TSI/Alnor 4073, 4072, 4075, 4078, 4076 User manual

Page 1
CERTIFIER® FA
TEST SYSTEM

OPERATOR’S MANUAL

P/N 1980436, REVISION L
FEBRUARY 2017
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 2
Caution
TSI flowmeters are not medical devices under FDA 510(k) and in no situation should be used for human measurements.
LIMITATION OF WARRANTY AND LIABILITY (effective April 2014)
Seller warrants the goods, excluding software, sold hereunder, under normal use and service as described in the operator's manual, to be free from defects in workmanship and material for 12 months, or if less, the length of time specified in the operator's manual, from the date of shipment to the customer. This warranty period is inclusive of any statutory warranty. This limited warranty is subject to the following exclusions and exceptions:
a. Hot-wire or hot-film sensors used with research anemometers, and certain other
components when indicated in specifications, are warranted for 90 days from the date of shipment;
b. Pumps are warranted for hours of operation as set forth in product or operator’s
manuals;
c. Parts repaired or replaced as a result of repair services are warranted to be free from
defects in workmanship and material, under normal use, for 90 days from the date of shipment;
d. Seller does not provide any warranty on finished goods manufactured by others or on
any fuses, batteries or other consumable materials. Only the original manufacturer's warranty applies;
e. This warranty does not cover calibration requirements, and seller warrants only that
the instrument or product is properly calibrated at the time of its manufacture. Instruments returned for calibration are not covered by this warranty;
f. This warranty is VOID if the instrument is opened by anyone other than a factory
authorized service center with the one exception where requirements set forth in the manual allow an operator to replace consumables or perform recommended cleaning;
g. This warranty is VOID if the product has been misused, neglected, subjected to
accidental or intentional damage, or is not properly installed, maintained, or cleaned according to the requirements of the manual. Unless specifically authorized in a separate writing by Seller, Seller makes no warranty with respect to, and shall have no liability in connection with, goods which are incorporated into other products or equipment, or which are modified by any person other than Seller.
The foregoing is IN LIEU OF all other warranties and is subject to the LIMITATIONS stated herein. NO OTHER EXPRESS OR IMPLIED WARRANTY OF FITNESS FOR
PARTICULAR PURPOSE OR MERCHANTABILITY IS MADE. WITH RESPECT TO SELLER’S BREACH OF THE IMPLIED WARRANTY AGAINST INFRINGEMENT, SAID WARRANTY IS LIMITED TO CLAIMS OF DIRECT INFRINGEMENT AND EXCLUDES CLAIMS OF CONTRIBUTORY OR INDUCED INFRINGEMENTS. BUYER’S EXCLUSIVE REMEDY SHALL BE THE RETURN OF THE PURCHASE PRICE DISCOUNTED FOR
i
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 3
REASONABLE WEAR AND TEAR OR AT SELLER’S OPTION REPLACEMENT OF THE GOODS WITH NON-INFRINGING GOODS.
TO THE EXTENT PERMITTED BY LAW, THE EXCLUSIVE REMEDY OF THE USER OR BUYER, AND THE LIMIT OF SELLER'S LIABILITY FOR ANY AND ALL LOSSES, INJURIES, OR DAMAGES CONCERNING THE GOODS (INCLUDING CLAIMS BASED ON CONTRACT, NEGLIGENCE, TORT, STRICT LIABILITY OR OTHERWISE) SHALL BE THE RETURN OF GOODS TO SELLER AND THE REFUND OF THE PURCHASE PRICE, OR, AT THE OPTION OF SELLER, THE REPAIR OR REPLACEMENT OF THE GOODS. IN THE CASE OF SOFTWARE, SELLER WILL REPAIR OR REPLACE DEFECTIVE SOFTWARE OR IF UNABLE TO DO SO, WILL REFUND THE PURCHASE PRICE OF THE SOFTWARE. IN NO EVENT SHALL SELLER BE LIABLE FOR LOST PROFITS, BUSINESS INTERRUPTION, OR ANY SPECIAL, INDIRECT, CONSEQUENTIAL OR INCIDENTAL DAMAGES. SELLER SHALL NOT BE RESPONSIBLE FOR INSTALLATION, DISMANTLING OR REINSTALLATION COSTS OR CHARGES. No Action, regardless of form, may be brought against Seller more than 12 months after a cause of action has accrued. The goods returned under warranty to Seller's factory shall be at Buyer's risk of loss, and will be returned, if at all, at Seller's risk of loss.
Buyer and all users are deemed to have accepted this LIMITATION OF WARRANTY AND LIABILITY, which contains the complete and exclusive limited warranty of Seller. This LIMITATION OF WARRANTY AND LIABILITY may not be amended, modified or its terms waived, except by writing signed by an Officer of Seller.
Service Policy
Knowing that inoperative or defective instruments are as detrimental to TSI as they are to our customers, our service policy is designed to give prompt attention to any problems. If any malfunction is discovered, please contact your nearest sales office or representative, or call TSI's Customer Service department at (800) 874-2811 / (1) 651 490-2811 (USA and International) or TSI Instruments in UK at: +44 (0) 1494 4 59200.
ii
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 4
Contents
1 INTRODUCTION ........................................................................... 1
1.1 Parts List ............................................................................. 3
1.2 Glossary .............................................................................. 6
2 SETUP ........................................................................................... 7
3 OPERATION ............................................................................... 13
3.1 Power Up .......................................................................... 13
3.2 Keypad Functions ............................................................. 14
3.3 Displaying Test Measurements ........................................ 14
3.4 Measurements .................................................................. 15
3.4.1 Flow Rate .............................................................. 15
3.4.2 Peak Flow Rate ..................................................... 16
3.4.3 Volume .................................................................. 16
3.4.4 Minute Volume ...................................................... 16
3.4.5 Stacked Volume .................................................... 17
3.4.6 Oxygen Concentration .......................................... 17
3.4.7 Low Pressure ........................................................ 17
3.4.8 Peak Pressure ....................................................... 18
3.4.9 Peep Pressure....................................................... 18
3.4.10 Breath Rate ........................................................... 18
3.4.11 I:E Ratio ................................................................. 18
3.4.12 I Time .................................................................... 19
3.4.13 Absolute Pressure ................................................. 19
3.4.14 Oxygen Concentration .......................................... 19
3.5 Required Pre-test Calibrations .......................................... 20
3.5.1 Low-Pressure Transducer Zero Calibration .......... 20
3.5.2 Oxygen Sensor Calibration ................................... 20
3.6 Breathing Cycles and Trigger Levels ................................ 21
3.6.1 Inhalation and Exhalation Timing .......................... 21
3.6.2 Flow Trigger Levels ............................................... 21
3.7 Display Information ........................................................... 22
4 TROUBLESHOOTING ................................................................ 25
5 MAINTENANCE .......................................................................... 29
5.1 Replacing the Batteries (as required) ............................... 29
5.2 Replacing the Oxygen Sensor (yearly) ............................. 29
iii
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 5
5.3 Cleaning (as required) ....................................................... 29
5.4 Factory Calibration (yearly) ............................................... 30
5.5 Return Procedure .............................................................. 30
6 SPECIFICATIONS....................................................................... 31
6.1 Physical ............................................................................. 31
6.2 Environmental ................................................................... 31
6.3 Power ................................................................................ 31
6.4 Test Measurements .......................................................... 31
6.5 Calibration Requirements.................................................. 33
6.6 Compliance and Approvals ............................................... 33
List of Figures
Figure 1. The Certifier Figure 2. Certifier
Figure 3. Connecting the Controller Module to a Flow Module ..... 7
Figure 4. Connecting the Bacteria Filters to the Flow Modules .... 8
Figure 5. Attaching Pressure Tubing to the High Flow Module ..... 9
Figure 6. Attaching the Oxygen Sensor to the High Flow
Module ......................................................................... 10
Figure 7. Installing a Flow Module into the Test Circuit .............. 11
Figure 8. Certifier® FA Test System Installed in an Oxygen
Concentrator Circuit .................................................... 12
Figure 9. Controller Module Keypad ............................................ 14
®
FA Test System ....................................... 2
®
FA Test System Parts ..................................... 3
List of Tables
Table 1. Certifier
Table 2. Keypad Functions and Operation .................................. 14
Table 3. Screen Displays ............................................................. 22
Table 4. Troubleshooting the Certifier
Table 5. Cleaning Recommendations ......................................... 29
iv
®
FA Test System Parts List ............................... 4
®
FA Test System ............ 25
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 6

1 Introduction

The Certifier® Flow Analyzer (FA) Test System allows you to test respiratory care or other devices. This portable tester makes it simple to test flows, volumes, pressures, oxygen concentration, and breath timing. The Certifier® FA Test System is designed for institutional, home care, field service, and laboratory settings.
Certifier® FA Test System components include:
Controller module:
The keypad and display allow you to select test measurements and units for display. The controller module connects to a high or Low Flow module.
High Flow module:
Measures air or 100% oxygen (O2) over a range of flows from 0 to 300 standard liters per minute (SLPM).
Low Flow module:
Measures air, 100% O2, or 100% nitrous oxide (N2O) over a range of flows from 0.01 to 15 SLPM with greater accuracy than the High Flow module at low flow rates.
Oxygen sensor:
Used with the High Flow module, allows the High Flow module to measure O2 concentration and other measurements for any mixture of air and O2.
You can connect or disconnect the flow modules and oxygen sensor at any time during normal operation without interrupting tester operation. Four AA batteries power the test system. The test system conserves power by automatically turning off if none of the keys are pressed for 15 minutes. See Section 3.1 to disable this feature.
1
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 7
2
Controller module
Low Flow module
High Flow module
Oxygen sensor
Figure 1. The Certifier® FA Test System
WARNING
To avoid the risk of explosion, do not use in the presence of
flammable anesthetic gases.
Only qualified and trained service technicians are authorized to
service the Certifier® FA Test System. Use only factory-approved parts and procedures to service the device.
CAUTION
To avoid inaccurate test readings, do not obstruct tubing or inlet
or exhaust ports, and always use dry gas.
To avoid damage to the Certifier® FA Test System components,
always use bacteria filters upstream of the flow modules, and always cap flow module ports when not in use.
TSI flowmeters are not medical devices under FDA 510(k) and in
no situation should be used for human measurements.
1: Introduction
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 8
3

1.1 Parts List

Carefully unpack the test system components from the shipping container. Check the individual parts against the packing list and notify TSI immediately if anything is missing or damaged. Table 1 summarizes the Certifier® FA Test System components and part numbers shown in Figure 2.
Figure 2. Certifier® FA Test System Parts
1: Introduction
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 9
4
Table 1. Certifier® FA Test System Parts List
Item
no.
Description
Part no.
Qty
High Flow standard kit (part no. 4070)
1
Controller module
4078
1 2 High Flow module
4071
1
3
Bacteria filter, 22-mm x 22-mm male/female, for use with High Flow module (single use)
1602341
1
4
Soft carrying case (holds Certifier® FA Test System and accessories)
1319289
1
5
Adapter, 15-mm ID x 22-mm OD
1102093
1 6 Airway pressure fitting with screen
1611330
1
7
Adapter, 22-mm x 6-mm (for interfacing High Flow module to Low Flow filter, for use with oxygen concentrator)
1102091
2
8
Pressure tubing, 1/8-in. ID x 1/4-in. OD x 48-in. length, silicone
3002053
1
9
Mounting bracket (includes mounting bracket, screws, and Velcro strap)
1040044
1
10
Pocket driver (used to remove/install the controller module battery cover)
3012034
1 11
AA batteries (not shown)
NA
4
12
Certifier® FA Test System Operator’s Manual (not shown)
1980436
1
1: Introduction
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 10
1: Introduction
5
Table 1. Certifier® FA Test System Parts List (cont)
Item
no.
Description
Part no.
Qty
Low Flow standard kit (part no. 4075)
1
Controller module
4078
1
13
Low Flow module
4074
1
14
Bacteria filter, barbed fittings, for use with Low Flow module (single use)
1040045
1 4 Soft carrying case (holds Certifier® FA Test
System and accessories)
1319289
1 9 Mounting bracket (includes mounting
bracket, screws, and Velcro strap)
1040044
1
10
Pocket driver (used to remove/install the controller module battery cover)
3012034
1 11
AA batteries (not shown)
NA
4
12
Certifier® FA Test System Operator’s Manual (not shown)
1980436
1
High Flow module kit (optional) (part no. 4076)
2
High Flow module
4071
1
3
Bacteria filter, 22-mm x 22-mm male/female, for use with High Flow module (single use)
1602341
1
5
Adapter, 15-mm ID x 22-mm OD
1102093
1 6 Airway pressure fitting with screen
1611330
1
7
Adapter, 22-mm x 6-mm (for interfacing High Flow module to Low Flow filter, for use with oxygen concentrator)
1102091
2
8
Pressure tubing, 1/8-in. ID x 1/4-in. OD x 48-in. length, silicone
3002053
1 9 Mounting bracket (includes mounting
bracket, screws, and Velcro strap)
1040044
1
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 11
6
Table 1. Certifier® FA Test System Parts List (cont)
Item
no.
Description
Part no.
Qty
Low Flow module kit (optional) (part no. 4072)
13
Low Flow module
4074
1
14
Bacteria filter, barbed fittings, for use with Low Flow module (single use)
1040045
1 9 Mounting bracket (includes mounting
bracket, screws, and Velcro strap)
1040044
1
Oxygen sensor kit (optional) (part no. 4073)
15
Oxygen sensor
2917019
1
16
Threaded tee
1313118
1
17
Cable, 8-in. length
1303741
1
Other accessories (optional)
18
Hard shell carrying case (holds Certifier® FA Test System and accessories)
1319288
1

1.2 Glossary

Refer to manual: see Certifier® FA Test System Operator’s Manual for important information.
Safety approvals for Canada by Underwriter’s Laboratories Inc. (UL).
These labels, terms, and symbols appear on the Certifier® FA Test System:
See Section 3.7 for definitions of symbols and abbreviations that appear on the Certifier® FA display.
1: Introduction
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 12
Controller module
Flow
module

2 Setup

Follow these steps to set up the Certifier® FA Test System:
CAUTION
To avoid damage to the Certifier® FA Test System components, always use bacteria filters upstream of the flow modules, and always cap flow module ports when not in use.
1. Connect the controller module to a flow module (Figure 3).
To remove the cable, pull its locking connector (not the cable) from the controller module.
Figure 3. Connecting the Controller Module to a Flow Module
7
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 13
2. Attach the bacteria filter to the flow module (Figure 4).
Controller module
High Flow module (note flow direction)
Low Flow module (note flow direction)
High Flow filter
Low Flow filter
Install the bacteria filter to the flow module inlet or upstream of all Certifier® FA Test System components. Make sure the filter is in the correct orientation by aligning the filter’s flow arrow with the direction of the flow, ensuring that the filter’s inlet indicator (labeled “INLET”, “I”, or other) faces upstream of the flow module, or that the outlet indicator (label “Patent side”, “Outlet” or other) faces towards the flow module.
Figure 4. Connecting the Bacteria Filters to the Flow Modules
8 2: Setup
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 14
3. High Flow module only: install pressure tubing (Figure 5).
Pressure tubing
High Flow module
Attach one end of the pressure tubing to the low-pressure port on the flow module, and the other to the pressure port in the circuit.
Figure 5. Attaching Pressure Tubing to the High Flow Module
2: Setup 9
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 15
4. High Flow module only: install the oxygen sensor (optional,
High Flow module
Threaded tee
Oxygen sensor
Oxygen sensor
cable
Figure 6).
Plug the oxygen sensor cable into the High Flow module and oxygen sensor. Turn the cable collar to secure the cable to the oxygen sensor. Use the threaded tee to install the oxygen sensor into the circuit. The oxygen sensor can only connect to the High Flow module.
Figure 6. Attaching the Oxygen Sensor to the High Flow Module
10 2: Setup
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 16
5. Install the flow module into the test circuit (Figure 7, Figure 8).
Ventilator
Align the flow direction of the flow module and filter with the direction of flow through the circuit.
Figure 7. Installing a Flow Module into the Test Circuit
2: Setup 11
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 17
Oxygen Concentrator
Figure 8. Certifier® FA Test System Installed in an Oxygen
Concentrator Circuit
12 2: Setup
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 18

3 Operation

3.1 Power Up

Pull the protective caps from the flow module ports before powering up. Do not apply pressure to the flow module at power up (this ensures accurate low-pressure transducer zero calibration).
Press the I/O (on/off) key on the controller module to power up the
Certifier® FA Test System (the controller module powers all of the attached Certifier® FA Test System components). At power up, the controller module shows information in this sequence:
1. All LCD segments light (about two seconds).
NOTE: If battery voltage is below the minimum operating level, the battery symbol lights and the controller module will turn off.
2. Firmware revision shown (about two seconds).
3. Controller module verifies connection and correct operation of the flow module and oxygen sensor (if installed). If the oxygen
sensor is installed, the controller module will display a symbol.
4. Default measurements displayed:
High Flow module: flow (top line), low pressure (bottom line). Low Flow module: flow (top line), respiratory rate (bottom line).
CAUTION
To ensure accurate measurements, wait about one minute for the
Certifier® FA Test System to warm up. If environmental conditions have changed significantly, more time may be necessary.
To avoid damage to the Certifier® FA Test System components,
always use bacteria filters upstream of the flow modules, and always cap flow module ports when not in use.
If liquid has penetrated any of its components, do not use, and
return to the factory for calibration.
The test system conserves battery life by automatically turning off if none of the keys are pressed for 15 minutes. To override the automatic turn off, press the GAS SELECT key with the I/O key when turning on. The will appear while the software revision is displayed during power up if the automatic turn off is disabled.
13
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 19

3.2 Keypad Functions

Top line
select key
Bottom line
select key
ON/OFF
key
DISPLAY UNITS key
GAS SELECT
key
Backlight
on/off key
Table 2. Keypad Functions
Key
Primary Function
Top line select
Selects the measurement shown on the
top line of the display.
Bottom line select
Selects the measurement shown on the
bottom line of the display.
ON/Off
Turns the system on (I) or off (O).
DISPLAY UNITS
Selects the unit of measure for
measurements.
GAS SELECT
Selects the supply gas type for the Certifier
®
FA.
Backlight on/off
Turns the controller module display
backlight on or off.
Table 2 summarizes the primary functions of the Certifier® FA Test System keypad (Figure 9).
Figure 9. Controller Module Keypad

3.3 Displaying Test Measurements

Follow these steps to display test measurements:
1. Set up the Certifier® FA Test System and install it into the circuit.
2. Press the I/O (on/off) key to power up the system.
14 3: Operation
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 20
3. Once power up is complete, press the Top line select and Bottom line select keys to select the measurements to be
displayed. See Section 3.4, Measurements, for details on each parameter.
You can change the display selections and connect or disconnect the flow module or oxygen sensor at any time during normal operation.
If you disconnect the flow module, the controller module display
is blank.
If you disconnect the oxygen sensor, the sensor symbol and any
oxygen-related measurements are not shown on the display.
Normal operation and display resume about one second after
reconnection.
NOTE: The Certifier® FA Test System automatically updates measurements that are recalculated at every breath. The Certifier® FA Test System uses a threshold flow to determine the beginning and end of each breath. The threshold flow calculated from the last three breaths, is updated at each breath, and then is applied to the following breath. Breath measurement displays begin after two full breaths.
CAUTION
To avoid damage to the Certifier® FA Test System components, always use bacteria filters upstream of the flow modules, and always cap flow module ports when not in use.

3.4 Measurements

Measurements on Top Line of Display
The following parameters are displayed on the top line of the display.
The Top Line Select key is used to scroll through parameters by momentarily pressing and then releasing the key.

3.4.1 Flow Rate

The flow rate can be displayed on the top line of control module display. Units of standard liters per minute, indicated by “SLPM”, and actual liters per minute, indicated by “LPM”, can be selected using the DISPLAY UNITS key. For the High Flow module air, O2, and air/O2 mixture (when oxygen sensor is attached) can be selected using the GAS SELECT key. For the Low Flow module air, O2, and N2O can be selected using the GAS SELECT key.
3:Operation 15
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 21

3.4.2 Peak Flow Rate

Maximum flow rate during the inhalation cycle of a breath can be displayed on the top line of the control module display. Units of standard liters per minute, indicated by “PEAK SLPM”, and actual liters per minute, indicated by “LPM”, can be selected using the DISPLAY UNITS key. See Section 3.7, Display Information, for definitions of these units of measure. For the High Flow module air, O2, and air/O2 mixtures can be selected using the GAS SELECT key. For the Low Flow module air, O2, and N2O can be selected using the GAS SELECT key.
NOTE: A valid flow cycle must occur for this value to be displayed. See Section 3.6, Breathing Cycles and Trigger Levels, for details.

3.4.3 Volume

Volume that occurs during the inhalation cycle of a breath can be displayed on the top line of the control module display. Units of actual liters, indicated by “ATP L”; standard liters, indicated by “STP L”; or liters at body temperature and pressure saturated, indicated by “BTPS L” can be selected using the DISPLAY UNITS key. See Section 3.7, Display Information, for definitions of these units of measure. For the High Flow module air, O2, and air/O2 mixture can be selected using the GAS SELECT key. For the Low Flow module air, O2, and N2O can be selected using the GAS SELECT key.
NOTE: A valid flow cycle must occur for this value to be displayed. See Section 3.6, Breathing Cycles and Trigger Levels, for details.

3.4.4 Minute Volume

Minute volume calculated from the inhalation cycle of the last breath and breath rate can be displayed on the top line of the control module display. Units of “MINUTE VOL ATP L” (actual liters), “MINUTE VOL STP L” (standard condition liters), or “MINUTE VOL BTPS L” (liters at body temperature and pressure saturated) can be selected using the DISPLAY UNITS key. See Section 3.7, Display Information, for definitions of these units of measure. For the High Flow module air, O2, and air/O2 mixture can be selected using the GAS SELECT key. For the Low Flow module air, O2, and N2O can be selected using the GAS SELECT key.
NOTE: A valid flow cycle must occur for this value to be displayed. See Section 3.6, Breathing Cycles and Trigger Levels, for details.
16 3: Operation
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 22

3.4.5 Stacked Volume

Total volume over a displayed number of inhalation cycles can be displayed on the top line of the control module display with the number of inhalation cycles displayed on the bottom line of the display. Units of actual liters, indicated by “ATP S L”; standard liters, indicated by “STP S L”; or liters at body temperature and pressure saturated, indicated by “BTPS S L” can be selected using the DISPLAY UNITS key. See Section 3.7, Display Information, for definitions of these units of measure. For the High Flow module air, O2, and air/O2 mixture can be selected using the GAS SELECT key. For the Low Flow module air, O2, and N2O can be selected using the GAS SELECT key.
NOTE: A valid flow cycle must occur for this value to be displayed. See Section 3.6, Breathing Cycles and Trigger Levels, for details.

3.4.6 Oxygen Concentration

When the oxygen sensor is attached to the High Flow module the oxygen concentration “%O2” can be displayed on the top line of the control module display. Oxygen concentration can also be displayed on the bottom line of the display, see Section 3.4.14.
Daily calibrations need to be done on the oxygen sensor. See Section 3.5.2 for instructions on oxygen sensor calibration.
NOTE: The symbol will appear on the display when the oxygen cable is attached.
Measurements on Bottom Line of Display
The following parameters are displayed on the bottom line of the display. The Bottom line select key is used to scroll through parameters by momentarily pressing and then releasing the key.

3.4.7 Low Pressure

Gauge pressure from the low-pressure port of the High Flow module can be displayed on the bottom line of the control module display. Select units of “cmH2O” or “mmHg” by pressing and holding the DISPLAY UNITS key for at least three seconds.
For best results check the zero on the low pressure transducer before measurements. See Section 3.5.1 for instructions on zeroing the pressure transducer.
3:Operation 17
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 23
NOTE: For distinguishing low pressure from absolute pressure, the resolution for low pressure is in 0.1 cmH2O or mmHg and absolute pressure measurement (Section 3.4.13) is displayed in resolution of 1 mmHg.

3.4.8 Peak Pressure

Peak gauge pressure from the low-pressure port of the High Flow module during the inhalation cycle can be displayed on the bottom line of the control module display. Select units of “PEAK cmH2O” or “PEAK mmHg” by pressing and holding the DISPLAY UNITS key for at least three seconds.
NOTE: A valid flow cycle must occur for this value to be displayed. See Section 3.6, Breathing Cycles and Trigger Levels, for details.
For best results check the zero on the low pressure transducer before measurements. See Section 3.5.1 for instructions on zeroing the pressure transducer.

3.4.9 Peep Pressure

Positive end expiratory pressure (PEEP) from the low-pressure port of the High Flow module can be displayed on the bottom line of the control module display. Select units of “PEEP cmH2O” or “PEEP mmHg” by pressing and holding the DISPLAY UNITS key for three seconds.
For best results check the zero on the low pressure transducer before measurements. See Section 3.5.1 for instructions on zeroing the pressure transducer.
NOTE: A valid flow cycle must occur for this value to be displayed. See Section 3.6, Breathing Cycles and Trigger Levels, for details.

3.4.10 Breath Rate

The breaths per minute (BPM) can be displayed on the bottom line of the control module display.
NOTE: A valid flow cycle must occur for this value to be displayed. See Section 3.6, Breathing Cycles and Trigger Levels, for details.

3.4.11 I:E Ratio

The ratio of the inhalation time and exhalation time (I:E RATIO) can be displayed on the bottom line of the control module display.
NOTE: A valid flow cycle must occur for this value to be displayed. See Section 3.6, Breathing Cycles and Trigger Levels, for details.
18 3: Operation
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 24
NOTE: The inhalation time is defined as the time of positive inhalation flow and does not include the breath hold time. If the vents inhalation time includes the inhalation pause time, the Certifier® FA’s I:E ratio will not match the vents I:E ratio.

3.4.12 I Time

The inhalation time (I TIME) can be displayed on the bottom line of the control module display.
NOTE: A valid flow cycle must occur for this value to be displayed. See Section 3.6, Breathing Cycles and Trigger Levels, for details.
NOTE: The inhalation time is defined as the time of positive inhalation flow and does not include the breath hold time. If the vents pause inhalation time includes a pause time, the Certifier® FA’s I time will not match the vents I time.

3.4.13 Absolute Pressure

The absolute pressure in the flow tubes can be displayed on the bottom line of the control module display in units of “mmHg”.
NOTE: For distinguishing absolute pressure from low pressure, the resolution for absolute pressure measurement is displayed in a resolution of 1 mmHg and low pressure (Section 3.4.7) is displayed in a resolution of 0.1 cmH2O or mmHg.

3.4.14 Oxygen Concentration

When the oxygen sensor is attached to the High Flow module, the oxygen concentration “%O2” can be displayed on the bottom line of the control module display. Oxygen concentration can also be displayed on the top line of the display. See Section 3.4.6.
Daily calibrations need to be done on the oxygen sensor. See
Section 3.5.2 for instructions on oxygen sensor calibration.
NOTE: The symbol will appear on the display when the oxygen
cable is attached.
3:Operation 19
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 25

3.5 Required Pre-test Calibrations

3.5.1 Low-Pressure Transducer Zero Calibration

The Certifier® FA Test System automatically performs a low-pressure transducer zero calibration at power up. Check the low pressure zero by disconnecting the pressure tubing from the flow module before each low-pressure measurement after initial power up to ensure the most accurate readings. If low pressure is not reading zero, perform the following steps to zero the transducer.
1. Disconnect the pressure tubing from the flow module to expose the flow module to ambient air.
2. Momentarily press the Bottom line select key until low pressure
(see Section 3.4.7) is shown on the display.
3. Press and hold the Bottom line select key for 2 to 3 seconds.
The display shows ZERO to indicate that the zero calibration is in progress.
4. When ZERO is no longer displayed, the low-pressure transducer zero calibration is complete.
NOTE: The barometric pressure transducer does not require a zero calibration.

3.5.2 Oxygen Sensor Calibration

Follow these steps daily and following an altitude change or sensor replacement to calibrate the oxygen sensor:
1. Power up the Certifier® FA Test System with the High Flow module and oxygen sensor attached, then allow about one minute to warm up. The oxygen sensor symbol flashes if the controller module detects the sensor needs calibration or has expired.
2. Momentarily press the Top line select or Bottom line select key
to show %O2 on the display.
3. Press and hold the GAS SELECT key for 2 to 3 seconds. Expose
the oxygen sensor to room air when you see 21.0 %O2 and CAL on the display. The 21% calibration is complete when 100.0 %O2 lights up on the top line and CAL flashes on the display. This can take several minutes while the oxygen concentration and sensor stabilize.
4. Expose the oxygen sensor to 100% oxygen of at least 5 liters per
minute. Press and release the GAS SELECT key to begin the
100% calibration. CAL stops flashing when the 100% calibration begins.
20 3: Operation
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 26
5. If the calibration is successful, CAL disappears and the O2 concentration is shown on the display. This can take several minutes while the oxygen concentration and sensor stabilize.
6. If the calibration is not successful (oxygen sensor symbol continues to flash, no O2 concentration is shown), repeat the calibration. If the repeated calibration is not successful, replace the oxygen sensor and repeat.
7. Expose the oxygen sensor to room air. The Certifier® FA Test System is ready to use when the %O2 reading returns to approximately 21%.

3.6 Breathing Cycles and Trigger Levels

The Certifier® FA uses flow rate to trigger the beginning and the end of a ventilator’s inhalation cycle.

3.6.1 Inhalation and Exhalation Timing

At the beginning of the inhalation cycle the flow rate must be above the trigger flow rate for at least 0.25 milliseconds. If this time is less than 0.25 milliseconds, the Certifier® FA considers this part of the exhalation cycle. When a valid inhalation cycle is ending and the flow rate goes below the trigger level, the flow rate must remain below the trigger level for at least 0.25 milliseconds, otherwise, this period is included in the inhalation cycle.
If testing of ventilator pressure modes is needed, a test lung must be used to ensure that a long enough flow delivery time is generated. If there is not enough volume in the breathing circuit, the pressure can be generated in less than 0.25 milliseconds.

3.6.2 Flow Trigger Levels

The default trigger level is automatically set on power up at the 20% point from the minimum flow to the peak flow. For example, if the maximum flow rate is 80 L/min and the minimum flow is 5 L/min, then the trigger flow rate is set to (80 L/min – 5 L/min) x 20% + 5 L/min = 20 L/min. The 20% auto trigger level will work for most ventilators, but some manufactures may instruct you to select a different trigger level.
A 10% of the peak flow rate and trigger can also be selected. The 10% auto trigger does not use the minimum flow rate in calculating the trigger level; zero is assumed for the minimum flow rate. In the above example the trigger level would be 80 L/min x 10% = 8 L/min.
Manually set trigger levels can also be used. Trigger can be manually set at 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 12, 15, 17, 20, 22, 24, 30, 35, or 40 L/min.
3:Operation 21
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 27
Follow these steps to view or adjust the flow trigger level.
Table 3. Screen Displays
Display
Meaning
- - -
An out-of-range measurement.
Oxygen sensor symbol: indicates presence of oxygen sensor. Flashes to indicate that sensor must be calibrated or replaced.
Low battery voltage symbol: indicates that batteries should be replaced.
%O2
Oxygen concentration can be displayed on either line if High Flow module and oxygen sensor are attached.
AIR
Air supply gas, selected using the GAS SELECT
key (when High Flow or Low Flow module is attached).
AIR O2
Mixed air and oxygen supply gas, selected using the
GAS SELECT key (when High Flow module and
oxygen sensors are attached).
1. Momentarily press the Top line select key until volume
parameter is on the display, see Section 3.4.3.
2. Press and hold the Top line select key until the “trig” appears on
the bottom of the display.
3. The trigger value will appear on the top line of the display which will be one of the following: “20 PEAK” (20% Auto trigger), “1 LPM”, “2 LPM”, “3 LPM”, “4 LPM”, “5 LPM”, “6 LPM”, “7 LPM”, “8 LPM”, “9 LPM”, “10 LPM”, “12 LPM”, “15 LPM”, “17 LPM”, “20 LPM”, “22 LPM”, “25 LPM”, “30 LPM”, “40 LPM”, or “10 PEAK” (10% Auto trigger).
4. To adjust the value, press and release the Top line select or Bottom line select to scroll through the list of values in step 3.
After 3 seconds the new trigger level will be set and the interface module will return to displaying volume.
NOTE: The default value of 20% auto trigger is restored once the controller module is turned off.

3.7 Display Information

Table 3 describes information that can appear on the controller
module screen.
22 3: Operation
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 28
Table 3. Screen Displays (cont.)
Display
Meaning
ATP
Atmospheric temperature and pressure: a condition of volume measurement. The Certifier® FA Test System calculates the ATP value by applying the actual gas temperature and pressure to the STP measurement.
BPM
Breaths per minute: a unit of respiratory rate. Can be displayed on the bottom line (either High Flow or Low Flow module attached).
BTPS
Body temperature and pressure, saturated: a condition of volume measurement. The Certifier® FA Test System calculates a BTPS value by compensating the STP measurement for BTPS conditions (37 °C (98.6 F), ambient pressure, 100% relative humidity).
CAL 100%
O2
Oxygen sensor calibration in progress (during exposure to 100% O2).
CAL 21% O2
Oxygen sensor calibration in progress (during exposure to room air).
cmH2O
Centimeters of water: a unit of pressure.
I TIME
Inspiratory time (in seconds). Can be displayed on the bottom line (either High Flow or Low Flow module attached).
I:E RATIO
Ratio of inspiratory time to expiratory time, can be displayed on the bottom line (when High Flow or Low Flow module is attached).
L
Liter: a unit of volume.
LPM
Liters per minute: a unit of flow. The Certifier® FA Test System calculates the LPM value by applying the actual gas temperature and pressure to the SLPM measurement.
MINUTE VOL
Minute volume: an estimate of exhaled volume for the next 60 seconds, based on the current breath. Can be displayed on the bottom line (when High Flow or Low Flow module is attached).
mmHg
Millimeters of mercury: a unit of pressure.
N2O
100% nitrous oxide supply gas, selected using the
GAS SELECT key (if Low Flow module is attached).
3:Operation 23
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 29
24 3: Operation
Table 3. Screen Displays (cont.)
Display
Meaning
O2
100% oxygen supply gas, selected using the GAS SELECT key (when High Flow or Low Flow module
is attached).
PEAK
Peak flow or pressure. Peak flow can be displayed on top line (when High Flow or Low Flow module is attached). Peak pressure can be displayed on bottom line (if High Flow module is attached).
PEEP
Positive end expiratory pressure: the minimum pressure measured in the circuit throughout the breath cycle. Can be displayed on the bottom line (if High Flow module attached).
REV
Software revision level, displayed at power up.
S
Stacked volume: a cumulative volume measurement for consecutive breaths. The top line shows the cumulative volume and the bottom line shows the number of consecutive breaths.
SEC
Seconds, a unit of time for I TIME measurements.
SLPM
Standard liters per minute: a unit of flow. The Certifier® FA Test System measures flows in SLPM. Standard conditions are defined as 21.1 C (70 F) at 101.3 kPa (14.7 psia).
STP
Standard temperature and pressure: a condition of volume measurement. The Certifier® FA Test System measures volumes at STP. Standard conditions are defined as 21.1 C (70 F) at
101.3 kPa (14.7 psia).
VOL
Volume: the exhaled volume for the most recent breath. Can be displayed on the top line (either High Flow or Low Flow module attached).
ZERO
Low-pressure transducer zero calibration in progress.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 30
Table 4. Troubleshooting the Certifier® FA Test System
Symptom
Possible Cause
Corrective Action
Controller module won’t turn on, or turns on and off.
Batteries are depleted or installed backwards.
Check that batteries are installed correctly. Replace batteries.
Measurements aren’t displayed even though controller module is on.
Flow module isn’t connected to controller module.
Connect flow module to controller module.
‘---‘ is shown on display.
Measurement is out of range.
Check range for displayed measurement, and only make measurements within that range.
Flow rate does not read zero when no gas flowing (Especially when set to N2O).
Meter was not purged with gas displayed on the controller.
Purge meter with gas displayed on controller or press gas select key to change to desired gas.
Unable to disconnect flow module from controller module.
Pulling on the cable rather than the connector.
Pull the locking connector (not the cable) to disengage connector lock.
Can’t display %O2.
Oxygen sensor not connected. Look for oxygen sensor
symbol, .
Connect oxygen sensor cable to High Flow module and oxygen sensor.

4 Troubleshooting

Table 4 lists the symptoms, possible causes, and recommended
corrective actions for problems you may encounter with the Certifier® FA Test System. If the symptom is not listed, or if none of the recommended corrective actions solves the problem, please contact TSI Customer Support at (800) 874-2811 or 651-490-2811.
25
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 31
Table 4. Troubleshooting the Certifier® FA Test System (cont.)
Symptom
Possible Cause
Corrective Action
Can’t select AIR O2 mixtures.
Oxygen sensor not connected. Look for oxygen sensor
symbol, .
Connect oxygen sensor cable to High Flow module and oxygen sensor.
Volume, minute volume, peak flow, peak pressure, PEEP, respiratory rate, or I:E ratio measurement isn’t updated.
Less than two consecutive full breaths have been supplied to flow module, or flow is not supplied as a breathing waveform.
Bias flow is greater than auto trigger level.
Inhalation cycle is less than 0.25 milliseconds
Testing pressure mode on ventilator and no test lung attached.
Wait for at least two consecutive full breaths to be supplied to the flow module.
Ensure that flow is supplied as a breathing waveform.
Manually set flow trigger level.
Certifier® FA cannot measure shorter than
0.25 millisecond flow delivery.
Use a test lung so that at least 0.25 milliseconds of flow delivery is generated to generate pressure.
Can’t change measurement units.
Measurement isn’t displayed while changing units.
Press the Top line
select or the Bottom line select key to
display measurement, then press the
DISPLAY UNITS key to
change units.
Can’t zero low­pressure transducer.
Transducer is connected to a pressure source.
Disconnect pressure tubing from flow module, then zero low­pressure transducer.
Can’t zero barometric pressure transducer.
Barometric pressure transducer does not require a zero calibration.
Resume normal system operation.
Find Quality Products Online at: sales@GlobalTestSupply.com
26 4: Troubleshooting
www.GlobalTestSupply.com
Page 32
Table 4. Troubleshooting the Certifier® FA Test System (cont.)
Symptom
Possible Cause
Corrective Action
Oxygen sensor calibration takes longer than 5 minutes.
21% oxygen and/or 100% oxygen not supplied for calibration.
Flow of 100% oxygen is too low.
Verify that calibration gases are 21% oxygen and 100% oxygen and repeat calibration.
Increase flow of 100% oxygen to at least 5 SLPM and repeat calibration.
Oxygen sensor calibration fails.
21% oxygen and/or 100% oxygen not supplied for calibration.
Oxygen sensor is expired.
Verify that calibration gases are 21% oxygen and 100% oxygen and repeat calibration.
Replace oxygen sensor.
4:Troubleshooting 27
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 33
Table 5. Cleaning Recommendations
Component
Cleaning
Controller module
Flow modules
Carrying cases
Oxygen sensor
Oxygen sensor cables
Clean exterior as required with a clean cloth and isopropyl alcohol, hydrogen peroxide (3%), or ammonia (15%).
Tee
Adapters
Steam autoclave after contact with any non-sterile breathing circuit components, and discard if any damage is visible.
Single use filters (high
and Low Flow modules)
Discard after contact with any non­sterile breathing circuit components or if damage is visible.

5 Maintenance

5.1 Replacing the Batteries (as required)

Replace the batteries when the low battery voltage symbol is displayed or instrument will not power up.
1. Turn off the controller module.
2. Use the pocket driver tool (supplied) to loosen the screw that holds the battery cover on the controller module back panel.
3. Remove the old batteries from the battery compartment, and install new batteries noting the polarity indicators.
4. Reinstall the battery cover and tighten the screw to hold the cover to the controller module.

5.2 Replacing the Oxygen Sensor (yearly)

The oxygen sensor will function for one year of normal operation if use begins before the expiration date. Replace the oxygen sensor every year of normal use, or if the sensor cannot be calibrated or sensor readings are erratic.

5.3 Cleaning (as required)

Table 5 summarizes recommended cleaning methods for
Certifier® FA Test System components.
29
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 34
Dimensions
Controller module: 13.2 cm x 7 cm x 3.3 cm (5.2 in. x 2.8 in. x 1.3 in.).
High Flow module: 18 cm x 6.6 cm x 4 cm (7.1 in. x 2.6 in. x 1.6 in.).
Low Flow module: 12.7 cm x 5.1 cm x 2.8 cm (5.0 in. x 2.0 in. x 1.1 in.).
Flow connectors
High Flow module:
flow inlet: 22-mm female ISO taper.
flow outlet: 22-mm male ISO taper.
Low Flow module:
flow inlet: 0.25-in. barb.
flow outlet: 0.25-in. barb.
Weight
Approximately 1.36 kg (3 lb) for the standard kits.
Temperature
Operating: 5 to 40 C (41 to 104 F). 15 to 80% relative humidity from 5 to 31 C decreasing linearly to 15 to 50% relative humidity at 40 C.
Storage: -40 to 70 C (-40 to 158 F) at 10 to 95% relative humidity.
Atmospheric Pressure
Operating: 57.1 to 106 kPa (8.28 to 15.37 psia).
Storage: 50 to 106 kPa (7.25 to 15.37 psia).
Conditions
Indoor Use Operating Altitude up to 4000 m (13,000 ft) Pollution degree I or II
Battery life
15 to 20 hours.
Battery type
Four commercially available AA batteries.
Measurement
High Flow Module
Low Flow Module
Flow and Peak Flow
Range
0 to 300.0 SLPM.
0.01 to 15.00 SLPM.
Accuracy
Air and oxygen: 2% of reading or 0.075 SLPM, whichever is greater.
Air/oxygen mixtures: 4% of reading or 0.1 SLPM, whichever is greater.
Air and oxygen: 2% of reading or 0.010 SLPM, whichever is greater.
Nitrous oxide: 4% of reading or 0.025 SLPM, whichever is greater.

6 Specifications

NOTE: Specifications are subject to change without notice.

6.1 Physical

6.2 Environmental

6.3 Power

6.4 Test Measurements (see notes at end of section)

31
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 35
32 6: Specifications
Measurement
High Flow Module
Low Flow Module
Pressure drop
Maximum between inlet and outlet ports at 101.3 kPa (14.7 psia) including filter: 0.50 cmH2O at 20 SLPM, 1.50 cmH2O at 50 SLPM, 5.0 cmH2O at 100 SLPM,
30.0 cmH2O at 300 SLPM.
Maximum between inlet and outlet ports at 101.3 kPa (14.7 psia) including filter: 4 cmH2O at 2 SLPM, 16 cmH2O at 5 SLPM, 45 cmH2O at 10 SLPM, 85 cmH2O at 15 SLPM.
Volume
Range
0.01 to 10.0 L STP.
0 to 9.999 L STP.
Accuracy
Air and oxygen: 2% of reading plus 0.020 L STP
Air/oxygen mixtures: 4% of reading plus 0.020 L STP.
Air and oxygen: 2% of reading or 0.010 L STP, whichever is greater.
Nitrous oxide: 4% of reading or 0.010 L STP, whichever is greater.
Minute Volume
Range
0 to 99.00 L STP.
0 to 9.999 L STP.
Accuracy
7% of reading.
7% of reading.
Stacked Volume
Range
0 to 99.00 L STP.
0 to 9.999 L STP.
Accuracy
Air and oxygen: 2% of reading plus 0.020 L STP
Air/oxygen mixtures: 4% of reading plus 0.020 L STP.
Air and oxygen: 2% of reading or 0.010 L STP, whichever is greater.
Nitrous oxide: 4% of reading or 0.010 L STP, whichever is greater.
Inspiratory Time
Range
0.25 to 60.00 seconds.
0.25 to 60.00 seconds.
Accuracy
0.01 seconds.
0.01 seconds.
I:E Ratio
Range
1:100.0 to 100.0:1.
1:15.0 to 15.0:1.
Accuracy
5% of reading.
5% of reading.
Respiratory Rate
Range
0.5 to 120.0 breaths per minute.
0.5 to 120.0 breaths per minute.
Accuracy
5% of reading.
5% of reading.
Low Pressure
Range
-25.0 to 150.0 cmH2O.
(-18.4 to 110 mmHg)
Not applicable.
Accuracy
0.75% of reading or 0.20 cmH2O (0.15 mmHg),
whichever is greater.
Not applicable.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 36
Measurement
High Flow Module
Low Flow Module
Peak & PEEP Low Pressure
Range
0 to 150.0 cmH2O.
(0 to 110 mmHg)
Not applicable.
Accuracy
0.75% of reading or 0.20 cmH2O (0.15 mmHg), whichever is greater.
Not applicable.
Barometric Pressure
Range
375 to 1500 mmHg.
375 to 1500 mmHg.
Accuracy
8 mmHg.
8 mmHg.
Oxygen Concentration
Range
0 to 100.0% O2.
Not applicable.
Accuracy
2% O2 at daily calibration conditions.
Not applicable.
NOTES
1. Standard conditions are defined as 21.1 C (70 F) and 101.3 kPa (14.7 psia).
2. Flow and volume accuracy is applicable in SLPM or STP mode only.
3. The temperature of the gas and the ambient air must be within 10 C (18 F) of each other and the gas must be less than 30% relative humidity at 21. 1 C (70 F).
4. Flow and volume accuracy de-rating: 0.075% of reading per 1 C (1.8 F) away from 21.1 C (70 F); 0.015% of reading per 1.03 kPa (0.15 psia) above 101.3 kPa (14.7 psia); 0.022% of reading per 1.03 kPa (0.15 psia) below 101.3 kPa (14.7 psia); 0.07% of reading per 1% relative humidity above 30% relative humidity.

6.5 Calibration Requirements

Flow modules
Factory calibration every year for normal use under normal conditions using the filter(s) provided.
Controller module
No calibration required.
Oxygen sensor
Daily, following sensor replacement or as required.
Complies with these standards:
CAN/CSA-C22.2 No. 1010. 1-92: Canadian Standard for the
Safety of Electrical Equipment for Measurement, Control and Laboratory Use, Part 1.
EN 55011 (1991) Class B, CISPR 11 (1990) Class B, FCC
(CFR 47, Part 15) Class B: Emissions, Radiated and Conducted.
EMC Directive 89/336/ECC, EN 61326-1 (1997 plus
Amendment A1 1998), IEC 1000-4-2 (1995), EN 61000-4-2, IEC 1000-4-3 (1995), EN 61000-4-3: Immunity.

6.6 Compliance and Approvals

6: Specifications 33
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com