Ventilation Testing
Positive and Negative Duct
Accreditation (PANDA) System
Model PAN200 Series
Operation and Service Manual
Copyright
TSI Incorporated / 2012 / All rights reserved.
Address
TSI Incorporated / 500 Cardigan Road / Shoreview, MN 55126 / USA
Fax No.
(651) 490-3824
LIMITATION OF WARRANTY AND LIABILITY (effective June 2011)
(For country-specific terms and conditions outside of the USA, please visit www.tsi.com.)
Seller warrants the goods sold hereunder, under normal use and service as described in the
operator's manual, shall be free from defects in workmanship and material for 24 months, or if
less, the length of time specified in the operator's manual, from the date of shipment to the
customer. This warranty period is inclusive of any statutory warranty. This limited warranty is
subject to the following exclusions and exceptions:
a. Hot-wire or hot-film sensors used with research anemometers, and certain other components
when indicated in specifications, are warranted for 90 days from the date of shipment;
b. Pumps are warranted for hours of operation as set forth in product or operator’s manuals;
c. Parts repaired or replaced as a result of repair services are warranted to be free from defects
in workmanship and material, under normal use, for 90 days from the date of shipment;
d. Seller does not provide any warranty on finished goods manufactured by others or on any
fuses, batteries or other consumable materials. Only the original manufacturer's warranty
applies;
e. Unless specifically authorized in a separate writing by Seller, Seller makes no warranty with
respect to, and shall have no liability in connection with, goods which are incorporated into
other products or equipment, or which are modified by any person other than Seller.
The foregoing is IN LIEU OF all other warranties and is subject to the LIMITATIONS stated
herein. NO OTHER EXPRESS OR IMPLIED WARRANTY OF FITNESS FOR PARTICULAR
PURPOSE OR MERCHANTABILITY IS MADE. WITH RESPECT TO SELLER’S BREACH OF
THE IMPLIED WARRANTY AGAINST INFRINGEMENT, SAID WARRANTY IS LIMITED TO
CLAIMS OF DIRECT INFRINGEMENT AND EXCLUDES CLAIMS OF CONTRIBUTORY OR
INDUCED INFRINGEMENTS. BUYER’S EXCLUSIVE REMEDY SHALL BE THE RETURN OF
THE PURCHASE PRICE DISCOUNTED FOR REASONABLE WEAR AND TEAR OR AT
SELLER’S OPTION REPLACEMENT OF THE GOODS WITH NON-INFRINGING GOODS.
TO THE EXTENT PERMITTED BY LAW, THE EXCLUSIVE REMEDY OF THE USER OR
BUYER, AND THE LIMIT OF SELLER'S LIABILITY FOR ANY AND ALL LOSSES, INJURIES,
OR DAMAGES CONCERNING THE GOODS (INCLUDING CLAIMS BASED ON CONTRACT,
NEGLIGENCE, TORT, STRICT LIABILITY OR OTHERWISE) SHALL BE THE RETURN OF
GOODS TO SELLER AND THE REFUND OF THE PURCHASE PRICE, OR, AT THE OPTION
OF SELLER, THE REPAIR OR REPLACEMENT OF THE GOODS. IN THE CASE OF
SOFTWARE, SELLER WILL REPAIR OR REPLACE DEFECTIVE SOFTWARE OR IF UNABLE
TO DO SO, WILL REFUND THE PURCHASE PRICE OF THE SOFTWARE. IN NO EVENT
SHALL SELLER BE LIABLE FOR LOST PROFITS OR ANY SPECIAL, CONSEQUENTIAL OR
INCIDENTAL DAMAGES. SELLER SHALL NOT BE RESPONSIBLE FOR INSTALLATION,
DISMANTLING OR REINSTALLATION COSTS OR CHARGES. No Action, regardless of form,
may be brought against Seller more than 12 months after a cause of action has accrued. The
goods returned under warranty to Seller's factory shall be at Buyer's risk of loss, and will be
returned, if at all, at Seller's risk of loss.
Buyer and all users are deemed to have accepted this LIMITATION OF WARRANTY AND
LIABILITY, which contains the complete and exclusive limited warranty of Seller. This
LIMITATION OF WARRANTY AND LIABILITY may not be amended, modified or its terms
waived, except by writing signed by an Officer of Seller.
Service Policy
Knowing that inoperative or defective instruments are as detrimental to TSI as they are to our
customers, our service policy is designed to give prompt attention to any problems. If any
malfunction is discovered, please contact your nearest sales office or representative, or call
Customer Service department at +44 (0) 149 4 459200 (UK), (800) 874-2811 (USA), or (1) 651490-2811 (International).
CONTENTS
CHAPTER 1 UNPACKING AND PARTS IDENTIFICATION ............. 1
CHAPTER 2 PREPARING PAN200 SYSTEM FOR
AIR DUCT LEAK TESTING ................................................................ 5
CHAPTER 3 PERFORMING A DUCT LEAKAGE TEST ................ 11
Measuring Duct Static Pressure ................................................... 11
Measuring Duct Leakage Flow ..................................................... 12
Turning on the PAN200 Duct Leakage Tester.............................. 12
Using Leakage Test Application in the Model 9565-P .................. 13
Troubleshooting Guide ................................................................. 16
APPENDIX A SPECIFICATIONS ..................................................... 17
APPENDIX B LEAKAGE TESTING STANDARDS HIGHLIGHTS ... 21
Standards Supported .................................................................... 21
EU Standards ............................................................................... 22
US Standards ............................................................................... 25
APPENDIX C TYPICAL SETUP....................................................... 29
APPENDIX D PROCEDURE FOR USING SMOKE PELLETS
IN LEAKAGE TESTERS ................................................................... 31
iii
(This page intentionally left blank)
iv
Primary duct adapter
spigot plus rubber bung
(to fit to test duct)
Cam lock primary spigot
(to connect flexi-duct to
PANDA)
Ø4-in. (100-mm)
adjustable over lock
straps
13-ft (4-m) long
Ø4-in. (100-mm) plastic
flexible duct
20-in. (500-mm) silicone
tubes (red)
Chapter 1
Unpacking and Parts Identification
Carefully unpack the PANDA system and instrument cases from the
shipping container. Check the individual parts against the list of
components below. If anything is missing or damaged, notify TSI
immediately.
The PANDA system consists of the following:
1
20-in. (500-mm) silicone
tubes (blue)
16-ft (5-m) silicone tube
(blue)
Smoke cap holder
assembly
K-type thermocouple
probe
Operation and User’s
manual
2 Chapter 1
The following two instruments should be used in conjunction with the
9565-P Multi-function
Instrument
Refer to Operation and Service
Manual (TSI P/N 6004851) supplied
with the instrument for additional
parts supplied as standard.
Refer to Operation and Service
Manual (TSI P/N 1980568) supplied
with the instrument for additional
parts supplied as standard.
PANDA unit:
Unpacking and Parts Identification
(This page intentionally left blank)
4 Chapter 1
Successfully completing a duct leakage test requires certain
information be compiled prior to starting the test. Refer to
Appendix B for a discussion of standards relating to duct
leakage testing. The list below indicates the information
required:
Type of leakage test to be performed (Positive or
negative).
Leakage standard to be followed.
Air tightness/leakage class to be achieved
Amount of ductwork to be tested, such as the complete
system or a statistical sample.
Chapter 2
Preparing PAN200 System for Air Duct Leak
Testing
The following procedure should be followed carefully so that safe and
accurate leakage testing can be achieved:
1. Select the section of the ductwork to be tested.
2. Calculate the surface area of the ductwork of the section to
be tested.
3. Temporarily seal the all openings of the ductwork except one,
which will be connected to the PAN200 duct leakage tester.
4. Position the PAN200 unit as close to the remaining opening in the
ductwork as possible to minimize the flexible tubing needed.
Minimize bends in the flexible tubing to reduce the pressure loss,
giving the best performance.
5. Make sure the Fan Control Switch on the Fan Speed Controller is
in the OFF position and the multi-turn Fan Speed Control
potentiometer is fully turned counter-clockwise using the pictures
of the VFD in Figure 1 as a reference. Plug the cord into the
PANDA unit as shown in Figure 2 and Figure 3. Then connect
the other end of the cord to a suitable electrical supply.
5
Figure 1. 110V and 230V Fan Speed Controllers
Figure 2. Receptacle for
Power Cord
Figure 3. Connected Power Cord
CAUTION: Remove the power cord from the PANDA duct
leakage tester before tilting it to the vertical
position to avoid damaging the cord.
Note: The control pad on the Fan Speed Controller has been
disabled.
6 Chapter 2
6. Fit the primary duct adapter
spigot (black sheet metal with
rubber bung) to one end of the
4-in. (100-mm) diameter flexitube. Make an air-tight seal
using one of the over lock
straps and lever-locking cam
provided as shown in Figure 4.
Adjust the fit of the over lock
strap with a screwdriver.
7. Securely attach the black
primary duct adapter
spigot/flexi-tube assembly to
the opening on the ductwork to
be pressure tested.
Figure 4. Flex Ductwork
Connected to Primary Adapter
Spigot. Bung not shown.
8. If the static pressure tap on the
black Primary Duct Adapter is
open to the duct, connect the
16-ft (5-m) long blue silicone
tube to it as shown in Figure 5.
If the static pressure tap on the
black Primary Duct Adapter is
not open to the duct, drill a
4-mm hole in the duct and
insert about 6 inches (10 mm)
of the silicone tube into the
duct. Seal around the hole with
putty.
9. Connect the other end of the
4-in. (100-mm) flexi-tube to
Figure 5. Connecting Pressure
Tubing to Tap on Primary Duct
Adapter. Bung not shown.
the cam lock connector (grey cast aluminum without nozzle).
Make an airtight seal using the other over lock strap and leverlocking cam provided. Adjust the fit of the over lock strap with a
screwdriver.
10. Determine if you are going to perform a high- or low-flow testing
and positive or negative testing. Set-up the duct leakage tester by:
Preparing PAN200 System for Air Duct Leak Testing
a. For positive pressure,
high-flow testing,
remove the low flow
nozzle if it is installed.
Then, connect the grey
cast-aluminum cam
lock connector to the
outlet side of the
blower per Figure 6.
Close both cam lock
arms at the same time
to ensure proper fit.
Connect the free end of
the 16-ft (5-m) silicone
Figure 6. Positive Pressure,
High-Flow Set-Up
tube to the pressure tap marked DUCT PRESSURE P3 (+) on
the black square box.
Finally, verify that the FLOW GRID MODE pressure taps on
the inlet tubes are connected to the FLOW GRID pressure
taps, i.e., P1 (+) to P1 (+) and P2 (-) to P2 (-).
b. For positive pressure,
low-flow testing, add
the low-flow nozzle to
the blower inlet if it is
not installed per
Figure 7. Then, connect
the grey cast-aluminum
cam lock connector to
the outlet side of the
blower per Figure 6.
Close both cam lock
arms at the same time
to ensure proper fit.
Figure 7. Positive Pressure,
Low-Flow Set-Up
Connect the free end of the 16-ft (5-m) silicone tube to the
pressure tap marked DUCT PRESSURE P3 (+) on the black
square box.
8 Chapter 2