UNI-T ut151a, ut151b, ut151c, ut151d, ut151e Operating Manual

P/N:110401102629
TABLE OF CONTENTS
Item Page I. Characteristics---------------------------------------------------------------------------------------- --------------3 II. Technical data------------------------------------------------------------------------------------------------------ 3 III. Operating method ------------------------------------------------------------------------------------------------9 IV. Maintenance of instrument ---------------------------------------------------------------------- ------------15 V. Accessories -------------------------------------------------------------------------------------------------------17
OVERVIEW
3 1/2-bit DDM in the brand-new UT151 series is a kind of hand-held 3 1/2-bit digital universal meter with stable performance and high reliability. The large-scale integrated circuits and dual integral A/D transducer are the core design of circuits of the overall instrument and overload protection with full range of functions is also provided. It can be used to measure the DC voltage and AC voltage, electric current, resistance, capacitance, diode, triode, temperature, frequency, battery and the circuit break-make. It's your ideal tool.
SAFETY RULES AND PRECAUTIONS
This instrument is designed and manufactured in strict accordance with safety requirements of electronic measuring instrument in GB4793 as well as safety standards of IEC61010-1 and IEC100-2-O32. It also complies with the safety standards of double insulation, over voltage CAT I 1000V, CAT II 600V and pollution class 2.F ollow t he i nstructions o f t his M anual, or the protection of the instrument will be damaged.
Ensure that it cannot be operated before the back cover is put in the place, otherwise there is a risk of electric shock.
The range switch shall be installed at the correct position.
Check and ensure the insulation layer of the pen is in good condition without damage and broken line.
The red and black pens should be inserted in the hole which meet the measurement requirement; ensure well contact.
The input signal is not allowed to exceed the limit value to avoid electric shock and damage.
Changing range switch shift is strictly prohibited during the measurement of voltage and current in order to avoid damaging the instrument.
If you want to change the bad fuse and you must use the one of the same type. In order to avoid electric shock, the potential difference between common port COM a nd
the earth is not exceeded 1000V.
When the voltage to be measured is more than 60 DC or 30 Vrms AC, you must be careful
to avoid electric shock.
When the LCD shows the symbol of , you must change the battery in time to ensure the
measuring accuracy.
Turn off the power after finishing the measurement. If you do not use it for a long time, take
out the battery.
Don't operate this instrument under high temperature or high humidity. Especially, it cannot be stored in damp environment. If the instrument is affected with damp, its performance will be damaged.
Don't change the circuits of the instrument at random, otherwise, you will damage the instrument and be in danger.
Maintenance: Clean the shell of the instrument with wet cloth or mild cleaning agent rather than abrasive objects or solvent.
Symbols
Lower battery level
Buzzer
Fuse
Grounding
AC
Double insulation
Warning
DC
Diode
I.CHARACTERISTICS
1. There are 30 ranges for function selection.
2. LCD display with visible zone 63 29mm
3. Over-range indication 1 .
4. The maximum value displayed of 1999(three and half digits)
5. Overload protection for the full range.
6. Automatic power cut-off
7. Temperature range Working temperature: 0 ~40 (32 ~104 ) Storage temperature:
-10 ~50 (14 ~122 )
8. Indication of lower battery level: The upper-left corner of LCD will show the symbol of
9. The function to keep number of all shifts.
10. Physical dimensions: 186mm x 91mm x 39mm
11. Weight: Approximate 300g (instrument +protection sleeve + bracket+ battery, excluding pens)
II.TECHNICAL DATA
Accuracy: (a% reading +digit); the guarantee period is 1 year. Ambient temperature: 23 5 Relative humidity: <75%
2-1. DC voltage
Range
200mV
2 V
20 V
200 V
1000 V
nput impedance: 10MÙ for all ranges Overloading protection: For 200mV range, 250V, DC or AC effective value. For other ranges, 750Vrms or 1000Vp-p peak value.
2-2. AC voltage
Range
2 V
20 V 200 V 750 V
Input impendence: 10 MÙ for all ranges Frequency range: 45Hz-400Hz Overloading protection: 750 Vrms or 1000Vp-p peak Display: Mean value (sine-wave effective value)
Resolution
100µV
1 mV
10mV
100mV
1 V
Resolution
1 mV
10mV
100mV
1 V
UT151A UT151B
UT151A UT151B
Accuracy
UT151C
(0.5%+1)
(0.8 %+2)
Accuracy
UT151C
(0.8%+3)
(1.5%+5)
UT151D UT151E
UT151D UT151E