TSI/Alnor 9535-A User manual

Page 1
VELOCICALC®
AIR VELOCITY METER
MODEL 9535/9535-A

OPERATION AND SERVICE MANUAL

P/N 1980563, REVISION H
FEBRUARY 2016
www.GlobalTestSupply.com
Page 2
CONTENTS
CHAPTER 1 UNPACKING AND PARTS IDENTIFICATION ............ 1
CHAPTER 2 SETTING-UP ................................................................. 3
Supplying Power to the Model 9535/9535-A .................................. 3
Installing the Batteries ............................................................... 3
Using the Optional AC Adapter ................................................. 3
Using the Telescoping Probe ......................................................... 3
Extending the Probe .................................................................. 3
Retracting the Probe.................................................................. 3
Connecting to a Computer .............................................................. 4
CHAPTER 3 OPERATION ................................................................. 5
Keypad Functions ........................................................................... 5
Common Terms .............................................................................. 5
Menus ............................................................................................. 6
DISPLAY SETUP ................................................................................. 6
SETTINGS ......................................................................................... 6
FLOW SET UP .................................................................................... 6
ACTUAL/STANDARD SETUP ................................................................ 6
DATA LOGGING ................................................................................. 7
Measurements ........................................................................... 7
Delete Data ................................................................................ 7
% Memory ................................................................................. 7
LogDat2™ Downloading Software ................................................. 7
CHAPTER 4 MAINTENANCE ............................................................ 9
Recalibration ................................................................................... 9
Cases .............................................................................................. 9
Storage ........................................................................................... 9
CHAPTER 5 TROUBLESHOOTING ................................................ 11
APPENDIX A SPECIFICATIONS ..................................................... 13
iii
www.GlobalTestSupply.com
Page 3
N o t e
For temperature measurements, make sure that at least 3 inches (7.5 cm) of the probe is in the flow to allow the temperature sensor to be in the air stream.
Chapter 2
Setting-up

Supplying Power to the Model 9535/9535-A

The TSI Model 9535/9535-A VelociCalc Air Velocity Meter is powered with four size AA batteries.

Installing the Batteries

Insert four AA batteries as indicated by the diagram located on the inside of the battery compartment. The Model 9535/9535-A is designed to operate with either alkaline or NiMH rechargeable batteries, although it will not recharge NiMH batteries. Battery life will be shorter if NiMH batteries are used. Carbon-zinc batteries are not recommended because of the danger of battery acid leakage.

Using the Optional AC Adapter

When using the AC adapter, the batteries (if installed) will be bypassed. Be sure to provide the correct voltage and frequency, which is marked on the back of the AC adapter. The AC adapter will not recharge the batteries.

Using the Telescoping Probe

The telescoping probe contains the velocity, temperature, and humidity sensors. When using the probe, make sure the sensor window is fully exposed and the orientation dimple is facing upstream.

Extending the Probe

To extend the probe, hold the handle in one hand while pulling on
the probe tip with the other hand. Do not hold the cable while
extending the probe as this prevents the probe from extending.

Retracting the Probe

To retract the probe, hold the handle in one hand while pulling gently on the probe cable with the other hand.
www.GlobalTestSupply.com
3
Page 4

Connecting to a Computer

C a u t i o n
This symbol is used to indicate that the data port of the Model 9535/9535-A is not intended for connection to a public telecommunications network. Connect the USB data port only to another USB port.
Use the Computer Interface USB Cable provided with the Model 9535/9535-A to connect the instrument to a computer for downloading stored data.
For more information on how to download stored data see Chapter 3 section titled LogDat2™ Downloading Software.
4 Chapter 2
www.GlobalTestSupply.com
Page 5
ON/OFF Key
Press to turn the Model 9535/9535-A on and off. During the power up sequence the display will show the following: Model Number, Serial Number, Software Revision and Last Date Calibrated.
Arrow () Keys
Press to scroll through choices while setting a parameter.
(Enter) Key
Press to accept a value or condition.
Arrow (or ) and Menu Soft Keys
Press arrow keys to change choices while setting a parameter. Press the Menu soft key to select the Menu selections, which are Display Setup, Settings, Flow Setup, Actual/Std Set up, Data Logging, and Calibration.
Sample
Consists of all of the measurement parameters stored at the same time.
Test ID
A group of samples. The statistics (average, minimum, maximum, and count) are calculated for each test ID. The maximum number of test IDs is
100.
Time Constant
The time constant is an averaging period. It is used to dampen the display. If you are experiencing fluctuating flows, a longer time constant will slow down those fluctuations. The display will update every second, but the displayed reading will be the average over the last time constant period. For example, if the time constant is 10 seconds, the display will update every second, but the displayed reading will be the average from the last 10 seconds. This is also referred to as a “moving average”.
Chapter 3
Operation

Keypad Functions

Common Terms

In this manual there are several terms that are used in different places. The following is a brief explanation of the meanings of those terms.
www.GlobalTestSupply.com
5
Page 6

Menus

N o t e
The horn numbers are the models of the horns. For example, 100 refers to a horn model number AM 100. Only horns with Model numbers as follows can be used with this function: AM 100, AM 300, AM 600, and AM 1200. If a horn model number is chosen, the instrument will return to measuring mode and use a preprogrammed curve to calculate flow rate from velocity.

DISPLAY SETUP

Display setup menu is where you will setup the desired parameters to be displayed on the running screen. With a parameter highlighted you can then use the ON soft key to have it show up on the running screen or select the OFF soft key to turn off the parameter. Use PRIMARY soft key to have a parameter show up on the running screen in a larger display. Only one parameter can be selected as a primary, and up to 2 secondary parameters can be selected at one time.

SETTINGS

Settings menu is where you can set the general settings. These include Language, Beeper, Select Units, Time Constant, Contrast, Set Time, Set Date, Time Format, Date Format, Number Format, Backlight and Auto Off. Use the or soft keys to adjust the settings for each option and use the key to accept settings.

FLOW SET UP

In Flow Setup mode, there are 4 types: Round Duct, Rectangle Duct, Duct Area and Horn. Use the or soft keys to scroll through the types and then press the key to accept the desired type. To change the value, highlight the Enter Settings option and press the key. Use the instructions below to change the value of the type that you selected.

ACTUAL/STANDARD SETUP

Choose Actual/Standard measurements and parameters in the Act/Std Setup menu. Within this menu, the user can also select Standard Temperature, Standard Pressure and a source for the actual temperature. The actual barometric pressure must be entered
to convert air velocity and volume measurements to actual conditions.
6 Chapter 3
www.GlobalTestSupply.com
Page 7

DATA LOGGING

Measurements

Measurements to be logged are independent of measurements on the display, and must therefore be selected under DATA LOGGING Measurements.

Delete Data

Use this to delete all data, delete test or delete sample.

% Memory

This option displays the memory available. Delete All, under Delete Data, will clear memory and reset the memory available.

LogDat2 Downloading Software

The TSI Model 9535/9535-A comes with special software called LogDat2 Downloading Software, which is designed to provide you with maximum flexibility and power. To install this software on your computer, follow the instructions on the label of the LogDat2 CD­ROM.
To download data from the Model 9535/9535-A, connect the supplied computer interface USB cable to the Model 9535/9535-A and to a computer USB port. Then run the LogDat2 software. Within the LogDat2 software, either select the tests to be downloaded or double­click on a test to open it.
Operation 7
www.GlobalTestSupply.com
Page 8
Symptom
Possible Causes
Corrective Action
No Display
Unit not turned on
Switch unit on.
Low or dead batteries
Replace batteries or plug in AC adapter.
Dirty battery contacts
Clean the battery contacts.
Velocity reading fluctuates unstable
Fluctuating flow
Reposition probe in less-turbulent flow or use longer time constant.
No response to keypad
Keypad locked out
Unlock keypad by pressing  keys simultaneously.
Instrument Error message appears
Memory is full
Download data if desired, then DELETE ALL memory.
Fault in instrument
Factory service required on instrument.
Chapter 5
Troubleshooting
Table 5-1 lists the symptoms, possible causes, and recommended solutions for common problems encountered with the Model 9535/9535-A. If your symptom is not listed, or if none of the solutions solves your problem, please contact TSI.
Table 5-1: Troubleshooting the Model 9535/9535-A
Remove the probe from excessive temperature immediately: excessive heat can damage the sensor. Operating temperature limits can be found in Appendix A, Specifications.
W A R N I N G !
11
www.GlobalTestSupply.com
Page 9
Appendix A
Specifications
Specifications are subject to change without notice.
Velocity:
Range: 0 to 6000 ft/min (0 to 30 m/s) Accuracy
Resolution: 1 ft/min (0.01 m/s)
Duct Size:
Range: 1 to 250 inches in increments of 0.1 in. (1 to 635 cm
Volumetric Flowrate:
Range: Actual range is a function of actual velocity, pressure,
Temperature:
Range: 0 to 200°F (-18 to 93°C) Accuracy3: ±0.5°F (±0.3°C) Resolution: 0.1°F (0.1°C)
1&2
: ±3% of reading or ±3 ft/min (±0.015 m/s), whichever
is greater
in increments of 0.1 cm)
duct size, and K factor
Instrument Temperature Range:
Operating (Electronics): 40 to 113°F (5 to 45°C) Operating (Probe): 0 to 200°F (-18 to 93°C) Storage: -4 to 140°F (-20 to 60°C)
Instrument Operating Conditions:
Altitude up to 4000 meters Relative humidity up to 80% RH, non-condensing Pollution degree 1 in accordance with IEC 664 Transient over voltage category II
Data Storage Capabilities:
Range: 12,700+ samples and 100 test IDs (one sample can
contain fourteen measurement types)
Time Constant:
User-selectable
Response Time:
Velocity: 200 msec Temperature: 2 minutes (to 66% of final value)
External Meter Dimensions:
3.3 in. 7.0 in. 1.8 in. (8.4 cm 17.8 cm 4.4 cm)
www.GlobalTestSupply.com
13
Page 10
Meter Probe Dimensions:
Probe length: 40 in. (101.6 cm) Probe diameter of tip: 0.28 in. (7.0 mm) Probe diameter of base: 0.51 in. (13.0 mm)
Articulating Probe Dimensions:
Articulating section length: 6.0 in. (15.24 cm) Diameter of articulating knuckle: 0.38 in. (9.5 mm)
Meter Weight:
Weight with batteries: 0.6 lbs (0.27 kg)
Power Requirements:
Four AA-size batteries (included) or AC adapter 9 VDC, 300 mA min.
1
Temperature compensated over an air temperature range of 40 to 150°F (5 to
65°C).
2
The accuracy statement of ±3.0% of reading or ±3 ft/min (±0.015 m/s), whichever is
greater, begins at 30 ft/min through 6000 ft/min (0.15 m/s through 30 m/s).
3
Accuracy with instrument case at 77°F (25°C), add uncertainty of 0.05°F/°F
(0.03°C/°C) for change in instrument temperature.
14 Appendix A
www.GlobalTestSupply.com
Loading...