UEi DMG150 Instruction Manual

DMG150
INSTRUCTION MANUAL
ENGLISH
DMG150
Set Leak Rate Temp Run
mbar Pa mm-Hg Microns mTorr /sec
MENU
MODE
99 Washington Street Melrose, MA 02176
Phone 781-665-1400 Toll Free 1-800-517-8431
Visit us at www.TestEquipmentDepot.com
IMPORTANT SAFETY WARNINGS
WARNING
Read entire Safety Notes section regarding potential hazard and proper instructions before using this meter. In this manual the word “WARNING” is used to indicate conditions or actions that may pose physical hazards to the user. The word “CAUTION” is used to indicate conditions or actions that may damage this instrument.
WARNING
To ensure safe operation and service of the tester, follow these instructions. Failure to observe these warnings can result in severe injury or death.
WARNING
This gauge is designed for trade professionals who are familiar with the hazards of their trade. Observe all recommended safety procedures that include proper use of personal protective equipment that includes safety
glasses and gloves.
FEATURES
• Measures in microns, pascals, Millibar, mmHg, mTorr
• Range 0 to 15,000 microns
• Maximum over pressure 700 psi
• Instant warm-up time
• Instant response time
• Backlit display
• Compact and light weight
• Oil contamination detection
• Sleep mode
• Ambient temperature measurement
• 5 units of measurement
• Real time leak measurement
• Time period leak measurement
• Sensor fail detection
GENERAL SPECIFICATIONS
• Operating Temperature: 10˚ to 122˚F (-12˚ to 50˚C)
• Backlight: Yes
• Over-range: “OL” is displayed
• Dimensions: 4.3’ X 1.3’ X 1’
• Item Weight: 2.7 oz
• Calibration: Recommended annually
• Accuracy: 0 to 9,999 ±5%rdg ±5 microns
10,000 to 15,000 ±10% rdg
OVERVIEW
DMG150
A. LCD Display
B. Power/ Backlight Button
C. Mode Button
Set Leak Rate Temp Run
mbar Pa mm-Hg Microns mTorr /sec
MENU
MODE
D. Menu/Check Button
E. Change/X Button
F. Dust Cap
G. Battery
Cover
I. Serial Number
H. Vacuum Port J. Dust
Cap Removed
OPERATING INSTRUCTIONS
Basic Operation Mode
1. Press Power button to power on the gauge.
2. Press Mode Button to change the operation to Vacuum
Reading, Temperature Reading, Leak Rate Test, Leak Test or Run mode.
Vacuum Reading Mode
1. The Vacuum reading is shown on the display with Vacuum unit.
CAUTION:
psi limit is exceeded (15,000 Micron/700 psi or unit reading of equal value).
2. User may press Mode Button to skip Vacuum Reading mode or enter Temperature Reading mode.
3. In Vacuum Reading mode, press Menu Button or Change Button to select the unit to mBar, Pa, mmHg, mTorr or Micron.
The Vacuum indicator will be shown to indicate the Vacuum unit selected.
Temperature Reading Mode
1. The Vacuum reading is shown on the display with Temperature measurement unit.
2. User may press Mode Button to skip Temperature Reading mode or enter Leak Rate Test mode.
3. In Temperature Reading mode, press Menu/Check Button or Change/X Button to select temperature measurement unit to ˚F or ˚C
Leak Rate Test Mode
1. In Leak Rate Test mode, the leak rate test time base is 1 second.
2. The indicator “Leak Rate” and “/sec” will be shown. The real time leak rate is updated on the display once per second.
3. User may press Mode Button to skip Leak Rate Test mode and enter Leak Test mode.
4. In Leak Rate Test mode, press Menu/Check Button or Change/X Button to select vacuum.
OL” will be shown if maximum Micron or
Leak Test Mode
1. In the Leak Test mode, the leak test time may be
programmed for a set duration. Only the icon “LEAK” will be shown.
2. In Leak Test mode, press the Menu Button to change the parameter to set in order. Vacuum unit, Minute, Second of time duration can be set. 3 parameters are shown on the display. The parameter flashes when it is selected to be set. If there are no changes to the 3 parameters, user may skip to step 6.
3. While the vacuum unit be set, press the Change/X Button to select the vacuum unit.
4. While the Minute of time duration can set, press the Change/X Button, to select the minute value.
5. While the Second of time duration can be set, press the Change/X Button to select the second value.
6. When all 3 parameters are confirmed, Press and hold the Menu/Check Button to start Leak Test. “Start” will be shown on the display for 2 seconds and then the Leak Test starts. During the Leak Test, the vacuum reading and time left on test are shown on the display in order.
7. User may press Mode Button to interrupt Leak Test mode before test time expires and go back to Vacuum Reading Mode. “Quit?” will be shown on display to confirm interrupting test. Press Menu/Check Button to end test or Change/X button to continue test.
8. When the test time ends, “End”, the start vacuum reading, end vacuum reading and vacuum drop value will be shown on the display in order. User may press Menu/Check Button or Change/X Button to scroll between 4 readings.
9. Press Mode button to go back to Vacuum Reading mode.
10. User may press Mode Button
to skip Leak Test mode
and enter Programming mode before doing leak test.
Programming Set Mode
1. In Programming mode, user sets the target vacuum value and time duration. The vacuum value and time duration are shown on the display in order.
2. User may press Mode Button to enter Run mode without setting in programming mode if the parameters are confirmed before setting.
3. In Programming mode, press Menu/Check Button to select the parameter to be set in order. Vacuum value, unit, Minute, Second of time duration can be set. The parameter flashes when it is selected to be set. If the 4 parameters are all set, user may press Mode Button at anytime to leave Programming mode and enter Run mode. “Save?” will be shown on display. Press Menu/Check Button to save or press Change/X Button to quit without saving.
4. User may press and hold Menu/Check Button to hold and save and enter Run mode without “Save?
5. While the target vacuum value can be set, press the
Change/X Button to select the vacuum value.
6. While the target vacuum unit can be set, press the
Change/X Button to select the vacuum unit.
7. While the minute of time can be set, press the Change/X
Button to select the minute value.
8. While the second of time can be set, press the
Change/X Button to select the second value.
Run Mode
1. In Run mode, the elapsed time is accumulated when vacuum reading is less than target vacuum value.
2. User may press Mode Button to enter Vacuum Reading mode without changes in Run mode.
3. In Run mode the icon “Run” is shown on the display.
Ready” is shown on the display after entering Run mode.
4. User powers on the vacuum source and then presses
the Menu/Check Button to start monitoring the
vacuum. “Run” is shown on the display for 2 seconds when monitoring begins. When the vacuum value is less than target value, the vacuum gauge counts elapsed time and accumulated. During the monitoring, the vacuum value and the time count are shown on the display in
order.
5. Press the Mode Button to skip monitoring before the elapsed time expires and go back to Vacuum Reading mode. “Quit?” is shown on the display for confirming the interrupt. Press Menu/Check Button to confirm quit or press Change/X Button to cancel quit. When the accumulated elapsed time is greater than the set time duration, “End” flashes on the display. Press the Mode Button to end monitoring and go back to Vacuum Reading mode.
6. The Sleep mode is disabled in Run mode. User should check battery life in before and during long monitoring
times.
BATTERY REPLACEMENT
New
(CR123A)
battery
When the batteries are too low for safe operation, the Low Battery indicator will display.
• Remove screw. Remove battery cover.
• Replace the old batteries with 1 new (CR123A) battery.
• Place the battery cover.
WARRANTY
The DMG150 is warranted to be free from defects in materials and workmanship for a period of one year from the date of purchase. If within the warranty period your instrument should become inoperative from such defects, the unit will be repaired or replaced at UEi’s option. This warranty covers normal use and does not cover damage which occurs in shipment or failure which results from alteration, tampering, accident, misuse, abuse, neglect or improper maintenance. Batteries and consequential damage resulting from failed batteries are not covered by warranty.
Any implied warranties, including but not limited to implied warranties of merchantability and fitness for a particular purpose, are limited to the express warranty. UEi shall not be liable for loss of use of the instrument or other incidental or consequential damages, expenses, or economic loss, or for any claim or claims for such damage, expenses or economic loss.
A purchase receipt or other proof of original purchase date will be required before warranty repairs will be rendered. Instruments out of warranty will be repaired (when repairable) for a service charge.
For more information on warranty and service, contact:
Sleep Mode
1. The gauge will enter Sleep mode if the reading is over limit and continues for 5 minutes. “Sleep” will be shown on the display when the gauge is in Sleep mode.
After 5 minutes of Sleep mode, Auto power off powers
off the gauge. The down-count timer is shown on the display to alert user of sleep mode and Auto power off
time.
Oil Detection
1. When the vacuum sensor has contamination like oil or other material, “Oil” is shown on the display.
2. The gauge will Auto power off after 5 minutes.
Sensor Fail
1. If the sensor wire is broken or fails, “Fail” will be shown on the display.
2. The gauge will Auto power off after 5 minutes.
Copyright © 2017 Kane USA Inc. All Rights Reserved.
This warranty gives you specific legal rights. You may also have other rights, which vary from state to state.
DISPOSAL
CAUTION: This symbol indicates that equipment
and its accessories shall be subject to separate collection and correct disposal.
CLEANING
Periodically clean your meter’s case using a damp cloth. DO NOT use abrasive, flammable liquids, cleaning solvents, or strong detergents as they may damage the finish, impair safety, or affect the reliability of the structural components.
STORAGE
Remove the batteries when instrument is not in use for a prolonged period of time. Do not expose to high temperatures or humidity. After a period of storage in extreme conditions exceeding the limits mentioned in the General Specifications section, allow the instrument to return to normal operating conditions before using it.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
Loading...