UVP White-UV Transilluminators User Manual

White/UV Transilluminators

Instruction Manual

UVP, LLC Ultra-Violet Products Ltd. 2066 W 11th Street, Upland, CA 91786 Unit 1,Trinity Hall Farm Estate, Tel: (800) 452-6788 / (909) 946-3197 Nuffield Road Cambridge CB4 1TG UK Fax: (909) 946-3597 Tel: +44(0)1223-420022 / Fax: +44(0)1223-420561
Web site: www.uvp.com
81-0286-01 Rev C
White/UV Transilluminators Page 2
UV Blocking Cover is
surface
Control knob for selection of Power switch
Long life filters: 20 x

Introduction

The UVP Transilluminator offers the researcher uniform and intense sources of ultraviolet light (radiation) and white light in one unit. UV and white light 20x20cm filter areas side by side.
The special UV design emits high intensity excitation UV wavelength for back-illumination of transparent fluorescent materials. The UV back-illumination provides a highly sensitive method to detect double-stranded nucleic acids that have been labeled with fluorescent dyes such as ethidium bromide or acridine orange. Single stranded nucleic acids may be detected, but with excitation wavelength sensitive for nucleic acid visualization.
White light is used to illuminate coommassie blue, silver stain and protein gels. The transilluminator is uniquely designed with increased UV intensity and uniformity, instant on
capabilities, no lamp flicker and reduced electrical consumption. NOTE: Though the manual refers to the midrange UV waveband as 302nm, others refer to this
region as 300nm or 312nm. The spectral output of all these regions is the same. A Word of Caution: UV Transilluminators are powerful sources of UV radiation that will cause
damage to unprotected eyes and skin. Before operating any unit, be sure all personnel in the area are properly protected. If not using the transilluminator with an imaging system darkroom, a UV Blocking Cover should be attached to the transilluminator. Even though this cover blocks the ultraviolet radiation emitted by the unit, UV Blocking Eyewear should be worn as well.
The White/UV Transilluminator features:
hinged and adjustable to varying heights for access to the transilluminator
20cm UV and white light areas side by side
UV or white light
White/UV Transilluminators Page 3
When the UV Blocking Cover is not being used, UV light may escape through the
Safety Plug
Specifications
White/UV Transilluminators
White/UV Transilluminator models are designed with white light and one or two UV wavelengths. Units are equipped with a high efficiency electronic ballast. Dimensions: 13.25D x 19.13W x 5.64H in. (337 x 486 x 143mm)

Model Part Number Volt/Hz Wavelength Filter Size No. of Tubes /Watts

TLW-20 95-0415-01 100-115/60 365nm 20 x 20cm 4 x 8W 95-0415-02 230/50 White 20 x 20cm 2 x 8W
TMW-20 95-0415-04 100-115/60 302nm 15 x 15cm 4 x 8W 95-0415-05 230/50 White 20 x 20cm 2 x 8W
LMW-20 95-0418-01 100-115/60 302/365nm 20 x 20cm 4 x 8W 95-0418-02 230/50 White 20 x 20cm 2 x 8W
Transilluminator Operation
Safety Precautions
holes dedicated to accepting the bracket pins of the UV Blocking Cover.
Remove the black safety plugs from
their package
Insert the safety plugs through the
holes as shown.
Setting Up the Transilluminator
Place the transilluminator on a level work surface. Be sure that an air space exists around the
bottom of the work surface. This space allows for the proper air circulation through the unit.
Plug the female end of the power cord into the transilluminator. For 230 volt models, or those
requiring special power cord connectors, ensure that the proper configuration of male connector or plug has been properly connected to the power cord.
Plug the male end of the power cord into a properly grounded electrical outlet. The proper
voltage of the transilluminator is found on the product information label. If using the transilluminator with an imaging system, a jumper cable is required for connecting to the darkroom. Refer to the imaging system documentation for additional instructions.
The transilluminator is equipped with a UV Blocking Cover. Remove the brown protective
paper from the cover. Insert the bracket pins on the cover into the holes on the front of the transilluminator. The cover is adjustable to varying angles for access to the filter surface.
White/UV Transilluminators Page 4
NOTE: Do not operate the unit without securing the cover. If the cover is missing, a UV Blocking Faceshield must be worn to avoid UV exposure to the skin. UV Blocking Eyewear should be worn even with the cover in place to avoid accidental UV exposure.
Using the Transilluminator
Place gel/sample on the filter area. It is recommended that researchers place the gels on a Gel­Tray to protect the filter surface from cuts and scratches. It is recommended that gloves be worn to prevent skin contact with gel and staining agents.
The transilluminator can be turned on with the wavelength selector knob poin tin g in any
position. Press the ON/OFF switch to ON. The tubes within the unit will begin glowing beneath the filter.
Dial the knob to the appropriate wavelength setting. After viewing the sample, turn the
transilluminator off.
Maintenance/Repair/Technical Assistance
Cleaning and Care of the Transilluminator
Clean unit surface with a damp soft cloth or sponge. Never use abrasive cleaners (can damage the filter surface). To protect the filter glass and minimize moisture and liquids on the glass, it is recommended that you use a UV transmitting Gel-Tray. Refer to the Replacement Parts for ordering information.
Replacing Tubes in the Transilluminator
Disconnect the transilluminator for the electrical supply. Remove the filter cover: Use a Phillips head screwdriver to remove the four screws on the
sides of the unit. Lift the filter cover off the unit.
Remove the reflectors on the left and right side. Slide the reflectors up out of the unit. Remove the tube: Carefully rotate the tube and slide out of the socket. Replace with a new
tube by sliding the tube into the socket and rotating into place. Insert the reflectors back into place and reattach the filter cover.
Replacement Parts/Accessories
For replacement parts or components not shown here, please call UVP Customer Service or place of purchase. Please have the transilluminator model number available when you call.
Replacement Part Description Part Number
Filter Assembly (LMW-20, TMW-20) 38-0189-05 Filter Assembly (TLW-20) 38-0189-08 Cover, UV blocking 19-0112-01 Tube, 8 Watt, 302nm midrange UV (FL8E) 34-0042-01 (Qty. 4 Required) Tube, 8 Watt, 365nm longwave UV (F8T5/BL) 34-0006-01 (Qty. 4 Required) Tube, 8 Watt, cool white (T8T5/CW) 34-0056-01 (Qty. 2 Required) Ballast, electronic universal voltage 42-0068-04 (115V) 42-0068-06 (230V) Fuse, 2 amp, 250V, Slo-Blo (all voltages) 56-0002-01 (Qty. 2 Required) White/UV Selector Switch 53-0196-02
White/UV Transilluminators Page 5
Accessories Description Part Number
Gel-Cutter 85-0002-01 Gel-Scooper 85-0006-01 Gel-Tray, UV Transmitting, Lg. 16.5”W x 10.5”D (42 x 27cm) 85-0005-01 Gel-Tray, High Performance UV Transmitting (25 x 26cm) 38-0296-03 Gel-Ruler, UV Fluorescing 85-0003-01 Spectacles, UV Blocking 98-0002-01 Goggles, UV Blocking 98-0002-02 Faceshield, UV Blocking 98-0002-04
Technical Support
UVP offers technical support for all of its products. If you have any questions about the product’s use or, operation, please contact UVP’s offices at the following locations.
If you are in North America, South America, If you are in Europe, Africa, the Middle East, East Asia or Australia: Western Asia:
Call (800) 452-6788 or (909) 946-3197 Call +44(0) 1223-420022 Customer Customer Service regular business regular business days, between 9 am days, between 7 am and 5 pm PST and 5:30 pm
E-Mail: info@uvp.com E-Mail: uvp@uvp.co.uk
Fax Customer Service: (909) 946-3597 Fax Customer Service: +44(0)1223-420561
Write to : UVP, L LC Write to : Ultra-Violet Products Ltd.
2066 W. 11th Street Unit 1, Trinity Hall Farm Est., Nuffield Rd.
Upland, CA 91786 USA Cambridge CB4 1TG UK
Note: A Returned Goods Authorization (RGA) number must be obtained from UVP Customer
Service before returning any product.
Warranty
UVP, LLC warrants its Ultraviolet Transilluminators to be free of detects in materials and workmanship for a period of two (2) years from date of purchase. The foregoing warranty of UVP shall be of no force and effect if buyer has modified or damaged the product. Tubes and filters are warranted for 90 days.
All warranties or merchantability and fitness for any purpose and all other warranties, expressed or implied, except those expressly set forth herein, are deemed waived and excluded.
UVP’s duty under the warranty is limited to replacement and/or repair of the defective part at the option of UVP, LLC. UVP shall not be liable for any expenses or damages incurred by the purchaser except as expressly set forth herein, and in no event shall UVP be liable for any special, incidental or consequential damages of any kind. This warranty does not supersede any statutory rights that may be available in certain countries.
Loading...