VEGA TOR256C User Manual

Operating Instructions
in
out
VEGATOR 256C Signal conditioning instrument
Signal conditioning instruments
and communication

Contents

1 About this document
1.1 Function .............................
1.2 Target group ..........................
1.3 Symbolism used .......................
2 For your safety
2.1 Authorised personnel ....................
2.2 Appropriate use........................
2.3 Warning about misuse ...................
2.4 General safety instructions ................
2.5 CE conformity .........................
2.6 Safety instructions for Ex areas ............
2.7 Environmental instructions ................
3 Product description
3.1 Conguration..........................
3.2 Principle of operation ....................
3.3 Operation ............................
3.4 Storage and transport ...................
4 Mount
4.1 General instructions.....................
4.2 Mounting instructions....................
4 4 4
5 5 5 5 5 6 6
7 7 7 8
9 9
5 Connecting to power supply
5.1 Preparing the connection .................
5.2 Wiring plan ...........................
6 Set up
6.1 Adjustment system .....................
6.2 Adjustment elements ....................
6.3 Switching point adjustment................
7 Maintenance and fault rectication
7.1 Maintenance ..........................
7.2 Remove interferences ...................
7.3 Instrument repair .......................
8 Dismounting
8.1 Dismounting steps ......................
8.2 Disposal .............................
2 VEGATOR 256C - Signal conditioning instrument
10 10
13 13 13
15 15 15
17 17
23409-EN-070316
9 Supplement
9.1 Technical data.........................
9.2 Dimensions ...........................
Contents
18 20
23409-EN-070316
VEGATOR 256C - Signal conditioning instrument 3
About this document

1 About this document

1.1 Function

This operating instructions manual has all the information you need for quick setup and safe operation. Please read this manual before you start setup.

1.2 Target group

This operating instructions manual is directed to trained, qualied personnel. The contents of this manual should be made available to these personnel and put into practice by them.

1.3 Symbolism used

Information, tip, note
This symbol indicates helpful additional information.
Caution: If this warning is ignored, faults or malfunctions can result. Warning: If this warning is ignored, injury to persons and/or serious damage to the instrument can result. Danger: If this warning is ignored, serious injury to persons and/or destruction of the instrument can result.
Ex applications
This symbol indicates special instructions for Ex applications.
l List
The dot set in front indicates a list with no implied sequence.
à Action
This arrow indicates a single action.
1 Sequence
Numbers set in front indicate successive steps in a procedure.
4 VEGATOR 256C - Signal conditioning instrument
23409-EN-070316
For your safety

2 For your safety

2.1 Authorised personnel

All operations described in this operating instructions manual must be carried out only by trained specialist personnel authorised by the operator. For safety and warranty reasons, any internal work on the instruments must be carried out only by personnel authorised by the manufacturer.

2.2 Appropriate use

VEGATOR 256C is a universal signal conditioning instrument for connection of a level switch.
You can nd detailed information on the application range of VEGATOR 256C in chapter "Product description".

2.3 Warning about misuse

Inappropriate or incorrect use of the instrument can give rise to application-specic hazards, e.g. vessel overll or damage to system components through incorrect mounting or adjustment.

2.4 General safety instructions

VEGATOR 256C is a high-tech instrument requiring the strict observance of standard regulations and guidelines. The user must take note of the safety instructions in this operating instructions manual, the country-specic installation standards (e.g. the VDE regulations in Germany) as well as all prevailing safety regulations and accident prevention rules.

2.5 CE conformity

VEGATOR 256C is in CE conformity with EMC (89/336/EWG) and LVD (73/23/EWG) as well as 93/68/EWG.
Conformity has been judged according to the following standards:
l EMC:
- Emission EN 50081-1: 1992
- Suceptibility EN 50082-2: 1995
l LVD: EN 61010-1: 1993
23409-EN-070316
VEGATOR 256C - Signal conditioning instrument 5
For your safety

2.6 Safety instructions for Exareas

Please note the Ex-specic safety information for installation and operation in Ex areas. These safety instructions are part of the operating instructions manual and come with the Ex­approved instruments.

2.7 Environmental instructions

Protection of the environment is one of our most important duties. That is why we have introduced an environment management system with the goal of continuously improving company environmental protection. The environment man­agement system is certied according to DIN EN ISO 14001.
Please help us full this obligation by observing the environ­mental instructions in this manual:
l Chapter "Storage and transport" l Chapter "Disposal"
6 VEGATOR 256C - Signal conditioning instrument
23409-EN-070316

3 Product description

3.1 Conguration
Product description
Scope of delivery
Components
Area of application
Functional principle
The scope of delivery encompasses:
l VEGATOR 256C signal conditioning instrument l Documentation
- this operating instructions manual
VEGATOR 256C consists of:
l VEGATOR 256C signal conditioning instrument
min.
1
max.
2
3
200...250VAC 3VApower supply
R
Relais: max 250V, 5A, 750VA
1
2
10
0
4
5
6
L1
7
N
8
Fig. 1: VEGATOR 256C 1 Signal lamp - Relay output 2 Potentiometer for switching point adjustment

3.2 Principle of operation

VEGATOR 256C is a single signal conditioning instrument for processing the signals of conductive probes.
VEGATOR 256C signal conditioning instrument can power connected instruments and process their measuring signals.
Supply
You can nd detailed information on the power supply in chapter "Technical data" in the "Supplement".

3.3 Operation

A potentiometer is located on the front plate of VEGATOR
256C. With this potentiometer, the switching point can be adjusted.
23409-EN-070316
VEGATOR 256C - Signal conditioning instrument 7
Product description

3.4 Storage and transport

Packaging
Storage and transport tem-
perature
Your instrument was protected by packaging during transport. Its capacity to handle normal loads during transport is assured by a test according to DIN EN 24180.
The packaging of standard instruments consists of environ-
ment-friendly, recyclable cardboard. For special versions, PE foam or PE foil is also used. Dispose of the packaging material via specialised recycling companies.
l Storage and transport temperature see "Supplement -
Technical data - Ambient conditions"
l Relative humidity 20 85 %
8 VEGATOR 256C - Signal conditioning instrument
23409-EN-070316

4 Mount

4.1 General instructions

Mount
Installation location
Mount
VEGATOR 256C signal conditioning instrument with plug-in socket for mounting on carrier rail according to EN 50022.

4.2 Mounting instructions

You can mount the VEGATOR 256C signal conditioning instruments either to the wall (2 screws) or plug it onto a carrier rail.
Carrier rail mounting
Place the signal conditioning instrument onto the carrier rail
(35x7.5 according to EN 50022) from below and press the instrument against the carrier rail until it snaps in.
Wall mounting
Fasten the instrument directly to the wall by means of two screws (max. ø 3 mm/0.12 in).
")
32
/
5
4 mm
(
")
64
/
23
60 mm (2
/64")
7,5 mm
19
(
/64")
Fig. 2: Drilling template VEGATOR 256C
22 mm
55
(
23409-EN-070316
VEGATOR 256C - Signal conditioning instrument 9
Connecting to power supply

5 Connecting to power supply

5.1 Preparing the connection

Note safety instructions
Select power supply
Selecting connection cable
Cable screening and ground-
ing
Always keep in mind the following safety instructions:
l Connect only in the complete absence of line voltage l If overvoltage surges are expected, overvoltage arresters
should be installed
Tip:
We recommend VEGA overvoltage arresters B61-300 (power supply VEGATOR 256C) and B62-36G (sensor supply).
You can nd detailed information on the power supply in chapter "Technical data" in the "Supplement".
Power supply of VEGATOR 256C is connected with standard cable according to the national installation standards.
Standard two-wire cable without screening can be used to connect sensors. If electromagnetic interference is expected, screened cable must be used.
From a cable length of 50 m use one cable for each signal conditioning instrument.
If you want to use a common cable, the min. and max. cables must be screened. The screen must be connected to ground on both sides.

5.2 Wiring plan

Level detection
3
Fig. 3: Level detection 1 Supply 2 Relay output 3 Mass 4 max.
min.
1
max.
4
2 1
2
3
0
10
4
5
6
7
L1
8
N
10 VEGATOR 256C - Signal conditioning instrument
23409-EN-070316
Pump control (min./max. con- trol)
Connecting to power supply
min.
1
max.
3
2
1
5
4
3
2 1
2
3
0
10
4
5
6
7
L1
8
N
6
2
3
1
Fig. 4: Pump control 1 Supply 2 Relay output 3 Mass 4 max. 5 min. 6 Probe, e.g. EL3
Note:
Mutiple rod probes connected to several signal conditioning instruments or to a multiple channel instrument need a ground rod to keep the signal conditioning instruments from inuenc­ing each other.
If there are several VEGATOR 256C, it is absolutely necessary that they be connected identically, i.e. the rst supply line to all no. 7 terminals and the second supply line to all no. 8 terminals. Exchanging no. 7 and no. 8 or connecting to dierent phases is not permitted.
Exchange of a VEGATOR 256B
If you want to replace a VEGATOR 256B by a VEGATOR 256C or if you want to use a VEGATOR 256C as well as a
VEGATOR 256B on a multiple rod probe, the supply voltage must be wired as follows.
VEGATOR 256C VEGATOR 256B
L17 7
N 85
Keep in mind that the function of the relay control lamp on VEGATOR 256C is inverted to that of VEGATOR 256B.
23409-EN-070316
VEGATOR 256C - Signal conditioning instrument 11
Connecting to power supply
VEGATOR 256C
The relay control lamp lights when the relay is energized and extinguishes when the relay is deenergized.
VEGATOR 256B
The relay control lamp extinguishes when the relay is energized and lights when the relay is deenergized.
12 VEGATOR 256C - Signal conditioning instrument
23409-EN-070316

6 Set up

6.1 Adjustment system

min.
1
max.
2
3
10
0
4
5
200...250VAC 3VApower supply
6
R
L1
7
8
Relais: max 250V, 5A, 750VA
N
Fig. 5: Indicating and adjustment elements 1 Signal lamp - Relay output 2 Potentiometer for switching point adjustment

6.2 Adjustment elements

Set up
1
2
Control lamp
The yellow relay control lamp (LED) shows the switching condition of the relay.
In general, the relay control lamp indicates the activated (energised) condition of the relay.
A dark relay control lamp means that the relay is deenergised.
Potentiometer for switching point adjustment
A potentiometer for switching point adaptation is located on the front plate of the signal conditioning instrument. With this potentiometer you can adapt the measuring system to the conductivity of the product.

6.3 Switching point adjustment

Level detection
23409-EN-070316
VEGATOR 256C - Signal conditioning instrument 13
l Connecting the signal conditioning instrument to voltage
supply
l Turn potentiometer (2) to position 0 l Fill the vessel until the max. probe is covered approx. 1 cm
by the product
l Turn the potentiometer (2) slowly clockwise until the yellow
LED extinguishes
The switching sensitivity of the signal conditioning instrument
is now adapted to the conductivity of the product.
Set up
Pump control (min./max.)
l at rst, connect only the ground and max. electrode to
VEGATOR 256C (terminal 2 and 3)
l Connecting the signal conditioning instrument to voltage
supply
l Turn potentiometer (2) to position 0 l Fill the vessel until the max. probe is covered approx. 1 cm
by the product
l Turn potentiometer (2) slowly clockwise until the yellow
LED extinguishes
l Connect min. electrode to terminal 1 of VEGATOR 256C
The switching sensitivity of the signal conditioning instrument is now adapted to the conductivity of the product, i.e. the relay output deenergises with max. level and the yellow control lamp extinguishes.
This switching condition remains until the level falls below the position of the min. electrode, i.e. the relay output energizes at min. level and the yellow control lamp lights.
The relay output deenergises again when the max. level is reached.
14 VEGATOR 256C - Signal conditioning instrument
23409-EN-070316
Maintenance and fault rectication
7 Maintenance and fault rectication

7.1 Maintenance

When used as directed in normal operation, VEGATOR 256C is completely maintenance free.

7.2 Remove interferences

Causes of malfunction
Remove interferences
24 hour service hotline
Failure
VEGATOR 256C oers maximum reliability. Nevertheless, faults can occur during operation. These may be caused by the following, e.g.:
l Measured value from sensor not correct l Voltage supply l Interference on the cables
The rst measures to be taken are to check the input/output signal. In many cases, the causes can be determined this way and the faults rectied.
However, should this measures not be successful, call the VEGA service hotline in urgent cases under the phone no. +49
1805 858550.
The hotline is available to you 7 days a week round-the-clock. Since we oer this service world-wide, the support is only
available in the English language. The service is free of charge, only the standard telephone costs will be charged.
? The signal conditioning instrument does not switch when
the respective probe is covered or uncovered
l Conductivity of the product too low
à Check if the conductivity value of your product is at
least 7.5 µS/cm
l Line break to the sensor
à Check the connection cables from the probe to the
signal conditioning instrument

7.3 Instrument repair

If a repair is necessary, please proceed as follows:
You can download a return form (23 KB) in the Internet from
our homepage
and Certicates - Repair form".
23409-EN-070316
VEGATOR 256C - Signal conditioning instrument 15
www.vega.com under: "Downloads - Forms
Maintenance and fault rectication
By doing this you help us carry out the repair quickly and
without having to call for needed information.
l Print and ll out one form per instrument l Clean the instrument and pack it damage-proof l Attach the completed form and probably a safety data
sheet outside on the packaging
l Please ask the agency serving you for the address of your
return shipment. You nd the respective agency on our website
www.vega.com under: "Company - VEGA world-
wide"
16 VEGATOR 256C - Signal conditioning instrument
23409-EN-070316
Dismounting

8 Dismounting

8.1 Dismounting steps

Take note of chapters "Mounting" and "Connecting to power supply" and carry out the listed steps in reverse order.

8.2 Disposal

The instrument consists of materials which can be recycled by specialised recycling companies. We use recyclable materials and have designed the electronics to be easily separable.
WEEE directive 2002/96/EG
This instrument is not subject to the WEEE directive 2002/96/ EG and the respective national laws (in Germany, e.g. ElektroG). Pass the instrument directly on to a specialised
recycling company and do not use the municipal collecting points. These may be used only for privately used products according to the WEEE directive.
Correct disposal avoids negative eects to persons and environment and ensures recycling of useful raw materials.
Materials: see chapter "Technical data"
If you cannot dispose of the instrument properly, please
contact us about disposal methods or return.
23409-EN-070316
VEGATOR 256C - Signal conditioning instrument 17
Supplement

9 Supplement

9.1 Technical data

General data
Series Module unit for mounting on carrier rail 35x7.5
or 35x15 according to EN 50022
Dimensions W = 37 mm (1.46 in), H = 68 mm (2.68 in), D =
66 mm (2.6 in)
Weight approx. 170 g (6 oz) Housing material Noryl SE100, Lexan 920A
Voltage supply
Supply voltage
- Standard 200 250 VAC, 50/60 Hz
- optional 24 V, 42 V, 48 V, 100 … 130 VAC(+10 %,
-15 %)
Power consumption 1 VA
Sensor input
Quantity 1x
Response resistor 1 … 200 kOhm adjustable Measuring circuit approx. 12 V eff., max. 1 mA Switching hysteresis approx. 20 %
level detection or 1x pump control (min./
max.)
Relay output
Quantity 1 (1x level detection)
Mode Max. detection or overll protection Integration time approx. 500 ms
Contact Changeover contact (spdt) Contact material AgNi 0.15 hard gold-plated Turn-on voltage min. 10 mVDC, max. 250 VAC, 250 VDC Switching current min. 10 µAD Breaking capacity max. 750 VA, max. 54 W
Adjustment elements
Control lamp to the indication of the relay switching status
Potentiometer to the adaptation of the product conductivity
18 VEGATOR 256C - Signal conditioning instrument
C, max. 5 AAC, 1 ADC
23409-EN-070316
Ambient conditions
Ambient temperature -20 +50 °C (-4 +122 °F) Storage and transport temperature -40 +70 °C (-40 +158 °F)
Electromechanical data
Spring-loaded terminals for wire cross-section up to 1.5 mm²
(0.0023 in²)
Electrical protective measures
Protection IP 20
Protection class II
Supplement
23409-EN-070316
VEGATOR 256C - Signal conditioning instrument 19
Supplement

9.2 Dimensions

")
64
/
43
68 mm (2
min.
1
max.
2
3
10
0
4
5
200...250VAC 3VApower supply
6
R
L1
7
N
Relais: max 250V, 5A, 750VA
8
Fig. 6: Dimensions VEGATOR 256C
66 mm (1 19/32")
37 mm
29
(1
/64")
20 VEGATOR 256C - Signal conditioning instrument
23409-EN-070316
Supplement
23409-EN-070316
VEGATOR 256C - Signal conditioning instrument 21
Supplement
22 VEGATOR 256C - Signal conditioning instrument
23409-EN-070316
Supplement
23409-EN-070316
VEGATOR 256C - Signal conditioning instrument 23
VEGA Grieshaber KG Am Hohenstein 113 77761 Schiltach Germany Phone +49 7836 50-0 Fax +49 7836 50-201 E-mail: info@de.vega.com
www.vega.com
ISO 9001
All statements concerning scope of delivery, application,
practical use and operating conditions of the sensors and
processing systems correspond to the information avail-
able at the time of printing.
© VEGA Grieshaber KG, Schiltach/Germany 2007
Subject to change without prior notice 23409-EN-070316
Loading...