Piezo Driver/Power Amplifier
The Trek Model PZD700A is a high-voltage DC-stable piezo driver/amplifier designed to provide precise control of output voltages in bipolar or unipolar ranges which are customer specified within a range of available settings. The instrument achieves the accurate output responses and high slew rates demanded by reactive loads by utilizing a four-quadrant active output stage that sinks or sources current into reactive or resistive loads.
Key Specifications
! Output Voltage Range Bipolar:
Unipolar: 0 to +1.4 kV DC or peak AC or 0 to -1.4 kV DC or peak AC
! Output Current Range Bipolar:
Unipolar: 0 to ±50 mA
! Slew Rate Bipolar: Greater than 380 V/
Unipolar: Greater than 370 V µs
Large Signal Bandwidth Bipolar: DC to greater than 125 kHz (-3 dB)
!
Unipolar: DC to greater than 120 kHz (-3 dB)
DC Voltage Gain: 0 to 300 V/V, adjustable using a front panel potentiometer
!
0 to ±700 V DC or peak AC
0 to ±100 mA
µs
/
Typical Applications Include
! Piezoelectric driving/control !
Laser modulation
!
MEMS
!
Semiconductor research
!
Piezoelectric vibration damping
Features and Benefits
Four-quadrant output for driving capacitive loads
!
Closed loop system for high accuracy
!
Short-circuit protected for equipment protection
!
All solid-state design for maintenance-free operation
!
DC-stable for programmable supply applications
!
Low output noise for ultra-accurate outputs
!
Model PZD700A M/S is also available with twice the current capability of the Model PZD700A
!
NIST-traceable Certificate of Calibration provided with each unit
!
compliant
!
TREK, INC. • 190 Walnut Street • Lockport, NY 14094 • USA • 800-FOR TREK
716-438-7555 • 716-201-1804 (fax) • www.trekinc.com • sales@trekinc.com
Features (cont.)
Input Configuration
The input is configured as a noninverting amplifier. An inverting amplifier is also available
Limit Indicator
An amber indicator warns when the PZD700A fails to produce the required HV output.
Automatic Power Limit
Automatically limits the internal power dissipation to protect the PZD700A from overheating.
Mechanical
Dimensions (single channel instrument)
110 mm H x 220 mm x W 445 mm D (4.3” H x 8.7” W x 17.5” D)
Weight
5 kg (11 lb) (Single channel unit)
HV Connector
SHV High Voltage Connector
Operating Conditions
Temperature
0°C to 40°C (32°F to 104°F)
Relative Humidity
To 85%, noncondensing
Altitude
To 2000 meters (6561.68 ft.)
Electrical
Line Voltage
Factory Set for one of two ranges: 90 to 127 V AC or 180 to 250 V AC, either at 48 to 63 Hz
AC Line Receptacle
Standard 3-prong with integral fuse holder
Power Consumption
90 VA, single channel 175 VA, dual channel
HV Cable
2 m, 66 pF per foot
Supplied Accessories
Operators’ Manual
PN: 23439
HV Output Cable Assembly
PN: 43874R cable and SHV mating connector Line Cord, Fuses
Selected per geographic destination
Ordering Information
90 to 127 V AC 90 to 127 V AC 180 to 250 V AC 180 to 250 V AC
Model PZD700A-1-L (single unit) Model PZD700A-2-L (dual unit) Model PZD700A-1-H (single unit) Model PZD700A-2-H (dual unit)
Note
The Model PZD700A comes from the factory with settings for an output voltage of ±700 V DC or peak AC, a voltage gain ratio of 200 V/V, with a noninverting input. Please specify voltage range (±700 V, +1400V or
-1400V) and input configuration (inverting or noninverting) when ordering.
Also available is the Model PZD700A M/S with twice the current capability of the standard PZD700A.
Model PZD700A Specifications
Performance
Output Voltage Bipolar Range
0 to ±700 V DC or peak AC
Output Voltage Unipolar Range
0 to +1.4 kV DC or 0 to -1.4 kV DC or peak AC
Output Current Bipolar Range
0 to ±100 mA
Output Current Unipolar Range
0 to ±50 mA Input Voltage Range
0 to ±10 V DC or peak AC
Input Impedance
90 kW, nominal (non-inverting) 1 MW nominal, (inverting)
DC Voltage Gain
0 to 300 V/V, adjustable using the front panel potentiometer
DC Voltage Gain Accuracy
Better than 0.1% for factory set gain of 200 V/V Offset Voltage
Less than ±500 mV
Output Noise (all ranges)*
Less than 50 mV rms to 20 kHz for a 1 nF load. Less than 100 mV rms to 20 kHz with no load.
Slew Rate (10% to 90%, typical)
Bipolar: Greater than 380 V/µs Unipolar: Greater than 370 V/µs
Large Signal Bandwidth (-3 dB)
Bipolar: DC to greater than 125 kHz Unipolar: DC to greater than 120 kHz
Small Signal Bandwidth (-3dB)
DC to greater than 200 kHz Settling Time
Less than 50 µs when critically damped
Stability
With a factory set gain of 200 V/V
Drift with Time
Less than 50 ppm/hr, noncumulative
Drift with Temp
Less than 100 ppm/°C
Voltage Monitor
Ratio
1/200th of the high voltage output
DC Accuracy
Better than ±0.1% of full scale
Current Monitor
Ratio
0.1 V/mA, ±1% of full scale
DC Accuracy
Better than ±1% of full scale
Features
Digital Enable
BNC connection for TTL compatible signal to turn ON/OFF the HV output for each channel.
Gain Control
The gain of the Model PZD700A is adjustable from 0 to 300 V/V
Dynamics Adjustment
A graduated 1-turn front panel potentiometer is used to optimize the AC response of the
for various load configurations.
output
signal
*Measured using the true rms feature of the HP Model 34401A digital multimeter)
Measurement and Power Solutions
Copyright ã 2012 TREK, INC. All specifications are subject to change. 1237/DEC
TM
www.tre k inc.com