Trek PZD2000A User Manual

Trek Model PZD2000A
High-Voltage, High-Bandwidth Power Amplifier
The Model PZD2000A
power amplifier used for precisio n high power applications. The amplifier incorporates an all-solid-state design for high reliability and low-noise operation. Its four­quadrant output stage sinks as well as sources load current throughout the output voltage range, thus achieving accurate output response and high slew rates, even into highly capacitive loads.
Key Specifications
! Output Voltage Range:
! Output Current Range:
! Slew Rate: Greater than 750 V/
Large Signal Bandwidth (3% distortion): DC to greater than 60 kHz
!
DC Voltage Gain: 200 V/V
!
0 to ±2 kV DC or peak AC 0 to ±200 mA DC or ±400 mA peak AC. Maximum duration for ±400 mA
current pulse is 2 ms at 50% duty cycle using a square wave
is a wide bandwidth, high-voltage
µs
Typical Applications Include
! Dielectric material characterization
Polymer and ceramic corona poling
!
Piezoelectric driving and control
!
Features and Benefits
DC accuracy is better than 0.1% of full scale
!
!
Precision voltage and current monitors provide buffered low-voltage representations of the high-voltage output
and load current for monitoring purposes, or for use as feedback signals in closed-loop systems
Remote high-voltage ON-OFF suitable for use with automated or computer controlled systems
!
Output stage fully protected against over voltage and over current conditions that may be generate by active loads,
!
overloads or arcing to ground
Adjustable current limit or current trip level
!
NIST-traceable Certificate of Certification provided with each unit shipped
!
TREK, INC. • 190 Walnut Street • Lockport, NY 14094 • USA • 800-FOR TREK
(fax) • www.trekinc.com • sales@trekinc.com 716-438-7555 • 716-201-1804
Current Monitor (cont.)
Output Impedance
47 W
Features
High Voltage On-Off
Switch selectable for local or remote. Local: Individual push-button switches; Remote: TTL compatible input. TTL High turns off high voltage. TTL low turns on high voltage.
Dynamic Adjustment
Graduated 1-turn panel potentiometer is used to optimize the AC response for various load parameters.
Current Limit/Trip
Switch selectable for either limit or trip. Grad­uated 1-turn potentiometer is used to adjust current limit or trip level from 10 to 200 mA
Out of Regulation Status
Indicator illuminates and BNC provides a TTL low when required high voltage is not provided such as during a current limit
Trip Status
Indicator illuminates and BNC provides a TTL low when high voltage output trips due to cur­rent trip, detection of fault or removal of cover
Fault Status
BNC provides TTL low when out of regulation for greater than 500 ms
Mechanical
Dimensions
266 mm H x 482 mm W 655 mm D (10.5” H x 19” W x 25.8” D)
Weight
24.9 kg (55 lb)
HV Connector
Alden high voltage connector
BNC Connector
Amplifier input, voltage monitor, current monitor, digital enable, fault/trip status, out of regulation status
Operating Conditions
Temperature
0°C to 40°C (32°F to 104°F)
Relative Humidity
To 75%, noncondensing
Altitude
To 2000 meters (6561.68 ft.)
Electrical
AC Line Receptacle
Standard three-prong AC line connector
Line Voltage
Factory set for one of two ranges: 104 to 126 V AC or 180 to 250 V AC, at 48 to 63 Hz
Power Consumption
1000 VA, maximum
Supplied Accessories
Operators’ Manual
PN: 23271
HV Output Cable
PN: 43406
Line Cord
PN: N5011 (104 to 126 V AC) Contact Factory: (180 to 250 V AC)
Optional Accessories
HV Output Cable
PN: 43406
Model PZD2000A Specifications
Performance
Output Voltage Range
0 to ±2 kV DC or peak AC
Output Current Range
0 to ±200 mA DC or ±400 mA peak AC. Maximum duration for ±400 mA current pulse is 2 ms at 50% duty cycle using a square wave.*
Maximum Power
500 W (real, apparent or reactive). Unit will trip off if internal power dissipation exceed 500 W
Input Voltage Range
0 to ±10 V DC or peak AC, noninverting
Input Impedance
25 kW, nominal
DC Voltage Gain
200 V/V
DC Voltage Gain Accuracy
Better than 0.1% of full scale DC Offset Voltage
Less than ±2 V
Output Noise
Less than 500 mV rms**
Slew Rate (10% to 90%, typical)
Greater than 750 V/µs Small Signal
Bandwidth (-3dB)
DC to greater than 100 kHz
Large Signal Bandwidth (3% distortion)
DC to greater than 60 kHz
Settling Time to 1%
Less than 50 µs for a 2 kV step
Stability
Drift with Time
Less than 50 ppm/hr, noncumulative
Drift with Temp
Less than 100 ppm/°C
Auto Power Limit
Limits internal power dissipation to protect from overheating
Voltage Monitor
Ratio
01/200th of the high voltage output
DC Accuracy
Better than 0.1% of full scale
DC Offset Voltage
Less than ±2 mV
Output Noise
Less than 5 mV rms**
Output Impedance
47 W
Current Monitor
Ratio
0.025 V/mA
DC Accuracy
Greater than 1% of full scale
Offset Voltage
Less than ±10 mV
Output Noise
Less than 10 mV rms**
Bandwidth (-3 dB)
DC to greater than 5 kHz
*See Automatic Power Limit feature for limitations **Measured using the true rms feature of the Hewlett Packard Model 34401A digital multimeter