This analyser extracts combustion gases that may be toxic in relatively low
concentrations. These gases are exhausted from the side of the instrument. This
instrument must only be used in well ventilated locations. It must only be used by
trained and competent persons after due consideration of all the potential hazards.
Protection Against Electric Shock (in accordance with EN 61010-1 :
1993)
This instrument is designated as Class III equipment and should only be connected to
SELV circuits. The battery charger is designated as:
Class II equipment
Installation category II
Pollution degree 2
Indoor use only
Altitude to 2000m
Ambient temperature 0°C-40°C
Maximum relative humidity 80% for temperatures up to 31°C decreasing linearly to
50%RH at 40°C
Mains supply fluctuations not to exceed 10% of the nominal voltage.
3. FIRST TIME USE
Charge the battery for 12 hours. Following this, an overnight charge should be
sufficient for an average 8 hour day. See Main Parameter displays for Battery
Indicator.
The KANE900 Plus has a rechargeable lead acid battery which uses a different
charger than other Kane analysers. Ensure the correct charger is used or damage
may occur to the instrument.
Check that you have all the items you have ordered.
Take time to read this manual fully. Be aware that the analyser configuration that you have purchased may not support all the features detailed in this manual.
When using the analyser for the first time you will need to choose from: Language selection
Calibration countdown time
CO gas alarm
NOx percentage for calculation
Time and Date
Printed header name and telephone number
The SET UP MENU (Section 5.2.5) gives details of how to change the above settings.
KANE900 Plus Operators Manual Page 5
4. NORMAL START UP SEQUENCE
4.1 Every Time You Use The Analyser
BEFORE SWITCH-ON CHECK THAT:
the particle filter is not dirty
the water trap and probe line are empty of water
all hose connections, etc, are properly made
the probe is sampling CLEAN AMBIENT air
the water trap is correctly fitted and the instrument upright
the flue temperature is connected
Switch ON the instrument by pressing
4.2 Automatic Calibration
During this sequence the analyser pumps fresh air into the sensors to allow toxic
sensors (if fitted) to be set to zero and the Oxygen sensor to be set to 20.9 %.
After switch-on the analyser will briefly display header information :-
Kane International
(44)-1707-375550
And then show the countdown screen :-
ZERO CAL
Time : 180
FRESH AIR PURGE
The calibration time will count down in seconds to zero. Calibration time may be
changed from 2 to 6 minutes. See Set-Up menu section 5.2.5.
Note ! Three minutes is recommended to allow the sensors to stabilise fully. Anything
less than this may result in drift of the toxic and oxygen sensors in clean
ambient air.
To obtain the quoted specification an instrument should be calibrated with clean
ambient air at standard temperature and pressure (STP).
KANE900 Plus Operators Manual Page 6
Once the time has reached zero an audible beep will be heard and will show the
selected fuel on the following display:-
NATURAL GAS
∗
PRESS -MENU- KEY
Press
This zeros the toxic sensor and sets Oxygen to 20.9%. The next screen is the MAIN DISPLAY of the analyser:-
∗
NETT C 0.0
O2 % 20.9
CO ppm 0000
EFF (G) % 0.0
Use andto change the display.
CO2 % 0.0
FLUE C 0.0
INLT NOT FITTED
AMBIENT C 21.5
All parameters are detailed in Appendix A - MAIN DISPLAY PARAMETERS.
KANE900 Plus Operators Manual Page 7
4.3 Main Displays
The main display can be changed to show either 4 or 8 parameters at one time. Two
options are available when 4 parameters are selected.
• 4 Page Mode displays 4 lines of data in set format, each page is predefined.
• Line scroll mode allows you to customise the display to show the data you
require.
•8 Page Mode displays 8 parameters on 4 lines in set format, the bottom two can be
changed.
Changing between the different modes is detailed in Display Menu Section 5.2.4.
4.3.1 4 Page Mode
Use the andkeys to change the information that is displayed on the
screen. The following pages are available.
NATURAL GAS
DATE 07-08-96
TIME 12:31:35
BATTERY % 54
This screen only displayed on an
analyser fitted with an NO sensor
NETT C 0.0
O2 % 20.9
CO ppm 0000
EFF (G) % 0.0
CO2 % 0.0
FLUE C 0.0
INLT NOT FITTED
AMBIENT C 21.5
CO/CO2 R 0.0001
P INDEX % 0.01
XAIR % 0.0
Prs mbar 0.00
NO ppm 0000
NOx ppm 0000
NOx calc% 5
O2 ref % 3.0
TIP - In 4 page mode onlyturns the backlight ON and OFF.
KANE900 Plus Operators Manual Page 8
4.3.2 Line Scroll Mode
Line scroll mode allows you to customise the display.
Use the andkeys to change the bottom line of the display. Once the
correct line is displayed pressto confirm and move the line up. Select the
next parameter and repeat until all lines display the desired parameters.
Change bottom line using
NETT C 0.0
O2 % 20.9
and
CO ppm 0000
CO2 % 0.0
O2 % . . . 20.9
to select and
move parameter up
Select next parameter.
Repeat above until
display reads desired data
CO ppm . . 0000
CO2 % . . . 0.0
CO2 % . . . 0.0
O2 % . . . 20.9
CO ppm . . 0000
CO2 % . . . 0.0
CO/CO2 R . . 0.0001
4.3.3 8 Page Mode
Displays 8 parameters on the screen at one time. Symbols used in this mode are
different from those used in 4 page and line scroll modes and are detailed in Appendix
A - MAIN DISPLAY PARAMETERS.
O2
CO
PI
λ
The bottom line of the display can be changed to display other parameters.
Use theandkeys to change this line.
KANE900 Plus Operators Manual Page 9
:20.9 %
: 0ppm
:
:
CO2
Eff
∆T
Tf
:
:
: 0C
: 21C
4.4 Sampling the Flue Gas
Once the automatic calibration procedure has been completed and the specific fuel has
been selected (See SELECT menu) the probe can be inserted into the desired
sampling point.
It is recommended that the sampling point be located at least two flue diameters
downstream of any bend and that the probe tip is in the centre of the flue. With
balanced flues and other domestic units the probe should be positioned far enough
into the flue so that no air can ‘back flush’ into the probe. This will be indicated by a
low oxygen reading and/or a low ‘Poison Index’ reading.
The probe depth stop cone provided with the instrument allows the probe to be used
in holes whose diameters range from 8 mm to 21 mm (5/16 to
The standard probe is rated at 650°C (1202oF). Temperatures of up to 1200°C
(2200°F) can be accommodated using an optional high temperature probe.
TIP To conserve battery power, switch off the pump when you are not taking a
measurement. Use the
key to turn ON and OFF the pump.
13
/16 inch).
4.5 Taking a Pressure Reading
With the optional pressure module fitted a flue draught measurement can be made at
any time.
Connect the standard probe to the pressure sensor inlet and the probe in the flue.
The pressure reading will be displayed :-
CO/CO2 R 0.0001
P INDEX % 0.01
XAIR % 0.0
Prs mbar 0.00
To perform a combustion test and display draught pressure at the same time a special
probe is required. Contact Kane International or Authorised Distributor for details.
KANE900 Plus Operators Manual Page 10
4.6 Regular Checks During Sampling
Care must be taken at all times not to exceed the analysers operating specifications, in
particular ensure the following :-
• Do not exceed the maximum temperature of the flue probe.
• The analyser internal temperature does not exceed normal operating range, typically 0-50°C.
• DO NOT PLACE THE INSTRUMENT ON A HOT SURFACE.
• The water trap is vertical at all times. Water condenses in the probe line and can
quickly fill the water trap when the probe is moved. Take care and watch the water
trap closely.
• The in-line particle filter is clean and does not become blocked.
4.7 Normal Shutdown Sequence
DO THIS EVERY TIME YOU USE THE ANALYSER
Remove the probe from the flue - TAKE CARE ! THE PROBE WILL BE HOT -
and allow it to cool naturally. Do not immerse the probe in water as this will be drawn
into the analyser and damage the pump and sensors.
Once the probe is removed from the flue press and the analyser will count
down from 30 to switch off.
OFF 30
MENU TO ESCAPE
If you have not finished but press by mistake, you can pressto
return to normal operation and not switch OFF.
KANE900 Plus Operators Manual Page 11
4.8 Electromagnetic Compatibility
The European Council Directive 89/336/EEC requires that electronic equipment does
not generate electromagnetic disturbances that exceed defined levels and has an
adequate level of immunity to enable it to be operated as intended. The specific
standards applicable to this product are detailed in the appendices.
Since there are many electrical products in use that pre-date this Directive and may
emit electromagnetic radiation in excess of the standards defined in the Directive
there may be occasions where it would be appropriate to check the analyser prior to
use.
The following procedure should be adopted:
Go through the normal start up sequence in the location where the equipment
is to be used.
Switch on all localised electrical equipment that might be capable of causing
interference.
Check that all readings are as expected. (A level of disturbance in the
readings is acceptable). If not adjust the position of the instrument to
minimise interference or switch off, if possible, the offending equipment for
the duration of the test.
N.B. Maximum cable lengths must be less than 3 metres.
At the time of writing this manual (March 2006) Kane International Ltd is not aware
of any field based situation where such interference has ever occurred and this advice
is only given to satisfy the requirements of the Directive.
KANE900 Plus Operators Manual Page 12
5. MOVING THROUGH THE MENUS
5.1 Basic Operation
From the MAIN DISPLAY
Press
to access the MAIN MENU
Press
andto move
cursor up and down
Press to access selected
Menu
NETT C 0.0
O2 % 20.9
CO ppm 0000
EFF (G) % 0.0
MAIN MENU
1. SELECT 3. DISPLAY
2. UNITS 4. SETUP
MAIN MENU
1. SELECT 3. DISPLAY
2. UNITS 4. SETUP
MAIN MENU
1. SELECT 3. DISPLAY
2. UNITS 4. SETUP
Press to select parameter
Use andto change
setting i.e. fuel selected
Press to enter value and
move to next parameter
Press
to save settings and return to the
MAIN MENU
FUEL
O2 Ref
SMOKE
RESET
FUEL
O2 Ref
SMOKE
RESET
FUEL
O2 Ref
SMOKE
RESET
: LIGHT OIL
: OFF
: OFF
: NO
: NATURAL GAS
: OFF
: OFF
: NO
: LIGHT OIL
: OFF
: OFF
: NO
MAIN MENU
1. SELECT 3. DISPLAY
2. UNITS 4. SETUP
Pressto return to the MAIN DISPLAY
KANE900 Plus Operators Manual Page 13
5.2 Menu Options and Settings
5.2.1 Main Menu
The MAIN MENU consists of 4 sub menus which are shown below and detailed on
the following pages.
MAIN MENU
1. SELECT 3. DISPLAY
2. UNITS 4. SETUP
All sub-menus are accessed usingand exited using
The andkeys move the cursor within a menu and allow
parameters to be changed.
TIP Holding down one of these keys scrolls through the data quicker.
5.2.2 Select Menu
FUEL
O2 Ref
SMOKE
RESET
This menu allows selections to be made for the parameters detailed below.
FUEL :Select the fuel being used by the boiler from either a standard fuel
stored in the analyser or by entering the user fuel. Once the correct fuel
has been selected pressto view the fuel constants.
: NATURAL GAS
: OFF
: OFF
: NO
NATURAL GAS
K1g
K_2
K_4
Calculation of fuel constants are detailed in the Appendix. Fuel constants will have to
be calculated before a user fuel can be entered.
: 0.350
: 11.89
: 32
K1n
K_3
O2r
: 0.390
: 9.83
: 3.0
KANE900 Plus Operators Manual Page 14
To enter the user fuel select ‘User Fuel’ and Press
USER FUEL
K1g
K_2
K_4
Use andto select the correct value.
: 0.000
: 0.00
: 00
K1n
K_3
O2r
: 0.000
: 0.00
: 00
USER FUEL
K1g
K_2
K_4
:0.350
: 0.00
: 0
K1n
K_3
O2r
: 0.000
: 0.00
: 00
Useto move to the next parameter, repeat above until all
parameters are correct. Press
O2 Ref :Toxic gas measurements can be referenced to defined oxygen levels.
Reference values can be set from 1-20%, to AUTO or more normally
to the default value - OFF. Setting to AUTO uses the figure in the
FUEL constants data.
Oxygen referencing is required by some regulations such as TA-LUFT.
If a reference value is selected then toxic gas measurements will be
displayed with the symbol (n) attached to the reading. i.e. CO(n)
What does Oxygen reference mean ?
If 3 % O2 reference is selected and 5 % O2 is measured in the flue then
toxic gas values will be recalculated as if 3 % were measured. The
equation for referencing is detailed in the Appendix.
Oxygen referencing prevents false readings being submitted, e.g.
allowing more air into the boiler will increase the oxygen level in the
flue and hence dilute any toxic gas reading. Oxygen referencing gives
readings as if they were undiluted.
SMOKE: Allows the user to enter a smoke test number from 0-9. This value will
be printed on the standard printout. Default value is OFF.
RESET : Allows theuser set the Oxygen to 20.9% and zero the toxic sensors
without turning the analyser off.
to return to SELECT menu.
KANE900 Plus Operators Manual Page 15
Selecting YES andwill display the following screen.
RESET SENSORS
O2 % : 20.9 CO & NO = 0
PRESS ENTER
MENU TO ESCAPE
After pressingthe analyser will count down for 5 seconds and
then return to the main display.
WARNING : The sensors must only be reset if you are sure they
have been sampling fresh air for at least 3 minutes.
Errors in measurement will occur if the sensors are
reset during or just after sampling.
5.2.3 Units Menu
TEMP
GAS
PRESS.
EFF.
Allows all displayed units to be changed.
TEMP : Choose from Centigrade, °C, or Fahrenheit, °F.
GAS : Changes the toxic gas measurement units. Select from volumetric
readings, parts per million (ppm) or mass flow reading milligrams per
cubic meter (mg/m3) .
PRESS. :Flue draught can be displayed in millibar (mbar), hectaPascals (hPa),
millimeters water gauge (mmWG) or inches water gauge (in WG).
EFF. :Efficiency can be selected for gross or net values. Gross efficiency
assumes latent heat of vaporisation is lost in the boiler and hence will
be lower than net efficiency. For natural gas the difference will be
approximately 11%.
Efficiency is displayed as EFF (G) or EFF (N) respectively. Should
the instrument detect that a condensing boiler is under test then it
automatically switches to a third mode that is displayed as EFF (C).
: C
: ppm
: mbar
: GROSS
KANE900 Plus Operators Manual Page 16
5.2.4 Display Menu
LIGHT
MODE
CONTRAST
Allows the configuration of the display to be changed.
LIGHT : Choose from ON or OFF.
MODE : Select 4 or 8 Page Mode or Line Scroll Mode as detailed in section 4.3
Main Displays.
CONTRAST :The contrast is set to a DEFAULT value or can be adjusted
↑ LIGHTER or ↓ DARKER. Use theandkeys to
adjust.
5.2.5. Set-Up Menu
The set up menu allows the following parameters to be set / altered.
• Language.
• Automatic calibration time
• CO gas alarm
• NOx percentage for calculation
• Date and time
• Printout Header
: OFF
: 8-PAGE
: DEFAULT
LANG
CO AL
CALEN
: ENGLISH
ARM : 400
DAR
ZERO : 3
NOx% : 5
HEADER
Parameter Description Settings
LANG :
ZERO :
CO ALARM :
Changes the analysers displayed and printed
language.
Allows setting of the Autocalibration time in
minutes. Care must be taken when changing this
parameter as sensors may drift from zero if too
short a time is used. Kane International advise 3
minute countdown.
Allows an alarm level to be set on for the CO
reading. This is set as a default at 1000 ppm.
ENGLISH
SPANISH
DUTCH
FRENCH
ITALIAN
2-6 minutes
OFF
0-4000 ppm
KANE900 Plus Operators Manual Page 17
Once an alarm has been exceeded the display will flash every two minutes warning
the user of an alarm state and display the gas concentration. A similar display will be
shown during a RECHARGE BATTERY and PUMP OFF alarms.
-- -- -- -- -- -- -- -- -- -- -- -- -- -CO ALARM
1010 ppm
-- -- -- -- -- -- -- -- -- -- -- -- -- --
NO REF:
CALENDAR:
Displayed on the Nitric Oxide unit only. Allows
the percentage P in the following calculation to be
set. The default value set is 5%. Note the
percentage allows for NO2 in a typical boiler.
NOx = NO + P% NO
Allows the user to change the date and time. (24
hour clock).
The following screen will be shown once the
parameter is entered:
OFF
1-9 %
TIME :
FORMAT :
DATE :
FORMAT :
To change the time position the cursor on Time and press. The cursor will
now be to the left of the 13 as shown below:
Changes the date format for display and printing.
TIME :
FORMAT :
DATE :
Usingandscroll through the setting options i.e. 0-23.
Once the correct hour is set pressto move to the next parameter, the cursor
will move to the left of minutes (53). Move to each parameter until the correct time is
set.
hh
: mm : ss
13
: 53 : 26
dd
: mm : yy
03
: 01 : 96
dd : mm : yy
yy : mm : dd
mm : dd : yy
hh
: mm : ss
13
: 53 : 26
dd
: mm : yy
03
: 01 : 96
Pressingafter setting the seconds will return the cursor to the left of the
screen.
Format and Date are set in a similar manner.
KANE900 Plus Operators Manual Page 18
Header :
Allows two lines of 20 characters to be
programmed into the analyser. The header appears
on the top of the standard printout. This can be used
to print your company name and/or phone number.
Name/Phone
K
ane International
(44)-1707-375550
‘LEFT’ USE STORE KEY
The screen above shows the standard header setting with the cursor now shown
underlining the K in Kane. By usingandany letter or number can be
chosen.
Once the correct character is displayed, useto move right to the next. Move
along until all characters spell the desired name or phone number. If you need to go
back and change a character useto move left.
Pressto return to the SET UP menu.
KANE900 Plus Operators Manual Page 19
6. PRINTING INFORMATION
Supplied as accessories for the KANE900 Plus are an infra-red thermal printer or a
dot matrix serial printer. Read the manual supplied with each printer prior to
operation. Connections to the KANE900 Plus are detailed below :
•Infra-red thermal printer - this does not require a cable to transmit the data but
uses an infra-red (IR) link similar to a TV remote control. The IR emitter is
positioned on the top of the KANE900 Plus and the bottom of the printer. Ensure
they are pointing at each other and within 300 mm, with no obstructions in the
way. Data may be lost if transmission is interupted. Keep the KANE900 Plus
pointing at the printer until the printout has finished.
•Dot matrix serial printer - requires the supplied serial cable to transmit data.
Connect the cable to the 8 pin DIN socket on the top of the KANE900 Plus and the
25 pin D-connector on the printer.
Data can either be printed from a ‘live’ test or from stored data. Printing of stored data
is detailed in STORING AND RETREIVING DATA.
6.1 Printing a ‘Live’ Test
During a combustion test the KANE900 Plus will print data on request. With the
analyser showing the MAIN DISPLAY pressand current data will be sent to
the printer.
The display will show the following until data transmission is complete.
∗∗∗∗∗
Printing
∗∗∗∗∗
KANE900 Plus Operators Manual Page 20
6.2 Standard Printout
The standard printout is:-
KANE900 Plus Operators Manual Page 21
7. STORING AND RETRIEVING DATA
The KANE900 Plus can store combustion tests. Once stored, the data can be viewed
on the display or downloaded to a PC or printer.
7.1 Storing a ‘Live’ Test
While performing a test and viewing the data on the MAIN display access the STORE
menu as follows :-
MODE
Press
to access the STORE MENU
Mode: Select from the following :-
Location: Automaticaly allocates a location in the memory of the instrument for
To store a test set MODE to STORE and press. The current readings will
be stored in the analysers memory.
Tip : Make a note of the location number for your particular test as it may be
LOCATION
PRESS ‘STORE’ TO LOG
• STORE - Allows data to be stored in memory.
• VIEW / PRINT - Stored data can be viewed or printed.
• DELETE - Clears all data in memory.
the next test. On the display shown above the next location will be 3.
useful when downloading or printing.
7.2 Viewing and Printing a ‘Stored’ Test
Multiple tests can be printed easily with the KANE900 Plus.
Select PRINT under MODE in the STORE menu. This feature is in addition to the
VIEW/PRINT, STORE and DELETE options.
STORE MENU
: STORE
: 3
MODE
Press
to access the STORE
LOCATION
MENU
The cursor will move to the first number, use the
location and start printing.
Press
print.
KANE900 Plus Operators Manual Page 22
to move the cursor to the second number, select the last location to
STORE MENU
: PRINT
: 1 TO10
PRESS ‘PRINT’
andto select the
To print the data press. In the screen shown above locations 1 to 10 will be
printed.
During printing the following will be shown.
PRINT TESTS
1 to 10
PRINTING TEST 1
NOTE While the display above is shown (i.e. the instrument is printing a test) the
keypad is disabled. To exit from printing wait until the current test has finished and
the display below is shown:
PRINT TESTS
Pressto exit the print routine. The
instrument will return to main display
1 to 10
PLEASE WAIT
MENU TO ESCAPE
7.3 Deleting Data
To delete the data in stored memory pressto obtain the STORE MENU (as
above) :-
Pressto access the STORE
MENU
Pressto access delete data
screen
Press
screen.
to delete data in memory, pressto exit delete data
STORE MENU
MODE
LOCATION
: DELETE
: 3
PRESS ‘ENTER’ TO DELETE
ENTER to ERASE DATA
MENU to ESCAPE
KANE900 Plus Operators Manual Page 23
8. AVERAGE OF THREE
8.1 Storing
Ensure the instrument is switched on and in the main screen. In the 4-page mode
below.
When you are ready to store the first reading, go into the STORE menu, use the
andto find the “Ave Store” item. Press . The first sample
has been stored.
The screen appears to go back to the main screen but there is a tiny difference: the
right-most column has just one ‘+’ there. The line that it is on tells you which sample
you have got to in the obvious way. Using the and keys, you can
look at all the usual screens. When you get to the screen with the fuel you will notice
that it is note there. In its place is a countdown (in seconds). (It started at 120
seconds. While it is active another sample cannot be stored. This is there to ensure
users cannot take samples any quicker than once every two minutes.
By and by, the countdown will countdown to zero. Then there is a beep and the
countdown line becomes a message saying you can press when you are
ready to store the next sample. In addition, the ‘+’ is now a ’*’. This is visible even
if the countdown line is not and is there to indicate that the instrument is ready to
accept the next sample. Once you have done this, you have stored the second sample.
Repeat for the third and final sample.
Then the instrument returns to its previous state.
It has stored the three samples together with the average thereof in EEPROM, so it is
available again after switching off and on. Only one such set can be stored at a time.
8.2 Viewing
With the instrument switched on and in the main screen, enter the STORE menu.
Find the “Ave view” item and select if by pressing. You can now look at
the first sample using the andkeys. The ‘*’s indicate that you are in
the Average of Three mode and by the position of the ‘*’ which sample you are
currently viewing. Press to pass to the next sample. Eventually you will be
looking at the averages. This is indicated by the bottom ‘*’ being shown. Press
a final time to exit from that mode. (To exit from this mode early, press
, then , then .
8.3 Printing
KANE900 Plus Operators Manual Page 24
Get into the average of three viewing mode, “Ave View” as above. Offer a printer to
the instrument in the usual way. Press
flagged as NOT FITTED.
. The print out omits any reading
KANE900 Plus Operators Manual Page 25
9. MAINTENANCE
9.1 Emptying and Cleaning the In-line Water Trap
The in-line water trap should be checked and emptied on a regular basis. Water
vapour will condense and gather in the probe line. This may move suddenly to the
trap when the probe is moved. Care should be taken at all times.
Emptying of the water trap is detailed below :-
Carefully remove the end cap from the in-line housing. Dispose of the condensate in a
suitable drain, care must be taken as it could be acidic. If condensate spills onto the
skin or clothing, clean off immediately using fresh water, seek medical advice if
problems occur.
9.2 Changing the Particle Filter
This is a very important part of the analyser and should be changed regularly. It
prevents dust and dirty particles entering the pump and sensors and hence causing
damage. The filter MUST be changed when it is discoloured.
Remove the end cap from the in-line filter housing. Carefully remove the paper filter
element and dispose of it. Clean the inside of the filter housing with a suitable soft
cloth. Insert a new filter element onto the spigot in the filter housing and carefully
replace the end cap.
KANE900 Plus Operators Manual Page 26
10. PROBLEM SOLVING
The following is a list of problems that may occur on the instrument through its
operating life. If the cause of the fault is not easy to identify then we advise you
contact Kane International Service Department or an International Distributor for
expert advice.
Fault symptom Causes
• Oxygen too high
• CO2 too low
• Oxygen Error (FAULT)
• Toxic sensor Error (FAULT)
• Analyser not holding charge
• Analyser not charging
• Analyser does not respond to flue
gas
• Flue temperature readings erratic • Temperature plug reversed in socket.
• Analyser automatically switches
off in operation.
• Display shows dark lines and no
response from ON/OFF key.
11. ANNUAL RE-CALIBRATION
• Air leaking into probe, tubing, water trap,
connectors or internal to instrument.
• Oxygen cell needs replacing.
• Calibration time set too short and
instrument not allowed to stabilise
• Instrument has been stored in a cold
environment and is not at normal working
temperature.
• Oxygen cell or toxic sensors needs
replacing.
• Battery exhausted.
• AC charger not giving correct output.
• Fuse blown in charger plug.
• Particle filter blocked.
• Probe or tubing blocked.
• Pump not working or damaged with
contaminants.
• Probe connected to pressure connector.
• Faulty connection or break in cable or plug.
• Battery below alarm level.
• Ambient temperature above 50°C.
• Battery quickly discharging and is faulty.
• Fault has occurred on the instrument
electronics and requires resetting. Contact
Kane International or Distributor.
Whilst the sensors have an expected life of more than two years in normal use it is
recommended that the analyser is re-calibrated at least annually. This is so that long
term drift on the sensors and electronics can be eliminated. Local regulation may
require more frequent re-calibration and users should check with appropriate
authorities to ensure they comply with relevant guidelines.
KANE900 Plus Operators Manual Page 27
12. PRODUCT SPECIFICATION
Parameter Resolution Accuracy Range
2
0.1°C/F
0.1°C/F
0.1%
1ppm,mg/m
1ppm,mg/m
1ppm,mg/m
1ppm,mg/m
0.1mbar
0.1%
0.1%
0.1%
0.1%
1.0°C/F
0.0001
0.01%
Temp Measurement
Flue Temperature
Inlet Temperature
Gas Measurement
Oxygen
Carbon Monoxide
Carbon Monoxide, H
compensated
Nitric Oxide
(optional)
Nitrogen Dioxide
(optional)
Pressure (optional)
Carbon Dioxide
*2
Losses
Efficiency
Excess Air
Temp (Nett)
CO/CO
ratio
2
Poison Index *2
*2
*2
*2
*2
*2
Pre-programmed Fuels
Dimensions
Weight
Handset
Probe
Ambient Operating Range
Power Supply (battery charger)
Battery Life
*1
Using dry gases at STP
*2
+2.0°C +0.3% reading
+1°C +0.3% reading
0.2%*1
+
3
20ppm <400ppm*1
+
5% >400ppm
+
3
20ppm <400ppm*1
+
5% <5000ppm
+
10% >5000ppm
+
3
5ppm <100ppm*1
+
5% >100ppm
+
3
5% full scale
+
5.0% full scale
+
0.3% reading
+
1.0% reading
+
1.0% reading
+
0.2%
+
2°C +0.3% reading
+
0.0001
+
0.01
+
0-1200°C/32 -2200°F
with suitable probe
0-50°C/32-122°F
0-25%
0-4000ppm, 0-5000mg/m
0-10,000ppm, 0-12,000mg/m
0-5000ppm, 0-6700mg/m
100ppm, 200mg/m
150 mbar
0-99.9%
0-99.9%
0-99.9%
0-2885.0%
0-1200°C/32-2200°F
0-0.9999
0-99.99
3
Natural gas, Town gas, Gascor, Light Oil, Heavy Oil,
Propane, Butane, Anthracite, Coke, Coal, Kinsale Gas.
1kg
220mm x 55mm x 120mm
L240mm x Dia8mm with 285mm long stainless steel shaft,
type K thermocouple and 1.5m long neoprene hose
-5°C to +50°C/10% to 90% RH non condensing
Input: 110Vac/220 Vac nominal
Output: 12 Vac off load
>8 hours from full charge
Calculated
3
3
3
KANE900 Plus Operators Manual Page 28
APPENDICES
A - Main Display Parameters
The parameters and their meanings are detailed as follows : -
DATE : Analyser date. See Set-Up menu section 5.2.5 to change.
TIME : Analyser time. Use Set-Up menu section 5.2.5 to change.
BATTERY : Displays the battery level from 0-100%. The analyser will flash
(BAT) RECHARGE BATTERY at less than 10 % of charge. With the
charger connected the display shows AC ON.
NETT : Nett temperature calculated by deducting the internal AMBIENT
temperature
(∆T) from the measured FLUE temperature. Displays in either °C (C) or
°F (F) and will display NOT FITTED (N/F) if flue probe is not
connected.
If an external INLET probe is used then INLET is deducted from
FLUE.
O2 : Oxygen reading in percentage %.
CO : Carbon Monoxide reading indicated in ppm or mg/m3. If the figures
are referenced to oxygen then the display will show CO(n). See
SELECT menu 5.2.2 for oxygen reference. The display will read
‘O2 > 20%’ if referenced values selected and instrument is in clean
ambient air.
EFF (G) :Combustion Efficiency calculation displayed in percentage. Gross G
or Net N can be set see SELECT menu 5.2.3. The calculation is
determined by fuel type see Appendix B for calculation. The
efficiency is displayed during a combustion test, ‘- -’ is displayed
while in fresh air.
CO2 :Carbon Dioxide calculation determined by the type of fuel. This
only shows a reading when a combustion test is being carried out. ‘-
-’ is displayed while in fresh air.
FLUE : Temperature measured by flue gas probe in Centigrade or
(Tf) Fahrenheit. Will show ambient temperature after fresh air calibration and NOT FITTED (N/F) or FAULT (FLT) if probe disconnected.
INLET : Temperature measured by the optional inlet air probe. This probe is
(Ti) plugged into the instrument through the RS232 socket. This figure is
used to calculate the NET temperature instead of AMBIENT when
fitted.
KANE900 Plus Operators Manual Page 29
AMBIENT : Temperature measured by the internal sensor, used in the NET
(Ta) temperature calculation if an INLET probe is not fitted.
CO/CO2 R : The CO/CO2 ratio, is the ratio of measured CO divided by
calculated CO2.
It gives an indication of the following :-
• How good a gas sample the instrument is reading.
• How clean the boiler is running.
For example : A new or clean domestic boiler will display a ratio of
less than 0.004, a unit in need of cleaning 0.004-0.008 and a unit in
need of major overhaul will show greater than 0.008.
This only shows a reading when a combustion test is being carried
out. ‘- -’ is displayed while in clean ambient air.
P INDEX : The CO/CO2 ratio expressed as a percentage %, called the ‘Poison
(PI) Index’ i.e. P INDEX % = 100 x CO/CO2. ‘- -’ is displayed while in
clean ambient air.
XAIR % : Excess air calculated from the measured oxygen and type of fuel
(λ) used. During a combustion test ‘O2 > 20%’ will be displayed while
in clean ambient air.
Prs :Flue draught pressure reading. Displayed when pressure sensor
fitted. See UNITS menu 5.2.3. for scales.
NO:Nitric Oxide reading in ppm or mg/m3. Displayed when Nitric
Oxide sensor fitted. Also displayed as NO (n) when referenced to
oxygen. The display will read ‘O2 > 20%’ if referenced values
selected and instrument is in clean ambient air.
NOx :Calculated total Nitric oxides displayed in ppm or mg/m3. Where
NOx = NO + P%NO, note P can be set from 0-9%, default = 5%.
See SELECT menu 5.2.2. Also displayed as NOx (n) referenced to
oxygen. The display will read ‘O2 > 20%’ if referenced values are
selected and instrument is sampling clean ambient air.
SO2 :Sulphur Dioxide reading in ppm or mg/m3. Displayed when Sulphur
Dioxide sensor fitted. Also displayed as SO2 (n) referenced to
oxygen. The display will read ‘O2 > 20%’ if referenced values
selected and instrument is in clean ambient air.
O2 ref % : Toxic gas measurements can be referenced to defined oxygen levels.
(O2r) See SELECT menu 5.2.2 for details.
KANE900 Plus Operators Manual Page 30
B. COMBUSTION EFFICIENCY CALCULATION
The efficiency calculation is based upon British Standard BS845.
This identifies three sources of loss associated with fuel burning:
Losses due to flue gasses: Dry Flue gas loss,
Moisture and hydrogen
Sensible heat of water vapour
Unburned gas
Losses due to refuse:Combustible in ash Combustible in riddlings
Combustible in dust
Other losses: radiation
convection
conduction
other unmeasured losses
Net efficiency calculations assume that the energy contained in the water vapour
(formed as a product of combustion and from wet fuel) is recovered and the wet loss
term is zero. Gross efficiency calculations assume that the energy contained in the
water vapour is not recovered.
Since the fuel air mixture is never consistent there is the possibility of
unburned/partially unburned fuel passing through the flue. This is represented by the
unburned carbon loss.
Losses due to combustible matter in ashes, riddlings, dust and grit, radiation,
convection and conduction are not included.
Efficiency Calculation:
Known Data - Fuel: Qgr = Gross Calorific Value (kJ/kg)
Qnet = Net Calorific Value (kJ/kg)
K1 = Constant based on Gross or Net Calorific Value:
K1g = ( 255 x %Carbon in fuel )/Qgr
K1n = ( 255 x %Carbon in fuel )/Q
net
K2 = % max theoretical CO2 (dry basis)
K3 = % Wet Loss
H
H
= % Hydrogen
2
O = % Water
2
Measured Data: Tf = Flue Temperature
Ti = Inlet Temperature
O
O
m = % Oxygen in flue gas
2
r = Oxygen reference %
2
KANE900 Plus Operators Manual Page 31
Calculated data: Tnet = Net Temperature
% CO2 content in flue gas
% Dry Flue Gas losses
% Wet losses
% Unburned carbon loss
% Efficiency
Tnet = Flue Temperature - Inlet Temperature
Dry flue gas loss % = 20.9 x K1 x (Tnet) / K2 x (20.9 - O2m)
Wet loss % = 9 x H2 + H2O / Qgr x [2488 + 2.1Tf - 4.2 Ti]
simplified = [(9 x H2 + H2O) / Qgr] x 2425 x [1 + 0.001 Tnet]
= 100 – {[20.9 x K1g x (Tnet) / K2 x (20.9 - O2m)]+
[K3 x (1 + 0.001 x Tnett)]}
Excess Air = [(20.9% / (20.9% - 02m%)) – 1] x 100%
CO2% = [(20.9 - O2m) x K2 / 20.9]
Unburned
fuel Loss % = K4 x CO / ( CO + CO2 ) Note: CO scaled in %
Where K4 = 70 for coke
= 65 for anthracite
= 63 for Bituminous coal
= 62 for coal tar fuel
= 48 for liquid petroleum fuel
= 32 for natural gas
The formula for K4 is based on the gross calorific value Qgr. To obtain the loss based
on net calorific value multiply by Qgr/Qnet. Since this loss is usually small this
conversion has been ignored. This loss is subtracted from the efficiency.
Oxygen Reference CO(n) = CO x
(20.9 - O
r)
2
(20.9 - O2m)
KANE900 Plus Operators Manual Page 32
C. CALCULATION OF FUEL DATA
For any fuel not specified by Kane International the net calorific value, gross calorific
value and composition should be obtained from the fuel supplier.
The following fuel data has been calculated with reference to the efficiency
calculation.
Example 1:
Chemical composition: C 25%
H
H
Q
Q
Max CO
K1n= (255 x % carbon in fuel) / Q= (255 x 25) / 8350 = 0.763
K1g= (255 x % carbon in fuel) / Qg (kJ/Kg) = (255 x 25) / 9300 = 0.685
K2 = Max % CO2 = 20.40
K3 = Wet Loss = [(9 x %H2 + %H2O) / 9300] x 2425
= [(9 x 3 + 50) / 9300] x 2425
= (77 / 9300) x 2425 = 20.08
K4 = 65 (an approximation for wood) *
The fuel values to program into the Analyser are as follows:
3%
2
O 50%
2
8.35 MJ/kg
net
9.3 MJ/kg *
g
20.4%
2
(kJ/Kg)
net
NATURAL GAS
K1g
K_2
K_4
* Assumed values in the absence of supplied data. See previous appendix for other
fuels.
KANE900 Plus Operators Manual Page 33
: 0.763
: 20.4
: 65
K1n
K_3
O2r
: 0.685
: 20.08
: 8.0
D. ELECTROMAGNETIC COMPATABILITY (CE)
STATEMENT
This product has been tested for
compliance with the following
generic standards:
EN 61000-6-3
EN 61000-6-1
and is certified to be compliant
Specification
EC/EMC/KI/KANE900 Plus details
the specific test configuration,
performance and conditions of use.
Please Note: Batteries used in this instrument should be disposed of in
accordance with current legislation and local guidelines.
KANE900 Plus Operators Manual Page 34
KANE900 Plus Operators Manual Page 35
Thank you for buying this
instrument.
Before use,
please register on our
website
www.kane.co.uk
KANE900 Plus Operators Manual Page 36
CONTENTS
1. ANALYSER LAYOUT AND FEATURES ................................................................................1-4
1.1 Instrument features and keypad .................................................................................1