Trek Model PZD350A
Piezo Driver/Power Amplifier
The Trek Model PZD350A is a high-voltage DC-stable piezo
driver/amplifier designed to provide precise control of output
voltages in bipolar or unipolar ranges that are customer specified
within a range of available settings. It is configured as a
noninverting amplifier with a variable DC gain. An inverting
amplifier configuration is also available.
The unit features an all-solid-state design, a high slew rate and a
four-quadrant active output stage which sinks or sources current
into reactive or resistive loads throughout the output voltage
range. This capability is essential for achieving the accurate
output responses and high slew rates demanded by reactive
loads.
Key Specifications
! Output Voltage Range Bipolar:
Unipolar: 0 to +700 V DC or peak AC or 0 to -700 VDC or peak AC
! Output Current Range Bipolar:
Unipolar: 0 to ±100 mA
! Slew Rate Bipolar: Greater than 550 V/
! /
Large Signal Bandwidth Bipolar: DC to greater than 250 kHz (-3 dB); DC to greater than 90 kHz (1% distortion)
!
! Unipolar: DC to greater than 200 kHz (-3 dB); DC to greater than 70 kHz (1% distortion)
! Small Signal Bandwidth Bipolar: DC to greater than 350 kHz (-3 dB)
! Unipolar: DC to greater than 250 kHz (-3 dB)
DC Voltage Gain: 0 to 150 V/V, adjustable using a front panel potentiometer
!
Unipolar: Greater than 440 V µs
0 to ±350 V DC or peak AC
0 to ±200 mA
µs
Typical Applications Include
! Piezoelectric driving/control
! Laser modulation
! MEMS
! Semiconductor research
! Piezoelectric vibration damping
Features and Benefits
!
Four-quadrant output for driving capacitive loads
!
Closed loop system for high accuracy
!
Short-circuit protected for equipment protection
!
All solid-state design for maintenance-free operation
!
DC-stable for programmable supply applications
!
Low output noise for ultra-accurate outputs
Model PZD350A M/S is also available with twice the current capability of the Model PZD350A
!
NIST-traceable Certificate of Calibration provided with each unit
!
compliant
!
TREK, INC. • 190 Walnut Street • Lockport, NY 14094 • USA • 800-FOR TREK
(fax) • www.trekinc.com • sales@trekinc.com 716-438-7555 • 716-201-1804
A graduated 1-turn front panel potentiometer is
used to optimize the AC response of the
for various load configurations
output
signal
The input is configured as a noninverting
amplifier. An inverting amplifier is also available
An amber indicator warns when the PZD350A
fails to produce the required HV output
Automatically limits the internal power
dissipation to protect the PZD350A from
overheating
Dimension - Single
Channel
Dual Channel
110 mm H x 220 mm W x 445 mm D
(4.3” H x 8.7” W x 17.5” D)
110 mm H x 432 mm W 445 mm D
(4.3” H x 17” W x 17.5” D)
5 kg (11 lb)
10 kg (22 lb)
SHV High Voltage Connector
0°C to 40°C (32°F to 104°F)
To 2000 meters (6561.68 ft.)
Factory Set for one of two ranges:
90 to 127 V AC or 180 to 250 V AC,
either at 48 to 63 Hz
Standard 3-prong with integral fuse holder
90 VA, single channel
175 VA, dual channel
PN: 43874R cable and SHV mating connector
Line Cord, Fuses
Selected per geographic destination
90 to 127 V AC
90 to 127 V AC
180 to 250 V AC
180 to 250 V AC
Model PZD350A-1-L (single unit)
Model PZD350A-2-L (dual unit)
Model PZD350A-1-H (single unit)
Model PZD350A-2-H (dual unit)
The Model PZD350A comes from the factory with settings for an output
voltage of ±350 V DC or peak AC, a voltage gain ratio of 100 V/V, with a
noninverting input. Please specify voltage range (±350 V, +700 V, or
-700 V) and input configuration (inverting or noninverting) when
ordering.
Also available is the Model PZD350A M/S with twice the current
capability of the standard PZD350A.
Model PZD350A Specifications
Output Voltage
Bipolar Range
0 to ±350 V DC or peak AC
Output Voltage
Unipolar Range
0 to +700 V DC or 0 to -700 V DC or peak AC
Output Current
Bipolar Range
Output Current
Unipolar Range
0 to ±100 mA
Input Voltage Range
90 kW, nominal (non-inverting)
1 MW nominal, (inverting)
Better than 0.1% for factory set gain of 100 V/V
(input to output)
Output Noise (all
ranges)*
Less than 100 mV rms to 20 kHz w/100 pF load
Less than 150 mV rms to 20 kHz with no load
Slew Rate
(10% to 90%, typical)
Bipolar: Greater than 550 V/µs
Unipolar: Greater than 440 v/µs
Large Signal
Bandwidth (-3 dB)
Bipolar: DC to greater than 250 kHz
Unipolar: DC to greater than 200 kHz
Large Signal
Bandwidth
(1% distortion)
Bipolar: DC to greater than 90 kHz
Unipolar: DC to greater than 70 kHz
Small Signal
Bandwidth (-3dB)
Bipolar: DC to greater than 350 kHz
Unipolar: DC to greater than 250 kHz
Less than 30 µs when critically damped
With a set gain of 100 V/V
Less than 50 ppm/hr, noncumulative
1/100th of the high voltage output
0.05 V/mA, ±1% of full scale
BNC connection for TTL compatible signal to
turn ON/OFF the HV output for each channel
The gain of the Model PZD350A is adjustable to
150 V/V
*Measured using the true rms feature of the HP Model 34401A digital multimeter
Measurement and Power Solutions
www.tr ekinc.com
Copyright ã 2012 TREK, INC. All specifications are subject to change. 1237/DEC
TM