choose on the DIL switch blocks (14 and 15) the mode (A - overfill protection, B - dry run protection). Make sure that switch 1 of the two DIL switch blocks (14 and 15) is adjusted to the same mode (both the mode A or both the mode B)
Set the min./max. changeover switch (12) to off.
After having selected mode A, make sure that the meas. probe of channel 1 is always mounted in a higher position than the meas. probe for channel 2.
In mode B, the meas. probe of channel 1 must be mounted lower than the meas. probe of channel 2.
Make sure that the integration time is switched off. The switches 2 - 6 of the DIL switch blocks (14 and 15) must be in Off position. Set the potentiometer (3) to 0.
Channel 1
Lower the level at least 100 mm below the measuring probe 1.
Turn the potentiometer (3) very slowly clockwise until the relay control lamp (4) lights (mode A) or extinguishes (mode B). Note the adjusted value now.
1
Fill the vessel until the measuring probe is covered completely. With measuring probes for very adhesive products, the sensitive tip (standard 100 mm) must be covered completely.
2
Turn the potentiometer (3) very slowly clockwise until the relay control lamp (4) lights (mode A) or extinguishes (mode B). Note the adjusted value now.
Horizontally mounted meas. probes or
Set the potentiometer (3) to the average
meas. probes for very adhesive products
Here, signals of two measuring probes, e.g. mounted to min. or max. levels, can be linked to one common pump control signal.
VEGATOR 19
value of the two noted values.
Channel 2
Carry out the adjustment like for channel 1. Then repeat the adjustment with the adjustment elements for channel 2.
Set the min./max. changeover switch (12) to "on".
The relay of channel 1 now has the function of a min./max. relay (hysteresis signal). The relay of channel 2 still signals the covering of electrode 2.
If necessary, choose the following adjustments on the DIL switch blocks (14 and 15):
- switch off delay, switch on delay
- integration time (0.2 … 20 s)
VEGATOR is now ready for operation.
Setup
20 VEGATOR
Diagnostics

5 Diagnostics

5.1 Maintenance

The instrument is maintenance free.

5.2 Repair

Repairs require accessing internal parts of the instrument to correct an instrument de­fect. For safety and warranty reasons, repair must only be carried out by VEGA personnel.
In case of a defect, please return the respec­tive instrument with a short description of the flaw to our repair department.
Failures, their possible causes and their rectification are listed under “5.3 Failure rectification“.
VEGATOR 21
Diagnostics

5.3 Failure rectification

Fai lur e Rectification / Measure
Instrument is dead / Green operating control lamp extinguishes Check the voltage supply and the mains connections as described
Red failure LED lights Check the sensor inputs for the following failures:
under “Connection instructions“ on page 8. If the instrument still does not function, please call our service department.
- short circuit on the input
- sensor not connected correctly or polarity-reversed
- sensor line interrupted
- supply voltage too low
Check if the sensor is connected correctly
- failures in the sensor, effecting a current change to below 2 mA or above 23 mA, generate a fault signal on VEGATOR.
Measure the current in the connection cable to the sensor.
4 … 20 mA 12 … 20 V
mA
+ ±
Sensor
V
28
+
+
30
-
-
32
VEGATOR
For Ex systems, make sure that the Ex protection is not de­graded by the measuring instruments.
22 VEGATOR
Diagnostics
a. Current value < 2 mA
- Check the supply voltage on the connection cable to the sensor. The voltage should be at least 18 V. Should a value below 18 V should be measured, it indicates a defect in the signal conditioning instrument. In this case, return the instrument for repair to VEGA.
- Should the red failure lamp continue to light, separate the connection cable from the signal conditioning instrument and connect to the sensor inputs of the signal conditioning instrument a resistor of 1 kOhm. If the failure lamp continues to light, the signal conditioning instrument is defective. In such a case, return the instrument for repair to VEGA.
- Should the failure lamp extinguish, reconnect the signal conditioning instrument. Separate the sensor from the connection cable and connect a resistor of 1 kOhm in its place.
- Should the failure lamp continue to light, there is probably a short circuit in the connection line. Check the connection cable to the sensor.
- If the failure lamp extinguishes, the sensor is defective. Check the connected sensor.
b. Current value > 22 mA
- Check all connections and the connection cable to the sensor.
- Should the red failure lamp continue to light, separate the sensor from the connection cable and connect a resistor of 1 kOhm in its place. If the failure lamp extinguishes, the sensor is defective. Check the connected sensor.
- Should the failure lamp continue to light, reconnect the sensor. Separate the signal conditioning instrument from the connection cable and connect a resistor of 1 kOhm to its sensor input.
- Should the failure lamp continue to light, the signal conditioning instrument is defective. In such a case, return the instrument for repair to VEGA.
- If the failure lamp extinguishes, there is probably a short circuit in the connection line. Check the connection cable to the sensor.
VEGATOR 23
VEGA Grieshaber KG Am Hohenstein 113 D-77761 Schiltach Phone (0 78 36) 50 - 0 Fax (0 78 36) 50 - 201 e-mail info@vega-g.de
ISO 9001
The statements on types, application, use and operating conditions of the sensors and processing systems correspond to the latest infor­mation at the time of printing.
Technical data subject to alterations
19312-EN-041110
Loading...