TSI Incorporated / 2007-2016 / All rights reserved.
LIMITATION OF WARRANTY AND LIABILITY (effective February 2015)
(For country-specific terms and conditions outside of the USA,)
Seller warrants the goods, excluding software, sold hereunder, under normal use and
service as described in the operator's manual, to be free from defects in workmanship and
material for 24 months, or if less, the length of time specified in the operator's manual, from
the date of shipment to the customer. This warranty period is inclusive of any statutory
warranty. This limited warranty is subject to the following exclusions and exceptions:
a. Hot-wire or hot-film sensors used with research anemometers, and certain other
components when indicated in specifications, are warranted for 90 days from the date
of shipment;
b. Pumps are warranted for hours of operation as set forth in product or operator’s
manuals;
c. Parts repaired or replaced as a result of repair services are warranted to be free from
defects in workmanship and material, under normal use, for 90 days from the date of
shipment;
d. Seller does not provide any warranty on finished goods manufactured by others or on
any fuses, batteries or other consumable materials. Only the original manufacturer's
warranty applies;
e. This warranty does not cover calibration requirements, and seller warrants only that the
instrument or product is properly calibrated at the time of its manufacture. Instruments
returned for calibration are not covered by this warranty;
f. This warranty is VOID if the instrument is opened by anyone other than a factory
authorized service center with the one exception where requirements set forth in the
manual allow an operator to replace consumables or perform recommended cleaning;
g. This warranty is VOID if the product has been misused, neglected, subjected to
accidental or intentional damage, or is not properly installed, maintained, or cleaned
according to the requirements of the manual. Unless specifically authorized in a
separate writing by Seller, Seller makes no warranty with respect to, and shall have no
liability in connection with, goods which are incorporated into other products or
equipment, or which are modified by any person other than Seller.
The foregoing is IN LIEU OF all other warranties and is subject to the LIMITATIONS stated
herein. NO OTHER EXPRESS OR IMPLIED WARRANTY OF FITNESS FOR
PARTICULAR PURPOSE OR MERCHANTABILITY IS MADE. WITH RESPECT TO
SELLER’S BREACH OF THE IMPLIED WARRANTY AGAINST INFRINGEMENT, SAID
WARRANTY IS LIMITED TO CLAIMS OF DIRECT INFRINGEMENT AND EXCLUDES
CLAIMS OF CONTRIBUTORY OR INDUCED INFRINGEMENTS. BUYER’S EXCLUSIVE
REMEDY SHALL BE THE RETURN OF THE PURCHASE PRICE DISCOUNTED FOR
REASONABLE WEAR AND TEAR OR AT SELLER’S OPTION REPLACEMENT OF THE
GOODS WITH NON-INFRINGING GOODS.
TO THE EXTENT PERMITTED BY LAW, THE EXCLUSIVE REMEDY OF THE USER OR
BUYER, AND THE LIMIT OF SELLER'S LIABILITY FOR ANY AND ALL LOSSES,
INJURIES, OR DAMAGES CONCERNING THE GOODS (INCLUDING CLAIMS BASED
ON CONTRACT, NEGLIGENCE, TORT, STRICT LIABILITY OR OTHERWISE) SHALL BE
THE RETURN OF GOODS TO SELLER AND THE REFUND OF THE PURCHASE PRICE,
OR, AT THE OPTION OF SELLER, THE REPAIR OR REPLACEMENT OF THE GOODS.
IN THE CASE OF SOFTWARE, SELLER WILL REPAIR OR REPLACE DEFECTIVE
SOFTWARE OR IF UNABLE TO DO SO, WILL REFUND THE PURCHASE PRICE OF
THE SOFTWARE. IN NO EVENT SHALL SELLER BE LIABLE FOR LOST PROFITS,
BUSINESS INTERRUPTION, OR ANY SPECIAL, INDIRECT, CONSEQUENTIAL OR
INCIDENTAL DAMAGES. SELLER SHALL NOT BE RESPONSIBLE FOR INSTALLATION,
DISMANTLING OR REINSTALLATION COSTS OR CHARGES. No Action, regardless of
form, may be brought against Seller more than 12 months after a cause of action has
accrued. The goods returned under warranty to Seller's factory shall be at Buyer's risk of
loss, and will be returned, if at all, at Seller's risk of loss.
Buyer and all users are deemed to have accepted this LIMITATION OF WARRANTY AND
LIABILITY, which contains the complete and exclusive limited warranty of Seller. This
LIMITATION OF WARRANTY AND LIABILITY may not be amended, modified or its terms
waived, except by writing signed by an Officer of Seller.
Service Policy
Knowing that inoperative or defective instruments are as detrimental to TSI as they are
to our customers, our service policy is designed to give prompt attention to any
problems.
Carefully unpack the instrument and accessories from the shipping
container. Check the individual parts against the list of components. If
anything is missing or damaged, notify TSI immediately.
For temperature measurements, make sure that at least 3 inches
(7.5 cm) of the probe is in the flow to allow the temperature sensor to
be in the air stream.
Chapter 2
Setting-up
Supplying Power to the Model 9535/9535-A
The TSI Model9535/9535-AVelociCalcAir Velocity Meteris powered
with four size AA batteries.
Installing the Batteries
Insert four AA batteries as indicated by the diagram located on
the inside of the battery compartment. The Model 9535/9535-A is
designed to operate with either alkaline or NiMH rechargeable
batteries, although it will not recharge NiMH batteries. Battery life
will be shorter if NiMH batteries are used. Carbon-zinc batteries
are not recommended because of the danger of battery acid
leakage.
Using the Optional AC Adapter
When using the AC adapter, the batteries (if installed) will be
bypassed. Be sure to provide the correct voltage and frequency,
which is marked on the back of the AC adapter. The AC adapter
will not recharge the batteries.
Using the Telescoping Probe
The telescoping probe contains the velocity, temperature, and
humidity sensors. When using the probe, make sure the sensor
window is fully exposed and the orientation dimple is facing upstream.
Extending the Probe
To extend the probe, hold the handle in one hand while pulling on
the probe tip with the other hand. Do not hold the cable while
extending the probe as this prevents the probe from extending.
Retracting the Probe
To retract the probe, hold the handle in one hand while pulling
gently on the probe cable with the other hand.
Use the Computer Interface USB Cable provided with the Model
9535/9535-A to connect the instrument to a computer for
downloading stored data.
For more information on how to download stored data see Chapter 3
section titled
C a u t i o n
This symbol is used to indicate that the data port of the Model
9535/9535-A is not intended for connection to a public
telecommunications network. Connect the USB data port only to
another USB port.
Press to turn the Model 9535/9535-A on and off.
During the power up sequence the display will
show the following: Model Number, Serial Number,
Software Revision and Last Date Calibrated.
Arrow () Keys
Press to scroll through choices while setting a
parameter.
(Enter) Key
Press to accept a value or condition.
Arrow (or ) and
Menu Soft Keys
Press arrow keys to change choices while setting a
parameter. Press the Menu soft key to select the
Menu selections, which are Display Setup,
Settings, Flow Setup, Actual/Std Set up, Data
Logging, and Calibration.
Sample
Consists of all of the measurement parameters
stored at the same time.
Test ID
A group of samples. The statistics (average,
minimum, maximum, and count) are calculated for
each test ID. The maximum number of test IDs is
100.
Time Constant
The time constant is an averaging period. It is used to
dampen the display. If you are experiencing
fluctuating flows, a longer time constant will slow
down those fluctuations. The display will update every
second, but the displayed reading will be the average
over the last time constant period. For example, if the
time constant is 10 seconds, the display will update
every second, but the displayed reading will be the
average from the last 10 seconds. This is also
referred to as a “moving average”.
Chapter 3
Operation
Keypad Functions
Common Terms
In this manual there are several terms that are used in different
places. The following is a brief explanation of the meanings of those
terms.
The horn numbers are the models of the horns. For example, 100
refers to a horn model number AM 100. Only horns with Model
numbers as follows can be used with this function: AM 100, AM 300,
AM 600, and AM 1200. If a horn model number is chosen, the
instrument will return to measuring mode and use a preprogrammed
curve to calculate flow rate from velocity.
DISPLAY SETUP
Display setup menu is where you will setup the desired parameters to
be displayed on the running screen. With a parameter highlighted you
can then use the ON soft key to have it show up on the running
screen or select the OFF soft key to turn off the parameter. Use
PRIMARY soft key to have a parameter show up on the running
screen in a larger display. Only one parameter can be selected as a
primary, and up to 2 secondary parameters can be selected at one
time.
SETTINGS
Settings menu is where you can set the general settings. These
include Language, Beeper, Select Units, Time Constant, Contrast,
Set Time, Set Date, Time Format, Date Format, Number Format,
Backlight and Auto Off. Use the or soft keys to adjust the
settings for each option and use the key to accept settings.
FLOW SET UP
In Flow Setup mode, there are 4 types: Round Duct, Rectangle Duct,
Duct Area and Horn. Use the or soft keys to scroll through the
types and then press the key to accept the desired type. To
change the value, highlight the Enter Settings option and press the
key. Use the instructions below to change the value of the type that
you selected.
ACTUAL/STANDARD SETUP
Choose Actual/Standard measurements and parameters in the
Act/Std Setup menu. Within this menu, the user can also select
Standard Temperature, Standard Pressure and a source for the
actual temperature. The actual barometric pressure must be entered
to convert air velocity and volume measurements to actual conditions.
Measurements to be logged are independent of measurements
on the display, and must therefore be selected under DATA
LOGGING Measurements.
Delete Data
Use this to delete all data, delete test or delete sample.
% Memory
This option displays the memory available. Delete All, under
Delete Data, will clear memory and reset the memory available.
LogDat2™ Downloading Software
The TSI Model 9535/9535-A comes with special software called
LogDat2 Downloading Software, which is designed to provide you
with maximum flexibility and power. To install this software on your
computer, follow the instructions on the label of the LogDat2 CDROM.
To download data from the Model 9535/9535-A, connect the supplied
computer interface USB cable to the Model 9535/9535-A and to a
computer USB port. Then run the LogDat2 software. Within the
LogDat2 software, either select the tests to be downloaded or doubleclick on a test to open it.
The Model 9535/9535-A requires very little maintenance to keep it
performing well.
Recalibration
To maintain a high degree of accuracy in your measurements, we
recommend that you return your Model 9535/9535-A to TSI for annual
recalibration. Please contact one of TSI’s offices or your local
distributor to make service arrangements and to receive a Return
Material Authorization (RMA) number. To fill out an online RMA form,
visit
The Model 9535/9535-A can also be recalibrated in the field using the
CALIBRATION menu. These field adjustments are intended to make
minor changes in calibration to match a user's calibration standards.
The field adjustment is NOT intended as a complete calibration
capability. For complete, multiple-point calibration and certification,
the instrument must be returned to the factory.
Cases
If the instrument case or storage case needs cleaning, wipe it off with
a soft cloth and isopropyl alcohol or a mild detergent. Never immerse
the Model 9535/9535-A. If the enclosure of the Model 9535/9535-A or
the AC adapter becomes broken, it must be replaced immediately to
prevent access to hazardous voltage.
Storage
Remove the batteries when storing the unit for more than one month
to prevent damage due to battery leakage.
Table 5-1 lists the symptoms, possible causes, and recommended
solutions for common problems encountered with the Model
9535/9535-A. If your symptom is not listed, or if none of the solutions
solves your problem, please contact TSI.
Table 5-1: Troubleshooting the Model 9535/9535-A
Symptom Possible Causes Corrective Action
No Display Unit not turned on Switch unit on.
Low or dead batteries Replace batteries or plug
in AC adapter.
Dirty battery contacts Clean the battery
contacts.
Velocity reading
fluctuates unstable
No response to
keypad
Instrument Error
message appears
Fault in instrument Factory service required
Fluctuating flow Reposition probe in
less-turbulent flow or use
longer time constant.
Keypad locked out Unlock keypad by
pressing keys
simultaneously.
Memory is full Download data if desired,
then DELETE ALL
memory.
on instrument.
W A R N I N G !
Remove the probe from excessive temperature immediately:
excessive heat can damage the sensor. Operating temperature limits
can be found in
Specifications are subject to change without notice.
Velocity:
Range: 0 to 6000 ft/min (0 to 30 m/s)
Accuracy
Resolution: 1 ft/min (0.01 m/s)
Duct Size:
Range: 1 to 250 inches in increments of 0.1 in. (1 to 635 cm
Volumetric Flowrate:
Range: Actual range is a function of actual velocity, pressure,
Temperature:
Range: 0 to 200°F (-18 to 93°C)
Accuracy3: ±0.5°F (±0.3°C)
Resolution: 0.1°F (0.1°C)
1&2
: ±3% of reading or ±3 ft/min (±0.015 m/s), whichever
is greater
in increments of 0.1 cm)
duct size, and K factor
Instrument Temperature Range:
Operating (Electronics): 40 to 113°F (5 to 45°C)
Operating (Probe): 0 to 200°F (-18 to 93°C)
Storage: -4 to 140°F (-20 to 60°C)
Instrument Operating Conditions:
Altitude up to 4000 meters
Relative humidity up to 80% RH, non-condensing
Pollution degree 1 in accordance with IEC 664
Transient over voltage category II
Data Storage Capabilities:
Range: 12,700+ samples and 100 test IDs (one sample can
contain fourteen measurement types)
Time Constant:
User-selectable
Response Time:
Velocity: 200 msec
Temperature: 2 minutes (to 66% of final value)
External Meter Dimensions:
3.3 in. 7.0 in. 1.8 in. (8.4 cm 17.8 cm 4.4 cm)