TSI Alnor 8392 Data sheet

Ventilation Testing
Certifier
®
Air Velocity Calibrator
TSI’s Model 8392 CERTIFIER®Air Velocity Calibrator gives you reliable and convenient in­house calibrations at a reasonable cost. You can have the confidence of high accuracies over a wide velocity range with the C
ERTIFIER. It enables you to
calibrate your probed air velocity meters or transduc­ers at velocities as low as 30 ft/min (0.15 m/s) and as high as 9000 ft/min (45 m/s) with typical accuracies of ±1 percent of reading.
The C
ERTIFIER is ideal whenever frequent,
accurate, air velocity calibrations are required. Having a C
ERTIFIER on site avoids the delays asso-
ciated with sending the instrument to an outside calibration facility. Calibrations can be performed at your convenience as often as necessary.
The C
ERTIFIER comes with a TSI Model 8495
T
RANSDUCER DISPLAY & MONITOR that will dis-
play actual air velocity. Entering a density correc­tion factor automatically converts the velocity to standard conditions. This allows the user to read air velocity directly from the T
RANSDUCER DISPLAY
& MONITOR instead of reading pressure and then calculating velocity.
The Model
8392 C
ERTIFIER Air
Velocity Calibrator provides a complete system to meet all of your calibration needs. Standard accessories furnished with the system include:
Model 8495 TRANSDUCER DISPLAY & MONITOR
MKS Pressure Transducer
Fittings which accept 0.188 in. (4.8 mm) through 0.625 in. (15.9 mm) diameter probes (other sizes can be accommodated with the prop­er
1
/2 NPT male compression fitting)
Controller for setting fan speed
Two nozzle plates for establishing low velocities
P/N 2980015 Rev. D . 3/05 Copyright ©2005 by TSI Incorporated
TSI Incorporated
United States: Tel : 651 490 2811 Toll Free: 1 800 874 2811 Fax: 651 490 3824 E-mail: answers@tsi.com Germany: Tel : +49-241 523030 Fax: +49-241 523 0349 E-mail: tsigmbh@tsi.com Sweden: Tel : +46-8-595-132 30 Fax: +46-8-595-132 39 E-mail: tsiab@tsi.com United Kingdom: Tel : +44-1275-847837 Fax: +44-1275-842437 E-mail: tsiuk@tsi.com
or visit www.tsi.com to find your closest TSI Representative or distributor.
Test Section Dimensions: 4 in.×4 in. (10.2 cm×10.2 cm) Feedthrough Hole: 1/2” NPT Overall Dimensions: 57 in. L
×
17 in. W×17 in. H
(145 cm
×
43 cm×43 cm)
Weight: 30 lbs (14.6 kg) Power: 115 or 230 VAC
(specify when ordering), 50/60 Hz
MKS Pressure Transducer Range: 0 to 10 mmHg Accuracy: 0.15% of reading Power: 115/230 VAC, 50/60 Hz switch selectable
Specifications
Model 8392 CERTIFIER Air Velocity Calibrator
Inlet Section
Inlet Pressure Tap
Orifice (Nozzle)
Test Section
Probe Fitting
Test Section Pressure Tap
Diffuser Section
Specifications
Model 8495 Transducer Display & Monitor
Accuracy(1&2) ±0.25% of reading, ±1 digit of resolu­tion
Time Constant 1 to 20 seconds in 1 second intervals Display 5 digit LED Input Power 11-30 VDC or 18-28 VAC, requires
500mA maximum
1
±.02% of reading/°C = .01% of reading /°F away from 25°C.
2
±.025% of full scale= effect of EMI per IEC801-3.
Specifications subject to change without notice. Specifications in parentheses ( ) indicate metric equivalents.
Nozzle Plate
None
No. 1
1500-9000
250-1500
No. 2
Velocity Range
(ft/min) (m/s)
7.5-45
1.25-7.5
30-250
0.15-1.25
1.0% of reading ±15 ft/min
2.0% of reading ±5 ft/min
2.0% of reading ±2 ft/min
Typical
Accuracy
(±.075 m/s)
(±.025 m/s)
(±.01 m/s)
Pressure Drop
(mmHg)
0.25-10.0
0.2-8.5
0.11-8.5
Turbulence
Intensity*
0.25%
1.0%
1.0%
Loading...