Vaisala BAROCAP Technical data

Vaisala BAROCAP® Pressure Module
Features/Benefits
Based on silicon capacitive
sensor technology - BAROCAP®
Excellent accuracy, repeatability,
and long-term stability
range
Calibration traceable to
international standards (NIST)
An optional barometer module
for the Vaisala Data Logger QML201
www.vaisala.com
The Vaisala BAROCAP Module has excellent accuracy, repeatibility and long-term stability over a wide range of operating temperatures. The total accuracy is ± 0.25 hPa with factory calibration. It is well suited for demanding synoptic applications, and an affordable choice for Vaisala automatic weather stations utilizing the Vaisala Data Logger QML201.
®
Pressure
The ne adjustment and calibration of the sensor are handled using the electronic working standards, which are traceable to international standards (NIST). The pressure module can be calibrated independently of the QML201 data logger.
The pressure module is located on the QML201 CPU board. The pressure range of the sensor is 500 … 1100 hPa and operating temperature range
-40 ...+60 °C.
Pressure module located on the central processing unit QML201.
Technical data
Property Description/value
Pressure range 500 ... 1100 hPa Linearity * ± 0.10 hPa Hysteresis * ± 0.03 hPa Repeatability * ± 0.03 hPa Calibration uncertainty ** ± 0.15 hPa Accuracy at +20 °C (+68 °F) *** ± 0.20 hPa Temperature dependence **** ± 0.10 hPa Total accuracy at -40 … +60 °C (-40 … +140 °F) ± 0.25 hPa Long-term stability ± 0.10 hPa/year Operating temperature range -40 ... +60 °C Resolution 0.10 hPa
* Defined as ± 2 standard deviation limits of endpoint non-linearity, hysteresis error or repeatability error. ** Defined as ± 2 standard deviation limits of inaccuracy of the working standard including traceability to NIST. *** Defined as the root sum of the squares (RSS) of endpoint non-linearity, hysteresis error, repeatability error and calibration uncertainty at room temperature. **** Defined as ± 2 standard deviation limits of temperature dependence over the operating temperature range.
For more information, visit www.vaisala.com or contact us at sales@vaisala.com
Ref. B211084EN-A ©Vaisala 2010
This material is subject to copyright protection, with all copyrights retained by Vaisala and its individual partners. All rights reserved. Any logos and/or product names are trademarks of Vaisala or its individual partners. The reproduction, transfer, distribution or storage of information contained in this brochure in any form without the prior written consent of Vaisala is strictly prohibited. All specifications — technical included — are subject to change without notice.