UEi DL429 Instruction Manual

DL429
True RMS
600V 600A
CAT Ill
CATIII
600V
MAX
T2
T1
TEMP
INRUSH
BT RF
T1-T2
NCV
MAX/ MIN
LRA Inrush
RANGE
T
e
m
p
/
°
F
°
C
μ
A
V
H
z
/
D
u
t
y
MFD
LINK
INSTRUCTION MANUAL
ENGLISH
DL429
1-800-547-5740 • Fax: (503) 643-6322
www.ueitest.com • email: info@ueitest.com
True RMS Digital Clamp-On Wireless with Dual Temperature
WARRANTY
The DL429 is warranted to be free from defects in materials and workmanship for a period of two years from the date of purchase.
For more information on warranty and service:
www.ueitest.com • Email: info@ueitest.com
1-800-547-5740 • FAX: (503) 643-6322
This warranty gives you specific legal rights. You may also have other rights which vary from state to state.
FCC INFORMATION
NOTE: This equipment has been tested and found to comply with the limits for
a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation.
This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the equipment and receiver.
• Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
• Consult the dealer or an experienced radio/TV technician for help.
CAUTION : Changes or modifications not expressly approved by the
manufacturer responsible for compliance could void the user’s authority to operate the equipment.
BATTERY REPLACEMENT
• When indicator is displayed on the LCD, batteries must be replaced.
• Remove the back screw and replace 6 x AAA batteries.
CLEANING
Turn instrument off and disconnect test leads. Clean the instrument by using a damp cloth. Do not use abrasive cleaners or solvents.
STORAGE
Remove the batteries when instrument is not in use for a prolonged period of time. Do not expose to high temperatures or humidity. After a period of storage in extreme conditions exceeding the limits mentioned in the Specifications section, allow the instrument to return to normal operating conditions before using it.
DISPOSAL / RECYCLE
Caution : This symbol indicates that equipment and its
accessories shall be subject to a separate collection and correct disposal.
10097 02/15
Copyright © 2015 UEi. All Rights Reserved.
iPhone® and iPad® are trademarks of Apple, Inc., registered in the U.S. and other countries. Android™ is a trademark of Google Inc. Other third party trademarks referenced are the property of their respective owners.
FEATURES
• True RMS
• Direct function access with push-button operation
• Dual Display
• Dual Temperature with Differential
• Locked Rotor Inrush Measurement
• Non-contact voltage
• Min / Max / Hold
• Backlit display with Clamp Work-light
• Magnetic mount for hands free use
• RF or Bluetooth® Wireless: BT app available for Android and iOS
• Accepts CH3: The Hook, or Fieldpiece accessory heads
• Input Jack Safety Lockout
• Auto-ranging measurement with manual ranging capability
• Auto Power Off: After 30 minutes of non-use.
• Low battery lockout: Displays “BATT” in display blocking potentially
inaccurate readings.
• Visible High-Voltage alert
MEASURES
• Amps AC to 600A
• µA DC 0.1µA~ 2000µA
• Volts DC 0.1mV ~ 1,000V
• Volts AC 0.1mV ~ 750V
• Resistance 0.1Ω ~ 60MΩ
• Capacitance 0.01nF ~ 2,000µF
• Temperature -22° ~ 752°F (-30° ~ 400°C) with Differential
• Frequency 0.01Hz ~ 99.99kHz
• Duty Cycle
GENERAL SPECIFICATIONS
Altitude: Operating - up to 2000m (6,561 ft.)
Storage - 10,000m (32,808 ft.)
Humidity: 80% max
Operating Temperature: 32°F to 122°F (0°C to 50°C) at < 75% R.H
Storage Temperature: -4°F to 140°F (-20°C to 60°C) at < 80% R.H
Relative humidity: 0% to 80% at 32°F to 95°F (0°C to 35°C),
0% to 70% at 32°F to 131°F (0°C to 55°C)
Temperature Coefficient: Nominal 0.1 x (Specified accuracy) / °C
(<18°C or >28°C ; <64°F or >82°F)
Pollution degree: 2
Display: 3-5/6 digits 6000 counts dual LCD display(s)
Refresh Rate: 3 times/sec
Overrange: "OL" is displayed
Polarity: Automatic(no indication for positive polarity) ; Minus(-) sign
for negative polarity
Dimensions: 10.2" x 2.5" x 1.5"
Weight: 15.2oz.
Calibration: Accurate for one year
CAT Rating: CAT III 600V
Certifications: ETL & cETL IEC61010-2-032
Battery type: 6 x 1.5V AAA or LR03
Silicon Test Lead: IEC61010-2-031
Accuracy: ± (% of reading + # of least significant digits)
WARNINGS
To ensure safe operation and service of the tester, follow these instructions. Failure to observe these warnings can result in severe injury or death.
• Before each use, verify meter operation by measuring a known voltage or current.
• Never use the meter on a circuit with voltages that exceed the category based rating of this meter.
• Do not use the meter during electrical storms, or in wet weather.
• Do not use the meter or test leads if they appear to be damaged.
• Ensure meter leads are fully seated, and keep fingers away from the metal probe contacts when making measurements.
• Do not open the meter to replace batteries while the probes are connected.
• Use caution when working with voltages above 60V DC, or 25V AC RMS. Such voltages pose a shock hazard.
• To avoid false readings that can lead to electrical shock, replace batteries if a low battery indicator appears.
• Unless measuring voltage or current, shut off and lock out power before measuring resistance or capacitance.
• Always adhere to local and national safety codes. Use Personal Protective Equipment (PPE) to prevent shock and arc blast injury.
SYMBOLS USED ON LCD
~
AC Measurement DC Measurement
Negative DC
AT
Auto Ranging
O.L.
Overload: Range Exceeded
Apo
Auto Power-Off Active
Low Battery
HOLD
Display not updating
MIN
Minimum measured value displayed
M
AX
Maximum measured value displayed
%
Units for duty cycle Units for frequency
V
Voltage measurement mode
A
Amperage measurement mode
Ω
Resistance measurement mode
Diode test mode
MFD
Capacitance measurement mode in nano farads or microfarads
VALUE MULTIPLIER
M
mega (x 106 or 1,000,000)
k
Kilo prefix (x 103 or 1,000)
m
milli (x 10-3 or 0.001)
n
nano (x 10-9 or 0.000000001)
µ
micro (x 10-6 or 0.000001)
INTERNATIONAL SYMBOLS
~
AC Alternating Current Warning or Caution
DC Direct Current Dangerous levels
DC/AC Voltage or Current Double Insulated Class II
Ground
Safe for disconnect from live conductors
NAVIGATION
• Press briefly to turn the meter on
• Press and hold “HOLD” while turning on to disable auto power off.
• Press and hold to turn the meter off
• The new user interface allows direct access from any mode.
• Press and hold to access functions in yellow text.
• Wireless app and meter display update simultaneously.
• Modes and functions can be selected from either the app or the meter.
• From wireless app press ' ' first
• Press to select DC µA measurement mode.
• Press to select T1.
• Press a second time for T2.
• Press a third time for T1 - T2.
• Press and hold to change from °F to °C.
MF
D
• Press to select continuity.
• Press a second time for resistance.
• Press a third time for capacitance and a fourth time for diode.
• Press to select VAC or VDC In VAC mode
• Press and hold for Hz
• Press and hold again for Duty Cycle.
NCV
• Non-Contact Voltage Detection key is used to detect power with a sensor located at the tip of the clamp head and indicates positive response with an audible alarm and visual LED indicator light just above the “NCV” button.
• Do not use non-contact voltage detector to determine if there is current in the wire. Detection operation could be affected by socket design, insulation thickness, type and other factors.
• Voltage indicator light may also light when voltage is present on the meter’s input jack or from external interference sources such as motors, flashlights etc.
• From wireless app press ‘ ’ first
MAX/ MIN
• Press to enter Max / Min mode; the largest and smallest values will be saved while in this mode.
• Press repeatedly to alternate between the maximum and minimum readings.
• Press for 2 seconds to return to live reading and clear the stored maximum and minimum values.
Note: Select range prior to selecting Min/Max to capture large values
LRA Inrush
The UEi LRA Inrush is programmed to properly capture the starting current for compressor motors.
• Select AC Amps
• Select the range capable of capturing the maximum value
• Press and hold LRA Inrush for two seconds – INRUSH will now be shown on the screen
• Activate compressor and read value on the display
• Press and hold LRA Inrush for two seconds to exit
RANGE
• Press repeatedly to cycle through manual ranges.
• Press for 2 seconds to return to auto ranging mode.
• AT is displayed on LCD only during auto ranging mode.
Note: Select range prior to Min/Max for best results.
LINK
• Press to activate Bluetooth®
• Press again to activate RF wireless ◊ Press and hold to turn radio off
• The RF will work with specific transmitter, while the BT will work with a tablet or smartphone with app installed
HOLD
• Press to hold the reading on the display. Press again to return to live reading.
• Press and hold "work light" key for 2 seconds to turn on. Press and hold again to turn off.
DL429
True RMS
600V 600A
CAT Ill
INRUSH
BT RF
T1-T
2
NCV
MAX/ MIN
LRA Inrush
RANGE
T
e
m
p
/
°
F
°
C
μ
A
V
H
z
/
D
u
t
y
MFD
LINK
x1x1
DL429
600V 600A
CAT Ill
NCV
MAX/ MIN
LRA Inrush
RANGE
T
e
m
p
/
°
F
°
C
μ
A
V
H
z
/
D
u
t
y
MFD
LINK
HOLD
Use CAT III rated leads or higher.
Do not attempt to measure more than 1000V DC or 750V AC.
• Press for AC Volts
Note: This is the default setting when first powered on. ◊ Press and hold in AC volts for Hz/Duty Cycle
• Press again for DC volts
• High Volt Visible Alert - Any input exceeding 30V (AC or DC) will light the NVC LED to alert users to potentially dangerous voltage levels
• Press for T1
• Press again for T2
• Press again for T1-T2
• Press and hold to change scale
Features:
NCV
LINK
MAX/ MIN
HOLD
Range Resolution Accuracy Overload Protection
-22°F ~ 14°F (-30°C ~ -10°C)
0.1°F (0.1°C)
± (1.0% + 5.4°F) ± (1.0% + 3.0°C)
30V RMS
15°F ~ 752°F (-9°C ~ 400°C)
0.1°F (0.1°C)
± (1.0% + 3.6°F) ± (1.0% + 2.0°C)
Sensor: K-Type Thermocouple, sensor accuracy not included
TEMP/ °F °C°
DL429
True RMS
600V 600A
CAT Ill
CATIII
600V
MAX
30V
T
2
T
1
TEMP
INRUSH
BT RF
T1-T
2
NCV
MAX/ MIN
LRA Inrush
RANGE
T
e
m
p
/
°
F
°
C
μ
A
V
H
z
/
D
u
t
y
MFD
LINK
T
e
m
p
/
°
F
°
C
When the meter is powered on, the upper display will always show
amperage, or adapter (ADP) output
• Center wire in guides for best accuracy.
• Opposing currents cancel (use line-splitter when necessary).
Keep hands below line when measuring high current levels.
Features:
NCV
LINK
MAX/ MIN
LRA Inrush
RANGE
HOLD
AC Amps Measurement - Jaw input
Range Resolution Accuracy Overload Protection
60A 0.01A ± (2.9% + 15 digits)
600V RMS
600A 0.1A ± (1.9% + 8 digits)
True RMS: 45Hz to 400Hz * Minimum Current for Clamp Measurement: 0.2A
AC Amps: < 600A
Single Conductor Only
DL429
True RMS
600V 600A
CAT Ill
CATIII
600V
MAX
30V
T
2
T
1
TEMP
INRUSH
BT RF
T1-T
2
NCV
MAX/ MIN
LRA Inrush
RANGE
T
e
m
p
/
°
F
°
C
μ
A
V
H
z
/
D
u
t
y
MFD
LINK
• Press for DC micro amps
Features:
NCV
LINK
MAX/ MIN
LRA Inrush
RANGE
HOLD
DC Low Amps Measurement -Test lead input
Range Resolution Accuracy Overload Protection
600µA 0.1µA
± (1.2% + 3 digits) 2000µA / 600V RMS
2000µA 1µA
DC Low Amps: < 2000uA
DL429
True RMS
600V 600A
CAT Ill
CATIII
600V
MAX
30V
T
2
T
1
TEMP
INRUSH
BT RF
T1-T2
NCV
MAX/ MIN
LRA Inrush
RANGE
T
e
m
p
/
°
F
°
C
μ
A
V
H
z
/
D
u
t
y
MFD
LINK
x1 =
Hz/Duty
DL429
True RMS
600V 600A
CAT Ill
CATIII
600V
MAX
30V
T
2
T
1
TEMP
INRUSH
BT RF
T1-T
2
NCV
MAX/ MIN
LRA Inrush
RANGE
T
e
m
p
/
°
F
°
C
μ
A
V
H
z
/
D
u
t
y
MFD
LINK
x1 =
Press and hold = Hz/Duty Cycle
x2 =
V
H
z
/
D
u
t
y
• Forward voltage drop if forward biased.
• “O.L.” if reverse biased.
Features:
NCV
LINK
MAX/ MIN
HOLD
Diode Test
Range Open Circuit Voltage Test Current (Typical) Overload Protection
2.0V< 1.6V DC 0.25mA 600V RMS
BAD DIODE
Open Diode
Displays "OL" Both directions
'0' Both directions
(Shorted)
or
Features:
NCV
LINK
MAX/ MIN
LRA Inrush
RANGE
HOLD
DC Voltage Measurement
Range Resolution Accuracy Overload Protection
600mV 0.1mV
± (0.5% + 4 digits)
1000V
6V 1mV 60V 10mV 600V 100mV 1000V 1V ± (0.8% + 10 digits)
AC Voltage Measurement
Range Resolution Accuracy Overload Protection
600mV 0.1mV
± (2.0% + 5 digits) 750V RMS
6V 1mV 60V 10mV 600V 100mV 750V 1V
True RMS: 45Hz to 400Hz
Frequency Measurement
Range Resolution Accuracy Overload Protection
99.99Hz 0.01Hz
± (0.1% + 4 digits) 600V RMS
999.9Hz 0.1Hz
9.999kHz 1Hz
99.99kHz 10Hz
Sensitivity: 1.8V RMS
Duty Cycle Measurement
Range Accuracy Overload Protection
1.0 ˜ 99.0% ±(0.2% per kHz + 0.1% + 5 digits) 600V RMS
Audible Continuity
• Press for continuity
• Display shows resistance.
• Buzzer sounds if less than 40Ω.
Features:
NCV
LINK
MAX/
MIN
LINK
HOLD
Overload Protection Open Circuit Voltage
600V RMS
<
0.44V
Threshold Approx : < 40Ω
Resistance
• Press again for resistance
• Do not measure resistance on a live circuit.
Features:
NCV
LINK
MAX/ MIN
LRA Inrush
RANGE
HOLD
Range Resolution Accuracy Overload Protection
600Ω 0.1Ω
± (1.0% + 4 digits)
600V RMS
6kΩ 1Ω
60kΩ 10Ω
600kΩ 100Ω
6MΩ 1kΩ
60MΩ 10kΩ
± (2.0% + 4 digits)
Capacitance
• Press again for capacitance
Features:
NCV
LINK
HOLD
Range Resolution Accuracy Overload Protection
60.00nF 0.01nF
± (3.5% + 6 digits) 600V RMS
600.0nF 0.1nF
6.000μF 0.001μF
60.00μF 0.01μF
600.0μF 0.1μF
2000μF 1μF
/ / MFD /
DL429
True RMS
600V 600A
CAT Ill
CATIII
600V
MAX
30V
T
2
T
1
TEMP
INRUSH
BT RF
T1-T
2
NCV
MAX/ MIN
LRA Inrush
RANGE
T
e
m
p
/
°
F
°
C
μ
A
V
H
z
/
D
u
t
y
MFD
LINK
x1 = x2 = x3 =
MFD
x4 =
MF
D
GOOD DIODE
Diode
• Press again for diode test
DL429
True RMS
600V 600A
CAT Ill
CATIII
600V
MAX
30V
T
2
T
1
TEMP
INRUSH
BT RF
T1-T
2
NCV
MAX/ MIN
LRA Inrush
RANGE
T
e
m
p
/
°
F
°
C
μ
A
V
H
z
/
D
u
t
y
MFD
LINK
MFD
Forward Bias
Displays approx. voltage drop
x1 = x2 = x3 =
MFD
x4 =
Reverse Bias
Displays "OL"
Connecting and Using the App
• Search for App as, “UEi Wireless”
Compatible with iPhone® 4X and up running iOS® 7 or higher, Galaxy S4®,
Nexus 5™, HTC One® running Android™ 4.4 or higher.
• To install or search on iPad® use "iPhone® only" to find App.
• Press "Link" button on meter to activate wireless "BT"
• Open app. Meter will connect automatically.
Menu
• Press “ ” to connect, disconnect, and
access settings.
Settings
• General settings adjust button sound, vibrate and refresh rate.
Recording settings
• Continuous reading
• Number of samples
• Sampling interval
Record
• Press to start, stop.
• The number of samples will show in real time
Logs
• Press to view recorded data.
• Press the entry you wish to view (yyyy-mm-dd hh:mm:ss)
• Functions are noted underneath respectively AMP-AMP (TOP-
BOTTOM) Display
• Press “ “ button for summary
• Press “ “ button for sample data
• Press “ “ button to export data via email in (.CSV .PNG or .JPG) formats
Graph
• Press “ ” to view trending data
in real time during measurement.
High Voltage Alert
Graph
Graph