TSI/Alnor 5815-AC User manual

Page 1
DP-CALC
MICROMANOMETER
MODEL 5815

OPERATION AND SERVICE MANUAL

P/N 1980567, REVISION E
www.GlobalTestSupply.com
Page 2
www.GlobalTestSupply.com
Page 3
Copyright TSI Incorporated / May 2007-2016 / All rights reserved.
LIMITATION OF WARRANTY AND LIABILITY (effective April 2014) Seller warrants the goods, excluding software, sold hereunder, under normal use and
service as described in the operator's manual, to be free from defects in workmanship and material for 24 months, or if less, the length of time specified in the operator's manual, from the date of shipment to the customer. This warranty period is inclusive of any statutory warranty. This limited warranty is subject to the following exclusions and exceptions:
a. Hot-wire or hot-film sensors used with research anemometers, and certain other
components when indicated in specifications, are warranted for 90 days from the date of shipment;
b. Pumps are warranted for hours of operation as set forth in product or operator’s
manuals;
c. Parts repaired or replaced as a result of repair services are warranted to be free
from defects in workmanship and material, under normal use, for 90 days from the date of shipment;
d. Seller does not provide any warranty on finished goods manufactured by others or
on any fuses, batteries or other consumable materials. Only the original manufacturer's warranty applies;
e. This warranty does not cover calibration requirements, and seller warrants only that
the instrument or product is properly calibrated at the time of its manufacture. Instruments returned for calibration are not covered by this warranty;
f. This warranty is VOID if the instrument is opened by anyone other than a factory
authorized service center with the one exception where requirements set forth in the manual allow an operator to replace consumables or perform recommended cleaning;
g. This warranty is VOID if the product has been misused, neglected, subjected to
accidental or intentional damage, or is not properly installed, maintained, or cleaned according to the requirements of the manual. Unless specifically authorized in a separate writing by Seller, Seller makes no warranty with respect to, and shall have no liability in connection with, goods which are incorporated into other products or equipment, or which are modified by any person other than Seller.
The foregoing is IN LIEU OF all other warranties and is subject to the LIMITATIONS stated herein. NO OTHER EXPRESS OR IMPLIED WARRANTY OF FITNESS FOR
PARTICULAR PURPOSE OR MERCHANTABILITY IS MADE. WITH RESPECT TO
SELLER’S BREACH OF THE IMPLIED WARRANTY AGAINST INFRINGEMENT,
SAID WARRANTY IS LIMITED TO CLAIMS OF DIRECT INFRINGEMENT AND EXCLUDES CLAIMS OF CONTRIBUTORY OR INDUCED INFRINGEMENTS.
BUYER’S EXCLUSIVE REMEDY SHALL BE THE RETURN OF THE PURCHASE PRICE DISCOUNTED FOR REASONABLE WEAR AND TEAR OR AT SELLER’S
OPTION REPLACEMENT OF THE GOODS WITH NON-INFRINGING GOODS. TO THE EXTENT PERMITTED BY LAW, THE EXCLUSIVE REMEDY OF THE USER
OR BUYER, AND THE LIMIT OF SELLER'S LIABILITY FOR ANY AND ALL LOSSES, INJURIES, OR DAMAGES CONCERNING THE GOODS (INCLUDING CLAIMS BASED ON CONTRACT, NEGLIGENCE, TORT, STRICT LIABILITY OR OTHERWISE) SHALL BE THE RETURN OF GOODS TO SELLER AND THE REFUND OF THE PURCHASE PRICE, OR, AT THE OPTION OF SELLER, THE REPAIR OR REPLACEMENT OF THE GOODS. IN THE CASE OF SOFTWARE, SELLER WILL
www.GlobalTestSupply.com
iii
Page 4
REPAIR OR REPLACE DEFECTIVE SOFTWARE OR IF UNABLE TO DO SO, WILL REFUND THE PURCHASE PRICE OF THE SOFTWARE. IN NO EVENT SHALL SELLER BE LIABLE FOR LOST PROFITS, BUSINESS INTERRUPTION, OR ANY SPECIAL, INDIRECT, CONSEQUENTIAL OR INCIDENTAL DAMAGES. SELLER SHALL NOT BE RESPONSIBLE FOR INSTALLATION, DISMANTLING OR REINSTALLATION COSTS OR CHARGES. No Action, regardless of form, may be brought against Seller more than 12 months after a cause of action has accrued. The goods returned under warranty to Seller's factory shall be at Buyer's risk of loss, and will be returned, if at all, at Seller's risk of loss.
Buyer and all users are deemed to have accepted this LIMITATION OF WARRANTY AND LIABILITY, which contains the complete and exclusive limited warranty of Seller. This LIMITATION OF WARRANTY AND LIABILITY may not be amended, modified or its terms waived, except by writing signed by an Officer of Seller.
Service Policy
Knowing that inoperative or defective instruments are as detrimental to TSI as they are to our customers, our service policy is designed to give prompt attention to any problems. If any malfunction is discovered, please contact your nearest sales office or
iv
www.GlobalTestSupply.com
Page 5
CONTENTS
CHAPTER 1 UNPACKING AND PARTS IDENTIFICATION ............ 1
CHAPTER 2 SETTING-UP ................................................................. 3
Supplying Power to the Model 5815 DP-Calc Micromanometer .... 3
Installing the Batteries ............................................................... 3
CHAPTER 3 OPERATION ................................................................. 5
Keypad Functions ........................................................................... 5
CHAPTER 4 MAINTENANCE ............................................................ 7
Recalibration ................................................................................... 7
Cases .............................................................................................. 7
Storage ........................................................................................... 7
CHAPTER 5 TROUBLESHOOTING .................................................. 9
APPENDIX A SPECIFICATIONS ..................................................... 11
www.GlobalTestSupply.com
v
Page 6
Chapter 1
Unpacking and Parts Identification
Carefully unpack the instrument and accessories from the shipping container. Check the individual parts against the list of components below. If anything is missing or damaged, notify TSI immediately.
1. Carrying case
2. Instrument
www.GlobalTestSupply.com
1
Page 7
Chapter 2
Setting-up

Supplying Power to the Model 5815 DP-Calc Micromanometer

The Model 5815 is powered with four size AA batteries.

Installing the Batteries

Insert four AA batteries as indicated by the diagram located on the inside of the battery compartment. The Model 5815 is designed to operate with either alkaline or NiMH rechargeable batteries, although it will not recharge NiMH batteries. Battery life will be shorter if NiMH batteries are used. Carbon-zinc batteries are not recommended because of the danger of battery acid leakage.
www.GlobalTestSupply.com
3
Page 8
ON/OFF Key
Press to turn the Model 5815 on and off. During the power up sequence the display will show the following: Model Number, Serial Number, Software Revision and Last Date Calibrated.
ft/min / m/s Key
Pressing this key changes the display to read air velocity, converted from velocity pressure when connected to a pitot probe.
in H2O / Pa / hPa / kPa / mm HG Key
Pressing this key changes the display to read differential pressure.
0 Key
Pressing the 0 key zeroes the pressure measurement.
Changing Units
To change units, first put the desired measurement (air velocity or pressure) on the display. Then press and hold the left, unlabelled key for five seconds. Finally, use the  and ENTER key to select the units.
Chapter 3
Operation

Keypad Functions

www.GlobalTestSupply.com
5
Page 9
U.S. & International
TSI Incorporated
Chapter 4
Maintenance
The Model 5815 DP-Calc Micromanometer requires very little maintenance to keep it performing well.

Recalibration

To maintain a high degree of accuracy in your measurements, we recommend that you return your Model 5815 to TSI for annual
recalibration. Please contact one of TSI’s offices or your local
distributor to make service arrangements and to receive a Return Material Authorization (RMA) number. To fill out an online RMA form,

Cases

If the instrument case or storage case needs cleaning, wipe it off with a soft cloth and isopropyl alcohol or a mild detergent. Never immerse the Model 5815. If the enclosure of the Model 5815 becomes broken, it must be replaced immediately to prevent access to hazardous voltage.

Storage

Remove the batteries when storing the unit for more than one month to prevent damage due to battery leakage.
7
www.GlobalTestSupply.com
Page 10
Symptom
Possible Causes
Corrective Action
No Display
Unit not turned on
Switch unit on.
Low or dead batteries
Replace batteries.
Dirty battery contacts
Clean the battery contacts.
Velocity reading fluctuates unstable
Fluctuating flow
Reposition probe in less-turbulent flow.
Instrument Error message appears
Fault in instrument
Factory service required on instrument.
Chapter 5
Troubleshooting
Table 5-1 lists the symptoms, possible causes, and recommended solutions for common problems encountered with the Model 5815. If your symptom is not listed, or if none of the solutions solves your problem, please contact TSI.
Table 5-1: Troubleshooting the Model 5815
WARNING!
The pressure sensor is protected from damage up to 7 psi (48 kPa or 360 mmHg). At higher pressure it can burst!
9
www.GlobalTestSupply.com
Page 11
Appendix A
Specifications
Specifications are subject to change without notice.
Static/Differential Pressure:
Range1: -15 to +15 in. H2O (-28.0 to +28.0 mm Hg, -3735 to
+3735 Pa) Accuracy: ±1% of reading ±0.005 in. H2O (±0.01 mm Hg, ±1 Pa) Resolution: 0.001 in. H2O (1 Pa, 0.01 mm Hg)
Velocity From a Pitot Tube:
Range2: 250 to 15500 ft/min (1.27 to 78.7 m/s) Accuracy3: ±1.5% at 2000 ft/min (10.16 m/s) Resolution: 1 ft/min (0.1 m/s)
Instrument Temperature Range:
Operating (Electronics): 40 to 113°F (5 to 45°C) Storage: -4 to 140°F (-20 to 60°C)
Instrument Operating Conditions:
Altitude up to 4000 meters Relative humidity up to 80% RH, non-condensing Pollution degree 1 in accordance with IEC 664 Transient over voltage category II
External Meter Dimensions:
3.3 in. 7.0 in. 1.8 in. (8.4 cm 17.8 cm 4.4 cm)
Meter Weight:
Weight with batteries: 0.6 lbs (0.27 kg)
Power Requirements:
Four AA-size batteries (included)
1
Overpressure range = 7 psi (190 in. H2O, 360 mmHg, 48 kPa).
2
Pressure velocity measurements are not recommended below 1,000 ft/min (5 m/s)
and are best suited to velocities over 2000 ft/min. Range can vary depending on barometric pressure.
3
Accuracy is a function of converting pressure to velocity. Conversion accuracy
improves when actual pressure values increase.
www.GlobalTestSupply.com
11