Model UT2000: OPERATING MANUAL
1
Table of Contents
Title
Page 2
ATTENTION
Please go through the instruction manual before using your meter and also pay attention to the section VI. SAFETY RULES.
INTRODUCTION£º
UT2000 series are the large LCD DMM with accurate reading, stable quality, multifunctional and modern design of 3 1/2 and 4 1/2 handheld digital multimeter used for measuring DC voltage and current, AC voltage and current, resistance, capacitance, frequency, temperature, positive diode resistance, transistor hFE and short circuit testing. It is suitable for engineering design, laboratory testing and industrial manufacturing and repair etc.
International Electrical Symbols:
AC (Alternating Current) DC (Direct Current) AC or DC Grounding Double Insulated Deficiency of Built-In Battery. Continuity Test. Diode. Fuse. Warning. Refer to the Operating Manual. Conforms to Standards of European Union.
2
Model UT2000: OPERATING MANUAL
Page 3
FEATURES
l DC basic accuracy: (31/2) 0.5%; (4 1/2) 0.05%. l Max. display:1999 (3 1/2 digit) ; 19999
(4 1/2 digit).
l Reading display frequency: 2-3 times per
second.
l Automatic indication of polarity, decimal place
and signs. Automatic indication of "1" for over-range."
" shows when the battery is
insufficient.
l Auto-zeroing for capacitance test. l Automatic circuitry protection and buzzer
sounds for over-range and mis-operation.
l With automatic power-off function. l Release the lock by press the cases top button.
Adopt large LCD display, 25mm height digit, the angle of the display can be rotated within 70' to select the most suitable reading angle..
l Size: 185x89x32mm. l Weigh: approx. 300g (including battery).
3
Model UT2000: OPERATING MANUAL
Page 4
ENVIRONMENTAL CONDITIONS
Guarantee accuracy: 23
o
C 5oC ; relative humidity:
<75%. Temperature range: working temperature: 0
o
C to
40
o
C (32oF to 104oF),
storing temperature: -10
o
C to 50 oC (14
o
F to 122oF) Battery: 9Vbattery (NEDA1604,6F22 or similar type). Relative humidity: 0
o
C-31oC,<80%; 31oC-40oC,<50%
Altitude: (operating) 2000 meters, and 10000 meters for storage. Replacement Fuse: 5x20mm, 0.2A/250V FAST (Except Model 2001, 2A/250V FAST)
4
Model UT2000: OPERATING MANUAL
Page 5
TECHNICAL SPECIFICATION
Accuracy: (a% reading + No. of digits), guaranteed for 1 year.
1. DC V oltage
4 1/2Digit ACCURACY
RESOLUTION
(0.5%of rdg+1digit)
3 1/2 Digit ACCURACY
RANGE
200mV
2V
20V
200V
1000V
2001 2002/5 2006 2007
(0.8%of rdg+2digits)
RESOLUTION
(0.05%of
rdg+3digits)
RANGE
200mV
2V
20V
200V
1000V
10 V
100 V
1mV
10mV
100mV
100 V
1mV
10mV
100mV
1V
2003 2004
(0.1%of
rdg+2digits)
(0.2%of
rdg+5digits)
(0.1%of
rdg+5digits)
Input impedance: 10M on all ranges. Overload protection: DC or AC peck value of 1000V. (expect 200mV range with the maximum value being 250V rms)
5
Model UT2000: OPERATING MANUAL
Page 6
2. AC Voltage
6
RESOLUTION
(0.8%of rdg+3digits)
3 1/2 Digit ACCURACY
RANGE
200mV
2V
20V 200V 750V
2001 2002/5 2006 2007
(1.2%of rdg+3digits)
100 V
1mV
10mV
100mV
1V
(1.2%of rdg+3digits)
RESOLUTION
(0.8%of rdg+10digits)
4 1/2 Digit ACCURACY
RANGE
2V
20V 200V 750V
2003 2004
(1%of rdg+15digits)
100 V
1mV
10mV
100mV
Input impedance: 10M on all ranges. (2003/4 2M ) Frequency all range: Below 200V; 40-400Hz; 750V;40-200Hz. Overload protection: AC 750V rms or 1000Vpeak continuous on all ranges. (expect 200mVrange with the maximum value being 250Vrms) Indication: verage value (rms of sine wave).
Model UT2000: OPERATING MANUAL
Page 7
7
3. DC Current
RESOLUTION
(0.8%of
rdg+1digit)
3 1/2 Digit
ACCURACY
RANGE
200 A
2mA
20mA
200mA
2A
10A
20 A
0.1 A 1 A
10 A
100 A
1mA
10mA 10n A
2007
2006
2002/5
2001
(1.2%of
rdg+1digit)
(2%of
rdg+5digits)
Overload protection: 0.2A/250V fused (Except 2001:2A/250V) 10A range is not fused. Maximum input current :10A
RESOLUTION
(0.5%of
rdg+2digit)
4 1/2 Digit
ACCURACY
RANGE
2mA
20mA
200mA
0.1 A 1 A
(2%of
rdg+10digits)
(0.75%of
rdg+5digit)
1mA
2004
2003
10 A
10A
Indication: Average value (rms of sine wave) For 10A input jack, it is non-fused. For safety, each measuring time should be equal to or less
than 10 seconds. Time intervals should be equal to or over 15 minutes.
Model UT2000: OPERATING MANUAL
Page 8
4. AC Current
Overload protection: 0.2A/250V fused (Except 2001:2A/250V) 10A range is not fused. Maximum input: 10A Frequency range: 40-400Hz. Indication: Average value (rms of sine wave)
For 10A input jack, it is non-fused. For safety, each measuring time should be equal to or less
8
than 10 seconds. Time intervals should be equal to or over 15 minutes.
RESOLUTION
(1%of
rdg+3digits)
3 1/2 Digit
ACCURACY
RANGE
2mA
20mA
200 A
200mA
2A
10A
20 A
1 A
10 A
0.1 A
100 A
1mA
10mA 10n A
2007
2006
2002/5
2001
(1.8%of
rdg+3digits)
(3%of
rdg+7digits)
RESOLUTION
(0.8%of
rdg+10digit)
4 1/2 Digit
ACCURACY
RANGE
2mA
20mA
200mA
0.1 A 1 A
(2%of
rdg+10digits)
1mA
2004
2003
10 A
10A
Model UT2000: OPERATING MANUAL
Page 9
9
5. Resistance
RESOLUTION
3 1/2 Digit ACCURACY
2001 2002/5 2006 2007
(0.8%of rdg+1digit)
0.1 1
10
100
1K
10K
(0.8%of rdg+3digits)
(1%of rdg+2digits)
200
2K
20K
200K
2M
20M
200M
RESOLUTION
4 1/2 Digit ACCURACY
0.01
(0.2%of
rdg+5digits)
200
2K
2003 2004
(0.5%of
rdg+5digits)
(0.2%of
rdg+1digit)
(0.5%of
rdg+1digit)
[5%(of rdg-10digits)+10digits]
100K
(0.5%of
rdg+5digits)
(1%of
rdg+5digits)
RANGE
20k
200K
2M
20M
0.1 1
10
100
1k
other range: 0oCto 35oC: 0~90%. * When testing on 200M range, the display will show 1.0 if connect the two test lead together. This reading is a fixed deviation
Overload protection: 250VDC or AC rms. Open circuit voltage: 0.7V (Except 2003,2004: 3V) 200M range is 3V.
Relative Humidity: 2M range: 0oCto 35oC: 0~75%,
Model UT2000: OPERATING MANUAL
RANGE
Page 10
10
which does not affect the accuracy of the reading. If the resistor under measurement is 100M , The reading will be 101.0, the actual value should be the display value minus
1.0: 101.0 - 1.0 = 100.0 M
6. Capacitance
4 1/2Digit ACCURACY
RESOLUTION
(2.5%of rdg+3digits)
3 1/2 Digit ACCURACY
RANGE
2nF
20nF
200nF
2 F
20 F
2002/5 2006 2007
RESOLUTION
1pF
10pF
100pF
1nF
10nF
2003 2004
RANGE
20nF
200nF
2 F
20 F
1pF
10pF
100pF
1nF
(2.5%of rdg+10digits)
(especially large capacitor) to the meter. The reading need some times to stabilize or return to zero when changing the range.
Testing frequency: 400Hz Testing voltage: 40mV * When using the capacitance range for testing, don’t connect a resistor or a charged capacitor
Model UT2000: OPERATING MANUAL
Page 11
11
7. Temperature
Temperature sensor: International standard K type (NiCr - NiSi) probe.
8. Frequency
RESOLUTION
RANGE
(1.5%of rdg+15digits)
(0.75%of rdg+3digits)
ACCURACY
2007
-40oC~+400oC
400oC~1000oC
1oC
1oC
RESOLUTION
RANGE
(1.5%of rdg+5digits)
ACCURACY
2006
10Hz
1Hz2kHz
20kHz
RESOLUTION
RANGE
(1.5%of rdg+5digits)
ACCURACY
1Hz
20kHz
2003 2004
Input sensitivity: 100mV rms. Overload protection: 250 DC or AC rms.
Model UT2000: OPERATING MANUAL
Page 12
12
9. Diode
Overload protection: 250V DC or AC rms.
10 . Transistor hFE
11. Open circuit test
Display approx. forward voltage
of diode
All
series
RANGE
DESCRIPTIONS
MODEL
Forward DC current
approx. 1mA
Reversed DC voltage
approx. 3V
TESTING CONDITION
Overload protection: 250V DC or AC rms.
All
series
RANGE
DESCRIPTIONS
MODEL
TESTING CONDITION
Open circuit voltage
approx. 3V
If the resistance
between the two testing point less than 30 , buzzer
sounds.
12. Battery test
RANGE
DESCRIPTIONS
MODEL
INTERNAL
RESISTANCE
2001
1.5V 9V
Display the value
of voltage between
the cathode and
anode of battery.
30
1.8k
All
series
RANGE
DESCRIPTIONS
MODEL
TESTING CONDITION
hFE
Display approx.
hFE value
(0~1000) of
transistor under test
Base current approx. 10 A,Vce approx.3V.
Model UT2000: OPERATING MANUAL
Page 13
13
13. Alarm
2007
2003
2002/5
2006
2001
2004
ALARM SOUNDS
Mis-
operation
Switch position
T est leads position
A or 10A
A,mA,2A
10A
10A
A
Over
range
Display >19999*
V, ,Hz, ,
BATT
The testing value for range is meaningless.
14. Square wave output Frequency approx. 50Hz, range Vp-p 2V (only for UT2001).
OPERATING INSTRUCTIONS
1. Check the 9V battery by press down the ON
- OFF SWITCH. If the battery is weak,”
“ sign will appear on left bottom side of the display. The sign next to the test lest lead jacks is for warning that the input voltage or current should not exceed the indicated values. This is used to prevent the damage to the internal circuitry. The function switch should be set to the range which you want to test before operation.
2. Except capacitance, transistor hFE and temperature which use the special test jack, the input terminal for all other range should be “V/ ”, “COM” being the input earth terminal.
Model UT2000: OPERATING MANUAL
Page 14
14
3. Input Current Jack: There is a 5x20mm fuse inside the “A” input jack. Over range measurement will burn the fuse and replacement should use the fuse of the same specification. “10A” input jack does not have fuse protection.
4. The meter will show the room temperature if the temperature probe is not connect to the object under testing. The meter will only shows the object temperature if the probe is connect to it.
5. When the meter is not in use for more than 15 minutes, the power will be automatically turned off. Just release & press the “POWER” button, the meter will be turned on again.
6. Battery or fuse replacement should only be done after the test leads have been disconnected and power is off, to open the battery door, see the following diagram.
Model UT2000: OPERATING MANUAL
Page 15
15
7. The choice of LCD display angle: see the following diagram.
SAFETY RULES
The Digit Multimeter is a precise electronic device. Do not tamper with the circuit and pay attention to the following:
1. The Meter complies with IEC1010-1 pollution Degree2 CAT I 1000V, CAT II 600V over voltage standards. Use the Meter only as specified in this manual, otherwise the protection provided by the Meter may be impaired.
2. CAT I- For signal level, telecommunication, electronic with small transient over voltage.
3. CAT II- For local level, appliances, main wall outlets, portable equipment.
4. The meter is designed to withstand the stated Max. Voltages. If it’s not possible to exclude without doubts that impulses, transients, disturbance or for other reasons, these voltages are exceeded a suitable prescale (10:1) must be used.
Model UT2000: OPERATING MANUAL
Page 16
5. Do not operate the Meter before the cabinet has been closed and screwed safely as terminal can carry voltage.
16
6. Make sure before each measurement the Meter is set to the suitable range.
7. Before using the Meter, please inspect the cabinet and test leads for damaged insulation or exposed metal.
8. Connect the red and black test lead to the correct measuring input jack properly.
9. Do not input values over the maximum range of each measurement to avoid damages of the Meter.
10. Do not turn the rotary function switch during Voltage or Current measurement, otherwise the Meter could be damaged.
11. Make sure to use new fuses with proper rating to replace the bad fuses.
12. To avoid electric shock or damages, do not apply more than 1000V between the “COM”
terminal and “
” earth ground.
13. Use caution when working with Voltages above 60V (DC) or 30Vrms (AC). These Voltages pose shock hazard.
14. Replace the battery as soon as the battery indicator “
” appears. With a low battery, the Meter might produce false reading that can lead to electric shock and personal injury .
15. Turn off the Meter once finished measuring. Fetch out the battery when the meter will not be used for long period.
Model UT2000: OPERATING MANUAL
Page 17
17
16. Do not operate the Meter under adverse environmental condition including high temperature and especially humid area as the Meter’s function may be ineffective after moisturizing.
17. To avoid damages and dangerous, do not change the circuit on your own.
18. Periodically wipe the cabinet with a damp cloth and mid detergent. Do not use abrasives or solvents.
19. Dispose the used battery proper.
20. The Meter is suitable for indoor use only.
Model UT2000: OPERATING MANUAL
Page 18
18
Maintenance(1)
This section provides basic maintenance information including battery and fuse replacement instruction.
Warning Do not attempt to repair or service your Meter unless you are qualified to do so and have the relevant calibration, performance test, and service information. To avoid electrical shock or damage to the Meter, do not get water inside the case.
1.General Service
l Periodically wipe the case with a damp cloth and mild detergent. Do not use abrasives or solvents. l To clean the terminals with cotton bar with detergent, as dirt or moisture in the terminals can affect readings. l Turn the Meter power off when it is not in use and take out the battery when not using for a long time. l Do not store the Meter in a place of humidity , high temperature, explosive, inflammable and strong magnetic field.
2.Replacing the Battery
Warning To avoid false readings, which could lead to possible electric shock or personal injury, replace the battery
as soon as the battery indicator “
”appears.
To replace the battery:
1. Disconnect the connection between the testing leads and the circuit under test, and remove the testing leads away from the input terminals of the Meter.
2. Turn the Meter power off.
3. Remove the screw from the battery and fuse compartment, and separate the battery and fuse compartment from the case bottom.
Model UT2000: OPERATING MANUAL
Page 19
~ END ~
This operating manual is subject to change without notice.
19
4. Remove the battery from the battery compartment.
5. Replace the battery with a new 9V battery (NEDA 1604 or 6F22 or 006P).
6. Rejoin the case bottom and battery and fuse compartment, and reinstall the screw.
Maintenance(2)
3.Replacing the Fuses
Warning To avoid electrical shock or arc blast, or personal injury or damage to the Meter, use specified fuses ONLY in accordance with the following procedure.
To replace the Meter’s fuse:
1. Disconnect the connection between the testing leads and the circuit under test, and remove the testing leads away from the input terminals of the Meter.
2. Turn the Meter power off.
3. Remove the screw from the battery and fuse compartment, and separate the battery and fuse compartment from the case bottom.
4. Remove the fuse by gently prying one end loose, and then take out the fuse from its bracket.
5. Install ONLY replacement fuses with the identical type and specification as follows and make sure the fuse is fixed firmly in the bracket. Model UT2002, UT2003, UT2004, UT2005, UT2006 and UT2007: 0.2A, 250V, fast type, 5x20mm. Model UT2001: 2A, 250V, fast type, 5x20mm.
6. Rejoin the case bottom and battery and fuse compartment, and reinstall the screw.
Replacement of the fuses is seldom required. Burning of a fuse always results from improper operation.
Model UT2000: OPERATING MANUAL
Page 20
Copyright 2001 Uni-Trend International Limited.
All rights reserved.
20
Manufacturer: UNI-TREND TECHNOLOGY(DONG GUAN)LIMITED Address: Dong Fang Da Dao, Bei Shan Dong Fang Industrial Development District, Hu Men Town, Dong Guan City, Guang Dong Province, China Headquarters: Uni-Trend International Limited Address: Rm901, 9/F, Nanyang Plaza 57 Hung To Road Kwun Tong Kowloon, Hong Kong Tel: (852) 2950 9168 Fax: (852) 2950 9303 Email: info@uni-trend.com http://www.uni-trend.com