TSI Alnor 8375 Data sheet

.
ACCUBALANCE
®
Capture Hood
Model 8375
The TSI 8375 Capture Hood is a multipurpose electronic air
balancing instrument for reading air volume flow at diffusers,
and grilles. It is ideally suited for commissioning agents,
facilities managers, health and safety specialists, test engineers,
and test and balance professionals. This ultra light weight,
ergonomically designed kit saves time and money while helping
to create a healthy and energy efficient environment.
Optional Accessories
Pitot tubes
16-point velocity matrix with telescoping handle
Air flow probe
Temperature probe
Temperature/humidity probe
Multiple hood sizes available
Bio-safety cabinet hood kit
Ventilation Test Instruments
Features and Benefits
Ergonomic design and ultra light weight for easy one person operation
Detachable digital manometer for use in other applications with Pitot, air flow, temperature, velocity matrix, or relative humidity probes
Back pressure compensated
Multiple hood sizes available
Bio-Safety hood kit available
ENERGY AND COMFORT
Model 8375
TRUST. SCIENCE. INNOVATION
ENERGY AND COMFORT
Specifications
ACCUBALANCE Model 8375
Volume
Range 25 to 2,500 ft3/min (42 to 4250 m3/h) capture
hood
Accuracy ±3% of reading ±7 ft
3
/min (±12 m3/h) at flows
>50 ft
3
/min (>85 m3/h)
Units ft
3
/min, m3/h, m3/min, l/s
Resolution 1 ft
3
/min (1 m3/h)
Velocity
Range (Pitot Probes)
25 to 8,000 ft/min (0.125 to 40 m/s)
Range (Airflow Probe)
25 to 5,000 ft/min (0.125 to 25 m/s)
Velocity Matrix 25 to 2,500 ft/min (0.125 to 12.5 m/s) Accuracy ±3% of reading ±7 ft/min (±0.04 m/s) at
velocities >50 ft/min (0.25 m/s)
Units ft/min, m/s Resolution 1 ft/min (0.01 m/s)
Pressure
Differential Pressure
±15 in. H
2
O (±3735 Pa ); 150 in. H2O
(37.5 kPa) maximum safe operating pressure
Absolute Pressure 15 to 40 in. Hg (356 to 1016 mm Hg) Accuracy ±2% of reading ±0.001 in. H
2
O (±0.25 Pa ) static and differential; ±2% of reading absolute
Units in. H
2
O, in. Hg, Pa, hPa, kPa, mm Hg, cm Hg,
mm H
2
O, cm H2O
Resolution 0.00001 in. H
2
O (0.001 Pa) static and
differential; 0.01 in. Hg (1 mm Hg) absolute
RH
Range 0 to 95% RH temperature/RH probe Accuracy ±3% RH Resolution 0.1% RH
Temperature
Sensor in Base 40 to 140°F (4.4 to 60°C) Temperature Probe -40 to 250°F (-40 to 121°C) Temperature/RH Probe
14 to 140°F (-10 to 60°C)
Accuracy ±0.5 °F (±0.3°C) from 32 to 160°F
(0 to 71°C)
Units °F, °C Resolution 0.1°F (0.1°C)
Specifications are subject to change without notice.
Ventilation Test Instruments
TSI Incorporated
- 500 Cardigan Road, Shoreview, MN 55126-3996 USA
USA Tel: +1 800 874 2811 E-mail: info@tsi.com Website: www.tsi.com UK Tel: +44 149 4 459200 E-mail: tsiuk@tsi.com Website: www.tsiinc.co.uk France Tel: +33 491 95 21 90 E-mail: tsifrance@tsi.com Website: www.tsiinc.fr Germany Tel: +49 241 523030 E-mail: tsigmbh@tsi.com Website: www.tsiinc.de Sweden Tel: +46 8 595 13230 E-mail: tsiab@tsi.com Website: www.tsi.se India Tel: +91 80 41132470 E-mail: tsi-india@tsi.com China Tel: +86 10 8260 1595 E-mail: tsibeijing@tsi.com
Contact your local TSI Distributor or visit our website
wwwwww..ttssii..ccoomm
for more detailed specifications.
P/N 5001003 Rev A Copyright © 2007 by TSI Incorporated Printed in U.S.A.
.