Copyright
TSI Incorporated / May 2007-2016 / All rights reserved.
Address
TSI Incorporated / 500 Cardigan Road / Shoreview, MN 55126 / USA
Fax No.
(651) 490-3824
LIMITATION OF WARRANTY AND LIABILITY (effective February 2015)
Seller warrants the goods, excluding software, sold hereunder, under normal use and
service as described in the operator's manual, to be free from defects in workmanship
and material for twenty-four (24) months, or the length of time specified in the
operator's manual, from the date of shipment to the customer. This warranty period is
inclusive of any statutory warranty. This limited warranty is subject to the following
exclusions:
a. Hot-wire or hot-film sensors used with research anemometers, and certain other
components when indicated in specifications, are warranted for 90 days from the
date of shipment.
b. Parts repaired or replaced as a result of repair services are warranted to be free
from defects in workmanship and material, under normal use, for 90 days from the
date of shipment.
c. Seller does not provide any warranty on finished goods manufactured by others or
on any fuses, batteries or other consumable materials. Only the original
manufacturer's warranty applies.
d. This warranty does not cover calibration requirements, and seller warrants only
that the instrument or product is properly calibrated at the time of its manufacture.
Instruments returned for calibration are not covered by this warranty;
f. This warranty is VOID if the instrument is opened by anyone other than a factory
authorized service center with the one exception where requirements set forth in
the manual allow an operator to replace consumables or perform recommended
cleaning;
g. This warranty is VOID if the product has been misused, neglected, subjected to
accidental or intentional damage, or is not properly installed, maintained, or
cleaned according to the requirements of the manual. Unless specifically
authorized in a separate writing by Seller, Seller makes no warranty with respect
to, and shall have no liability in connection with, goods which are incorporated into
other products or equipment, or which are modified by any person other than
Seller.
The foregoing is IN LIEU OF all other warranties and is subject to the LIMITATIONS
stated herein. NO OTHER EXPRESS OR IMPLIED WARRANTY OF FITNESS FOR
PARTICULAR PURPOSE OR MERCHANTABILITY IS MADE.
TO THE EXTENT PERMITTED BY LAW, THE EXCLUSIVE REMEDY OF THE USER
OR BUYER, AND THE LIMIT OF SELLER'S LIABILITY FOR ANY AND ALL LOSSES,
INJURIES, OR DAMAGES CONCERNING THE GOODS (INCLUDING CLAIMS
BASED ON CONTRACT, NEGLIGENCE, TORT, STRICT LIABILITY OR
OTHERWISE) SHALL BE THE RETURN OF GOODS TO SELLER AND THE REFUND
OF THE PURCHASE PRICE, OR, AT THE OPTION OF SELLER, THE REPAIR OR
REPLACEMENT OF THE GOODS. IN THE CASE OF SOFTWARE, SELLER WILL
REPAIR OR REPLACE DEFECTIVE SOFTWARE OR IF UNABLE TO DO SO, WILL
REFUND THE PURCHASE PRICE OF THE SOFTWARE. IN NO EVENT SHALL
SELLER BE LIABLE FOR LOST PROFITS, BUSINESS INTERRUPTION, OR ANY
SPECIAL, INDIRECT, CONSEQUENTIAL OR INCIDENTAL DAMAGES. SELLER
SHALL NOT BE RESPONSIBLE FOR INSTALLATION, DISMANTLING OR
REINSTALLATION COSTS OR CHARGES. No Action, regardless of form, may be
brought against Seller more than 12 months after a cause of action has accrued. The
goods returned under warranty to Seller's factory shall be at Buyer's risk of loss, and
will be returned, if at all, at Seller's risk of loss.
www.GlobalTestSupply.com

Buyer and all users are deemed to have accepted this LIMITATION OF WARRANTY
AND LIABILITY, which contains the complete and exclusive limited warranty of Seller.
This LIMITATION OF WARRANTY AND LIABILITY may not be amended, modified or
its terms waived, except by writing signed by an Officer of Seller.
Service Policy
Knowing that inoperative or defective instruments are as detrimental to TSI as they are
to our customers, our service policy is designed to give prompt attention to any
problems. If any malfunction is discovered, please contact your nearest sales office or
representative, or call Customer Service department at (800) 874-2811 (USA) or (1)
651-490-2811 (International).
www.GlobalTestSupply.com

CONTENTS
CHAPTER 1 UNPACKING AND PARTS IDENTIFICATION ............ 1
CHAPTER 2 SETTING-UP ................................................................. 3
Supplying Power to the Model 7515 ............................................... 3
Installing the Batteries ............................................................... 3
Using the Integral Probe ............................................................ 3
CHAPTER 3 OPERATION ................................................................. 5
Keypad Functions ........................................................................... 5
CHAPTER 4 MAINTENANCE ............................................................ 7
Recalibration ................................................................................... 7
Cases .............................................................................................. 7
Storage ........................................................................................... 7
CHAPTER 5 TROUBLESHOOTING .................................................. 9
APPENDIX A SPECIFICATIONS ..................................................... 11
www.GlobalTestSupply.com
iii

Chapter 1
Unpacking and Parts Identification
Carefully unpack the instrument and accessories from the shipping
container. Check the individual parts against the list of components
below. If anything is missing or damaged, notify TSI immediately.
1. Carrying case
2. Instrument
3. Calibration collar
www.GlobalTestSupply.com
1

Chapter 2
Setting-up
Supplying Power to the Model 7515 IAQ-CALC Indoor Air Quality
Meter
The Model 7515 is powered with four size AA batteries.
Installing the Batteries
Insert four AA batteries as indicated by the diagram located on
the inside of the battery compartment. The Model 7515 is
designed to operate with either alkaline or NiMH rechargeable
batteries, although it will not recharge NiMH batteries. Battery life
will be shorter if NiMH batteries are used. Carbon-zinc batteries
are not recommended because of the danger of battery acid
leakage.
Using the Integral Probe
The sensing probe relies on the diffusion of air. For best results,
try to keep the sensing probe surrounded by moving air. Do not
breathe on the probe. Humans exhale CO2 levels exceeding
10,000 ppm and it may take time for the probe to re-stabilize.
www.GlobalTestSupply.com
3

Press to turn the Model 7515 on and off.
During the power up sequence the display will
show the following: Model Number, Serial
Number, Software Revision, and Last Date
Calibrated.
Press and release to start the sample, then
press the END key to stop the sample.
Press to view the minimum, maximum and
average of the last sample taken.
Press to begin the field-calibration process.
To re-calibrate in the field, first connect zero
gas to the instrument and open the regulator.
The instrument will automatically make zero
itself after one minute.
After zeroing the sensor, connect span gas to
the instrument and open the regulator. The
instrument will take one minute to make a
reading. Then use the and ENTER keys
to adjust the concentration displayed on the
instrument to match the concentration on the
gas cylinder.
Chapter 3
Operation
Keypad Functions
www.GlobalTestSupply.com
5

Chapter 4
Maintenance
The Model 7515 requires very little maintenance to keep it performing
well.
Recalibration
To maintain a high degree of accuracy in your measurements, we
recommend that you return your Model 7515 to TSI for annual
recalibration. Please contact one of TSI’s offices or your local
distributor to make service arrangements and to receive a Return
Material Authorization (RMA) number. To fill out an online RMA form,
visit TS
The Model 7515 can also be recalibrated in the field using the
CALIBRATION menu. These field adjustments are intended to make
minor changes in calibration to match a user's calibration standards.
The field adjustment is NOT intended as a complete calibration
capability. For complete, multiple-point calibration and certification,
the instrument must be returned to the factory.
Cases
If the instrument case or storage case needs cleaning, wipe it off with
a soft cloth and isopropyl alcohol or a mild detergent. Never immerse
the Model 7515. If the enclosure of the Model 7515, it must be
replaced immediately to prevent access to hazardous voltage.
Storage
Remove the batteries when storing the unit for more than one month
to prevent damage due to battery leakage.
7
www.GlobalTestSupply.com

Clean the battery
contacts.
Shield probe from
breath and allow
instrument to restabilize.
Instrument Error
message appears
Factory service
required on instrument.
Chapter 5
Troubleshooting
Table 5-1 lists the symptoms, possible causes, and recommended
solutions for common problems encountered with the Model 7515. If
your symptom is not listed, or if none of the solutions solves your
problem, please contact TSI.
Table 5-1: Troubleshooting the Model 7515
Remove the probe from excessive temperature immediately:
excessive heat can damage the sensor. Operating temperature limits
can be found in Appendix A, Specifications.
WARNING!
9
www.GlobalTestSupply.com

Appendix A
Specifications
Specifications are subject to change without notice.
CO2:
Range: 0 to 5000 ppm
Accuracy1: ±3% of reading or ±50 ppm, whichever is greater
Resolution: 1 ppm
Sensor type: Non-Dispersive Infrared (NDIR)
Instrument Temperature Range:
Operating (Electronics): 40 to 113°F (5 to 45°C)
Storage: -4 to 140°F (-20 to 60°C)
Instrument Operating Conditions:
Altitude up to 4000 meters
Relative humidity up to 80% RH, non-condensing
Pollution degree 1 in accordance with IEC 664
Transient over voltage category II
External Meter Dimensions:
3.3 in. 9.6 in. 1.8 in. (8.4 cm 24.4 cm 4.4 cm)
Meter Weight:
Weight with batteries: 0.6 lbs (0.27 kg)
Meter Display Dimensions:
Primary display: 4-digit LCD, 0.6 in. (15 mm) digit height
Secondary display: 3.5-digit LCD, 0.3 in. (8 mm) digit height
Power Requirements:
Four AA-size batteries (included)
1
At 77°F (25°C). Add uncertainty of ±0.2%/°F (±0.2%/°C) away from calibrated
temperature