ALNOR® LOFLO BALOMETER®
CAPTURE HOOD
P/N 116620001, REV 08
SEPTEMBER 2014
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 2

TABLE OF CONTENTS

Section 1 General Description ................................................ 1
Section 2 Theory of Operation ................................................ 1
Section 3 Safety ...................................................................... 2
Section 4 Instrument Setup and Storage ................................ 2
Section 5 About the Instrument Display .................................. 4
Section 6 Getting Started ........................................................ 5
Section 7 Detailed Operation .................................................. 5
Section 8 Troubleshooting ...................................................... 8
Section 9 Maintenance ........................................................... 9
Section 10 Service Information ............................................... 10
Appendix A Correction Factors for Standard
Flow Rate to Actual Flow Rate ............................. 11
Appendix B Model 6200 LoFlo Balometer® Capture Hood Field Setting Instruction for User Programmable
Correction Factors ................................................ 12
v
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 3
SECTION 1
General Description
The Alnor® LoFlo Balometer® Capture Hood is a compact instrument that measures very low air flow rates in HVAC systems. It measures from 10 to 500 cfm (17 to 850 m3/h, 4.7 to 236 l/s). The LoFlo Balometer® Capture Hood displays standard air volume rate when placed at supply or return diffusers, registers, or grilles located on the ceiling, wall, or floor.
The LoFlo Balometer® Capture Hood has a vent mechanism which reduces the restriction of air flow caused by the instrument. This unique venting system (patent pending) allows the instrument to remain compact while still measuring a wide range
SECTION 2
Theory of Operation: Capture Hoods
and Thermo-anemometry Sensing
As air exits a diffuser, it is captured in the hood or base and directed over a manifold. The manifold averages the velocity pressure and directs air over the sensors. The velocity and temperature of the air passing through the manifold are sensed, using a constant temperature thermo-anemometer principle. The voltage output from the sensor is then calibrated for the entire volume rate passing through the LoFlo Balometer® Capture Hood.
All capture hoods are susceptible to “odd” air flow patterns. “Odd” could refer to any air flow pattern different from the pattern where it was calibrated. Substantial negative effects can be observed when using a large hood on a small diffuser. For
of volume rates. The LoFlo Balometer® Capture Hood should be used with the vents closed at lower volume rates, but higher volume rates should be measured with the vents open to minimize any resistance effects. A 2-point measurement may be taken to compensate for the resistance effect of the instrument at higher volume rates.
The LoFlo Balometer® Capture Hood is battery powered and may be used with or without optional hoods. It weighs just over 6 pounds with the 2' x 2' hood, helping to reduce worker fatigue.
example, using a 2' x 2' hood on a 10" x 10" diffuser. This creates large recirculation regions on the sides of the fabric hood and causes an “odd” air flow pattern as it passes over the manifold. It is recommended to match the hood closely to the size of the diffuser.
NOTE: Throughout this manual, the units cfm,
m3/h and l/s are used for simplicity. Note, however, that the LoFlo Balometer® Capture Hood actually reads in standard cubic feet per minute (scfm), standard cubic meters per hour (std. m3/h), or standard liters per second (std. l/s).
1
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 4
2
C A U T I O N
WHILE USING THE LOFLO BALOMETER
®
INSTRUMENT TO TEST AIR FLOW IN DUCTS, YOU MAY COME INTO CONTACT WITH OR BE EXPOSED TO DUST, POLLEN, MOLD, FUNGUS, OR OTHER AIRBORNE CONTAMINANTS. IF YOU ARE OR MAY BE SENSITIVE TO DUST, POLLEN, MOLD, FUNGUS, OR OTHER AIRBORNE CONTAMINANTS, ALWAYS USE AN APPROPRIATE MASK OR RESPIRATOR WHILE EMPLOYING THE LOFLO BALOMETER® INSTRUMENT.
SECTION 3
Safety
When using the LoFlo Balometer® Capture Hood to check air flow at ceiling diffusers, make certain that you can safely raise and hold the unit while making measurements. Be especially careful when working on a ladder.
Observe all necessary precautions so that the unit does not become caught in moving machinery or touch any exposed electrical wiring.
SECTION 4
Instrument Setup and Storage

Hood Setup

1. Locate the four (4) aluminum channels for the hood and position them into a square with the foam facing up. Slide the “U” opening on the channel end (Fig. 1) and corner bracket toward the brass eyelets until you feel it latch.
2. Continue with the other two channels until you have a square frame that is locked tightly in all four corners.
3. Locate the hood for the frame size you have assembled.
The LoFlo Balometer® Capture Hood is not designed for gas mixtures other than air. Use with corrosive or other dangerous or explosive gas mixtures is not recommended and is at the user’s risk.
the base. Rotate the hood assembly until the metal cups are above the spring-loaded pins.
8. Locate the hood support rods. Extend the hood so it is taut and insert one support rod onto the spring-loaded pin closest to you.
9. Push the support rod down and insert the top end of the rod into the metal cup on the hood frame. Fig 1.
10. Install the rod on the opposite side then install the last two rods.
4. Position a corner seam of the hood onto the corner of the frame. Stretch the nylon hood to an adjacent corner until the hood corner meets the frame corner.
5. Push the elastic cord into the channel along the side of the frame.
6. Repeat steps 4 and 5 until the hood assembly is complete and ready to mount onto the base.
7. Place hood on a table or clean floor with the base inside the opening. Pull the hood over
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Hood Frame Channel
Figure 1
Page 5
3

Instrument Storage

The hood and frame assemblies can be detached from the instrument base and placed inside their own carrying case without disassembly.
The LoFlo Balometer® Capture Hood should always be turned off before re-packing the instrument. If storing the LoFlo Balometer® Capture Hood for an extended period of time, remove the batteries to prevent damage due to leakage.

Battery Installation

To install the 4 C-size batteries that ship with the LoFlo Balometer® Capture Hood:
1. Make sure the LoFlo Balometer® Capture Hood switch is off.
2. Remove the battery cover located on top of the meter.
3. Slide the battery holder up until the snap connector is exposed.
4. Disconnect the battery holder from the LoFlo Balometer® Capture Hood.
5. Install the 4 C size batteries in the holder.
6. Reconnect the snap connector and slide the battery holder into the battery compartment.
7. Replace battery cover.
3
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 6
4
SECTION 5
Index No.
Name
1
LCD
2
On/Off Switch
3
Supply/Return Button
4
Manual Range Button
5
Handle Button
6
Battery Holder
7
Vents
8
Vent Open/Close Slide
Index No.
Name
1
Digital Readout
2
Simulated Analog Display
3
Manual Range Indicator
4
Supply/Return Indicator
5
2-Point Mode Indicator
6
Vent Indicator
7
Units Indicator
8
Low Battery Indicator
About the Instrument Display
This display is an analog display, simulated with 26 LCD segments to give the user the response of a mechanical Balometer® Capture Hood. It also displays an exact digital number, using 3½ digit numeric section. The LCD has indicators for supply, return, manual range, vent modes, and low battery.
Figure 3 shows the front of the LoFlo Balometer® Capture Hood.
Figure 4 shows the LCD.
Figure 3
Front View of LoFlo Balometer® Capture Hood
Figure 4
LoFlo Balometer® Capture Hood LCD
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 7
6

Supply/Return Button

The LoFlo Balometer® Capture Hood is in supply mode when turned on. This is indicated on the display with the supply indicator . To take return
air measurements, press the supply/return button on the front of the meter. Return air mode will be indicated on the display with the return indicator
.
Significant errors may result if:
Instrument is not in proper mode.
Fabric hood is not used with vent open
mode.

Vent Closed/Vent Open Modes

When the instrument is turned on, it is in the vent closed mode . When the vent closed indicator is on, the vents must be closed to take the measurement. Close the vents by loosening the thumbscrew and sliding the vent baffle to the fully closed position and locking the vent in place again.
The vent open mode is indicated with the vent open indicator on the LCD . Open the vents by loosening the thumb screw and sliding the vent baffle to the fully open position and locking it into place again. Then press and hold the handle switch for more than 3 seconds to switch the LoFlo Balometer® Capture Hood to the open vent mode.
The vent open mode was developed for higher volume rates on larger diffusers; the 2' x 2' or 650 mm x 650 mm hood should always be used with vent open mode.

1-Point Measurement Mode

Use Vent Closed Mode for measurements between:
10 to 150 cfm (17 to 255 m3/h, 4.7 to 71 l/s).
NOTE: In Vent Closed Mode, volume rates
below 8 cfm (13 m3/h, 3.6 l/s) are displayed as 0; volume rates above 517 cfm (879 m3/hr, 244 l/s) are displayed as Or.
Use Vent Open Mode for measurements between:
150 to 500 cfm (255 to 850 m3/h, 71 to 236 l/s)
NOTE: In Vent Open Mode, volume rates
below 150 cfm (255 m3/h, 71 l/s) are displayed as Ur; volume rates above 517 cfm (879 m3/h, 244 l/s) are displayed as Or.

2-Point Measurement Mode—Supply Only

To activate the 2-point measurement mode, turn the instrument off. While pressing the handle button, turn the instrument back on. The 2-point mode indicator will show.
The 2-point measurement mode takes both vent open and vent closed measurements to calculate a resistance-compensated volume flow rate.
The process starts with the vents closed and the display showing the vent closed indicator.
1. Place the LoFlo Balometer® Capture Hood
over the diffuser to be measured and press the handle switch once to take the vent closed measurement.
2. The instrument stores that reading.
3. The display will now show the open vent
indicator and you should open the vents.
NOTE: Make certain you are using a fabric
hood when in 2-point mode.
4. Again place the LoFlo Balometer® Capture
Hood over the diffuser to be measured and press the handle switch to take the vent open measurement.
The default mode of the LoFlo Balometer® Capture Hood is the 1-point measurement mode. If you are in 2-point measurement mode, simply turn the instrument off and turn back on.
The 1-point measurement mode is described in the Getting Started section earlier in this manual. This is the faster, simpler way to take measurements.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
5. The instrument calculates and displays the
compensated reading.
6. The numeric display will flash; press the
handle switch again and the instrument will display the vent closed indicator. Close the vents.
7. Return to Step 1 to take another resistance
compensated reading.
NOTE: Return to 1- point measurement mode
at any time by turning the power off and on.
Page 8
7
Condition
Reason
The vent closed or vent open measurement is less than 150 cfm (70 l/s, 255 m3/h).
Calculations are not made at the low end since
the effect is small and can generally be neglected. Use 1-point mode.
The vent closed or vent open measurement is greater than 500 cfm (235 l/s, 850 m3/h).
Calculations are not made beyond the calibrated
range.
The vent open measurement minus the vent closed measurement is too small (< 0).
A mistake has occurred in the process. Repeat
the two-point measurement.
The resistance effect is small and was
overwhelmed by natural fluctuations. Use single­point mode.
The vent open measurement minus the vent closed measurement is too large.
A mistake has occurred in the process. The vent
may not have been open for the open vent measurement. Repeat the two-point measurement.
The effect is beyond the range of the
compensation equation. Perform a traverse of the system instead.
When “–––” is displayed at the last step of the 2-point mode, the calculation for the resistance effect cannot be made. It can be caused by any of the conditions listed here.
NOTE: The display will also show “–––” when the return mode is selected. The 2-point measurement
mode is for supply only.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
7
Page 9
8
Symptom
Possible Cause and Corrective Action
Meter does not turn on.
Battery holder snap connector not connected.
Connect battery holder to instrument.
Batteries may be discharged. Charge or replace
them.
Switch failure, wiring failure, or circuit failure. Call
TSI.
Meter reading lower than expected.
Wrong vent mode chosen. Choose correct vent
mode, open or closed.
Wrong air direction mode chosen. Choose
correct direction, supply or return.
Back pressure effects could be significant. Use 2-
point measurement mode.
Hood frame not sealing properly around diffuser
or grill. Press hood evenly against diffuser.
Odd air flow pattern present. Perform traverse
and use proportional balancing.
Hood torn. Replace, or repair tear with duct tape
or other non-porous material.
Meter out of calibration. Call TSI.
Damage to manifold. Call TSI.
Meter reading higher than expected.
Wrong vent mode chosen. Choose correct vent
mode, open or closed.
Wrong air direction mode chosen. Choose
correct direction, supply or return.
Odd air flow pattern present. Perform traverse
and use proportional balancing.
Meter out of calibration. Call TSI.
Meter not at zero at zero flow.
Instrument not at room temperature. Allow
instrument to acclimate before turning on the power.
The LoFlo Balometer
®
Capture Hood is reading real room air currents. Place on floor with a piece of cardboard over the top to determine if it is measuring real air currents.
Meter out of calibration. Call TSI.
Meter displays Er.
Self test failed. Call TSI.
Handle button not working.
Call TSI.
SECTION 8
Troubleshooting
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 10
9
SECTION 9
Maintenance

Hoods

Hoods may be cleaned in cool water with a mild detergent.

Sensing Manifold

The manifold should be checked visually before using to be certain that the sensing holes have not become clogged with dirt or dust particles. Do not immerse the manifold in water. It is recommended that cleaning be done with extreme care with the manifold in place. Do not use highly compressed air to clean the manifold tubes.

Calibration

It is recommended that your LoFlo Balometer® Capture Hood be returned to the factory once a year for a calibration check. When shipping the LoFlo Balometer® Capture Hood for factory calibration, pack it carefully, and follow the Instructions for Return in this manual.

Vent Operation

If the vent becomes difficult to operate, remove both thumbscrews from the vent mechanism and spray the contacting surfaces with silicone spray lubricant. Wipe off excess and reassemble vent mechanism.

Performance Check

If the calibration is to be checked, the best way is to use a reference flow standard more accurate than the LoFlo Balometer® Capture Hood. Flow standards that may be used for this purpose include orifice plates, venturis, nozzles, and laminar flow elements. These devices require accurate differential pressure, barometric pressure, and temperature reading instruments to measure either actual or standard volume rate.
If the LoFlo Balometer® Capture Hood is checked against a velocity instrument such as a pitot probe/manometer or thermo-anemometer, there may be errors. If a velocity standard is used, the average velocity must be obtained by taking a traverse.* The accuracy of the traverse is dependent on the flow uniformity, the number of readings taken, the accuracy of the velocity instrument, and operator expertise. This average velocity reading must then be multiplied by the area over which the traverse was taken.
Finally, air flow instrumentation is dependent on environmental conditions such as temperature, atmospheric pressure, humidity, and even turbulence. These conditions can have very different effects on various instrument types. Caution must be exercised when making comparisons.
*A traverse is a set of velocity readings taken in a prescribed pattern which will provide an overall velocity value when averaged together. Any of the Alnor velocity measuring instruments can be used for this purpose.
9
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 11
10
SECTION 10
Service Information

Service and Repair

Please return your Product Registration Card immediately. This allows us to send you service reminders, special offers, and important information about your product.
Before sending your instrument for calibration or repair, you should call TSI Customer Service. The Service Department will provide you with the cost of service or calibration, Return Material Authorization (RMA) number, and shipping instructions.
Please have the following information available when you call:
Owner’s name, address, and phone
number
Billing address, if different and applicable  Instrument name or model  Serial number  Date of purchase  Where purchased
TSI recommends that you keep a “calibration log” and keep all records of service on your instrument.

Instructions for Return

Send the instrument prepaid. Securely package your instrument in a strong container surrounded by at least two inches (5 cm) of suitable shock­absorbing material. Include the Purchase Order showing instrument model number, cost of service and/or calibration, and the RMA number. Mark the outside of the shipping container with the RMA number. This will expedite processing of your instrument when we receive it.

Damaged in Transit

All orders are carefully packed for shipment. On receipt, if the shipping container appears to have been damaged during shipment, the instrument should be thoroughly inspected. The delivering carrier’s papers should be signed noting the apparent damage. DO NOT DISCARD THE BOX.
If the instrument itself has been damaged, a claim should be promptly filed against the carrier by the customer. The selling agent will assist the customer by supplying all pertinent shipping information; however, the claim must be filed by the insured. If the instrument is damaged beyond use, a new order should be placed with TSI while awaiting reimbursement from the carrier for the damaged instrument.
Call TSI directly for assistance if necessary.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 12
11

APPENDIX A

Correction Factors for Standard Flow Rate to Actual Flow Rate
The LoFlo Balometer® Capture Hood measures standard volumetric flow rate. Standard flow is defined as the flow rate at standard conditions, 70°F (21.1°C) and 14.7 psia (29.92 inHg, 760 mmHg). Actual flow rate is the true volumetric rate of air at the local temperature and barometric pressure. If you desire standard flow rate, use the displayed reading. However, if the actual flow rate is desired, use the information below.
To correct the standard flow rate to actual flow rate use the equation below.
V
= V
act
V
= Actual flow rate
act
V
= Standard flow rate (this is displayed
std
T
= Actual Temperature
act
= Actual barometric pressure
act
= Standard density 0.075 lb/ft3
std
a = Actual density in lb/ft3
x CF
std
by the instrument)
(at 70°F and 14.7 psia)
CF =
Using the equations or chart below, you can bypass the full density calculation.
Calculations in imperial units
CF = (14.7/
Where
Calculations in imperial units
CF = (29.92/
Where
Calculations in metric units
CF = (760/
Where
For your convenience, we have made a chart for determining the correction factor given temperature and pressure.
NOTE: The atmospheric pressure as reported
by the National Weather Service is corrected to sea level and cannot be used if measurements are not at sea level.
(Correction Factor)
std/a
)*(460+ T
act
is in psia and T
act
)*(460+ T
act
is in inHg and T
act
)*(273.15+ T
act
is in mmHg and T
act
act
is in °F
act
act
act
act
act
)/530
)/530
is in °F
)/294.25
is in °C
11
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 13
12

APPENDIX B

Model 6200 LoFlo Balometer® Capture Hood Field Setting Instructions for
User Programmable Correction Factors
HISTORY:
It is well known that capture hoods are susceptible to diffuser styles that are different from that upon which they were calibrated. Understanding that these productivity tools may require adjustment in the field, the LoFlo Balometer® Capture Hood is equipped with two separate sets of field
programmable correction factors, FLD Aand FLD B, that allow for a field calibration of the
instrument to different diffusers.
DETERMINING FIELD PROGRAMMABLE CORRECTION FACTORS:
Correction factors are simply the relation between the actual volume flow rate and that which is displayed on the LoFlo Balometer® Capture Hood. The actual volume flow rate is most accurately established through performing a duct traverse. The correction factor is then determined as the ratio of the actual volume flow rate divided by the volume flow displayed on the LoFlo Balometer® Capture Hood set in the factory calibration setting (default from factory). The number obtained is used as the multiplier for that particular diffuser.
Example:
Duct Traverse yields 120 CFM LoFlo Balometer® Capture Hood in Supply Mode
with Vents Closed yields 107 CFM
Correction Factor = 120/107 = 1.121
Because there are no decimal points available for the correction factor values, this number will be programmed into the LoFlo Balometer® unit as
1121.
If all available operating modes of the LoFlo instrument are to be used, a separate multiplier must be determined for each mode. The four available modes are:
• Supply, Vent Closed (10-150 CFM)
• Supply, Vent Open (150-500 CFM)
• Return, Vent Closed (10-150 CFM)
• Return, Vent Open (150-500 CFM)
PROGRAMMING THE CORRECTION FACTORS TO THE INSTRUMENT:
Once the multipliers have been determined, the correction factors are programmed to the LoFlo instrument using the following steps:
1. Turn the instrument off.
2. Press and hold the range selector button while simultaneously switching the instrument
on. The display will show “FAC” with the seven
segment characters, indicating the unit is in its default factory setting.
3. Use the supply/return button to cycle
through the available storage fields, FLD A”, FLD B” or “FAC”. Use the handle hold switch to select the desired field. If FLD A” or FLD B” are selected, the field letter chosen will be
displayed at the upper right of the LCD throughout the programming cycle. Note that
the factory calibration setting, “FAC”,
cannot be altered.
4. Once the desired storage field has been selected, the unit will display the symbols for the Supply, Vent Closed mode, along with the current correction factor programmed for that mode. The factory default coefficient for each mode within both of the programmable storage fields is 1000 (recall that decimal points are not available, and thus the stored value of 1000 is actually 1.000). Use the supply/return button to increase the coefficients value, or the range selector button to decrease the value as desired. When the correct multiplier is displayed, press the handle hold switch to store the value. The display will then cycle to the Supply, Vent Open mode (indicated by the appropriate symbols), again allowing for increasing/decreasing of the multiplier to the desired set-point as described above. This process will repeat for the remaining two operating modes (Return, Vent Closed; and Return, Vent Open).
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 14
5. Once a correction factor has been selected and stored for each of the four operating modes, the unit will proceed into operation using the programmed coefficients of the storage field
that was selected. Once again, use of FLD A” or FLD B” will be indicated by the field letter
chosen being displayed at the upper right of the LCD throughout the operating cycle. The selected storage field will remain as the start­up default setting unless changed by the user.
6. To return to the factory calibration setting,
FAC”, press and hold the range selector
button while simultaneously switching the
instrument on. Select the “FAC” setting by
depressing the handle hold switch. The factory calibration has just been restored as the default
operating setting. Stored coefficients in “FLD A” and FLD B” will remain for future use.
NOTES:
• Always use the factory calibration setting,
FAC”, in determining the correction factors.
• Correction factors must be in the range of
0.250 to 1.999 (0250 to 1999 without the decimal point).
• Use of correction factors greater than 1.000
may cause the display to indicate a flow value at zero flow (typically 12-13 CFM at zero flow). This is normal and will limit the use of correction factors at the extreme low-end of the operating range.
• Do not use measurements which are less than
10 CFM * correction factor. For example, if you program a correction factor of 1.5, do not use measurements less than 15 CFM (10 * 1.5 =
15).
13
13
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 15
USA Tel: +1 800 874 2811 India Tel: +91 80 67877200
China Tel: +86 10 8219 7688 Singapore Tel: +65 6595 6388
Alnor Products, TSI Incorporated
Visit our website www.alnor.com for more information.
P/N 116-620-001 Rev. 08 ©1998–2014 TSI Incorporated Printed in U.S.A.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Loading...