TSI Alnor 6203 Operating Manual

Page 1
ENERGY AND COMFORT
Combustion Analysis
CA-CALC
Combustion Analyzers
Model 6203 Series
Operation and Service Manual
August 2010
Page 2
Page 3
3
Copyright © TSI Incorporated / 2002–2010 / All rights reserved.
Address
TSI Incorporated / 500 Cardigan Road / Shoreview, MN 55126 / USA
Fax No.
(651) 490-3824
LIMITATION OF WARRANTY AND LIABILITY (effective July 2000)
Seller warrants the goods sold hereunder, under normal use and service as described in the operator's manual, shall be free from defects in workmanship and material for twenty-four (24) months, or the length of time specified in the operator's manual, from the date of shipment to the customer. This warranty period is inclusive of any statutory warranty. This limited warranty is subject to the following exclusions:
a. Batteries are warranted for 90 days from the date of shipment to the customer.
Electrochemical gas sensors are warranted for a period of twelve (12) months from the date of shipment to the customer.
b. Parts repaired or replaced as a result of repair services are warranted to be free from defects
in workmanship and material, under normal use, for 90 day s from the date of shipment.
c. Seller does not provide any warranty on finished goods manufactured by others or on any
fuses, batteries or other consumable materials. Only the original manufacturer's warranty applies.
d. Unless specifically authorized in a separate writing by Seller, Seller makes no warranty with
respect to, and shall have no liability in connection with, goods which are incorporated into other products or equipment, or which are modified by any person other than Seller.
The foregoing is IN LIEU OF all other warranties and is subject to the LIMITATIONS stated herein. NO OTHER EXPRESS OR IMPLIED WARRANTY OF FITNESS FOR PARTICULAR PURPOSE OR MERCHANTABILITY IS MADE.
TO THE EXTENT PERMITTED BY LAW, THE EXCLUSIVE REMEDY OF THE USER OR BUYER, AND THE LIMIT OF SELLER'S LIABILITY FOR ANY AND ALL LOSSES, INJURIES, OR DAMAGES CONCERNING THE GOODS (INCLUDING CLAIMS BASED ON CONTRACT, NEGLIGENCE, TORT, STRICT LIABILITY OR OTHERWISE) SHALL BE THE RETURN OF GOODS TO SELLER AND THE REFUND OF THE PURCHASE PRICE, OR, AT THE OPTION OF SELLER, THE REPAIR OR REPLACEMENT OF THE GOODS. IN NO EVENT SHALL SELLER BE LIABLE FOR ANY SPECIAL, CONSEQUENTIAL OR INCIDENTAL DAMAGES. SELLER SHALL NOT BE RESPONSIBLE FOR INSTALLATION, DISMANTLING OR REINSTALLATION COSTS OR CHARGES. No Action, regardless of form, may be brought against Seller more than 12 months after a cause of action has accrued. The goods returned under warranty to Seller's factory shall be at Buyer's risk of loss, and will be returned, if at all, at Seller's risk of loss.
Buyer and all users are deemed to have accepted this LIMITATION OF WARRANTY AND LIABILITY, which contains the complete and exclusive limited warranty of Seller. This LIMITATION OF WARRANTY AND LIABILITY may not be amended, modified or its terms waived, except by writing signed by an Officer of Seller.
Service Policy
Knowing that inoperative or defective instruments are as detrimental to TSI as they are to our customers, our service policy is designed to give prompt attention to any problems. If any malfunction is discovered, please contact your nearest sales office or representative, or call TSI's Customer Service department at (800) 874-2811 or 651-490-2811.
Page 4
4
(This page intentionally left blank)
Page 5
i

Contents

Introduction ........................................................................................ 1
Manual Purpose ............................................................................ 1
Using This Manual ......................................................................... 1
Warnings and Cautions ................................................................. 1
Chapter 1 Instrument Description .................................................. 3
CA-6203 CA-CALCTM Combustion Analyzer Measurements
and Calculations ......................................................................... 3
Chapter 2 Unpacking ....................................................................... 5
List of Standard Components ........................................................ 5
Accessories and Replacement Parts ............................................ 5
Chapter 3 Component Identification .............................................. 7
The Gas Sensors .......................................................................... 8
The Sampling Probe ...................................................................... 9
Flue Probe Thermocouple ............................................................. 9
On-Board Temperature Measurement .......................................... 9
Diaphragm Pump .......................................................................... 9
Draft Sensor .................................................................................. 9
Water Trap ..................................................................................... 9
Optional Combustion Supply Air Thermocouple Probe ................ 9
Schematic Representation of CA-CALC Analyzer ........................ 10
Chapter 4 Getting Started ............................................................. 11
Supplying Power ......................................................................... 11
Installing Batteries ....................................................................... 11
Connecting the Sampling Probe ................................................. 11
Connecting the Optional Combustion Supply Air Temperature
Probe ........................................................................................ 12
Connecting the Optional Portable Printer .................................... 12
Connecting to a Computer .......................................................... 12
Default Instrument Settings ......................................................... 13
Factory Defaults and Language Selection .................................. 13
Chapter 5 Basic Operation ............................................................ 15
Buttons and Button Operations ................................................... 15
Startup ......................................................................................... 16
Characters and Display Icons ..................................................... 18
Measurements and Calculations ................................................. 20
Data Display, Viewing Measurements and Calculations............. 22
Error Indications .......................................................................... 22
Using the Sampling Probe ........................................................... 22
Making a Draft Measurement ...................................................... 24
Page 6
ii
Printing to the Portable Printer and to a Computer ..................... 24
Handling Data .............................................................................. 27
Chapter 6 MENU Options .............................................................. 29
BASE.L MENU—Option Update Sensor Baselines .................... 30
FUEL MENU Option—Select Fuel .............................................. 30
FP MENU Option—Fuel Parameters .......................................... 31
UNITS MENU Option (changing units) ........................................ 31
INT.SA MENU Option—Sampling Interval for Saving Data ........ 31
TIME MENU Option—Set Time and Date for Printout ................ 32
CALIB MENU Option (select sensor for calibration) ................... 32
O2 Gas Sensor Calibration .......................................................... 33
Calibration of the CO (hydrogen compensated) Sensor ............. 33
Calibration Steps ......................................................................... 33
NO Gas Sensor Calibration ......................................................... 34
Draft Calibration ........................................................................... 34
Temperature Calibration .............................................................. 35
CAL.FA MENU Option—Setting the Calibration Factor for a
Replacement Gas Sensor ........................................................ 35
CO Sensor ................................................................................... 35
NO Sensor ................................................................................... 36
BAUD Rate MENU Option ........................................................... 37
COMM MENU Option—Set the Output Communications
Device ....................................................................................... 37
FP MENU Option—Fuel Parameters .......................................... 37
LANG MENU Option .................................................................... 40
Chapter 7 Setup for Gas Calibration ............................................ 41
The Calibration Setups ................................................................ 41
Chapter 8 Maintenance and Troubleshooting ............................. 45
Emptying Water Trap ................................................................... 45
Changing the Water Trap Filter ................................................... 45
Cleaning the Sample Probe ......................................................... 46
Lithium Battery Replacement ...................................................... 46
Replacing a Gas Sensor .............................................................. 47
Appendix A Error Codes ............................................................... 49
Appendix B Calculations ............................................................... 51
Undiluted Gas Concentration Calculation (COu, NOu, NOxu) ...... 51
Excess Air Calculation ................................................................. 51
Calculating Combustion Efficiency for U.S. Fuels ....................... 51
Siegert Formula ........................................................................... 52
Determining CO2 Using the O2 Concentration ............................ 53
Emission Rate Calculations Using Emission Factors .................. 53
Emission Rate Factors ................................................................ 53
Page 7
iii
A General Equation for the Combustion of a Simple
Hydrocarbon in Air ................................................................... 54
Calculating CO2 Max From the Carbon Content ......................... 54
Calculation of Combustion Air Requirement ............................... 54
Appendix C Series CA-6203 CA-CALCTM Combustion
Analyzer Specifications ............................................................ 55
TÜV Reg. No. 229
Tested and approved for Measurement of O2, CO, and Temperature.
Page 8
iv
(This page intentionally left blank)
Page 9
1
!
!

Introduction

Manual Purpose

This manual describes the operation and maintenance of TSI
ALC™ portable combustion analyzer.
CA-C

Using This Manual

®
CA-6203
Before using your CA-C its entirety.

Warnings and Cautions

The manual assumes that you have a basic understanding of combustion safety and are thoroughly familiar with the fuel burning equipment being tested. If you are using measurements as the basis for equipment adjustments, rely on your good judgment and experience together with the measured data. This is especially important for safety. Remember when adjusting equipment always follow the manufacturer’s recommendations.
ALC™ combustion analyzer, review this manual in
WARNING
High temperatures and toxic gases are produced when
fossil fuels are burned. Only qualified individuals, thoroughly familiar with operating and adjusting fuel­burning equipment, should use gas measurements for the purpose of making equipment adjustments.
Note
The CA-CALC analyzer is not intended for use as a
continuous monitor.
Notes: Best results are obtained if the CA-C
allowed to stabilize at the temperature of the test environment before using.
To reduce sensor exposure to gas and to reduce build up of water
vapor in the sampling lines and water trap, turn the pump off when not making measurements.
ALC™ combustion analyzer is
Page 10
2
(This page intentionally left blank)
Page 11
Oxygen concentration
O2, electrochemical
Carbon monoxide
CO, electrochemical
Nitric oxide
NO, electrochemical
Flue temperature
Thermocouple
Supply air temperature
On-board sensor or Thermocouple probe
Draft pressure
Pressure transducer, internal solid state
CO2, Carbon Dioxide
Excess air / λ
Stack/Flue loss / qA
Combustion efficiency
Differential temperature
NOx CO undiluted
NO undiluted
COr index
NOx undiluted
Emission Rates CO, NO
Chapter 1
Instrument Description
The Model CA-6203 CA-CALC combustion analyzer is used to evaluate the performance of burners in boilers, furnaces, and hot water tanks. The Model CA-6203 has sensors for the direct measurement of combustion gases; O Model CA-6203 performs a complete list of combustion and emission parameters for a variety of system performance evaluations. Different units and different fuel types are readily selectable.
CA-6203 CA-C Calculations
Measurement Sensor
, CO, NO; combustion gas temperatures and draft pressure. The
2
TM
ALC
Combustion Analyzer Measurements and
Calculations
The Model 6203 is supplied with a sampling probe having an in-line water trap and particulate filter.
The Model 6203 operates using either an AC power supply or C-size batteries.
3
Page 12
4
The Model 6203 stores individual data samples (up to 100), and prints the data to a portable printer or computer. Stored data can be saved over a user­defined interval and averaged.
The Model 6203 has a variety of standard fuels, and enables you to modify the fuel parameters, or install your own user-defined fuel.
Loss and efficiency are calculated from standard ASME heat-loss calculations or using the Siegert formula (refer to Appendix B
).
Chapter 1
Page 13
1
Sample probe with water trap, draft line and
801989 4
C cell alkaline batteries
1 Operation and Service manual
1980472
Combustion Supply Air thermocouple
3013003
Portable printer kit
801994
Computer cable
8940
Hard side Carrying case
1319319
CO replacement sensor
802014
O2 replacement sensor
802012
NO replacement sensor
802007
Water trap filters
801947
Calibration Kit O2 (N2 gas) and CO
802003
Calibration Kit CO
801999
Calibration Kit NO
801937
Replacement bottled gas, CO (requires Cal. Kit)
801941
Replacement bottled gas, NO (requires Cal. Kit)
801930
Replacement bottled gas, N2 (requires Cal. Kit)
801980
Bottled gas, CO with H2 (requires Cal. Kit)
801942
Chapter 2
Unpacking
Carefully unpack your CA-CALC combustion analyzer and accessories from the carrying case. Check the individual parts against the list of components in the table below. If items are missing or damaged, notify TSI immediately.

List of Standard Components

Qty. Item Component Part/ Model
1 Series CA-6203 CA-CALC combustion analyzer 1083618
temperature
1 Universal Power supply 7.2 V with assorted plug
types

Accessories and Replacement Parts

2182003
Item Part/Model
5
Page 14
6
(This page intentionally left blank)
Chapter 2
Page 15
1. Data buttons
9. Battery cover
2. LCD display
10. Battery cover tab
3. On-Off button
11. Port for flue gas thermocouple
4. Control buttons
12. Port for supply air thermocouple
5. Icon buttons
13. RS-232 serial port
6. Magnets
14. Power connection
7. Vent
15. Draft sample port
8. Sensor cover
16. Gas sample port
Chapter 3
Component Identification
Key components of the CA-CALC combustion analyzer and sampling probe are identified in Figures 1 and 2, and under section headings in the text that follows.
Figure 1: CA-6203 CA-CALCTM Combustion Analyzer
7
Page 16
8
1. Tube retaining fitting 6. Thermocouple connector
2. Probe handle 7. Flexible sample line
3. Position collar 8. Water Trap
4. Sample tube 9. Water trap filter
5. Probe tip with thermocouple
Figure 2: CA-C
ALC Sampling Probe Components

The Gas Sensors

The CA-6203 CA-C oxygen (O
), carbon monoxide (CO) and nitric oxide (NO). CO and NO
2
sensors are capable of operating from two to three years. The O
ALC™ analyzer uses three gas sensors for measuring
sensor
2
typically lasts one year. Best accuracy is obtained by routine user calibration (automatic for the O
sensor). Calibrated replacement CO and NO sensors
2
are also available from the factory.
Chapter 3
Page 17
9

The Sampling Probe

Your combustion analyzer comes equipped with a sampling probe, depicted in Figure 2. A thermocouple extends to the end of the sampling tube for making flue temperature measurements. Two flexible sampling lines are used for gas sampling and draft measurement. The sampling tube is removable for cleaning.

Flue Probe Thermocouple

The flue temperature (TS) is measured using a Type K probe, which extends through the sampling tube to the probe tip. The thermocouple probe measures temperatures up to 1000 degrees C (1800 degrees F).

On-Board Temperature Measurement

Your CA-C
ALC™ analyzer uses an on-board temperature detector to provide
the Combustion Supply Air (TA) temperature when no supply-air accessory probe is present.

Diaphragm Pump

The CA-C
ALC™ analyzer samples exhaust gases from the flue and delivers
them to the electrochemical sensors using a diaphragm sampling pump. Typical pump life is 5000 hours.

Draft Sensor

A differential pressure transducer in the CA-C
ALC™ analyzer is used to
measure draft pressure.

Water Trap

The water trap, shown in Figure 2, is used to remove moisture that collects in the sample tubing when combustion gases are sampled. The water trap has a two-chamber design with a particulate filter.

Optional Combustion Supply Air Thermocouple Probe

A measurement of the Combustion Supply Air (TA) temperature at the source can be made using an optional thermocouple accessory probe (TSI PN 3013003). This probe is connected to the supply air temperature port shown in Figure 1.
When a Supply Air temperature probe is not used, the supply air temperature is determined from the on-board temperature sensor.
Component Identification
Page 18
10

Schematic Representation of CA-CALC Analyzer

Figure 3. Schematic Representation of CA-C
ALCAnalyzer
Chapter 3
Page 19
Chapter 4
Getting Started

Supplying Power

The CA-C or an AC adapter. Quality alkaline batteries enable the instrument to operate for at least 24 hours. Use of the plug-in AC adapter conserves battery life, and can be substituted for batteries.

Installing Batteries

Turn the combustion analyzer over and remove the battery cover by lifting up and out on the battery cover tab shown below. Remove the old batteries.
Note: It is not necessary to remove the
Install four new C-cell batteries, noting the battery orientation depicted on the base of the battery holder. Install spring-side batteries first.

Connecting the Sampling Probe

The sampling probe depicted in Figure 2, is connected to the instrument by pushing the brass sample and silver draft connectors over the ports on the instrument. Refer to the figures below showing the proper connection of the probe tubes to the instrument sample ports. Note that the thermocouple connector can be inserted only one way. The thermocouple connector is oriented with the large spade to the right. Do not force the connector.
ALCportable combustion analyzer operates using 4 C batteries
battery holder when removing or installing batteries. Best results are obtained if the batteries opposite the contact springs are removed first.
11
Page 20
12
Thermocouple Connector
Sample Connector (Brass)
Draft Connector (Silver)
Figure 4: Sampling Probe Connections

Connecting the Optional Combustion Supply Air Temperature Probe

A type K thermocouple probe may be used to measure the temperature of the air supplied to the burner. Connect the optional supply air thermocouple (see “Accessories and Replacement Parts”) to the supply air thermocouple port depicted in Figure 1. The thermocouple connector can be inserted only one way—large spade to the right. Do not force the connector.
When the optional Supply Air Temperature Probe is not used, supplied combustion air temperature is determined from the instrument temperature using an on-board sensor.

Connecting the Optional Portable Printer

Find the printer interface cable included with the optional portable printer. For the serial printer, connect the large 9-pin connector on the cable to mating connector on the printer. Connect the opposite end to the instrument’s RS-232 communications and printer port. See Figure 1 for port location.
The printer and CA-C
ALC™ combustion analyzer have both been factory
set for a baud rate of 1200. If baud rates are not matched, the printer will print random characters, question marks or asterisks. Printer settings are described in the printer manual, along with illustrations identifying the correct DIP-switch configuration. To set the CA-C device settings, refer to Chapter 6, “MENU Options
ALC™ baud rate and
.”

Connecting to a Computer

Use the optional computer interface cable, Model 8940, to transfer (download) data serially from the CA-C
ALC™ analyzer to a computer. Your
printer cable will not work for downloading to the computer. Connect the
Chapter 4
Page 21
13
O2
%
%
2
λ
2
Fuel
Baud Rate
1200
1200
large 9-pin connector on the computer interface cable to the 9-pin serial connector on your computer. Connect the opposite end to the instrument’s RS-232 communications and printer port. See Figure 1 for port location. Set the baud rate of the CA-C described in Chapter 6, “MENU Options
ALC™ analyzer to that of your computer, as
.” The factory preset baud rate is
1200. Set the COMM option to COMP so serial data is formatted for output to the computer. The alternative is PRNTR, indicating output formatted for printer output. Press the PRINT button to send data to the computer.

Default Instrument Settings

The CA-C
ALC™ combustion analyzer uses a number of parameter settings
for presenting measured data, performing calculations, and controlling instrument operation. These include the measurement units, the fuel used, the baud rate, and so on. When shipped, your instrument has factory preset parameter settings and the selection of E (English) as the Default language. When another language is chosen from the language MENU option, different settings are automatically installed. These are indicated in the table below. Specific settings are easily changed as described in Chapter 6, “MENU Options
”. However, settings are reset to those listed below if th e
language is changed.

Factory Defaults and Language Selection

Language Selection
CO Draft units Temp units Excess Air Effc./Loss basis Effc./loss units Fuel parameters Fuel list
Sampling interval Communications Time
E (Default) D, FI, S, N, I
PPM (parts per million) PPM Inches H Degrees Fahrenheit Degrees Centigrade %EA ASME Siegert formula % (net) % qA C, H, H U.S. Fuels Siegert Fuels Natural Gas Natural Gas
1 second 1 second Printer format Printer format MM / DD / YY DD / MM / YY
O mbar
= %EA/100 +1
O, HHV, CO2max A2 B
Getting Started
Page 22
14
(This page intentionally left blank)
Chapter 4
Page 23
Chapter 5
Basic Operation

Buttons and Button Operations

ON-OFF Control Button Turns the instrument on and off. The ENTER Control Button

Button Indicators

Press the ENTER button to execute a command, such as selecting a menu item.
The ESC Control Button
Used to return to the previous screen or cancel a process.
ARROW Control Buttons
Use the Left and Right arrow buttons to step between options.
Use the Up and Down arrow buttons to change number values.
Back Light ON/OFF
Pump ON/OFF
Print
Make draft measurement
Zero draft transducer
Select fuel
15
Page 24
16

Data Buttons

Clear stored data
Review stored data
Save data

Startup

Remove the sampling probe from the flue or disconnect the sample tubing from the sampling port. Press the ON-OFF button. Refer to Figure 5 showing screens displayed during the start up sequence. If no errors are detected, the Main Data Display screen appears when the sequence is complete. Error codes are presented in Appendix A
. ESC will bypass the
error messages. While the WARM screen is displayed, the O
sensor is automatically
2
calibrated using ambient air, and the CO and NO sensors are zeroed . A t the end of the WARM sequence (45 seconds), the pump stops and the draft sensor is zeroed.
The WARM sequence can be bypassed by pressing the ESC button. Do this if a new O
calibration and new zeros are not desired.
2
Chapter 5
Page 25
17
Time and Date
Error Codes
Warm-up Countdown
Model Number
Serial Number
Version Number
Basic Operation
Main Data Display
Figure 5. Startup Sequence
Page 26
18

Characters and Display Icons

Refer to Figures 6 and 7 that follow. These identify the icons and characters found on the Model CA-6203 display. The display icons indicate measurements made, units, or functions performed. Some icons are not used with the Model CA-6203.
Figure 6. LCD Display Showing Icons
Chapter 5
Page 27
19
CO
2
°C
L
OSS
COr NOxu
qA
T
NOx
λ
EA
°F
MENU
Eff
η
CO
NO
O
2
mg/m
3
mmH
2
O
x
mbar
hPa in.H
2
O
%
COu
ng/J
lb/MBtu
PPM
NOu
#
Sample number
CO undiluted
Pounds / million Btu
Nano grams / Joule
Parts per million
Milligrams / M3
Inches of water
Percent
Millimeters of water
Millibars
Battery life
Clear data
Save data
Pump on/off
Printer
Zero draft sensor
Draft
Supply air temperature
Efficiency (Siegert)
Efficiency (ASME)
Carbon Dioxide
Menu
Fuel
Flue Temperature
Differential temp.
Loss (Siegert)
Loss (ASME)
CO ratio
NOX undiluted
Oxygen Carbon monoxide Nitric oxide
Lambda (excess air) Excess air % Degrees F. Degrees C. NOX
Figure 7. LCD Display and Description of Icons
Average of samples
NO undiluted
Hecto Pascals
Review data
Light on/off
Basic Operation
Page 28
20

Measurements and Calculations

Oxygen Measurement

An O
electrochemical sensor measures oxygen concentration in the
2
range of 0 to 25%. The O concentration of carbon dioxide (CO
measurement is also used to calculate the
2
in the exhaust, and is used in
2)
calculations of combustion efficiency and undiluted CO and NO concentrations.

Carbon Monoxide Measurement

Carbon monoxide is measured in the range of 0 to 5000 PPM.
Measurement
CO
2
%CO
calculated from the O2 concentration in the flue. Refer to
2
Appendix B for a description of the calculation.

NO Measurement

Nitric Oxide is measured in the range of 0 to 4000 PPM.

Flue/stack Temperature, TS

Flue temperature is measured with sampling probe.

Combustion Air Temperature, TA

Measurement of combustion air temperature is determined using the temperature probe accessory, TSI PN 3013003. In the absence of this probe, the combustion air temperature is assumed to be the instrument temperature, and determined from a temperature sensor in the instrument case.

ΔT, Temperature Difference

The flue/stack temperature minus the combustion air temperature.
Excess Air or Lambda (
Calculations of these values are found in Appendix B
λ)
.

Loss

ASME based heat loss from the hot gases exiting the flue/stack. Includes latent heat loss from the formation of water vapor. See
Appendix B
.
Chapter 5
Page 29
21
qA
Heat loss using the Siegert formula. See Appendix B
.

Efficiency

Combustion efficiency, expressed as a percent. Calculated as 100 percent minus ASME heat loss (Loss) described earlier. Also see
Appendix B
.

ŋ Siegert Efficiency

Combustion efficiency, expressed as a percentage. Calculated as 100 percent minus qA. Also refer to Appendix B
CO
, Undiluted Carbon Monoxide Concentration
u
Calculation of the CO concentration, undiluted by excess air. This calculation uses a measurement of the O the CO measurement. Refer to Appendix B
CO
, CO Ratio
r
Ratio of CO to CO
NO
, Undiluted Nitric Oxide Concentration
u
. See Appendix B.
2
Calculation of the NO concentration, undiluted by excess air. This calculation requires a measurement of the O with the NO measurement. Refer to Appendix B
.
concentration together with
2
.
concentration together
2
.
NO
Concentration
x
NO
is an expression of the total oxides of nitrogen in the flue gas
x
stream. NO estimated at 5% of the NO concentration. NO
NO
xu
Calculation of the NO calculation requires a measurement of the O measurement of the NO concentration. Refer to Appendix B
combines NO and the theoretical NO2 concentration
x
is not measured.
2
, Undiluted Nitric Oxide concentration
concentration, undiluted by excess air. This
x
concentration and the
2
.

Draft

Draft pressure is measured with the sampling probe inserted in the flue. The draft measurement is referenced to room pressure.
Basic Operation
Page 30
22

Data Display, Viewing Measurements and Calculations

Measurements and calculations are presented in the Data Display screens. The Data Displays appear once the Startup sequence is complete (illustrated earlier in Figure 5).
Data appears on three lines, with an icon indicating the measurement or calculation presented above or to the left of the data. Units are presented on the right side.
Data are viewed using the ARROW buttons, in groups of two or three. The table below shows the order of the data displayed together in vertical columns. Use the right ARROW button to scroll through the data as shown below.
Order of displayed measurements and calculations.
ARROW button
O2 Draft Combustion Air
Efficiency CO
NO NOu
2
Temperature
CO Fuel
Type
COu Differential
Flue Temperature
Excess
CO NOx NOxu
Air Loss COr
Temperature

Error Indications

A series of ERR (error) codes indicate a problem with instrument operation. Often the problem has a simple remedy. If the ERR code persists, contact the factory. Press ESC to bypass the error message.
Error codes are described in Appendix A
.

Using the Sampling Probe

Gas and Temperature Measurements

Connect the sampling probe to the CA-C in Chapter 4
.
ALC™ analyzer as described
Place the Sampling probe through a hole in the exhaust flue, following recommendations presented below. Placement of the probe is
Chapter 5
Page 31
23
important and certain considerations must be given when choosing a sampling location.
To ensure that the gas measurements are not diluted or cooled by outside air, place the probe before any draft damper or regulator as illustrated in Figure 8. Tilt the probe tip up slightly so vapor condensing in the sampling tube does not run back to the probe tip and cool the thermocouple tip.
The sampling probe tip should be placed at the point of highest exhaust gas temperature. This means at the base of the flue, before heat is lost to the flue sidewalls, and towards the center, especially for small ducts. If the flue/stack gas temperature is underestimated, the operating efficiency will be overstated. When an economizer or air heater is used, the flue/stack temperature is measured after these devices
Important: Orient the sampling tube to ensure that the thermocouple tip is exposed directly to exhaust flow (see Figure 8 below).
1. Sampling probe 3. Exhaust flue
2. Draft regulator 4. Hot exhaust gases
Figure 8: Sampling Probe Location
Basic Operation
Page 32
24
!
Cautions
Hot probe! When removed, the sampling probe will be extremely hot. Avoid touching the probe tip, and avoid placing the probe on or near plastic materials such as the instrument case. These will melt. Maintain a minimum 2” (5 cm) clearance between the probe handle and position collar when the probe is mounted in flue.
Empty Water Trap! Watch the water trap and empty it frequently to prevent the possibility of flooding the instrument. See Chapter 8 instructions.
for
Making a Draft Measurement
Use the DRAFT button to the right of the DRAFT button indicator to make draft measurements. When the DRAFT button is pressed, the sampling pump turns off and the draft measurement is displayed and updated each second.

Zeroing the Draft Sensor

For the most accurate draft reading, the draft sensor must be zeroed prior to the draft measurement. To zero the draft sensor, remove the sampling probe from the flue, or separate the sampling tube from the draft port. Press the ZERO draft button. When ZERO is displayed, the draft sensor is zeroed.

Saving Draft

Whenever a sample is saved using the SAVE DATA button, a draft reading is taken. For this to occur, the pump is automatically turned off and a few seconds are allowed for the draft reading to stabilize. Draft is recorded at the end of the data-sampling interval.
Printing to the Portable Printer and to a Computer

Printing to the Portable Printer

Instrument data can be output to the optional portable printer through the RS-232 serial port. Refer to “Connecting the Optional Portable
Printer” in Chapter 4.
Chapter 5
Page 33
25
To print the information on the Data Display, press the PRINT button, to the right of the print button icon. The printer responds immediately, producing a printout of the current data. An example of this printout is shown in Figure 9.
Saved data can be printed too. Hold the Print button down until a countdown from three (3) begins. Release the button at zero (0), and all saved data is printed.
Printing individual saved data samples is described in the next section, “Handling Data
.”

Printing to a Computer

Use the PRINT button to output data to a computer as well as to the portable printer. Refer to “Connecting to a Computer
” in Chapter 4.
Data transferred to a computer is the same as that output to a printer (see Figure 9); however, it is formatted differently and uses the
®
Windows
character set rather than DOS characters. You will need to
set appropriate COMM MENU option before sending data to the computer. Refer to the Chapter 6, “MENU Options
.”
Data can be downloaded to a terminal emulator program such as the HyperTerminal, which accompanies the Windows
®
operating system
program. Look for HyperTerminal in the Accessories folder. In HyperTerminal, use the Capture Text option from the Transfer menu for recording instrument data.
Your instrument comes configured with the following communications protocol.

Communications Protocol

Baud rate 1200 (default) Data bits 8 Parity None Stop bit 1 Flow None
®
Windows and Excel are a registered trademarks of Microsoft Corporation.
Basic Operation
Page 34
26
-------------------------------­MODEL: 6203 SERIAL:
---------------------------------­Current Data
--------------------------------­DATE: 01/01/06 TIME: 15:00:18 Fuel: Nat Gas
Fuel Parameters: C: 75 H: 25 CO2MX: 11.8 KBTU: 23.8 H2O: 0 O2 6.0 CO 5 PPM NO: 0 PPM COu 7 PPM COr NA NO 40 PPM NOu 56 PPM NOx 42 PPM TA 70 °F TS 300 °F Draft 0.01 inH2O CO2 8.4 % Lambda OVER Effc 84% LOSS 16%
Figure 9: Example Printout
Chapter 5
Page 35
27

Handling Data

Saving Data
Up to one hundred (100) separate measurements can be made and saved to the instrument memory as data Samples. Press the SAVE DATA button located below the SAVE DATA icon at the bottom of the display. When SAVE DATA is pressed, measurements are sampled and averaged over the interval set in the INT.SA MENU option described in Chapter 6
. As data is sampled, the sample number is
displayed, and the save data icon blinks.
Note: The pump turns off while the draft is measured and saved.
CLEAR Data
While in the Data Display screen, press and hold the CLEAR DATA button for three (3) seconds to erase all saved data from the instrument memory. The button must be released when zero is displayed. If the button is held longer than three (3) seconds, nothing is cleared.
To clear the last Sample only, release the CLEAR data button before the end of the countdown, during 3, 2, or 1.
REVIEW Data Samples
Press the REVIEW DATA button briefly to recall saved Samples. The number of the last Sample appears. Select the specific data Sample you wish to view using the ARROW buttons and press ENTER. When the Sample number (#) is displayed, use the ARROW buttons to review the different measurements.
Printing
To print individual saved Samples, press the PRINT button while the saved Sample is shown. To print all saved Samples, hold the PRINT button down until a countdown from three (3) begins. Release the button at zero, (0) and all saved Samples are printed.
Averaging Saved Samples
To average the last three (3) saved Samples, press and hold the REVIEW DATA for a couple of seconds. The screen blanks during this interval. When averaged, the X-bar icon appears indicating that the data presented is an average of the Samples. Data averaging works only from the main Data Display screen.
Basic Operation
Page 36
28
(This page intentionally left blank)
Chapter 5
Page 37
Press the
Use
to
Press
Display.
Chapter 6
MENU Options
The Series 6203 CA-CALC instrument has a variety of user selectable parameters, available as MENU options. The user-selectable MENU options are shown in the schematic below. Press ENTER to access the MENU options from the Main Data Display screen. View the MENU options by using the right/left ARROW buttons. Once the screen of interest is displayed, press the ENTER button again to choose the MENU option. Use up/down buttons to change values. Refer to Figure 10, which shows the
MENU options.
ESC
to return to Data
ENTER button from the Main Display to enable the Menu options.
ENTER
select an option.
Figure 10. Menu Option
29
Page 38
30

BASE.L MENU—Option Update Sensor Baselines

Use the BASE.L MENU option to update the zero concentration baselines for your electrochemical sensors. Make sure the sampling probe is removed from the flue or disconnected from the instrument during zeroing. If ESC is pressed before zeroing is complete, previous baseline values are restor ed.

FUEL MENU Option—Select Fuel

Use the FUEL option to select from seven preset U.S. fuels or six Siegert fuels, or select the USER fuel. The fuels parameter list presented (U.S. or Siegert), is determined by the selection, LOSS or qA, respectively, made from the UNITS MENU option
Figure 11. FUEL Options
Chapter 6
Page 39
31

FP MENU Option—Fuel Parameters

See later.

UNITS MENU Option (changing units)

It is possible to display data in different measurement un its as indicated in the table below.
Optional Units
Measurement UNIT Options
TEMP (temperature) Degrees F, Degrees C GAS (for CO, NO, NOx
concentrations) PRESS (draft) In.H2O, mbar, hPa, mmH2O LOSS (flue gas heat loss) Loss (ASME), qA (Siegert) Excess Air % (EA), λ (Lambda) DECIMI Periods or commas for printing DATE Date format, MM/DD or DD/MM for printing
PPM, mg/m3, lb/Mbtu, ng/J
To change units, press the ENTER button from the Data Display screen. Using the right/left ARROW buttons, find UNITS from the MENU options
and press ENTER. Continue using the right/left ARROW buttons to find the measurement for which units are to be changed. Use the up/down ARROW buttons to select the desired unit and press ENTER.
Use ESC when complete to return to the Data Display screen.

INT.SA MENU Option—Sampling Interval for Saving Data

This option enables you to change the sampling interval over which data is averaged and saved. Select this option using the ENTER button. Use the up and down ARROW buttons to choose the sampling interval from the following: 1, 5, 10, 15, 20, 25, and 30 seconds. Press ENTER to select the interval.
Use the SAVE DATA button to save data over the sampling interval.
Menu Options
Page 40
32

TIME MENU Option—Set Time and Date for Printout

Press the ENTER button when the TIME MENU item is displayed. Use the ARROW buttons to select the field to be changed. The f ield will blink as it
is changed with the right and left ARROW button. To change the value, use the up and down ARROW buttons. Press ENTER to install the new value. Press ESC to abort and return to the previous screen. Use the DATE option in UNITS to change the month/day presentation.
Time or and Date fields

CALIB MENU Option (select sensor for calibration)

The CALIB MENU option is used for performing calibrations of the on­board sensors. When this option is chosen using the ENTER button, characters and icons associated with the installed sensor will be displayed, one at a time. Press an ARROW button and scroll through the options until the desired sensor for calibration appears. Press ENTER to begin the calibration process.
Example, Draft calibration initiated.
HR:MN MM.DD 2 0 Y Y
HR:MN DD.MM 2 0 Y Y
Icons below appear one at a time. See
Chapter 5
for a description of icons.
Use ARROW buttons to choose sensor for calibration. Press ENTER to initiate the calibration.
Chapter 6
Page 41
33

O2 Gas Sensor Calibration

The Oxygen (O
) sensor is SPAN calibrated in room air whenever the
2
instrument is turned on and allowed to complete the Startup sequence. Using the BASE.L MENU option also spans the O
sensor. Additional calibration
2
is generally not required.
Calibrating the O
sensor is done to determine a new zero value (no O2
2
present) for the sensor. This is done if a drift in the sensor zero value is suspected.
Select CALIB from the MENU options using the ENTER button. Select the
icon from the icons displayed using an ARROW button, and press
O
2
ENTER. Attach the zero gas (nitrogen, N described in Chapter 7, “Setup for Gas Calibration
) to the sampling probe as
2
.” When ready, press
ENTER to begin the zero calibration. Once complete, the SPAN calibration screen appears. The O
concentration can be adjusted using the ARROW
2
buttons. If room air is used as the calibration gas, use the value displayed initially, 20.9%. Press the ENTER button to begin the SPAN calibration. Once complete, use the ESC button to return to the Data Display screen.

Calibration of the CO (hydrogen compensated) Sensor

The presence of hydrogen gas (H electrochemical CO sensors. The CO sensor in your CA-C has a second internal sensor that detects the H
) produces a false signal on
2
ALC™ analyzer
concentration and enables the
2
instrument software to subtract the error caused by the presence of hydrogen gas.
You can elect to calibrate using both the CO and H only. In the second case the sensor retains the previous correction for H
, or calibrate with CO
2
.
2

Calibration Steps

From the MENU screen select the CALIB option. Press ENTER. Use the ARROW buttons to select the CO icon and press ENTER. Sampling from
the room air, press ENTER to perform the ZERO calibration. When complete, connect your CO calibration gas as depicted in Chapter 7, “Setup
for Gas Calibration.” Use the ARROW buttons and adjust the CO gas
calibration number to match the concentration of your gas bottle. Press ENTER to initiate the CO calibration. Once complete, con nect your calibration gas bottle containing both carbon monoxide and hydrogen (see note below). Use the ARROW buttons to adjust to the CO bottle concentration under the SPAN.CO heading, then press ENTER. Under the
SPAN.H
heading adjust the number using the ARROW buttons to match
2
the hydrogen concentration. Press ENTER to start the calibration.
Menu Options
Page 42
34
!
Note: It is possible to do only the CO part of the calibration. To
bypass the calibration using the combined CO and H press the ESC button after the CO portion of the calibration is complete.
bottle,
2
CAUTION
Toxic Gases!
Familiarize yourself with the toxic properties of the calibration gases by reading the supplied Material Safety Data Sheets (MSDS) accompanying the gas cylinders. Do not perform calibrations in a confined space. Make sure calibrations are performed in an area with proper ventilation, or under an exhaust hood.

NO Gas Sensor Calibration

From the MENU screen select the CALIB option. Press ENTER. Use the ARROW buttons to select the NO icon and press ENTER. Sampling from
the room air, press ENTER to perform the ZERO calibration. When complete, connect your NO calibration gas as depicted in Chapter 7, “Setup
for Gas Calibration.” Use the ARROW buttons and adjust the NO gas
calibration number to match the concentration of your gas bottle. Press ENTER to initiate the NO calibration.

Draft Calibration

Span calibration of the draft sensor is normally not required. The draft sensor should be routinely zeroed, however, using the DRAFT ZERO button. This is especially important prior to making measurements at low draft pressures. When zeroing, make sure the sample probe is disconnected or that the probe is not in the flue.
Calibration Procedure Required: An apparatus to provide stable calibration pressures
of between plus 10–30 inches of water (25–75 hPa), and stable calibration pressures of between minus 10–30 inches of water (minus 25–75 hPa).
From the Main Data Display screen, press the ENTER button. Select CALIB from the MENU options presented using the ENTER button.
Select the draft icon from the icons displayed using an ARROW button and press ENTER. Begin the calibration by pressing the ENTER button to perform a ZERO calibration. When the zero is complete, use the arrow buttons to adjust your positive supply reference pressure. Connect
Chapter 6
Page 43
35
your positive pressure to the draft port and press ENTER. Once the positive calibration is complete, connect your negative pressure reference to the draft port. Adjust the negative calibration pressure value using the ARROW buttons and press ENTER. Once the calibration is complete, use the ESC button to return to the main Data Display screen.

Temperature Calibration

Calibration of the flue and ambient thermocouple sensors is not recommended. Thermocouples are very repeatable, and even if replacement of a sampling probe is required, or a combustion air temperature probe is purchased as an accessory, it is unnecessary to calibrate it.
Thermocouple accuracy can be improved for a specific temperature range if a calibration is preformed in that range. You will need to provide an accurate temperature reference. Refer to the steps below if calibration of a thermocouple is desired.

Calibration Procedure

Select the CALIB MENU option. Select the appropriate thermometer icon in the CALIB screen using the ARROW buttons. Install your sampling probe thermocouple or accessory thermocouple at the appropriate TS or TA connector. Press ENTER and set the calibration temperature to that of your supplied standard. Bring your thermocouple to the entered temperature and initiate the calibration by pressing ENTER.

CAL.FA MENU Option—Setting the Calibration Factor for a Replacement Gas Sensor

Install your new sensor as described in Chapter 8, “Maintenance and
Troubleshooting.”

CO Sensor

Find the Calibration Factor sheet. Use the MENU option CAL.FA and press ENTER. Change the AVAL, BVAL, CVAL and DVAL to values on the
calibration sheet using the up and down ARROW buttons. Follow each change by pressing the ENTER button. Figure 12 shows the steps. Pressing ESC takes you out of the entry sequence but keeps values which were saved with the ENTER button.
Once complete, review entered values by stepping through the calibration factor screens using ENTER.
Menu Options
Page 44
36
Use
buttons to
Use the ARROW buttons to find the CAL.FA MENU option. Press the ENTER button.
ARROW
change AVAL, BVAL, CVAL and DVAL calibration factors. Press the ENTER after changing each value.
CO icon
Choose the CO icon using the ARROW buttons and press ENTER.
Complete, return with the ENTER button followed by ESC.
Figure 12. Setting the Calibration Factor

NO Sensor

After installing your new sensor, as described in Chapter 8, “Maintenance
and Troubleshooting,” find the NO calibration factor sheet. Use the
ARROW buttons in the MENU option and find CAL.FA. Press ENTER. Use the ARROW buttons and select the NO sensor icon, press ENTER. Adjust the number displayed to match the value on the calibration sheet. Press ENTER to finish. Press the ESC button to return to the Data Display.
Chapter 6
Page 45
37

BAUD Rate MENU Option

Baud rate can be set to match your computer or portable printer. Your instrument is delivered with a default baud rate of 1200 to match the printer.
Baud rate values are displayed in the BAUD rate MENU option divided by
1000. The following baud rates can be set. 1.2 (1200), 2.4 (2400), 4.8 (4800), 9.6 (9600), 19.2 (19200).
To set the baud rate, select BAUD from the MENU option using the ENTER button. Step through the BAUD rates using the ARROW buttons and choose the desired value with the ENTER button. Use the ESC button to return.

COMM MENU Option—Set the Output Communications Device

Your CA-C
ALC™ analyzer transfers current or saved data to a serial printer
or computer. The COMM MENU option is used to select the device that you want to communicate with. Failure to select the correct device may cause incorrect characters to appear on your printout or computer file.
COMM options are COMP and PRNTR, for computer and printer, respectively. Set the device COMM from the MENU options using the ENTER button. Use the ARROW buttons and choose the desired option followed by pressing the ENTER button. Use the ESC button to return.

FP MENU Option—Fuel Parameters

U.S. Fuel Parameters Note: For U.S. fuels and fuel parameters to be displayed, the
fuel heat LOSS selection must be set to Loss, not qA. Refer to the section “UNITS MENU Option (changing
units)” in this chapter.
For calculations of flue losses, maximum %CO
, fuel composition and
2
fuel heat content are used. These are the Fuel Parameters. In your
ALC™ analyzer, U.S. fuel parameters are the carbon and hydrogen
CA-C content, moisture content, sulfur content, and maximum %CO
,
2
CO2MX.
Menu Options
Page 46
38
Press
button from the Data Display screen.
Change the parameter
value using the ARROW buttons. Press the ENTER button to
accept the new value
.
Press ESC to abort. The example to the right is for changing the hydrogen fuel percent.
Figure 13. Fuel Parameters for U.S. Fuels.
ENTER
Find FP using the ARROW buttons.
Press ENTER to access the fuel parameter list. Find the parameter to change using the
ARROW buttons
.
Chapter 6
Page 47
39
Refer to the diagrams (Figures 13 and 14) and tables below for information on viewing and changing fuel parameters.
Your CA-C
ALC™ analyzer has parameters for seven common U.S. fuels
in instrument memory. Fuel parameters for these fuels can be changed, if you know, for example, that your fuel has a different composition than that stored. The instrument fuel parameters values are presented in
Appendix B
, with additional technical information.
Fuel Parameter Description
C % carbon by weight H % hydrogen by weight H2O % moisture content by weight S % sulfur by weight CO2MX CO2 maximum % KBTU Heating value in kiloBTUs, (BTU/1000) F.CO F.NOx Emission Factors

FP Fuel Parameters for Siegert Calculation—Siegert Fuel Parameters

Note: For Siegert fuels and fuel parameters to be displayed, the
fuel heat LOSS selection must be set to qA not LOSS. Refer to the section “UNITS MENU Option (changing
units)” in this chapter.
The Siegert value for flue loss, given the designation qA, is used widely in Europe. Two coefficients are used in the Siegert formula for flue loss, derived from typical fuel compositions. These are given the designations, A2 and B.
Fuel Parameter Description
A2 Siegert factor, fuel specific B Siegert factor, fuel specific CO2MX CO2 maximum % F.CO F.NOx Emission Factors
Menu Options
Page 48
40
The default Siegert coefficient values in your CA-CALC analyzer, are those used in Germany. Siegert coefficients used in other countries may be different, reflecting differences in local fuel compositions.
Figure 25 diagrams the steps in changing the Siegert fuel parameters.
Figure 14. Changing Fuel Parameters for
Siegert Fuels

LANG MENU Option

Use this option to select between the following languages: E English, D German, N Dutch, FI Finish, I Italian or S Swedish. English is the
language (default) initially installed on the instrument.
When the language is changed, instrument settings are automatically changed too. Refer to Chapter 4 “Default Instrument Settings
” for more
information on this feature.
Chapter 6
Page 49
Chapter 7
Setup for Gas Calibration
The Calibration Setups Note: To perform your gas sensor calibration, you will also need to
refer back to Chapter 6, and the section, “CALIB MENU
Option.”
CO, NO and O best accuracy. Gas sensors do drift over time, depending upon the operating environment and gas exposure history. CO and NO gas sensors can lose sensitivity by as much as by 10% per year. Since O room air, loss in sensitivity poses no real problem. However, a drift in the zero (no O
concentrations.
O
2
With the proper equipment, such as that shown in the figures below, it is easy to calibrate your CA-C wish, you may also return your instrument to TSI for a new factory calibration.
The equipment needed to calibrate individual gas sensors can be purchased from TSI as calibration kits. Model numbers for these kits are found in Chapter 2, “Unpacking calibration system. Two calibration setups are presented in Figures 26 and
27. A brief discussion of these calibration setups is presented in the following section.
gas sensors can be calibrated periodically to maintain the
2
sensors are calibrated in
2
present) can occur, impacting the accuracy when measuring low
2
ALC™ analyzer gas sensors. However, if you
.” You may also elect to put together your own
41
Page 50
42
Figure 15. Calibration with TSI Calibration Kit
A TSI supplied calibration kit (Figure 15) uses a demand flow regulator to supply gas to the CA-C
ALC™ analyzer in response to the draw of the
instrument sampling the pump. If a conventional regulator and valve are used (Figure 16), the setup requires a tee to a bleed-off extra gas. This prevents a forced flow at the instrument inlet. The bead-type flow meter depicted in the figure is used to verify there is extra flow (.5 to 2 L/min recommended). Extra flow is required to prevent room air from being drawn in, diluting the sample.
Chapter 7
Page 51
43
Figure 16. Alternative Calibration Setup
Setup for Gas Calibration
Page 52
44
(This page intentionally left blank)
Chapter 7
Page 53
Chapter 8
Maintenance and Troubleshooting
For additional troubleshooting information, please visit TSI’s website
http://combustion.tsi.com.

Emptying Water Trap

Refer back to Figure 2 showing the water trap in the sample line and to Figure 17. Liquid water forms in the first chamber of the water trap as gases are sampled from the flue. The water trap is designed so even when shaken, or when its orientation is changed, water does not pass to the second chamber. The water level must remain below the level depicted in the figure, however.
To empty the water trap:
1. First separate it from the sampling tube by pulling the tube ends off the
barbs on the end caps.
2. Remove the probe side end cap by pulling outward with a twisting
motion.
3. Po ur out the water and replace the end cap. Re-install the trap.
Important: Make sure the water trap is oriented so that end-cap 1 below is toward the instrument.

Changing the Water Trap Filter

Identify the water trap filter (refer to Figure 17). This filter is designed to remove soot particles before they contaminate the instrument. The filter can be removed for cleaning or replacement by following these steps:
1. Remove the in s trument side end cap by pulling it out with a twisting
motion.
2. Grasp the filter using a needle-nose pliers and pull it out.
3. To clean the filter, remove the bulk of the soot by tapping the filter. The
soot may be removed by rinsing with water or isopropyl alcohol. The effectiveness of the rinse depends on the soot composition—is it dry or oily. Avoid rubbing, which may drive contaminates into the filter causing permanent plugging.
45
Loading...