TSI Alnor 4000 Specifications

UNDERSTANDING, ACCELERA
TED
Learn more at
www.tsi.com /general-purpose-owmeters/
MEASURE FLOW, PRESSURE,
AND TEMPERATURE…
Designed for Performance
TSI thermal mass owmeters incorporate a proprietary platinum lm
sensor design for measuring gas ows in applications demanding fast
response and high accuracy over a wide ow range. TSI owmeters have
turn-down ratios greater than 1000:1 due to our thermal ow sensing
technology and extensive gas calibration process. The TSI 4000 Series
was designed for ultra-low pressure loss to minimize any undesirable
eects the owmeter can have on the readings when installed in-circuit.
Industries
+ Medical
– Ventilators
– Anesthesia
– CPAP
+ Environmental
+ Analytical
+ Aerosol Science
Applications
+ Product Development
+ Manufacturing
+ Research
+ Field Service
+ Quality Assurance
Features
+ 4 millisecond ow response
+ High accuracy ±2% of reading
+ High turndown ratio
+ Low pressure drop
+ Convenient analog output of ow rate
+ Versatile digital output of ow rate, volume, pressure, temperature
+ Built-in temperature and pressure compensation
+ NIST-traceable calibration certicate included at no additional cost
RS232 Interface For Digital Outputs and Congurable Device Options
+ Set analog output zero and scaling
+ Specify start/stop trigger levels for volume measurement
+ Set update rate for LCD display
+ Set sampling rate for analog and digital outputs
+ Select gas calibration
+ Select either standard or volumetric ow measurement
+ Set display units for Model 4140/4143 to L/min or cm3/min
+ Compute volume
Accurate
A owmeter specied as ±2% of full scale is most accurate at full scale. If full scale is 300 L/min, then the uncertainty for all readings is ±6 L/min. TSI owmeters are specied as ±2% of reading and have an uncertainty of ±2% of the actual reading from full scale all the way down to a specied lower limit. TSI owmeters, therefore, provide dependable accuracy over a wide range of ow rates. One TSI owmeter covers the same range as three or more “percent of full scale” devices... with better accuracy at all points!
Fast
Fast 4 millisecond response ensures accuracy in uctuating ows. This fast response is ideal for closed-loop control systems and integrated volume measurements. Pressure and temperature measurements are also extremely fast.
Low Pressure Drop
Low pressure drop minimizes ow circuit back pressure and its impact on the system under test.