if pocket has clip retaining groove, push
clip over pocket
spring
brass brush
clip engages
in groove
probe is forced
against end of
pocket
If using compatible pocket
If using Universal Fixing Kit
Adjust probe length
c2
pocket
length
if pocket has grub screw, discard spring and clip
tighten screw
over bush
tighten
use 2 mm
hexagonal socket key
*If used for chilled water ensure pocket is sealed around probe or fill pocket with thermally
conducting oil to avoid the build up of condensation in bottom of pocket.
It is recommended to use SET (Software Tool) for the setting of the sensor type module.
For all IQ2 series controllers with firmware version 2.1 or greater, or IQ3 series controllers,
the following SET Unique Sensor References should be used:
PRT I -10+40(T/PI/40, °C)
PRT I +14+104 F (T/PI/40, °F)
PRT I -10+110 (T/PI/110, °C)
PRT I +14+230 F (T/PI/110, °F)
PRT I -40+50(T/PI/-40, °C)
PRT I -40+122 F (T/PI/-40, °F)
Alternatively set scaling mode to 5 (characterise) and enter scaling manually as defined in
the appropriate table below. Note that for IQ3, the scaling mode and exponent do not need
to be set up.
For all other IQ controllers see Sensor Scaling Reference Card TB100521A
At the end of their useful life the packaging
and product should be disposed of by a
Do not dispose of with normal household waste.
Do not burn.
suitable recycling centre.
Manufactured for and on behalf of the Environmental and Combustion Controls Division of Honeywell Technologies Sàrl, Ecublens, Route
du Bois 37,Switzerland by its Authorized Representative, Trend Control Systems Limited.
Trend Control Systems Limited reserves the right to revise this publication from time to time and make changes to the content hereof
without obligation to notify any person of such revisions or changes.