UEi DL479 Instruction Manual

DL479
Visit us at www.TestEquipmentDepot.com
AC 600A TRMS HVAC/R Clamp Meter
INSTRUCTION MANUAL
ENGLISH
1000V 600A
DL479
°C°F
NCV
µA
A
99 Washington Street Melrose, MA 02176
Phone 781-665-1400 Toll Free 1-800-517-8431
True RMS
600V
CAT IV
MAX/MIN
Hz / Duty
1000V
CAT III
CAT IV 600V
CAT III 1000V
MAX
CATEGORY DEFINITIONS
Measurement
Category
II < 10 Circuits connected to mains socket
III < 50 Mains distributions parts of the building
IV > 50 Source of the mains installation in the
Short-Circuit
(typical) kA
a
outlets and similar points in the MAINS installation
building
Location in the
building installation
WARRANTY
The DL479 is warranted to be free from defects in materials and workmanship for a period of two years from the date of purchase. If within the warranty period your instrument should become inoperative from such defects, the unit will be repaired or replaced at UEi’s option. This warranty covers normal use and does not cover damage which occurs in shipment or failure which results from alteration, tampering, accident, misuse, abuse, neglect or improper maintenance. Batteries and consequential damage resulting from failed batteries are not covered by warranty.
Any implied warranties, including but not limited to implied warranties of merchantability and fitness for a particular purpose, are limited to the express warranty. UEi shall not be liable for loss of use of the instrument or other incidental or consequential damages, expenses, or economic loss, or for any claim or claims for such damage, expenses or economic loss.
A purchase receipt or other proof of original purchase date will be required before warranty repairs will be rendered. Instruments out of warranty will be repaired (when repairable) for a service charge.
This warranty gives you specific legal rights. You may also have other rights, which vary from state to state.
DISPOSAL
CAUTION: This symbol indicates that equipment and its accessories
shall be subject to separate collection and correct disposal.
CLEANING
Periodically clean your meter’s case using a damp cloth. DO NOT use abrasive, flammable liquids, cleaning solvents, or strong detergents as they may damage the finish, impair safety, or affect the reliability of the structural components.
STORAGE
Remove the batteries when instrument is not in use for a prolonged period of time. Do not expose to high temperatures or humidity. After a period of storage in extreme conditions exceeding the limits mentioned in the General Specifications section, allow the instrument to return to normal operating conditions before using it.
Copyright © 2016 UEi. All Rights Reserved.
FEATURES
• True RMS
• 600A AC
• 750V AC/600V DC
• Resistance 60MΩ
• Capacitance 2000µF
• Temperature -31˚ to 752˚F (-35˚ to 400˚C)
• DC Microamps 2000µA
• Frequency/Duty Cycle
• Data hold
• Min/Max (25 ms fast capture)
• Manual ranging option
• Backlight
• Worklight
• Audible voltage indicator
• Low battery indicator
• Auto power off
• Test lead storage
• Overmolded grip
• Magnetic mount
• Battery compartment latches
GENERAL SPECIFICATIONS
• Operating Temperature: 32° to 104°F (0° to 40°C)
• Storage Temperature 14° to 122°F (-10° to 50°C)
• Operating Humidity: <75%
• Operating Altitude: 6,562 ft. (2000m)
• Pollution Degree: 2
• Display: 3 5/6" Digits 6,000 Count
• Backlight: Yes
• Refresh Rate: 3/sec
• Over-range: “OL” is displayed
• Dimensions: 8.75” X 3.5” X 1.75”
• Item Weight: 0.65 lb
• Calibration: Recommended Annually
• CAT Rating: CAT IV 600V / CAT III 1000V
• Certifications: cETLus 3rd Edition, CATIII 1000V, IEC 61010-1, 61010-2-032,
• Battery Type: (AAA) X 2
• Test Leads: Test lead w/Alligator Clips
• Accuracy: ± (% of reading + # of least significant digits)
61010-2-033, CSA C22.2, NO. 61010-1, 61010-2-032, 61010-2-033, 6’ Drop Protection, RoHS Compliant, CE Conformity
IMPORTANT SAFETY WARNINGS
WARNING
Read entire Safety Notes section regarding potential hazards and proper instructions before using this meter. In this manual the word “WARNING” is used to indicate conditions or actions that may pose physical hazards to the user. The word “CAUTION” is used to indicate conditions or actions that may damage this instrument.
WARNING
To ensure safe operation and service of the tester, follow these instructions. Failure to observe these warnings can result in severe injury or death.
WARNING
• Before each use, verify meter operation by measuring a known voltage or current.
• Never use the meter on a circuit with voltages that exceed the category based rating of this meter.
• Do not use this meter during electrical storms or in wet weather.
• Do not use the meter or test leads if they appear damaged.
• Ensure meter leads are fully seated and keep fingers away from the metal probe contact when making measurements. Always grip the leads behind the finger guards molded into the probe.
• Do not open the meter to replace batteries while the probes are connected.
• Use caution when working with voltages above 60V DC or 25V AC RMS. Such voltages pose shock hazards.
• To avoid false readings that can lead to electrical shock, replace batteries if a low battery indicator appears.
• Unless measuring voltage or current, shut off and lockout power before measuring resistance or capacitance.
• Always adhere to national and local safety codes. Use proper personal protective equipment (PPE) to prevent shock and arc blast injury where hazardous live conductors are exposed.
• Always turn off power to a circuit or assembly under test before cutting, unsoldering or breaking the current path. Even small amounts of current can be dangerous.
• Always disconnect the live test lead before disconnecting the common test lead from the circuit.
• In the event of electrical shock, ALWAYS bring the victim to the emergency room for evaluation, regardless of victim’s apparent recovery. Electrical shock can cause unstable heart rhythms that may need medical attention.
• If any of the following occur during testing, turn off the power source to the circuit being tested: arching, flame, smoke, extreme heat, smell of burning materials or discoloration or melting of components.
WARNING
Higher voltages and currents require greater awareness of physical safety hazards. Before connecting the test leads, turn off power to the circuit under test, set meter to the desired function and range, connect the test leads to the meter first, then connect to the circuit under test. Reapply power. If an erroneous reading is observed, disconnect power immediately and recheck all settings and connections.
WARNING
This meter is designed for trade professionals who are familiar with the hazards of their trade. Observe all recommended safety procedures that include proper lock-out utilization and use of personal protective equipment that includes safety glasses, gloves and flame resistant clothing.
SYMBOLS
AC (Alternating current) DC (Direct current)
Negative DC AC/DC Voltage
Auto-ranging Overload: Range Exceeded
Auto power off Active Low Battery
Non-Contact Voltage
Hold/Capture Value Duty Cycle
Maximum measured value displayed
Hertz/Frequency Ohms/Resistance
Amperage
Diode Degrees Celsius
Degrees Fahrenheit Continuity
High Voltage Indication Milli (x10-3 or 0.001)
Mega (x106 or 1,000,000) Nano (x10-9 or 0.000000001)
Kilo (x103 or 1,000) Warning or Caution
Micro (x10-6 or 0.000001) Dangerous Levels
Ground
Double Insulation (Protection to Class II)
Minimum measured value displayed
Voltage
Capacitance mode in nanofarads or microfarads
Safe for disconnect from live conductors