SECTION III Alarms .....................................................................................................................................
SECTION IV Cleaning ...................................................................................................................................
Important User Information ......................................................................................................
General Cleaning Directions ......................................................................................................
Directions for Cleaning Pump Housing .......................................................................................
English
Directions for Cleaning Pump Power Cord ...................................................................................
Directions for Cleaning Drop Detector ........................................................................................
Directions for Cleaning OPTIX Lock .............................................................................................
Directions for Cleaning Rotor Assembly ......................................................................................
SECTION VIII Warranty ..................................................................................................................................
kANGAROO 924
11/10/2006 Artwork#s: WN 182493-01 Job#: 14761
Enteral Feeding Pump
Status: Review Routing On: CR10000186 Doc#: AW10001678
Section I General Information
The Tyco Healthcare Group LP KANGAROO™ 924 Enteral Feeding Pump is a rotary peristaltic pump designed to regulate the flow rate of
enteral feedings. Among the features provided by this pump are:
• Portable operation-built in rechargable battery
• Wide flow rate range 1-300 ml/hr in 1 ml/hr increments
• Small and lightweight for easy transport
• Language Selection (13 languages available)
• Touch panel and knob operation
• Shut off and activation of audible and visible alarms when:
− Feeding container is empty
− Feeding tube becomes occluded
− Drop sensors blocked
− Battery low (alarms 15 minutes before pump shut off )
− Unit is left on “HOLD” longer than 5 minutes
− Pump set is improperly loaded
− Self diagnosed problem which requires technical servicing
− Pump motor malfunction
− Ambient light error
• Automatic systems check after pump has been turned ON
• Large LED display
• Low occlusion pressure. Pump will not overcome back pressure greater than approximately 12 psi (82.7 kPa)
• Integral pole clamp
• Use of OPTIX pump sets to prevent free flow from inadvertent misload of pump set onto pump
• Pump running indicator lights
• DC motor with intermittent operation
• Adjustable alarm volume
• Memory of flow rate setting is retained for 16 or 24 hrs. after pump shut off, depending on AC connection status.
WARNING This KANGAROO 924 Pump requires the use of a KANGAROO 924 Pump Set. The pump will not operate if
sets other than KANGAROO 924 Pump Sets are loaded into the pump.
kANGAROO™ 924
1
11/10/2006 Artwork#s: WN 182493-01 Job#: 14761
Enteral Feeding Pump
7
8
12
11
13
1
2
3
4
5
6
14
10
15
9
1
2
3
4
5
6
10
14
10
15
Status: Review Routing On: CR10000186 Doc#: AW10001678
Status: Review Routing On: CR10000186 Doc#: AW10001678
Section II Operating Procedures
NOTE: To disconnect AC power it is necessary to unplug
power cord from the AC wall outlet.
AC Light
On whenever pump is plugged in.
Vol CLR Light
On when volume delivered has been cleared.
ON
Powers the pump (See Figure 3 for location).
OFF
Shuts pump off. NOTE: To disconnect AC power it is
necessary to unplug power cord from the AC wall outlet
(See Figure 3 for location)
Increases rate or VTBD desired.
Decreases rate or VTBD desired.
NOTE: NUMBERS DISPLAYED WILL CHANGE MORE RAPIDLY IF
EITHER ARROW BUTTON IS HELD IN.
Adjustment Knob
Increases or decreases rate or VTBD desired.
Figure 3
VOL
Displays volume delivered for 5 seconds when pressed.
When the pump has been off for 24 hours (16 hours battery),
the volume delivered is automatically reset to zero.
VOL CLR
Clears volume delivered when pressed after VOL has been pressed. The VOL CLR light will go on when volume delivered has been
reset to zero. Volume delivered cannot be cleared until the preset VTBD has been delivered or cleared.
VTBD (Volume To Be Delivered)
When pressed, allows user to preset a volume to be delivered by pressing or or rotating adjustment knob. Holding for 5
seconds will clear VTBD setting and volume. VOL CLR light flashes 3 times to signal that VTBD has been cleared.
START/HOLD
Starts pump rotor to deliver feeding formula to patient and alternately puts unit on HOLD. HOLD is used to stop fluid flow to:(1)
correct alarm condition-alarm is silenced (2) change the flow rate(3) change pump set. When on HOLD, the display flashes on and off
and the total delivery is retained.
NOTE: IF THE PUMP HAS BEEN PLACED ON HOLD AND NOT RESTARTED WITHIN 5 MINUTES, AN ALARM WILL SOUND.
Normal Operation
Insert power cord into AC wall outlet.
kANGAROO™ 924
3
11/10/2006 Artwork#s: WN 182493-01 Job#: 14761
Enteral Feeding Pump
A
Status: Review Routing On: CR10000186 Doc#: AW10001678
Battery Operation
Unplugging the pump from the AC line outlet will automatically put the pump on battery power. A new battery, when fully charged,
will operate the instrument for more than 16 hours at 125 milliliters per hour. Approximately 15 minutes prior to battery discharge, a
low battery alarm will occur, (see alarm section). When complete discharge occurs, the pump will automatically turn itself off.
NOTE: AS THE BATTERIES GET OLDER, THE TIME FROM LOW BATTERY ALARM TO COMPLETE DISCHARGE MAY BE LESS THAN 15 MINUTES.
To recharge battery, insert power cord into AC wall outlet. Battery will automatically begin charging. The recharge time for a fully
discharged battery is approximately 12 hours.
The battery will charge whenever the pump is plugged into a wall outlet, even when the OFF button has been pressed. To be certain that
power to all circuits has been disconnected, the power cord must be unplugged from the wall outlet.
Battery
Replacement battery packs are available from Tyco Healthcare Group LP (See Section VI - Factory Service)
WARNING Replace the battery pack only with the same type and rating of battery.
Beginning Administration
Below are general instructions for inserting all Tyco Healthcare Group LP Pump Sets into the KANGAROO 924 Enteral Feeding Pump.
1. Press
2. Wait while pump performs system check. (Displays “
3. Check
4. Rate of delivery may be changed when the delivery rate is shown on the display by using the adjustment knob or by
ON.
“ for a moment.)
VOL CLR light or press VOL to insure that volume delivered is clear, if desired.
pressing or button. Holding button in will cause numbers to continue to increase or decrease.
Figure 4A Figure 4B
5. Slide drip chamber into upper drip chamber guide (See Figure 4A).
6. Seat and secure chamber in lower drip chamber guide (A).
7. Remove distal connector protective cover from the feeding set..
8. Follow 924 feeding set instructions to prepare for feeding.
9. Slowly open control clamp and fill entire line with fluid, leaving a few drops of fluid in the drip chamber.
NOTE: AVOID FILLING THE DRIP CHAMBER MORE THAN HALF FULL. Close control clamp.
10. Attach distal connector to feeding tube.
11. Inspect lens in OPTIX lock area (B). Clean any contaminants from lens before proceeding to the next step (Figure 5B).
kANGAROO™ 924
4
11/10/2006 Artwork#s: WN 182493-01 Job#: 14761
Enteral Feeding Pump