When using the TRANSDUCER DISPLAY & MONITOR with the
TSI C
ERTIFIER
®
Air Velocity Calibrator, the unit will display
actual air velocity. Entering a density correction factor
automatically converts the velocity to standard conditions.
This allows the user to read air velocity directly from the
Transducer Display & Monitor instead of reading pressure
and then calculating velocity.
TSI instruments are backed by more than 38 years of experience in air flow measurement technology. It’s this type of
experience and innovation that provides you with accurate
and reliable instruments. Along with TSI’s expertise, each
instrument is backed by a two year limited
warranty and a great service policy. In addition to fast service,
calibrations are NIST* traceable and a free certificate of calibration is included.
P/N 2980075 Rev.B
Copyright ©2004 by TSI Incorporated Printed in U.S.A.
TSI Incorporated
Corporate Headquarters—Tel: 651 490 2811 Toll Free: 1 800 874 2811 Fax: 651 490 3824 E-mail: answers@tsi.com
Contact TSI or visit www.tsi.com for information on specific office locations worldwide.
Specifications
Model 8495
Full Scale Value Resolution
0.1000 to 1.0000 0.0001
1.000 to 10.000 0.001
10.00 to 100.00 0.01
100.0 to 1000.0 0.1
1000 to 10000 1
10010 to 99990 10
Accuracy
(1&2)
±.25% of reading ±1 digit of resolution
Time Constant 1 to 20 seconds in 1 second intervals
Transducer Connections Detachable terminal strip
Display 5 digit LED
Output Power 12 VDC unregulated, supplies 500 mA
minimum with TSI-supplied AC Adapter
Signal Input Ranges 0-5VDC, 1-5VDC, 0-10VDC, 2-10VDC,
0-20mA, or 4-20mA
Temperature Stability 0.0014%/°F (0.003%/°C)
Temperature Range 32 to 122°F (0 to 50°C)
Input Impedance Greater than 500KΩ for 0-5 V input,
greater than 1MΩ for 0-10 V input, 250Ω
for current input
Input Power 11-30 VDC or 18-28 VAC, requires 500
mA maximum
Power Consumption 19VA maximum
Housing ABS plastic
Dimensions 8.0 in.
×
6.2 in.× 2.5 in.
(20.3 cm
×
15.8 cm× 6.5 cm)
* National Institute of Standards and Technology.
1
±.02% of reading/°C = .01% of reading/°F away from 25°C.
2
±.025% of full scale = effect of EMI per IEC801-3.
Specifications subject to change without notice.
Specifications in parentheses ( ) indicate metric equivalents.